    510k Summary Text (Full-text Search) :

    VAC-DUAL C); Kiwi® Complete Vacuum Delivery System, ProCup® (VAC-DUAL S)

    Regulation Number: 21 CFR 884.4340
    Name:** Fetal vacuum extractor
    Regulation Name: Fetal vacuum extractor
    Regulation Number: 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kiwi Complete Vacuum Delivery System (Kiwi® Complete Vacuum Delivery System, OmniCup® (VAC-DUAL M); Kiwi® Complete Vacuum Delivery System, OmniCup® with Traction Force Indicator (VAC-DUAL MT); Kiwi® Complete Vacuum Delivery System, OmniC Cup® for Cesarean Section (VAC-DUAL C); Kiwi® Complete Vacuum Delivery System, ProCup® (VAC-DUAL S)) is indicated for use in the following conditions:

    • Prolonged second stage of labor (arrest of descent) where fetopelvic relationships are adequate
    • Presumed fetal jeopardy that is not considered to be severe
    • Elective shortening of the second stage for selected maternal or fetal conditions
    Device Description

    The subject device is a fetal vacuum extractor (Kiwi® Complete Vacuum Delivery System) that includes four different models: OmniCup® (VAC-DUAL M); OmniCup® with Traction Force Indicator (VAC-DUAL MT); OmniC Cup® for Cesarean Section (VAC-DUAL C); ProCup® (VAC-DUAL S). All models use the same pump assembly, which allows the user to generate vacuum by squeezing the handle repeatedly. The pump/handle assembly includes a vacuum gauge that is color-coded to indicate safe operating pressures for the device. Models M and MT differ only in the inclusion of a traction force indicator (MT). Model C differs from models M/MT in the cup design, which has a lower profile cup with finger grooves to facilitate use in cesarean sections, as well as a baffle instead of an inner filter. The model S design is suggested for use in vaginal births and includes a combined tube and cup portion that is flexible and fits over the fetal head.

    For vaginal delivery, the device is operated by inserting the cup into the vagina and placing the cup at the flexion point on the fetal skull. The Omni-C is designed specifically for the confined abdominal space of C-section deliveries. It is designed with a slimmer cup profile, finger grooves and baffle filter on the inside. For cesarean delivery, the cup is inserted into the incision over the flexion point.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K170329

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-03-24

    (50 days)

    Product Code
    Regulation Number
    878.5040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CLASSIFICATIONS:

    System, suction, lipoplasty - MUU, 21 CFR 878.50406 Extractor, vacuum, fetal- HDB, 21 CFR 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use for Medela's Basic and Dominant Flex Suction Pumps is to provide a vacuum source for use in hospitals and clinics.

    Basic: The intended use of the Basic suction pump is the creation of a constant vacuum for use in hospitals and clinics. This vacuum can be used for general suction, to aspirate and remove: surgical fluids, tissue (including bone), gases, bodily fluids or infectious materials and during specific procedures which may include, vacuum extraction, aspiration during flexible endoscopy, use with cardiac tissue stabilizers during off-pump coronary artery bypass and epicardial ablation probes.

    Dominant Flex: The intended use of the Dominant Flex suction pump is the creation of a constant vacuum for use in hospitals and clinics. This vacuum can be used for general suction, to aspirate and remove: surgical fluids, tissue (including bone), gases, bodily fluids or infectious materials and during specific procedures which may include, vacuum extraction, aesthetic body contouring, aspiration during flexible endoscopy, use with cardiac tissue stabilizers during off-pump coronary artery bypass, and epicardial ablation probes.

    Device Description

    The Medela" Basic and Dominant Flex are high vacuum and high flow, AC-powered suction pumps that can be used when large quantities of fluid must be suctioned quickly in the hospital, ambulatory surgery center, clinic, and doctors' practice. Their construction includes a piston and cylinder system which provides strong suction performance and quiet, dependable operation and a microcontroller which is used to control motor speed and the user interface. Additional advantages of the Medela" Basic and Dominant Flex are user friendliness and simple cleaning.

    Medela® Basic and Dominant Flex cover the 4 main functions of

    • . Powerful and high suction capacity
    • . Rapid vacuum build-up
    • Low vibrations and quiet
    • . Design; smooth surface and easy to use and clean

    The Medela" Basic provides a fixed airflow of 30 l/m and the Medela" Dominant Flex offers the innovative function of selectable airflow of 40, 50 or 60 l/m.

    All operating elements for nurses or doctors are located on the front side of both pumps. These include the vacuum gauge, vacuum regulator knob, operating elements and Safety Set. The Safety Set consists of a 0.25l jar, lid and float and prevents an overflow into the pump. The Medela" Basic and Dominant Flex suction pumps can be operated via capacitive sensors (called "CleanTouch") to turn the pumps on/off. Additionally the Dominant Flex has capacitive sensors for adjusting the flow between 40, 50 and 60 l/m. Both pumps have a knob for the vacuum regulator. The vacuum inside the tubing is displayed on the vacuum gauge. The Medela" Basic and Dominant Flex use three indicator lights to provide information to the user on the status of the pump. All operating elements for the biomedical technicians are located on the back side. This is where the appliance inlet for plugging in the power cord is located.

    Both pumps are available either as rack or portable version can be combined with the trolley to create a mobile version.

