Search Results
Found 86 results
510(k) Data Aggregation
(220 days)
Image Management and Processing System | Yes, same as predicate |
| Regulation Number | §892.2050 | §892.1170
AutoDensity is a post-processing software intended to estimate spine Bone Mineral Density (BMD) from EOSedge dual energy images for orthopedic pre-surgical assessment applications. It is an opportunistic tool that enables immediate assessment of bone density from EOSedge images acquired for other purposes.
AutoDensity is not intended to replace DXA screening. Suspected low BMD should be confirmed by a DXA exam.
Clinical judgment and experience are required to properly use the software.
Based on EOSedge™ system's images acquired with the dual energy protocols cleared in K233920, AutoDensity software provides an estimate of the Bone Mineral Density (BMD) for L1-L4 in EOSedge AP radiographs of the spine. These values are used to aid in BMD estimation in orthopedic surgical planning workflows to help inform patient assessment and surgical decisions. AutoDensity is opportunistic in nature and provides BMD information with equivalent radiation dose compared to the EOSedge images concurrently acquired and used for general radiographic exams. AutoDensity is not intended to replace DXA screening.
Here's a breakdown of the acceptance criteria and the study details for the AutoDensity device, based on the provided FDA 510(k) clearance letter:
1. Acceptance Criteria and Reported Device Performance
Device Name: AutoDensity
Intended Use: Post-processing software to estimate spine Bone Mineral Density (BMD) from EOSedge dual energy images for orthopedic pre-surgical assessment applications.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Vertebral Level Identification Accuracy | |
| Percent of levels correctly identified ≥ 90% | Testing confirms that the AutoDensity ROI detection algorithm meets performance thresholds. (Specific percentage not provided, but stated to meet criterion). |
| Spine ROI Accuracy (Dice Coefficient) | |
| Lower boundary of 95% CI of mean Dice Coefficient ≥ 0.80 | Testing confirms that the AutoDensity ROI detection algorithm meets performance thresholds. (Specific value not provided, but stated to meet criterion). |
| BMD Precision (Phantom - CV%) | |
| CV% < 1.5% (compared to reference device) | Results met the acceptance criterion (CV% < 1.5%). |
| BMD Agreement (Phantom - max difference) | |
| (Specific numeric criterion not explicitly stated, but implies clinical equivalence to reference device) | Maximum BMD difference of 0.057 g/cm² for the high BMD phantom vertebra, and a difference of < 0.018 g/cm² for clinically relevant BMD range. |
| BMD Precision (Clinical - CV%) | |
| (Specific numeric criterion not explicitly stated, but implies acceptable clinical limits) | AutoDensity precision CV% was 2.23% [95% CI: 1.78%, 2.98%], which is within the range of acceptable clinical limits for the specified pre-surgical orthopedic patient assessment. |
| BMD Agreement (Clinical - Bland-Altman) | |
| (Specific numeric criterion not explicitly stated, but implies equivalence to other commercial bone densitometers) | Bland-Altman bias was 0.045 g/cm², and limits of agreement (LoA) were [-0.088 g/cm², 0.178 g/cm²]. Stated as equivalent to published agreement between other commercial bone densitometers. |
2. Sample Sizes and Data Provenance
Test Set (for ROI Performance Evaluation):
- Sample Size: 129 patients.
- Data Provenance: All cases obtained from EOSedge systems (K233920). The document does not specify the country of origin but mentions a clinical study with 65% US subjects and 35% French subjects for clinical performance testing, which might suggest a similar distribution for the test set, though it's not explicitly stated for the ROI test set. The data was retrospective as it was "obtained from EOSedge systems."
3. Number of Experts and Qualifications for Ground Truth
For ROI Performance Evaluation Test Set:
- Number of Experts: At least 3 (implied by "3 truther majority voting principle") plus one senior US board certified expert radiologist who acted as the gold standard adjudicator.
- Qualifications:
- Two trained technologists (for initial ROI and level identification).
- One senior US board-certified expert radiologist (for supervision, review, selection of most accurate set, and final adjustments).
4. Adjudication Method for the Test Set
For ROI Performance Evaluation Test Set:
- Adjudication Method: A "3 truther majority voting principle" was used, with input from a senior US board-certified expert radiologist (acting as the "gold standard"). The radiologist reviewed results, selected the more accurate set, and made necessary adjustments. This combines elements of majority voting with expert adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No, the provided document does not mention an MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The performance data presented focuses on the standalone performance of the AI algorithm and its agreement/precision with a reference device or clinical measurements.
6. Standalone Performance Study (Algorithm Only)
- Was a standalone study done? Yes. The "Region of Interest (ROI) Performance Evaluation" section explicitly states: "To assess the standalone performance of the AI algorithm of AutoDensity, the test was performed with..." This section details the evaluation of the algorithm's predictions against ground truth for vertebral level identification and spine ROI accuracy.
7. Type of Ground Truth Used
For ROI Performance Evaluation Test Set:
- Type of Ground Truth: Expert consensus with adjudication. Ground truths for ROIs and level identification were established by two trained technologists under the supervision of a senior US board-certified radiologist. The radiologist made the final informed decision, often described as a "gold standard."
8. Sample Size for the Training Set
- Training Set Sample Size: The AI algorithm was trained using 4,679 3D reconstructions and 9,358 corresponding EOS (K152788) or EOSedge (K233920) biplanar 2D X-ray images.
9. How Ground Truth for the Training Set was Established
- The document implies that the training data was "selected to only keep relevant images with the fields of view of interest." However, it does not explicitly detail how the ground truth for the training set was established (e.g., whether it used expert annotations, a similar adjudication process, or other methods). It primarily focuses on the test set ground truth establishment.
Ask a specific question about this device
(156 days)
BARCELONA, 08007
SPAIN
Re: K251022
Trade/Device Name: 3D-SHAPER
Regulation Number: 21 CFR 892.1170
:** 3D Analysis software for Bone Densitometers
DEVICE CLASS: Class II
REGULATION NUMBER: 892.1170
Bone Densitometer | Bone Densitometer | Bone Densitometer | Identical |
| Regulation Number | 21 CFR 892.1170
| 21 CFR 892.1170 | 21 CFR 892.1170 | Identical |
| Product Code | KGI | KGI | KGI | Identical |
| 510
3D-SHAPER uses data from conventional Dual Energy X-Ray Absorptiometry (DXA) scans to measure the distribution of bone mineral mass at the hip and allows the physician to estimate structural properties of the hip, such as cortical sBMD, trabecular vBMD and integral vBMD.
These values can then be compared to an adult reference population at the sole discretion of the physician. The results can assist the physician in the assessment of relative fracture risk.
3D-SHAPER is indicated to be used by qualified medical professionals to assist healthcare professionals in the bone health evaluation & changes, and cannot fully substitute their clinical judgement.
3D-SHAPER is a software reconstructing in 3D, the shape and bone mineral density from 2D DXA images. The software incorporates a DXA-based 3D reconstruction algorithm and measures the distribution of bone mineral mass at the hip. It provides physicians with estimates of the structural properties of the hip, such as cortical sBMD, trabecular vBMD and integral vBMD. The software is not suitable for use with pediatric patients or for subjects with hip prosthesis such as total hip replacement prosthesis, hip resurfacing prosthesis or an osteosynthesis system, since the software has not been tested against these conditions. 3D-SHAPER is indicated to be used by qualified medical professionals to assist healthcare professionals in the bone health evaluation & changes, and cannot fully substitute their clinical judgement.
The 3D-SHAPER measurements include the following parameters calculated at the total femur:
- Cortical surface Bone Mineral Density (Cortical sBMD, mg/cm²)
- Trabecular volumetric Bone Mineral Density (Trabecular vBMD, mg/cm³)
- Integral volumetric Bone Mineral Density (Integral vBMD, mg/cm³), the integral compartment being the union of the cortical and trabecular compartments.
