K Number
DEN150029

Validate with FDA (Live)

Device Name
Permaseal
Date Cleared
2016-07-27

(401 days)

Product Code
Regulation Number
870.4510
Type
Direct
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Permaseal is indicated for soft tissue approximation of cardiac apical tissue during transcatheter valve replacement procedures.

Device Description

The Permaseal device is composed of a delivery device and an implant that is deployed to facilitate access and closure. The deployment site is accessed via minimally invasive surgery or percutaneous surgery. The device is compatible with a 0.025" guide wire. The implant is designed to close punctures, incisions or ostomies in the cardiac apical tissue up to 30F, or 10mm in diameter.

The device features a handle designed to facilitate proper device placement and allow for singlehanded deployment of the implant. The handle contains a trigger that actuates the device and incorporates a mechanical safety. The handle functions as a tool to grip and position the device. It contains the actuating mechanism to control deployment of the anchor using a trigger mechanism.

The implant is composed of an array of eight polypropylene anchors connected by a 2-0 braided, coated polyester suture U.S.P, as seen in Figures 2 and 3. The suture was separately cleared to market through K021019.

The suture is threaded through the eyelets of the eight anchors in a circular pattern and terminates in a surgical knot. The excess suture provided is referred to as the loop limb. The opening that is formed in the center of the implant is referred to as the 'operative window.' The knot is tightened at the end of the procedure by advancing the knot pusher toward the implant. The knot pusher is shown in Figure 4. above. The knot pusher is affixed to the insertion tube by the knot pusher retainer and is removed from the tube before the implant is deployed.

Upon release of the safety and actuation of the trigger, a spring-loaded pusher tube connected distally to 8 pusher pins is released, transmitting the force of the spring through the pusher pins to the anchors, simultaneously embedding the 8 anchors firmly into the tissue at the target site.

The anchors are deployed at a pre-determined depth with the sutures resting on the surface of the heart. Once deployed, the polyester suture and eight anchors create an 'operative window' in a pattern approximating that of a purse-string suture pattern. The anchors provide a secure attachment site for the suture that connects the multiple anchors together. The suture serves as a means to bring the incised tissue edges into apposition so as to close the puncture. Advancing the surgical knot in the suture with the knot pusher creates sufficient tension on the anchors to pull them and the tissue together so as to close punctures, incisions or otomies in the cardiac apical tissue up to 30F, or 10mm in diameter.

AI/ML Overview

Below is a summary of the acceptance criteria and the studies that demonstrate the Permaseal device meets these criteria.

