K Number
DEN150059

Validate with FDA (Live)

Date Cleared
2016-10-06

(287 days)

Product Code
Regulation Number
866.2190
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APAS Compact is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use:

The APAS Compact, when using its urine analysis module, automates urine culture plate imaging and interpretation to detect the presence of microbial growth on sheep blood and MacConkey agar culture plates that are inoculated with a 1uL sample volume. The APAS Compact, when using its urine analysis module, provides a semiquantitative assessment of colony counts that are used as an aid in the diagnosis of urinary tract infection. All urine culture plates that are identified as positive for growth by the APAS Compact, when using its urine analysis module, must be reviewed by a trained microbiologist.

Device Description

The APAS Compact with Urine Analysis Module is an instrumented system that is designed for screening of urine culture plates for the presence of microbial growth. The device comprises an imaging station for capture of digital images of the culture plates, together with software for analysis of the images, the determination and enumeration of microbial growth and reporting of results.

AI/ML Overview

This document outlines the acceptance criteria and study findings for the APAS Compact with Urine Analysis Module.

1. Acceptance Criteria and Reported Device Performance

CriteriaAcceptance CriteriaReported Device Performance (Overall)
Detection of Growth on Blood Agar (Clinical Study)False negative results for growth detection should be acceptably low.99.0% (95% CI: 98.7-99.2%) correct designation of growth. False negative rate for growth detection ranged from 0% to 2.9% depending on colony count.
Detection of Growth on MacConkey Agar (Clinical Study)False negative results for growth detection should be acceptably low.99.5% (95% CI: 99.2-99.7%) correct designation of growth. False negative rate for growth detection ranged from 0% to 1.3% depending on colony count.
Colony Count Accuracy (Analytical Study)Low counts obtained by APAS Compact should be within 1-log10 of manual reference. APAS Compact should not incorrectly designate any cultures with growth as "Negative".All low counts were within 1-log10 of manual reference. APAS Compact did not incorrectly designated any of the cultures with growth as "Negative". Instances where APAS Compact designated growth while reference method reported no growth were acceptable as these require microbiologist review.
Reproducibility of Colony CountsAcceptable reproducibility of colony counts between instruments and rotations.Demonstrated acceptable reproducibility (detailed in Tables 5 and 6, with %CVs generally low, though some higher %CVs for very low counts or specific dilutions).
Analytical Specificity (Expected Colony Morphology - Pure Cultures)High agreement for detection of expected colony morphology.Blood Agar: Generally 100% agreement, with exceptions for Aerococcus urinae (98.1%), Lactobacillus rhamnosus (22.2%), Pseudomonas aeruginosa (98.0%), and Staphylococcus epidermidis (70.4%). All classifications were "Positive" or "Review" for all but A. urinae and L. rhamnosus. MacConkey Agar: Generally 100% agreement, with Serratia marcescens at 77.8% (but 100% at 20 & 22 hrs). All classifications were "Positive".
Analytical Specificity (Expected Colony Morphology - Mixed Cultures)High agreement for detection of both expected colony types.Blood Agar (1:1 ratio): 100% for all combinations. Blood Agar (1:10 ratio): 100% for E. coli combinations, but S. agalactiae in presence of E. faecalis was 44.4%. All classifications were "Positive". MacConkey Agar: Generally 100% agreement, with one image for E. coli/M. morganii (1:10 ratio) at 98.8%. All classifications were "Positive".
Daily Quality Control PerformanceExpected results for positive controls; acceptable rate for negative controls.100% of positive controls yielded expected results on both blood and MacConkey agar. 98.4% of negative controls (blood agar) and 100% (MacConkey agar) yielded expected results. Failed negative controls were due to single contaminating colonies or artifacts.
Detection Limit (LOD)Demonstrated limits of colony size for reliable detection.Provided specific LODs in mm (95% CI) for various organisms on blood and MacConkey agar (Tables 10 and 11).

2. Sample Sizes and Data Provenance

  • Test Set (Clinical Study):

    • Total Enrolled Samples: 10,100 urine samples (5835 from Site 1, 2117 from Site 2, 2148 from Site 3).
    • Included in Performance Analysis: 9,224 samples (5634 from Site 1, 1769 from Site 2, 1821 from Site 3).
    • Data Provenance: Three clinical sites: one in the US (Site 1) and two ex-US (Sites 2 and 3). The study used remnant urine samples from standard of care culture. This indicates a retrospective data collection approach for the clinical study.
  • Training Set: The document does not explicitly state the sample size for the training set.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

  • Number of Experts: Three microbiologists.
  • Qualifications: "Trained microbiologist" is mentioned multiple times. For the clinical study, each microbiologist was "trained to read the urine culture plates in a standard fashion." Specific years of experience are not provided.

4. Adjudication Method (Test Set)

  • The document states that for the reference method in the clinical study, "each microbiologist... was blinded to the results from the other panel members and to those obtained by the APAS Compact."
  • The ground truth for the clinical study was established by a "reference microbiologist panel," meaning the consensus among these three microbiologists. The exact method of achieving consensus (e.g., 2+1, 3+1, none) is not explicitly detailed, but it implies a shared understanding to form the "reference method."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study was not done. The clinical study compared the APAS Compact's performance against a panel of microbiologists (serving as the reference standard), not against human readers with and without AI assistance to measure improvement. The primary goal of the APAS Compact is to automate initial screening to reduce microbiologist workload, rather than directly augment human reading for every plate.

6. Standalone Performance Study

  • Yes, a standalone performance study was done. The entire clinical study (Section L.3) and analytical performance studies (Section L.1) evaluate the APAS Compact's performance as an algorithm-only device (without human-in-the-loop during the initial assessment and designation phase). The performance tables (e.g., Tables 19-24) directly report the APAS Compact's designations and colony counts compared to the reference standard. While the device's indications for use emphasize that growth-positive plates "must be reviewed by a trained microbiologist," the reported performance metrics are for the algorithm's initial assessment before this mandatory human review.

7. Type of Ground Truth Used

  • Clinical Study: Expert Consensus (Microbiologist Panel): The ground truth for the clinical study was established by "an independent panel of three microbiologists," who read the urine culture plates in a "standard fashion" and were blinded to other results. Their combined assessment served as the reference method.
  • Analytical Study (Linearity/Assay Reportable Range): Expert Consensus (Manual Colony Counts): The mean of two independent manual colony counts performed by two microbiologists was used as the reference result for each plate.

8. Sample Size for the Training Set

  • The document does not explicitly state the sample size for the training set. It describes the "APAS Controller Software" and "Urine Analysis Module Software" functions, implying that these algorithms were developed and likely trained, but the specific training data volume is not provided in this excerpt.

9. How Ground Truth for the Training Set Was Established

  • The document does not explicitly describe how the ground truth for the training set was established. While it details the methods for establishing ground truth in the analytical and clinical validation studies (using expert microbiologist consensus), it does not provide information specific to the data used for training the algorithms.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR THE AGAR PLATE ASSESSMENT SYSTEM (APAS) COMPACT

DECISION MEMORANDUM

A. 510(k) Number:

DEN150059

B. Purpose for Submission:

De novo request for evaluation of automatic class III designation for the Agar Plate Assessment System (APAS) Compact

C. Measurand:

Digital images of microbial colonies cultured on blood and MacConkey agar plates

D. Type of Test:

The APAS Compact when using its Urine Analysis Module (the "APAS Compact" or the "APAS Compact with Urine Analysis Module") is an in vitro diagnostic test system for automated assessment and enumeration of microbial colonies on solid culture media. The system is for use on urine cultures from suspected cases of urinary tract infection (UT).