N/A
Ask a specific question about this device
(249 days)
, CA 94107
Re: K242295
Trade/Device Name: BunkerHill BMD
| Regulation Number: 21 CFR 892.1170 |
|---|
| Classification Name |
| Regulation Number |
| Notification** |
| Classification Name |
| Regulation Number |
| Summary** |
| --- |
| Product code |
| Regulation number |
| 21 CFR §892.1170 |
| Modality |
The Bunkerhill BMD Algorithm is a post-processing AI-powered software intended for adults 30 years and above to assess estimated DXA-measured average areal bone mineral density of spinal bones from existing CT scans and outputs a flag for low bone density below a pre-specified threshold. It is not intended to replace DXA or any other tests dedicated to BMD measurement.
Bunkerhill BMD is an opportunistic AI-powered tool that enables:(1) retrospective assessment of bone density from CT scans acquired for other purposes, (2) assessment of bone density in conjunction with another medically appropriate procedure involving CT scans, and (3) assessment of bone density without a phantom as an independent measurement procedure
The Bunkerhill BMD application is a software only medical device (SaMD) that includes deep- learning-based computer vision and post-processing algorithms that estimates the bone mineral density from previously obtained computed tomography (CT) images.
The results from Bunkerhill BMD are not intended to be used as the primary input for clinical decision making, but rather are intended to provide information that may assist the clinician to identify 'findings of interest' within existing imaging studies.
Here's a breakdown of the acceptance criteria and study details for the BunkerHill BMD device, based on the provided FDA 510(k) clearance letter:
Acceptance Criteria and Device Performance Study for BunkerHill BMD
1. Table of Acceptance Criteria and Reported Device Performance
| Metric | Acceptance Criteria | Reported Device Performance (95% Confidence Interval) | Status |
|---|---|---|---|
| Sensitivity | Lower bound > 70% | 81.0% (74.0% - 86.8%) | Pass |
| Specificity | Lower bound > 70% | 78.4% (72.3% - 83.7%) | Pass |
| Pearson Correlation Coefficient | Not explicitly stated, but implicitly supported by "further supporting the robustness and reliability" | 0.791 (0.752–0.830) | N/A (Secondary) |
| AUROC | Not explicitly stated, but implicitly supported by "further supporting the robustness and reliability" | 0.883 (0.849–0.916) | N/A (Secondary) |
| PPV (Positive Predictive Value) | Not explicitly stated, but implicitly supported by "further supporting the robustness and reliability" | 73.6% (66.4%–79.9%) | N/A (Secondary) |
| NPV (Negative Predictive Value) | Not explicitly stated, but implicitly supported by "further supporting the robustness and reliability" | 84.8% (79.0%–89.5%) | N/A (Secondary) |
2. Sample Size and Data Provenance for the Test Set
- Test Set Sample Size: 371 CT studies
- Data Provenance: The studies were collected from four (4) geographically diverse sites. The retrospective nature of the study is explicitly stated ("stand-alone retrospective study").
3. Number of Experts and Qualifications for Ground Truth
The document does not explicitly state the number of experts used to establish the ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience").
4. Adjudication Method for the Test Set
The document does not explicitly state the adjudication method (e.g., 2+1, 3+1, none) used for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not reported in the provided text. The study described is a standalone performance evaluation of the algorithm.
6. Standalone Performance Study
Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The document states: "Bunkerhill BMD performance was validated in a stand-alone retrospective study for overall agreement of the device output compared to the established ground truth."
7. Type of Ground Truth Used
The type of ground truth used is implied to be based on DXA-measured average areal bone mineral density of spinal bones, as the device is intended to "assess estimated DXA-measured average areal bone mineral density." The text refers to "established ground truth" in relation to this assessment.
8. Sample Size for the Training Set
The document does not provide the sample size for the training set. It only describes the test set.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set was established. It only refers to "established ground truth" for the test set evaluation.
Ask a specific question about this device
(105 days)
Plan-les-Ouates, 1228 SWITZERLAND
Re: K243218
Trade/Device Name: TBS iNsight (V4) Regulation Number: 21 CFR 892.1170
----------|-------------------|
| Common Name: | Bone densitometer |
| Regulation Number : | 892.1170
|
| Regulation Number : | 892.1170
| Comparisonto predicatedevice |
| Classification Rule /Name | 21 CFR 892.1170
| 21 CFR 892.1170
TBS iNsight is a software provided for use as a complement to both DXA analysis and clinical examination. It computes the antero-posterior spine DXA examination file and calculates a score (Trabecular Bone Score - TBS) that is compared to those of the age matched controls. The TBS is derived from the texture of the DXA image and has been shown to be related to bone microarchitecture.
TBS iNsight provides as an option an assessment of 10-year fracture risk. It provides an estimate of 10-year probability of hip fracture and 10-year probability of a major osteoporotic fracture (clinical spine, forearm, hip or shoulder fracture). This estimate is based on the WHO's FRAX® Fracture Risk Assessment Tool, after adjustment for the TBS. The tool has been validated for Caucasian and Asian men and post-menopausal women between 40 and 90 years old.
TBS provides information independent of BMD value: it is used as a complement to the data obtained from the DXA analysis and the clinical examination (questioning by the clinician about patient history, bioassay of bone resorption markers...).
The results can be used by a physician in conjunction with other clinical risk factors as an aid in the diagnosis of osteoporosis and other medical conditions leading to altered trabecular bone microarchitecture, and ultimately in the assessment of fracture risk.
The TBS score can assist the health care professional in monitoring the effect of treatments on patients across time. Overall fracture risk will depend on many additional factors that should be considered before making diagnostic or therapeutic recommendations. The software does not disease or recommend treatment regimens. Only the health care professional can make these judgments.
TBS iNsight is a software application provided for use as a complement to bone mineral density (BMD) acquired from dual energy X-ray absorptiometry (DXA) and other clinical risk factors for osteoporosis and fragility fracture. It calculates a score (Trabecular Bone Score - TBS) derived from the texture of the DXA image of the anterior-posterior (AP) lumbar spine and has been shown to be related to bone microarchitecture. The method analyzes X-ray based images acquired by DXA imaging systems and produces the TBS based on the computation of an Adapted Experimental Variogram (modified fractal-like approach). This variogram is used to measure the degree of spatial variation between pairs of data points in a spatial dimension of a region of a digital image.
The absolute TBS values for the same equivalent tissue thickness vary slightly between GE and Hologic systems. Additionally, within each system, variability in TBS values can be observed across different scan modes. To address these differences, corrections are applied that are both device-specific and mode-specific. For instance, TBS is corrected differently for GE and Hologic systems to account for inherent differences in tissue thickness assessment. Furthermore, corrections are also tailored separately for each scan mode within the same system, ensuring that TBS measurements remain consistent and reliable regardless of the scan mode used. These device-specific and mode-specific corrections are necessitated by differences in the dynamic range of tissue thickness measurements between GE and Hologic devices. The variations arise due to differences in the methodologies used to assess tissue thickness. To harmonize these discrepancies and ensure measurement accuracy, correction fits derived from ex-vivo data are applied individually to each device and scan mode. This approach ensures the accuracy and consistency of TBS measurements across all configurations.
The device is intended to be used for bone health assessment in medical facilities employing one or more DXA system(s) to which the subject device is connected. These facilities are usually hospitals, clinics, healthcare centers, radiology practices and medical imaging centers. The software is designed to be used by qualified clinical professionals (including physicians, radiologists and DXA technicians) and the physicians are solely responsible for making all final patient management decisions.
The provided text does not contain a discrete table of acceptance criteria nor explicit reported device performance metrics against such criteria in a tabular format. However, based on the Performance Testing Summary & Conclusions section (pages 9-10), the acceptance criteria can be inferred from the studies described. The reported performance is woven into the narrative of the study results.