Acceptance Criteria and Reported Device Performance

Test/CharacteristicPurposeAcceptance CriteriaReported Device Performance/Results
Biocompatibility
Delivery System CytotoxicityAssessment of biological reactivity of mammalian cell cultures following incubation with test device extractsNon-cytotoxicNon-cytotoxic
Delivery System SensitizationDetermine the potential for the test device extract to elicit contact dermal allergenicityNon-sensitizingNon-sensitizing
Delivery System IrritationAssess potential of the device to produce irritation following a single intradermal injection of specific extracts prepared from a test deviceNon-irritantNon-irritant
Delivery System HemolysisEvaluate the hemolytic potential of test articlesNon-hemolyticNon-hemolytic
Delivery System Systemic ToxicityScreen test article extracts or solutions for potential toxic effects as a result of a single-dose systemic injection in mice.Non-toxicNon-toxic
Delivery System PyrogenicityEvaluate the test device extract for leachates that have the potential to induce material-mediated pyrogenicity following a single dose injectionNot pyrogenicNot pyrogenic
Implant CytotoxicityEvaluate an extract of a test article for cytotoxicity to mammalian cells in cultureNon-cytotoxicNon-cytotoxic
Implant SensitizationEvaluate the potential of test article to cause delayed dermal contact sensitizationNon-sensitizingNon-sensitizing
Implant IrritationEvaluate the local dermal irritation of a test article extract following intracutaneous injection in rabbitsNon-irritantNon-irritant
Implant Acute Systemic ToxicityEvaluate acute systemic toxicity of test article extract following injection in miceNon-toxicNon-toxic
Implant PyrogenicityDetermine whether an extract of the test article induced a pyrogenic response following intravenous injection in rabbitsNon-pyrogenicNon-pyrogenic
Implant HemocompatibilityEvaluate the potential to cause hemolysisNon-hemolyticNon-hemolytic
Implant GenotoxicityEvaluate the mutagenic potential of the device test article by measuring its ability to induce DNA reverse mutations in S. typhimurium and E. coli in the presence and absence of microsomal enzymesNon-mutagenicNon-mutagenic
Shelf Life/Sterility
Packaging ValidationProduce objective evidence that the package meets requirements for seal visual inspection, peel strength, burst values, and bubble/dye penetration.Seal Visual Inspection: No burns, channels, voids, pleats or foreign matter; Peel Samples: Minimum peel force 1 lb/in; Burst Samples: Minimum burst values recorded; Bubble Samples: No seal leaks/seal bubbles accepted; Dye Samples: No complete seal dye penetration acceptedAll tests passed
Sterilization ValidationEvaluate the sterilization procedure for the Permaseal device and its ability to achieve a Sterility Assurance Level of 10-6.Bioburden, Bacteriostasis/Fungistasis, EO Residual, Bacterial Endotoxin, EtO Acceptable Limit met pre-identified criteria.All tests passed
Shelf Life (1 year)Evaluate the impact of one year real time aging on the Permaseal device.Visual and Functional assessments (i.e., successful ex-vivo device deployment) at baseline and one year.All visual inspections and functional tests passed
Environmental Conditioning & Package IntegrityEvaluate devices after being subjected to a range of temperatures, humidity and simulated distribution testing.Verification testing, performance testing (i.e., successful deployment in porcine heart, compatibility with sheath, functionality of suture grip, knot pusher, knot and anchor eyelets), packaging verification testing, average peel force >1.0lbf/in.All tests passed, any deviations appropriately explained
Package IntegrityDemonstrate that the Permaseal device packaging will provide sufficient protection ensuring device and package integrity are not compromised.Seal Peel: No irregularities or delamination of the seals observed, peel strength greater than 1 lbf/in; Dye Penetration: No dye penetrates through seal.All tests passed
Bench Performance Testing
Design Verification TestingEnsure that the design meets the design inputs as defined in the Product Performance Specifications.All acceptance criteria met, any deviations appropriately justified.All passed
Device WeightEnsure that the device weight meets specification.Device weight meets specification.All passed
Dimensional VerificationEnsure that overall device length, height, width; insertion tube length, diameter; anchor overall length; knot location on suture meet specification.All specified measurements meet specification.All passed
Color/Appearance VerificationEnsure that the handle, insertion tube and safety meet color/ appearance specifications.Handle, insertion tube and safety meet color/ appearance specifications.All passed
Device Operation VerificationEnsure that the device operates to specification by evaluating trigger safety functionality, guidewire compatibility, trigger pull force, safety disengagement force, inability to fire with an engaged safety, anchor deployment depth.Device operates to specification for all listed parameters.All passed
Implant Performance VerificationEvaluate if the implant meets criteria for accommodating intended sheath sizes, suture grip detachment, knot pusher functionality, suture knot functionality and slip force, anchor eyelet functionality, anchor barb extraction, and peak anchor insertion force.Implant meets all listed criteria.All passed
Performance Verification TestingConduct functional performance verification testing in porcine ex vivo hearts.All criteria passed.All criteria passed
Anchor Pull Out TestNo anchor pull out from the tissue at a tensioning force of 1N.No anchor pull out from the tissue at a tensioning force of 1N.Passed, no anchor pull out
Leakage TestNo leakage of solution shall occur at a pressure of 180 mmHg.No leakage of solution shall occur at a pressure of 180 mmHg.Passed, no leakage
Animal Study
Chronic Animal Study (30 Days)Demonstrate the performance of the Permaseal Device compared to traditional mattress suture closure in facilitating transapical access and wound healing chronically in a porcine model.Success in facilitating transapical access and respect to wound healing chronically (30-day survival) in a porcine model with no significant adverse clinical observations or procedural complications.All animals survived to the designated endpoint with no significant adverse clinical observations or procedural complications related to the test devices. Pathological changes were minimal and expected.
Acute Investigational AnalysisEvaluate the performance of the Permaseal in an in vivo porcine model, assessing polypropylene anchors, suture/insertion tube interaction, tensioning for wound closure, implant deployment, sheath passage, and hemostasis.Assessment of performance of polypropylene anchors, suture and insertion tube interaction during deployment, tensioning process for wound closure, implant deployment performance, passage of transapical sheath, and achieving hemostasis.Three separate deployments were completed in one animal with acceptable results.
Clinical Study (STASIS)
Primary Effectiveness EndpointRate of pulsatile bleeding requiring significant surgical intervention (more than one pledgeted suture) at discharge and at 30-day follow-up.Literature derived performance goal was 15%.6.5% (2/31), meeting the primary effectiveness endpoint target.
Safety EndpointAll adverse events (AEs) and serious adverse events (SAEs) occurring during the TA-TAVR procedure and follow-up periods of 30 days, 90 days and 12 months, using VARC 2 definitions.Favorable safety profile (low rates of deaths, myocardial infarctions, strokes; minimal device related SAEs).No deaths, myocardial infarctions or strokes at 30 days. Favorable safety profile maintained at 90 days. Two device-related SAEs early in the study (resolved with training).