    A variety of reusable and disposable accessories are available for use with the Medela" Basic and Dominant Flex suction pumps or are intended to be marketed with these pumps.

    AI/ML Overview

    This document is an FDA 510(k) summary for the Medela Basic and Dominant Flex suction pumps. It states that no clinical testing was performed for this submission, as it relates primarily to re-phrased and slightly expanded indications for use. Therefore, the information typically found in acceptance criteria tables and study results for clinical performance is not present.

    However, based on the provided document, here's what can be inferred and stated regarding the device and its compliance, specifically focusing on the non-clinical aspects and the lack of a clinical study for this 510(k):

    Acceptance Criteria (Non-Clinical) and Reported Device Performance

    Since this 510(k) is an administrative change to re-phrase and expand existing indications for use, and explicitly states "The Basic and Dominant Flex Suction Pumps have not been the subject of clinical testing," there isn't a table of clinical acceptance criteria and performance data as one would expect from a new device submission or a performance study.

    Instead, the acceptance criteria for this specific submission relate to non-clinical performance and compliance with relevant standards. The document highlights:

    Acceptance Criteria (Non-Clinical)Reported Device Performance
    Compliance with voluntary standards for electrical safetyDemonstrated compliance with standards.
    Compliance with voluntary standards for electromagnetic compatibilityDemonstrated compliance with standards.
    Compliance with voluntary standards for safety of electrically powered suction pumpsDemonstrated compliance with standards.
    No impact on compliance to standards due to changes to product labeling/indications for useChanges to labeling did not impact compliance.
    Maintenance of performance as intended despite re-phrased indicationsVerification and Validation testing demonstrated no adverse effects and device performs as intended.
    Equivalence to predicate devices for new indicationsStated as "substantially equivalent" based on nonclinical testing and unchanged intended use.
    Compliance with IEC 60601-1-2 (4th edition)Demonstrated compliance with the newer 4th edition.

    Study Details (Focusing on the 510(k) submission, not a clinical study):

    For this particular 510(k) submission (K170329), a traditional clinical study with patient data was not conducted. The basis for substantial equivalence relies on:

    1. Predicate Device Equivalence: The Medela Basic and Dominant Flex were previously cleared under K150134. This submission re-phrases indications and adds "general suction" which is stated as being identical to the predicate device, Penumbra Aspiration Pump (K051758).
    2. Non-Clinical Testing: Verification and Validation testing were performed to ensure that the labeling changes did not adversely affect the device's performance or compliance with safety standards.

    Therefore, many of the typical questions for a clinical study are not applicable here.

    1. Sample sizes used for the test set and the data provenance: Not applicable, as no new clinical test set was used for this 510(k) submission. Non-clinical testing (e.g., electrical safety, EMC) would involve testing units of the device, not patient data. The document implies prior testing for the original clearance (K150134) is still valid.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth typically refers to clinical diagnosis or outcomes. For non-clinical testing, expert judgement might be involved in reviewing test protocols and results, but it's not "ground truth" in the clinical sense.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a suction pump, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithmic device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For this specific submission, the "ground truth" for the device's performance is adherence to established engineering and safety standards, and functional verification that the re-phrased indications do not alter the physical operation or safety of the pump. The [K150134](https://510k.innolitics.com/device/K150134) submission, which initially cleared these devices, would have relied on previously established predicate device performance and non-clinical testing.
    7. The sample size for the training set: Not applicable. This refers to machine learning algorithms, which is not what this device is.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K150134

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-04-21

    (90 days)

    Product Code
    Regulation Number
    878.5040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CLASSIFICATIONS:

    System, suction, lipoplasty - MUU, 21 CFR 878.50406 Extractor, vacuum, fetal- HDB, 21 CFR 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use for Medela's Basic and Dominant Flex Suction Pumps is to provide a vacuum source for use during surgery.

    The Basic Suction Pump is indicated for vacuum extraction, aspiration during flexible endoscopy, and aspiration and removal of surgical fluids, tissue (including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system either during surgery or at the bedside. The Basic Pump can also be used to provide the vacuum required for use with cardiac tissue stabilizers used in off-pump coronary artery bypass and for use with epicardial ablation probes.

    The Dominant Flex Suction Pump is indicated for vacuum extraction, aesthetic body contouring, aspiration during flexible endoscopy, and aspiration and removal of surgical fluids, tissue (including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system either during surgery or at the bedside. The Dominant Flex Pump can also be used to provide the vacuum required for use with cardiac tissue stabilizers used in off-pump coronary artery bypass and for use with epicardial ablation probes.

    Device Description

    The Medela" Basic and Dominant Flex are high vacuum and high flow, AC-powered suction pumps that can be used when large quantities of fluid must be suctioned quickly in the hospital, ambulatory surgery center, clinic, and doctors' practice. Their construction includes a piston and cylinder system which provides strong suction performance and quiet, dependable operation and a microcontroller which is used to control motor speed and the user interface. Additional advantages of the Medela " Basic and Dominant Flex are user friendliness and simple cleaning.

    Medela" Basic and Dominant Flex cover the 4 main functions of

    • . Powerful and high suction capacity
    • . Rapid vacuum build-up
    • . Low vibrations and quiet
    • . Design; smooth surface and easy to use and clean

    The Medela" Basic provides a fixed airflow of 30 l/m and the Medela" Dominant Flex offers the innovative function of selectable airflow of 40, 50 or 60 l/m.

