Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133355

    Validate with FDA (Live)

    Date Cleared
    2014-01-29

    (90 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cadwell Flex system is indicated for prescription use to acquire, record, transmit and display physiological and environmental data for electroencephalographic (EEG) and polysomnographic (PSG) ambulatory and/or clinical studies of patients of all ages.

    Device Description

    The Cadwell Flex EEG/PSG system consists of: (1) acquisition hardware that can acquire, record, store, and transfer up to 32 channels of EEG data, 8 active-reference signal, and audio signals from a microphone, and has EzNet-module-compatibility, (2).software_that_allows_a_user to acquire, store, transmit, view and print data, and to create reports based on the data, (3) optional video recording, (4) an optional photic stimulator, and (5) a host electronic device (typically a PC) capable of running the software and interfacing with the acquisition, video recording, and photic stimulation devices.

    The Cadwell Flex EEG/PSG system is capable of two modes of operation: (1) Ambulatory, where the acquisition hardware acquires, records, and stores physiological and/or environmental data to onboard mass-storage non-volatile memory, and said data is later transferred to a host electronic device, and (2) Clinical, where the host electronic device is connected to the acquisition hardware during data acquisition and recording, with mass-storage occurring on the host.

    The Cadwell Flex EEG/PSG system software is capable of interfacing with additional software used to perform EEG/PSG waveform analysis.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Cadwell Flex device, focusing on acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text for the Cadwell Flex (K133356) does not explicitly state quantitative "acceptance criteria" for its performance in the typical sense of a medical device study (e.g., specific sensitivity, specificity, accuracy thresholds). Instead, the submission focuses on demonstrating substantial equivalence to predicate devices. This means that the device is deemed acceptable if it performs comparably to or better than existing, legally marketed devices.

    The "performance" is largely described by comparing its technical specifications and capabilities with those of its predicate devices.

    CharacteristicPredicate 1: EBNeuro BEmicro (K093728)Predicate 2: Kilowin (K97214)Cadwell Flex (Applicant Device)
    Device classClass IIClass IIClass II
    Target PopulationPediatric through adultPatients of all agesPatients of all ages
    Recording ModalityAttended and unattendedAttendedAttended or unattended
    Use EnvironmentHospital and homeOperating room or clinicHospital or home
    Prescription StatusAvailable only on the order of a physicianBy prescription onlyBy prescription only
    PowerBattery powered or USB poweredAC powerBattery powered or USB powered
    Number of EEG Channels2124-6432
    Other Inputs1 digital12 ECG and 1-16 EMG8 active/reference, 2 other
    A/D Resolution16 bit16 bit16 bit
    Sampling RateUp to 2048 samples/s3200 Hz3200 Hz
    Impedance CheckYesYesYes

    Conclusion regarding performance: The text states, "The Cadwell Flex system is substantially equivalent to [The Cativen Tres] in design, use, technology, and function." This implies that its performance is considered acceptable because it meets or exceeds the capabilities of the predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a clinical study with a "test set" in the context of diagnostic accuracy, sensitivity, or specificity. Instead, the testing appears to be primarily engineering verification and validation.

    The relevant section states: "Software and hardware verification and validation, and electromagnetic compatibility and safety testing were performed."

    • Sample Size for Test Set: Not applicable in the context of a typical clinical performance study. The device's functionality was evaluated through engineering tests, likely involving a single or limited number of prototype devices and simulated or real-world signal inputs to verify its specifications.
    • Data Provenance: The document does not specify the origin of any data used for performance testing beyond the statement of "software and hardware verification and validation." It implies internal company testing rather than external clinical data. It is not retrospective or prospective clinical data in the usual sense.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. As there's no clinical "test set" for diagnostic performance being evaluated against a ground truth, there's no mention of experts establishing a ground truth. The evaluation focused on technical specifications and safety through engineering tests, not clinical interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical test set requiring adjudication of findings by experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document makes no mention of an MRMC study or any comparison of human readers' performance with and without AI assistance. The device is an EEG/PSG acquisition system, not an AI-driven interpretation tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No. The Cadwell Flex is an EEG/PSG system designed to acquire and display physiological data for human interpretation. It is not an algorithm performing standalone diagnostic tasks. The "software" mentioned allows the user "to acquire, store, transmit, view and print data, and to create reports based on the data," and also "is capable of interfacing with additional software used to perform EEG/PSG waveform analysis," but the core device itself is not presented as an AI-driven interpretation tool.

    7. Type of Ground Truth Used

    Not applicable in the context of diagnostic accuracy. The "ground truth" for the engineering tests would be the established technical specifications, safety standards, and electromagnetic compatibility requirements.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that is "trained" on a dataset. It's a hardware and software system for data acquisition and display.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI/ML algorithm, no ground truth was established for this purpose.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1