Study Details: PERMASEAL Device

2. Sample Size and Data Provenance

  • Test Set (Clinical Study - STASIS):
    • Sample Size: 34 patients were enrolled. 31 patients were included in the primary effectiveness analysis (1 roll-in and 2 protocol violations excluded).
    • Data Provenance: Prospective, multi-center, conducted at five clinical sites in Germany and the Netherlands.

3. Number of Experts and Qualifications (for Ground Truth)

  • Clinical Study (STASIS): The document does not explicitly state the number or specific qualifications of experts used to establish "ground truth" for the clinical endpoints (e.g., diagnosis of pulsatile bleeding, assessment of AEs). However, it is implied that clinical trial investigators (most likely interventional cardiologists or cardiac surgeons) at the participating sites made these assessments based on standard clinical practice and the Valve Academic Research Consortium (VARC) 2 definitions for safety events. The study design is observational, and the endpoints are objective clinical outcomes.

4. Adjudication Method (for Test Set)

  • The document does not explicitly mention a specific adjudication method (e.g., 2+1, 3+1). Clinical events and outcomes would typically be recorded by the study investigators at each site. For Serious Adverse Events (SAEs), causality assessment (relation to device or procedure) was performed, indicating a review by the study team or an independent committee. The two device-related SAEs were attributed to user error and a need for additional training, suggesting internal review of events.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. The STASIS study was designed to evaluate the Permaseal device's performance against a literature-derived performance goal (for the primary effectiveness endpoint) and observed clinical outcomes, rather than comparing human readers with and without AI assistance. The Permaseal is a physical medical device, not an AI diagnostic tool.

6. Standalone Performance Study (Algorithm Only)

  • No, a standalone (algorithm only) study was not conducted. The Permaseal is a physical medical device for apical closure, not an algorithm or AI system. Its performance was evaluated in bench tests, animal studies, and a clinical study with human use.

7. Type of Ground Truth Used (for Test Set)

  • Clinical Study (STASIS): The ground truth for the primary effectiveness endpoint was based on clinical observation and intervention (rate of pulsatile bleeding requiring significant surgical intervention). For safety endpoints, the ground truth was based on clinical diagnoses and outcomes data as defined by the Valve Academic Research Consortium (VARC) 2 definitions, including mortality, myocardial infarction, stroke, vascular complications, etc.
  • Animal Studies: Ground truth was based on direct observation (e.g., survival, freedom from complications, pathological changes) and functional assessment in porcine models.
  • Bench Studies: Ground truth was based on engineering specifications and measurements and functional performance tests.

8. Sample Size for the Training Set

  • The document does not describe a "training set" in the context of an AI algorithm, as the Permaseal is a physical device. All described studies are for evaluating the device's safety and performance itself.

9. How Ground Truth for the Training Set Was Established

  • Not applicable, as there is no "training set" for an AI algorithm. The studies described are for validation of the physical device.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR PERMASEAL

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Apical Closure Device. An apical closure device is a prescription device consisting of a delivery system and implant component that is used for soft tissue approximation of cardiac apical tissue during transcatheter valve replacement procedures.

NEW REGULATION NUMBER: 21 CFR 870.4510

CLASSIFICATION: II

PRODUCT CODE: PNQ

BACKGROUND

DEVICE NAME: PERMASEAL

SUBMISSION NUMBER: DEN150029

DATE OF DE NOVO: JUNE 25, 2015

CONTACT: MICRO INTERVENTIONAL DEVICES, INC. 5 Caufield Pl. SUITE 102 NEWTOWN, PA 18940

REQUESTER'S RECOMMENDED CLASSIFICATION: II

INDICATIONS FOR USE

The Permaseal is indicated for soft tissue approximation of cardiac apical tissue during transcatheter valve replacement procedures.

LIMITATIONS

The sale, distribution and use of the Permaseal device is limited to prescription use onlv.

Limitations on the device use are also achieved through the following statements included in the Instructions for Use:

Contraindications:

  • · Patients who are allergic to contrast agent.

{1}------------------------------------------------

  • · Patients who are allergic to platelet inhibitor therapy.
  • · Patients where a substantial risk of complications due to concomitant therapy, disease state or other condition exists.
  • Patients whose target site for deployment has a myocardial wall thickness less than 10mm.
  • · Tissue that has been compromised due to previous adverse events such as infarction.
  • · Instances where excessive fat is present at the target site for deployment.
  • · Procedures where sheath introducers or catheters of more than 30F in outer diameter are needed.

Warnings:

DO NOT use the Permaseal device if the sterile barrier of the packaging has previously been broken, damaged or if the contents of the package appear to be damaged or defective.

DO NOT handle the Permaseal device without ensuring the safety is in the ON position.

