Name:** CaseBio™ Culture w/HSA (CMH5); CaseBio™ Handling w/HSA (WHH5)
Regulation Number: 21 CFR 884.6180
Reproduction Media
Regulation Name: Reproductive Media and Supplements
Regulation Number: 884.6180
CaseBio™ Culture w/HSA is intended for use for culture of embryos from fertilization to the blastocyst stage. This device can be used for transfer of embryos to the uterus.
CaseBio™ Handling w/HSA is intended for use in short-term handling and manipulating gametes and embryos, including washing and intracytoplasmic sperm injection (ICSI). This medium is not intended for transferring embryos to the uterine cavity.
CaseBio™ Culture w/HSA (CMH5) and CaseBio™ Handling w/HSA (WHH5) are intended for use during in vitro procedures for gamete/embryo culture and one step media.
CaseBio™ Culture w/HSA is intended for use for culture of embryos from fertilization to the blastocyst stage. This device is a one-step media and can be used for transfer of embryos to the uterus.
CaseBio™ Handling w/HSA is intended for use in short-term handling and manipulation of gametes and embryos, including washing and intracytoplasmic sperm injection (ICSI). This medium is not intended for transferring embryos to the uterine cavity.
CaseBio™ Culture w/HSA (CMH5) and CaseBio™ Handling w/HSA (WHH5) have almost identical composition with differences in amount of NaHCO3 and Phenol red and presence of HEPES in CaseBio™ Handling w/HAS. Both formulations contain gentamicin and human serum albumin.
The two solutions are aseptically filtered (storage vials sterilized by radiation) and provided in PETG square media bottles with HDPE screw-style caps. They have a shelf-life of 90 days when stored at 2-8°C. Media in the vials can be used for up to seven days after vial opening.
N/A
Ask a specific question about this device
(210 days)
California 92705
Re: K250445
Trade/Device Name: Fast Warm - NX
Regulation Number: 21 CFR 884.6180
blastocysts
Regulation Name: Reproductive Media and Supplements
Regulation Number: 21 CFR 884.6180
Fast Warm - NX is intended for use in assisted reproductive procedures for the thawing of vitrified human blastocyst stage embryos.
Fast Warm – NX is a dual-buffered solution (HEPES & MOPS) of Continuous Single Culture Medium (CSCM) containing gentamicin sulfate, 20% (v/v) dextran serum supplement (DSS), and 1.0M trehalose. DSS is a protein supplement consisting of 50 mg/mL human serum albumin (HSA) and 20 mg/mL dextran. DSS is used at 20% (v/v) in Fast Warm – NX for a final concentration of 10 mg/mL HSA and 4 mg/mL dextran. The product is supplied as a ready to use liquid in two (2) different configurations, a kit containing multiple vials of either two (2) x 20mL or six (6) x 2mL. The medium can be used for up to 14 days after bottle opening.
N/A
Ask a specific question about this device
(90 days)
Flushing Solution; Dewin One Step medium [with HSA and without HSA])
Regulation Number: 21 CFR 884.6180
Reproduction Media
Regulation Name: Reproductive Media and Supplements
Regulation Number: 884.6180
Dewin Gamete Buffer is intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI).
Dewin Follicle Flushing Solution is intended for use during ovarian follicle flushing and oocyte collection procedures for use in in vitro fertilization procedures.
Dewin One Step Medium is intended for use for culture of embryos from fertilization to the blastocyst stage. This device can be used for transfer of embryos to the uterus.
The DonneVie Dewin Reproductive Media are solutions used in assisted reproductive procedures. The family of Reproductive Media include the following 3 types and their variants:
- Dewin Gamete Buffer [with and without Human Serum Albumin (HSA)];
- Dewin Follicle Flushing Solution;
- Dewin One Step Medium [with and without HSA].
All variants contain gentamicin, an antibiotic agent that suppresses bacterial growth. The Gamete Buffer and Follicle Flushing media are offered in 100mL volume, while the One Step Medium is offered in volumes of 25mL and 50mL. The three Dewin Reproductive Media are used together during an assisted reproductive procedure. The Follicle Flushing Solution is used to first flush/wash the Cumulus-Oocyte Complex after it's removed from the ovary; it is then transferred to the Gamete Buffer for cleaning. After the prepared oocyte has been fertilized, it is transferred to the One Step Medium to culture the embryo from fertilization to blastocyst stage.
The primary glass bottle containers are washed, then sterilized and depyrogenated via dry heat. The bottle caps are supplied sterile by electron beam irradiation. The media are sterile filtered through a 0.2µm filter in an aseptic isolator at DonneVie Medical Technology Co., Ltd.
The provided FDA 510(k) clearance letter describes a medical device called "Dewin Reproductive Media" and presents non-clinical performance testing to demonstrate its substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Dewin Reproductive Media are primarily derived from the specifications compared to the predicate device and the results expected from specific non-clinical tests.