    All operating elements for nurses or doctors are located on the front side of both pumps. These include the vacuum gauge, vacuum regulator knob, operating elements and Safety Set. The Safety Set consists of a 0.25l jar, lid and float and prevents an overflow into the pump. The Medela" Basic and Dominant Flex suction pumps can be operated via capacitive sensors (called "CleanTouch") to turn the pumps on/off. Additionally the Dominant Flex has capacitive sensors for adjusting the flow between 40, 50 and 60 l/m. Both pumps have a knob for the vacuum regulator. The vacuum inside the tubing is displayed on the vacuum gauge. The Medela " Basic and Dominant Flex use three indicator lights to provide information to the user on the status of the pump. All operating elements for the biomedical technicians are located on the back side. This is where the appliance inlet for plugging in the power cord is located.

    Both pumps are available either as rack or portable version can be combined with the trolley to create a mobile version.

    A variety of reusable and disposable accessories are available for use with the Medela" Basic and Dominant Flex suction pumps or are intended to be marketed with these pumps.

    AI/ML Overview

    The provided text describes the Medela AG Basic and Dominant Flex Suction Pumps (K150134). The submission is a 510(k) for new Indications for Use, therefore, the study focuses on bench testing to demonstrate performance for these new indications, rather than a clinical study establishing new acceptance criteria for the entire device.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal acceptance criteria in a table format with corresponding reported performance for the entire device. Instead, it describes performance characteristics demonstrated through bench testing for the new indications.

    Parameter/IndicationAcceptance Criteria (Implicit)Reported Device Performance
    Aesthetic Body Contouring (Dominant Flex)Equivalent performance to predicate devices (Dominant 50 Lipo K063336 and HK Liposuction Aspirator Model AP-III K032802) in:
    Free air flowEquivalent to predicates.Demonstrated equivalent free air flow to predicate devices.
    Vacuum build upEquivalent to predicates.Demonstrated equivalent vacuum build up to predicate devices.
    Suctioning capacity (fluids with varying consistencies and cannulas)Equivalent to predicates.Demonstrated equivalent suctioning capacity of fluids with varying consistencies (sterile water, apple sauce, solidifier mix) and various cannulas to predicate devices.
    Epicardial Ablation Probes (Basic and Dominant Flex)Maintain vacuum at specified level when the pump is running.Maintained vacuum at -600 mmHg ± 10% when the pump is running with the Estech Cobra Fusion Ablation System.
    Cardiac Tissue Stabilizers (Basic and Dominant Flex)Maintain vacuum at specified levels for a sustained period.Maintained vacuum at settings of -250 or -400 mmHg for up to 8 hours with the Medtronic Octopus Evolution and Urchin Evo Positioners.

    2. Sample Size Used for the Test Set and Data Provenance

    • The document primarily describes bench testing, which typically involves a limited number of test units (samples) of the device itself rather than a large human subject test set.
    • Sample size for the device under test: Not explicitly stated, but bench testing generally uses a small number of production or representative units.
    • Data Provenance: The testing was conducted by Medela AG, likely at their facilities, indicating in-house (prospective) testing for the specific purpose of this 510(k) submission. No country of origin for test data is specified beyond the manufacturer's location (Switzerland).

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

    • This information is not applicable as the studies are entirely non-clinical (bench testing). No human experts were involved in establishing "ground truth" for diagnostic or clinical outcomes.

    4. Adjudication Method for the Test Set

    • This information is not applicable as the studies are entirely non-clinical (bench testing).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The submission focuses on non-clinical bench testing to demonstrate substantial equivalence for new indications.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • This question is not applicable as the device is a physical suction pump, not an AI algorithm. The performance evaluation is inherently a standalone assessment of the device's physical capabilities.

    7. The Type of Ground Truth Used

    • The ground truth for the bench testing was based on objective measurements of physical parameters such as free air flow, vacuum build-up, suctioning capacity, and vacuum maintenance. This is a form of objective physical measurement rather than expert consensus, pathology, or outcomes data. Comparison was made against the specifications and performance of legally marketed predicate devices.

    8. The Sample Size for the Training Set

    • This information is not applicable as the device is a physical suction pump, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable as the device is a physical suction pump, not an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130123

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-10-11

    (267 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Suction Pump

    PRODUCT CODE: BTA

    SECONDARY CLASSIFICATIONS:

    Extractor, vacuum, fetal- HDB, 21 CFR 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Basic and Dominant Flex Suction Pumps are indicated for vacuum extraction, aspiration during flexible endoscopy and aspiration and removal of surgical fluids, tissue (including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system either during surgery or at the bedside.

    Device Description

    The Medela Basic and Dominant Flex are high vacuum and high flow, AC-powered suction pumps that can be used when large quantities of fluid must be suctioned quickly in the hospital, ambulatory surgery center, clinic, and doctors' practice. Their construction includes a piston and cylinder system which provides strong suction performance and quiet, dependable operation and a microcontroller which is used to control motor speed and the user interface. Additional advantages of the Medela Basic and Dominant Flex are user friendliness and simple cleaning.