DO NOT look into the distal end of the device or point the device at another individual.

DO NOT use rapid pacing when deploying the Permaseal device for cardiac applications.

DO NOT deploy the device at the true apex.

DO NOT attempt to pass sheath introducers or catheters of more than 30F (10mm) in outer diameter through the "operative window" of the Permaseal implant.

DO NOT use the Permaseal device if the expiration date has elapsed, as either sterility or performance may be compromised.

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

The Permaseal device is composed of a delivery device and an implant that is deployed to facilitate access and closure. The deployment site is accessed via minimally invasive surgery or percutaneous surgery. The device is compatible with a 0.025" guide wire. The implant is designed to close punctures, incisions or ostomies in the cardiac apical tissue up to 30F, or 10mm in diameter.

The device features a handle designed to facilitate proper device placement and allow for singlehanded deployment of the implant. The handle contains a trigger that actuates the device and incorporates a mechanical safety. The handle functions as a tool to grip and position the device. It contains the actuating mechanism to control deployment of the anchor using a trigger mechanism.

{2}------------------------------------------------

With the safety in the ON position, it prevents movement of the trigger and reduces the potential for accidental firing. The insertion tube extends from the handle and contains the implant. The insertion tube is 14cm long to ensure sufficient access to the target site. The device handle and insertion tube can be viewed below in Figure 1.

Image /page/2/Figure/1 description: This image shows a diagram of a surgical tool with several labeled parts. The tool has a handle, a trigger, and a safety mechanism. The front of the tool has a guidewire channel, a suture slot, a knot pusher, a knot pusher retainer, and an insertion tube.

Figure 1. Permaseal™ Delivery System

The implant is composed of an array of eight polypropylene anchors connected by a 2-0 braided, coated polyester suture U.S.P, as seen in Figures 2 and 3. The suture was separately cleared to market through K021019.

Image /page/2/Figure/4 description: The image shows two figures, Figure 2 and Figure 3, with labels pointing to different parts of the figures. Figure 2, titled "Permaseal Implant", shows a circular structure with labels such as "Eight Polypropylene Anchors", "Operative Window", "One Coated, Braided Polyester Suture", "Loop Limb", and "Surgical Knot". Figure 3, titled "Anchor", shows a pointed structure with labels such as "Eyelet", "Flexible Shaft", "Barb", "Three Identical Barbs", and "Tip".

{3}------------------------------------------------

Image /page/3/Figure/0 description: The image shows a diagram of a surgical tool. The tool has several labeled parts, including an implant, knot, loop limb, knot pusher, suture grip, and score. The implant is a circular structure with small protrusions around its perimeter. The suture grip is located at the end of the tool.

Figure 4. Knot Pusher

The suture is threaded through the eyelets of the eight anchors in a circular pattern and terminates in a surgical knot. The excess suture provided is referred to as the loop limb. The opening that is formed in the center of the implant is referred to as the 'operative window.' The knot is tightened at the end of the procedure by advancing the knot pusher toward the implant. The knot pusher is shown in Figure 4. above. The knot pusher is affixed to the insertion tube by the knot pusher retainer and is removed from the tube before the implant is deployed.

Upon release of the safety and actuation of the trigger, a spring-loaded pusher tube connected distally to 8 pusher pins is released, transmitting the force of the spring through the pusher pins to the anchors, simultaneously embedding the 8 anchors firmly into the tissue at the target site. A properly deployed anchor is demonstrated in Figures 5 through 7, below.

Image /page/3/Figure/4 description: The image shows three diagrams of a medical device being deployed into the epicardium. The first diagram, labeled "Figure 5. Properly Deployed", shows the device fully inserted into the epicardium. The second diagram, labeled "Figure 6. Partially Deployed", shows the device partially inserted into the epicardium. The third diagram, labeled "Figure 7. Un-Deploy", shows the device not inserted into the epicardium.

Image /page/3/Figure/5 description: The image is a title that reads "Figure 5. Properly Deployed Anchors". The text is centered on the page. The font is a serif font.

Figure 6. Partially Deployed Anchors

Figure 7. Un-Deployed Anchors

The anchors are deployed at a pre-determined depth with the sutures resting on the surface of the heart. Once deployed, the polyester suture and eight anchors create an 'operative window' in a pattern approximating that of a purse-string suture pattern. The anchors provide a secure attachment site for the suture that connects the multiple anchors together. The suture serves as a means to bring the incised tissue edges into apposition so as to close the puncture. Advancing the surgical knot in the suture with the knot pusher creates sufficient tension on the anchors to pull them and the tissue together so as to close punctures, incisions or otomies in the cardiac apical tissue up to 30F, or 10mm in diameter.