Here's an attempt to derive the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Inferred from study objectives) | Reported Device Performance (TBS iNsight V4) |
|---|---|
| Accuracy of TBS V4 in reflecting bone microarchitecture (vs. microCT) | Strong and statistically significant correlations with parameters such as trabecular number and separation. Maintained or exceeded correlations established with the predicate. |
| Agreement between TBS V4 and Predicate Software | Excellent agreement, with correlation coefficients exceeding 0.99. Bland-Altman statistics confirmed negligible differences between versions across multiple DXA systems. |
| Reproducibility (Precision and Least Significant Change - LSC) in ex vivo settings | Precision and LSC values comparable to or slightly improved over the predicate, consistent across various DXA systems. |
| Reproducibility (Precision and LSC) in in vivo settings (clinical conditions with repositioning) | Met or exceeded precision standards established by the International Society for Clinical Densitometry (ISCD), with LSC values within acceptable thresholds. |
| Validity of Tissue Thickness (TT) Range | Validated in-vivo, establishing a valid tissue thickness range of 7-30 cm, aligning with physiological parameters for TBS calculations. |
| Applicability across diverse demographic groups (population-specific reference curves) | Generation of population-specific reference curves for age, sex, and ethnicity. Comparisons with the predicate software confirmed statistical equivalence, reinforcing reliability across diverse demographic groups, including multi-ethnic populations in the United States. |
| Performance in fracture risk assessment (TBS-adjusted FRAX and BMD T-scores) | Demonstrated similar TBS-adjusted FRAX and BMD T-scores for major osteoporotic and hip fractures compared to the predicate, using data from over 17,000 individuals across 14 international cohorts. |
| Effectiveness for treatment monitoring (TBS-adjusted FRAX and BMD T-scores over time) | Demonstrated effectiveness for accurately outputting TBS-adjusted FRAX and BMD T-scores over time for treatment monitoring, with substantial equivalence to the predicate, in a large cohort of postmenopausal women with osteoporosis. |
| Software Verification and Validation | Demonstrated specified requirements are met and the software functions as intended. |
| Cybersecurity Compliance | Controls and processes implemented; documentation and testing demonstrate alignment with FDA guidance ("Cybersecurity in Medical Devices") and Section 524B of the FD&C Act. |
| Substantial Equivalence (Overall Device Performance) | TBS version 4.0 demonstrates substantially equivalent performance in the assessment of bone quality and TBS-adjusted FRAX and BMD T-scores compared to the predicate device. Standalone validation studies confirm performance, reliability, and utility across diverse populations and DXA systems. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Clinical Association Study 1 (Correlation with microCT):
- Sample Size: 30 human cadaver lumbar vertebrae.
- Data Provenance: Not explicitly stated, but "human cadaver lumbar vertebrae" implies a laboratory-based, retrospective study.
- Clinical Association Study 2 (Agreement between V4 and Predicate):
- Sample Size: 15 human cadaver vertebrae.
- Data Provenance: Not explicitly stated, but "human cadaver vertebrae" implies a laboratory-based, retrospective study. Measurements were performed on "multiple DXA systems."
- Analytical Reproducibility (Ex Vivo Study):
- Sample Size: Not explicitly stated, but "dried human lumbar vertebrae," likely similar to the cadaver studies.
- Data Provenance: Retrospective, laboratory-based.
- Analytical Reproducibility (In Vivo Study):
- Sample Size: 132 participants.
- Data Provenance: Not explicitly stated, but "scanned on four different DXA systems" with "repositioning between scans" indicates a prospective, controlled clinical study.
- Performance and Utility (Fracture Risk Assessment):
- Sample Size: Over 17,000 individuals across 14 international cohorts.
- Data Provenance: International cohorts, implying retrospective data from multiple countries.
- Performance and Utility (Treatment Monitoring):
- Sample Size: A "large cohort" of postmenopausal women with osteoporosis. Specific number not provided.
- Data Provenance: Not explicitly stated, but likely retrospective clinical data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number or qualifications of experts used to establish ground truth for the various test sets.
- For microCT correlation: Micro-computed tomography (microCT) is an objective imaging technique, so human expert interpretation for ground truth might not be directly applicable in the same way as, for example, image interpretation by radiologists.
- For in vivo studies: The ground truth for bone microarchitecture or fracture risk assessment, as implied by the use of ISCD standards and FRAX, are typically established by clinical diagnosis and follow-up, rather than a panel of experts explicitly reviewing the test set images for ground truth.
4. Adjudication Method for the Test Set
The document does not describe any adjudication methods (e.g., 2+1, 3+1) for establishing ground truth from expert readers for the test sets.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- No, an MRMC comparative effectiveness study was not explicitly described in the provided text. The studies focus on the device's standalone performance, its agreement with the predicate device, and its correlation with bone microarchitecture ground truth (microCT), as well as its utility for fracture risk assessment and treatment monitoring.
- The text describes the device as a "complement to both DXA analysis and clinical examination" and states that "The results can be used by a physician in conjunction with other clinical risk factors." While it aids physicians, the document does not include a comparative study evaluating how human reader performance (e.g., diagnosis or risk assessment accuracy) improves with TBS iNsight V4 assistance versus without it.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, a standalone performance assessment was conducted. The "Standalone Performance Testing Summary" section (page 9) explicitly states that "Software Verification and Validation demonstrate specified requirements are met and the software functions as intended."
- Furthermore, the "Clinical Association" and "Analytical Reproducibility" studies directly evaluate the performance of TBS version 4.0 in calculating TBS values and their correlation with microCT independently, and their agreement/reproducibility compared to the predicate, without human intervention in the calculation process.
7. The Type of Ground Truth Used
- Pathology/Objective Measurement: For establishing the relationship between TBS and bone microarchitecture, micro-computed tomography (microCT) measurements were used as the ground truth (e.g., for trabecular number and separation). This is an objective measurement of bone structure.
- Clinical Diagnosis/Outcomes Data: For "fracture risk assessment" and "treatment monitoring," the ground truth implicitly refers to the clinically established reference values (e.g., those used in FRAX) and potentially actual fracture outcomes data from the cohorts. The studies confirm that TBS V4 provides "similar TBS-adjusted FRAX and BMD T-scores" and is effective for monitoring, suggesting that it aligns with known clinical indicators and outcomes.
- Predicate Device Performance: For demonstrating "agreement" and "substantial equivalence," the performance of the predicate device (TBS iNsight V3) serves as a comparative ground truth or benchmark.
8. The Sample Size for the Training Set
The document does not provide information regarding the sample size used for the training set for the TBS iNsight V4 algorithm.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for any training set was established, as the details of algorithm development and training are not included. The focus is on the performance testing of the final device version.
Ask a specific question about this device
(462 days)
Ave., Suite 280 BERKELEY.CA 94704
Re: K220402
Trade/Device Name: VirtuOst Regulation Number: 21 CFR 892.1170
---------------|--|
| Common Name: | QCT Bone Densitometer | |
| Regulation Number: | 21 CFR 892.1170
---------------------|
| Common Name: | QCT Bone Densitometer |
| Regulation Number: | 21 CFR 892.1170
VirtuOst uses data from computed tomography scans to estimate bone mineral density, bone strength, and a load-to-strength ratio. This information can be used by a physician to assess fracture risk, identify osteoporosis, and monitor therapy. For pediatric patients, VirtuOst provides these estimates without any classifications and should be used only when the benefit of obtaining these estimates outweighs the risk of radiation.
VirtuOst is a software-only medical device that analyzes data in computed tomography (CT) scans to measure bone mineral density (BMD), bone strength, and a load-to-strength ratio at the proximal femur and vertebral body. BMD is measured from both a 2D projection (in g/cm2) and a volumetric scan reconstruction (in mg/cm³) of the CT scan. VirtuOst measurements can be used by a physician to identify osteoporosis, assess fracture risk, and monitor therapy. The VirtuOst analysis is performed on previously physician-acquired image data and is unrelated to acquisition equipment and clinical workstations.
