    510k Summary Text (Full-text Search) :

    Linn, OR 97068

    Re: K251876
    Trade/Device Name: cleadew GP hydra one
    Regulation Number: 21 CFR 886.5918
    Linn, OR 97068

    Re: K251876
    Trade/Device Name: cleadew GP hydra one
    Regulation Number: 21 CFR 886.5918
    Lens Care Solution

    Classification Name: Rigid Gas Permeable Contact Lens Care Products (21 CFR 886.5918
    acrylate and silicone acrylate rigid gas permeable contact lenses. |
    | Classification | 21 CFR 886.5918
    | 21 CFR 886.5918 |
    | Product Code | MRC | MRC |
    | Class | Class II | Class II |
    | *Volume

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    cleadew GP hydra one is indicated for cleaning, removing protein, rinsing, disinfecting, conditioning, and storing fluoro silicone acrylate and silicone acrylate rigid gas permeable contact lenses.

    Device Description

    Ophtecs cleadew GP hydra one is a sterile, aqueous, buffered solution that contains Boric acid, Disodium Phosphate, Trometamol, Glutamic Acid, Sodium Chloride, Polyoxyethylene 9 Lauryl Ether, Tetrasodium (1-hydroxy-1,1-ethanediyl)bis(phosphonate), C12-13 Alkyl Glyceryl Hydrolyzed Hyaluronate, Sodium Hyaluronate and Hypromellose; preserved with Hydrogen peroxide (0.004%) and Polyhexamethylene biguanide hydrochloride (0.0005%). The solution is for cleaning, removing protein, rinsing, disinfecting, conditioning, and storing fluoro silicone acrylate and silicone acrylate rigid gas permeable contact lenses.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K222541

    Validate with FDA (Live)

    Date Cleared
    2022-12-06

    (106 days)

    Product Code
    Regulation Number
    886.5928
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    contact lens care product (21 CFR §886.5928),Rigid gas permeable contact lens care product (21 CFR §886.5918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bausch + Lomb Preservative Free Lubricating and Rewetting Drops are indicated for use to lubricate and rewet soft (hydrophilic) contact lenses, including silicone hydrogel as well as silicone acrylate (SA) and fluorosilicone acrylate (FSA) rigid gas permeable (RGP) contact lenses during wear. The product may be used with daily or extended wear, and disposable lenses.

    Device Description

    Bausch + Lomb Preservative Free Lubricating and Rewetting Drops is a sterile, preservative free, buffered, isotonic, aqueous solution that contains boric acid, sodium borate, potassium chloride, erythritol, poloxamine, glycerol and hyaluronan for lubricating and rewetting soft (hydrophilic) contact lenses including silicone hydrogel contact lenses and rigid gas permeable contact lenses during wear.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    The document describes the device, Bausch + Lomb Preservative Free Lubricating and Rewetting Drops, as being substantially equivalent to the predicate device, OPTI-FREE Replenish Rewetting Drops. The "acceptance criteria" are implicitly defined by demonstrating non-inferiority to the predicate device in terms of safety and effectiveness.

    Here's a table summarizing the implicit acceptance criteria (non-inferiority to the predicate) and the reported device performance, as evidenced by the clinical study:

    Acceptance Criteria (Implicit: Non-inferiority to Predicate Device)Reported Device Performance (Bausch + Lomb Preservative Free Lubricating and Rewetting Drops)
    Safety:
    Non-inferiority to predicate for slit lamp findings > Grade 2Achieved. The Bausch + Lomb Preservative Free Lubricating and Rewetting Drops group demonstrated non-inferiority to the OPTI-FREE Replenish Rewetting Drops group Over All Follow-up Visits for slit lamp findings greater than Grade 2.
    Acceptable adverse event profileMinimal. Five ocular treatment-emergent adverse events (TEAEs) in three eyes of two subjects (eye irritation (2), eye pruritus (2), eye complication associated with ocular discomfort with device (contact lens) (1)). All resolved with no action required. No discontinuations due to AEs. No corneal infiltrates or changes to pre-existing corneal scars. No notable differences in lens wear/performance parameters or visual acuities compared to control.
    Effectiveness:
    Non-inferiority to predicate for overall comfortAchieved. Treatment differences indicated that the Bausch + Lomb Preservative Free Lubricating and Rewetting Drops is non-inferior to the OPTI-FREE Replenish Rewetting Drops solution for overall comfort averaged over all follow-up visits.
    Non-inferiority to predicate for drynessAchieved. Treatment differences indicated that the Bausch + Lomb Preservative Free Lubricating and Rewetting Drops is non-inferior to the OPTI-FREE Replenish Rewetting Drops solution for dryness averaged over all follow-up visits.
    Non-inferiority to predicate for optimal (none/light) lens depositsAchieved. Treatment differences indicated that the Bausch + Lomb Preservative Free Lubricating and Rewetting Drops is non-inferior to the OPTI-FREE Replenish Rewetting Drops solution for optimal (none or light) lens deposits at all follow-up visits.
    Acceptable performance for additional effectiveness assessmentsAchieved. No notable differences between the two treatment groups with regard to additional effectiveness assessments of symptoms/complaints, worn lens characteristics, and dispensed lens characteristics.
    Compatibility with various lens typesDemonstrated. Compatible with soft contact lenses (including silicone hydrogel), rigid gas permeable, silicone acrylate (SA), and fluorosilicone acrylate (FSA) lenses, as evidenced by lens compatibility studies and inclusion of these lens types in the clinical study.