    Medela Basic and Dominant Flex cover the 4 main functions of

    • Powerful and high suction capacity .
    • Rapid vacuum build-up .
    • Low vibrations and quiet .
    • Design; smooth surface and easy to . use and clean

    The Medela Basic provides a fixed airflow of 30 l/m and the Medela "Dominant Flex offers the innovative function of selectable airflow of 40, 50 or 60 I/m.

    All operating elements for nurses or doctors are located on the front side of both pumps. These include the vacuum gauge, vacuum regulator knob, operating elements and Safety Set. The Safety Set consists of a 0.25l jar, lid and float and prevents an overflow into the pump. The Medela Basic and Dominant Flex suction pumps can be operated via capacitive sensors (called "CleanTouch") to turn the pumps on/off. Additionally the Dominant Flex has capacitive sensors for adjusting the flow between 40, 50 and 60 l/m. Both pumps have a knob for the vacuum regulator. The vacuum inside the tubing is displayed on the vacuum gauge. The Medela Basic and Dominant Flex use three indicator lights to provide information to the user on the status of the pump. All operating elements for the biomedical technicians are located on the back side. This is where the appliance inlet for plugging in the power cord is located.

    Both pumps are available either as rack or portable version can be combined with the trolley to create a mobile version.

    A variety of reusable and disposable accessories are available for use with the Medela ` Basic and Dominant Flex suction pumps or are intended to be marketed with these pumps.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Medela Basic and Dominant Flex Suction Pumps. It aims to demonstrate substantial equivalence to predicate devices rather than conducting a de novo study to establish new acceptance criteria and prove performance against them.

    Therefore, the submission does not contain information typically found in a study designed to prove a device meets specific acceptance criteria using defined performance metrics, sample sizes, expert ground truth, or MRMC studies. Instead, it focuses on comparing the new device's technological characteristics and intended use to previously cleared predicate devices.

    Here's a breakdown of why the requested information is largely absent or not directly applicable to this type of submission, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is not presented in the context of a de novo study the device meets. The submission primarily compares the new device's specifications (flow, maximum vacuum) to predicate devices and asserts equivalence.

    CharacteristicAcceptance Criteria (from study)Reported Device Performance (from study)Why not available/applicable
    Vacuum Performance(Not explicitly defined as acceptance criteria for a new study)Basic: -675mmHg (-90kPa)Dominant Flex: -713mmHg (-95kPa)These are specifications of the new device, compared to predicate device specifications for equivalence, not new acceptance criteria proven by a study. The "Discussion of Differences" states "Equivalent. Dominant Flex now offers a slightly higher vacuum."
    Flow Rate(Not explicitly defined as acceptance criteria for a new study)Basic: 30 l/minDominant Flex: Selectable 40, 50, 60 l/minThese are specifications of the new device, compared to predicate device specifications for equivalence. The "Discussion of Differences" states "Equivalent."
    Noise Level(Not explicitly defined, but "Low vibrations and quiet" is a claimed feature)"Bench Testing demonstrated that the specifications for vacuum, flow, noise and endurance were met."Specific quantitative noise acceptance criteria and measured performance are not provided.
    Endurance(Not explicitly defined)"Bench Testing demonstrated that the specifications for vacuum, flow, noise and endurance were met."Specific quantitative endurance acceptance criteria and measured performance are not provided.
    Electrical SafetyCompliance with IEC 60601-1Compliance demonstratedNon-quantitative (compliance with standard)
    EMCCompliance with IEC 60601-1-2Compliance demonstratedNon-quantitative (compliance with standard)
    Electrically Powered Suction Pump SafetyCompliance with ISO 10079-1Compliance demonstratedNon-quantitative (compliance with standard)

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. The submission does not describe a clinical study with a "test set" in the sense of patient data or samples used to evaluate device performance against clinical endpoints. The "non-clinical tests" mentioned (Risk Analysis, Software Validation, Electrical Safety, etc.) are engineering and regulatory compliance activities, not clinical performance studies on a specific dataset.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. No multi-reader, multi-case study with ground truth established by experts is described.

    4. Adjudication Method for the Test Set

    Not applicable. No test set requiring expert adjudication is described.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "The Basic and Dominant Flex Suction Pumps have not been the subject of clinical testing."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a powered suction pump, a physical medical device, not an algorithm or AI software. Therefore, the concept of "standalone algorithm performance" is not relevant.

    7. The Type of Ground Truth Used

    Not applicable. There is no clinical study described that required a "ground truth" (e.g., pathology, outcomes data) to evaluate diagnostic or therapeutic accuracy. The closest concept is the "Bench Testing" which verified engineering specifications.

    8. The Sample Size for the Training Set

    Not applicable. As this is not an AI/ML device, there is no "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set was Established

    Not applicable for the same reason as point 8.


    In summary, the provided document is a 510(k) submission which focuses on demonstrating substantial equivalence to pre-existing, legally marketed predicate devices through a comparison of technological characteristics, intended use, and adherence to relevant voluntary standards. It explicitly states that no clinical testing was performed on the new device, and thus, does not detail studies designed to establish and meet specific performance acceptance criteria in the manner requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041579

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-04-21

    (311 days)

    Product Code
    Regulation Number
    884.4340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Vacuum Extractor, Obstetrical and Gynecological Surgical(Classified Class II, per 21 CFR Section § 884.4340
    MN 55112-1891 Re: K041579

    Trade/Device Name: Medela® Single Use Silc Cup Regulation Number: 21 CFR 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medela® Single Use Silc Cup is intended to be used to facilitate the delivery of the fetus during childbirth. The device enables traction to be applied to the fetal head (in the birth canal) by means of a suction cup attached to the scalp. It is larmedd by an external vacuum source and indicated for using during vaginal or cesarean delivery.