E. Applicant:

Clever Culture Systems AG

F. Proprietary and Established Names:

APAS Compact

G. Regulatory Information:

    1. Regulation section:
      21 CFR 866.2190 Automated image assessment system for microbial colonies on solid culture media
    1. Classification:
      Class II
    1. Product code:
      PPU

{1}------------------------------------------------

4. Panel:

  • 83 Microbiology

H. Indications for Use:

1. Indication(s) for Use:

The APAS Compact is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use:

The APAS Compact, when using its urine analysis module, automates urine culture plate imaging and interpretation to detect the presence of microbial growth on sheep blood and MacConkey agar culture plates that are inoculated with a 1uL sample volume. The APAS Compact, when using its urine analysis module, provides a semiquantitative assessment of colony counts that are used as an aid in the diagnosis of urinary tract infection. All urine culture plates that are identified as positive for growth by the APAS Compact, when using its urine analysis module, must be reviewed by a trained microbiologist.

2. Special conditions for use statement(s):

For prescription use only.

The performance of APAS Compact with Urine Analysis Module has not been evaluated with urine samples from pregnant women. The APAS Compact with Urine Analysis Module will detect colonies of GBS if they are present but should not be used for the primary screening for Group B Streptococcus (GBS) carriers. Follow recommended guidelines for identification of pregnant women who are colonized with GBS.

The performance of the APAS Compact with Urine Analysis Module has not been evaluated with urine samples from suspected cases of complicated urinary tract infection e.g., those with underlying urinary tract pathology, suspected cases of persistent urinary tract infection, urine collected by invasive procedures or urine samples from immunocompromised subjects.

Slow growing organisms such as Corynebacterium spp. and Gardnerella vaginalis may not exhibit detectable growth within 18-22 hours. If infection with a slow growing species is suspected, extension of the incubation time may be required, followed by manual interpretation of the culture result. APAS Compact with Urine Analysis Module should not be used to interpret cultures incubated for greater than 22 hours.

The APAS Compact with Urine Analysis Module is qualified for use with the following culture plates manufactured by Remel: Tryptic Soy Sheep Blood Agar and MacConkey Agar with Crystal Violet.

1 CDC Guidelines, 2010: Prevention of perinatal Group B Streptococcal disease.

{2}------------------------------------------------

3. Special instrument requirements:

APAS Compact

I. Device Description:

The APAS Compact with Urine Analysis Module is an instrumented system that is designed for screening of urine culture plates for the presence of microbial growth. The device comprises an imaging station for capture of digital images of the culture plates, together with software for analysis of the images, the determination and enumeration of microbial growth and reporting of results. A list of the major sub-components of the APAS Compact is depicted in Table 1. The Urine Analysis Module software is compatible with the following types of culture media:

Tryptic Soy Sheep Blood Agar MacConkey Agar with Crystal Violet

ComponentFunction
Imaging StationLight Emitting Diode (LED) illumination of cultureplates and image capture using a Charge CoupledDevice (CCD) camera
APAS Controller Personal Computer (PC)Urine Analysis Module SoftwareImage capture, storage and analysisInstalled on the APAS Controller PC to provide theconfiguration and instructions for image capture andanalysis
Instrument Controller SoftwareLIMS Interface SoftwareUser interface for operation of the APAS CompactInstalled on the Instrument Controller PC and used toimport other diagnostic information, such asmicroscopy or chemistry results, and provide contextfor interpretation of urine culture results.Imported information may be applied to the system:"LIMS Force Flag": automatically forces an APASresult to "Review" irrespective of the growthcharacteristics observed"LIMS Complementary Test Flag": automaticallychanges a "Negative” designation to “Review" basedon user defined rules applied to additional diagnosticinformation
Color Calibration ToolMulticolored disk for calibration of the system optics

Table 1. Sub-components of the APAS Compact

The imaging station comprises a fixed CCD camera, top and bottom LED illumination for the culture plates and a plate handling mechanism, all of which are housed in a light-sealed chassis. The Instrument Controller PC provides the user interface and sends instructions to the APAS Controller PC which in turn controls the imaging station.

The plate handling mechanism of the imaging station comprises a manual lever that allows

{3}------------------------------------------------

the operator to load a single culture plate into the system and position it for imaging. The operator then initiates image capture using the Instrument Controller PC. Once an image is taken by the camera it is sent to the APAS Controller PC for assessment and storage. APAS Compact reports are viewed on a separate computer through a dedicated user interface.

J. Standard/Guidance Document Referenced (if applicable):

IEC 61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements, 3rd Edition.

IEC 61326-2-6: Electrical equipment for measurement, control and laboratory use - EMC Requirements - Part 2-6: Particular Requirements - In vitro diagnostic (IVD) medical Equipment, 2nd Edition.

IEC 62304 Medical device software - software lifecycle processes, 1st Edition.

ISO 14971 Medical devices - Application of Risk Management to Medical Devices, 2007.

CISPR 11 Industrial, scientific and medical equipment - radio frequency disturbance characteristics - Limits and methods of measurement, Version 5.1.

K. Test Principle:

The APAS Compact with Urine Analysis Module is designed for the assessment of microbial colonies on urine culture plates. The device comprises an imaging station for capture of images of the culture plates, together with software for analysis of the images, the assessment and enumeration of microbial colonies and result designation. Plates with growth are designated as either "Positive" or "Review" depending on the number of colonies present and their morphologic characteristics. All samples that exhibit growth therefore require follow up according to standard laboratory practice by a trained microbiologist. Plates that are designated by the APAS Compact to have no growth may be discarded without further review.

The APAS Compact with Urine Analysis Module is indicated for screening paired sheep blood and MacConkey agar plates that are each inoculated with 1ul of urine and which are incubated at 35°C ± 2°C for 18 to 22 hours. For each urine sample, both culture plates must be processed on the APAS Compact. The system takes digital images of each plate which are analyzed automatically to determine the number of colonies present and their associated morphologies. For each plate, the APAS Compact performs the following tasks:

  • a) Differentiates between areas of growth (colonies) and no growth (e.g., agar, labels, handwriting):
  • b) Differentiates areas that are required for interpretation of colony morphology (e.g., α- and β-hemolysis);
  • c) Assigns to each area of growth a code that corresponds to a morphology type;
  • d) Estimates the number of colonies present, converts this number to an organism concentration in the original urine sample and reports it in terms of colony forming units per milliliter (CFU/mL).

{4}------------------------------------------------

The interpretation of the growth on a pair of plates follows a decision tree whereby the two plate reports are combined and a designation is made for the sample.

The colony morphologies reported by the APAS Compact with Urine Analysis Module are shown in Table 2 for sheep blood agar and Table 3 for MacConkey agar. A summary of the result logic used for automated interpretation of the colony counts and morphology is shown in Table 4. A "significant organism" at 103 CFU/mL on the plate will trigger designation of that plate as "Positive," as opposed to "Review." Plates that are designated as "Positive" or "Review" undergo further evaluation by a clinical microbiologist.

Table 2. Colony morphologies on sheep blood agar identified by APAS Compact with Urine Analysis Module

Morphology NameDescription
ColiformGram-negative and coliform-like colonies
Cream-whiteStaphylococcus and related species
GranularGranular morphologies (e.g., Pseudomonas spp.)
SmallEnterococcus spp. and related species
Small α-hemolyticSmall colonies with α-hemolysis or very small colonies
Small ß-hemolyticSmall colonies with ß-hemolysis
Swarming organismProteus spp. and related, high-motility species

Table 3. Colony morphologies on MacConkey agar identified by APAS Compact with Urine Analysis Module

Morphology NameDescription
Lactose fermenterPink/red colonies
Non-fermenterColonies without red/pink pigment
Non-fermenter with green pigmentPseudomonas spp.
Red pigmented coloniesPigmented Serratia marcescens

Table 4. Result interpretation for APAS Compact with Urine Analysis Module

Colony Count (CFU/mL)MorphologyAPAS Compact Designation
Not ApplicableSwarming organism
$>=10^4$AnyPositive
$10^3$Significant organism1
$10^3$No significant organismReview
0Not ApplicableNegative

1 Defined as growth of coliform or small colonies with B-hemolysis on blood agar or lactose fermenting colonies on MacConkey agar

{5}------------------------------------------------

L. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

The reproducibility of colony counts performed by the APAS Compact with Urine Analysis Module was evaluated on different instruments over multiple runs. Testing was performed using logio-fold dilutions of 0.5 McFarland standard suspensions of representative uropathogenic species that produced from 0 to >100 colonies per plate. Each dilution was plated in triplicate and incubated for 18 hours at 35°C. Images of each plate were taken at 5 different orientations (0, 60, 120, 180 and 270°) on each of 3 APAS Compact instruments. The mean and standard deviation of the resulting colony counts was determined for each dilution on each instrument and overall (Tables 5 and 6). The results demonstrated acceptable reproducibility of colony counts between instruments and rotations of the same plate.