{4}------------------------------------------------

SUMMARY OF NONCLINICAL/BENCH STUDIES

BIOCOMPATIBILITY/MATERIALS

The Permaseal implant is tissue contacting and the delivery system is a limited contacting, external communicating device that comes in contact with tissue.

The distal 10cm of the delivery system, including the implant, was tested to assess the biocompatibility of the delivery system. This was appropriate because all limited contact exposure materials are contained within the distal 10cm of the device. In accordance with ISO 10993-1 and relevant subparts, cytotoxicity, sensitization testing was performed to assess the delivery system.

The Permaseal implant was tested separately using the following biocompatibility tests: cvtotoxicity, intracutaneous reactivity, sensitization, material mediated pyrogenicity, acute systemic toxicity, hemolysis and bacteria reverse mutation. The implant did not require implantation, sub-chronic toxicity and chronic toxicity based on existing animal study data and OUS clinical data. Carcinogenicity testing was omitted considering that the suture is cleared and that there were no relevant concerns with the animal or clinical data reviewed as part of the de novo submission. An overview of the biocompatibility testing performed to evaluate the biocompatibility of Permaseal is provided in Table 1. The Permaseal implant and delivery system were determined to be biocompatible based on these tests.

TestPurposeResults
Delivery System
CytotoxicityMEM Elution Using L-929 MouseFibroblast Cells (GLP)Assessment of biological reactivity ofmammalian cell cultures following incubationwith test device extractsNon-cytotoxic
SensitizationGuinea Pig MaximizationSensitization Test (GLP)Determine the potential for the test deviceextract to elicit contact dermal allergenicityNon-sensitizing
IrritationIntracutaneous Irritation Test(GLP)Assess potential of the device to produceirritation following a single intradermal injectionof specific extracts prepared from a test deviceNon-irritant
HemolysisHemolysis Assay - Direct ContactMethod (GLP)Evaluate the hemolytic potential of test articlesNon-hemolytic
Systemic ToxicityAcute Systemic Injection Test(GLP)Screen test article extracts or solutions forpotential toxic effects as a result of a single-dosesystemic injection in mice.Non-toxic
PyrogenicityMaterials Mediated RabbitPyrogen (USP and GLP)Evaluate the test device extract for leachates thathave the potential to induce material-mediatedpyrogenicity following a single dose injectionNot pyrogenic
Implant
Cytotoxicity (In VitroCytotoxicity Study)Evaluate an extract of a test article forcytotoxicity to mammalian cells in cultureNon-cytotoxic
Sensitization (MaximizationSensitization Study)Evaluate the potential of test article to causedelayed dermal contact sensitizationNon-sensitizing

Table 1. Biocompatibility Tests for Permaseal Delivery System and Implant

{5}------------------------------------------------

TestPurposeResults
Irritation (IntracutaneousReactivity Study)Evaluate the local dermal irritation of a testarticle extract following intracutaneous injectionin rabbitsNon-irritant
Acute Systemic Toxicity(Systemic Toxicity Study)Evaluate acute systemic toxicity of test articleextract following injection in miceNon-toxic
Pyrogenicity (Material Mediated(USP and GLP))Determine whether an extract of the test articleinduced a pyrogenic response followingintravenous injection in rabbitsNon-pyrogenic
Hemocompatibility (Hemolysis –Direct (GLP))Evaluate the potential to cause hemolysisNon-hemolytic
Genotoxicity (AMES AssayBacterial Reverse Mutation)Evaluate the mutagenic potential of the devicetest article by measuring its ability to induceDNA reverse mutations in S. typhimurium andE. coli in the presence and absence ofmicrosomal enzymesNon-mutagenic

SHELF LIFE/STERILITY

Packaging validation, sterilization validation and shelf life testing were performed to evaluate the Permaseal device, as summarized in Table 2 below. The Permaseal device was determined to have a 1-year shelf life, based on the real time aging.

Table 2. Packaging Validation, Sterilization Validation and Shelf Life Testing Overview for the

Permaseal
TestPurposeAcceptance CriteriaResults
PackagingValidationProduce objective evidencethat the package obtainedusing the design for theindividual package ofMicro InterventionalDevices (MID) meets therequirements of MID, themanufacturer's drawingsand product specificationswhen the package is made.Seal Visual Inspection: No burns,channels, voids, pleats or foreignmatterPeel Samples: Minimum peel force 1lb/inBurst Samples: Minimum burstvalues recordedBubble Samples: No seal leaks/sealbubbles acceptedDye Samples: No complete seal dyepenetration acceptedAll tests passed