    The Indications for use are

    • Fetal distress in the second stage
    • Maternal delay in the second stage
    • Maternal conditions requiring a short second stage
    Device Description

    The Medela® Single Use Silc Cup is a sterile, disposable device made of an elastomer and available in one size, 60mm inner diameter. The instrument is elastomer and consists of a handle and a suction cup to be used with an external vacuum pump and tubing.

    The cup has a smooth external surface with a small longitudinal ridge, which makes it possible to observe any rotation. The inside of the cup is lined with mance it possible to objections enable the air between the cup and the child's Small projoctions: The projections that the cup sucks firmly against the child's head. The soft suction cup deforms, thereby ensuring optimal adhesion and minimizing trauma to the fetal scalp.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Medela Single Use Silc Cup, which is an Obstetrical Vacuum Delivery System. The submission aims to demonstrate substantial equivalence to predicate devices, rather than establishing new acceptance criteria or proving performance through a new clinical study.

    Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria, device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or training set details.

    The core of this 510(k) submission is to assert that the Medela Single Use Silc Cup is substantially equivalent to legally marketed predicate devices, meaning its technological characteristics and intended use are similar enough that it does not raise new questions of safety and effectiveness.

    Here's what can be extracted from the document based on your request, with the caveat that many points cannot be addressed due to the nature of a 510(k) submission for substantial equivalence:

    1. Table of acceptance criteria and the reported device performance:

    This information is not provided in the document. The submission focuses on demonstrating substantial equivalence to predicate devices rather than setting new acceptance criteria and reporting performance against them for this specific device.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    No test set data or sample sizes are mentioned. The submission relies on the established safety and effectiveness of predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. No new test set data requiring ground truth establishment is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is not an AI-assisted diagnostic tool, and no MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable. The submission relies on the established safety and effectiveness of predicate devices, not new ground truth data for this device.

    8. The sample size for the training set:

    Not applicable. This is not an AI/algorithm-based device.

    9. How the ground truth for the training set was established:

    Not applicable.

    Summary of the document's argument for meeting "acceptance criteria":

    The "acceptance criteria" in this context are primarily the regulatory requirements for Substantial Equivalence under 21 CFR 884.4340, a Class II medical device. The study that "proves" the device meets these criteria is the comparison to predicate devices and the assertion that "The Technology of the Medela® Single Use Silc Cup is identical to the predicate... There are no technical differences which would raise new aspects regarding safety and effectiveness."

    The key points of the submission are:

    • Device Name: Medela® Single Use Silc Cup, Fetal Vacuum Extractor
    • Common Name: Obstetrical Vacuum Delivery System
    • Classification Name: Fetal Vacuum Extractor, Obstetrical and Gynecological Surgical (Class II, per 21 CFR Section § 884.4340)
    • Predicate Devices:
      • Medela Inc. OB Vacuum Extractor (K841492)
      • Hollister Inc. Ameda/Egnell Dolphin Dispo-Soft Vacuum Extractor (K895700)
    • Technological Characteristics: The submission explicitly states, "The Technology of the Medela® Single Use Silc Cup is identical to the predicate... There are no technical differences which would raise new aspects regarding safety and effectiveness."
    • Biocompatibility: The materials are considered appropriate for their intended use.
    • Conclusion: Based on the information presented, it is concluded that the proposed device is safe and effective for its intended use and is substantially equivalent to the predicate devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K043614

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-04-14

    (105 days)

    Product Code
    Regulation Number
    884.4340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Product Name:

    This is an accessory to a fetal vacuum extractor, which is a Class II device per 884.4340
    McDermott Rd., #200-207 PLANO TX 75025

    Re: K043614

    Trade/Device Name: Vaculink Regulation Number: 21 CFR 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VacuLink is intended for obstetric use in the measurement and recording of data related to vacuum-assisted delivery (such as vacuum, number and duration of pulls and number of pop-offs) on the strip chart.

    Device Description

    The VacuLink is a small, battery powered single-use vacuum measuring device. It is connected in parallel to the vacuum chamber of commercial handheld vacuum assisted delivery systems (i.e. Mityvac & Kiwi PalmPump). As a redundant gauge to the vacuum gauges on those devices, it electronically measures the vacuum applied to the fetal head during a vacuum assisted delivery. It connects to the vacuum port of the maternal/fetal monitor (i.e. HP, Corometrics and Spacelabs) via electrical cable. The vacuum value is numerically displayed in centimeters of mercury (cm Hg). The vacuum value is also recorded from the uterine activity channel of the maternal/fetal monitor and it is permanently stored in the hospital database. The information may also be displayed on a video monitor as well.

    The VacuLink incorporates a light sequence of red, yellow, and green. When the VacuLink is first connected to the maternal/fetal monitor, all three lights will flash in sequence as a quick test of the internal components. At the successful completion of the test, the green LED illuminates until the VacuLink is disconnected from the maternal/fetal monitor. A flashing yellow LED illuminates when the battery voltage is low and the VacuLink should be replaced. If a VacuLink that has been used previously and is reconnected, the red LED illuminates indicating the VacuLink is no longer operational.