{6}------------------------------------------------

SpeciesDilution1ReplicateInstrument 1Instrument 2Instrument 3Overall
Mean%CVMean%CVMean%CVMean%CVMean%CV
E. coli3176.21.983.24.474.02.177.86.078.37.0
3285.83.682.84.282.61.883.73.6
3372.82.573.62.673.43.773.32.8
418.610.38.25.58.25.58.37.518.639.2
E. coli4223.21.923.81.923.00.023.32.1
4323.81.923.86.224.42.324.03.9
515.69.85.69.85.69.85.69.13.275.0
524.00.04.00.04.00.04.00.0
530.0NA0.0NA0.0NA0.0NA
31207.85.54242.46.16165.05.7205.116.9202.238.0
32141.86.09135.610.18108.64.3128.713.6
33345.610.60297.64.23175.64.0272.928.2
4114.46.1816.413.3515.27.215.310.719.629.6
E. coli/S. agalactiae4218.87.8719.212.9714.217.517.417.9
4328.85.1429.47.4520.213.726.118.4
512.846.43.453.56.026.34.150.07.879.5
5215.617.317.69.514.813.916.014.8
533.433.53.023.73.650.63.337.3
31326.09.0317.63.6292.81.8312.17.2271.118.6
32208.03.0201.63.3203.21.8204.32.9
33310.21.8285.02.6295.41.7296.94.1
4133.611.033.26.524.215.330.317.936.719.3
E. faecalis4245.010.141.813.041.24.742.710.0
4341.611.935.614.933.85.737.014.3
518.418.17.020.14.434.56.633.39.638.5
528.427.49.417.88.814.88.919.4
5314.418.814.620.911.417.113.521.1
SalineNA10.0NA0.22250.0NA0.1260.00.0NA
SalineNA20.0NA0.0NA0.0NA0.0NA
SalineNA30.0NA0.0NA0.2225.00.1260.0

Table 5. Reproducibility of colony counts on blood agar obtained by the APAS Compact with Urine Analysis Module

NA: Not Applicable; %CV: Percent Coefficient of Variation
¹ log10-fold dilution of 0.5 McFarland standard suspension

¹log₁₀-fold dilution of 0.5 McFarland standard suspension

{7}------------------------------------------------

SpeciesDilution1ReplicateInstrument 1Instrument 2Instrument 3Overall
Mean%CVMean%CVMean%CVMean%CV
E. coli31152.23.0140.23.2140.22.3144.24.8
2174.64.5162.41.3160.01.8165.74.9
3247.22.1233.03.4231.82.8237.34.0
E. coli4148.25.441.01.742.45.943.98.6
248.83.949.22.247.63.248.53.3
344.22.544.67.943.63.544.14.9
E. coli514.220.04.619.33.828.94.222.4
22.00.01.00.01.00.01.337.7
38.821.87.620.08.215.98.219.1
Saline110.0NA0.0NA0.0NA0.0NA
20.0NA0.0NA0.0NA0.0NA
31.071.00.2225.00.0NA0.4157.5
182.422.5
45.57.5
4.665.2
0.1400.0

Table 6. Reproducibility of colony counts on MacConkey agar obtained by the APAS Compact with Urine Analysis Module

NA: Not Applicable; %CV: Percent Coefficient of Variation

1 log10-fold dilution of 0.5 McFarland standard suspension

{8}------------------------------------------------

b. Linearity/assay reportable range:

To determine the accuracy of colony counts obtained using the APAS Compact with Urine Analysis Module, a study was conducted using 10-fold serial dilutions of bacterial cultures that were plated on blood and/or MacConkey agar, as appropriate to the organism/organism combination. The plates were incubated and read independently by two microbiologists who each estimated colony counts. The mean of the two manual counts was used as the reference result for each plate. Each plate was also analyzed in 5 different orientations using APAS Compact with Urine Analysis Module (0, 60, 120, 180 and 270°). Note: plates that exhibited no growth on initial reading were imaged once and were not rotated to different orientations. The individual APAS Compact colony counts obtained with each plate in each orientation were compared to those obtained by the manual reference method. For analysis of the data, counts obtained by both APAS Compact and the reference method were each grouped into four categories as shown below:

Colonies/plateCategory (CFU/mL)1
0No Growth
1-9$10^3$
10-99$10^4$
≥100≥ $10^5$

1 Based on a 1uL inoculum volume

The results presented in Tables 7 and 8 show that all low counts obtained by APAS Compact with Urine Analysis Module were within 1-log10 of those obtained by the manual reference method. APAS Compact did not incorrectly designate any of the cultures with growth as "Negative" and there were several cases in which APAS Compact designated plates as positive for growth (i.e., ≥10° CFU/mL) but for which no growth was reported by the reference method. Together with the results of the Clinical Study in Section L(3), these data indicate that there is low probability that the APAS Compact with Urine Analysis Module will produce false negative results due to the failure to detect visible microbial colonies. In addition, because any growth detected by the APAS Compact is subject to follow-up by a trained microbiologist. the risk to patients from incorrect assignment of colony counts by the instrument is mitigated. The results of this study are therefore acceptable and provide evidence of the accuracy of colony counts obtained using the APAS Compact with Urine Analysis Module.

{9}------------------------------------------------

E. coliReference Method (CFU/mL)
No Growth103104≥105Total
No Growth00000
1033372042
APAS(CFU/mL)1040850563
≥1050013110123
Total34565115228
% APAS < ManualNA03.14.3
% APAS = Manual082.276.995.7
% APAS > Manual10017.820.0NA
E. faecalisReference Method (CFU/mL)
No Growth103104≥105Total
No Growth00000
10314430057
APAS(CFU/mL)10403710047
≥1050045105150
Total148055105254
% APAS < ManualNA000
% APAS = Manual053.818.2100
% APAS > Manual10046.381.8NA
P. vulgarisReference Method (CFU/mL)
No Growth103104≥105Total
No Growth00000
1038542064
APAS(CFU/mL)104317330107
≥1050008080
Total115575110251
% APAS < ManualNA02.727.3
% APAS = Manual098.297.372.7
% APAS > Manual1001.80.0NA
P. aeruginosaReference Method (CFU/mL)
No Growth103104≥105Total
No Growth00000
1030380038
APAS(CFU/mL)10402132053
≥1050023125148
Total05955125239
% APAS < ManualNA000
% APAS = ManualNA64.458.2100
S. saprophyticusReference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth00000
10310790089
1045150561
≥1050009595
Total158050100245
% APAS < ManualNA005.0
% APAS = Manual098.810095.0
% APAS > Manual1001.30NA
E. coli/E. faecalis 1Reference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth00000
1039475061
10402165894
≥1050009292
Total96870100245
% APAS < ManualNA07.18.0
% APAS = Manual069.192.992.0
% APAS > Manual10030.90NA
E. coli/P. vulgaris/G. vaginalis 2Reference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth00000
1035610066
104014551584
≥1050008585
Total57555100235
% APAS < ManualNA0015.0
% APAS = Manual081.310085.0
% APAS > Manual10018.70NA

Table 7. Correlation of colony count categories obtained on blood agar using the APAS Compact with Urine Analysis Module and the manual reference method

{10}------------------------------------------------

  1. I we spection of culture plates determined the presence of colonies of both 2. coli and be ratio of

equal numbers (1:1 ratio). However, no colonies with morphology consistent with G. vaginalis were observed.