1. Table of Acceptance Criteria and Reported Device Performance
| Feature/Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Hip Areal BMD Equivalence (Modified Device vs. DXA) | Standard error of the regression ≤ 0.051 g/cm² for femoral neck areal BMD. | For the subject device, the standard error of the regression for femoral neck areal BMD was 0.048 g/cm² (overall). Stratified results were 0.044 g/cm² for women (N=244) and 0.051 g/cm² for men (N=80). All met the acceptance criterion. |
| Phantomless Calibration (Fat/Air vs. External Phantom) | Mean values of measurements from fat/air phantomless calibration and external phantom calibration are not significantly different (p>0.05) and show a high degree of agreement (e.g., R² close to 1) with no fixed or proportional bias. | Mean values of measurements using the two calibration methods were not significantly different (p>0.05) and showed a high degree of agreement (R² = 0.98-0.99) with no fixed or proportional bias. |
| CT-to-DXA Areal BMD Conversion for Hip (T- and Z-scores) | Standard error of the regression of less than 0.051 g/cm² for the updated sex-specific conversion equations. | The standard errors for the new sex-specific equations ranged from 0.025-0.047 g/cm², which were considered acceptable and met the criterion. |
| Slice Spacing Adjustment for Femurs Effectiveness | Measurements from images with thin and thick slice spacing are highly correlated (R > 0.98). Adjustments lead to measurements equivalent to 1.25 mm scans. No evidence of a slice spacing effect on hip fracture prediction. | Measurements from images with thin and thick slice spacing were highly correlated (R > 0.98). The adjustments resulted in equivalent measurements. An observational study demonstrated that hazard ratios for hip fracture using femoral strength or hip areal BMD from VirtuOst were at least as good as DXA, and there was no evidence of a slice spacing effect on these results. |
| Non-clinical Software Functionality | All software requirements are met. The device performs as intended for all modified functionalities. | Automated unit and regression testing, as well as manual component testing (black box testing), confirmed that software requirements were met and the device performed as intended for all modified functionalities (worklist, new input types, improved femoral head segmentation, fat/air calibration, slice spacing adjustments, range checks, results report elements, DICOM output). All tests passed. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Hip Areal BMD Equivalence:
- Test Set Size: 324 women and men (overall). Specifically, 244 women and 80 men.
- Data Provenance: Not explicitly stated, but implies clinical data. Retrospective or prospective is not specified.
- Phantomless Calibration:
- Development of reference data for visceral fat attenuation: 268 patients and a custom torso phantom scanned on 31 and 35 different CT scanners respectively.
- Comparison of fat/air phantomless calibration vs. phantom calibration: 40 independent patient CT scans.
- Data Provenance: Not explicitly stated, but implies clinical data from various CT manufacturers (GE Healthcare, Siemens Medical Solutions, Philips Healthcare, Toshiba International) at various settings (80-140 kVp). Retrospective or prospective is not specified.
- CT-to-DXA Areal BMD Conversion for Hip:
- Test Set Size: 200 subjects scanned on both CT and DXA.
- Data Provenance: Not explicitly stated, but implies clinical data. Retrospective or prospective is not specified.
- Slice Spacing Adjustment for Femurs:
- Development of adjustments: Pairs of CT images (unspecified number).
- Observational study: 2,783 women and men (1,306 with hip fracture). This was a case-cohort study design.
- Data Provenance: Not explicitly stated, but implies clinical data. The observational study implies retrospective data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document does not describe the use of human experts to establish ground truth for the test sets. Instead, it relies on:
- Clinical reference standards: FDA-cleared DXA devices for hip areal BMD equivalence.
- External calibration phantoms: For validating phantomless calibration.
- Outcomes data: Hip fracture prediction in an observational study for slice spacing adjustment.
- Regression analysis and statistical comparison: For quantitative metrics.
4. Adjudication Method for the Test Set
Not applicable, as expert adjudication was not used to establish ground truth. Ground truth was based on objective measurements from clinical reference standards (DXA, external phantoms) or clinical outcomes.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study involving human readers with and without AI assistance was mentioned. The study focused on the equivalence of the device's measurements to established clinical standards (DXA, phantom) and its ability to predict outcomes, rather than human reader performance improvement.
6. Standalone (Algorithm Only) Performance Study
Yes, the studies presented primarily demonstrate the standalone performance of the VirtuOst algorithm.
- The comparison of hip areal BMD measurements to DXA devices is a standalone assessment.
- The phantomless calibration validation compares the algorithm's output to a known standard (external phantom).
- The CT-to-DXA conversion study assesses the algorithm's ability to produce DXA-equivalent T- and Z-scores.
- The slice spacing adjustment validation and the observational study on hip fracture prediction also assess the algorithm's standalone performance in providing clinically relevant measurements.
7. Type of Ground Truth Used
The ground truth used was a combination of:
- Clinical reference standards: Measurements from FDA-cleared DXA devices for hip areal BMD.
- Physical standards: External calibration phantoms for validating phantomless calibration.
- Outcomes data: Hip fracture occurrence in a case-cohort study for evaluating the slice spacing adjustment and overall clinical utility.
- Regression analysis and statistical comparison: Used to define the relationship between the device's measurements and the chosen ground truths.
8. Sample Size for the Training Set
The document does not explicitly state the sample sizes used for training the VirtuOst algorithm. The information provided relates to testing and validation of the modified device.
9. How the Ground Truth for the Training Set Was Established
The document does not provide details on how the ground truth for the training set was established. The studies described are primarily focused on the validation of the modified device's performance against established benchmarks and clinical data, rather than the initial development and training of the core algorithms. It is implied that the predicate device (K113725) was developed and trained prior to this submission, and this submission focuses on demonstrating that modifications have not negatively impacted safety or effectiveness.
Ask a specific question about this device
(263 days)
Barcelona, Barcelona 08036 SPAIN
Re: K220822
Trade/Device Name: 3D-SHAPER Regulation Number: 21 CFR 892.1170
|
| REGULATION NUMBER | 892.1170
| Identical |
| RegulationNumber | 21 CFR 892.1170
| 21 CFR 892.1170
3D-SHAPER uses data from conventional Dual Energy X-Ray Absorptiometry (DXA) scans to measure the distribution of bone mineral mass at specific cross sections of the hip and allows the physician to estimate structural properties of the hip, such as CSA, CSMI, Z, Buckling Ratio, cortical sBMD, trabecular vBMD and integral vBMD.
3D-SHAPER is indicated to be used by qualified medical professionals to assist healthcare professionals in the bone health evaluation & changes, and cannot fully substitute their clinical judgement.
3D-SHAPER is a software reconstructing in 3D, the shape and bone mineral density from 2D DXA images. The software incorporates a DXA-based 3D reconstruction and measures the distribution of bone mineral mass at specific cross sections of the hip. It provides physicians with estimates of the structural properties of the hip, such as densities or geometrical parameters. The software is not suitable for use with paediatric patients or for subjects with hip prosthesis such as total hip replacement prosthesis, hip resurfacing prosthesis or an osteosynthesis system, since the software has not been tested against these conditions. 3D-SHAPER is indicated to be used by qualified medical professionals to assist healthcare professionals in the bone health evaluation & changes, and cannot fully substitute their clinical judgement.
The 3D-SHAPER® measurements include:
Cortical surface Bone Mineral Density (Cortical sBMD, mg/cm2)
Trabecular volumetric Bone Mineral Density (Trabecular vBMD, mg/cm³)
Integral volumetric Bone Mineral Density (Integral vBMD, mg/cm³), the integral compartment being the union of the cortical and trabecular compartments.
