K Number
DEN130046

Validate with FDA (Live)

Date Cleared
2015-03-06

(477 days)

Product Code
Regulation Number
874.5900
Type
Direct
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reza Band® UES Assist Device is indicated for patients 18 years and older to reduce the symptoms of larvngopharyngeal reflux (LPR) disease by reducing the regurgitation of stomach contents from passing through the upper esophageal sphincter. The device is worn by the patient when sleeping.

Device Description

The Reza Band® UES Assist Device (Reza Band) (Figure 1) is a non-invasive, non-sterile device worn by the patient and is designed to provide a set pressure (20-30 mm Hg) on the cricoid cartilage, which increases the luminal pressure within the upper esophageal sphincter (UES). The patient wears the Reza Band after being fit by the physician (Figure 2).

The Reza Band has 5 main components:

  • Frame
  • Cushion
  • Comfort Band
  • Clasp
  • Comfort Dial

The Reza Band® External Manometer (External Manometer) is a hand-held device and is connected to the Reza Band® Pressure Sensor (Pressure Sensor). It is used by the treating physician, is powered by a AAA battery and displays the pressure being applied to the cricoid cartilage region by the Reza Band in millimeters of Mercury (mm Hg) as the Reza Band® is being fitted (Figure 3).

The Pressure Sensor connected to the External Manometer is regulated as a miniature pressure transducer (21 CFR Part 890.1615) which is Class I exempt. The Pressure Sensor and the External Manometer were assessed to ensure that the Reza Band® was accurately applying the specified pressure.

AI/ML Overview

The Reza Band® Upper Esophageal Sphincter (UES) Assist Device did not undergo a study that assessed its performance against specific, pre-defined acceptance criteria with a quantifiable metric (e.g., accuracy, sensitivity, specificity). Instead, the studies focused on demonstrating the safety and effectiveness of the device.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

As mentioned, there isn't a direct table of "acceptance criteria" where a specific quantitative metric (like accuracy or AUC) is listed alongside a target value and then the device's performance against that target for a classification task. The studies provided focus on demonstrating safety and symptomatic improvement.

Bench Test Acceptance Criteria and Results (from Table 2):

TestAcceptance Criteria (Implicit)Reported Device Performance
Reza Band Neck Ring Lifetime TestRetention of intended functionality over 6-month use period.All established acceptance criteria were met.
Reza Band Mechanical Test- Comfort Band: No significant change in length after pull testing. - Frame Pusher: No damage/failures; Comfort Dial functions as intended after pull testing. - Frame Body: No failure, damage, or deformation after force application.- Comfort Band: Length not significantly different after being pulled. - Frame Pusher: No damage or failures; Comfort Dial continued to function. - Frame Body: Did not fail; no damage noted after being subjected to force.
Reza Band Accelerated Aging Shelf Life TestingRetention of intended functionality over simulated 6-month shelf life.With the exception of observations regarding Frame Body magnets (rationale provided), acceptance criteria were met.
Reza Band, External Manometer and Pressure Sensor Shipping TestWithstand damage and retain intended functionality after simulated shipping conditions.Demonstrated continued functionality and conformance with specifications.
External Manometer Pressure Measurement Accuracy TestingPressure measurement accuracy within design specifications.Within (b)(4) mmHg as specified over the (b)(4) range.
External Manometer Life TestMaintain functionality over its intended 600-use lifetime.All established acceptance criteria were met following (b)(4) (in excess of 600-use requirement).

Clinical Study Effectiveness Endpoint (from Study #1):

Acceptance Criteria (Implicit)Reported Device Performance
Primary Effectiveness Endpoint: Percent reduction in Reflux Symptom Index (RSI) from Baseline to Visit 3 (Week 4)- Mean % change in RSI from Baseline to Visit 3: -54.3% (p<0.0001) - 60.7% of subjects reported RSI score below 13 (suggesting LPR absence) at the last post baseline visit. - Statistically improved RSI at 2 and 4 weeks (p<0.0001 for both). - p-value (one-sided one sample t-test to test hypothesis of mean change >25%): <0.0001 for all post-baseline visits.

2. Sample Size Used for the Test Set and the Data Provenance

  • Study #1 (Clinical Effectiveness Study):

    • Sample Size:
      • Enrolled: 95 patients
      • Analyzed for Effectiveness: 89 patients (who wore the device for at least 2 weeks and provided one post-baseline RSI assessment). 6 subjects discontinued and were excluded from this specific effectiveness analysis, but a worst-case sensitivity analysis confirmed no impact on treatment effect.
    • Data Provenance: Prospective, multi-center (5 investigational sites). Country of origin is not explicitly stated, but implicitly US given the FDA submission context.
  • Study #2 (Safety of Intentionally Displaced Device):

    • Sample Size: 20 subjects
    • Data Provenance: Prospective. Country of origin not explicitly stated, but implicitly US given the FDA submission context.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not explicitly provided in the document for either study.