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance | Comments |
|---|---|---|---|
| Appearance | Clear, particulate free | Met after real-time aging (Shelf-life testing) | Confirmed through visual inspection as part of shelf-life testing. |
| pH (per USP <791>) | 7.2 - 7.5 | Met after real-time aging (Shelf-life testing) | Within the specified range after aging. |
| Osmolality (per USP <785>) | 260 - 295 mOsm/kg | Met after real-time aging (Shelf-life testing) | Within the specified range after aging. |
| Endotoxin (per USP <85>) | < 0.25 EU/mL | Met after real-time aging (Shelf-life testing) | Below the specified limit after aging. |
| Sterility (per USP <71>) | No microbial growth | Met after real-time aging (Shelf-life testing) | Confirmed no microbial growth after aging. |
| Mouse Embryo Assay (MEA) for Dewin One Step Medium | ≥ 80% embryos developed to expanded blastocyst at 120 hours | Met after real-time aging (Shelf-life testing) | Demonstrated equivalent or better performance than predicate specifications. |
| Mouse Embryo Assay (MEA) for Dewin Gamete Buffer | ≥ 80% developed to expanded blastocyst at 96 hours after 30 min exposure | Met after real-time aging (Shelf-life testing) | Demonstrated equivalent or better performance than predicate specifications. |
| Mouse Embryo Assay (MEA) for Dewin Follicle Flushing Solution | ≥ 80% embryos developed to expanded blastocyst at 96 hours after 30 min exposure | Met after real-time aging (Shelf-life testing) | Demonstrated equivalent or better performance than predicate specifications. |
| Sterilization Process (Aseptic Filtration) | Aseptic filtration validated per ISO 13408-1:2008 & A1:2013 and ISO 13408-2:2018. | Validated, with filter challenge test using worst-case conditions (excluding antimicrobials). | Ensures sterility of the media. |
| Sterilization Process (Glass Bottles) | Dry heat sterilization per 20857:2010/(R)2015. | Validated. | Ensures sterility of the primary packaging. |
| Sterilization Process (Bottle Caps) | Radiation sterilization per ISO 11137-1:2006 and ISO 11137-2:2013. | Validated. | Ensures sterility of the bottle closures. |
| Shelf-life | Gamete Buffer: 7 months; Follicle Flushing Solution: 12 months; One Step Medium: 4 months. | Met for respective products (Shelf-life testing) | Performance maintained for the specified duration. |
| Transportation Testing | Resistance to transportation stresses per ASTM D4169-22. | Performed. | Ensures product integrity during shipping. |
| Cap/Seal Leak Testing | Method equivalent to USP <1207.2> on transportation-conditioned devices. | Performed. | Ensures product integrity and sterility is maintained. |
| Biocompatibility (Dewin Follicle Flushing Solution & Dewin One Step Medium) | Cytotoxicity (ISO 10993-5:2009); Guinea Pig Maximization Sensitization Test (ISO 10993-10:2021); Intracutaneous Reactivity (ISO 10993-23:2021). | Results support biocompatibility. | Confirmed safety for patient contact. |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not specify the exact sample sizes for most of the non-clinical tests conducted. For example, it doesn't state how many batches were tested for sterility, pH, osmolality, or endotoxin, nor does it specify the number of mouse embryos used in the MEA tests.
- Provenance: The studies were conducted by DonneVie Medical Technology (Shanghai) Co. Ltd. This implies the data provenance is likely from China, as that is the submitter's location. The studies are non-clinical, meaning they are laboratory tests, not human trials. They are implicitly prospective in the sense that they were designed and executed to test the device's performance against predefined criteria.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This section is not applicable as the studies described are non-clinical laboratory tests, not studies requiring expert interpretation of medical images or patient outcomes. The "ground truth" for these tests (e.g., pH value, osmolality, microorganism presence, blastocyst development) is established through established laboratory methods and validated instrumentation, not by expert consensus in a clinical diagnostic context.
4. Adjudication Method for the Test Set
This section is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies, particularly in cases where human readers or diagnosticians are involved and their interpretations need to be reconciled to establish a ground truth. For laboratory tests, the results are typically quantitative or qualitative outcomes determined by a predefined method, eliminating the need for expert adjudication in this typical sense.
5. If a Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. The device is a "Reproductive Media," which are solutions used in assisted reproductive procedures (e.g., for washing gametes, culturing embryos). It is not an AI-powered diagnostic device or an imaging system. Therefore, no MRMC study involving human readers or AI assistance would be relevant or conducted for this type of product.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This section is not applicable. The device is a reproductive media solution, not an algorithm or an AI system. Its performance is evaluated through physical, chemical, and biological laboratory tests, not through algorithmic output.
7. The Type of Ground Truth Used
The ground truth for the non-clinical tests is established through:
- Physicochemical Measurements: For parameters like pH, osmolality, and endotoxin, the ground truth is the numerical value obtained using validated analytical instruments and methods (e.g., pH meter, osmometer, LAL assay).
- Biological Endpoints: For the Mouse Embryo Assay (MEA), the ground truth is the observed development of mouse embryos to the expanded blastocyst stage within a specified timeframe, as assessed under a microscope by trained personnel following a standardized protocol.
- Microbiological Endpoints: For sterility testing, the ground truth is the absence of microbial growth in culture, determined by standard microbiological techniques.
- Regulatory Standards: Compliance with ISO standards for sterilization, biocompatibility, and transportation testing.
8. The Sample Size for the Training Set
This section is not applicable. The Dewin Reproductive Media is a chemical/biological product, not a machine learning model or AI system that requires a "training set." The product's formulation and manufacturing processes are developed through research and development, not by training an algorithm on a dataset.
9. How the Ground Truth for the Training Set was Established
This section is not applicable as there is no "training set" for this type of device.
Ask a specific question about this device
(120 days)
Re: K251305
Trade/Device Name: Ultra-Fast Vitri; Ultra-Fast Warm
Regulation Number: 21 CFR 884.6180
vitrified oocytes
Regulation Name: Reproductive media and supplements
Regulation Number: 884.6180
Reproductive | Class II Media, Reproductive |
| Product Code | MQL | MQL, MQK |
| Regulation No. | 21 CFR 884.6180
| 21 CFR 884.6180 |
| Manufacturer | Kitazato Corporation | Kitazato Corporation |
| Intended Use |
Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII).
Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes(MII).
The Ultra-Fast Vitri and Ultra-Fast Warm is composed of a set of three media to vitrify and warm oocytes for assisted reproductive technology (ART) procedures.
The Ultra-Fast Vitri includes two components, Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol and dimethyl sulfoxide. There are two vitrification procedures to choose from. During the vitrification process, oocytes are first exposed to ES then in VS within several minutes. Using this methodology, permeating cryoprotectants can replace water in the oocytes prior to vitrification and storage in liquid nitrogen. The Ultra-Fast Vitri comes prepackaged with 1.5 mL vial or 4 mL vial of ES, three 1.5 mL vials or three 4 mL vials of VS.
Ultra-Fast Warm is composed of one media used for warming and removing cryoprotectants from vitrified oocytes. It is composed of Thawing Solution (TS). The Ultra-Fast Warm comes pre-packaged with four 4.0 ml vials of TS.
All the media in the Ultra-Fast Vitri and Ultra-Fast Warm contain Gentamicin. The media undergoes aseptic filtration, while the vials are sterilized by radiation.
Based on the provided FDA 510(k) Clearance Letter, here's a description of the acceptance criteria and the study that proves the device meets them:
Device: Ultra-Fast Vitri; Ultra-Fast Warm
Description: A set of three media used for the vitrification (cryopreservation) and warming of oocytes (MII) for Assisted Reproductive Technology (ART) procedures.