{11}------------------------------------------------

E. coliReference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth00000
1030560056
10401965084
≥105005105110
Total07570105250
% APAS < ManualNA000
% APAS = ManualNA74.792.9100
% APAS > ManualNA25.37.1NA
P. aeruginosaReference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth00000
1030290029
10402640571
≥105003569104
Total0557574204
% APAS < ManualNA006.8
% APAS = ManualNA52.753.393.2
% APAS > ManualNA47.346.7NA
P. vulgarisReference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth00000
1030649073
1040161769
≥105000103103
Total06570110245
% APAS < ManualNA012.96.4
% APAS = ManualNA98.587.193.6
% APAS > ManualNA1.50NA

Table 8. Correlation of colony count categories obtained on MacConkey agar using the APAS Compact with Urine Analysis Module and the manual reference method

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The instructions for use indicate that the APAS Compact requires daily calibration using a dedicated optical tool that is provided with the instrument. Daily calibration of the instruments was performed throughout the analytical and clinical studies that were conducted to evaluate the performance of the device. None of the instruments failed calibration during the course of these studies.

In addition to daily optical calibration, the instructions for use indicate that operators should also perform a daily functional quality control check with cultures of Enterococcus faecalis and Escherichia coli grown on blood and MacConkey agar, respectively. In order to pass, the results from both culture media must meet the prescribed acceptance criteria for colony count and morphological designation. In

{12}------------------------------------------------

addition, uninoculated plates of both types should exhibit no growth. Table 9 shows the results obtained from Quality Controls runs performed during the APAS Compact clinical studies. The expected results were obtained with all (100%) positive controls on both blood and MacConkey agar. For the negative controls, the expected results were obtained with 98.4% on blood agar and 100% on MacConkey agar.

SiteInstrumentPositive ControlNegative Control
TestedExpected Result (%)Blood MacConkeyTestedExpected Result (%)Blood MacConkey
1115050 (100) 50 (100)5049 (98.0)3 50 (100)
1224949 (100) 49 (100)4949 (100) 49 (100)
231313 (100) 13 (100)Note Done4Not Done Not Done
342929 (100) 29 (100)3029 (96.7)3 30 (100)
Total141141 (100) 141 (100)129127 (98.4) 129 (100)

Table 9. Results of Quality Control testing performed during the APAS Compact clinical studies

1 Negative Control for blood agar and 1 for MacConkey agar gave positive results due to artifacts on the plate. Positive and Negative QC runs for both days were repeated successfully.

2 Negative Controls for blood agar gave positive results due to artifacts on the plate. Positive and Negative QC runs for that day were repeated successfully.

3 Both Negative Control failures on blood agar were due to the presence of a single contaminating colony.

4 The study at Site 2 was completed prior to the implementation of daily Negative Control testing.

d. Detection Limit:

Detection of colonies by the APAS Compact is, in part, dependent upon colony size. A study was therefore conducted using a representative panel of UTI pathogens to determine the effect of colony size of the ability to detect growth. The organisms in the study were selected to include species that exhibit each of the characteristic morphologies used by the Urine Analysis Module software for designation of results.

Each organism in the study was diluted to different concentrations and inoculated onto blood and/or MacConkey agar plates which were incubated at 35±2°C until pin point colonies were visible to the naked eye. Images of the plates were then captured by the APAS Compact and the operator digitally labeled multiple isolated colonies to measure their diameter. The image of each plate was then analyzed using the Urine Analysis Module software to generate the corresponding APAS Compact result. The measurements of the colonies that were successfully detected by the APAS Compact were used to calculate the minimum colony size for each organism that could be detected with 95% probability. Results of the study are presented in Tables 10 and 11 and demonstrate the limits of colony size that can be detected reliably by the APAS Compact with Urine Analysis Module.

{13}------------------------------------------------

Table 10. Limits of detection (LOD) of the APAS Compact with Urine Analysis Module for organisms cultured on blood agar

DetectedNot DetectedLOD in mm(95% CI)
NColony Diameter (mm)NColony Diameter (mm)
MinMaxMinMax
Aerococcus viridans1410.2721.268810.2721.1781.258(1.049, 1.692)
Enterococcus faecalis2250.1811.1781650.1811.0870.856(0.776, 0.980)
Escherichia coli2400.1811.7211230.1810.9060.899(0.821, 1.024)
Proteus mirabilis1400.6342.8082810.1812.7172.501(2.168, 3.091)
Pseudomonas aeruginosa2380.4533.4421530.1812.3554.037(2.881, 7.121)
Staphylococcussaprophyticus1210.3620.996980.1810.9060.877(0.784, 1.049)
Streptococcus agalactiae1200.3620.7251380.1810.7250.625(0.584, 0.697)

N: number of colonies; Min: minimum; Max: maximum

Table 11. Limits of detection (LOD) of the APAS Compact with Urine Analysis Module for organisms cultured on MacConkey agar

DetectedNot Detected
NColony Diameter (mm)Colony Diameter (mm)LOD in mm(95% CI)
MinMaxNMinMax
Escherichia coli1120.9962.2641360.3621.1781.253(1.214, 1.324)
Morganella morganii1700.6341.7211700.2721.2681.262(1.164, 1.421)
Pseudomonas aeruginosa480.6342.989780.1811.0871.170(1.020, 1.596)
Serratia marcescens760.7251.0872960.1811.0871.613(1.398, 2.093)

N: number of colonies; Min: minimum; Max: maximum

e. Analytical Specificity:

The ability of the APAS Compact with Urine Analysis Module to assign the appropriate colony morphology to different clinically relevant organisms associated with UTIs was evaluated by testing serial dilutions of both pure and mixed cultures. Dilutions of each organism or organism combination were inoculated in triplicate onto the appropriate culture plates which were incubated at 35±2℃ for 18 hours. The APAS Compact was used to capture images of plates that exhibited 1-100 individual colonies or areas of confluent growth and >100 isolated colonies. Each plate was read on the APAS Compact instrument in 3 different orientations (0, 120 and 270°). The plates were then re-incubated and imaged again after 20 and 22 hours. Results for assignment of colony morphology are shown for pure and mixed cultures below in Tables 12-15. The percentage of plate images with at least one colony of the expected morphology is reported as agreement.

{14}------------------------------------------------

Pure Cultures

With pure cultures on blood agar, the APAS Compact with Urine Analysis Module reported detection of the expected colony morphology on each image for all the species tested at each time point except with Aerococcus urinae, Lactobacillus rhamnosus, Pseudomonas aeruginosa and Staphylococcus epidermidis (Table 12). For all species except A. urinae and L. rhamnosus, the APAS Compact designation of each plate at each time point was either "Positive" or "Review."

For S. marcescens on MacConkey agar, the expected morphology was observed on all plates after 20 and 22 hours of incubation but not after 18 hours (Table 13). All images of cultures on MacConkey agar were designated as "Positive" by the APAS Compact.