  • Study #1: The primary effectiveness endpoint, Reflux Symptom Index (RSI), is a "validated nine-item, patient-administered outcome questionnaire." This suggests the ground truth for effectiveness is based on patient-reported outcomes, rather than expert assessment of the RSI score itself. Clinical diagnosis of LPR was made by healthcare professionals, but the evaluation of effectiveness relied on the patient-completed RSI.
  • Study #2: Ground truth for safety was established by direct physiological measurements (heart rate, blood pressure, cardiac rhythm, IOP). These are objective measurements, not requiring expert consensus for ground truth establishment in the dataset.

4. Adjudication Method for the Test Set

Not applicable to these studies as the primary endpoints are either patient-reported (RSI) or direct physiological measurements, not requiring expert adjudication of a "test set" in the context of diagnostic performance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, an MRMC comparative effectiveness study was not done. This device is not an AI/diagnostic imaging device, so this type of study is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. The Reza Band is a physical medical device, not an algorithm. Its performance is tied to its physical application and effect on the patient.

7. The Type of Ground Truth Used

  • Study #1 (Effectiveness): Patient-reported outcomes (Reflux Symptom Index - RSI). The RSI is a validated questionnaire.
  • Study #2 (Safety in Displaced Position): Objective physiological measurements (heart rate, blood pressure, cardiac rhythm, intraocular pressure).

8. The Sample Size for the Training Set

Not applicable. The Reza Band is a physical medical device. It does not utilize machine learning or AI models that require a "training set" in the conventional sense. The "training" of the device itself refers to its design, testing, and refinement, not an algorithmic training process.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" for an AI algorithm.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR REZA BAND® UPPER ESOPHAGEAL SPHINCTER (UES) ASSIST DEVICE

REGULATORY INFORMATION

FDA identifies this generic type of device as:

External Upper Esophageal Sphincter (UES) Compression Device: An external upper esophageal sphincter (UES) compression device is a prescription device used to apply external pressure on the cricoid cartilage for the purpose of reducing the symptoms of laryngopharyngeal reflux (LPR) disease.

NEW REGULATION NUMBER: 874.5900

CLASSIFICATION: CLASS II

PRODUCT CODE: PKA

BACKGROUND

DEVICE NAME: REZA BAND® UPPER ESOPHAGEAL SPHINCTER (UES) ASSIST DEVICE

SUBMISSION NUMBER: DEN130046

DATE OF DE NOVO: NOVEMBER 27, 2013

CONTACT: SOMNA THERAPEUTICS, LLC W175 N11081 Stonewood Drive GERMANTOWN, WI 53022

REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS I, MODIFIED TO CLASS II

INDICATIONS FOR USE

The Reza Band® UES Assist Device is indicated for patients 18 years and older to reduce the symptoms of larvngopharyngeal reflux (LPR) disease by reducing the regurgitation of stomach contents from passing through the upper esophageal sphincter. The device is worn by the patient when sleeping.

LIMITATIONS

Prescription Use only: Federal (USA) law restricts this device to sale by or on the order of a physician.

Limitations on device use are included in the Instructions for Use as Warnings and Precautions.

{1}------------------------------------------------

Warnings

  • The safety and effectiveness has not been demonstrated for the following conditions:
    • Patient with implants or implant parts that reside in the area where the Reza o Band is applied.
    • Patient with an implanted pacemaker, implanted cardioverter defibrillator o (ICD), vagus nerve stimulator, or other such similar devices implanted in the neck.
    • Patient diagnosed with glaucoma. O
    • Patient had a malignancy of the neck, including neck surgery. O
    • Patients that may have an altered mental status including due to the use of O sedative drugs or narcotics.
    • Patients with carotid artery disease, thyroid disease, a history of o cerebrovascular disease, or any disorder of connective tissues (e.g., Marfan's Syndrome or Ehlhers-Danlos Syndrome).
  • The Reza Band is packaged and designed for single patient use only.
  • The Pressure Sensor is packaged and designed for a one-time use only.

Precautions

  • The Reza Band, the External Manometer or the Pressure Sensor should only be ● used by health care providers trained to prescribe and apply the device.
  • The safety of the Reza Band for use during pregnancy, in breastfeeding females or in patients under 18 years of age has not been established.
  • The safety of the Reza Band® UES Assist Device in patients currently receiving ● treatment with continuous positive airway pressure (CPAP) has not been established.
  • If the patient feels unusual pain, skin irritation or the worsening of symptoms, they should discontinue the use of the Reza Band and contact their health care provider.
  • . If the Reza Band does not stay in place, the patient should stop using the Reza Band and contact their health care provider. The patient may need to be re-fitted for the Reza Band.
  • The External Manometer is not designed for liquid immersion. The External . Manometer should be cleaned with a damp cloth using warm water or presaturated isopropyl alcohol wipes.