Hip structural measurements: Cross-Sectional Area (CSA, in cm²), Cross-Sectional Moment of Inertia (CSMI, in cm4), section modulus (Z, in cm3) and buckling Ratio (BR).
Cortical sBMD, trabecular VBMD and integral vBMD are calculated at the total femur. CSA, CSMI, Z and BR are calculated in cross-sections at the neck, intertrochanteric region and lower shaft.
Here's a breakdown of the acceptance criteria and study information for the 3D-SHAPER device, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance:
The document primarily focuses on demonstrating substantial equivalence to a predicate device through correlation studies rather than establishing absolute performance metrics against predefined acceptance criteria for each measurement. However, it does outline the type of performance expected in the "Performance Bench Testing" section.
| Feature/Metric | Acceptance Criteria (Implied by correlation values) | Reported Device Performance (Correlation to QCT) |
|---|---|---|
| Integral vBMD | Very strong correlation (r ≥ 0.9) | r = 0.96 |
| Cortical sBMD | Very strong correlation (r ≥ 0.9) | r = 0.95 |
| Trabecular vBMD | Very strong correlation (r ≥ 0.9) | r = 0.93 |
| CSA | Very strong correlation (r ≥ 0.95) | r ≥ 0.95 |
| CSMI | Very strong correlation (r ≥ 0.92) | r ≥ 0.92 |
| Z | Very strong correlation (r ≥ 0.91) | r ≥ 0.91 |
| Buckling Ratio (BR) | Strong correlation (r ≥ 0.7) | r ≥ 0.71 |
| Computing time | Not explicitly stated as a target | Lower than 2 minutes and 30 seconds |
| Software convergence | Not explicitly stated as a target | Superior to 95% |
2. Sample Size Used for the Test Set and Data Provenance:
The document mentions two main studies:
- Correlation Study with Predicate Device:
- Sample Size: 740 men and women (94% women).
- Data Provenance: Not explicitly stated (e.g., country of origin). The study is retrospective in the sense that existing DXA scans were analyzed by both the 3D-SHAPER and the predicate device.
- Performance Bench Testing (Correlation to QCT):
- Sample Size: Not explicitly stated for this particular correlation study with QCT. It refers to "subject-specific models of the femoral shape and measurements compared to Quantitative Computed Tomography (QCT)." This implies data from human subjects but the exact number isn't provided.
- Data Provenance: Not explicitly stated. Likely retrospective, as it compares measurements to an established imaging modality (QCT).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Correlation Study with Predicate Device: Ground truth for this study was the measurements produced by the predicate device (Hologic Hip Structural Analysis software, K061561). No human experts were involved in establishing this specific ground truth; it relied on a previously cleared, validated software.
- Performance Bench Testing (Correlation to QCT): Ground truth was established by Quantitative Computed Tomography (QCT) measurements. QCT is a quantitative imaging technique that provides volumetric bone mineral density and structural information. No human experts are described as manually establishing this ground truth; it's a direct comparison to QCT output.
4. Adjudication Method:
- Correlation Study with Predicate Device: There was no adjudication method described. The study compared the output of the 3D-SHAPER software directly against the output of the predicate device.
- Performance Bench Testing (Correlation to QCT): There was no adjudication method described. The study compared the output of the 3D-SHAPER software directly against QCT measurements.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study described focuses on the comparison of the 3D-SHAPER software's measurements against a predicate software and against QCT, not on how human readers' performance improves with or without AI assistance.
6. Standalone Performance:
Yes, a standalone performance study (algorithm only without human-in-the-loop) was done. The document consistently describes the 3D-SHAPER software's ability to measure structural properties from DXA scans and correlates these measurements with a predicate device and QCT. This is considered standalone performance.
7. Type of Ground Truth Used:
- Correlation Study with Predicate Device: The ground truth was the output of a legally marketed predicate device (Hologic Hip Structural Analysis software).
- Performance Bench Testing (Correlation to QCT): The ground truth was Quantitative Computed Tomography (QCT) measurements.
8. Sample Size for the Training Set:
The document does not provide any information regarding the sample size used for the training set. It focuses on the validation of the device.
9. How the Ground Truth for the Training Set Was Established:
The document does not provide any information on how the ground truth for the training set was established, as the training process itself is not detailed.
Ask a specific question about this device
(240 days)
St TORRANCE CA 90502
Re: K213760
Trade/Device Name: ABMD Software Regulation Number: 21 CFR 892.1170
Density Software Trade Name: ABMD Software Common Name: ABMD Software Classification: Class II, 21 CFR 892.1170
QCT Bone Mineral Density Analysis Software Predicate 510(k): K894854 Classification: Class II, 21 CFR 892.1170
Asynchronous Calibration Module, CliniQCT Predicate 510(k): K140342 Classification: Class II, 21 CFR 892.1170
The Automated Bone Mineral Density Software Module (ABMD) is a post-processing AI-powered software intended to measure bone mineral density (BMD) from existing CT scans by averaging Hounsfield units in the trabecular region of vertebral bones. ABMD is not intended to replace DXA or any other tests dedicated to BMD measurement. It is solely designed for measuring BMD in existing CT scans ordered for reasons other than BMD measurement. In summary, ABMD is an opportunistic AI-powered tool that enables: (1) retrospective assessment of bone density from CT scans acquired for other purposes, (2) assessment of bone density in conjunction with another medically appropriate procedure involving CT scans, and (3) assessment of bone density without a phantom as an independent measurement procedure.
The Automated Bone Mineral Density (ABMD) Software is a software module that estimates bone mineral density in the vertebral bones by averaging Hounsfield Units (HU) in the trabecular area. ABMD Software is a post-processing software that works on existing CT scans. ABMD Software measurements are to be reviewed by radiologists and should be used by healthcare providers in conjunction with clinical evaluation.
Here's a breakdown of the acceptance criteria and study information for the ABMD Software based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Correlation with Manual QCT-based BMD Measurement | Strong correlation reported (t = 0.97, p<0.01). |
| Correlation with DXA BMD Measurement | Significant correlation reported (r = 0.72, p<0.01). This closely matched correlations reported in literature between DXA and manual QCT (r=0.5 to r=0.75). |
| Sample Volume Placement (relative to cortical bone) | Software passes if the sample volume is at least 1 pixel away from the cortical border. |
| Functional Requirements and Performance Specifications | All functional requirements and performance specifications were met. |
| Agreement with Manual QCT-based BMD Measurement | Strong agreement reported. |
| Agreement with DXA BMD Measurement | Modest but significant agreement reported. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Study 1 (Reference Dataset):
- Sample Size: 993 cases.
- Data Provenance: Not explicitly stated, but indicated as a "cohort of asymptomatic cases who underwent CT scans." The geographical origin (country) is not specified. It is retrospective as it uses "existing CT scans."
- Study 2:
- Sample Size: 172 asymptomatic cases.
- Data Provenance: Not explicitly stated, but indicated as cases who underwent "whole-body DXA scans as well as CT scans." The geographical origin (country) is not specified. It is retrospective as it uses "existing CT scans."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: Not explicitly stated in terms of a specific count. However, the ground truth was established by "trained operators."
- Qualifications of Experts: Described as "trained operators" for both manual QCT measurements and DXA scan derivations. Specific qualifications like "radiologist with 10 years of experience" are not provided.
4. Adjudication Method for the Test Set
- The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the test set's ground truth. It states that "QCT BMD, T-score, and Z-score values derived from manual measurement by trained operators" were used for ground truthing. This implies a single measurement by a "trained operator" formed the ground truth for QCT, and DXA values were also used, presumably from standard reports.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance?