    Study Details:

    2. Sample size used for the test set and the data provenance:

    • Sample Size (Clinical Study):
      • Enrolled: 369 subjects
      • Completed: 365 subjects
      • Test Group (Bausch + Lomb Preservative Free): 188 subjects
      • Control Group (OPTI-FREE Replenish): 181 subjects
    • Data Provenance: The study was a "multicenter, randomized, masked, parallel, bilateral study conducted in the USA." This indicates prospective data collection from multiple sites within the contiguous United States.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not specify the number of experts used or their qualifications for establishing ground truth for the clinical study's endpoints. The "ground truth" for safety and effectiveness endpoints (e.g., slit lamp findings, comfort, dryness, lens deposits) would typically be assessed by the clinicians (e.g., optometrists or ophthalmologists) participating in the multicenter study.

    4. Adjudication method for the test set:

    The document does not specify an explicit adjudication method (e.g., 2+1, 3+1). Since it was a "masked" study, assessors were likely blinded to the treatment assignment. The primary endpoints were compared statistically between the two groups. It's implied that the findings from the clinicians at each site contributed to the data for analysis.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. This study is for contact lens rewetting drops, not an AI-powered diagnostic device. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is a medical product (rewetting drops) and not an algorithm or AI system.

    7. The type of ground truth used:

    For the clinical study, the ground truth was based on clinical assessments and subjective patient reporting.

    • Clinical Assessments: E.g., slit lamp findings (graded by clinicians).
    • Subjective Patient Reporting: E.g., overall comfort and dryness (self-reported by subjects).
    • Objective Observations: E.g., lens deposits (likely observed and graded by clinicians).

    8. The sample size for the training set:

    Not applicable. This is a medical product (rewetting drops), not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. As above, there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221263

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-10-12

    (163 days)

    Product Code
    Regulation Number
    886.5928
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Accessories, Soft Lens Products (21 CFR 886.5928)Products, Contact Lens Care, Rigid Gas Permeable (21 CFR 886.5918
    Accessories, Soft Lens Products (21 CFR 886.5928); Products, Contact Lens Care, Rigid Gas Permeable (21 CFR 886.5918
    Accessories, Soft Lens Products (21 CFR 886.5928); Products, Contact Lens Care, Rigid Gas Permeable (21 CFR 886.5918
    | 21 CFR886.5928;21 CFR886.5918
    | 21 CFR886.5928;21 CFR886.5918 | 21 CFR886.5918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Aqua Naina Plus Sterile Saline Solution: The Aqua Naina Plus Sterile Saline Solution for single use is indicated for rinsing soft (hydrophilic) and rigid gas permeable (hard) contact lenses prior to lens insertion. This solution may be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses. The Aqua Naina Plus Sterile Saline Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner.

    Aqua Naina Sterile Saline Solution: The Aqua Naina Sterile Saline Solution is indicated for rinsing soft (hydrophilic) and rigid gas permeable (hard) contact lenses prior to lens insertion and following proper lens disinfection as recommended by the eye care practitioner. The Aqua Naina Sterile Saline Solution may be used as needed throughout the day to rinse contact lenses and contact lens cases.

    Device Description

    Aqua Naina Plus Sterile Saline Solution is a sterile, preservative-free, buffered, isotonic, clear (colorless) aqueous solution containing isotonic solution of sodium chloride, boric acid and sodium borate in purified water USP. Aqua Naina Plus Sterile Saline Solution has pH of 7.00 to 7.40 and Osmolality of 290 to 320 mOsmol/kg. The Sterile Saline Solution is packed in single dose 1 fl oz (30 ml) high density polyethylene (HDPE) container with a tamper evident seal. Aqua Naina Plus Sterile Saline Solution is for single use only.

    Aqua Naina Sterile Saline Solution is a sterile, preservative-free, buffered, isotonic, clear (colorless) aqueous solution containing isotonic solution of sodium chloride, boric acid and sodium borate in purified water USP. Aqua Naina Sterile Saline Solution has pH of 7.00 to 7.40 and Osmolality of 290 to 320 mOsmol/kg. The Sterile Saline Solution is packed in a multidose 4 fl oz (118ml) high density polyethylene (HDPE) container with a tamper evident seal.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Aqua Naina Sterile Saline Solution and Aqua Naina Plus Sterile Saline Solution). This document outlines the device's characteristics and its substantial equivalence to existing predicate devices, but it does not contain information about acceptance criteria or specific performance studies as would be typically found for AI/ML-based diagnostic devices.

    The relevant section, "VII. PERFORMANCE DATA," explicitly states:
    "The package volume change was validated using the same methods identified in K193441. No other performance data (clinical or non-clinical) are required for this special 510(k) submission."