DEVICE DESCRIPTION

Reza Band® UES Assist Device

The Reza Band® UES Assist Device (Reza Band) (Figure 1) is a non-invasive, non-sterile device worn by the patient and is designed to provide a set pressure (20-30 mm Hg) on the cricoid cartilage, which increases the luminal pressure within the upper esophageal sphincter (UES). The patient wears the Reza Band after being fit by the physician (Figure 2).

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a headband with a dark blue band and a white plastic frame. The headband appears to be designed for comfort and stability, with the soft band providing cushioning against the head. The white plastic frame likely provides structure and support, possibly housing electronic components or sensors. The headband is likely used for sports, fitness, or medical purposes.

Figure 1: Reza Band® UES Assist Device

Figure 2: Fitting the Reza Band® UES Assist Device to the Patient

The Reza Band has 5 main components:

  • Frame Is semi-flexible so that it can apply the prescribed pressure as determined by the ● physician, is comfortable for the patient, and allows limited pressure adjustments by the patient. The Cushion is attached to the Frame. The magnet in the Frame attaches to the Clasp.
  • . Cushion - Positioned over the cricoid cartilage and designed to conform to the anatomical structures when being worn.
  • Comfort Band - Holds the Reza Band in place.
  • Clasp - Attaches to the Reza Band Frame by the use of a magnet. This allows for the patient to easily and quickly take the device on and off during routine use and in emergency situations if needed.
  • . Comfort Dial - Allows the patient a limited adjustment to the applied pressure.

Reza Band® Pressure Sensor and the Reza Band® External Manometer

The Reza Band® External Manometer (External Manometer) is a hand-held device and is connected to the Reza Band® Pressure Sensor (Pressure Sensor). It is used by the treating physician, is powered by a AAA battery and displays the pressure being applied to the cricoid cartilage region by the Reza Band in millimeters of Mercury (mm Hg) as the Reza Band® is being fitted (Figure 3).

The Pressure Sensor connected to the External Manometer is regulated as a miniature pressure transducer (21 CFR Part 890.1615) which is Class I exempt. The Pressure Sensor and the

{3}------------------------------------------------

External Manometer were assessed to ensure that the Reza Band® was accurately applying the specified pressure.

Image /page/3/Picture/1 description: The image shows a medical device with a pressure sensor assembly and an external manometer. The pressure sensor assembly is a blue, bulbous component connected to a clear tube. The tube leads to a handheld external manometer, which has a digital display showing "0.0" and a blue button.

Figure 3: Reza Band® Pressure Sensor Connected to the Reza Band® External Manometer

  • . External Manometer – Contains the "On" button and displays the pressure being applied by the Reza Band in mmHg.
  • Pressure Sensor – A non-sterile, one-time-use component that is connected to the External Manometer with a luer lock connector. It is positioned between the Reza Band and the patient, at the time of the fitting by the physician. When the Pressure Sensor is compressed during the fitting of the Reza Band, air is forced down the tubing to an internal pressure sensor. This, in turn, provides the input for the display.

SUMMARY OF NONCLINICAL/BENCH STUDIES

Non-clinical/bench studies conducted on the Reza Band® UES Assist Device to demonstrate a reasonable assurance of safety and effectiveness of the device are summarized in the sections below.

BIOCOMPATIBILITY/MATERIALS

The Reza Band contacts the patient's intact skin surfaces. The Reza Band is categorized as a permanent contact device as the device is intended to be used more than 30 days. In accordance with ISO 10993-1: Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing within a Risk Management Process, the following biocompatibility testing was conducted on the Reza Band. The biocompatibility assessment was deemed adequate.

Table 1: Reza Band® UES Assist Device Biocompatibility Testing

Company of Concession Comparis and Comments of Comments of Concession Comer College of Concession Come of Children Come of Children Come of Children Comer Collection Comer CoC month a mallently and a long of comments of the finance of the finance of the final
Cﺎﺭﮦ ﮐﺎ ﺭﺑﻂ ﻭﯾﮑﯽ ﮈﯾﭩﺎ ﺳﮯ ﺑﻠﻨﺪﯼ ﭘﺮ ﻭﺍﻗﻊ ﮨﮯ۔Purpose1997 3TAUDULU

{4}------------------------------------------------

Cytotoxicity Evaluation –Fluid Extract/L929 MouseFibroblastTo assess the biological activity of L-929mouse fibroblast cells (grown in culture) afterexposure to extracts prepared from thecompleted Reza Band deviceNon-cytotoxic
Primary Skin Irritationwith 0.9% NaCl &Cottonseed Oil ExtractsTo assess the irritating potential of extracts ofthe Reza Band to produce dermal irritationNegligible
Dermal SensitizationMaximization with 0.9%NaCl & Cottonseed OilTo estimate the potential for sensitization ofthe Reza Band extract using the guinea pig asthe animal modelNon-Sensitizer

SHELF LIFE/STERILITY

Reza Band® UES Assist Device is provided non-sterile. The Reza Band and External Manometer components are reusable and intended to be used for up to 6 months. The Pressure Sensor component is for single use. The sponsor provided reprocessing instructions (i.e., cleaning with mild soap and warm water) for the Reza Band and External Manometer for the patient and health care facility users, respectively. The information provided is adequate.