    AI/ML Overview

    Here's an analysis of the provided text regarding the VacuLink device, focusing on acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CriteriaAcceptance ValueReported Device Performance
    Claimed accuracy/toleranceAt 70 cm Hg = 3.26% on the VacuLink (inherent)At 70 cm Hg = 3.26% (matches claim)
    Comparative AccuracyAt least as accurate as comparable vacuum gauges"More accurate by a factor of 10" than comparable gauges in clinical use.
    Electrical Safety & EMC ConformanceConformance to applicable standards"Electrical safety and EMC testing was also performed to demonstrate conformance to applicable standards."
    Software ValidationValidated"Software validation...have also been provided."
    Sterilization Process InformationProvided and validated (implied for a single-use device)"Sterilization process information have also been provided."

    Note: The document primarily focuses on the device's accuracy and comparison to predicate devices, rather than explicitly listing "acceptance criteria" as pass/fail thresholds for a primary clinical endpoint. The "Claimed accuracy/tolerance" is presented as a characteristic of the device itself, which then needs to be demonstrated. The comparative accuracy against existing devices could be considered an acceptance criterion for market clearance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Phase): Not explicitly stated. The document mentions an "Initial clinical phase (feasibility study)" but does not provide details on the number of subjects or cases included in this study.
    • Data Provenance: Not explicitly stated. Given that the device is manufactured by Medevco, Inc. in Dallas, TX, and the submission is to the FDA, it's likely the clinical feasibility study (if conducted with patients) was in the US. The nature of "Initial bench testing" and "Secondary bench testing" implies laboratory or engineering environments. It's a prospective study as implies "Initial clinical phase (feasibility study)" to study a new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not explicitly stated. The document does not provide details on who established the "ground truth" for the clinical feasibility study, or if "ground truth" for vacuum measurements was established by a separate, highly accurate reference device independent of human interpretation.

    4. Adjudication Method for the Test Set

    Not explicitly stated. The document does not describe any expert adjudication process for the clinical feasibility study. For the bench testing, the "ground truth" would likely be established by a highly accurate, calibrated reference instrument, not human adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. The VacuLink is a vacuum measuring device, not an imaging or diagnostic device that typically involves human interpretation of results in an MRMC study. Its function is to provide an objective measurement (vacuum pressure) and record data. Therefore, an MRMC study assessing human reader improvement with AI assistance is not relevant to this device.

    6. Standalone Performance Study

    Yes, a standalone performance study was done implicitly through the "Performance Data & Conclusions" section.

    • The initial and secondary bench testing phases directly assess the standalone performance of the VacuLink device against certified reference instruments. The statement "Performance testing of VacuLink... reveals that the VacuLink is as accurate as the claimed accuracy" and "VacuLink was more accurate by a factor of 10" (in clinical use, compared to uncalibrated existing gauges) directly addresses its standalone measurement capability.
    • The "Initial clinical phase (feasibility study)" would also have assessed its standalone functionality in a real-world setting, though specific data from this phase is not provided in detail.

    7. Type of Ground Truth Used

    • Bench Testing: Certified reference instruments (for vacuum measurement).
    • Clinical Feasibility Study: Implicitly, the actual vacuum applied during an assisted delivery, as measured by the VacuLink and potentially compared to existing devices. The document highlights the "lack of calibration of the comparable vacuum gauges" in clinical use, suggesting the VacuLink aimed to provide a more reliable ground truth in that setting.

    8. Sample Size for the Training Set

    Not applicable. The VacuLink is a hardware device for measuring vacuum pressure and recording data. It is not an AI/ML-based device that requires a "training set" in the conventional sense. The "training" for such a device involves engineering design, component selection, and calibration, not machine learning model training on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of AI/ML for this device. The accuracy of the device itself would be established through a rigorous calibration process using highly stable and accurate reference pressure gauges.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021193

    Validate with FDA (Live)

    Date Cleared
    2003-02-12

    (303 days)

    Product Code
    Regulation Number
    884.4340
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Hsien 252 TAIWAN

    Re: K021193

    Trade/Device Name: All Silicone Vacuum Suction Regulation Number: 21 CFR 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Silicone Vacuum Cup is to assist a clinician in delivery of an infant during childbirth.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) clearance letter from the FDA for a medical device (All Silicone Vacuum Suction) and does not contain the specific information requested in your prompt regarding acceptance criteria, study details, sample sizes, ground truth establishment, or expert involvement. The letter simply states that the device has been found substantially equivalent to a legally marketed predicate device.

    Therefore, I cannot extract the following information from the given text:

    1. A table of acceptance criteria and the reported device performance: This information is not present. The letter confirms substantial equivalence, not performance against specific criteria.
    2. Sample size used for the test set and the data provenance: Not mentioned.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
    4. Adjudication method for the test set: Not mentioned.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not mentioned.
    6. If a standalone study was done: Not mentioned.
    7. The type of ground truth used: Not mentioned.
    8. The sample size for the training set: Not mentioned.
    9. How the ground truth for the training set was established: Not mentioned.