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described in the provided text. The studies focused on comparing the ABMD software to manual QCT measurements and DXA measurements, not on how human readers perform with or without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
- Yes, standalone performance was assessed. The studies directly compare the ABMD Software's measurements (which are algorithm-driven) to established ground truth methods (manual QCT and DXA). The text states, "The ABMD Software strongly correlated with manual QCT-based BMD measurement" and "The ABMD Software also correlated with DXA BMD measurement." This indicates the algorithm's performance independent of human-in-the-loop interaction for the measurement itself, though the results are intended for review by radiologists.
7. The Type of Ground Truth Used
- Study 1:
- Type: Expert consensus (implicitly, from "trained operators") for QCT BMD, T-score, and Z-score values derived from manual measurements.
- Study 2:
- Type: Combined expert consensus (implicitly, from "trained operators") for QCT BMD, T-score, and Z-score values derived from manual measurements, and clinical outcomes/measurements from DXA scans (DXA T-score and Z-score values).
8. The Sample Size for the Training Set
- The document does not explicitly state the sample size for the training set. It mentions the ABMD Software uses "an AI trained model," but the details of the training data are not provided in this summary.
9. How the Ground Truth for the Training Set Was Established
- The document does not explicitly state how the ground truth for the training set was established. It only mentions the ground truthing process for the test/reference datasets.
Ask a specific question about this device
(178 days)
|
| K082317 | 21 CFR 892.1170 | FRAX® 10-Year Fracture RiskSoftware Option for
- EchoS Family is a non-invasive ultrasound (US) bone sonometer. EchoS Family works together with EchoStudio software. EchoStudio analyzes the ultrasound signals in order to compute the diagnostic parameters (BMDus, T-score, and Z-score). The BMDus Index is a clinical measure based on ultrasound variables of the lumbar spine or femoral neck which is highly correlated with the value of BMD of the same anatomical location as provided by DXA (BMDDXA), with a standard error of the estimate of 0.044grams/cm^2 for lumbar spine and 0.038 grams/cm^2 for femoral neck measurements.
BMDus Index is expressed in grams/cm^2 and as a T- and Z-score, derived from comparison to a normative x-ray absorptiometry reference database. BMDus Index has a precision comparable to that of x-ray absorptiometry, which makes it suitable for monitoring bone changes in women and men.
- The EchoStudio software includes an optional tool called Fragility Score, which is intended to provide an assessment of 5-year fracture risk. The optional tool Fragility Score provides an estimate of 5-year probability of hip fracture and 5-year probability of a major osteoporotic fracture (clinical spine, forearm, hip or shoulder fracture). This estimate takes into account the patient's age, sex, ethnicity, height, and the vertebraffemur neck ultrasound spectra and is computed using a proprietary algorithm. The tool has been validated for men and women between 30 and 90 years old.
The output is provided in a separate screen display and report that can be viewed or exported to an optional physician report generator tool. The results can be used by a physician, in conjunction with other clinical risk factors, as an aid in the diagnosis of osteoporosis and medical conditions leading to increased bone fragility, and ultimately in the assessment of fracture risk.
- EchoS Family, when used with the optional Body Composition module of EchoStudio software, is indicated to estimate total body fat percentage (%BF). EchoS Family, together with Body Composition software module, is intended to be used only on generally healthy adults and is not for disease or condition. Body Composition software is indicated for the calculation of the Body Mass Index (BMI) and the Basal Metabolic Rate (BMR). Body Composition software module generates a report which displays the calculated values of Body FAT, BMR, and Body Mass Index (BMI).
The EchoS Family is an ultrasound device intended primarily for the diagnosis of osteoporosis. EchoS, through the ultrasound scan of the lumbar or femoral site of interest, picks up the ultrasound signal (RF) and performs an estimate of the bone mineral density (BMD).
The device therefore allows not only the visualization of ultrasound images, but also the real-time sampling of the RF signal and its appropriate treatment to make it usable for diagnostic algorithms.
The EchoS Family consists in two different configurations: EchoS (portable version) and the EchoStation (cart version). Each version consists of two main parts: the equipment device (EchoS and EchoStation) with its own probe and the software EchoStudio.
EchoStudio is a biomedical software that, used in combination with EchoS Family, allows the evaluation of bone mineral density (BMD) by means of the proprietary method REMS (Radiofrequency Echographic Multi Spectrometry) densitometry.
By using EchoStudio, it is possible to assess the key diagnostic parameters directly on the anatomical sites with increased fracture risk, such as lumbar spine and proximal femur.
EchoStudio analyzes the ultrasound signals and echographic images in order to compute the diagnostic parameters (BMD, T-score, and Z-score) and to estimate the fracture risk by means of the Echolight diagnostic algorithms and non-ionizing technique.
The provided FDA 510(k) summary for the EchoS Family device describes clinical performance studies related to its three Indications for Use (IFUs):
IFU 1: Diagnostic parameters (BMDus, T-score, and Z-score)
This IFU relates to the device's ability to compute diagnostic parameters based on ultrasound signals from the lumbar spine or femoral neck, correlating with DXA BMD.
IFU 2: Fragility Score for 5-year fracture risk assessment
This IFU relates to an optional tool that provides an estimate of 5-year probability of hip fracture and major osteoporotic fracture.
IFU 3: Body Composition module for total body fat percentage (%BF), BMI, and BMR
This IFU relates to an optional module for estimating body fat percentage, and calculating Body Mass Index (BMI) and Basal Metabolic Rate (BMR).
However, the summary does not provide a table of and does not explicitly state numerical acceptance criteria for each IFU, nor does it present the reported device performance in a direct comparative table. Instead, it describes general claims of correlation and mentions clinical trial reports.
Here's a breakdown of the available information based on your requested categories, acknowledging the limitations from the provided text:
Acceptance Criteria and Reported Device Performance
As noted, the document does not contain a specific table of acceptance criteria with numerical targets. However, the descriptions of the clinical studies imply the intended performance.
| Indication for Use | Implied Acceptance Criteria (from text) | Reported Device Performance (from text) |
|---|---|---|
| 1. Diagnostic parameters (BMDus, T-score, Z-score) | High correlation with DXA BMD. Precision comparable to DXA. | BMDus Index has a standard error of the estimate of 0.044 g/cm² for lumbar spine and 0.038 g/cm² for femoral neck measurements compared to DXA. BMDus Index has a precision comparable to that of x-ray absorptiometry. |
| 2. Fragility Score for 5-year fracture risk assessment | Ability to identify patients at risk for osteoporotic fracture within 5 years. | "demonstrating the ability of the Fragility Score (FS) parameter... to identify patients... at risk for osteoporotic fracture within 5 years." (No specific metrics provided in this summary). |
| 3. Body Composition (%BF, BMI, BMR) | Feasibility and accuracy of body fat percentage calculation by ultrasound (correlation with BIA). Feasibility and accuracy of BMR calculation (correlation with BIA). Validation of BMI calculation (with BIA and pocket calculator). | For %BF: "assesing the correlation with the corresponding bioelectrical impedance analysis (BIA)-measured parameter, assumed as the reference." For BMR: "assesing the correlation between basal metabolic rate (BMR) measured by the new US technology and by bioelectrical impedance analysis (BIA)-based technologies, assumed as the reference." For BMI: "Validate the calculation of the body mass index (BMI) provided by a novel US-based device with respect to measurements obtained by a bioelectrical impedance analysis (BIA)-based body composition monitor and to the results obtained by a pocket calculator, taken as reference." (No specific correlation coefficients or accuracy metrics provided in this summary). |
Study Details
The document mentions several clinical studies but often refers to "Performance Evaluation Protocol" and "Data Evaluation Report" documents rather than providing the detailed study results within the 510(k) summary itself. Therefore, specific details for all your requested points are limited or not available in the provided text.
1. Sample sizes used for the test set and the data provenance:
-
IFU 1 (Diagnostic parameters):
- One study produced a publication "Radiofrequency echographic multi spectrometry for the prediction of incident fragility fractures: A 5-year follow-up study."