PERFORMANCE TESTING - BENCH

The sponsor conducted finite element analysis (FEA) of the structural behavior to support the design of the Reza Band UES Assist Device. Bench testing was also conducted to demonstrate the mechanical performance of the Reza Band for the intended conditions of use, and compatibility of use with the External Manometer, and Pressure Sensor. The functionality of the devices was tested over a simulated 6-month shelf life period. All bench testing was conducted on the final, finished devices that met established specifications. Table 2 summarizes each of the bench tests.

TestPurposeMethodResults
Reza Band NeckRing LifetimeTestTo demonstratethat the RezaBand retains itsintendedfunctionality overits intended 6-month use period(b)(4)All the establishedacceptance criteria weremet.
functionality, in accordancewith establishedspecifications.
Reza BandMechanical TestTo demonstratethe mechanicalintegrity of threekey componentsof the Reza Bandfor theirintended use:tensile strengthof the ComfortBand;functionality andattachment ofthe frame pusherafter exposure topull testing; andattachment ofthe ComfortBand attachmentportions of theframe body afterpull testingComfort Band Evaluation:The Comfort Band was(b)(4)thelength was (b)(4)testingThe length of the ComfortBand was not significantlydifferent after being pulledby the (b)(4)when comparing thelength before and after thetesting.
Frame Pusher Evaluation:The Frame Pusher was(b)(4)theComfort Dial actuation.The Frame Pusher did notexhibit any damage orfailures and the ComfortDial continued to functionas intended (b)(4)was appliedto the Frame Pusher.
Frame Body Evaluation:The Frame Body wasapplied (b)(4)visuallyexamined for evidence ofbreakage, damage ordeformation.The Frame Body did notfail and there was nodamage noted, after beingsubjected (b)(4)of force.
Reza BandAcceleratedAging Shelf LifeTestingTo demonstratethat the RezaBand retains itsintendedfunctionality overa simulated 6-month shelf life(storage) periodDevices were subjected to(b)(4)andevaluated for continuedfunctionality andconformance withspecifications.With the exception of theobservations regarding theFrame Body magnetsfollowing exposure toaccelerated agingconditions (b)(4)acceptance criteria weremet.A rationale (b)(4)
Reza Band,ExternalManometer andPressure SensorShipping TestTo demonstratethat the packagedReza Band,ExternalManometer, andPressure Sensorwithstanddamage fromDevices were subjected tosimulated shippingconditions includingtemperature, humidityconditions, compression,vibration, shock, and repeatvibration in accordance withThe Reza Band, ExternalManometer and PressureSensor demonstratedcontinued functionality andconformance withspecifications.
and retain theirintendedfunctionalityafter, simulatedshippingconditions(b)(4)
ExternalManometerPressureMeasurementAccuracy TestingTo demonstratethat the pressuremeasurementaccuracy of theExternalManometermeets designspecificationsKnown pressures (b)(4)weremeasured by the ExternalManometer.The pressure measurementaccuracy of the ExternalManometer is within (b)(4)mmHg as specified overthe (b)(4)
ExternalManometer LifeTestTo demonstratethat the ExternalManometer iscapable ofmaintainingfunctionalityover its intended600-use life timePressure measurementaccuracy was evaluated atbaseline (b)(4)withpressure accuracy assessedat from (b)(4) mmHg.All established acceptancecriteria were met following(b)(4) (in excess of the600-use requirement)

Table 2. Reza Band® UES Assist Device Bench Testing

{5}------------------------------------------------

{6}------------------------------------------------

SUMMARY OF CLINICAL INFORMATION

The following is a summary of two clinical studies performed by the sponsor to support a reasonable assurance of safety and effectiveness for the Reza Band® UES Assist Device.

Study #1: Multi-Center, Non-Randomized, Prospective Study of the Reza Band® Upper Esophageal Sphincter (UES) Assist Device for the Treatment of Esophagopharyngeal Reflux

The sponsor conducted a non-randomized, prospective, open label study of 95 patients treated at 5 investigational sites with the Reza Band. The study enrolled subjects that had been clinically diagnosed with laryngopharyngeal reflux (LPR) (i.e., chronic cough, choking, aspiration, chronic post nasal drip, globus, sore throat, throat clearing) and who met the inclusion and exclusion criteria.