    This indicates that for this specific type of device (saline solution), comprehensive performance studies with acceptance criteria, sample sizes, ground truth establishment, expert reviews, and MRMC studies, as typically requested for AI/ML medical devices, were not required by the FDA for this submission. The submission relies on establishing substantial equivalence to previously cleared predicate devices based on chemical composition, intended use, and production methods, rather than demonstrating performance against specific clinical metrics.

    Therefore, I cannot provide the requested information in the format of acceptance criteria and study details because such data is not present in the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210300

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-01

    (57 days)

    Product Code
    Regulation Number
    886.5928
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: Accessories, Solution, Ultrasonic Cleaners For Lenses

    Regulation Number: 21 CFR 886.5928/21 CFR 886.5918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VueSonic One Contact Lens Cleaning System is a cleaning system for Soft and Hybrid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.

    The VueSonic Advance 2 Contact Lens Cleaning System is a cleaning system for Rigid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.

    Device Description

    The VueSonic One Contact Lens Cleaning System and VueSonic Advance 2 Contact Lens Cleaning Systems are cleaning systems for contact lenses. Both devices consist of a device body, a contact lens holder, which houses the cleaning cushions and the damping heads. The contact lens holder, cleaning cushions and damping heads are made of soft medical silicone, which will not damage the lens. The systems are designed to mimic human hand rubbing. This is done by the 3D multi-frequency modulated stereo power source and corresponding intelligent control circuit, which make a uniform three-dimensional relative movement between the damping heads and cleaning cushion.

    The proposed device consists of two versions:

    • For Rigid Lenses, VueSonic Advance 2 Contact Lens Cleaning System; and
    • For Soft and Hybrid Lenses, VueSonic One Contact Lens Cleaning System.
    AI/ML Overview

    The provided document describes the VueSonic One Contact Lens Cleaning System and VueSonic Advance 2 Contact Lens Cleaning System. It does not contain a study explicitly proving the device meets acceptance criteria. Instead, it relies on a battery of non-clinical tests and a comparison of technological characteristics with predicate devices to establish substantial equivalence.

    Therefore, many of the requested details about a specific study, acceptance criteria for performance, and ground truth establishment cannot be extracted directly from this document.

    However, I can provide the information that is present:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly define "acceptance criteria" for device performance in the form of specific thresholds or metrics for cleaning effectiveness. Instead, it lists the non-clinical tests performed, implying that satisfactory results from these tests constitute meeting the requirements for substantial equivalence. The "reported device performance" in terms of cleaning effectiveness is described through the successful completion of a "Protein and Lipid Removal Test."

    Test CategorySpecific TestRelevant Standard/GuidanceImplied Acceptance: Satisfactory Result
    Cleaning PerformanceProtein and Lipid Removal TestCitation (specific not provided)The test results demonstrate adequate protein and lipid removal, supporting the device's cleaning function (implied by "test results show that the devices are substantially equivalent").
    BiocompatibilityOcular IrritationISO 10993-10, 12Device materials are not ocular irritants.
    In Vitro Cytotoxicity TestISO 10993-5Device materials are not cytotoxic.
    SensitizationISO 10993-10Device materials do not cause sensitization.
    Systemic ToxicityISO 10993-11, 12Device materials do not cause systemic toxicity.
    Electrical Safety and EMCGeneral Requirements electromagnetic compatibilityIEC 60601-1-2The device meets electromagnetic compatibility requirements.
    General Requirements electrical safetyIEC 60601-1The device meets general electrical safety requirements.
    General Requirements home healthcare environmentIEC 60601-1-11The device meets safety requirements for home healthcare use.
    UsabilityHuman Factor Engineering and Usability Engineering Evaluation ReportIEC62366-1:2015The device is user-friendly and safe for its intended use, based on human factors principles.
    Material SafetyTotal Extractive ResiduesFDA 21 CFR 177.2600The device's materials do not release unacceptable levels of extractive residues.
    SoftwareSoftware ValidationFDA GuidanceThe software components of the device are validated and function as intended.
    Lens CompatibilityLens Compatibility TestISO 18369-2:2017The device is compatible with the specified types of contact lenses (soft, hybrid, rigid) without causing damage.

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size for the "Protein and Lipid Removal Test" or any other non-clinical test. It also does not explicitly state the country of origin of the data or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The document describes non-clinical tests, which typically rely on standardized protocols and laboratory measurements rather than expert consensus for ground truth.

    4. Adjudication method for the test set

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned. The device is a contact lens cleaning system, not an AI-powered diagnostic or interpretive tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable to a contact lens cleaning device. The "algorithm" in this context refers to the device's mechanical cleaning mechanism (3D sonic vibration technology), which operates in a "standalone" manner in that it performs its cleaning function without human intervention during the cleaning cycle. However, the overall process requires human interaction for placement of lenses and solution.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the "Protein and Lipid Removal Test," the ground truth would likely be established through quantitative laboratory measurements of protein and lipid levels on the contact lenses before and after cleaning, using standardized analytical methods. For other non-clinical tests (e.g., biocompatibility), the ground truth is based on the adherence to specified ISO standards and the absence of adverse biological reactions or safety failures.