{6}------------------------------------------------

TestPurposeAcceptance CriteriaResults
SterilizationValidationEvaluate the sterilizationprocedure for thePermaseal device and theability to sterilize thedevice according to the pre-identified criteria, includingachieving a SterilityAssurance Level of 10-6Bioburden: b(4)Bacteriostasis/Fungistasis: NeitheracceptableEO Residual: Total recoverable massb(Bacterial Endotoxin: b(4)EtO Acceptable Limit: b(All tests passed
Shelf LifeEvaluate the impact of oneyear real time aging on thePermaseal deviceVisual and Functional assessments(i.e., successful ex-vivo devicedeployment) at baseline and one yearAll visualinspections andfunctional testspassed
EnvironmentalConditioningand PackageIntegrityEvaluate the Permasealdevices after beingsubjected to a range oftemperatures, humidity andsimulated distributiontestingVerification testing, performancetesting (i.e., successful deploymentin porcine heart, compatibility withsheath, functionality of suture grip,knot pusher, knot and anchoreyelets), packaging verificationtesting, average peel force >1.0lbf/inAll tests passed,any deviationsappropriatelyexplained
PackageIntegrityDemonstrate that thePermaseal devicepackaging will providesufficient protection suchdevice and packageintegrity are notcompromised as a result oftypical package handlingprocessesSeal Peel: No irregularities ordelamination of the seals observed,peel strength greater than 1 lbf/inDye Penetration: No dye penetratesthrough sealAll tests passed
TestPurposeResults
Design VerificationTestingEnsure that the design meets the design inputs asdefined in the Product PerformanceSpecifications for the PermasealAll acceptance criteria met,any deviationsappropriately justified
Device WeightEnsure that the device weight meets specificationAll passed
DimensionalVerificationEnsure that the following measurements meetspecification:- Overall device length, height, width- Insertion tube length, diameter- Anchor overall length- Knot location on sutureAll passed
Color/AppearanceVerificationEnsure that the handle, insertion tube and safetymeet color/ appearance specificationsAll passed
Device OperationVerificationEnsure that the device operates to specificationby evaluating the following:- Trigger safety functionality- Guidewire compatibility- Trigger pull force- Safety disengagement force- Inability to fire with an engaged safety- Anchor deployment depth using thedelivery systemAll passed
ImplantPerformanceVerificationEvaluate if the implant meets the followingcriteria:- Accommodates intended sheath sizes- Suture grip detaches from knot pusher- Knot pusher functionality- Suture knot functionality and slip force- Anchor eyelet functionality undertension and axial pull force- Anchor barb extraction from tissuewithout breakage- Peak anchor insertion forceAll passed
Performance VerificationTestingConduct functional performance verificationtesting in porcine ex vivo heartsAll criteria passed
Anchor Pull OutTestNo anchor pull out from the tissue at a tensioningforce of 1NPassed, no anchor pull out
Leakage TestNo leakage of solution shall occur at a pressureof 180 mmHgPassed, no leakage

PERFORMANCE TESTING - BENCH

The Permaseal device functionality was assessed using two sets of bench testing, including design verification and a performance verification test, as described in Table 3.

{7}------------------------------------------------

Table 3. Bench Testing Overview
-----------------------------------------

The Permaseal also underwent verification of component materials, packaging configuration and labeling verification. These verifications demonstrated the conformance of the packaging, labeling and component material to predetermined acceptance criteria, thus further supporting the device safety profile.

{8}------------------------------------------------

PERFORMANCE TESTING - ANIMAL

The Permaseal was rigorously evaluated under a chronic animal study comparing the Permaseal device to conventional mattress suture in a porcine model at 30 days as summarized in Table 4. The device was also assessed using an investigational analysis, specifically in vivo testing of the Permaseal configuration with the molded polypropylene anchors.

TestPurposeAcceptance CriteriaResults
ChronicAnimal Studyat 30 DaysDemonstrate theperformance of thePermaseal Device(test article) totraditional mattresssuture closureb(4)cardiovascularsurgical applicationsSuccess in facilitating thetransapical access following atransapical puncture of theleft ventricle in a simulatedtranscatheter heart valveimplantation procedureacutely and with respect towound healing chronically(30 day survival) in a porcinemodelb(4)all animalssurvived to the designatedend point with nosignificant adverse clinicalobservations or proceduralcomplications related to thetest devices. Pathologicalchanges were minimal andexpected for this type ofprocedure.
AcuteInvestigationalAnalysis, in vivotesting ofPermasealconfigurationsEvaluate theperformance of thePermaseal in an invivo porcine modelAssessment of performanceof the polypropylene anchors,the interaction between thesuture and the insertion tubeduring deployment, thetensioning process to achievewound closure, theperformance of implantdeployment, passage oftransapical sheath andachieving hemostasisThree separate deploymentswere completed in oneanimal with acceptableresults.