The inclusion criteria were:

  • . 18 years of age or older
  • . The patient must be willing and able to provide informed consent.
  • Understands the clinical study requirements and is able to comply with follow-up schedule. .
  • . Clinically diagnosed with esophagopharyngeal reflux with extra-esophageal symptoms (i.e., chronic cough, choking, aspiration, chronic post nasal drip, globus, sore throat, throat clearing) for 6 months or longer.
  • . Reflux Symptom Index (RSI) >13

{7}------------------------------------------------

The exclusion criteria were:

  • Currently being treated with another investigational medical device and/or drug
  • . Currently receiving treatment for sleep apnea with continuous positive airway pressure (CPAP)
  • The patient is female and is of child bearing potential and is not using an acceptable method of birth control, or is pregnant or breast feeding.
  • Previous head or neck surgery or radiation
  • Carotid artery disease, thyroid disease, or history of cerebral vascular disease
  • Suspected esophageal cancer
  • Nasopharyngeal cancer
  • Has either a pacemaker or implanted cardioverter defibrillator (ICD)
  • Previously undergone Nissen Fundoplication

The study population consisted of 30.5% males and 69.5% females, (81.1% Caucasian, 8.4% Hispanic, 4.2% African-American, 1.1% Asian, 1.1% other subjects, and 4.2% not reported), with a mean age of 48.8 years, a mean body weight of 160.0 pounds, and a mean Body Mass Index (BMI) of 25.5.

The objective of the study was to assess the safety and effectiveness of the Reza Band when worn by subjects that had been clinically diagnosed with LPR (i.e., chronic cough, choking, aspiration, chronic post nasal drip, globus, sore throat, throat clearing). The subjects applied the Reza Band at bedtime and took it off upon waking.

The safety of the Reza Band was evaluated by assessing the incidence, type, duration and severity of adverse events observed in all patients.

The primary effectiveness endpoint was the percent reduction in the Reflux Symptom Index (RSI) from Baseline to Visit 3 (Week 4). The RSI is a validated nine-item, patient-administered outcome questionnaire designed to document the symptoms and severity of LPR, Patients were instructed to rate how the nine symptoms affected them on a scale of 0 (no problem) to 5 (severe problem), with a maximum total score of 45. Effectiveness was also assessed by the evaluation of patient and physician satisfaction, patient response to the RSI, the SF-36° Health Survey, the Functional Outcomes of Sleep Ouestionnaire and patient diaries.

Safety Results:

The adverse events that were reported during the study were generally mild, short in duration and the majority of those events were not related to the device. Those events that were reported as related to the device were also generally mild and short in duration. Device-related adverse events did not result in reduced outcomes as assessed by the change of the RSI score from baseline to Visit 3 (Week 4), as they were consistent with the overall population. There were no deaths or unexpected adverse events in the study. Six of ninety five (6/95) subjects discontinued use of the device due to reported symptoms of discomfort, and/or difficulty sleeping with the device in place, and therefore did not complete the study (categorized under Reza Band Problem in table below).

{8}------------------------------------------------

Table 3 below summarizes the incidence of adverse events in all subjects (including the 6 who did not complete the study) along with the reported relationship of the adverse event to the device.

Definitely(n/%)PossiblyProbably NotDefinitely Not
Dysphagia0 (0.0%)0 (0.0%)1 (1.1%)0 (0.0%)
Odynophagia0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Pain5 (5.3%)2 (2.1%)1 (1.1%)0 (0.0%)
Globus1 (1.1%)0 (0.0%)1 (1.1%)5 (5.3%)
Laryngospasm0 (0.0%)0 (0.0%)0 (0.0%)1 (1.1%)
Regurgitation0 (0.0%)0 (0.0%)1 (1.1%)2 (2.1%)
Choking5 (5.3%)2 (2.1%)1 (1.1%)2 (2.1%)
Nausea1 (1.1%)0 (0.0%)2 (2.1%)0 (0.0%)
Dyspepsia0 (0.0%)1 (1.1%)1 (1.1%)1 (1.1%)
Hoarseness0 (0.0%)2 (2.1%)3 (3.2%)3 (3.2%)
Cough0 (0.0%)2 (2.1%)2 (2.1%)7 (7.4%)
Difficulty Breathing0 (0.0%)1 (1.1%)3 (3.2%)5 (5.3%)
Unable to Belch0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Unable to Vomit0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Skin Reaction5 (5.3%)1 (1.1%)2 (2.1%)1 (1.1%)
Reza Band Problem14 (14.7%)0 (0.0%)0 (0.0%)1 (1.1%)
External Manometer Problem1 (1.1%)0 (0.0%)0 (0.0%)0 (0.0%)