    8. The sample size for the training set

    This information is not applicable and not provided. The device is a mechanical cleaning system and does not employ a machine learning model that would require a training set in the conventional sense.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided, as there is no mention of a machine learning model or a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200416

    Validate with FDA (Live)

    Date Cleared
    2020-09-23

    (216 days)

    Product Code
    Regulation Number
    886.5918
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: Bausch + Lomb Boston One Step Liquid Enzymatic Cleaner Regulation Number: 21 CFR 886.5918
    classification: | Class II |
    | Regulation Number: | 886.5918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bausch + Lomb Boston One Step Liquid Enzymatic Cleaner is indicated for weekly enzymatic cleaning of fluorosilicone acrylate and silicone acrylate rigid gas permeable contact lenses during (wetting, soaking, and disinfecting) with Boston Conditioning Solution or Boston ADVANCE Comfort Formula Conditioning Solution.

    Device Description

    Bausch + Lomb Boston One Step Liquid Enzymatic Cleaner is a preservative free sterile, aqueous solution containing proteolytic enzyme (subtilisin) as the active ingredient, and glycerol.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a contact lens care product, the "Bausch + Lomb Boston One Step Liquid Enzymatic Cleaner." The submission seeks to add an alternative manufacturing site and change packaging resin and raw material suppliers.

    However, the document does not contain information about acceptance criteria for a device's performance based on AI/algorithm/software output, nor does it detail a study proving the device meets such criteria. The "Preclinical testing" mentioned is related to biocompatibility and confirms the safety and effectiveness of the product itself, not a software device.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as the necessary information (specifically points 1-9 regarding an AI/algorithm-based device and its validation) is absent from the provided text.

    The text focuses on the following:

    • Device Type: A contact lens care solution (not an AI/software device).
    • Purpose of 510(k) Submission: To add an alternative manufacturing site and modify packaging components, not to introduce or validate a new AI/software feature or performance.
    • Testing Mentioned: Biocompatibility, cytotoxicity, and ocular irritation studies related to the chemical solution and its packaging components. These are standard tests for medical devices that come into contact with human tissue, not for evaluating AI performance metrics like sensitivity, specificity, or reader improvement.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201069

    Validate with FDA (Live)

    Date Cleared
    2020-07-31

    (100 days)

    Product Code
    Regulation Number
    886.5918
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Linn, OR 97068

    Re: K201069

    Trade/Device Name: VibrantVue Scleral Saline Regulation Number: 21 CFR 886.5918
    |
    | ClassificationName: | Rigid Gas Permeable Contact Lens Care Products (21 CFR 886.5918
    Indications for use

    • 트 Actions
    • Classification - Rigid Gas Permeable Contact Lens Care Products (21 CFR 886.5918
      day torinse contact lenses. |
      | Classification | 21 CFR 886.5918
      | 21 CFR 886.5918
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vibrant Vue Scleral Saline is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The VibrantVue Scleral Saline is for rinsing large diameter (scleral) rigid gas permeable (RGP) contact lenses prior to lens insertion. This solution may also be used as an insertion for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.

    Device Description

    The VibrantVue Scleral Saline is a sterile, preservative-free, unbuffered saline solution in a single dose 5 ml vial. The VibrantVue Scleral Saline is a 0.9 % saline solution that conforms to the requirements of USP saline. The rinsing solution removes loose debris and lens cleaners off rigid gas permeable (RGP) contact lenses following proper disinfection as recommended by the eye care practitioner. The sterile solution can be used to rinse contact lens cases and contact lenses as needed throughout the day.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the VibrantVue Scleral Saline:

    This document describes a medical device, VibrantVue Scleral Saline, and its clearance through the FDA's 510(k) pathway. The 510(k) process is for devices found to be "substantially equivalent" to a legally marketed predicate device, rather than requiring extensive clinical trials to prove efficacy from scratch. As such, the "acceptance criteria" discussed below relate to demonstrating substantial equivalence and meeting established standards for saline solutions and contact lens care products.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal table of "acceptance criteria" with numerical performance metrics in the way one might expect for a diagnostic or AI device. Instead, the acceptance criteria are implicit in demonstrating substantial equivalence to the predicate device and meeting recognized standards for saline solutions and contact lens care products.

    Acceptance Criterion (Implicit)Reported Device Performance
    Intended Use Equivalence: The new device's intended use should be substantially equivalent to the predicate device.The VibrantVue Scleral Saline has an identical intended use to the predicate, Menicon Saline Rinse Solution, for rinsing and insertion of large diameter RGP contact lenses, rinsing cases, and rinsing lenses throughout the day. The only difference noted is the predicate's additional indication for soft and hard contact lenses, which the new device does not claim.
    Indications for Use Equivalence: The new device's indications for use should be substantially equivalent to the predicate device.The indications for use are described as being substantially equivalent.
    Actions Equivalence: The mechanism/function of the new device should be substantially equivalent to the predicate device.The "actions" are deemed substantially equivalent. The device removes loose debris and lens cleaners.
    Classification Equivalence: The new device should fall under the same regulatory classification as the predicate.Both devices are classified under 21 CFR 886.5918 (Rigid Gas Permeable Contact Lens Care Products) and are Class II devices. (Predicate also had 21 CFR 886.5928; new device did not claim it). Both have product code MRC. (Predicate also had LPN; new device did not claim it).
    Formulation Equivalence: The new device should have a substantially equivalent formulation to the predicate, particularly for key characteristics.Both are described as sterile, preservative-free, unbuffered, single-dose, 0.9% saline solutions conforming to USP monograph. Both are supplied in 5 mL single-dose vials and plastic resin containers with twist-off caps.
    Physical Compatibility: The device should be physically compatible with currently marketed rigid gas permeable (RGP) contact lenses.Demonstrated through non-clinical studies.
    Packaging Safety: The packaging should be non-toxic and non-irritating.Demonstrated through non-clinical studies.
    Stability and Sterility: The product inside its packaging should remain stable and sterile throughout its proposed expiration date.Demonstrated through non-clinical studies.
    Safety and Effectiveness: The device must be as safe and effective as the legally marketed predicate device. This is the overarching goal of the substantial equivalence determination.Concluded based on composition and non-clinical testing results, supporting labeling directions and proposed indication.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that only non-clinical studies were conducted. There was no "test set" in the context of human patient data for the VibrantVue Scleral Saline device.