Table 4. Nonclinical Animal Study Testing Overview

SUMMARY OF CLINICAL INFORMATION

The STASIS (Secure Transapical Access and Closure) study, a non-randomized, multi-center, prospective, open-label, observational safety and performance study, was conducted to evaluate the Permaseal device. There were 34 patients undergoing transcatheter aortic valve replacement (TA-TAVR) at five clinical sites in Germany and the Netherlands. Table 5 details the population demographics and baseline characteristics of the patients in STASIS.

{9}------------------------------------------------

Number of subjects enrolled34
Age79.4 +/- 7
Female47.1%
Hypertension88.2%
Coronary artery disease70.6%
Mitral valve disease41.2%
COPD11.8%
Coagulopathy2.9%
Cancer11.8%

Table 5. STASIS Demographics and Baseline Characteristics

The STASIS endpoints were as follows:

Primary Effectiveness Endpoint:

Rate of pulsatile bleeding requiring significant surgical intervention (more than one pledgeted suture) at discharge and at 30-day follow-up. The literature derived performance goal was 15%.

Secondary Effectiveness Endpoint:

Treatment parameters that include procedure time, any reported assessment of ease of use, and the appearance of new hypo- or akinesis at 90 days and 12 months.

Safety Endpoint:

All adverse events (AEs) and serious adverse events (SAEs) occurring during the TA-TAVR procedure and follow-up periods of 30 days, 90 days and 12 months. The safety analysis involved reporting of all SAEs and AEs as well as their severity and relation to the Permaseal device and transapical approach. Valve Academic Research Consortium (VARC) 2 (J Thorac Cardiovasc Surg 2013;145:6-23) definitions were used.

Results

Primary Effectiveness:

The rate of pulsatile bleeding requiring more than one pledgeted suture at hospital discharge and 30-day follow-up was 6.5% (2/31), meeting the primary effectiveness endpoint target. Of the 34 total patients in the STASIS study, only 31 patients were included in the effectiveness analysis. One of the excluded patients was a roll-in, while the remaining two patients were excluded due to protocol violations.

Secondary Effectiveness:

The results of STASIS with respect to secondary endpoints and clinical outcomes are provided in Table 6 and Table 7.

{10}------------------------------------------------

Number of Subjects34
TAVR Procedural Success100%
Mortality (30-days)0.0%
Myocardial Infarction (30-day)0.0%
Stroke (30-day)0.0%
Vascular Complications2.9%
Conversion to Sternotomy0.0%
New Permanent Pacemaker12.1%
Procedure Time (min)86.0 +/- 19.6
Need for Transfusion (>2 units)8.8%
Hospital Stay10.6 +/- 3.7

Table 6. STASIS Procedural Secondary Endpoints and 30-Day Clinical Outcomes

Table 7. STASIS Clinical Outcomes and Secondary Endpoints at 90 Days
Number of Subjects3
Mortality3.1%
Myocardial Infarction0.0%
Stroke0.0%

One patient death due to non-cardiovascular causes subsequent to orthopedic surgery was reported prior to 90-day follow-up. There was no change in wall motion reported for any patient between screening and hospital discharge. At the 90-day follow-up one patient improved from moderate hypokinesis to normal and one improved from moderate to mild hypokinesis. One patient developed mild hypokinesis and one developed moderate hypokinesis. One patient with moderate hypokinesis at screening was reported akinetic at 90-day follow-up.

Safety Endpoints:

There were no deaths, myocardial infarctions or strokes for any patients during the 30-day follow-up period. This favorable safety profile was maintained at 90-days clinical follow-up. One patient died at day 37 due to pulmonary embolism subsequent to a hip replacement surgery. This death was determined to not be related to either the procedure or device.

Thirty-three (33) SAEs were reported during the study, of which two were considered devicerelated. The two device-related SAEs involved setting the pre-tied knot before achieving hemostasis, requiring the addition of extra sutures. These events occurred at one site, early in the study, indicating a need for additional training on the Permaseal IFU, which was implemented and no further device-related SAEs were reported.

LABELING

Labeling provided for the Permaseal includes Instructions for Use and packaging labels. The labeling provided satisfies the requirements of 21 CFR § 801.109 Prescription devices, and

{11}------------------------------------------------

includes information regarding specifications, instructions for use, contraindications, warnings, and cautions, as well as a prescription statement.

Important components of the labeling include:

  • Contraindication to exclude patients with myocardial wall thickness less than 10mm, due l to the risk of myocardial wall perforation; and
  • ー Detailed instructions explaining each step of Permaseal use

RISKS TO HEALTH

Table 8 identifies the risks to health that may be associated with use of the Apical Closure Device and the measures necessary to mitigate these risks.