Table 3. Adverse events relationship to device through 1 month follow-up (N = 95)

Effectiveness Results:

The Primary Effectiveness endpoint was defined as the percent change in the RSI from Baseline (Visit 1) to End of Study (4 weeks). Table 4 below provides the percent change in the RSI from the baseline for Visit 2 (2 weeks) and Visit 3 (4 weeks) for the 89 of 95 (93.7%) subjects that wore the Reza Band® for at least 2 weeks and provided one post-baseline RSI assessment. The six (6) subjects that discontinued use of device were excluded from the analysis as they did not return for Visit 2 (2 week) or Visit 3 (4 week). A worst-case sensitivity analysis demonstrated that the exclusion of the 6 subjects who did not return has no impact on the treatment effect or the outcome of the study.

The change in the RSI from baseline to each post treatment visit, as well as any post-treatment visit, were found to be statistically improved for both Visit 2 and Visit 3 (p<0.0001).

Table 4. Reflux Symptom Index (RSI) percent change in RSI from Visit 1 (N = 89)

{9}------------------------------------------------

Time PointStatisticRSI% Change fromVisit 1 - Baseline
Visit 1n89
Mean26.0
Standard Deviation6.8N/A
Median25.0
Minimum14.0
Maximum42.0
Visit 2n8686
Mean14.6-43.8
Standard Deviation9.231.1
Median12.5-50.0
Minimum0.0-100.0
Maximum41.035.7
p-Value (1)<0.0001
p-Value (2)<0.0001
Visit 3n8686
Mean12.0-54.3
Standard Deviation8.6-27.1
Median10.5-57.8
Minimum0.0-100.0
Maximum42.07.7
p-Value (1)<0.0001
p-Value (2)<0.0001
Last Post-Baseline Visitn8989
Mean12.1-54.2
Standard Deviation8.627.5
Median10.0-57.9
Minimum0.0-100.0
Maximum42.07.7
p-Value (1)<0.0001
p-Value (2)<0.0001

(1) p-Value based on paired t-test to test hypothesis of no mean percent change.

(2) p-value based on one-sided one sample t-test to test hypothesis of mean change >25%.

All study subjects had an RSI score of greater than 13 at baseline. In the literature, an RSI score of greater than 13 suggests the presence of LPR . In this study, 54 out of 89 subjects (60.7%) reported that their RSI score was below 13 at the last post baseline visit.

1 Belafsky PC, Postma GN, Koufman JA. Validity and reliability of the reflux symptom index (RSI). J Voice. 2002 Jun;16(2):274-7.

Of the 95 subjects that were enrolled in the study, 79 (83.2%) reported taking acid reducing medications with 16 subjects (16.8%) not taking acid reducing medications. Seventy-five of the 89 subjects (84.3%) that completed the study reported taking acid reducing medications at the time of enrollment. Of these 75 subjects, 61 (81.3%) continued taking acid reducing medication during the study and 14 (18.7%) discontinued taking acid reducing medications during the study.

{10}------------------------------------------------

The analysis in Table 5 below shows that the percent change in RSI score from baseline is no different whether acid reducing medications were used during the study. The study was not intended to show nor did the analysis of the data show that the Reza Band® provides a significant reduction in use of acid reducing medication.

Enrollment2 Week4 Week4 Week% Change
MedsNo MedsMedsNo MedsMedsNo MedsMedsNo Meds
Mean25.826.214.116.611.712.855.7%51.7%
Std Dev6.67.79.19.38.510.127.6%29.8%
Median252712159.511.559.4%56.2%

Table 5. RSI scores subjects taking acid reducing medications at enrollment (N = 79) compared to subjects not taking acid reducing medications at enrollment (N = 16)

Study #2: Safety of an Intentionally Displaced Reza Band® Upper Esophageal Sphincter (UES) Assist Device

Objective

The safety of the Reza Band® UES Assist Device was evaluated when worn as intended over the cricoid cartilage, as well as when intentionally displaced laterally over each side of the neck, based on changes in heart rate, blood pressure, cardiac rhythm and intraocular pressure (IOP), as compared to when the Reza Band was not being worn. The intentional lateral displacement was studied to evaluate potential safety issues that might occur if the device was unintentionally displaced at night during sleep.

Methods

Twenty (20) subjects who met the inclusion and exclusion criteria were enrolled in the study, and were representative of the intended Reza Band patient population. The inclusion and exclusion criteria were similar with those specified in the pivotal clinical study (Study #1), but in this study patients with a history of glaucoma were excluded. The study population consisted of 60% males and 40% females (95% Caucasian and 5% African-American subjects), with a mean age of 46.3 years, and a mean Body Mass Index (BMI) of 28.7.