Acceptance Criteria and Reported Device Performance
The core acceptance criteria for this device, as demonstrated through non-clinical and clinical performance data, revolve around its biological compatibility and effectiveness in preserving and recovering oocytes without compromising their viability or subsequent reproductive outcomes.
| Acceptance Criteria Category | Specific Metric (Unit) | Acceptance Criteria | Reported Device Performance (Ultra-Fast Vitri/Warm) |
|---|---|---|---|
| Non-Clinical Performance | |||
| Color/Appearance | Visual inspection | Acceptable appearance | Passed |
| pH Testing | pH value (range) | 7.20 – 7.60 | Passed (7.20 – 7.60) |
| Endotoxin Testing | Endotoxin level (EU/mL) | <0.25 EU/mL (LAL methodology) | Passed (<0.25 EU/mL LAL methodology) |
| Osmolality Testing | Osmolality value | (Not explicitly stated in document, but implied as conforming to media standards) | Passed |
| Sterility Testing | Sterility | Passes USP <71> | Passed (Passes USP <71>) |
| Gentamicin Test | Gentamicin presence/level | (Not explicitly stated, but implied as conforming to specification) | Passed |
| Initial Media Dispensing Validation | Functional dispensing/packaging | (Not explicitly stated, but implied as successful) | Passed |
| Mouse Embryo Assay (MEA) | One-cell embryo development (96 hours) | >80% | Passed (>80%) |
| Biocompatibility | Biocompatibility with cells | Passes | Passes |
| Storage Stability | Temperature range (°C) | 2 – 8°C | 2 – 8°C |
| Shelf Life | Duration (months) | 12 months | 12 months |
| Clinical Performance | |||
| Oocyte Survival Rate | % | (Implied to be comparable to conventional protocol) | 100.0% (Ultra-Fast) vs. 90.9% (Conventional) |
| Clinical Pregnancy Rate | % | (Implied to be comparable to conventional protocol) | 65.2% (Ultra-Fast) vs. 54.3% (Conventional) |
| Live Birth Rate | % | (Implied to be comparable to conventional protocol) | 56.5% (Ultra-Fast) vs. 52.2% (Conventional) |
Study Proving Device Meets Acceptance Criteria
The provided document describes both non-clinical (bench) and clinical performance studies to demonstrate the safety and effectiveness of the Ultra-Fast Vitri and Ultra-Fast Warm device and its substantial equivalence to the predicate device.
1. Non-Clinical Performance Data (Bench Testing):
- Description: A series of laboratory tests conducted directly on the media to confirm its physical, chemical, and biological properties.
- Specific Tests: Color/Appearance, pH Testing, Endotoxin testing, Osmolality Testing, Sterility Testing, Gentamicin Test, Initial Media Dispensing Validation, Mouse Embryo Assay (MEA), and Biocompatibility.
- Proof of Concept: The device passed all these tests, including achieving >80% one-cell development in the Mouse Embryo Assay, which is a critical biological performance indicator for reproductive media as per FDA guidance.
2. Clinical Performance Data:
- Study Design: A comparative study referenced from literature that evaluated the effectiveness of the ultra-fast vitrification and warming protocols using the subject device against conventional protocols (presumably using the predicate or similar conventional vitrification/warming solutions).
- Study Objective: To demonstrate comparable outcomes (oocyte survival, clinical pregnancy, live birth rates) between the ultra-fast protocol and conventional protocols.
Here's a breakdown of the specific requested information about the clinical study:
-
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: 1,077 mature oocytes in total.
- 519 oocytes for the conventional vitrification and warming protocols group.
- 558 oocytes for the ultra-fast protocols (subject device) group.
- Data Provenance: The document states "The referenced literature used Kitazato's vitrification and warming solutions (K171748 and K160864)".
- It does not explicitly state the country of origin of the data.
- It does not explicitly state if the study was retrospective or prospective. However, given it's a "study" comparing protocols, it's typically prospective, but this cannot be confirmed from the text.
- Sample Size: 1,077 mature oocytes in total.
-
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- This information is not provided in the document. The study focuses on clinical outcomes (survival, pregnancy, live birth rates) rather than human interpretation of images or other subjective assessments that would require expert consensus for ground truth.
-
4. Adjudication Method for the Test Set:
- This information is not applicable/not provided. Adjudication methods (like 2+1, 3+1) are typically relevant for studies involving human interpretation where reviewer disagreement needs to be resolved (e.g., radiology studies). This study measures biological/clinical outcomes.
-
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
- No, an MRMC study was not done. MRMC studies are used to assess the impact of a device (often AI) on human reader performance, typically in diagnostic imaging. This study evaluated the direct clinical effectiveness of the media itself.
- Therefore, an effect size of how much human readers improve with AI vs. without AI assistance is not applicable.
-
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:
- Yes, in a sense, the clinical study assessed the "standalone" performance of the media with its associated protocols. It compared the outcomes from using the media (with its specific ultra-fast protocol) to conventional media/protocols. There isn't an "algorithm" in the traditional sense for this device; it's a chemical formulation and protocol. The "performance" is the biological outcome achieved by the oocytes.
-
7. The Type of Ground Truth Used:
- The ground truth was based on clinical outcomes data:
- Oocyte survival rate after vitrification and thawing.
- Clinical pregnancy rate (following embryo transfer resulting from these oocytes).
- Live birth rate (following clinical pregnancy).
- These are considered objective biological and clinical endpoints.
- The ground truth was based on clinical outcomes data:
-
8. The Sample Size for the Training Set:
- This information is not applicable/not provided. This device is a media (consumable), not an AI algorithm that requires a separate "training set" of data. The "development" of the media and protocols would be based on laboratory research and refinement rather than a data training paradigm.
-
9. How the Ground Truth for the Training Set Was Established:
- This information is not applicable/not provided for the same reasons as #8.
Ask a specific question about this device
(269 days)
Flushing medium; FertiCult Flushing medium with phenol red and gentamicin
Regulation Number: 21 CFR 884.6180
Reproduction Media
Regulation Name: Reproductive Media and Supplements
Regulation Number: 884.6180
FertiCult™ Flushing medium are intended for in vitro procedures involving human gametes (sperm and oocytes), including washing of gametes, sperm swim-up procedures, intra-uterine insemination (IUI) of the spermatozoa and sperm injection during intracytoplasmic sperm injection (ICSI). FertiCult™ Flushing medium can also be used human embryo washing and holding, and for embryo transfer (ET) in the uterine cavity. The medium is NOT intended for oocyte pick-up and follicle flushing. FertiCult Flushing medium is used during assisted reproductive Technology (ART) procedures of patients and couples undergoing infertility treatments.