Identified RiskMitigation Measure
InfectionSterilization ValidationShelf Life TestingLabeling
Adverse Tissue ReactionBiocompatibility EvaluationIn vivo Performance Testing
Bleeding• At ventricular puncture or anchordeployment sitesNon-clinical Performance TestingIn vivo Performance TestingLabeling
Tissue Damage• Apical tearing• Myocardial tearing (local or diffuse)Non-clinical Performance TestingIn vivo Performance TestingLabelingTraining
New Hypokinesia or Akinesis of ApexIn vivo Performance TestingLabeling
Thromboemboli and Full Thickness InjuryIn vivo Performance TestingLabelingTraining
Pericardial TamponadeIn vivo Performance TestingLabeling

Table 8. Identified Risks to Health and Mitigation Measures

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the Apical Closure Device is subject to the following special controls:

    1. The patient contacting materials must be evaluated to be biocompatible.
    1. Performance data must validate the sterility of the patient-contacting components of the device.
    1. Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
    1. Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

{12}------------------------------------------------

  • Consistent and reliable implant deployment; a.
  • b. Assessment of implant pull-out force; and
  • c. Sheath size compatibility with implant.
    1. In vivo evaluation of the device must demonstrate device performance, including device operation resulting in closure of the myocardial wound.
    1. Labeling must include the following:
    • a. Detailed information explaining how the device operates;
    • b. Sheath size that device can accommodate;
    • c. Identification of the minimum myocardial wall thickness to ensure optimal device function; and
    • d. A shelf life.

BENEFIT/RISK DETERMINATION

The risks of the device are based on nonclinical laboratory tests, animal studies, and data collected in a clinical study as described above. In the STASIS clinical study, SAEs occurring in the population included vascular complications (2.9%) and new permanent pacemaker (related to TAVR placement) (12.1%) at 30 days post-procedure. There were no deaths, myocardial infarctions, or strokes for any patients during the 30-day follow-up. Within 90 days postprocedure, the mortality rate was 3.1%. No tearing or psuedoaneurysm of the apex was observed. No myocardial infarctions were observed and apical myocardial function was preserved. Considering the STASIS data, the probability of any adverse event related to the device are extremely low. Any harmful events requiring additional therapy would be expected to be observed within the 90-day STASIS study. There were no additional patient risks associated with the Permaseal over the standard of care, including the risks of bleeding, aneurysm, or pseudoaneurysm.

The probable benefits of the device are also based on nonclinical laboratory, animal studies as well as data collected in a clinical study as described above. The Permaseal provides a minimally invasive reproducible technique for apical closure that distributes tension equally and circumferentially around a trans-apical access point. Apical device closure, defined as freedom from pulsatile bleeding requiring significant surgical intervention (more than one pledgeted suture), was achieved in 93.5% of patients. No patients required re-operation for bleeding at the apical access site and results were durable through 90 days. There was also no death or stroke observed in the 30-day follow-up interval of the STASIS trial.

Additional factors to be considered in determining the probable risks and benefits for the Permaseal include that the placement of the Permaseal device is at least as safe and effective as the current standard of care. The additional probable benefits of the Permaseal device include reduction of apical bleeding complications, the minimization of apical myocardial injury that might result in apical akinesis or dyskinesis, reduction in procedural time, and reduction in procedural bleeding related to apical leak. Furthermore, patients would likely value a secure and less traumatic procedure available with the use of the Permaseal device. It is also important to note that the standard of care, which is closure using manually placed purse-strings or mattress sutures during surgery, is not precluded by the Permaseal if failure were to occur. Finally, the Permaseal is placed epicardially as a cinching mechanism and no device components remain in contact with the bloodstream or intra-ventricular cavity.

{13}------------------------------------------------

Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

Benefit/Risk Conclusion

In conclusion, given the available information above, the data supports that for soft tissue approximation of cardiac apical tissue during transcatheter valve replacement, the probable benefits outweigh the probable risks for the Permaseal device. The device provides substantial benefits and the risks can be mitigated by the use of general and the identified special controls.

CONCLUSION

The de novo request for the Permaseal is granted and the device is classified under the following:

Product Code: PNQ Device Type: Apical Closure Device Class: II Regulation: 21 CFR 870.4510

§ 870.4510 Apical closure device.

(a)
Identification. An apical closure device is a prescription device consisting of a delivery system and implant component that is used for soft tissue approximation of cardiac apical tissue during transcatheter valve replacement procedures.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient contacting materials must be evaluated to be biocompatible.
(2) Performance data must validate the sterility of the patient-contacting components of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
(4) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Consistent and reliable implant deployment;
(ii) Assessment of implant pull-out force; and
(iii) Sheath size compatibility with implant.
(5) In vivo evaluation of the device must demonstrate device performance, including device operation resulting in closure of the myocardial wound.
(6) Labeling must include the following:
(i) Detailed information explaining how the device operates;
(ii) Sheath size that device can accommodate;
(iii) Identification of the minimum myocardial wall thickness to ensure optimal device function; and
(iv) A shelf life.