Measurements of heart rate, blood pressure, cardiac rhythm and IOP were taken at baseline, immediately after the Reza Band was placed on the neck in each of the specified scenarios (cricoid, right displacement and left displacement). 5 minutes after placement (except for IOP) and again 15 minutes after placement. The Reza Band was removed and 3 minutes later, IOP was again measured in each placement scenario.

After the fitting of the Reza Band, the Comfort Dial was fully actuated until the Comfort Dial came to the stop, to achieve the maximum pressure. The maximum pressure with the Comfort Dial fully actuated was also measured with the Reza Band displaced laterally over the vascular structures on both sides of the neck.

Results

The results showed that there was no effect on heart rate, blood pressure, cardiac rhythm or IOP. when the Reza Band was worn as intended as well as when the Reza Band was intentionally

{11}------------------------------------------------

displaced laterally, as compared to the baseline. No adverse events were reported during this study.

Analyses of Variance (One Way) were conducted on heart rate, blood pressure and intraocular pressure to determine whether there was any effect of the Reza Band being placed as intended or laterally displaced, when compared to not wearing the Reza Band. There was no significant difference for any of the measurements at any point in time, when compared to the baseline values.

Additionally, electrocardiograms (ECGs) were evaluated to determine whether there was any impact on cardiac rhythm when the Reza Band was placed as intentionally displaced over the jugular and carotid vasculature. It was determined that there was no change in the ECGs for any of the subjects at any time during the study, when compared to the baseline ECG.

The following summary table (Table 6) shows that when the Reza Band is set to apply its maximal pressure with the Comfort Dial fully actuated, and then displaced laterally, the pressure being applied to the left and right sides of the neck by the Reza Band is significantly reduced (mean of 59.2% (13.3 mm Hg), when displaced laterally to the left and a mean of 55.2% (14.6 mm Hg), when displaced laterally to the right), as compared to the applied pressure over the cricoid (20.9 mm Hg). This further reduces the risk of increased pressure to the sensitive structures of the neck.

VariableCricoid Pressure(Comfort DialOFF)(mmHg)Cricoid Pressure(Comfort DialMAXIMUM)(mmHg)Displaced to Left(Comfort DialMAXIMUM)(mmHg)Displaced to Right(Comfort DialMAXIMUM)(mmHg)
Mean20.932.613.314.6
Standard Deviation1.43.34.94.0
Median20341215
Range20-2426-397-226-22

Table 6. Comparison of Reza Band Applied Pressure at Cricoid and Laterally Displaced

Conclusion

In the event that the Reza Band is unintentionally displaced laterally during sleep, the likelihood of impacting sensitive structures of the neck is low. There was no effect on heart rate, blood pressure, cardiac rhythm, or intraocular pressure, either when the Reza Band is placed as intended over the cricoid cartilage, or when it is laterally displaced over the jugular and carotid vasculature. Based on the results of this study, the Reza Band is safe, when worn as intended and if it were to be unintentionally displaced laterally.

LABELING:

The sponsor provided labeling information which includes Physician Instructions for Use, Patient Information Handbook, and package labels for the Reza Band, External Manometer, and Pressure Sensor.

{12}------------------------------------------------

The labeling is sufficient and satisfies the requirements of 21 CFR 801.109 Prescription devices. The patient labeling also follows the principles identified in FDA's guidance entitled "Medical Device Patient Labeling" (April 2001).

RISKS TO HEALTH:

Table 7 below identifies the risks to health that may be associated with use of External UES Compression Device type and the measures necessary to mitigate these risks.

Identified RiskMitigation Method
Adverse tissue reaction- Biocompatibility assessment
Risk of over-compression- Clinical study- Labeling- Technical specifications
Device misuse/ incorrect fitting/ malfunctions- Technical specifications- Clinical study- Labeling- Performance testing(mechanical integrity and shelflife testing)

Table 7: Identified Risks to Health and Mitigation Methods

SPECIAL CONTROLS:

In combination with the general controls of the FD&C Act, the External UES Compression Device is subject to the following special controls:

    1. The patient contacting components must be demonstrated to be biocompatible.
    1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
    • (A) Mechanical integrity testing (e.g., tensile strength testing, fatigue testing)
    • (B) Shelf life testing
    1. The technical specifications must include pressure measurement accuracy to characterize device performance.
    1. Clinical performance testing must document any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.
    1. Labeling must include the following:
    • (A) Appropriate warnings and precautions.
    • (B) A detailed summary of the clinical testing pertinent to use of the device including a detailed summary of the device-related complications or adverse events.

{13}------------------------------------------------

  • (C) Detailed instructions on how to fit the device to the patient.
  • (D) Instructions for reprocessing of any reusable components.
    1. Patient labeling must be provided and must include:
    • (A) Relevant warnings, precautions, and adverse effects/complications.
    • (B) Information on how to correctly wear the device.
    • (C) The potential risks and benefits associated with the use of the device.
    • (D) Alternative treatments.
    • (E) Reprocessing instructions.