FertiCult Flushing medium and FertiCult Flushing medium with phenol red and gentamicin intended for in vitro procedures involving human gametes (sperm and oocytes), including washing of gametes, sperm swim-up procedures, intra-uterine insemination (IUI) of the spermatozoa and sperm injection during intracytoplasmic sperm injection (ICSI). FertiCult Flushing media can also be used for human embryo washing and holding, and for embryo transfer (ET) in the uterine cavity.
The base formulation contains Water, Sodium Chloride, Potassium Chloride, Calcium Chloride Dihydrate, Magnesium Sulphate Heptahydrate, Sodium Dihydrogen Phosphate Dihydrate, Sodium Pyruvate, Glucose Monohydrate, Sodium Lactate, Sodium hydrogen carbonate, HEPES, Human Serum Albumin.
A FertiCult Flushing media variant contains Phenol red and Gentamicin.
The products can be used up to 7 days after opening, when sterile conditions are maintained, and the products are stored at 2-8°C.
The provided FDA 510(k) clearance letter pertains to the FertiCult Flushing medium and FertiCult Flushing medium with phenol red and gentamicin. This document demonstrates the substantial equivalence of the new device to a predicate device, as required for FDA clearance.
Here's an analysis of the acceptance criteria and the studies performed, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria listed are primarily for the physical, chemical, and biological properties of the media, ensuring its quality and suitability for use in Assisted Reproductive Technology (ART) procedures.
| Acceptance Criteria | Reported Device Performance (from "Shelf-life testing" and "In-use testing") |
|---|---|
| Appearance | All solutions should be without precipitates (Met at Time 0, during shelf-life, and after 7 days of in-use testing) |
| pH (USP <791>) | 7.3-7.9 (Met at Time 0, during shelf-life, and after 7 days of in-use testing) |
| Osmolality (USP <785>) | 270-290 mOsmol/kg (Met at Time 0, during shelf-life, and after 7 days of in-use testing) |
| Sterility (USP <71>) | No microbial growth (Met at Time 0, during shelf-life, and after 7 days of in-use testing) |
| Bacterial Endotoxin (USP <85>) | < 0.25 EU/mL (Met at Time 0, during shelf-life, and after 7 days of in-use testing) |
| Mouse Embryo Assay (MEA) | One-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 96 hours after a 1-hour exposure to test medium (Met at Time 0, during shelf-life, and after 7 days of in-use testing) |
Note: The document states that the product specifications were "met at time 0 and after real-time aging" for shelf-life, and "met seven days after opening" for in-use testing.
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes used for each test (e.g., how many batches were tested for pH, how many mouse embryos for MEA). It indicates that the tests were conducted for "validation and routine control" and to "support a shelf-life of 18 months" and "stability after bottle opening."
Regarding data provenance:
- The studies are non-clinical performance tests conducted in support of substantial equivalence.
- The data would originate from laboratory testing of the manufactured FertiCult media.
- The document does not specify the country of origin of the data or whether it was retrospective or prospective, but given it's for a new product obtaining FDA clearance, it would be prospective testing of the device for submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
These types of tests (chemical assays, sterility, endotoxin, MEA, biocompatibility) do not typically involve human experts to establish "ground truth" in the way clinical studies often do. The "ground truth" is established by adherence to recognized scientific and regulatory standards (e.g., USP monographs, ISO standards, FDA guidance documents) by trained laboratory personnel. The document does not specify the number or qualifications of individuals performing these specific tests, but it can be assumed they are qualified laboratory technicians or scientists.
4. Adjudication Method for the Test Set
Adjudication methods (like 2+1, 3+1) are typically used in clinical studies with human interpretation (e.g., imaging reads). For the non-clinical laboratory tests described here, adjudication is not applicable. The results are quantitative measurements or qualitative observations (e.g., presence/absence of microbial growth, precipitates) interpreted against defined numerical or descriptive criteria set by the relevant standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. The provided document is for the premarket notification (510(k)) of a reproductive media product. MRMC studies are typically performed for diagnostic imaging devices or other devices where human interpretation is a key component and the AI system's impact on human performance is being evaluated. This document focuses on the safety and effectiveness of the media itself through laboratory and biocompatibility testing, not on human-AI interaction.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No. This device is a culture medium, not an algorithm or AI-powered system. Therefore, standalone algorithm performance is not applicable.
7. The Type of Ground Truth Used
The ground truth for these non-clinical tests is based on:
- Established scientific and regulatory standards: USP monographs (<791> for pH, <785> for Osmolality, <71> for Sterility, <85> for Bacterial Endotoxin), ISO standards (e.g., ISO 13408, ISO 11137, ISO 17665, ISO 10993 series), ASTM D4169-22.
- Historically accepted biological performance indicators: Mouse Embryo Assay (MEA) results (development to expanded blastocyst), as specified by the 2021 FDA guidance.
- Observed physical and chemical characteristics: Absence of precipitates.
These are objective, quantifiable, or universally accepted biological endpoints, rather than expert consensus on a subjective interpretation.
8. The Sample Size for the Training Set
Not applicable. This device is a culture medium, not an AI/ML algorithm that requires a training set. The performance testing described validates the manufacturing quality and biological compatibility of the product.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as this is not an AI/ML device requiring a training set.
Ask a specific question about this device
(267 days)
Vitrification Freeze Kit (RFD-0101); Vitrification Thaw Kit (RFD-0201, RFD-0202)
Regulation Number: 21 CFR 884.6180
Reproduction Media
Classification Name: Reproductive media and supplements
Regulation: 884.6180
Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
Vitrification Thaw Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.
The Vitrification Freeze Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos during the in vitro assisted reproduction. The Vitrification Freeze Kit includes two solutions: Equilibration Solution (ES) and Vitrification Solution (VS). Both solutions contain ethylene glycol, dimethyl sulfoxide (DMSO), human serum albumin, and Quinn's Advantage® Medium with HEPES. The VS also contains additional sucrose to confer intracellular and extracellular cryoprotective effects. The solutions are available in three specifications: 2 mL, 4.5 mL, and 5 mL, all packaged in polypropylene (pp) vials. The kit includes one vial of each solution.
The Vitrification Thaw Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos during in vitro assisted reproduction. The Vitrification Thaw Kit includes 4 solutions: Thawing Solution 1 (TS1), Thawing Solution 2 (TS2), Thawing Solution 3 (TS3), and Thawing Solution 4 (TS4). All four solutions contain human serum albumin and Quinn's Advantage® Medium with HEPES. TS1 and TS2 also contain sucrose. The solutions are available in three specifications: 2 mL, 4.5 mL, and 5 mL, and 5 mL, all packaged in polypropylene (pp) vials. Model RFD-0201 includes one vial of each solution in the kit, and Model RFD-0202 contains two vials of TS1 and one vial of TS2, TS3, and TS4 each in the kit.
The provided FDA 510(k) clearance letter and summary describe the Vitrification Freeze Kit (RFD-0101) and Vitrification Thaw Kit (RFD-0201, RFD-0202) manufactured by Jiangsu Ruifuda Medical Device Co., Ltd. This device falls under the classification of Reproductive Media and Supplements (Regulation Number: 21 CFR 884.6180, Product Code: MQL).
This document focuses on the in vitro performance of the device rather than a clinical study involving human readers or a standalone AI algorithm. Therefore, many of the typical acceptance criteria and study elements associated with AI-driven medical devices (like MRMC studies, ground truth established by expert consensus, and human reader performance improvement) are not applicable here.
The device's performance is demonstrated through a series of non-clinical performance tests designed to show that it meets specific quality and functional criteria, ensuring its safety and effectiveness for its intended use in assisted reproduction.
Here's a breakdown of the acceptance criteria and study that proves the device meets them:
Acceptance Criteria and Reported Device Performance
The device's performance was assessed against specific product specifications through shelf-life testing. The criteria are outlined below:
| Acceptance Criteria (Specification) | Reported Device Performance (at Time 0 and after Real-Time Aging) |
|---|---|
| Appearance | Met (implicitly, as pH, osmolality, endotoxin, and MEA were tested, suggesting visual appearance was acceptable) |
| pH (USP <791>) | 7.30-7.70 (Met) |
| Osmolality (USP <785>) (mOsm/kg) | |
| Equilibration Solution (ES) | 700±70 (1:3 dilution) (Met) |
| Vitrification Solution (VS) | 1000±80 (1:5 dilution) (Met) |
| Thawing Solution 1 (TS1) | 1500-1900 (Met) |
| Thawing Solution 2 (TS2) | 840-920 (Met) |
| Thawing Solution 3 (TS3) | 240-300 (Met) |
| Thawing Solution 4 (TS4) | 240-300 (Met) |
| Endotoxin testing (USP <85>) | <0.5 EU/mL (Met) |
| Mouse Embryo Assay (MEA) | One-cell system: ≥ 80% embryos developed to expanded blastocyst at 96 hours (Met) |
| Sterility testing (USP <71>) | No microbial growth (Met) |
Study Details
-
Sample size used for the test set and data provenance:
This is not a clinical study involving a "test set" in the traditional sense of patient data. The "samples" refer to batches of the manufactured media. The document doesn't specify the exact number of batches or vials tested for each assay (e.g., how many vials for pH, how many embryos for MEA).- Data Provenance: The studies were conducted by the manufacturer, Jiangsu Ruifuda Medical Device Co., Ltd., in China, as indicated by the contact details. The data is prospective, generated from testing the manufactured product.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
Not applicable. The "ground truth" for this in vitro diagnostic/media device is established by the results of standardized laboratory assays (pH, osmolality, endotoxin, sterility, and Mouse Embryo Assay), not by expert human interpretation of images or observations. These assays have predefined acceptance criteria based on scientific and regulatory standards (e.g., USP and ISO standards). -
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable, as no human interpretation requiring adjudication was involved. The tests are quantitative laboratory measurements. -
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a vitrification and thawing media kit, not an AI or imaging device that would involve human readers. -
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. There is no algorithm involved with this device. Its performance is chemical and biological. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this device's performance is based on standardized laboratory assay results and biological functionality (specifically, the Mouse Embryo Assay, which demonstrates the media's ability to support embryo development). These methods are scientifically accepted ways to assess the quality and performance of reproductive media. -
The sample size for the training set:
Not applicable. This device does not involve a "training set" for an algorithm. Its development and validation are based on chemical formulation and in-house laboratory testing. -
How the ground truth for the training set was established:
Not applicable. No "training set" as understood in a machine learning context. The product formulation and manufacturing processes are established through standard chemical and biological development protocols, and validated against the performance criteria listed above.
Ask a specific question about this device
(225 days)
: Dewin Reproductive Media (Dewin Vitrification Kit, Dewin Thawing Kit)
Regulation Number: 21 CFR 884.6180
Reproduction Media
Regulation Name: Reproductive Media and Supplements
Regulation Number: 884.6180
Dewin Vitrification Kit is intended for use in the vitrification of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
Dewin Thawing Kit is intended for use in the thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
Dewin Vitrification Kit and Dewin Thawing Kit include a set of five media intended for use in the vitrification and thawing of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos as part of InVitro Fertilization (IVF) procedures.
The Vitrification Kit and Thawing Kit media are offered in 2mL and 5mL volumes. Dewin Vitrification Kit includes two media components, Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol, trehalose (only in VS), and dimethyl sulfoxide. During the vitrification process, embryos are first exposed to ES and then to VS. Using this methodology, the permeating cryoprotectants can replace water in the oocyte or PN through blastocyst stage embryos prior to vitrification and storage in liquid nitrogen.
Dewin Thawing Kit is composed of three media used stepwise for thawing and removing cryoprotectants from vitrified oocytes and PN through blastocyst stage embryos. It is composed of Thawing Solution (TS), Dilution Solution (DS) and Washing Solution (WS).
The primary ingredients of the vitrification and thawing media are water, sodium chloride, potassium chloride, sodium dihydrogen phosphate dihydrate, magnesium sulphate heptahydrate, calcium chloride, sodium bicarbonate, gentamicin sulfate, glucose, sodium-s-lactate, sodium pyruvate, alanyl glutamine, taurine, EDTA, phenol red, HEPES, HSA, trehalose, ethylene glycol, DMSO, non-essential amino acids. Cryoprotectants in the media include ethylene glycol (7.5% in ES, 15% in VS), DMSO (7.5% in ES, 15% in VS), and trehalose in VS, TS and DS.
The five solutions in the Dewin Vitrification Kit and Dewin Thawing Kit are aseptically filtered (storage vials sterilized by dry heat and radiation) and provided in glass bottles capped with a polypropylene screw-top cap. They are single-use only and have a shelf-life of 7 months when stored at 2-8°C.