Table 12. Detection of expected colony morphology by the APAS Compact with Urine Analysis Module: pure cultures on blood agar

SpeciesExpected ColonyMorphologyAgreement1
Number/TotalPercent
Aerococcus urinaeSmall, alpha hemolytic53/5498.12
Aerococcus viridansSmall, alpha hemolytic54/54100
Candida albicansSmall54/54100
Enterobacter cloacaeColiform54/54100
Escherichia coli (strain 1)Coliform54/54100
Escherichia coli (strain 2)Coliform54/54100
Enterococcus faecalisSmall54/54100
Klebsiella pneumoniaeColiform52/52100
Lactobacillus rhamnosusSmall6/2722.23
Morganella morganiiColiform54/54100
Proteus mirabilisSwarming54/54100
Pseudomonas aeruginosaGranular50/5198.04
Staphylococcus aureusCream white54/54100
Staphylococcus epidermidisCream white38/5470.45
Staphylococcus saprophyticusCream white51/51100
Streptococcus agalactiaeBeta-hemolytic54/54100
Streptococcus dysgalactiaeBeta-hemolytic51/51100

Note: The APAS Compact designation of each plate at each time point was either "Positive" for all species except A. urinae and L. rhamnosus

Number of plate images with at least one colony of the expected morphology (all time points and plate rotations combined)/total

ର Expected morphology detected: 18 hours: 18/18 (88.9%): 20 hours: 18/18 (100%): 22 hours: 17/18 (94.4%)

3 Expected morphology detected: 18 hours: 0/9 (0%); 20 hours: 0/9 (0%); 22 hours: 6/9 (66.7%)

4 Expected morphology detected: 18 hours: 18/18 (100%); 20 hours: 17/18 (94.4%); 22 hours: 15/15 (100%)

్ Expected morphology detected: 18 hours: 9/18 (50.0%); 20 hours: 11/18 (61.1%); 22 hours: 18/18 (100%)

{15}------------------------------------------------

Table 13. Detection of expected colony morphology by the APAS Compact with Urine Analysis Module: pure cultures on MacConkey agar

SpeciesExpected ColonyMorphologyAgreement1
Number/TotalPercent
Citrobacter koseriNon-fermenter53/53100
Escherichia coliLactose fermenter54/54100
Klebsiella pneumoniaeLactose fermenter54/54100
Morganella morganiiNon-fermenter54/54100
Serratia marcescens 2Red pigment21/2777.8

Note: All plate images of pure cultures on MacConkey agar were designated by APAS Compact as "Positive"

Number of plate images with at least one colony of the expected morphology (all time points and plate rotations combined)/total

2 Expected morphology detected: 3/9 (33.3%); 20 hours: 9/9 (100%); 22 hours: 9/9 (100%)

Mixed Cultures

For blood agar plates inoculated with two organisms at a 1:1 ratio, the APAS Compact reported detection of both the expected colony types at each time point in all but two images (Table 14). In both cases, the APAS Compact failed to detect the expected morphology for S. agalactiae in the presence of E. faecalis, although both images were reported as "Positive" indicating of the need for microbiological followup. With plates inoculated using a 1:10 ratio of S. agalactiae:E. faecalis there was a higher proportion of images in which colonies with the characteristic morphology of S. agalactiae were not detected. Nevertheless, all the images were still reported as "Positive."

For mixed cultures plated on MacConkey agar, both the expected colony types were detected in all but 1 image, although this image was still reported "Positive" by the instrument (Table 15).

Note: Because any growth detected by the APAS Compact with Urine Analysis Module is subject to follow-up by a trained microbiologist, the risk to patients from incorrect assignment of colony morphology by the instrument is mitigated.

Instructions for appropriate laboratory follow-up of any growth detected by the APAS Compact with Urine Analysis Module are included in the device labeling.

{16}------------------------------------------------

Table 14. Detection of expected colony morphology by the APAS Compact with Urine Analysis Module: mixed cultures on blood agar

Species Name & Expected MorphologyAgreement (%) 2
Species 1Species 2Mixture 1Species 1Species 2
Escherichia coliColiformEnterococcus faecalisSmall1:163/63(100)63/63(100)
1:1081/81(100)81/81(100)
Streptococcus agalactiaeBeta-hemolyticEnterococcus faecalisSmall1:181/81(100)81/81(100)
1:1028/63(44.4) 363/63(100)
Escherichia coliColiformStaphylococcussaprophyticusCreamwhite1:154/54(100)54/54(100)
1:1081/81(100)81/81(100)
Escherichia coliColiformAerococcus viridansAlpha-1:181/81(100)81/81(100)
hemolytic1:1081/81(100)81/81(100)

Note: All plate images of mixed cultures on blood agar were designated by APAS Compact as "Positive"

Approximate ratio of concentration, Species 1:Species 2

ରା Number of plate images with at least one colony of the expected morphology (all time points and plate rotations combined)/total

3 Expected morphology detected: 18 hours: 12/21 (57.1%); 20 hours: 10/21 (47.6%); 22 hours: 6/21 (28.6%)

Table 15. Detection of expected colony morphology by the APAS Compact with Urine Analysis Module: mixed cultures on MacConkey agar

Species Name & Expected MorphologyAgreement (%) 2
Species 1Species 2Mixture 1Species 1Species 2
Escherichia coliLactoseFermenterMorganella morganiiNon-fermenter1:153/53(100)53/53(100)
Escherichia coliLactoseFermenterMorganella morganiiNon-fermenter1:1079/80(98.8) 380/80(100)
Escherichia coliLactoseFermenterPseudomonas aeruginosaNon-fermenter1:181/81(100)81/81(100)
Escherichia coliLactoseFermenterPseudomonas aeruginosaNon-fermenter1:1072/72(100)72/72(100)

Instances of agreement with the expected morphology <100% are shaded

Note: All images of mixed cultures on MacConkey agar were designated by APAS Compact as "Positive"

Approximate ratio of concentration, Species 1:Species 2

2 Number of plate images with at least one colony of the expected morphology (all time points and plate rotations combined)/total

ന Expected morphology detected: 18 hours: 25/26 (96.2%); 20 hours: 27/27 (100%); 22 hours: 27/27 (100%)

  • Assay Cut-off: .f.
    Not applicable.

2. Comparison studies:

  • a. Method comparison with predicate device:
    Not applicable. Refer to Section L(3), Clinical Studies.

{17}------------------------------------------------

  • b. Matrix comparison:
    Not applicable.

3. Clinical Studies:

  • a. Clinical Sensitivity:
    To evaluate the performance of the APAS Compact with Urine Analysis Module, clinical studies were conducted at three sites (1 US and 2 ex-US) using remnant urine samples that were leftover from standard of care culture for suspected UTI. Each urine sample was inoculated onto one sheep blood agar and one MacConkey agar culture plate (both Remel) that were incubated for 18 hours at 35±2℃ prior to analysis by the APAS Compact and by an independent panel of three microbiologists. Results were reported by the Urine Analysis Module software as "Negative," "Review" or "Positive" as described in Table 4, above. For the analysis of performance, APAS Compact designations of "Review" and "Positive" were regarded as positive. For the reference method, each microbiologist was trained to read the urine culture plates in a standard fashion and was blinded to the results from the other panel members and to those obtained by the APAS Compact.

Only urine samples from "uncomplicated" cases of suspected UTI were included in the analysis of performance (i.e., those not obtained from subjects with known underlying urinary tract pathology, suspected cases of persistent urinary tract infection, urine collected by invasive procedures or urine samples from immunocompromised subjects). A summary of the reasons for exclusion of samples from the performance calculations is shown in Table 16.

DescriptionSite1Total
123
Enrolled58352117214810100
Included5634176918219224
Excluded201348327876
Reasons for exclusion
Complicated UTI290029
Read time11560235391
Incomplete record16692114
LIMS Flag203420342

Table 16. Number of urine samples enrolled at each clinical site and reasons for exclusion from the analysis of performance

1 Site 1: US; Sites 2 and 3: ex-US

2 Culture plates not read within specified timeframe after incubation

2 Additional data from Laboratory Information Management System (LIMS) used in algorithm for sample disposition (e.g., white cell count)

Tables 17 and 18 summarize the demographic characteristics of the subjects enrolled at each clinical site. The majority of urine samples were from female donors, which is consistent with the higher prevalence of UTIs in the female population. Samples were

{18}------------------------------------------------

enrolled from subjects ranging from <20 to >80 years of age, with the majority from subjects between 20 and 79 years of age.