BENEFIT/RISK DETERMINATION

The observed risks of the device are based on data collected in the clinical studies described above. There were no device-related serious adverse events. Device movement was the most common adverse event. Device-related, non-serious adverse events also include patient discomfort including feeling hot with the device in place. Some patients reported disliking the sensation of the device around their neck. However, given that this device is externally worn, non-surgical, and prescribed by a physician, there is a high likelihood that patients would typically experience only non-serious adverse events. Duration of most adverse events would only last as long as the device is worn. Most adverse events would be reversible by removal of the device.

The observed benefits of the device are also based on data collected in clinical studies as described above. The benefit of the device is patient satisfaction. From the patient satisfaction survey at least 75% of the subjects reported satisfaction (extremely satisfied, 16.5%; very satisfied, 38.9%; satisfied, 20.0%). As this study was an open-label, single arm study, it is difficult to characterize the device's actual treatment effect. In post-hoc analysis, the patient demographics and characteristics were evaluated against outcomes. The analyses revealed that there was no difference among those sub-groups. Although this study was not statistically powered to assess benefit across the subpopulations, the study revealed that all patient subgroups did well, and therefore, the results appear to be applicable to the overall intended population. However, the current data only support the device's benefit while the device is worn at night. Once the device is removed, then it no longer provides the external pressure to the UES (i.e., there is likely no residual effect).

Additional factors that were considered in determining probable risks and benefits for the Reza Band UES Assist Device include:

  • . The sponsor conducted additional studies using their prototype (early version) device to demonstrate the reduction in LPR by extra-esophageal pH, impedance, and direct endoscopic videography to support preliminary effectiveness of this device.
  • . In the survey conducted by the sponsor as part of the pivotal study, the following satisfaction scores were noted: 16.5% of the subjects were extremely satisfied, 38.9% were very satisfied, 20.0% were satisfied, and 9.4% were somewhat satisfied; 5.9% of the subjects were dissatisfied, 8.2% were very dissatisfied, and 1.2% were extremely dissatisfied.

{14}------------------------------------------------

  • . Considering that the device can be customized, removable, and non-implantable, the patients would tolerate even a smaller benefit given the low potential for risks.
  • LPR is a chronic disease/condition. The disease is usually managed by strict adherence to behavioral life style modification, and/or oral medication (PPI twice a day for a minimum of 6 months). If LPR becomes life-threatening, surgical therapy of fundoplication (either open or laparoscopic) is indicated.
  • Proper application of the Reza Band to the neck and detailed instructions for the physician as well as for the patients help to mitigate the risks. The physician instruction includes clear indications, warnings and precautions, and an appropriate discussion of the use of this device for patients given condition. The patient instruction includes appropriate use of the device, clear instructions to identify the potential risks and an understanding to report to their respective physician if the device is not functioning properly or intolerable.
  • . The quickest intervention is for the patient to remove the device if it is causing any difficulty or adverse effects.

In conclusion, given the available information above, the data support that the probable benefits outweigh the probable risks for the Reza Band Upper Esophageal Sphincter Assist Device for reducing the symptoms of LPR in patients 18 years of age and older. The device provides measurable benefit and the risks can be mitigated by the use of general and the identified special controls.

CONCLUSION

The de novo for the Reza Band Upper Esophageal Sphincter Assist Device is granted and the device is classified under the following:

Product Code: PKA Device Type: External Upper Esophageal Sphincter Compression Device Class: II Regulation: 21 CFR 874.5900

§ 874.5900 External upper esophageal sphincter compression device.

(a)
Identification. An external upper esophageal sphincter compression device is a prescription device used to apply external pressure on the cricoid cartilage for the purpose of reducing the symptoms of laryngopharyngeal reflux disease.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient contacting components must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
(i) Mechanical integrity testing (
e.g., tensile strength testing, fatigue testing) and(ii) Shelf life testing.
(3) The technical specifications must include pressure measurement accuracy to characterize device performance.
(4) Clinical performance testing must document any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.
(5) Labeling must include the following:
(i) Appropriate warnings and precautions,
(ii) A detailed summary of the clinical testing pertinent to use of the device including a detailed summary of the device-related complications or adverse events,
(iii) Detailed instructions on how to fit the device to the patient, and
(iv) Instructions for reprocessing of any reusable components.
(6) Patient labeling must be provided and must include:
(i) Relevant warnings, precautions, and adverse effects/complications,
(ii) Information on how to correctly wear the device,
(iii) The potential risks and benefits associated with the use of the device,
(iv) Alternative treatments, and
(v) Reprocessing instructions.