- A second study produced a "Performance Evaluation Protocol" and a "Data Evaluation Report" titled "ECHOLIGHT: Comparative Performance Evaluation Report Echos System Vs. DXA in a male population aged 30-90 years".
- Sample Size: Not explicitly stated for either study in this document.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the publication title "5-year follow-up study" suggests a prospective or longitudinal study in some capacity for the first study mentioned. The second study compared to DXA, implying live patient data.
-
IFU 2 (Fragility Score):
- "a dedicated clinical study was conducted."
- Sample Size: Not explicitly stated.
- Data Provenance: Not explicitly stated.
-
IFU 3 (Body Composition):
- Three clinical studies were conducted for %BF, BMR, and BMI respectively.
- Sample Size: Not explicitly stated for any of these three studies.
- Data Provenance: Not explicitly stated. The comparison to BIA and pocket calculator suggests live patient data.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not explicitly stated in the provided text for any of the studies.
- For IFU 1, DXA is used as the reference/ground truth, which is a widely accepted diagnostic method, implying established medical standards rather than expert consensus on images.
- For IFU 3, Bioelectrical Impedance Analysis (BIA) and a pocket calculator (for BMI) are used as references, which are also established methods, not typically requiring expert image review.
3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not explicitly stated in the provided text for any of the studies. Given the nature of the ground truth (DXA, BIA), expert adjudication of images or diagnoses might not have been the primary method.
4. Multi-reader multi-case (MRMC) comparative effectiveness study:
- No MRMC comparative effectiveness study involving human readers with/without AI assistance is mentioned. The device appears to be a diagnostic tool providing objective measurements, not an AI assisting human interpretation of complex images.
5. Standalone (i.e., algorithm only without human-in-the-loop performance) study:
- Yes, the performance data presented is for the algorithm/device itself. The device computes diagnostic parameters, fracture risk estimates, and body composition values. This implies standalone performance, where the output is directly generated by the EchoS Family and EchoStudio software.
6. Type of ground truth used:
- IFU 1 (Diagnostic parameters): DXA (Dual-energy X-ray Absorptiometry) for BMD, which is a clinical gold standard for bone mineral density measurement.
- IFU 2 (Fragility Score): Clinical outcomes (incident fragility fractures within 5 years).
- IFU 3 (Body Composition): Bioelectrical Impedance Analysis (BIA) for %BF and BMR; results from a pocket calculator for BMI.
7. Sample size for the training set:
- Not explicitly stated in the provided text for any of the algorithms (BMD, Fragility Score, Body Composition).
8. How the ground truth for the training set was established:
- Not explicitly stated in the provided text. However, it can be inferred that for algorithms correlating with DXA or BIA, similar methods to the test set ground truth would have been used for training. For the Fragility Score, based on its purpose, the training likely used historical patient data with documented fracture outcomes and associated clinical/ultrasound data.
Ask a specific question about this device
(146 days)
Trade/Device Name: GEHC DXA Bone Densitometers with enCORE version 18 Regulation Number: 21 CFR 892.1170
| Bone Densitometer (21CFR 892.1170
The proposed device: GE Lunar DXA Bone Densitometers with the enCORE version 18 software is intended for medical purposes to measure bone density, bone mineral content, and fat and lean tissue content by x-ray transmission measurements through the bone and adjacent tissues.
The changes proposed in this Premarket Notification will be used with the existing GEHC DXA Bone Densitometers, which utilize the enCORE software.
The GEHC DXA Bone Densitometers are composed of a scanner and a computer. The scanner comprises the x-ray source and detector, the patient scan table, the mechanical drive system, and the lowest level portions of the control system. The scanner is in communication with the computer, which is a standard PC. The computer runs the enCORE software, and thus controls the scanner, acquires scan data from the scanner, stores and analyzes the data, and interacts with the human operator.
GEHC DXA Bone Densitometers are used healthcare facilities and hospitals to measure bone mineral density (BMD) and body composition (%fat, fat mass, lean mass) using a technique called Dual-energy X-ray Absorptiometry or DXA. DXA measures the attenuation of x-rays of two different energy levels after they pass through the body of a subject. As bone, fat tissue, and lean tissue absorb the different energy x-rays at different rates, the relative attenuation of each x-ray energy is measured and used to calculate the composition of each pixel.
The provided text describes the GEHC DXA Bone Densitometers with enCORE version 18. This device is a bone densitometer that measures bone mineral density (BMD) and body composition using Dual-energy X-ray Absorptiometry (DXA). The submission is a 510(k) premarket notification, indicating that the device claims substantial equivalence to existing legally marketed predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with specific numerical thresholds for performance. Instead, it describes various software features and states that bench performance testing confirmed that design outputs met design input requirements and that results are "comparable" or "precise and accurate."
Here's an attempt to structure the information into a table based on the provided text, recognizing that precise criteria are often implied rather than explicitly stated with quantifiable targets in this type of summary:
| Feature/Study | Acceptance Criteria (Implied from text) | Reported Device Performance |
|---|---|---|
| DXAVision and Adult TBLH (Bone & Body Composition) | Differences in DXA results for neck-to-knees scans and total body measurements should be slightly larger but similar to precision studies in literature. | Differences are "slightly larger but similar to the results of precision studies of duplicate scans reported in the literature for both iDXA and Prodigy." |
| Integrated TBS iNsight (TBS and FRAX adjusted for TBS) | Results should be comparable between the integrated TBS and the previously cleared standalone TBS iNsight application. | "demonstrates that TBS and FRAX adjusted for TBS results are comparable between results calculated with the TBS integrated into enCORE 18 software and previously cleared TBS iNsight application." |
| CoreScan VAT/SAT (Visceral and Subcutaneous Adipose Tissue) | VAT/SAT results should be precise and accurate. | "demonstrate[s] that the VAT/SAT results for the updated CoreScan software option are precise and accurate." |
| Small Body Composition ROIs (Bone and Body Composition of Arms/Legs) | Accuracy of bone and body composition for upper and lower arms and legs should be confirmed. | "confirm the accuracy of bone and body composition of the upper and lower arms and legs." |
| Advanced Analytics | The feature should incorporate results of previously cleared features into user-defined equations and not affect safety or effectiveness. | "incorporates results of previously cleared features into user defined equations and does not affect the safety or effectiveness of the system." |
| Software Verification and Validation | Software should meet design specification requirements. | "confirmed that the software met the design specification requirements." |
2. Sample Size Used for the Test Set and Data Provenance
- DXAVision/Adult TBLH: The study involved a "heterogeneous sample of men and women with a diverse range of age, BMI, BMD, and body fat." No specific numerical sample size is provided.
- Integrated TBS iNsight: "internal engineering DXA data sets" were used. No specific numerical sample size.
- CoreScan: An "anthropomorphic phantom" was used. This is not human data.
- Small Body Composition ROIs: No specific sample size or provenance is mentioned for the verification results.
- Advanced Analytics: No specific sample size or provenance is mentioned.
- Software Verification and Validation: No specific sample size or provenance is mentioned besides "software verification and validation testing."
The provenance of human data, based on the description, is internal to GE Healthcare ("internal engineering DXA data sets"). It is implied to be retrospective, as it refers to existing data sets.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not explicitly mention the number or qualifications of experts used to establish ground truth for any of the described tests. The studies focus on comparing the updated software features to previous cleared versions or literature, or phantom data, rather than independent expert-adjudicated ground truth.