    The non-clinical studies would have involved laboratory testing of the saline solution itself, its packaging, and its interaction with contact lens materials. The data provenance for these non-clinical studies is not specified (e.g., in-house lab, third-party lab), nor are specific sample sizes for these tests detailed in this summary.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. No clinical test set involving human data was used, therefore no experts were required to establish ground truth for such a set. The "ground truth" for the non-clinical tests would be defined by established scientific methods, standards (e.g., USP monograph for 0.9% saline), and regulatory requirements for sterility, compatibility, and toxicity.


    4. Adjudication Method for the Test Set

    Not applicable. As there was no clinical test set requiring human interpretation, no adjudication method was used. Non-clinical studies rely on standardized test protocols and objective measurements.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used to assess human reader performance, often with and without AI assistance, for diagnostic or interpretive tasks. The VibrantVue Scleral Saline is a contact lens care solution, not a diagnostic or AI-powered device, so such a study is not relevant or required for its clearance.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. The VibrantVue Scleral Saline is a physical medical device (a saline solution), not an algorithm or AI system. Therefore, "standalone (algorithm only)" performance is not relevant.


    7. The Type of Ground Truth Used

    For the non-clinical studies, the "ground truth" was established based on:

    • USP Monograph for 0.9% Saline: This defines the chemical composition and purity standards for saline.
    • Industry Standards and Regulatory Requirements: For sterility, packaging safety (non-toxicity, non-irritating), and physical compatibility with RGP contact lenses. These are typically objective, measurable criteria.

    8. The Sample Size for the Training Set

    Not applicable. As VibrantVue Scleral Saline is a physical medical device and not an AI/ML algorithm, there was no "training set."


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200747

    Validate with FDA (Live)

    Date Cleared
    2020-05-15

    (53 days)

    Product Code
    Regulation Number
    886.5928
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    , Soft Lens Products Products, Contact Lens Care, Rigid Gas Permeable |
    | Regulation Number: | 886.5918
    |
    | Regulation Number | 886.5918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PuriLens Plus Preservative-Free Saline is indicated for use following proper lens disinfection as recommended by the eye care practitioner. PuriLens Plus Preservative-Free Saline is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.

    Device Description

    The PuriLens Plus Preservative Free Saline is a sterile, buffered saline solution for rinsing and storing contact lenses. The solution is buffered to roughly physiological pH and has an osmolarity similar to that of human tears. The product is sold over-the-counter in various packaging configurations.

    The current generation of the product, Purilens Saline Solution, was cleared under K002319 in 2000, and has been legally marketed in the US since that time. This 510(k) application is to expand the indications for use to include use with rigid gas permeable lenses. There is no change in the device design, formulation, or manufacturing. The labeling will be revised with the new indications.

    AI/ML Overview

    This submission describes a 510(k) for a contact lens solution, PuriLens Plus Preservative Free Saline, and argues for its substantial equivalence to a predicate device based on similar formulation, manufacturing, and packaging, and expands its indications for use.

    Since this is a submission for substantial equivalence based on a predicate device and there have been no changes to the device design, formulation, or manufacturing, no performance data or studies are presented. Therefore, I cannot provide information on acceptance criteria, specific study details, sample sizes, ground truth establishment, or expert involvement as these were not required or reported in this 510(k) summary.

    The summary states: "Because the subject is identical to the predicate in terms of formulation, manufacturing, and packaging, no bench-testing is required to demonstrate substantial equivalence." (Page 5)

    As such, for all the requested points regarding acceptance criteria, performance data, sample sizes, ground truth, expert involvement, and comparative effectiveness studies, the answer is that this information is not applicable or not provided in this 510(k) summary due to the nature of the submission (substantial equivalence based on an identical product with expanded indications, requiring no new bench testing).

    Ask a Question

    Ask a specific question about this device

    K Number
    K191872

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-02

    (234 days)

    Product Code
    Regulation Number
    886.5928
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Classification name: | 21 CFR 886.5928Soft (hydrophilic) Contact Lens Care Product21 CFR 886.5918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution is indicated for use in cleaning, daily protein removal, disinfection, and storing of soft (hydrophilic) contact lenses (including silicone hydrogel lenses) and rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses, as recommended by your eye care professional.

    Device Description

    The Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution system consists of: Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution and the special lens case. The lens case consists of a transparent cup and a lens holder-lens case cap assembly with a neutralizer catalyst disk. Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution and the special lens case must always be used together.

    The preservative free, aqueous Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution contains: hydrogen peroxide 3%, phosphonic acid compound (stabilizer), citric acid, polyethylene glycol, sodium chloride, sodium hydroxide, and phosphate (buffer system).

    AI/ML Overview

    The provided document describes the FDA 510(k) clearance for Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution. Here's a breakdown of the requested information based on the text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical "acceptance criteria" for performance metrics in a table. Instead, it states that tests were conducted according to established guidance and protocols, and the results "demonstrated" effectiveness or compatibility. The general acceptance criterion for most tests appears to be "similar to the predicate device" or "effective/compatible."

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Lens CompatibilityCompatible with various commercially available hydrogel, silicone hydrogel, and RGP lenses."Results demonstrated that the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution was compatible with the test contact lenses."
    Neutralization ProfileSimilar neutralization profile to the predicate device; neutralizer catalyst disk stability."The Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution showed a similar neutralization profile to the predicate device. In addition, stability of the neutralizer catalyst disk was confirmed for 90-cycles treatments."
    Disinfecting EfficacyHarmful microorganisms reduced to an ISO level at specified time points."Results demonstrated that harmful microorganisms were effectively reduced to an ISO level at 6 and 24 hours and 7 days after neutralization."
    Preservative EfficacyEnsure 3-month open bottle discard date."Preservative Efficacy test was conducted to ensure 3-month open bottle discard date of the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution."
    Cleaning EfficacySimilar performance to the predicate device in protein and lipid removal."Results suggested that both solutions performed similarly in the removal of lens deposition."
    Toxicology/BiocompatibilityNon-toxic and non-irritating for the neutralized solution, lens case, and catalyst."Results demonstrated that the neutralized solution, the lens case materials and the neutralizer catalyst were non-toxic and nonirritating under the experimental conditions."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes (e.g., number of lenses, microorganisms, or test runs) for each non-clinical test. It mentions testing with:

    • Lens Compatibility: Six commercially available hydrogel/silicone hydrogel soft contact lenses and two commercially available RGP lenses.
    • The data provenance (country of origin, retrospective/prospective) and specific details about the laboratories conducting these non-clinical tests are not provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is a cleaning and disinfecting solution, not an AI or diagnostic imaging device that requires human expert review for "ground truth." The "ground truth" for its performance is established through objective laboratory tests and scientific protocols that measure chemical properties, microbiological reduction, and material compatibility. Therefore, the concept of "experts establishing ground truth" in the way it applies to AI diagnostic studies is not relevant here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or AI diagnostic evaluations where human readers interpret data. This is a non-clinical, laboratory-based performance study for a chemical solution. Therefore, such adjudication methods are not applicable and not mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This is a chemical cleaning solution, not an AI-assisted diagnostic device, so comparison with human readers is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to an algorithm's performance. As the device is a chemical solution, this concept is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance of this device is established through:

    • Scientific Protocols and Laboratory Measurements: For lens compatibility, neutralization profiles (residual hydrogen peroxide levels), disinfecting efficacy (microbial reduction counts against ISO standards), preservative efficacy, cleaning efficacy (quantification of protein/lipid removal), and toxicology (in-vitro and in-vivo biocompatibility tests based on GLP regulations). This is essentially objective, measurable data generated by laboratory instruments and techniques following validated methods.

    8. The sample size for the training set

    This device is not an AI algorithm requiring a "training set." Therefore, the concept of a training set sample size is not applicable. The performance studies are validation tests, not machine learning model training.

    9. How the ground truth for the training set was established

    As there is no training set for an AI algorithm, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190674

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-08

    (146 days)

    Product Code
    Regulation Number
    886.5928
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Lens Care Products (21 CFR886.5928) and Rigid Gas Permeable Contact Lens CareProducts (21 CFR 886.5918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    REFRESH® RELIEVA™ FOR CONTACTS is indicated to lubricate and rewet soft and rigid gas permeable contact lenses, to help relieve dryness, discomfort and irritation that may be associated with lens wear and to cushion lenses by placing a drop on the lens prior to the application on the eye.

    Device Description

    REFRESH® RELIEVA™ FOR CONTACTS is a sterile, buffered, isotonic, preserved solution. This aqueous formulation includes carboxymethylcellulose sodium, glycerin, boric acid, calcium chloride dihydrate, erythritol, magnesium chloride hexahydrate, potassium chloride, purified water, sodium borate decahydrate, sodium citrate dihydrate, sodium hyaluronate and the preservative PURITE® (stabilized oxychloro complex 0.01%). The solution may also contain hydrochloric acid and/or sodium hydroxide to adjust pH. All ingredients except for Purite are compendial grade (USP/NF/Ph Eur). The product is supplied in a teal colored, low density polyethylene (LDPE) multi-dose bottle and tip with a teal colored high impact polystyrene (HIPS) cap. The bottles are further packaged in paper cartons with a patient information insert.