    The document is a regulatory approval, not a clinical study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020447

    Validate with FDA (Live)

    Date Cleared
    2002-04-12

    (60 days)

    Product Code
    Regulation Number
    884.4340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    . § 884.4340)

    Product Code

    HDB

    Predicate Devices

    The Mityvac® Merlin™ System is a modification
    K020447 Trade/Device Name: MityVac Merlin, Model 10027 Obstetrical Vacuum Cup Regulation Number: 21 CFR 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Prism Enterprises, Inc.'s Mityvac® Merlin™ is intended to be utilized to assist a clinician in the delivery of an infant during childbirth. This device is indicated for use during vaginal delivery and Cesarean sections.

    Device Description

    The Mityvac® Merlin™ System is a modification of the cleared Mityvac® MityOne™ System with M-Style® or MitySoft® Bell Cups (K011532). The primary difference between the Mityvac® Merlin™ System and the MityOne™ System is that the Merlin™ System utilizes a CO2 canister to create the energy for vacuum, while the MityOne™ uses a builtin hand driven pump as the energy source.

    AI/ML Overview

    Let's break down the information about the Mityvac® Merlin™ Vacuum Assist Delivery System based on the provided document.

    Acceptance Criteria and Device Performance

    The provided document (K020447) is a 510(k) summary for a medical device. In this context, "acceptance criteria" and "device performance" are typically evaluated by establishing substantial equivalence to a predicate device, rather than explicit numerical performance metrics like sensitivity/specificity for a diagnostic AI model.

    The core acceptance criterion for a 510(k) device is demonstrating that the new device is substantially equivalent in intended use, technological characteristics, and safety and effectiveness to a legally marketed predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (for 510(k) Substantial Equivalence)Reported Device Performance (Summary Basis for Finding)
    Intended Use: Device has the same intended use as the predicate device.The Merlin™ System and the MityOne™ System have the same intended use (to facilitate the delivery of the fetus during childbirth) and the same indications (for use during vaginal delivery and caesarean sections).
    Indications for Use: Device has the same indications for use as the predicate device.The Merlin™ System and the MityOne™ System have the same intended use and same indications.
    Technological Characteristics: Device has similar technological characteristics to the predicate device, or if there are differences, those differences do not raise new questions of safety or effectiveness.The primary difference is that the Merlin™ uses a CO2 canister for vacuum, while the MityOne™ uses a built-in hand pump. This difference "does not raise new questions of safety or effectiveness." The principles of operation are "very similar," with the only difference being the activation of the CO2 canister.
    Safety and Effectiveness: Data demonstrates the device is as safe and effective as the predicate device, or that any differences in technological characteristics do not raise new questions of safety or effectiveness. (This is generally inferred from the above points for 510(k)s, rather than requiring extensive clinical trials for a modified device).The document states, "Merlin™ has the same intended use and indications and very similar principles of operation and technological characteristics as the predicate device. Therefore, Merlin™ is substantially equivalent." This implicitly means it is considered as safe and effective as the predicate.

    2. Sample size used for the test set and the data provenance:

    • Not applicable (N/A). This 510(k) submission for the Mityvac® Merlin™ is for a physical medical device (obstetrical vacuum assist delivery system), not an AI/ML-based device that would typically rely on a "test set" of data for performance evaluation in the way an AI model does.
    • The evaluation is based on a comparison of the device's design, intended use, and technological characteristics to an existing predicate device (Mityvac® MityOne™). No specific patient "data" or "test set" in the computational sense is mentioned or implied.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • N/A. As explained above, this is not an AI/ML device, so there is no "ground truth" established by experts for a test set in this context. The FDA's review process itself involves expert evaluation by their staff, but this is an regulatory assessment, not a ground-truth labeling process.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • N/A. No test set or associated adjudication is relevant to this type of device submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • N/A. This is a physical medical device, not an AI-assisted diagnostic or treatment device. Therefore, an MRMC study is not applicable.

    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done:

    • N/A. This is a physical medical device; there is no "algorithm" or "standalone" performance in the context of AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • N/A. For this device, the "ground truth" for demonstrating safety and effectiveness relies on the legally marketed status and established safety/effectiveness profile of the predicate device (Mityvac® MityOne™). The manufacturer demonstrates that the new device is fundamentally similar and does not introduce new risks.

    8. The sample size for the training set:

    • N/A. No training set is relevant for this type of device.

    9. How the ground truth for the training set was established:

    • N/A. No training set is relevant for this type of device.

    In summary, the provided document outlines a 510(k) submission for a physical medical device, not an AI/ML device. Therefore, many of the requested details concerning test sets, ground truth, experts, and AI-specific studies are not applicable to this type of regulatory submission. The device's "performance" and "acceptance" are determined through its substantial equivalence to a legally marketed predicate device based on its intended use and technological characteristics, as reviewed by the FDA.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013460

    Validate with FDA (Live)

    Date Cleared
    2002-01-17

    (91 days)

    Product Code
    Regulation Number
    884.4340
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    004 without the vacuum release valve; 005 with the vacuum release valve Regulation Number: 21 CFR §884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDICATIONS FOR USE:

      1. Non-reassuring fetal status.
    • Failure of the fetus to deliver spontaneously following an appropriately 2. managed second stage.
    • Maternal need to avoid voluntary expulsive effort. 3.
      1. Inadequate maternal expulsive effort.
    • Selective shortening of the second stage when the fetal head is at the outlet 5. as defined by ACOG technical bulletin # 196 Aug. 1994.
    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification letter from the FDA regarding the Swift Low Profile Cup (Pro Cup) fetal vacuum extractor. It does not contain any information about acceptance criteria, device performance, or details of a study.