Site1MaleFemaleNot KnownTotal% Male% Female
197246593563417.382.7
270210661176939.760.3
381010074182144.655.4
Total248467328922427.073.0

Table 17. Gender distribution of donors of urine samples enrolled in the clinical studies

  • Site 1: US; Sites 2 and 3: ex-US
Site1
Age (years)123Total%
≤207471549799810.8
20-391611415458238425.9
40-591247440450213723.2
60-791460548614262228.4
≥80567212300107911.7
Not Known20240.0
5634176918219224100

Table 18. Age distribution of urine sample donors in the clinical studies

1 Site 1: US; Sites 2 and 3: ex-US

A summary of the APAS Compact designation for samples from the clinical studies in relation to colony counts as determined by the reference microbiologist panel is shown in Tables 19-21. Colony counts by both methods were categorized as described in Section L(1)(b). Across all sites combined, the APAS Compact correctly designated 99.0% (95% confidence interval: 98.7-99.2%) of cultures on blood agar and 99.5% (95% CI: 99.2-99.7%) of cultures on MacConkey agar as positive for growth. Depending on the colony count observed by the microbiologist panel, the false negative rate for detection of microbial growth on blood agar ranged from 0% to 2.9% and on MacConkey agar from 0% to 1.3%.

False-negative results with the APAS Compact for the detection of growth could lead to discard of culture plates without further review. The results of this study demonstrate that the likelihood of such results is acceptably low.

Tables 22, 23 and 24 show the correlation of colony count categories obtained by the APAS Compact and by the reference method. In general, the colony count categories reported by the APAS Compact were the same or higher than those obtained by the reference method, and there were multiple instances in which APAS Compact designated plates that exhibited "No Growth" according to the reference method as "Positive" or "Review." This is acceptable because all cultures designated by APAS Compact as "Positive" or "Review" are subject to further microbiological follow-up and any such incorrect designations do not increase risks to patients.

{19}------------------------------------------------

Site 1Reference ResultAPAS DesignationTotalCorrect 2% Correct(95% CI)
1No Growth1220142561490122081.9 (79.8, 83.8)
1103 CFU/mL373938451275123897.1 (96.0, 97.9)
1104 CFU/mL41089741167116399.7 (99.1, 99.9)
1≥105 CFU/mL01700217021702100 (99.8, 100)
1Any Growth4131829214144410399.0 (98.7, 99.3)
2No Growth438185751343885.4 (82.1, 88.2)
2103 CFU/mL1315230847346097.3 (95.4, 98.4)
2104 CFU/mL24612548848699.6 (98.5, 99.9)
2≥105 CFU/mL02950295295100 (98.7, 100)
2Any Growth159083331256124198.8 (98.0, 99.3)
3No Growth57426163757490.1 (87.5, 92.2)
3103 CFU/mL810126237136397.8 (95.8, 98.9)
3104 CFU/mL23552338037899.5 (98.1, 99.9)
3≥105 CFU/mL3422843343099.3 (98.0, 99.8)
3Any Growth138782931184117198.9 (98.1, 99.4)
TotalNo Growth2232343742640223284.5 (83.1, 85.9)
Total103 CFU/mL4664614152119206197.3 (96.5, 97.9)
Total104 CFU/mL819051222035202799.6 (99.2, 99.8)
Total≥105 CFU/mL32417102430242799.9 (99.6, 100)
TotalAny Growth69496815476584651599.0 (98.7. 99.2)

Table 19. Performance of the APAS Compact with Urine Analysis Module for detection of growth on blood agar

CI: 95% score Confidence Interval

Site 1: US; Sites 2 and 3: ex-US

4 For the purposes of data analysis, APAS Compact designations of "Positive" were both considered indicative of microbial growth.

For a Reference Result of No Growth, an APAS Compact designation of "Negative" was considered correct.

For Reference Results of 10, 10, ≥10 CFU/nL or Any Growth, an APAS Compact designation of "Positive" or "Review" was considered correct.

{20}------------------------------------------------

Site 1Reference ResultAPAS DesignationTotalCorrect% Correct 2 (95% CI)
1No Growth327423373334327498.2 (97.7, 98.6)
1103 CFU/mL64066747947398.7 (97.3, 99.4)
1104 CFU/mL1469847847799.8 (98.8, 100)
1≥105 CFU/mL2134101343134199.9 (99.5, 100)
1Any Growth92216752300229199.6 (99.3, 99.8)
2No Growth1257441265125799.4 (98.8, 99.7)
2103 CFU/mL21364818618498.9 (96.2, 99.7)
2104 CFU/mL1152115415399.4 (96.4, 99.9)
2≥105 CFU/mL01640164164100 (97.7, 100)
2Any Growth34524950450199.4 (98.3, 99.8)
3No Growth1241201243124199.8 (99.4, 100)
3103 CFU/mL31053914714498.0 (94.2, 99.3)
3104 CFU/mL1144114614599.3 (96.2, 99.9)
3≥105 CFU/mL02850285285100 (98.7, 100)
3Any Growth45344057857499.3 (98.2, 99.7)
TotalNo Growth577229415842577298.8 (98.5, 99.1)
Total103 CFU/mL1164715481280198.6 (97.6, 99.2)
Total104 CFU/mL37651077877599.6 (98.9, 99.9)
Total≥105 CFU/mL2179001792179099.9 (99.6, 100)
TotalAny Growth1632021643382336699.5 (99.2, 99.7)

Table 20. Performance of the APAS Compact with Urine Analysis Module for detection of growth on MacConkey agar

CI: 95% score Confidence Interval ¹ Site 1: US; Sites 2 and 3: ex-US

2 For the purposes of data analysis, APAS Compact designations of "Review" and "Positive" were both considered indicative of microbial growth.

For a Reference Result of No Growth, an APAS Compact designation of "Negative" was considered correct.

For Reference Results of 10, 10, ≥10 CFU/nL or Any Growth, an APAS Compact designation of "Positive" or "Review" was considered correct.

{21}------------------------------------------------

Table 21. Performance of the APAS Compact with Urine Analysis Module for detection of growth by sample (blood and MacConkey agars combined)

Site 1Reference ResultAPAS DesignationTotalCorrect% Correct 2(95% CI)
1No Growth1166252611452116680.3 (78.2, 82.3)
103 CFU/mL374637801280124397.1 (96.0, 97.9)
1104 CFU/mL41104641172116899.7 (99.1, 99.9)
≥105 CFU/mL01729117301730100 (99.8, 100)
Any Growth4132968454182414199.0 (98.7, 99.3)
2No Growth429195750542985.0 (81.6, 87.8)
103 CFU/mL1316630047946697.3 (95.4, 98.4)
2104 CFU/mL14642549048999.8 (98.9, 100)
≥105 CFU/mL02950295295100 (98.7, 100)
Any Growth149253251264125098.9 (98.1, 99.3)
3No Growth55536061855589.8 (87.2, 92.0)
103 CFU/mL911526238637797.7 (95.6, 98.8)
3104 CFU/mL23562338137999.5 (98.1, 99.9)
≥105 CFU/mL3426743643399.0 (98.0, 99.8)
Any Growth148972921203118998.8 (98.1, 99.3)
TotalNo Growth2150473782575215083.5 (96.7, 99.1)
103 CFU/mL5974413422145208697.2 (96.5, 97.9)
Total104 CFU/mL719241122043203699.7 (99.3, 99.8)
≥105 CFU/mL3245082461245899.9 (99.6, 100)
Any Growth69511814626649658099.0 (98.7, 99.2)

CI: 95% score Confidence Interval

¹ Site 1: US; Sites 2 and 3: ex-US

2 For the purposes of data analysis, APAS Compact designations of "Review" and "Positive" were both considered indicative of microbial growth.

For a Reference Result of No Growth, an APAS Compact designation of "Negative" was considered correct.

For Reference Results of 10, 10, ≥10 CFU/nL or Any Growth, an APAS Compact designation of "Positive" or "Review" was considered correct.