The provided FDA clearance letter and 510(k) summary pertain to the Dewin Reproductive Media (Dewin Vitrification Kit, Dewin Thawing Kit). This document does not describe an AI/ML-enabled device or a study involving human readers or comparative effectiveness with AI. Therefore, most of the requested information (such as effect size of human readers with AI assistance, expert qualifications, and adjudication methods) is not applicable or cannot be extracted from the given text.
The information below focuses solely on the device and its testing as described in the provided document.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Dewin Reproductive Media are primarily based on physical, chemical, and biological performance measures. The reported device performance matches these criteria for the supported shelf-life.
| Acceptance Criteria | Reported Device Performance (at 7 months shelf-life) |
|---|---|
| Appearance: Clear, particulate free | Met |
| pH per USP <791>: 7.20 - 7.60 (all solutions) | Met |
| Osmolality per USP <785> | |
| ES: 2300-3000 mOsm/Kg | Met |
| VS: 4900-6600 mOsm/Kg | Met |
| TS: 1255-1535 mOsm/Kg | Met |
| DS: 745-911 mOsm/Kg | Met |
| WS: 260-295 mOsm/Kg | Met |
| Sterility per USP <71>: No microbial growth | Met |
| Bacterial Endotoxin per USP <85>: < 0.25 EU/mL | Met |
| MEA (Mouse Embryo Assay): ≥ 80% embryos developed to expanded blastocyst at 96 hours | Met |
2. Sample size used for the test set and the data provenance
The document does not explicitly state specific sample sizes for each test mentioned (e.g., how many batches were tested for pH, osmolality, sterility, or bacterial endotoxin).
For the Mouse Embryo Assay (MEA), the criteria indicate the measurement is "≥ 80% embryos developed to expanded blastocyst at 96 hours." While not explicitly stated as a "test set" in the context of AI, this biological performance test would involve a certain number of mouse embryos. The document does not specify this number.
Data Provenance: The studies were conducted by DonneVie Medical Technology (Shanghai) Co. Ltd. (China) to support the substantial equivalence claim. The document does not specify the country of origin for the biological samples (e.g., mouse embryos) used in the MEA. The studies are assumed to be "prospective" in the sense that they were designed and performed to generate data for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is reproductive media, not an AI/ML diagnostic or image analysis device that requires expert ground truth establishment in the traditional sense. The performance criteria are objective laboratory measurements (pH, osmolality, sterility, endotoxin) and a biological assay (MEA).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as expert adjudication is not relevant for the type of objective laboratory tests performed for this device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is reproductive media, not an AI/ML diagnostic tool involving human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an algorithm-based device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the performance evaluation of this reproductive media is based on:
- Objective Laboratory Measurements: pH, Osmolality, Sterility, Bacterial Endotoxin, Appearance. These have established scientific and regulatory standards (e.g., USP monographs).
- Biological Functionality Assay: Mouse Embryo Assay (MEA), which assesses the ability of the media to support embryonic development in vitro to a certain stage (expanded blastocyst). This assay serves as a functional "ground truth" for the media's biological suitability.
8. The sample size for the training set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device that requires a training set.
Ask a specific question about this device
(246 days)
HEPES (WHMO); CaseBioscience® HTF (HTFC); CaseBioscience® HTF w/HEPES (HHTF) Regulation Number: 21 CFR 884.6180
|
| Regulation Number: | 884.6180
CaseMONO™ Culture is intended for use for culture of embryos from fertilization to the blastocyst stage. This device can be used for transfer of embryos to the uterus.
CaseBioscience® HTF is intended for use for the culture of embryos to the cleavage stage (Day 3). This device can be used for transfer of embryos to the uterus.
CaseMONO w/HEPES™ is intended for use in short-term handling and manipulating gametes and embryos, including washing and intracytoplasmic sperm injection (ICSI). This medium is not intended for transfering to the uterine cavity.
CaseBioscience® HTF w/HEPES is intended for use in short-term handling gametes embryos, including washing and intracytoplasmic sperm injection (ICSI). This medium is not intended for transferring embryos to the uterine cavitv.
CaseMONO Culture, CaseMONO w/HEPES, CaseBioscience HTF, and CaseBioscience HTF w/HEPES are intended for use during in vitro procedures for gamete/embryo culture and one step media.
CaseMONO Culture is intended for use for culture of embryos from fertilization to the blastocyst stage. This device is a one-step media and can be used for transfer of embryos to the uterus.
CaseMONO w/HEPES is intended for use in short-term handling and manipulating gametes and embryos, including washing and intracytoplasmic sperm injection (ICSI).
CaseBioscience HTF is intended for use for the culture of embryos to the cleavage stage (Day 3). This device is not a one-step medium but can be used for transfer of embryos to the uterus.
CaseBioscience HTF w/HEPES is intended for use in short-term handling and manipulating gametes and embryos, including washing and intracytoplasmic sperm injection (ICSI).
CaseMONO Culture and CaseMONO w/HEPES have almost identical composition with differences in amount of NaHCO3 and Phenol red and presence of HEPES in CaseMONO w/HEPES. Similarly, CaseBioscience HTF and CaseBioscience HTF w/HEPES have an almost identical composition with the exception of HEPES present in CaseBioscience HTF w/HEPES and minor difference in composition for NaCl, Sodium L-lactate and Phenol red. CaseMONO Culture and CaseMONO w/HEPES contain EDTA. All media formulations contain antibiotic gentamicin.
The four solutions are aseptically filtered (storage vials sterilized by radiation) and provided in PETG square media bottles with HDPE screw-style caps. They have a shelf-life of 90 days when stored at 2-8°C. Media in the vials can be used for up to seven days after vial opening.