    AI/ML Overview

    This document is a 510(k) Summary for the REFRESH® RELIEVA™ FOR CONTACTS device. It describes the device, compares it to predicate devices, and summarizes testing performed to support its substantial equivalence. The document is a regulatory submission for a medical device and does not inherently describe a study that uses AI, nor does it provide detailed acceptance criteria related to AI performance.

    Therefore, many of the requested sections (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable as this document pertains to a medical device (contact lens solution), not an AI/ML powered device.

    However, I can extract the information relevant to the performance of the device as described in this document.


    Device Name: REFRESH® RELIEVA™ FOR CONTACTS

    Device Description: A sterile, buffered, isotonic, preserved solution containing carboxymethylcellulose sodium, glycerin, boric acid, calcium chloride dihydrate, erythritol, magnesium chloride hexahydrate, potassium chloride, purified water, sodium borate decahydrate, sodium citrate dihydrate, sodium hyaluronate and the preservative PURITE® (stabilized oxychloro complex 0.01%).

    Indications for Use: To lubricate and rewet soft and rigid gas permeable contact lenses, to help relieve dryness, discomfort and irritation that may be associated with lens wear and to cushion lenses by placing a drop on the lens prior to the application on the eye.

    Predicate Devices:

    • REFRESH® Contacts Lubricating and Rewetting Drops (K992028)
    • REFRESH OPTIVE CONTACTS Lubricating and Rewetting Drops (K083812)

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" for performance in the typical sense of numerical thresholds for sensitivity, specificity, or similar metrics for an AI/ML device. Instead, the "acceptance criteria" are implied by the requirement for "substantial equivalence" to predicate devices, which the studies aimed to demonstrate. The performance is reported in terms of safety and efficacy.

    Acceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device Performance
    Stability: Chemical, physical, and microbiological stability to support the claimed shelf-life and preservative effectiveness.Demonstrated: Product showed chemical and physical stability for a 24-month shelf-life. Microbiology testing confirmed sterility maintenance and preservative effectiveness throughout the shelf-life.
    Biocompatibility: Safety and effectiveness according to FDA guidance for contact lens care products.Demonstrated: In-vitro and in-vivo studies, conducted per FDA guidance (Premarket Notification [510(k)] Guidance Document for Contact Lens Care Products dated 1 May 1997), indicated physical, chemical, and microbiological properties are substantially equivalent to predicate devices.
    Clinical Performance: Safety and efficacy for lubricating, rewetting, and cushioning contact lenses, comparable to marketed predicate devices.Demonstrated: A 3-month clinical study showed substantial equivalence to REFRESH CONTACTS® for lubricating and rewetting during lens wear, and cushioning upon lens insertion.
    Technological Characteristics: No new significant safety or effectiveness concerns compared to predicate devices.Demonstrated: Components are not novel, similar concentrations to cleared products, no additional significant safety or effectiveness concerns identified. The intended use is identical to predicate devices.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set:

      • Clinical Study: The document states "A 3-month clinical study was conducted... in contact lens wearers." However, the exact number of participants (sample size) in this clinical study is not provided in the available text.
      • Biocompatibility/Stability: Not applicable as these are bench/lab tests, not human test sets.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The clinical study was conducted to evaluate the device against a marketed predicate, implying a controlled clinical trial setting. The studies are prospective as they were specifically performed to support the 510(k) submission.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This section is not applicable as the described studies are for a contact lens solution, not an AI/ML powered device where expert readers typically establish ground truth. Clinical endpoints in this context would likely be based on participant reporting of symptoms and potentially objective ophthalmological assessments, not "expert ground truth" for image interpretation.


    4. Adjudication Method for the Test Set

    This section is not applicable for the reasons stated above. Clinical study endpoints would be collected and analyzed, but "adjudication" in the sense of resolving discrepancies between expert interpretations of data (as typically seen in AI/ML performance studies) is not relevant here.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. This document describes the safety and effectiveness testing for a contact lens comfort solution, not an AI/ML diagnostic or assistive device. There are no "human readers" interpreting images with or without AI assistance in this context.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable. The device is a physical product (contact lens solution), not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the clinical study, the "ground truth" or primary endpoints would likely be based on:

    • Patient-reported outcomes (PROs): Subjective relief of dryness, discomfort, and irritation; comfort upon lens insertion.
    • Clinical assessments: Potentially objective measures of ocular surface health or lens-eye interaction, as assessed by an optometrist or ophthalmologist, though specific measures are not detailed in this summary.
    • The study aimed to demonstrate substantial equivalence in safety and efficacy compared to a predicate device, meaning its performance was judged against the established acceptable performance of the predicate.

    For biocompatibility, the ground truth involves adherence to and successful completion of established test methods (e.g., cytotoxicity, irritation, sensitization) as per FDA guidance.

    For stability, the ground truth involves chemical, physical, and microbiological specifications being met over time.


    8. The sample size for the training set

    This section is not applicable. The device is a contact lens solution, not an AI/ML model that requires a training set.