{22}------------------------------------------------

Site 1 (US)Reference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth122037401261
10326210498721400
1047182828271044
≥1051724816731929
Total14901275116717025634
% APAS < ReferenceNA2.97.81.7
% APAS = Reference81.982.371.098.3
% APAS > Reference18.114.821.3NA
Site 2 (ex-US)Reference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth4381320453
10368388311488
10476336821459
≥1050987273369
Total5134734882951769
% APAS < ReferenceNA2.76.87.5
% APAS = Reference85.482.075.492.5
% APAS > Reference14.615.217.8NA
Site 3 (ex-US)Reference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth574823587
10363312278410
10404524512302
≥10506106410522
Total6373713804331821
% APAS < ReferenceNA2.27.65.3
% APAS = Reference90.184.164.594.7
% APAS > Reference9.913.727.9NA
All Sites CombinedReference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth223258832301
1033931749145112298
104142901441601805
≥10512244123562820
Total26402119203524309224
% APAS < ReferenceNA2.77.53.0
% APAS = Reference84.582.570.897.0
Site 1 (US)Reference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth32746123283
10356440110507
10443340738482
≥105005913031362
Total333447947813435634
% APAS < ReferenceNA1.32.53.0
% APAS = Reference98.291.985.197.0
% APAS > Reference1.86.912.3NA
Site 2 (ex-US)Reference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth12572101260
103816840180
1040161368160
≥1050013156169
Total12651861541641769
% APAS < ReferenceNA1.13.24.9
% APAS = Reference99.490.388.395.1
% APAS > Reference0.68.68.4NA
Site 3 (ex-US)Reference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth12413101245
103212730132
1040171153135
≥1050027282309
Total12431471462851821
% APAS < ReferenceNA2.02.71.1
% APAS = Reference99.886.478.898.9
% APAS > Reference0.211.618.5NA
All Sites CombinedReference Method (CFU/mL)
No Growth103104≥105Total
APAS(CFU/mL)No Growth577211325788
10366735180819
10446665849777
≥105009917411840
Total584281277817929224
% APAS < ReferenceNA1.42.72.8
% APAS = Reference98.890.584.697.2
Site 1 (US) Reference Method (CFU/mL)
APAS (CFU/mL)No Growth103104≥105Total
No Growth116637401207
10327610598621423
APAS(CFU/mL)1049178836201043
≥1051624617081961
Total14521280117217305634
% APAS < ReferenceNA2.97.71.3
% APAS = Reference80.382.771.398.7
% APAS > Reference19.714.421.0NA
Site 2 (ex-US) Reference Method (CFU/mL)
APAS (CFU/mL)No Growth103104≥105Total
No Growth4291310443
10369393321495
APAS(CFU/mL)10476436716454
≥1050990278377
Total5054794902951769
% APAS < ReferenceNA2.76.75.8
% APAS = Reference85.082.074.994.2
% APAS > Reference15.015.218.4NA
Site 3 (ex-US) Reference Method (CFU/mL)
APAS (CFU/mL)No Growth103104≥105Total
No Growth555923569
10363326267422
APAS(CFU/mL)10404524512302
≥10506108414528
Total6183863814361821
% APAS < ReferenceNA2.37.35.0
% APAS = Reference89.884.564.395.0
% APAS > Reference10.213.228.3NA
All Sites Combined Reference Method (CFU/mL)
APAS (CFU/mL)No Growth103104≥105Total
No Growth215059732219
1034081778144102340
APAS(CFU/mL)104162871448481799
≥10512144424002866
Total25752145204324619224
% APAS < ReferenceNA2.87.42.5
% APAS = Reference83.582.970.997.5

Table 22. Correlation of colony count categories obtained on blood agar by the APAS Compact with Urine Analysis Module and the manual reference method

{23}------------------------------------------------

Table 23. Correlation of colony count categories obtained on MacConkey agar by the APAS Compact with Urine Analysis Module and the manual reference method

{24}------------------------------------------------

Table 24. Correlation of colony count categories obtained by APAS Compact with Urine Analysis Module and the manual reference method (blood agar and MacConkey agar combined)

{25}------------------------------------------------

  • b. Clinical specificity:
    Refer to Section 3(a), above.

  • c. Other clinical supportive data (when a. and b. are not applicable):
    Not applicable.

    1. Clinical cut-off:
      Not applicable.
    1. Expected values/Reference range:
      Tables 25-27 show the percentage samples with microbial growth at different levels observed during the clinical studies by the reference method (manual inspection of plates) and by the APAS Compact with Urine Analysis Module. Refer to Section L(1)(b) for categorization of colony counts.

Table 25. Percentage of samples with microbial growth at different levels on blood agar as determined by the reference method and APAS Compact with Urine Analysis Module

CategoryPercentage of SamplesReference Method 1APAS Compact
Site 1Site 2Site 3OverallSite 1Site 2Site 3Overall
No growth26.429.035.028.622.425.632.224.9
103 CFU/mL22.626.720.423.024.827.622.524.9
104 CFU/mL20.727.620.922.118.525.916.619.6
≥105 CFU/mL30.216.723.826.334.220.928.730.6
Any Growth73.671.065.071.477.674.467.875.1

Site 1: US; Sites 2 and 3: ex-US

1 Manual review of urine cultures by a panel of microbiologists (refer to Section L(3)(a))

Table 26. Percentage of samples with microbial growth at different levels on MacConkey agar as determined by the reference method and APAS Compact with Urine Analysis Module

Percentage of Samples
CategoryReference Method 1APAS Compact
Site 1Site 2Site 3OverallSite 1Site 2Site 3Overall
No growth59.271.568.363.358.371.268.462.7
103 CFU/mL8.510.58.18.89.010.27.28.9
104 CFU/mL8.58.78.08.48.69.07.48.4
≥105 CFU/mL23.89.315.719.424.29.617.019.9
Any Growth40.828.531.736.741.728.831.637.3

Site 1: US; Sites 2 and 3: ex-US

4 Manual review of urine cultures by a panel of microbiologists (refer to Section L(3)(a))

{26}------------------------------------------------

Table 27. Percentage of samples with microbial growth at different levels on blood agar and MacConkey agar combined as determined by the reference method and APAS Compact with Urine Analysis Module

Percentage of Samples
CategoryReference Method 1APAS Compact
Site 1Site 2Site 3OverallSite 1Site 2Site 3Overall
No growth25.828.533.927.921.425.031.224.1
103 CFU/mL22.727.121.223.325.328.023.225.4
104 CFU/mL20.827.720.922.118.525.716.619.5
≥105 CFU/mL30.716.723.926.734.821.329.031.1
Any Growth74.271.566.172.178.675.068.875.9

Site 1: US; Sites 2 and 3: ex-US

1 Manual review of urine cultures by a panel of microbiologists (refer to Section L(3)(a))

M. Instrument Name:

APAS Compact

N. System Descriptions:

1. Modes of Operation:

Manual loading/unloading of individual culture plates and image capture, followed by automated analysis and compilation of the test report.

    1. Software:
      FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes _____________or No _______________________________________________________________________________________________________________________________________________________

The sponsor provided the results of verification and validation testing for the following software components:

Component NameFunction
Instrument Controller SoftwareGraphical User Interface (GUI) application thatsends and receives messages to and from theAPAS Controller.
LIMS Importer SoftwareGUI application that imports LIMS data and addsit the database that is accessed by the InstrumentController Software.
APAS Controller SoftwareEncodes the image processing and assessmentalgorithms.
Urine Analysis Module SoftwareComprised of the Assessment and DecisionPackages (AP and DP) which direct the captureand processing of culture plate images,respectively.

{27}------------------------------------------------

3. Specimen Identification:

Sample identification is entered into the Instrument Controller Software either by typing manually or using a barcode scanner.

4. Specimen Sampling and Handling:

Culture plates are loaded and unloaded manually, one at a time.

    1. Calibration:
      The APAS Compact requires daily color calibration prior to use. Calibration is performed using the Color Calibration Tool provided with the instrument. Instructions for calibration are included in the APAS Compact User Manual.

6. Quality Control:

Instructions for daily Quality Control testing using reference culture plates inoculated with Enteroccocus faecalis and Escherichia coli are provided in the User Manual for the Urine Analysis Module. Both the appropriate colony morphology and colony count must be obtained in order for the results of the Quality Control testing to be considered acceptable. Appropriate instructions are provided in the User Manual in the event of control failure.

O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not applicable.

P. Proposed Labeling:

The labeling is sufficient and satisfies the requirements of 21 CFR parts 801 and 809 as well as the Special Controls for this type of device.