4. Adjudication Method for the Test Set
No adjudication method is mentioned for any of the described tests. The evaluations appear to be based on comparisons to predicate technology, internal software testing, or phantom measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
No MRMC comparative effectiveness study is mentioned. The device's improvements are focused on streamlined workflow, integration of features, and accuracy/precision of measurements, not on direct human reader improvement with AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, the studies described are primarily standalone algorithmic performance evaluations. For example, the CoreScan bench testing used an anthropomorphic phantom, and the TBS iNsight validation compared algorithmic outputs. The DXAVision and Adult TBLH studies compare algorithm results to literature-reported precision, implying a standalone assessment of the algorithm's measurement precision.
7. The Type of Ground Truth Used
- DXAVision/Adult TBLH: The "ground truth" used for comparison appears to be the precision of "duplicate scans reported in the literature" for existing iDXA and Prodigy devices. This implies a comparison to established performance benchmarks rather than an independently adjudicated 'ground truth' for each case.
- Integrated TBS iNsight: The "ground truth" is the results from the "previously cleared TBS iNsight application."
- CoreScan: The "ground truth" is derived from the known characteristics of an "anthropomorphic phantom."
- Small Body Composition ROIs: "Accuracy of bone and body composition" is confirmed, suggesting comparison to a known standard or previously validated measurements, but the specific nature of this ground truth is not detailed.
- Advanced Analytics: No explicit ground truth is mentioned, as this feature is about user-defined equations using existing data.
8. The Sample Size for the Training Set
The document does not provide details about training sets for any of the software features. This submission is for updates to existing software (enCORE version 18) and integration of previously cleared applications, suggesting development was not based on a new, distinct training phase described in this summary.
9. How the Ground Truth for the Training Set Was Established
Since no training set details are provided, the method for establishing ground truth for a training set is not described. The focus of this 510(k) submission is on changes and integrations based on verified algorithms and features, rather than the initial development of a new AI model requiring a training set.
Ask a specific question about this device
(445 days)
|
| Secondary Regulation Number: | 21 CFR 892.1170
|
| Primary Regulation Number: | 21 CFR 892.1170
Virtu Ost VFA uses sagittal sections from a spine-containing CT scan, with or without contrast enhancement, to visualize and measure vertebral deformities, classify the type and grade of any existing vertebral fracture, and from this identify patients at high risk of a future osteoporosis-related fracture. This information can be interpreted by a physician to diagnose existing vertebral fractures and to manage patients for osteoporosis.
VirtuOst Vertebral Fracture Assessment (VFA) software is used to perform vertebral morphometry and is an integrated component of VirtuOst (K113725). VirtuOst VFA accepts as input a sagittal section of a vertebral body, obtained from a spine-containing computed tomography (CT) scan, and provides semi-automated, interactive tools with which the user can perform six-point quantitative vertebral morphometry according to well-established guidelines. The process can be repeated for multiple vertebral bodies. For each vertebral body analyzed, the quantitative vertebral morphometry algorithm locates three points each along the superior and inferior endplate on a sagittal section through the vertebra, typically a mid-sagittal section. The user verifies or modifies point placement. Based on these six morphometry points, anterior, posterior and middle vertebral heights are measured. Percent deformities are calculated from these heights and are then compared against well-established criteria for vertebral deformities typical of osteoporosis-related vertebral fracture in order to classify types and grades of any existing osteoporosis-related vertebral fracture, from which a patient can be classified as being at high risk of a future osteoporosis-related fracture. The VFA algorithm can be applied to any complete vertebral body captured in the CT scan, and the scan can be contrast-enhanced or not. Consistent with the recommended use of the well-established fracture-classification criteria, deformity types and fracture grades and fracture-risk classifications are only reported for vertebral levels T4 through L4. A report is generated containing these results, along with images of the sagittal sections from which the measurements were acquired. A physician then interprets the report to make any medical diagnoses or treatment decisions.
Here's a breakdown of the acceptance criteria and the study details for the VirtuOst Vertebral Fracture Assessment device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state quantitative acceptance criteria in percentages or thresholds in a dedicated section. However, it implicitly uses agreement and precision metrics compared to manual measurements and a predicate device as evidence of acceptable performance for market clearance.
| Acceptance Criteria (Implied) | VirtuOst VFA Reported Performance |
|---|---|
| Accuracy for fracture determination (vs. manual) | kappa = 0.94 (95% CI = 0.82-1.06) |
| Accuracy for fracture grading (Grades 0-3 vs. manual) | kappa = 0.76 (95% CI = 0.60-0.91) |
| Mean difference in height-ratio (VirtuOst VFA vs. manual) | 0.005 (small compared to range 0.17-1.13) |
| No significant difference in fracture determination (vs. manual) | p = 0.32 |
| No significant difference in fracture grades (vs. manual) | p = 0.32 |
| Intra-operator precision (SDRMS) | 0.018 (similar to predicate K103475: 0.021-0.039) |
| Interoperator precision (SDRMS) | 0.023 (similar to predicate K103475: 0.025-0.048) |
| Agreement in fracture determination (1.25mm vs 3mm slice) | kappa = 0.88 (95% CI = 0.72-1.04) |
| Agreement in fracture grading (1.25mm vs 3mm slice) | kappa = 0.66 (95% CI = 0.47-0.84) |
| Intra-operator precision (variable slice thickness) | SDRMS = 0.034 (within range of predicate K103475 at constant slice thickness) |
| Intra-operator precision (one enhanced image) | SDRMS = 0.020 (similar to unenhanced images) |
| Mean difference in height-ratio (unenhanced vs. enhanced) | 0.007 (small compared to range in height-ratios) |
Summary of Device Performance: The study concludes that VirtuOst VFA is highly accurate, and its precision is substantially equivalent to that of the predicate device. Precision errors are small and do not significantly affect fracture determination or grading, even with varying CT slice thicknesses (up to 3mm) and the use of IV contrast enhancement.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: n=40 women and men (35 with potential fractures, 5 without).
- Data Provenance:
- Country of Origin: Not explicitly stated, but the study used "computed tomography scans acquired as part of standard care for 7000 women and men age 65 and older."
- Retrospective or Prospective: The selection of patients from an existing database of 7000 CT scans, where scans were within a 90-day period (implying existing sets of scans), indicates a retrospective study design for data collection.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Not explicitly stated. The text mentions "vertebral heights were also manually measured from high resolution (0.5 mm) printouts of midsagittal sections of the L1 vertebra from each CT scan." It does not specify who performed these manual measurements or their qualifications. It simply refers to these "manual measurements" as the reference standard.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly described. The manual measurements are presented as a reference standard, but the process by which multiple experts might have agreed upon these measurements (e.g., 2+1, 3+1) is not detailed. The study mentions two operators performing VFA, but this is for repeatability analysis of the device, not for establishing ground truth directly.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
-
Was an MRMC study done?: No. The study conducted was primarily a precision and accuracy study, comparing the device's measurements against manual measurements and assessing intra- and inter-operator variability. It did not involve comparing human readers' diagnostic performance with AI assistance versus without AI assistance.
-
Effect size of human reader improvement with AI vs. without AI assistance: Not applicable, as no MRMC comparative effectiveness study was performed in this manner.
6. Standalone Performance Study
- Was a standalone study done?: Yes. The study investigated the "Accuracy of VirtuOst VFA relative to the manual measurements" and its repeatability (intra-operator, interoperator, short-term, combined precision), as well as the impact of slice thickness and contrast enhancement on the device's performance. This directly assesses the algorithm's performance independent of real-time human interpretation loop changes in diagnostic workflow.
7. Type of Ground Truth Used
- Ground Truth Type: "Manual measurements" of vertebral heights from high-resolution printouts of midsagittal sections using six-point morphometry methods. This constitutes an expert-derived or reference standard measurement based on established definitions of vertebral fracture. The specific "experts" who performed these manual measurements are not explicitly qualified in the text.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not specified. The document focuses on the validation study and does not provide details about the training data used to develop the VirtuOst VFA algorithm.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth for Training Set Was Established: Not specified in the provided text.
Ask a specific question about this device
Page 1 of 9