    9. How the ground truth for the training set was established

    This section is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191795

    Validate with FDA (Live)

    Date Cleared
    2019-07-31

    (28 days)

    Product Code
    Regulation Number
    886.5928
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Classification: | Class II (21 CFR 886.5928) Soft (hydrophilic) contact lens careproducts; (21 CFR 886.5918

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For storage of soft (hydrophilic), and rigid gas permeable (RGP) hard contact lenses during disinfection with buffered 3% hydrogen peroxide contact lens solution.

    Device Description

    The device is a specialized lens case consisting of a transparent cup with a connected unit of screw cap, lens holders (baskets) and platinum-coated neutralization disc. The platinum catalyst coating and shape of the plastic disc are designed to effectively neutralize hydrogen peroxide in the lens care system to harmless water and oxygen gas.

    AI/ML Overview

    This looks like a 510(k) summary for a medical device with a minor modification (change in silicone material of a gasket). Generally, devices with such minor changes do not contain extensive studies as they are not typically required for demonstrating substantial equivalence. The document confirms this in Section VII, "Performance Data," stating that the device was evaluated in "biocompatibility, physicochemical and design functionality tests."

    Given the information provided in the 510(k) summary:

    • No extensive clinical study was conducted or is reported for the device. The primary purpose of this 510(k) (K191795) is to demonstrate substantial equivalence to a legally marketed predicate device after a change in a gasket material.
    • The "performance data" mentioned refers to nonclinical testing, which is generally not a clinical trial involving human subjects or AI algorithm performance as you might expect in more complex device submissions.
    • The information you are asking for (e.g., sample size for training set, MRMC study, number of experts for ground truth) is typical for AI/ML-driven devices or devices requiring extensive clinical validation. This device (AOCup Lens Case with AODisc) is a contact lens case, which falls under a different regulatory pathway and testing requirements.

    Therefore, I cannot provide the detailed information you requested about acceptance criteria and study particulars in the format you specified, as those types of studies and criteria are not described in this 510(k) submission for this specific device. The provided text outlines a regulatory submission for a contact lens case after a material change, not an AI or imaging-based diagnostic device.

    However, I can interpret the "Performance Data" section to infer general acceptance criteria and the "study" (nonclinical testing) that proves the device meets them:

    General Acceptance Criteria and "Study" (Nonclinical Testing) for AOCup Lens Case with AODisc (K191795)

    The primary acceptance criterion for this 510(k) submission is to demonstrate that the modified device (with the new silicone gasket material) maintains the same safety and effectiveness as the predicate device. This is achieved through nonclinical testing to ensure the material change does not negatively impact the device's intended function or biocompatibility.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Criteria (Inferred)Reported Device Performance (Inferred)
    BiocompatibilityThe new gasket material must be biocompatible (e.g., non-cytotoxic, non-irritating, non-sensitizing) and not leach harmful substances."The AOCup Lens Case with AODisc and the changed gasket was evaluated in biocompatibility... tests." (Implicitly, these tests were successful as the submission concludes substantial equivalence and safety.)
    Physicochemical PropertiesThe new gasket material must maintain appropriate physical and chemical properties (e.g., durability, chemical stability) for its function as a seal in the lens case, particularly in contact with hydrogen peroxide solution."The AOCup Lens Case with AODisc and the changed gasket was evaluated in... physicochemical... tests." (Implicitly, these tests were successful, indicating the material change did not compromise the physical or chemical integrity of the gasket or its interaction with the solution.)
    Design FunctionalityThe device, with the new gasket, must continue to function as intended (e.g., seal properly, prevent leaks, accommodate lenses, allow neutralization)."The AOCup Lens Case with AODisc and the changed gasket was evaluated in... design functionality tests." "Dimensions and technical function of the gasket are unchanged." "The modifications of gasket material does not change any indications for use nor the basic technical principle of the device functions." (Implicitly, these tests were successful, confirming the gasket still performs its sealing role and the overall lens case maintains its designed function for storage and disinfection.)
    Substantial Equivalence (Overall)The modified device must be as safe and effective as the predicate device, and the material change must not raise new questions of safety or effectiveness."Successful results of all nonclinical testing supported the substantial equivalence and therefore safety and efficacy of the modified AOCup Lens Case with AODisc to the existing product for its intended use."

    Regarding the other points in your request:

    • 2. Sample sized used for the test set and the data provenance: Not applicable. These were nonclinical (bench) tests, not involving a "test set" in the context of clinical data or AI algorithms.
    • 3. Number of experts used to establish the ground truth... and qualifications: Not applicable. Ground truth for nonclinical tests is based on established testing standards and specifications, not expert interpretation of data.
    • 4. Adjudication method: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No. This is not an imaging or diagnostic device that would require such a study.
    • 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No. This device does not involve an algorithm.
    • 7. The type of ground truth used: For biocompatibility, physicochemical, and design functionality tests, the "ground truth" would be defined by the established industry standards, test methods (e.g., ISO standards for biocompatibility), and internal product specifications for the performance of the gasket and the lens case.
    • 8. The sample size for the training set: Not applicable. There is no training set for this type of device.
    • 9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document describes a regulatory submission for a physical medical device (contact lens case) with a minor material change. The "study" refers to nonclinical laboratory and bench testing rather than clinical trials or AI performance evaluations, therefore, most of your requested information is not relevant to this specific type of device and submission.