Acceptance Criteria and Device Performance Study for CaseBioscience Assisted Reproduction Media
This document outlines the acceptance criteria and the study conducted to prove that the CaseBioscience® Assisted Reproduction Media (CaseMONO™ Culture, CaseMONO™ w/HEPES, CaseBioscience® HTF, and CaseBioscience® HTF w/HEPES) meet these criteria, as detailed in the K242107 510(k) Summary.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for each variant of the CaseBioscience Assisted Reproduction Media are based on various non-clinical performance tests. The reported device performance met these acceptance criteria.
| Test Category | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Shelf-Life Testing | Product specifications (Appearance, pH, Osmolality, Sterility, Bacterial Endotoxin, MEA) met at time 0 and after aging at 2-8°C for 90 days. | All specified product parameters (Appearance, pH, Osmolality, Sterility, Bacterial Endotoxin, MEA) were met at time 0 and after 90 days of storage at 2-8°C, supporting a 90-day shelf-life. |
| Appearance | All solutions without precipitates. | Solutions were observed to be clear and free of precipitates. |
| pH (USP <791>) | CaseMONO™ Culture & CaseBioscience® HTF: 7.2-7.4CaseMONO™ w/HEPES & CaseBioscience® HTF w/HEPES: 7.2-7.6 | CaseMONO™ Culture & CaseBioscience® HTF: Measured pH values were within 7.2-7.4.CaseMONO™ w/HEPES & CaseBioscience® HTF w/HEPES: Measured pH values were within 7.2-7.6. |
| Osmolality (USP <785>) | CaseMONO™ Culture: 260-270 mOsm/kgCaseBioscience® HTF: 285-295 mOsm/kgCaseMONO™ w/HEPES: 260-270 mOsm/kgCaseBioscience® HTF w/HEPES: 285-295 mOsm/kg | CaseMONO™ Culture: Measured osmolality was within 260-270 mOsm/kg.CaseBioscience® HTF: Measured osmolality was within 285-295 mOsm/kg.CaseMONO™ w/HEPES: Measured osmolality was within 260-270 mOsm/kg.CaseBioscience® HTF w/HEPES: Measured osmolality was within 285-295 mOsm/kg. |
| Sterility (USP <71>) | No microbial growth. | No microbial growth was detected. |
| Bacterial Endotoxin (USP <85>) | < 0.1 EU/mL | Measured bacterial endotoxin levels were < 0.1 EU/mL. |
| Mouse Embryo Assay (MEA) | One-Cell MEA:- CaseMONO™ Culture: ≥ 80% embryos developed to expanded blastocyst at 120 hours.- CaseBioscience® HTF: ≥ 80% embryos developed to expanded blastocyst after 96 hours exposure.- CaseMONO™ w/HEPES: ≥ 80% embryos developed to expanded blastocyst at 96 hours after a 1-hour exposure.- CaseBioscience® HTF w/HEPES: ≥ 80% embryos developed to expanded blastocyst at 96 hours after a 1-hour exposure. | One-Cell MEA:- CaseMONO™ Culture: Achieved ≥ 80% expanded blastocyst development at 120 hours.- CaseBioscience® HTF: Achieved ≥ 80% expanded blastocyst development after 96 hours exposure.- CaseMONO™ w/HEPES: Achieved ≥ 80% expanded blastocyst development at 96 hours after a 1-hour exposure.- CaseBioscience® HTF w/HEPES: Achieved ≥ 80% expanded blastocyst development at 96 hours after a 1-hour exposure. |
| Biocompatibility | Meet criteria for Cytotoxicity, Sensitization, and Irritation per ISO 10993 standards. | The results demonstrated that the patient-contacting subject media (CaseMONO™ Culture and CaseBioscience® HTF) were biocompatible, meeting the requirements of ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), and ISO 10993-23 (Vaginal Irritation). |
| Aseptic Processing | Validation of aseptic filtration and aseptic filling according to ISO 13408-1:2008 & A1:2013 and ISO 13408-2:2018. | Aseptic filtration and filling were validated according to the specified ISO standards, including a filter challenge test using worst-case production conditions (excluding antimicrobials). |
| Sterilization | Validation of vial radiation sterilization per ISO 11137-1:2006 (Amd 1 (2013), Amd 2 (2018)) and ISO 11137-2:2013 (Amd 1 (2022)). | Vial radiation sterilization was validated according to the specified ISO standards. |
| Transportation Testing | Pass per ASTM D4169-22 and cap/seal leak testing using a method equivalent to USP <1207.2>. | The devices passed transportation testing per ASTM D4169-22 and subsequent cap/seal leak testing. |
2. Sample Size and Data Provenance
The provided document details non-clinical performance testing for assisted reproduction media. The sample sizes for each specific test (e.g., number of embryos in MEA, number of samples for pH/Osmolality) are not explicitly stated in this summary.
The data provenance is from non-clinical laboratory testing and is retrospective in nature, as it was conducted to support a 510(k) submission for an existing product. The country of origin of the data is not specified, but the testing would have been conducted by or for CaseBioscience, Inc., which is based in Canada.
3. Number of Experts and Qualifications for Ground Truth
This submission pertains to in vitro reproductive media. The ground truth for performance (e.g., embryo development, chemical properties, sterility) is established through standardized laboratory assays and specifications, rather than expert human interpretation of images or patient data. Therefore, the concept of "experts establishing ground truth for a test set" in the context of image interpretation or clinical diagnosis by multiple medical professionals is not applicable here. The "experts" would be the scientists and quality control personnel in the laboratory performing the assays, adhering to validated methods.
4. Adjudication Method for the Test Set
As the evaluation is based on objective laboratory measurements and standardized assays (e.g., pH meters, osmometers, bacterial growth cultures, embryo development rates), an "adjudication method" in the sense of a consensus process among human readers (e.g., 2+1, 3+1) is not applicable. The results are quantitative and binary (pass/fail against a threshold).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC comparative effectiveness study is typically performed for AI/CADe systems where human readers interpret medical images. Since this submission is for in vitro assisted reproduction media, which do not involve human interpretation of images or AI assistance in the diagnostic process, an MRMC comparative effectiveness study was not performed. Therefore, effect size of human improvement with AI assistance is not relevant or measurable in this context.
6. Standalone (Algorithm Only) Performance Study
This submission is for a medical device that is a consumable media, not a software algorithm. Therefore, a "standalone algorithm only" performance study is not applicable. The performance is assessed based on the physical, chemical, and biological properties of the media itself.
7. Type of Ground Truth Used
The ground truth used for proving the device meets the acceptance criteria is based on:
- Laboratory-established specifications and thresholds: For parameters like pH, osmolality, bacterial endotoxin, and appearance.
- Standardized biological assays: Specifically, the Mouse Embryo Assay (MEA), where the "ground truth" for success is a predefined percentage of embryos reaching a certain developmental stage (expanded blastocyst) within a specified timeframe.
- Compliance with international standards: For sterilization, aseptic processing, and biocompatibility testing (e.g., ISO, USP, ASTM).
8. Sample Size for the Training Set
This product is an in vitro medical device, not an AI/ML software. Therefore, there is no "training set" in the context of machine learning model development. The product is manufactured and tested according to established quality control procedures and specifications.
9. How the Ground Truth for the Training Set Was Established
Since there is no training set for an AI/ML model, this question is not applicable. The "ground truth" for the performance of the media is established by validated laboratory methods and industry/regulatory standards.
Ask a specific question about this device
(269 days)
Trade/Device Name: V-GRAD (V-GRAD Kit, V-GRAD 40, V-GRAD 80, V-GRAD 100) Regulation Number: 21 CFR 884.6180
|
| Regulation | 884.6180
V-GRAD is intended for separation of motile sperm from ejaculates by the density gradient method for use in assisted reproduction procedures.
V-GRAD is a sterile colloidal suspension with silicate particles, stabilized with covalently bound hydrophilic silanes that is intended for separation of motile sperm from ejaculates by the density gradient method for use in assisted reproduction procedures. Available as a 100 % stock solution (V-GRAD 100) and as 40% or 80% ready-to-use solutions (V-GRAD 40 and V-GRAD 80) with HEPES buffered Human Tubular Fluid (HTF) medium. V-GRAD 100 is a stock solution for preparing a density gradient system for semen preparation. The V-GRAD Kit consists of two bottles of V-GRAD 40 and V-GRAD 80. V-GRAD is aseptically filtered and provided in pre-sterilized 12 mL glass bottles closed with flourotec rubber stoppers and flip-tear off caps or 100 mL PET(G) bottles closed with HDPE screw caps. V-GRAD has a shelf-life of one year when stored at 2-8°C and can be used for up to seven days after bottle opening.
The provided text describes the regulatory clearance of a medical device, V-GRAD, which is a reproductive medium used for sperm separation. It outlines the device's technical specifications and compares them to a legally marketed predicate device (ORIGIO Gradient). However, the document primarily focuses on non-clinical performance testing to support substantial equivalence and does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/human-in-the-loop system.
The device, V-GRAD, is a laboratory reagent (colloidal suspension for density gradient sperm separation), not an AI-powered diagnostic or assistive tool. Therefore, the concepts of "acceptance criteria" and "study" as typically applied to AI-driven medical devices (e.g., using a test set, expert ground truth, MRMC studies, standalone performance) are not applicable to the V-GRAD submission as described.
The "performance testing" mentioned in the document refers to traditional lab-based tests for reproductive media:
- Aseptic filtration and aseptic filling validation: Ensures the sterility of the manufacturing process.
- Shelf-life testing: Verifies that the product maintains its specifications (appearance, pH, osmolality, endotoxin, density, Human Sperm Survival Assay (HSSA), sterility) over its stated shelf life after accelerated aging and simulated use.
- Transportation testing: Ensures the product integrity during shipping.
- Sperm evaluation: Confirms that sperm separated using the device exhibit comparable motility, morphology, and purity to those separated with the predicate device.
Since the prompt asks for details related to AI/human-in-the-loop medical device evaluation, and the provided text does not describe such a device or study, I cannot fulfill the request using only the given input.
Therefore, for each point, the answer will be "Information not available in the provided text" or "Not applicable as the device is not an AI/human-in-the-loop system."
Based on the provided text, here is the information:
-
A table of acceptance criteria and the reported device performance
-
Acceptance Criteria and Reported Device Performance: The document lists "product specifications" that are met during shelf-life testing, which implicitly serve as acceptance criteria for the quality and performance of the medium itself. These are not performance metrics for an AI system.
Criteria/Specification Acceptance Limit (Implied) Reported Performance (Implied, "met specifications") Appearance Pink rose (V-GRAD 40), light pink (V-GRAD 80), colorless (V-GRAD 100) Met pH 7.2-7.9 Met Osmolality (mOsm/kg) 310-340 (V-GRAD 40/80), 300-330 (V-GRAD 100) Met Endotoxin (EU/mL) <0.5 Met Density (g/cm³) 1.050-1.061 (V-GRAD 40), 1.100-1.111 (V-GRAD 80), 1.125-1.136 (V-GRAD 100) Met Human Sperm Survival Assay (HSSA) ≥ 80% of control motility at 24 hours after 1-hour exposure to test medium Met Sterility (USP <71>) No growth Met Sperm Evaluation (Motility, Morphology, Purity) Comparable performance with predicate device and other cleared devices Demonstrated comparable performance Shelf-Life 1 Year Supported (met specifications at time 0 and after accelerated aging)
-
-
Sample sizes used for the test set and the data provenance: Information not available in the provided text. The studies described are non-clinical lab tests of the medium itself, not a clinical test set for an AI device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as the device is not an AI/human-in-the-loop system requiring expert-adjudicated ground truth. The "ground truth" for V-GRAD's performance is established by standardized laboratory assays (e.g., pH meters, osmometers, endotoxin testing, HSSA).
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable as the device is not an AI/human-in-the-loop system.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as the device is not an AI/human-in-the-loop system.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as the device is not an AI/human-in-the-loop system.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the performance of the V-GRAD medium, the "ground truth" is defined by established laboratory methods and controls for measuring characteristics like pH, osmolality, sterility, and the biological response in an HSSA. For "Sperm separated using the subject device were evaluated for motility, morphology, and purity," the ground truth would be the measurement outcomes from standard semen analysis techniques, compared to controls (predicate device or other cleared devices).
-
The sample size for the training set: Not applicable as the device is not an AI/machine learning system.
-
How the ground truth for the training set was established: Not applicable as the device is not an AI/machine learning system.
Ask a specific question about this device
(180 days)
Shenzhen. 518107 CHINA
Re: K242043
Trade/Device Name: Sperm Freezing Medium Regulation Number: 21 CFR 884.6180
Sperm Freezing Medium is intended for the cryopreservation of human semen samples to sperm samples during cryopreservation and thawing prior to use in assisted reproductive technology procedures.
Not Found
I am sorry, but based on the provided text, there is no information available about the acceptance criteria or a study proving that a device meets such criteria.
The document is an FDA 510(k) clearance letter for a device named "Sperm Freezing Medium." This letter confirms that the device is substantially equivalent to a legally marketed predicate device. It discusses regulatory requirements, such as general controls, quality systems, and UDI, but it does not include any details about performance criteria or studies conducted by the manufacturer to demonstrate the device's performance.
Therefore, I cannot provide the requested information, including the table of acceptance criteria, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth details, or training set size.
Ask a specific question about this device
Page 1 of 23