O. Identified Risks to Health and Identified Mitigations:

Identified Risks to HealthIdentified Mitigations
False positive results (i.e., incorrect designation of plates for "Review" or as "Positive")General controls and special controls: (1), (2), (3), (4), (5), (6), (7)
False negative results (i.e., failure to detect growth and incorrect designation of plates as "Negative")General controls and special controls: (1), (2), (3), (4), (5), (6), (7)

{28}------------------------------------------------

R. Benefit/Risk Analysis:

Summary
Summary of the Benefit(s)The primary benefits of the APAS Compact are to the testing laboratory,rather than directly to the patient. Use of the technology to screen urinecultures will allow laboratories to focus attention on those that exhibitgrowth and eliminate negative cultures from the workflow, therebyreducing the overall workload. This may have a remote benefit topatients by reducing the time to result for urine cultures that exhibitgrowth, although this was not evaluated under the current de novosubmission.
Summary of the Risk(s)The risks associated with the APAS Compact are the potential for falsepositive and false negative results.False positive results (i.e., incorrect designation of plates for “Review” oras “Positive”) do not appreciably increase the risk of harm to patientsbecause all plates designated as such are required to undergo work-upaccording to Standard Of Care (SOC) practices. The main consequenceof false positive results would therefore be unnecessary expenditure oflaboratory resources.False negative results (i.e., failure to detect growth and incorrectdesignation of plates as “Negative”) could lead to the discard of cultureplates without further review. In turn this may cause a delay in diagnosisor failure to detect infection and/or under treatment of a UTI. UntreatedUTIs can lead to pyelonephritis, renal abscesses, bacteremia and/or

{29}------------------------------------------------

Summary
ConclusionsDo the probable benefitsoutweigh the probable risks?Yes. There are no benefit-risk considerations that would precludegranting the sponsor's de novo application. The probable benefits of theAPAS Compact are likely to outweigh the risk in the light of the specialcontrols and applicable general controls, including design controls.The primary risk associated with the APAS Compact is false negativeresults. False positive results do not substantially increase the risk ofharm to patients relative to SOC testing. All plates designated by APASCompact as “Review” or “Positive” are required to undergo additionalevaluation and appropriate work-up by a trained microbiologistaccording to current SOC practices. APAS Compact results of “Review”or “Positive” would therefore not be reported if the SOC evaluation was“Negative.”The performance observed during the clinical studies indicates that theAPAS Compact accurately identifies the presence of growth on urineculture plates (overall sensitivity for detection of growth was 99.0%[95% CI: 98.7-99.2%]), and that false negative results are unlikely tooccur.The APAS Compact with Urine Analysis Module presents a potentialbenefit to patients by reducing laboratory turnaround time for cultureresults for positive specimens, leading to improvements in patientmanagement and more timely administration of appropriate antimicrobialtreatment.

Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

S. Conclusion:

The information provided in this submission is sufficient to classify this device into Class II under regulation 21 CFR.866.2190. FDA believes that the stated special controls, and applicable general controls, including design controls, provide a reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:

Product Code:PPU
Device Type:Automated image assessment system for microbial colonies on solidculture media
Class:II (special controls)
Regulation:21 CFR 866.2190

{30}------------------------------------------------

(a) Identification.

An automated image assessment system for microbial colonies on solid culture media is an instrument system that is intended to assess the presence or absence of microbial colonies on solid microbiological culture medium, and to interpret their number, phenotypic and morphologic characteristics through analysis of two dimensional digital images as an aid in diagnosis of infectious disease.

(b) Classification.

Class II (special controls). An automated image assessment system for microbial colonies on solid culture media must comply with the following special controls:

    1. Pre-market notification submissions must include a detailed description of the device, including the technology employed, components and software modules, as well as a detailed explanation of the result algorithms and any expert rules that are used to assess colony characteristics and enumerate colonies from image capture through end result.
    1. Pre-market notification submissions must include detailed documentation of the analytical studies performed to characterize device performance to support the intended use, as appropriate.
    1. Pre-market notification submissions must include detailed documentation from clinical studies performed on a population that is consistent with the intended use population.
    • i. The clinical studies must establish the device performance based on comparison to results obtained by an acceptable reference method, as appropriate.
    • ii. The clinical study documentation must include the study protocol with a predefined statistical analysis plan and the final report documenting support for the Indications for Use and the results of the statistical analysis, as appropriate.
    1. Pre-market notification submissions must include detailed documentation for device software, including but not limited to software applications and hardware based components that incorporate software, and any decision making thresholds used to generate results for the device. If a part of a Total Laboratory Automation System, the pre-market notification submission must include detailed documentation addressing the instrument and software system integration.
    1. Pre-market notification submissions must include detailed documentation of appropriate instructions for use regarding the intended user's device quality control procedures for the instrument system and components, as appropriate.
    1. The 21 CFR 809.10 compliant device labeling must include:

{31}------------------------------------------------

  • i. Detailed user instructions to mitigate the risk of failure to operate the instrument correctly.
  • ii. A detailed explanation of the interpretation of results and limitations regarding the need for review of culture plates by a qualified microbiologist, as appropriate.
  • iii. A summary of performance data obtained from the analytical studies used to support device performance, as appropriate.
  • iv. A summary of performance data obtained from clinical studies performed on a population that is consistent with the intended use population, as appropriate.
    1. Under 21 CFR 820.30 compliant design control, device manufacturers must, as appropriate:
    • i. Conduct human factors/usability validation testing with the final version of the labeling and related materials to adequately mitigate the risk of failure to operate the instrument correctly.
    • Document a device training program that will be offered to the end ii. user to adequately mitigate the risk of failure to operate the instrument correctly.

§ 866.2190 Automated image assessment system for microbial colonies on solid culture media.

(a)
Identification. An automated image assessment system for microbial colonies on solid culture media is a system that is intended to assess the presence or absence of microbial colonies on solid microbiological culture medium, and to interpret their number, and phenotypic and morphologic characteristics through analysis of two dimensional digital images as an aid in diagnosis of infectious disease.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include a detailed description of the device, including the technology employed, components and software modules, as well as a detailed explanation of the result algorithms and any expert rules that are used to assess colony characteristics and enumerate colonies from image capture through end result.
(2) Premarket notification submissions must include detailed documentation of the analytical studies performed to characterize device performance to support the intended use, as appropriate.
(3) Premarket notification submissions must include detailed documentation from clinical studies performed on a population that is consistent with the intended use population.
(i) The clinical studies must establish the device performance based on comparison to results obtained by an acceptable reference method, as appropriate.
(ii) The clinical study documentation must include the study protocol with a predefined statistical analysis plan and the final report documenting support for the Indications for Use and the results of the statistical analysis, as appropriate.
(4) Premarket notification submissions must include detailed documentation for device software, including but not limited to software applications and hardware based components that incorporate software, and any decision-making thresholds used to generate results for the device. If a part of a Total Laboratory Automation System, the premarket notification submission must include detailed documentation addressing the instrument and software system integration.
(5) Premarket notification submissions must include detailed documentation of appropriate instructions for use regarding the intended user's device quality control procedures for the instrument system and components, as appropriate.
(6) The 21 CFR 809.10 compliant device labeling must include:
(i) Detailed user instructions to mitigate the risk of failure to operate the instrument correctly.
(ii) A detailed explanation of the interpretation of results and limitations regarding the need for review of culture plates by a qualified microbiologist, as appropriate.
(iii) A summary of performance data obtained from the analytical studies used to support device performance, as appropriate.
(iv) A summary of performance data obtained from clinical studies performed on a population that is consistent with the intended use population, as appropriate.
(7) Under 21 CFR 820.10(c) design and development, device manufacturers must, as appropriate:
(i) Conduct human factors/usability validation testing with the final version of the labeling and related materials to adequately mitigate the risk of failure to operate the instrument correctly.
(ii) Document a device training program that will be offered to the end user to adequately mitigate the risk of failure to operate the instrument correctly.