    Therefore, I cannot provide the requested information in the table or answer the subsequent questions based on the provided text. The letter only states that the device is substantially equivalent to legally marketed predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K981260

    Validate with FDA (Live)

    Date Cleared
    1998-09-22

    (169 days)

    Product Code
    Regulation Number
    884.4340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    6000S (Fetal Vacuum Extractors) Dated: July 8, 1998 Received: July 10, 1998 Regulatory Class: II 21 CFR 884.4340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use for fetal vacuum delivery assistance in conditions of: 1) dystocia, 2) uterine inertia, 3) maternal exhaustion (ineffective voluntary effort), or 4) maternal or fetal distress.

    Standard vacuum extraction

    Use for fetal vacuum extraction in conditions of 1) prolonged second stage of labor (arrest of descent) where fetopelvic relationships are adequate, 2) presumed fetal jeopardy which is not considered to be severe, or 3) elective shortening of the second stage for selected maternal or fetal conditions.

    Trial of vacuum extraction

    Vacuum delivery should be regarded as a "trial" 1) if there is arrest of descent in the second stage and fetopelvic relationships are considered to be boderline, or 2) in a mid-pelvic extraction when position and station are known.

    Vacuum extraction should be abandoned and birth completed by cesarean section 1) if no descent (progress) of the head occurs after 2 tractions. 2) if delivery is not achieved or imminent after 4 tractions, or 3) if the vacuum cup detaches ("pops-off") twice.

    Device Description

    The Kiwi Complete Vacuum Delivery Systems are fetal vacuum extractors having a cup to attach to the fetal head, a handle with which to apply withdrawal force, and a vacuum pump built into the handle of the cup/handle combination. This palm pump includes a method for easily generating a vacuum by pumping the handle with the palm. A vacuum indicator and a vacuum relief valve are also integral to the palm pump. Two styles of cups are available; one of soft silicone (VFE-6000S) and the other of the Malmstrom design (VFE-6000M).

    AI/ML Overview

    The provided text describes the Kiwi Fetal Vacuum Extractor and its 510(k) submission for market clearance. While it discusses the device's intended use, basic technological characteristics, and comparison to predicate devices, it does not provide detailed acceptance criteria or a specific study that proves the device meets such criteria.

    The text generally states:

    • "The safety and effectiveness are similar to existing devices as demonstrated in the laboratory and in clinical testing."
    • "Biocompatibility testing shows that the materials used in the Kiwi Fetal Vacuum Extractors are safe for this application."
    • "Effectiveness is the same as the predicate devices."
    • "The laboratory testing verified the performance in terms of vacuum measurement and device integrity."

    However, it lacks the specific data points to fill a comprehensive table of acceptance criteria and reported performance, nor does it detail the methodology, sample sizes, ground truth establishment, or expert involvement for the referred "clinical testing" or "laboratory testing."

    Therefore, I cannot fulfill all parts of your request based on the provided input.

    Here's what can be inferred and what is missing:


    Acceptance Criteria and Study Details for Kiwi Fetal Vacuum Extractor

    Missing Information: The provided document is a 510(k) summary and FDA clearance letter. It states that "clinical testing" and "laboratory testing" were performed to demonstrate safety and effectiveness similar to predicate devices, and that "laboratory testing verified the performance in terms of vacuum measurement and device integrity." However, it does not provide specific acceptance criteria, detailed results, study designs, sample sizes, or ground truth methodologies for these tests.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria for Performance/Effectiveness (Inferred/General statements):

    CriterionReported Device Performance (General Statement/Inferred)
    Safety: BiocompatibilityMaterials shown to be biocompatible and safe for application.
    Effectiveness: Equivalence to PredicateEffectiveness is the same as predicate devices.
    Performance: Vacuum MeasurementPerformance in terms of vacuum measurement verified by laboratory testing.
    Performance: Device IntegrityDevice integrity verified by laboratory testing.
    Clinical Performance: SuccessfulDemonstrated in clinical testing (details of success rates, adverse events, etc., are not provided in this summary).
    Clinical Performance: Similar to PredicateSafety and effectiveness are similar to existing devices as demonstrated in laboratory and clinical testing.

    Note: Specific quantitative acceptance thresholds (e.g., vacuum pressure range, force limits, successful delivery rates) and actual measured performance values are not provided in this document.


    The remaining sections cannot be addressed with specific details from the provided document as the information is not present.


    2. Sample size(s) used for the test set and the data provenance

    • Sample Size (Test Set): Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective nature of the clinical/laboratory testing).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not specified, as details of clinical or laboratory study design are absent.

    4. Adjudication method for the test set

    • Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/Not specified. This device is a physical medical instrument (fetal vacuum extractor), not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical instrument for use by a clinician.

    7. The type of ground truth used

    • Not explicitly stated for the clinical or laboratory tests. For an interventional device like a vacuum extractor, ground truth would typically relate to successful delivery, Apgar scores, maternal/fetal complications, and physical integrity/functionality of the device.

    8. The sample size for the training set

    • Not applicable as this is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable.