510(k) Data Aggregation

    K Number
    K242704

    Validate with FDA (Live)

    Date Cleared
    2025-06-05

    (269 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    44074

    Re: K242704
    Trade/Device Name: Synchrony (20-3000)
    Regulation Number: 21 CFR 882.5810
    Device Classification: Class 2
    Product Code: GZI, IPF, and KQX
    Regulation number: 882.5810
    Classification Regulation(s) | 21 CFR 882.5810 | 21 CFR 882.5810 | 21 CFR 890.5850 | Same.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synchrony (20-3000) is an electrical stimulation device indicated for the following uses:

    Functional Electrical Stimulation (FES).

    • Improvement of hand function and active range of motion in patients with hemiplegia due to stroke or upper limb paralysis due to C5 spinal cord injury.

    NeuroMuscular Electrical Stimulation (NMES).

    • Maintenance and/or increase of hand range of motion.
    • Prevention and/or retardation of disuse atrophy.
    • Increase in local blood circulation.
    • Reduction of muscle spasm.
    • Re-education of muscles.
    Device Description

    The Synchrony (20-3000) system is a functional electrical stimulation (FES) device and a powered muscle stimulator. It is specifically designed to enhance hand functionality in patients with arm paresis through a range of customizable therapies. The system integrates five major components: the Stimulator, Sensor, Patient Interface (electrodes), Motor Point Probe, and Clinician App. The therapies are directed and programmed by a clinician through an iOS based application and then the patient may apply the therapy at their home with the Stimulator and Sensor.

    The Synchrony system supports three primary therapeutic modes: Contralaterally Controlled Electrical Stimulation (CCFES), cyclical Neuromuscular Electrical Stimulation (CNMES), and Functional Task Practice (FTP). The CCFES mode uses sensor data from the patient's unimpaired side to control the timing, intensity, and movement of the paretic hand/arm, synchronizing stimulation with natural motor patterns. The cyclical NMES mode delivers repetitive stimulation to targeted muscles, aiding in muscle re-training, as described in the systematic review of research studies using cyclic NMES. The FTP mode allows for direct control of hand movements during task-oriented exercises, using real-time sensor feedback and stimulator output.

    Stimulation is delivered through up to four electrodes placed on identified motor points, using symmetric balanced biphasic pulses. The system offers preset stimulation amplitudes of 20mA, 40mA, and 60mA, with a pulse width adjustable between 1-250 µsec, providing high precision with 1 µsec resolution. The device's enclosure is constructed from biocompatible ABS Cycoloy, ensuring durability and patient safety, and is rated IP22 for ingress protection.

    The stimulator can operate independently or in conjunction with the Synchrony Sensor, depending on the therapy mode selected. The Sensor, which communicates wirelessly with the Stimulator via Bluetooth Low Energy (BLE), uses a Class I laser that is compliant with 21 CFR 1040.10 and 21 CFR 1040.11 to measure hand opening and closure distances, enabling accurate therapeutic adjustments. Both the Stimulator and Sensor are wirelessly charged using an FCC Part 15 compliant charging pad, enhancing ease of use and handling.

    The Synchrony system is engineered for ease of use with user-friendly control interfaces for both patients and clinicians. It incorporates a range of safety mechanisms to ensure effective and safe operation, making it a versatile and reliable tool for improving hand functionality through personalized therapeutic interventions.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Synchrony (20-3000) device does not contain information about acceptance criteria or a specific study proving the device meets those criteria.

    The letter primarily focuses on establishing substantial equivalence to predicate devices, referencing adherence to recognized standards, and outlining performance differences between the new device and the predicates without providing specific acceptance criteria or study details.

    However, based on the information provided, we can infer some details and highlight the missing information:

    Key Takeaways from the document relevant to performance and safety:

    • Substantial Equivalence: The primary method for demonstrating safety and effectiveness is through substantial equivalence to predicate devices (K123636 and K960457).
    • Standards Compliance: The device claims compliance with numerous IEC, ISO, ASTM, and ANSI standards related to safety, essential performance, usability, electromagnetic compatibility, software lifecycle, risk management, biological evaluation, laser safety, battery testing, shipping, cleaning processes, wireless coexistence, and cybersecurity. These standards inherently contain their own acceptance criteria that the device must meet.
    • Performance Differences: The comparison tables highlight differences in technical characteristics and output specifications. The rationale for these differences consistently asserts "No impact on safety and effectiveness."
    • Clinical Data: The document explicitly states: "No clinical data were reviewed in this submission in support of the subject device." This is a crucial piece of information indicating that clinical trials were not part of this 510(k) submission for performance validation against specific clinical outcomes or acceptance criteria.

    Inferred and Missing Information:

    Since dedicated acceptance criteria and a study proving the device meets them are not explicitly detailed in the provided text, the table below will reflect what is implied by the substantial equivalence and standards compliance, and explicitly state where information is missing.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred from Standards & Equivalence)Reported Device Performance
    Safety:
    IEC 60601-1 Basic Safety & Essential Performance CompliantCompliant
    IEC 60601-1-2 Electromagnetic Disturbances CompliantCompliant
    IEC 60601-1-6 Usability CompliantCompliant
    IEC 60601-1-11 Home Healthcare Environment CompliantCompliant
    IEC 60601-2-10 Nerve and Muscle Stimulators CompliantCompliant
    IEC 60825-1 Safety of Laser Products (Class I laser) CompliantCompliant
    ISO 10993-1 Biological Evaluation CompliantCompliant
    ISO 10993-5 Cytotoxicity CompliantCompliant
    ISO 10993-10 Sensitization CompliantCompliant
    ISO 10993-23 Irritation CompliantCompliant
    Patient Leakage Current below 100 microamperes (normal) & 500 microamperes (single fault)Within acceptable limits (Implied by "No impact on Safety and Effectiveness" and compliance with IEC 60601-1)
    Max Current Density < 2 mA/cm² (for IPF product code)2x2" electrodes: 0.34 mA/cm²2x4" electrodes: 0.17 mA/cm²MPP: 1.33 mA/cm² (All comply with < 2 mA/cm² criterion)
    Automatic Overload Trip functionalityYes
    Automatic Shut Off functionalityYes
    Patient Override Control functionalityYes
    Charged balanced biphasic waveform (0 μC Net Charge)0 μC @ 500Ω
    Effectiveness/Performance:
    Functional Equivalence to Predicate Devices (H200, EMS-2C)Demonstrated via substantial equivalence comparison of technical characteristics and indications for use. Differences described as having "No impact on safety and effectiveness."
    Output Voltage within predicate rangeMax output voltage at 500 Ohms, 2000 Ohms, and 10,000 Ohms is within the maximum output voltage range for predicate devices.
    Output Current effective for muscle contractions within safe densitySynchrony applied peak and RMS current are lower at nominal and maximum load compared to predicate. "Both current densities comply with FDA expectations that maximum current density must be below 2 mA/cm²" and provide effective muscle contractions over their range.
    Pulse Width programmable (1-250 µsec)Achieved (1-250 µsec, 1 µsec resolution)
    Frequency range within predicate ranges (32-37 Hz)Achieved (32-37 Hz)
    ON time within predicate range for safety and effectiveness0-500 µsec (within range)
    OFF time (interpulse delay) allowing action potential propagation100 µsec
    Other:
    IEC 62304 Software Life Cycle Processes CompliantCompliant
    ISO 14971 Risk Management CompliantCompliant
    IEC 62366-1 Usability CompliantCompliant
    ISO 62133-2 Battery testing, Lithium systems CompliantCompliant
    ASTM D4169 Shipping Containers and System CompliantCompliant
    ANSI C63.27 Wireless Coexistence CompliantCompliant
    47 CFR Part 15 Subpart B RF Devices CompliantCompliant
    AAMI TIR57 Cybersecurity Risk Mgt CompliantCompliant
    AAMI TIR97 Cybersecurity Post Market Risk Mgt CompliantCompliant
    ASTM F899-20 Wrought Stainless Steels for Surgical Instruments CompliantCompliant

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable / Not Provided: The document explicitly states: "No clinical data were reviewed in this submission in support of the subject device." Therefore, there is no "test set" of patient data in a clinical sense for this 510(k) submission. Performance testing was primarily based on engineering and bench testing to demonstrate compliance with standards and equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable / Not Provided: As no clinical test set was used, no experts were utilized to establish ground truth in this context for this specific submission. The "ground truth" for compliance was adherence to various engineering, electrical, and biocompatibility standards, and demonstrating substantial equivalence to known predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable / Not Provided: No clinical test set means no adjudication method was employed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable / Not Provided: This device is an external functional neuromuscular stimulator, not an AI-assisted diagnostic imaging device. An MRMC study is irrelevant for this type of device and was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Partially Applicable / Inferred: The device is a physical medical device that delivers electrical stimulation. Its "standalone" performance is assessed through its compliance with electrical output specifications, safety standards, and functional specifications without a human operator. This is implied by the extensive list of performance testing against standards. For example, the stimulator's ability to deliver a specific pulse width and frequency, or for the sensor to accurately measure hand opening, would be part of its standalone performance. However, there's no "algorithm only" in the sense of a pure software diagnostic or inferential AI system. The "Clinician App" and "Sensor" are components of the system functionality, but their performance tests are part of the overall system validation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Standards and Predicate Device Specifications: The "ground truth" for this 510(k) submission is derived from:
      • Recognized Consensus Standards: The extensive list of IEC, ISO, ASTM, and ANSI standards (e.g., IEC 60601-1 for basic safety, ISO 10993 for biocompatibility). The acceptance criteria are defined within these standards.
      • Predicate Device Performance: The technical and output specifications of the Bioness H200 and Med Labs EMS-2C served as benchmarks for comparison, implying that the performance of the predicate devices is an established "truth" for safe and effective similar functionality.

    8. The sample size for the training set

    • Not Applicable: As no clinical data was used and this is not an AI/ML device relying on a training set of patient data, this concept does not apply to this submission. "Training" refers to the engineering development and testing inherent in any device.

    9. How the ground truth for the training set was established

    • Not Applicable: See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243828

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-02

    (140 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K243828**
    Trade/Device Name: Cionic Neural Sleeve (NS-200)
    Regulation Number: 21 CFR 882.5810
    | 21 CFR 882.5810 | 21 CFR 890.5850 |
    | Product codes | GZI, IPF, HCC | GZI, IPF | IPF, GZJ, HCC, GZI
    | 21 CFR 882.5810 | 21 CFR 890.5850 | |
    | Product codes | GZI, IPF, HCC | GZI, IPF | IPF, GZJ, HCC,
    | 21 CFR 882.5810 | 21 CFR 890.5850 | |
    | Product codes | GZI, IPF, HCC | GZI, IPF | IPF, GZJ, HCC,
    | 21 CFR 882.5810 | 21 CFR 890.5850 | |
    | Product codes | GZI, IPF, HCC | GZI, IPF | IPF, GZJ, HCC,

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cionic Neural Sleeve NS-200 is intended to provide ankle dorsiflexion and/or plantarflexion and/or eversion in adult individuals with foot drop and/or to assist knee flexion or extension in adult individuals with muscle weakness related to upper motor neuron disease/injury (e.g. stroke, damage to pathways to the spinal cord). The Cionic Neural Sleeve NS-200 electrically stimulates muscles in the affected leg to provide ankle dorsiflexion and/or plantarflexion and/or eversion of the foot and/or knee flexion or extension; thus, it also may improve the individual's gait.

    The Cionic Neural Sleeve NS-200 may also:

    • Facilitate muscle re-education
    • Prevent/retard disuse atrophy
    • Maintain or increase joint range of motion
    • Increase local blood flow

    As a powered muscle stimulator the Cionic Neural Sleeve NS-200 is indicated for the following conditions:

    • Relaxation of muscle spasm

    As a biofeedback device the Cionic Neural Sleeve NS-200 is indicated for the following conditions:

    • Biofeedback, relaxation and muscle re-education purposes
    Device Description

    Same as the primary predicate device, the Cionic Neural Sleeve NS-100 (K221823), the Cionic Neural Sleeve NS-200 is a platform for the measurement and augmentation of lower limb mobility composed of a body-worn legging, a battery-powered electronic controller and a mobile application. The Cionic Neural Sleeve NS-200 has embedded sensors to measure limb movement and muscle activity. These data are used by the control unit to generate stimulation intended to activate muscles for exercise or functional assistance.

    The Cionic Neural Sleeve NS-200 system sales carton consists of the following components:

    1. SL-200 a fabric sleeve covering the upper and lower leg containing embedded motion sensors and skin-contacting electrodes. Left and right leg sleeves are available in four sizes: extra-small, small, medium, and large.

    2. DC-200 a portable battery-powered Control and Stimulation Unit that connects to, and is worn within the SL-200. The DC-200 communicates over Bluetooth™Low Energy protocol to the Cionic mobile application ("Cionic app").

    3. Power supply and cable to recharge the DC-200 and connect the DC-200 to a user's computer when required.

    4. Adhesive, electrically conductive and replaceable electrode pads.

    5. Electrode cover sheets.

    6. Instructions for Use documents.

    Components are available as accessories to the Cionic Neural Sleeve NS-200 system:

    • Replacement electrode pads.

    The Cionic Neural Sleeve NS-200 requires a password-protected Cionic mobile application that is exclusively available to Cionic Neural Sleeve NS-200 iOS and Android users.

    The Cionic Neural Sleeve NS-200 system consists of a software and hardware architecture that enables users to access a library exercise and programs. Programs can be added and removed from the user's mobile app by the user. All exercise and assistance programs utilize a standard calibration and stimulation user interface that is extendible to future exercise and augmentation programs.

    The following minor hardware/mechanical changes have been made to the NS-100 (K221823) and incorporate in the NS-200:

    1. Simplified Sleeve Electronics
      By relocating the electromyography (EMG) circuitry to the Control Unit (DC-200), the redesigned Sleeve (SL-200) is now lighter, more streamlined, and requires less power—delivering a more comfortable and efficient experience for users.

    2. Enhanced Control Unit Grip
      The Control Unit (DC-200) now features integrated side grips, making it easier to connect and disconnect from the sleeve cable—while maintaining the same compact form factor users love.

    3. More Secure Fit, All Day Comfort
      Enhanced Velcro sections on the Sleeve (SL-200) offer improved adherence and a more secure fit, supporting confident wear and optimal electrode positioning throughout the day.

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for the Cionic Neural Sleeve NS-200. It describes the device's technical characteristics and its substantial equivalence to predicate devices. However, the document does NOT contain information about acceptance criteria or specific study results that prove the device meets performance criteria, especially concerning the new functionality of advanced biofeedback within the NS-200.

    The document states:

    • "The tests listed have been conducted to demonstrate that the Cionic Neural Sleeve NS-200 performs as intended and is substantially equivalent to both predicate devices."
    • "The Cionic Neural Sleeve NS-200 has been verified and validated successfully for its intended use through a combination of original bench testing and verification and validation of all software and firmware."

    This indicates that testing was done, but the details of the acceptance criteria, specific performance metrics, and the results of those tests are not included in this publicly available clearance letter. For example, while it mentions "Stimulation Output Waveforms" and "Stimulation Output Specifications" as tests conducted, it doesn't provide the target waveform characteristics (acceptance criteria) or the measured waveform characteristics (reported performance).

    Similarly, for the "New Functionality in NS-200" related to biofeedback (Visual Biofeedback Mode and Muscle Activated Stimulation Mode), the document describes how it works but doesn't provide any quantitative acceptance criteria for its performance (e.g., accuracy of sEMG detection, latency of stimulation trigger).

    Therefore, based solely on the provided text, I cannot fulfill most of your request for specific acceptance criteria and study results. The information is simply not present in this type of FDA document, which focuses on substantial equivalence for market clearance rather than detailed performance study reporting.

    However, I can extract and infer some high-level information and point out what is missing:


    Inferred/Missing Information based on the Provided Text:

    Based on the nature of a 510(k) clearance and the information provided, the "acceptance criteria" here are implicitly tied to demonstrating substantial equivalence to predicate devices and meeting relevant safety and performance standards. The "study" mentioned is primarily focused on non-clinical bench testing and software validation.

    Table of Acceptance Criteria and Reported Device Performance (Inferred/Missing)

    Feature/Test AreaAcceptance Criteria (Implicit for Substantial Equivalence/Standards)Reported Device Performance (Specifics NOT provided in document)
    Stimulation Output WaveformsShould match specified waveform characteristics (e.g., rectangular, monophasic with hybrid stimulation) and demonstrate stability/consistency."Tests conducted" but specific measurement values (e.g., actual waveform shape, consistency over time) not provided.
    Stimulation Output SpecificationsShould meet specified ranges for Max Output Voltage, Max Output Current, Pulse Width, Frequency, Net Charge, Max Phase Charge.Values provided in comparison tables (e.g., 50V @ 500Ω, 100mA @ 500Ω, 100-400μs, 5-125Hz). Implicitly met, but no specific "performance" results given as "was measured at X."
    Stimulation Virtual Output ChannelsProper functioning and selection of all 24 virtual channels."Tests conducted" but no specific performance data given (e.g., success rate of channel selection).
    Stimulation Output Channel IsolationMeet specified isolation levels (e.g., -33 dB typical, -70 dB crosstalk) and demonstrate no unintended stimulation."Tests conducted" and implied to meet, citing IC manufacturer datasheet and bench test. No specific quantified result.
    Hybrid StimulationOperates as designed, delivering 0 µC net charge using its hybrid method."Tests conducted" and implied to meet. No specific quantified result.
    Stimulation Electrodes Short and/or Open DetectionSystem detects and responds appropriately to short/open circuit conditions."Tests conducted" but no specific performance data given (e.g., detection accuracy, response time).
    Wireless CoexistenceDevice functions without interference from/to other wireless devices."Tests conducted" but no specific performance data given (e.g., interference levels, reliability).
    Electrical Safety (IEC 60601-1, -1-11; -2-10)Compliance with all relevant clauses of these standards (e.g., leakage current limits, patient protection, functional safety for nerve stimulators)."Passed" (stated compliance with standards). No specific measured values.
    Electromagnetic Compatibility (IEC 60601-1-2)Compliance with EMC requirements (e.g., emission limits, immunity to interference)."Passed" (stated compliance with standards). No specific measured values.
    Software Validation (IEC 62304)Software performs as intended, free from critical bugs, meets design specifications (including new biofeedback features)."Verified and validated successfully." No specific test results, bug counts, or performance metrics.
    Usability (IEC 62366, IEC 60601-1-6)Device is safe and effective for use by intended users, minimal use errors."Leveraged from the predicate device submission." No new specific usability study results for NS-200.
    EMG Performance (Sensitivity, Sampling Rate, Range, Bandwidth, Processing)Meet specified values/ranges and accurately detect muscle activity as intended for biofeedback.Values provided (e.g., 0.0298 μV sensitivity, 2kHz sampling rate, +/-2.5*10⁶ μV range, 10-500 Hz bandwidth, RMS processing). Implicitly met, but no specific "performance" results given as "was measured at X."
    Biofeedback Mode FunctionalityVisual biofeedback accurately reflects muscle activation; Muscle Activated Stimulation triggers consistently at 20% max activation.Functionality described, but no quantitative performance data (e.g., accuracy of visual display, trigger latency, false positive/negative rates for stimulation).
    Clinical Performance (Gait Improvement, Muscle Re-education, etc.)Device provides intended therapeutic effects as described in Indications for Use.Supported by "Summary of Literature Clinical Data" referencing an article on "Adaptive Functional Electrical Stimulation." No direct clinical trial results for NS-200 itself are presented here.

    Specific Information Requested and What is Available/Missing:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria: Mostly implicit compliance with safety standards (IEC 60601 series, 21 CFR 898) and technical specifications (e.g., voltage/current ranges, EMG parameters) for substantial equivalence to predicates. Explicit, quantitative acceptance thresholds for performance tests are not provided.
      • Reported Device Performance: Specific quantitative results from testing (e.g., "The measured output voltage was XV, meeting the Y-V criterion") are not provided. The document states that tests "have been conducted" and the device "has been verified and validated successfully," implying compliance, but no actual performance data is listed. The tables compare specifications between devices, not measured performance results.
    2. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: Not provided. The testing described is primarily non-clinical bench testing and software validation, not typically expressed in terms of "sample size" of subjects/patients.
      • Data Provenance: The "Summary of Literature Clinical Data" references an article (https://www.medrxiv.org/content/10.1101/2022.04.27.22273623v2.full), which would be the source of any clinical data used to support the new indication for foot eversion. The document itself does not provide details on the study design or provenance of the data within that article.
      • Retrospective/Prospective: Not specified for the non-clinical tests. For the referenced clinical article, it would depend on the study design of that external publication.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable / Not provided. The document describes non-clinical and software validation tests. There is no mention of "ground truth" adjudicated by experts for a test set in the context of AI/ML performance evaluation, as this is a medical device (neuromuscular stimulator) rather than an AI diagnostic algorithm.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable / Not provided. As above, this type of adjudication is typically for establishing ground truth in image interpretation or similar expert-driven data labeling, which is not described for this device's clearances.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable / Not done (as described). An MRMC study is relevant for AI-assisted diagnostic tools. The Cionic Neural Sleeve is a functional neuromuscular stimulator; its clearance focuses on its mechanism of action, safety, and functionality, not its ability to assist human "readers" in interpreting data.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not explicitly stated as such. The device does have a software component with "advanced biofeedback capabilities." The software validation (IEC 62304) would assess the standalone performance of the algorithms. However, "standalone performance" in the context of AI often refers to metrics like AUC, sensitivity, specificity, etc., which are not provided here for any specific task. The document only confirms that the software was "verified and validated successfully."
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the non-clinical bench testing: The "ground truth" would be established by reference standards, calibrated equipment, and engineering specifications. For instance, a voltage meter would provide "ground truth" for output voltage.
      • For the software validation: "Ground truth" would be defined by the software's design specifications and expected behavior.
      • For the clinical support: The document references a published article ("Augmenting Gait in a Population Exhibiting Foot Drop with Adaptive Functional Electrical Stimulation"). The ground truth within that study would depend on its specific methodology (e.g., kinematic analysis, clinical outcome measures). The 510(k) does not perform primary ground truth establishment for clinical outcomes itself but rather leverages existing literature.
    8. The sample size for the training set:

      • Not provided (and likely not applicable in the traditional AI/ML sense). The software features described (Visual Biofeedback Mode, Muscle Activated Stimulation Mode) are based on sEMG functionality. While such systems might involve some level of calibration or adaptive algorithms, the document doesn't mention a "training set" in the context of a machine learning model that learns from large datasets. Any internal calibration routines would be part of the software's inherent design, not a separate "training set" for a generalizable AI model.
    9. How the ground truth for the training set was established:

      • Not applicable / Not provided for the reasons stated above.

    In summary, this FDA 510(k) clearance document primarily serves to demonstrate substantial equivalence to existing devices based on technical specifications and compliance with recognized safety and performance standards via non-clinical bench testing and software validation. It is not a detailed clinical study report or an AI/ML algorithm validation report that would contain the specific performance metrics and ground truth establishment details you requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241488

    Validate with FDA (Live)

    Device Name
    TrainFES Advance
    Manufacturer
    Date Cleared
    2025-02-05

    (257 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Functional |
    | | Regulation Number | 890.5850 IPF / 882.5810
    |

    Power Muscle Stimulator

    876.5320 882.5810

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For Prescription and Home Use by prescription from a medical professional:
    The TrainFES advanced is a neuromuscular electrical stimulator indicated for use under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions.

    As a powered muscle stimulator, TrainFES advanced is indicated for the following conditions:

    • Relaxation of muscle spasms
    • Prevention or retardation of disuse atrophy
    • increasing local blood circulation
    • immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
    • Maintaining or increasing range of motion
    • Muscle re-education

    As an external functional neuromuscular stimulator (FES), TrainFES Advanced is indicated for the following conditions:

    • Helps to relearn voluntary motor functions of the extremities.

    TrainFES' Advanced intended population is anybody aged 22 or over.

    Environments of use: TrainFES Advanced devices can be used by both therapists and patients, in the clinic, hospitals or at home.

    Platform: TrainFES is a battery-powered, wireless device, accessible through software.

    Device Description

    TrainFES Advanced is a portable functional electrostimulator with 6 channels designed for use in the clinics and hospitals by the medical professionals as well as in the home environment by the patient. This device generates electrical impulses to stimulate the musculature of paralyzed seqments and facilitate both the relearning of movement and neuromodulation of tone.

    TrainFES Advanced is a battery-powered, wireless device, configurable from the TRAINFES App, available for Smartphone and Tablets, which allows you to adjust different parameters and follow a training plan from your smartphone. Session settings can be retrieved from the PC or Cloud.

    AI/ML Overview

    The provided document describes the FDA 510(k) premarket notification for the "TrainFES Advanced" device, a neuromuscular electrical stimulator. The purpose of this submission is to demonstrate substantial equivalence to a predicate device (Stella BIO, K210002).

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for the device's performance in a quantitative manner (e.g., target accuracy, sensitivity, specificity). Instead, the substantial equivalence justification relies on demonstrating that the TrainFES Advanced device performs similarly to or meets the safety and effectiveness standards of the predicate device, K210002.

    The table below summarizes the comparison of key technical characteristics between the TrainFES Advanced and its predicate, Stella BIO, highlighting where "performance" is discussed in terms of meeting relevant standards or being considered equivalent. The reported "performance" for TrainFES Advanced is intrinsically linked to its compliance with these standards and the assertion that differences do not raise new safety or effectiveness concerns.

    Characteristic / SpecificationAcceptance Criteria (implied by predicate comparison)Reported Device Performance (TrainFES Advanced)
    Basic Device Characteristics
    ClassificationClass IIClass II
    Prescription/OTC UsePrescription and Home UsePrescription and Home Use
    Environment of UseClinics, hospitals, and homeClinics, hospitals, and home
    Indications for UseSimilar to predicate (specific conditions)Similar to predicate, with specific conditions listed, and the functions of powered muscle stimulator and external functional neuromuscular stimulator are exactly the same
    Power SourceBattery-powered, compliant with IEC 62133Battery: Li-Ion 3.7V (4000mAh), compliant with IEC 62133
    Method of Line Current IsolationN/A (Battery)N/A (Battery)
    Patient leakage current (Normal)<100μA<100μA
    Patient leakage current (Single fault)<100μA<100μA
    Number of Output ModesCompliant with IEC 60601-1 and IEC 60601-2-102, compliant with IEC 60601-1 and IEC 60601-2-10
    Software/Microprocessor controlYesYes
    Automatic Overload TripYesYes
    Automatic No-Load TripYesYes
    Automatic Shut OffYesYes
    Patient Override ControlYesYes (Button)
    Indicator Display On/Off StatusYes (Led)Yes (Led)
    Indicator Display Low BatteryYesYes
    Indicator Display Voltage/Current LevelYesYes
    StandardsCompliance with relevant IEC standardsISO 13485:2016, Bluetooth: FCC 15.2473, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11, IEC 62133, IEC 62304:2015, IEC 62366
    Compliance with 21 CFR 898YesYes
    WeightConsidered acceptable if within safety/effectiveness218 g (considered similar enough to 112g of predicate not to raise safety issues)
    DimensionConsidered acceptable if within safety/effectiveness130 x 79 x 33mm (considered similar enough to 91.5 x 68.4 x 24mm of predicate not to raise safety issues)
    Housing material & ConstructionPlastic (ABS)ABS Plastic
    Apply PartBFBF
    Powered Muscle Stimulator (PMS)
    Number of channelsCompliant with IEC 60601-1 and IEC 60601-2-106 Channels, compliant with IEC 60601-1 and IEC 60601-2-10
    Synchronous or alternatingAlternating or Synchronous (if safety/effectiveness not impacted)Alternating (difference from predicate not impacting essential performance or safety)
    Method of channel isolationMultiplexerMultiplexer
    Regulated Current or voltageRegulated CurrentRegulated Current
    Time Range1 - 60 min1 - 60 min
    WaveformBiphasicBiphasic
    ShapeRectangular, also other shapes for predicateRectangular (tested and compliant with IEC 60601-2-10)
    Symmetrical or asymmetricalSymmetricalSymmetrical
    Maximum voltage outputTested and compliant with IEC 60601-2-1050V @500Ω, 164V @2kΩ, 170V @10kΩ (tested and compliant)
    Maximum output CurrentTested and compliant with IEC 60601-2-10100mA @500Ω, 82 mA @2kΩ, 17 mA @10kΩ (tested and compliant)
    Current output channel0 - 100 mA0 - 100 mA (steps 1mA)
    Pulse Width50 µs - 400 µs50 µs - 400 µs
    Frequency1 - 140 Hz1 - 140 Hz
    Net charge (µC per pulse)0µC (Balanced pulses)0µC (Balanced pulses)
    Max phase charge @500Ω40 µC40 µC
    Maximum current density (mA/cm²) @500ΩComplies with FDA guidance for PMS 510(k)s0.224 mA/cm² (25 cm² electrode area) (complies with FDA guidance)
    max. power density (W/cm²)Complies with FDA guidance for PMS 510(k)s0.00062 W/cm² (25 cm² electrode area) (complies with FDA guidance)
    Functional Electrical Stimulation (FES)
    All FES parametersSame as PMS parameters, with differences justified as not impacting essential performance or safety.Same as PMS parameters, with differences justified as not impacting essential performance or safety.

    2. Sample Size for the Test Set and Data Provenance:

    The document explicitly states: "No clinical testing was performed in support of this submission."
    Therefore, there is no test set in the sense of clinical data from human subjects described in this document. The evaluation relies on non-clinical (bench) testing, comparison to a predicate device, and compliance with recognized standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    Given that no clinical testing was performed, there were no "experts" involved in establishing ground truth for a clinical test set from human subjects. The assessment relies on the expertise of the engineers and regulatory professionals who performed the non-clinical testing and drafted the 510(k) submission, confirming compliance with engineering standards.

    4. Adjudication Method for the Test Set:

    Not applicable, as no clinical test set was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. The submission relies entirely on non-clinical testing and comparison to a predicate device, without involving human readers or a comparative effectiveness study with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    No, this question doesn't directly apply here. The "TrainFES Advanced" is a physical medical device (a neuromuscular electrical stimulator) that includes software control, not an AI algorithm intended for diagnostic or prognostic purposes on its own. The "device performance" is assessed through its electrical and mechanical characteristics and compliance with safety standards, not through an algorithm's standalone interpretative accuracy.

    7. The Type of Ground Truth Used:

    For this device, the "ground truth" for the non-clinical testing is based on:

    • Engineering specifications and regulatory standards: The device is tested against established limits and requirements defined in standards like IEC 60601 series, IEC 62133, IEC 62304, IEC 62366, and ISO 14971.
    • Predicate device characteristics: The predicate device (Stella BIO, K210002) serves as a benchmark for comparison, implying that its established safety and effectiveness characteristics represent a "ground truth" for what is acceptable in this device class.

    8. The Sample Size for the Training Set:

    Not applicable, as no AI/machine learning algorithm training using a specific dataset is referenced or implied in the context of this 510(k) submission for this type of medical device.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as no AI/machine learning algorithm training using a specific dataset is referenced.

    In summary: The submission for TrainFES Advanced demonstrates substantial equivalence primarily through comprehensive non-clinical (bench) testing against recognized industry standards (IEC series, ISO 14971) and a detailed comparison of its technical specifications to a previously cleared predicate device (K210002). The core argument is that any differences in characteristics do not raise new questions of safety or effectiveness. No clinical studies, human reader studies, or AI algorithm performance evaluations against clinical ground truth were conducted or presented in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240632

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-22

    (261 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Connecticut 06905

    Re: K240632

    Trade/Device Name: Neuvotion NeuStim (NN-01) Regulation Number: 21 CFR 882.5810
    neuromuscular stimulator Classification Name: Stimulator, Neuromuscular, External Functional Regulation Number: 882.5810

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeuStim NN-01 is an electrical stimulation device indicated for the following uses:

    Functional Electrical Stimulation (FES):

    Improvement of hand function and active range of motion in patients with hemiplegia due to stroke or upper limb paralysis due to C5 spinal cord injury.

    NeuroMuscular Electrical Stimulation (NMES):

    • · Maintaining or increasing hand range of motion;
    • · Prevention or retardation of disuse atrophy;
    • · Increasing local blood circulation;
    • · Relaxation of muscle spasms; and
    • · Muscle re-education.

    The NeuStim NN-01 is intended for use on the right arm/right hand only.

    Device Description

    The NeuStim Model NN-01 (herein Device) consists of a voltage-controlled generating source that connects to disposable electrodes that are worn by the patient on the anterior and posterior forearm and around the thumb, these are named the Flexor, Extensor and Thumb Disposable Electrodes respectively. The disposable electrodes consist of multiple contacts that can stimulate the muscles in different areas of the forearm and thumb with 10kHz voltage pulses generated by the NeuStim voltage-controlled stimulator.

    The Device is classified as a Functional Stimulation (FES) and NeuroMuscular Electrical Stimulation (NMES) device. The Device uses galvanic electrical energy to evoke muscle contractions.

    The Device has 181 addressable electrode contacts enabling fine spatial resolution of stimulation to the patient. The Device allows manual electrode scanning to facilitate rapid mapping of effective stimulation points, which can be stored for use in future sessions. This rapid scanning capability is enabled through the use of solid-state relay technology embedded in the flexible patches that mate directly with the disposable electrodes. The motion of the user's (i.e. licensed healthcare professional) finger on the screen of the NeuStim" Tablet electronically steers the stimulation point through the use of a solid-state relay matrix in each patch and thumb dongle, while lighting LED indicators on the back of the patch/dongle under control. This allows immediate, dynamic visual indication of the stimulation location which facilitates rapid mapping (i.e. user can quickly find and return to useful stimulation points). When the user observes desired muscle activations and associated wrist, hand, or finger movements, the related stimulation locations can be stored as a pattern. Up to three of these stored patterns can then be combined into a compound move where the stimulator will continuously time multiplex between them to allow for compound motion. Finally, up to three compound moves or patterns can be combined into a sequence where each of the selected compound moves or patterns are applied for a user configurable number of seconds.

    The Device uses a stimulation pulse repetition rate of 50 Hz. The stimulation pulse is comprised of a multiphasic (5 cycle) waveform with a total duration of 0.5 ms. The stimulation output (peak) voltage is less than 65V (500 Ohm load) and less than 120V (10,000 Ohm load). The stimulation output power will not exceed 0.25 W/cm^2 (averaged over one second) into a 500 Ohm Load.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study information for the Neuvotion NeuStim (NN-01) device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided FDA 510(k) summary does not explicitly detail a table of acceptance criteria with specific quantitative thresholds. Instead, it references compliance with various national and international standards. The "Results" column consistently states "Passed" for each test, indicating that the device met the requirements of these standards.

    PurposeTestingReported Device Performance (Results)
    Verify compliance with ANSI/AAMI ES60601-1 and IEC 60601-1 requirements for basic safety and essential performanceBattery powered device, Type BF Applied PartsPassed
    Verify compliance with IEC 60601-1-2 requirements for basic safety and essential performance Electromagnetic DisturbancesBattery powered device; radiated emissions, radiated field immunity, proximity filed immunity, magnetic field immunity, electrostatic discharge immunityPassed
    Verify compliance with IEC 60601-2-10 requirements for the safety and essential performance of nerve and muscle stimulatorsBattery powered muscle stimulator, Type BF Applied Parts, normal and single-fault, energy and power limitsPassed
    Verify compliance with ISO 10993-1 Biologic Evaluation of Medical DevicesPart 5: Tests for in vitro cytotoxicity; Part 10: Tests for irritation and skin sensitizationPassed
    Verify compliance with IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systemsBattery individual cells and packs certified via third party testingPassed
    Verify software meets requirementsSoftware Verification and ValidationPassed
    Verify user needs are metSystem Design ValidationPassed
    Verify device meets requirementsDesign VerificationPassed
    Verify device meets IEEE/ANSI C63.27 Coexistence TestingTier 2Passed
    Verify device meets labeled Shelf-Life.Shelf-Life Validation StudyPassed
    Verify device meets Adhesive Performance CriteriaAdhesive Performance Validation StudyPassed
    Verify device meets Current Distribution (Dispersion) and Electrode Impedance CriteriaPower and Current Density Validation StudyPassed

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not include information about the sample size used for a test set, nor does it specify the data provenance (e.g., country of origin, retrospective or prospective). The studies listed are primarily non-clinical, focusing on device safety, performance, and compliance with standards. There is a "Summary of Clinical Testing: Not Applicable" statement, which further confirms the absence of a test set derived from human subjects.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Since "clinical testing is not applicable" and no human test set is described, there is no information provided regarding experts used to establish ground truth or their qualifications. The ground truth for the non-clinical tests is established by meeting the requirements of the specific engineering and safety standards referenced.

    4. Adjudication Method for the Test Set

    As there is no described test set involving human subjects and clinical outcomes, an adjudication method is not applicable and not mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned. The submission explicitly states "Summary of Clinical Testing: Not Applicable," indicating that no clinical studies (which would include MRMC studies) were performed for this 510(k). The comparison provided is between the subject device's technical specifications and the predicate device's general characteristics, not a clinical effectiveness study.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    The Neuvotion NeuStim (NN-01) is an external functional neuromuscular stimulator, a physical device that delivers electrical stimulation. It leverages software/firmware control for its operation but is not an AI algorithm that makes diagnostic or treatment decisions in a standalone manner without human interaction. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" as typically applied to AI/ML devices is not directly applicable here. The software verification and validation are part of the overall device performance, which is intrinsically human-controlled (by a licensed healthcare professional).

    7. Type of Ground Truth Used

    The ground truth used for the reported studies is primarily compliance with established regulatory standards and engineering specifications. For example:

    • Safety and Essential Performance: Defined by standards like ANSI/AAMI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10.
    • Biocompatibility: Defined by ISO 10993-1 (specifically Parts 5 and 10).
    • Battery Safety: Defined by IEC 62133-2.
    • Software Requirements: Defined by the software verification and validation process against predefined requirements.
    • User Needs, Device Requirements, Coexistence, Shelf-Life, Adhesive Performance, Current Distribution/Electrode Impedance: These are validated against internal design specifications, user requirements, and relevant standards (e.g., IEEE/ANSI C63.27 for coexistence).

    8. Sample Size for the Training Set

    There is no mention of a training set sample size. Training sets are typically associated with machine learning or AI models. The NeuStim NN-01, while having software, is not described as an AI/ML device that requires a data training set in the context of this 510(k) submission. Its functionality relies on pre-programmed parameters and user-defined stimulation patterns.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no training set mentioned, there is no information on how its ground truth was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233006

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-06-25

    (277 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Ontario L5N 3C9 Canada

    Re: K233006

    Trade/Device Name: MyndMove, MyndMove 2.0 Regulation Number: 21 CFR 882.5810
    Regulation | 21 CFR 882.5810 (External functional neuromuscularstimulator)21 CFR 890.5850

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MyndMove and MyndMove 2.0 are electrical stimulation devices indicated for the following uses:

    Functional electrical stimulation (FES)

    Improvement of arm and hand function and active range of motion in patients with hemplegia due to stroke or upper limb paralysis due to C3-T1 spinal cord injury.

    NeuroMuscular Electrical Stimulation (NMES) for general rehabilitation for:

    • · maintenance and/or increase of arm and hand range of motion,
    • · prevention and/or retardation of disuse atrophy,
    • · increase in local blood circulation,
    • · reduction in muscle spasm, and
    • · re-education of muscles.

    MyndMove and MyndMove 2.0 therapies can only be administered by Occupational or Physical Therapy professionals that have completed MyndMove training by MyndTec on the use of the MyndMove 2.0 Systems.

    Device Description

    The MyndMove and MyndMove 2.0 Systems are neuromodulation devices that deliver short electrical pulses to stimulate muscle contractions and enhance motor recovery following Stroke or Spinal Cord Injury. The MyndMove and MyndMove 2.0 systems comprise the Stimulator device, stimulation electrodes and cables, hand and foot switches, and integrated software.

    The Functional Electrical Stimulators in the MyndMove 2.0 are eight-channel devices with a touch screen user interface (UI). The Stimulator software allows the User to choose from a number of pre-programmed stimulation protocols, to customize the stimulation intensities for each patient, and to document the treatments provided.

    Each MyndMove/MyndMove 2.0 Stimulator is connected, via an IEEE 802.11 a/b/g/n wireless radio adapter, to a cloud-based backend data management system. The primary function of the backend is to manage the various elements of the MyndMove 2.0 ecosystem including but not limited to patient unique IDs and prescriptions, therapist profiles and stimulation protocols.

    MyndMove and MyndMove 2.0 use surface electrical stimulation to produce muscle contractions in different combinations to achieve a wide range of reaching and grasping functions. To achieve these movements, the single use electrodes are placed in various combinations on the surface of muscles in the arm and hand.

    MyndMove and MyndMove 2.0 are based on advanced functional electrical stimulation (FES) principles designed to promote neuroplasticity by influencing the efferent and afferent pathways. Proper sequencing of the muscle contractions as per the MyndMove or MyndMove 2.0 protocols achieve a wide range of functions. During the delivery of MyndMove 2.0 therapy, the patient executes functional tasks with assistance from the MyndMove/MyndMove 2,0 device. The patient attempts to perform the movement and is then assisted to complete the movement when the therapist activates the MyndMove 2.0 stimulation protocol, which generates bursts of short electrical pulses, using surface electrodes, to produce muscle contraction. In this manner, the patient gets both efferent and afferent input as they are actively attempting a movement. Non-damaged pathways can be activated and non-damaged areas of the central nervous system can be trained to substitute for injured areas.

    AI/ML Overview

    The provided text is a 510(k) summary for the MyndMove and MyndMove 2.0 electrical stimulation devices. It aims to demonstrate substantial equivalence to previously cleared predicate devices (MyndMove K170564 and MyndMove 2.0 K212149).

    The key takeaway is that no new acceptance criteria, performance studies, or clinical data were generated or required for this 510(k) submission (K233006). The submission explicitly states: "No additional non-clinical performance testing was conducted because there was no change to the hardware or software since clearances of MyndMove (K170564) and MyndMove 2.0 (K212149)."

    Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" for this specific submission (K233006) refer to the previously established criteria and studies for the predicate devices. The document essentially argues that since the new submission's devices are identical in technological characteristics and therapy protocols to the cleared predicates, they inherently meet the acceptance criteria validated by the predicate studies.

    However, to answer your request fully, I will extrapolate the "acceptance criteria" and "reported device performance" based on the "Substantial Equivalence Discussion" table, which compares the subject devices (MyndMove and MyndMove 2.0) to their predicates across various technical specifications. The acceptance criteria, in this context, are the specifications of the predicate devices that the subject devices are claiming to be identical to.


    1. Table of Acceptance Criteria and the Reported Device Performance

    Since the submission explicitly states there are no changes to the technological characteristics of MyndMove or MyndMove 2.0, the "reported device performance" for the subject devices (MyndMove and MyndMove 2.0 in K233006) is asserted to be identical to the predicate devices. The "acceptance criteria" are the specifications of the predicate devices.

    Feature / Acceptance Criteria (from Predicate Device)Reported Device Performance (MyndMove & MyndMove 2.0 in K233006)SE Comparison
    Indications for Use: FES for improvement of arm/hand function and active ROM in hemiplegia (stroke) or upper limb paralysis (C3-T1 SCI). NMES for general rehabilitation (maintenance/increase ROM, prevention disuse atrophy, increase local blood circulation, reduction muscle spasm, re-education muscles). Administered by trained PT/OT professionals.Identical to PredicateSimilar (minor wording change for NMES definition)
    Anatomical Sites for Stimulation (Upper Limb): Extensor Digitorum, Extensor Carpi Radialis & Ulnaris, Thenar Eminence (Opponens Pollicis Brevis, Flexor Pollicis Brevis & Abductor Pollicis Brevis), Flexor Digitorum Superficialis and Profundus, Biceps, Triceps, Posterior Deltoid, Middle Deltoid, Anterior Deltoid, Pectoralis Major, 1st, 2nd, and 3rd Lumbricals, 2nd Dorsal Interosseous, Serratus Anterior, Lower Trapezius, Upper Trapezius (Scapula).Identical to PredicateIdentical
    Where Used: MyndMove: clinical setting only. MyndMove 2.0: clinical setting and home use by qualified user.Identical to PredicateIdentical
    Regulated current or regulated voltage? Current RegulatedCurrent RegulatedIdentical
    Software/firmware/microprocessor control? YesYesIdentical
    Automatic overload trip? YesYesIdentical
    Automatic no-load trip? YesYesIdentical
    Automatic shut off? YesYesIdentical
    User override control? YesYesIdentical
    Indicator display: Yes (On/off status, Low battery, Voltage/current level)YesIdentical
    Timer range (minutes): Max duration of stimulation program = 120 minutes (configurable in the backend).Max duration of stimulation program = 120 minutes (configurable in the backend).Identical
    Compliance with voluntary standards: MyndMove: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10. MyndMove 2.0: ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-10.Identical to PredicateIdentical
    Compliance with 21 CFR 898? YesYesIdentical
    Power Sources: MyndMove: Mains OR Rechargeable Li-Ion Battery (specified specs). MyndMove 2.0: AHM100PS24C2-8 manufactured by XP-Power (specified specs), Tripp Lite P012-006.Identical to PredicateIdentical
    Method of line current isolation: Galvanic Isolation (transformer) 4000 VACGalvanic Isolation (transformer) 4000 VACIdentical
    Normal condition (uA) (Leakage): Earth Leakage: < 500 uA, Lead to Ground: < 100 uAEarth Leakage: < 500 uA, Lead to Ground: < 100 uAIdentical
    Single fault condition (uA) (Leakage): Earth Leakage: < 1000 uA, Lead to Ground: < 500 uA, Patient on Mains: < 5000 uAEarth Leakage: < 1000 uA, Lead to Ground: < 500 uA, Patient on Mains: < 5000 uAIdentical
    Average DC current through electrodes when device is on but no pulses are being applied (uA): Patient auxiliary current: Normal condition: < 100 uA, Single fault condition: < 500 uAPatient auxiliary current: Normal condition: < 100 uA, Single fault condition: < 500 uAIdentical
    Number of output modes (stimulation protocols): MyndMove: Two Modes (Biphasic Symmetric, Biphasic Asymmetric). MyndMove 2.0: Biphasic Asymmetric.Identical to PredicateIdentical
    Number of output channels: 88Identical
    Synchronous or alternating? Asynchronous (channels are staggered)Asynchronous (channels are staggered)Identical
    Method of channel isolation: Galvanic isolation (transformer): 1500VAC/3kV DCGalvanic isolation (transformer): 1500VAC/3kV DCIdentical
    Weight (Ibs., oz.): 14lbs 6oz.14lbs 6oz.Identical
    Dimensions (in.) [W x H x D] (including accessories): MyndMove: Stimulator: 9.0 in x 4.3 in x 9.8 in. MyndMove 2.0: Stimulator: L 33 cm x W 25 cm x H 13.5 cm.Identical to PredicateIdentical
    Housing materials and construction: Enclosure = Lexan PCB HPX4R, Plastic Bezel = ABS Plastic, Rubber Membrane = Silicone.Identical to PredicateIdentical
    Additional Features: Timing of stimulation controlled by user via hand or foot switch.Identical to PredicateIdentical
    Design and Human Factors: FES, Single Use Gel Electrodes, 8 stimulation channels (up to 16 electrodes), Controlled by embedded touchscreen GUI and hand/foot switches, Preprogrammed stimulation protocols, No fitting required, instructions for electrode placement included, Used by Clinician Only.Identical to PredicateIdentical
    Biphasic Symmetrical Output Mode (MyndMove only - MyndMove 2.0 N/A):
    Waveform: Balanced Biphasic SymmetricalBalanced Biphasic SymmetricalIdentical
    Shape: RectangularRectangularIdentical
    Max Output Voltage (+/- 10%) @ 500, 2k, 10kohms: 160V160VIdentical
    Max Output Current (+/- 10%) @ 500, 2k, 10kohms: 20mA (500/2k), 16mA (10k)20mA (500/2k), 16mA (10k)Identical
    Pulse width: Positive: 150-400 us, Negative: 150-400 usPositive: 150-400 us, Negative: 150-400 usIdentical
    Frequency: 1Hz or 40Hz1Hz or 40HzIdentical
    Symmetrical phases? YesYesIdentical
    Phase Duration: Positive: 150-400 us, Negative: 150-400 usPositive: 150-400 us, Negative: 150-400 usIdentical
    Net Charge (uC per pulse): 2.00 uC2.00 uCIdentical
    Maximum Phase Charge (uC): 9.02 uC9.02 uCIdentical
    Maximum Current Density (RMS) (mA/cm2): 1.36 mA/cm2 (for 1cm x 3cm electrode)1.36 mA/cm2 (for 1cm x 3cm electrode)Identical
    Maximum Power Density (W/cm2): 0.044W/cm20.044W/cm2Identical
    Burst Mode (Pulses per burst, Bursts per second, Burst duration, Duty Cycle): Not Applicable, No Burst ModeNot Applicable, No Burst ModeIdentical
    ON Time (seconds): Not Fixed - User ControlledNot Fixed - User ControlledIdentical
    OFF Time (seconds): Not Fixed - User ControlledNot Fixed - User ControlledIdentical
    Biphasic Asymmetrical Output Mode (MyndMove & MyndMove 2.0):
    Waveform: Balanced Biphasic AsymmetricalBalanced Biphasic AsymmetricalIdentical
    Shape: RectangularRectangularIdentical
    Max Output Voltage (+/- 10%): Positive: 160V; Negative: 40VPositive: 160V; Negative: 40VIdentical
    Max Output Current (+/- 10%): Positive: 20mA (500/2k), 16mA (10k); Negative: 5mA (500/2k), 4mA (10k)Positive: 20mA (500/2k), 16mA (10k); Negative: 5mA (500/2k), 4mA (10k)Identical
    Pulse width: Positive: 150-400 us, Negative: 600-1600 usPositive: 150-400 us, Negative: 600-1600 usIdentical
    Frequency: 1Hz or 40Hz1Hz or 40HzIdentical
    Symmetrical phases? NoNoIdentical
    Phase Duration: Positive: 150-400 us, Negative: 600-1600 usPositive: 150-400 us, Negative: 600-1600 usIdentical
    Net Charge (uC per pulse): 1.85uC1.85uCIdentical
    Maximum Phase Charge (uC): MyndMove: Positive: 9.02 uC, Negative: 8.86 uC. MyndMove 2.0: 9.02 uCIdentical to PredicateIdentical
    Maximum Current Density (RMS) (mA/cm2): MyndMove: 1.07 mA/cm2 (for 1 cm x 3 cm electrodes). MyndMove 2.0: Values for various electrode sizes given.Identical to PredicateIdentical
    Maximum Power Density (W/cm2): 0.044W/cm20.044W/cm2Identical
    Burst Mode (Pulses per burst, Bursts per second, Burst duration, Duty Cycle): Not Applicable, No Burst ModeNot Applicable, No Burst ModeIdentical
    ON Time (seconds): Not Fixed - User ControlledNot Fixed - User ControlledIdentical
    OFF Time (seconds): Not Fixed - User ControlledNot Fixed - User ControlledIdentical

    Regarding the Study that Proves the Device Meets the Acceptance Criteria:

    The current 510(k) (K233006) for MyndMove and MyndMove 2.0 relies entirely on the previous clearances of K170564 (MyndMove) and K212149 (MyndMove 2.0). The submission explicitly states: "No additional non-clinical performance testing was conducted because there was no change to the hardware or software since clearances of MyndMove (K170564) and MyndMove 2.0 (K212149).". This indicates that the "study" proving the device meets acceptance criteria was performed for the predicate devices. The current submission's argument is based on the substantial equivalence of the subject devices to these previously cleared predicates, meaning they share the same technological characteristics, indications for use, and therapy protocols.

    Therefore, the following information relates to the predicate devices' original submissions, as K233006 did not involve new testing. This document does not provide details on the original studies for the predicate devices, but typically such studies would have addressed:

    1. Sample size used for the test set and the data provenance: Not provided in K233006 document. The original predicate submissions (K170564 and K212149) would have outlined any clinical studies, their sample sizes, and whether the data was retrospective or prospective, and its geographical origin.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided in K233006 document. If clinical performance was assessed (likely in the predicate submissions), the ground truth for patient outcomes or assessments would typically be established by qualified medical professionals (e.g., physical therapists, occupational therapists, neurologists), but the specific number and qualifications are not detailed in this current summary.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not provided in K233006 document. This would depend on the nature of any clinical studies for the predicate devices.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable or provided. These devices are functional electrical stimulators for rehabilitation, not imaging analysis tools that typically use MRMC studies with "human readers" and "AI assistance." The "assistance" is the stimulation aiding patient movement, not an AI interpreting medical images.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable in the AI sense. The device requires human professionals (Occupational or Physical Therapy professionals) to administer the therapy, making it a human-in-the-loop system. There's no "algorithm-only" performance for the therapeutic effect. However, the device's software and hardware components would have undergone standalone verification and validation to ensure they meet their design specifications and safety standards as detailed under "Summary of Non-Clinical Testing" (e.g., IEC 62304 for software life cycle processes).
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not explicitly stated for the predicate studies in K233006. For an FES/NMES device, ground truth for clinical effectiveness would typically involve outcomes data such as changes in objective measures of arm/hand function (e.g., Fugl-Meyer Assessment, Action Research Arm Test), active range of motion, muscle strength, and patient-reported outcomes, as assessed by trained therapists or clinicians.
    7. The sample size for the training set: Not provided in K233006 document. If any machine learning components were involved in the development of the predicate devices (which is typical for "coded therapeutic algorithmic protocols"), a training set would have been used. This summary does not give such details.
    8. How the ground truth for the training set was established: Not provided in K233006 document. Similar to point 8, this information would be specific to the development of the predicate devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230997

    Validate with FDA (Live)

    Date Cleared
    2023-12-28

    (265 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Oakland, California 94607

    Re: K230997

    Trade/Device Name: EvoWalk 1.0 System Regulation Number: 21 CFR 882.5810
    Classification: Name: External functional neuromuscular stimulator Product Code: GZI Regulation Number: 21 CFR 882.5810

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EvoWalk 1.0 System is intended to provide ankle dorsiflexion in adult individuals with foot drop and/or to assist knee flexion or extension in adult individuals with muscle weakness related to upper motor neuron disease/injury (e.g., stoke, damage to pathways to the spinal cord). The EvoWalk 1.0 System electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the flexion or extension; thus, it also may improve the individual's gait.

    Device Description

    The EvoWalk 1.0 system is a wearable non-invasive functional electrical stimulation (FES) device intended to provide ankle dorsiflexion in adult individuals with foot drop and/or to assist knee flexion in adult individuals with muscle weakness related to upper motor neuron diseaselinjury (e.g., stroke, damage to pathways in the spinal cord), thus providing treatment for foot drop. EvoWalk 1.0 can be used in a healthcare facility and home settings.

    The EvoWalk 1.0 system consists of the EvoWalk device with an elastic band and built-in rechargeable lithium polymer battery, and the accessories including the EvoWalk mobile application, micro USB type charge cable with charge port, lead cables (one single channel and one dual channel), eight sets of electrodes for each channel, instructions for use (IFU), and a carrying case.

    AI/ML Overview

    The provided text describes the EvoWalk 1.0 System, an external functional neuromuscular stimulator, and its submission for FDA 510(k) clearance. However, it does not include specific acceptance criteria and detailed study results to demonstrate the device meets those criteria in a format that allows for a direct population of the requested table. The document primarily focuses on establishing substantial equivalence to a predicate device (Bioness L300 Go System) through a comparison of technological characteristics and a summary of nonclinical tests performed.

    Therefore, I cannot directly extract and provide the exact acceptance criteria and reported device performance from the given text in a tabular format, nor can I provide information on sample sizes for test and training sets, data provenance, expert ground truth establishment, or clinical study details as requested. The document explicitly states "Summary of Nonclinical Tests Submitted," indicating that only nonclinical (bench and software) tests were submitted, not clinical studies involving human performance metrics against specific acceptance criteria.

    The information that can be extracted relates to the nonclinical testing performed to demonstrate safety and effectiveness and substantial equivalence.

    Here's a breakdown of what is available and what is missing from the provided text according to your request:


    1. Table of Acceptance Criteria and Reported Device Performance:

    This information is not present in the provided document. The document lists the general nonclinical tests performed but does not provide specific quantitative acceptance criteria or the numerical performance results of the EvoWalk 1.0 System against such criteria.

    2. Sample size used for the test set and the data provenance:

    • Sample size for the test set: Not specified. The document mentions "nonclinical testing" and "bench testing" but does not detail sample sizes for any specific components or tests.
    • Data provenance: Not applicable as no clinical test set data is described. The tests are nonclinical (bench and software validation).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The document describes nonclinical tests and software validation, which do not typically involve expert review for ground truth in the same way clinical image analysis or diagnostic AI studies would.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This relates to clinical studies involving human observers, which are not detailed in this document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done according to the provided text. The document focuses on the device's technical performance and substantial equivalence to a predicate, not how human readers improve with AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly, for specific functionalities. The document mentions "Motion analysis to validate the gait event detection algorithm" and "Software validation according to IEC 62304." These are standalone algorithm performance evaluations without human-in-the-loop. However, specific performance metrics are not provided. The device itself is a functional electrical stimulator, and its core function is to deliver stimulation based on algorithms detecting gait events.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the nonclinical tests, the ground truth would be based on engineering specifications, established standards (e.g., IEC 60601 series for electrical safety, EMC, and neuromuscular stimulators), and internal verification protocols. For the "gait event detection algorithm," the ground truth would likely be established through sensor data from controlled movements where gait events are precisely known or measured independently. No specific external expert consensus, pathology, or outcomes data are mentioned for ground truth.

    8. The sample size for the training set:

    • Not specified. The document does not discuss a training set for the device's algorithms, implying that the algorithm might be rule-based or trained using internal data without explicit mention of size.

    9. How the ground truth for the training set was established:

    • Not specified, as no training set is explicitly discussed.

    Summary of Nonclinical Tests Mentioned:

    The document primarily relies on nonclinical testing to support the safety and effectiveness and substantial equivalence of the EvoWalk 1.0 System. The tests performed include:

    • Technological Characteristics of Powered Muscle Stimulation Stimulation Output Waveforms
    • Motion analysis to validate the gait event detection algorithm
    • Electrical Safety according to IEC 60601-1; IEC 60601-1-11
    • Muscle and Nerve Stimulators according to IEC 60601-2-10
    • Electromagnetic compatibility according to IEC 60601-1-2
    • Software validation according to IEC 62304

    The document concludes that "Evolution Devices believes the EvoWalk 1.0 system is substantially equivalent to the Bioness L300 GO system (K173682), and does not raise any new issues or concerns of safety or effectiveness. The evidence presented in the bench testing and non-clinical testing is sufficient to support the safety and effectiveness of the device."

    In essence, the provided text details the regulatory context (510(k) summary) and the types of nonclinical tests conducted to support substantial equivalence, but it does not contain the detailed quantitative results and specific acceptance criteria that would be found in a full clinical study report or a detailed test report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221823

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-21

    (28 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 94133

    Re: K221823

    Trade/Device Name: Cionic Neural Sleeve NS-100 Regulation Number: 21 CFR 882.5810
    California 94133

    Re: K221823

    Trade/Device Name: Cionic Neural Sleeve NS-100 Regulation Number: 21 CFR 882.5810
    Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Requlatory

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cionic Neural Sleeve NS-100 is intended to provide ankle dorsiflexion in adult individuals with foot drop and/or to assist knee flexion or extension in adult individuals with muscle weakness related to upper motor neuron disease /injury (e.g. stroke, damage to pathways to the spinal cord). The Cionic Neural Sleeve NS-100 electrically stimulates muscles in the affected leg to provide ankle dorsiflexion of the foot and/or knee flexion or extension; thus, it also may improve the individual's gait.

    The Cionic Neural Sleeve NS-100 may also:

    • Facilitate muscle re-education
    • Prevent/retard disuse atrophy
    • Maintain or increase joint range of motion
    • Increase local blood flow
    Device Description

    The Cionic Neural Sleeve NS-100 is a platform for the measurement and augmentation of lower limb mobility composed of a body-worn legging, a battery-powered electronic controller and a mobile application. The Cionic Neural Sleeve NS-100 has embedded sensors to measure limb movement and muscle activity. These data are used by the control unit to generate stimulation intended to activate muscles for exercise or functional assistance.

    The Cionic Neural Sleeve NS-100 system sales carton consists of the following components:

      1. SL-100 a fabric sleeve covering the upper and lower leg containing embedded motion sensors and skin-contacting electrodes. Left and right leg sleeves are available in two sizes: small, and medium.
      1. DC-100 a portable battery-powered Control and Stimulation Unit that connects to, and is worn within the SL-100. The DC-100 communicates over Bluetooth™Low Energy protocol to the Cionic mobile application ("Cionic app").
      1. Power supply and cable to recharge the DC-100 and connect the DC-100 to a user's computer when required.
      1. Adhesive, electrically conductive and replaceable electrode pads.
      1. Electrode cover sheets.
      1. Instructions for Use documents.

    Components are available as accessories to the Cionic Neural Sleeve NS-100 system:

    • o Replacement electrode pads.
      The Cionic Neural Sleeve NS-100 requires a password-protected Cionic mobile application that is exclusively available to Cionic Neural Sleeve NS-100 iOS and Android users.

    The Cionic Neural Sleeve NS-100 system consists of a software and hardware architecture that enables users to access a library exercise and augmentation programs. Programs can be added and removed from the user's mobile app. All exercise and assistance programs utilize a standard calibration and stimulation user interface that is extendible to future exercise and augmentation programs.

    AI/ML Overview

    This 510(k) submission (K221823) from Cionic, Inc. is a Special 510(k) submission to add compatibility of the Cionic Neural Sleeve to support the Android operating system and minor software updates to improve overall stability and performance. As such, the submission primarily focuses on software verification and validation, along with wireless coexistence testing. The claim of substantial equivalence is made with respect to the predicate device, also the Cionic Neural Sleeve NS-100 (K213622).

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria / TestReported Device Performance
    New tests performed for this submission:
    Wireless Coexistence (according to FDA Guidance Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: August 14, 2013)Performed to ensure no loss of original functionality. (Specific results not detailed in this summary, but implies successful completion)
    Software Validation (according to IEC 62304)Performed to ensure no loss of original functionality. (Specific results not detailed in this summary, but implies successful completion)
    Leveraged from predicate device (K213622) as modifications did not affect these areas:
    Stimulation Output Waveforms(Assumed to meet criteria from predicate)
    Biological evaluation of medical devices (according to ISO 10993-1)(Assumed to meet criteria from predicate)
    Stimulation Output Specifications(Assumed to meet criteria from predicate)
    Stimulation Virtual Output Channels(Assumed to meet criteria from predicate)
    Stimulation Output Channel Isolation(Assumed to meet criteria from predicate)
    Hybrid Stimulation(Assumed to meet criteria from predicate)
    Stimulation Electrodes Short and/or Open detection(Assumed to meet criteria from predicate)
    Electrical Safety (according to IEC 60601-1; IEC 60601-1-11)(Assumed to meet criteria from predicate)
    Muscle and Nerve Stimulators (according to IEC 60601-2-10)(Assumed to meet criteria from predicate)
    Electromagnetic compatibility (according to IEC 60601-1-2)(Assumed to meet criteria from predicate)
    Usability (according to IEC 62366; IEC 60601-1-6)(Assumed to meet criteria from predicate)
    Mobile Application OS Compatibility (specifically, addition of Android compatibility, risk analysis and testing)Risk analysis and testing demonstrate substantial equivalence. (Implies successful testing for Android OS)

    2. Sample Size Used for the Test Set and Data Provenance:

    The provided 510(k) summary does not detail specific sample sizes for the test sets. For software validation and wireless coexistence, these typically involve testing on the device itself and against relevant communication standards. No patient data or clinical study details are provided in this summary. Therefore, there's no information on the country of origin of data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. This submission is for a software/OS compatibility update and does not involve clinical performance assessment requiring expert opinion for ground truth establishment.

    4. Adjudication Method for the Test Set:

    Not applicable. The tests performed are primarily engineering and software validation tests (wireless coexistence, software validation) and do not involve subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. This submission is for a software and OS compatibility update, not for establishing clinical effectiveness or comparing human reader performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm-only without human-in-the-loop performance) was done:

    Not explicitly stated in terms of "standalone performance" in the context of an algorithm assessing medical images or data without human input. However, the software validation and wireless coexistence tests verify the system's technical functionality, which is an evaluation of the system's independent operation (without direct human user intervention during the specific test runs, although the device is designed for human interaction). There's no separate mention of an algorithm's standalone performance in a diagnostic or clinical decision-making capacity.

    7. The Type of Ground Truth Used:

    For the software and wireless coexistence tests, the ground truth would be defined by the technical specifications and standards (e.g., IEC 62304 for software, FDA guidance for wireless coexistence, Bluetooth Low Energy specifications for communication). The purpose of the tests is to verify that the device's software and communication capabilities adhere to these established technical requirements.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/ML device submission in the sense of a new algorithm being trained. The submission refers to "minor software updates" and "stability and performance" improvements. Therefore, there's no mention of a training set or machine learning model training in this context.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for an AI/ML algorithm mentioned in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213925

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-04-25

    (130 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: MyoCycle MC-2 (Home / Home + / Pro / Pro +)
    Regulation Number: 21 CFR 882.5810
    Classification product code: GZI
    Classification panel: Neurology
    Regulation number: 882.5810

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MyoCycle is intended for general rehabilitation for:

    1. Relaxation of muscle spasms
    2. Prevention or retardation of disuse atrophy
    3. Increasing local blood circulation
    4. Maintaining or increasing range of motion
    5. Muscle re-education
    Device Description

    The MyoCycle is a stationary Functional electrical stimulation (FES) cycle ergometer which is composed of:

    1. a motorized cycle ergometer
    2. an FES controller /stimulator
    3. a stimulation cable which connects the controller / stimulator to cutaneous electrodes (at maximum 20 cutaneous electrodes for 10 channels)
    4. cutaneous electrodes

    This system allows a person with impaired lower extremity movement to undertake cycle ergometry both actively (utilized FES evoked lower extremity muscle contractions) and passively (utilizing power developed by the ergometer's motor). The electric motor inside the MyoCycle ensures that the pedals always rotate at 35 revolutions per minute (r/min) (isokinetic cycling), so it automatically assists or resists the patient, depending on the magnitude and direction of the torque applied to the crank by the patient.

    The system software has four different configurations:

    1. Pro – allows for multiple users by storing unique user data that can be accessed by a unique user identification code; extended FES parameter adjustability; up to six channels of stimulation
    2. Home – only stores data for one user; limited FES parameter adjustability; up to six channels of stimulation
    3. Pro Plus (+) – same as Pro but allows for up to ten channels of stimulation and uses patient cabling with additional leads
    4. Home Plus (+) – same as Home but allows for up to ten channels of stimulation and uses patient cabling with additional leads
    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for the MyoCycle MC-2 device. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria, reported device performance metrics (numerical results), details about a test set (sample size, provenance), expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance data.

    The document primarily focuses on demonstrating substantial equivalence through:

    • Comparison of Indications for Use: Showing the MyoCycle MC-2 has substantially equivalent indications to the predicate device, with an added "muscle re-education" indication that is standard for this type of device and does not raise new safety/effectiveness questions.
    • Comparison of Technological Characteristics: Highlighting similarities and minor differences (e.g., increased stimulation channels, wider parameter range, larger touchscreen, wireless communication, database interface) that do not raise new safety/effectiveness questions.
    • Performance Data (Bench Testing): Stating that testing was conducted to ensure equivalent functionality and verify performance to specifications, primarily through a review of documentation and bench testing against recognized consensus standards (e.g., electrical safety, EMC, usability).

    Therefore, I cannot populate most of the requested sections as the information is not present in the provided text.

    Here's what can be extracted based on the document:


    Acceptance Criteria and Study Details for MyoCycle MC-2

    The FDA 510(k) submission for the MyoCycle MC-2 primarily relies on demonstrating substantial equivalence to a predicate device (MyoCycle MC-HP-1, K170132) rather than presenting new clinical performance data with numeric acceptance criteria for specific outcomes. The "performance data" referred to in the document pertains to bench testing against recognized consensus standards to ensure safety and functionality, not a clinical study to evaluate device effectiveness in patients.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes no specific quantitative acceptance criteria or reported device performance metrics in a clinical context (e.g., "device achieved X% accuracy," "device improved range of motion by Y degrees"). Instead, the "acceptance criteria" are implied to be the successful demonstration of:

    • Substantially equivalent Indications For Use.
    • Similar technological characteristics that do not raise new questions of safety or effectiveness.
    • Compliance with recognized consensus standards for device safety and electrical performance.
    Acceptance Criteria CategoryReported Device Performance (as per document)
    Indications for Use Equivalence"The MyoCycle MC-2 device's Indications for Use (IFU) Statement is substantially equivalent to that of the predicate device... addition of this indication... does not raise any new questions of safety and effectiveness."
    Technological Characteristics Equivalence"The MyoCycle MC-2 has similar technological characteristics and output specifications to those of the predicate device... The different characteristics and specifications do not raise new questions of safety and effectiveness."
    Safety and Functional Performance"MYOLYN's non-clinical testing, including bench testing against recognized consensus standards, have demonstrated that the MC-2 device is as safe and effective as the predicate device and is therefore substantially equivalent."
    Stimulator OutputOutput characteristic measurement of new device confirmed technical specifications: 0-140 mA charge balanced stimulator (predicate: 0-126 mA).
    Compliance with StandardsSystem testing made general use of recognized consensus standards (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2, 60601-1-6, 60601-1-11, 60601-2-10).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a test set in the conventional sense of a patient-based clinical study. The "testing" mentioned is primarily non-clinical bench testing and review of documentation. Therefore, information on sample size and data provenance for a clinical test set is not provided.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable, as no clinical test set requiring expert ground truth or adjudication is described.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not mentioned or described in the provided document. The submission focuses on substantial equivalence based on technological characteristics and bench testing, not on comparative clinical effectiveness with human readers/users.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. The MyoCycle MC-2 is a physical medical device (an FES cycle ergometer) rather than a software algorithm intended for standalone diagnostic or interpretive performance. Its functionality is tied to its interaction with a user.

    7. The Type of Ground Truth Used

    For the bench testing and design verification, the "ground truth" would be established by the specifications defined for the device and the requirements of the recognized consensus standards. For example, an electrical safety standard dictates specific performance criteria that the device must meet, and this standard serves as the "ground truth" for compliance. There is no mention of patient-level "ground truth" (e.g., pathology, outcomes data) for clinical efficacy.

    8. The Sample Size for the Training Set

    Not applicable. The document describes a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212149

    Validate with FDA (Live)

    Device Name
    MyndMove 2.0
    Manufacturer
    Date Cleared
    2022-03-03

    (237 days)

    Product Code
    Regulation Number
    882.5810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Mississauga, Ontario Canada

    Re: K212149

    Trade/Device Name: MyndMove 2.0 Regulation Number: 21 CFR 882.5810
    Regulation Number | 882.5810 External functional neuromuscularstimulators, and890.5850 Powered

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MYNDMOVE is an electrical stimulation device indicated for the following uses:

    Functional Electrical Stimulation (FES)

    Improvement of ann and hand functions and active range of motion in patients with hemplegia due to stroke or upper limb paralysis due to C3-T1 spinal cord injury.

    NeuroMuscular Electrical Stimulation (NMES)

    · Maintenance and/or increase of arm and hand range of motion

    • · Prevention and/or retardation of disuse atrophy
    • · Increase in local blood circulation
    • · Reduction of muscle spasm
      · Re-education of muscles.

    Myddlove therapy can only be administered by Occupational or Physical Therapy professionals that have completed MyndMove training by Mynd Tec on the use of the MyndMove System.

    Device Description

    The MyndMove 2.0 System is a neuromodulation device that delivers short electrical pulses to stimulate muscle contractions and enhance motor recovery following Stroke or Spinal Cord Injury. The MyndMove system comprises the Stimulator device, stimulation electrodes and cables, hand and foot switches, and integrated software. The MyndMove 2.0 Functional Electrical Stimulator is an eight-channel device with a touch screen user interface (UI). The Stimulator software allows the User to choose from a number of pre-programmed stimulation protocols, to customize the stimulation intensities for each patient, and to document the treatments provided.

    Each MyndMove Stimulator is connected, via an IEEE 802.11 a/b/q/n wireless radio adapter, to a cloud-based backend data management system. The primary function of the backend is to manage the various elements of the MyndMove ecosystem including but not limited to patient unique IDs and prescriptions, therapist profiles and stimulation protocols. MyndMove 2.0 Therapy can only be administered by trained MyndMove Therapy professionals (TMMT) with assigned Therapist Log-In IDs. TMMTs are Occupational or Physical Therapy professionals that have completed MyndMove training by MyndTec on the use of the MyndMove system.

    MyndMove 2.0 uses surface electrical stimulation to produce muscle contractions in different combinations to achieve a wide range of reaching and grasping functions. To achieve these movements, the single use electrodes are placed in various combinations on the surface of muscles in the arm and hand.

    MyndMove 2.0 is based on advanced functional electrical stimulation (FES) principles designed to promote neuroplasticity by influencing the efferent and afferent pathways. Proper sequencing of the muscle contractions as per the MyndMove 2.0 protocols achieve a wide range of reaching and grasping functions. During the delivery of MyndMove 2.0 therapy, the patient executes functional tasks with assistance from the therapist and the MyndMove 2.0 device. The patient attempts to perform the movement and is then assisted to complete the movement when the therapist activates the MyndMove 2.0 stimulation protocol, which generates bursts of short electrical pulses, using surface electrodes, to produce muscle contraction. In this manner, the patient gets both efferent and afferent input as they are actively attempting move and as the arm and/or hand is moved into the instructed position. Non-damaged pathways can be activated and non-damaged areas of the central nervous system can be trained to substitute for injured areas.

    AI/ML Overview

    The provided text is an FDA 510(k) summary for the MyndMove 2.0 device, which is an electrical stimulation device. The document primarily focuses on demonstrating substantial equivalence to a predicate device (MyndMove K170564) rather than providing a detailed study proving the device meets general acceptance criteria in a clinical performance context.

    Therefore, many of the requested details about acceptance criteria for clinical performance and specific study methodologies (like sample size for test sets, number of experts, adjudication, MRMC studies, standalone algorithm performance, and training set details) are not present in this type of regulatory submission document. This 510(k) summary focuses on technical equivalence and safety standards compliance due to minor modifications to an already cleared device.

    However, I can extract information related to technical performance and safety compliance acceptance criteria, and the study (testing) that proves the device meets these criteria, based on the provided text.

    Here is a breakdown of the available information:


    Acceptance Criteria and Device Performance (primarily technical and safety-related):

    Acceptance Criterion (Category)Reported Device Performance (MyndMove 2.0)
    Indications for UseIdentical to predicate device (MyndMove): Functional Electrical Stimulation (FES) for improvement of arm and hand functions and active range of motion in patients with hemiplegia due to stroke or upper limb paralysis due to C3-T1 spinal cord injury. NeuroMuscular Electrical Stimulation (NMES) for maintenance/increase of arm/hand ROM, prevention/retardation of disuse atrophy, increase in local blood circulation, reduction of muscle spasm, and re-education of muscles.
    Output CharacteristicsBiphasic Asymmetrical Output Mode:- Waveform Shape: Rectangular- Max Output Voltage (+/- 10%): Positive: 160V @ 500Ω, 2kΩ, 10kΩ; Negative: 40V @ 500Ω, 2kΩ, 10kΩ (Identical to predicate)- Max Output Current (+/- 10%): Positive: 20mA @ 500Ω, 2kΩ, 16mA @ 10kΩ; Negative: 5mA @ 500Ω, 2kΩ, 4mA @ 10kΩ (Identical to predicate)- Pulse Width: Positive: 150-400 µs; Negative: 600-1600 µs (Identical to predicate)- Frequency: 1Hz or 40Hz (Identical to predicate)- Symmetrical phases: No (Identical to predicate)- Net Charge (per pulse): 1.85µC (Identical to predicate)- Maximum Phase Charge: 9.02µC (Equivalent to predicate)- Maximum Current Density (RMS) using smallest electrode: 0.93 mA/cm2 (for 1x3 cm electrode). (Meets criteria < 2mA/cm2 limit identified in IEC 60601-2-10).- Maximum Power Density: 0.044W/cm2 (Identical to predicate).Note: Biphasic Symmetrical Output Mode removed in MyndMove 2.0, deemed "Equivalent" based on overall safety and performance.
    Electrical Safety Standards ComplianceComplies with:- IEC 60601-2-10 Ed 2.1 2016-04 (nerve & muscle stimulators)- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (basic safety & essential performance)- IEC 60601-1-2 Ed 4.0 2014-02 (EMC)- IEC 60601-1-11 Ed 2.0 2015-01 (home healthcare environment)- IEC 62304:2006 (software life cycle)- 60601-1-6:2010 + A1:2013 (usability).Reported performance: Earth Leakage (< 500 uA normal, < 1000 uA single fault); Lead to Ground (< 100 uA normal, < 500 uA single fault); Patient on Mains (< 5000 uA single fault); Patient auxiliary current (< 100 uA normal, < 500 uA single fault). All meet specified limits.
    Mechanical/Physical Changes- Change of stimulator tablet material (due to availability)- Removal of internal battery- Addition of IP22 rated ingress protection- Addition of an optional cart- Max 20mA output current for all electrode sizes (was varied by size in predicate)- Addition of scapula protocols
    UsabilityNo additional usability testing required as no change to hardware or software design affecting usability.
    Maximum net charge per pulse & phase chargeNet charge per pulse: 1.85 µC (Identical to cleared device)Maximum phase charge: 9.02 µC (Identical to cleared device)
    Peak current densityIn all cases, peak current density does not exceed 2mA/cm2 (limit identified in IEC 60601-2-10).

    Study (Testing) Proving Device Meets Acceptance Criteria:

    1. Sample size used for the test set and the data provenance:

      • For the technical performance and safety standard compliance, no specific sample size for a "test set" in a clinical trial sense is mentioned. Compliance was verified through Third Party testing for the relevant IEC standards and internal performance testing.
      • For the scapula anatomical site addition: "Testing was performed... on 3 healthy volunteers."
      • Data Provenance: Not explicitly stated for all testing. For the scapula protocol validation, it was performed at "a clinical site called Lyndhurst and the Toronto Rehabilitation Institute of the University Health Network of the University of Toronto" (Canada). This implies prospective testing for this specific modification. General IEC compliance testing is typically laboratory-based.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document as it's not a clinical performance study involving expert reads for ground truth.
      • For the scapula protocol validation, the "objective was to test each protocol on each subject to ensure that the induced motions achieve the desired motions per the described use cases." This likely involved observation by a qualified professional (e.g., physiotherapist), but no specific "expert" panel or "ground truth" establishment in a traditional clinical validation sense is described.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not provided as it's not a clinical performance study with subjective assessments requiring adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done or reported. The device is an electrical stimulator, not an AI imaging or diagnostic tool where MRMC studies are typically performed.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The device is a medical device, not an AI algorithm intended for standalone diagnostic performance. Therefore, no such standalone algorithmic performance study was done or reported.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For technical and safety performance: Compliance with recognized international standards (e.g., IEC 60601 series) serves as the "ground truth" or benchmark. This is verified through objective testing (oscilloscope tracings, leakage current measurements, etc.).
      • For the scapula protocol addition: The "ground truth" was whether the "induced motions achieve the desired motions per the described use cases." This appears to be a functional verification, likely observed by therapists, rather than a diagnostic ground truth.
    7. The sample size for the training set:

      • This information is not applicable as the device is not an AI or machine learning model that requires a training set in the conventional sense.
    8. How the ground truth for the training set was established:

      • This information is not applicable for the reasons stated above.

    Summary of the Study/Testing:

    The "study" or testing described in this 510(k) summary is primarily focused on demonstrating technical and safety compliance of the modified MyndMove 2.0 device with relevant medical device standards, and confirming that the modifications (e.g., removed battery, new tablet material, home-use capability, new anatomical protocols, altered current for electrodes) do not negatively impact its safety or essential performance compared to the predicate device.

    • Main tests: Third-party testing for compliance with IEC 60601-2-10, ES60601-1, 60601-1-2, 60601-1-11, 62304, and 60601-1-6. Internal performance testing.
    • Specific validation for new protocols: For the addition of scapula protocols, there was "planned validation testing on 3 healthy volunteers" to ensure "induced motions achieve the desired motions." This was done at a clinical site, with "All verification and validation... successfully performed per the design and development plans and under REB decision."

    The document concludes that the MyndMove 2.0 device is "substantially equivalent" to its predicate based on these technical and functional assessments, not on clinical performance metrics typically associated with diagnostic AI or imaging devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213043

    Validate with FDA (Live)

    Date Cleared
    2022-02-25

    (156 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | 21 CFR 890.585021 CFR 876.532021 CFR 882.505021 CFR 882.589021 CFR 882.5810

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NMES Relaxation of muscle spasms Prevention or retardation of disuse atrophy Increasing local blood circulation Muscle Re-education Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Maintaining or increasing range of motion

    EMG Triggered Stimulation (ETS) (nonimplanted electrical continence device only) Acute and ongoing treatment of stress, urge or mixed urinary incontinence and where the following results may improve urinary control: Inhibition of the detrusor muscles through reflexive mechanisms of pelvic floor muscles Incontinence treatment for assessing EMG activity of the pelvic floor and accessory muscles (abdominal or gluteal)

    TENS Symptomatic relief and management of chronic (long-term), intractable pain Adjunctive treatment in the management of post-surgical pain and post traumatic acute pain

    EMG Biofeedback, relaxation muscle training and muscle re-education

    Device Description

    Levator Elite(Model LE9011) is a single channel, hand-held, non-sterile, battery-powered, multi-patient device intended to be used by adult patients under the supervision of a trained clinical healthcare provider. The device contains EMG biofeedback, TENS (Transcutaneous Electrical Nerve Stimulation), ETS (EMG triggered stimulation) and NMES (Neuromuscular Electrical Stimulator). Each of them has pre-set and custom programs. The parameters of the device are controlled by the buttons, the levels of intensity are adjustable to the needs of the patient and treatments prescribed by their healthcare providers.

    The device is designed to provide safe and effective electrical stimulation by sending small electrical currents to underlying nerves and muscle groups via electrode pads applied on the skin or through a vaginal probe/rectal probe (for incontinence treatment protocols only). It can be used with or without linkage to a PC. Connecting the device with the PC via USB cable, the data can be transmitted between PC and device (It needs purchase the PC software Nu-Tek System and USB connection cable).

    For EMG biofeedback, EMG is for detecting the signal of muscle, which display muscle strength via EMG biofeedback bar graph or waveform format viewed on the LCD screen of the unit. Surface EMG is used for recording from superficial muscles in clinical or kinesiological protocols, where intramuscular electrodes are used for investigating deep muscles or localized muscle activity.

    For NMES is the elicitation of muscle contraction using electric impulses. The impulses are generated by a device and delivered through the electrodes in direct proximity to the muscles to be stimulated or via the probe. The impulses mimic the action potential coming from the central nervous system, causing the muscles to contract. NMES is both a form of electrotherapy and of muscle training. Neuromuscular Stimulation has been used to stimulate muscle and nerve fibers for muscle strengthening, maintenance of muscle mass and strength during prolonged periods of immobilization, selective muscle retraining.

    For ETS (i.e. EMG triggered stimulation), ETS involves initiating a voluntary contraction for a specific movement until the muscle activity reaches a threshold level. As soon as the EMG activity reaches a target threshold then an assisting electrical stimulus begins which helps to support the contracted muscle. A microprocessor connected to the surface electrodes, vaginal probe or rectal probe monitors the EMG activity levels as well as administers the neuromuscular stimulation. The target threshold could be set to automated regime, when it goes up and down depending on the running muscle performance.

    For TENS, the device provides a non-invasive, low-risk nerve stimulation in order to reduce pain (both acute and chronic). In TENS, mild electrical impulses are transmitted through the skin via surface electrodes to relieve muscle pain by modifying the body's pain perception. TENS does not cure problematic physiological conditions: it only helps to control the pain perception.

    LE9011 consists of the following elements:

    • Main device
    • Pedestal
    • Lead wire
    • Electrode pad
    • Vaginal probe
    • Rectal probe (optional)
    • PC Software(optional)
    • USB Cable(optional)
    AI/ML Overview

    The provided text does not contain acceptance criteria or a study that proves the device meets acceptance criteria in the way typically expected for a medical device's clinical performance.

    Instead, this document is a 510(k) Summary for a medical device (Levator Elite Model LE9011), which aims to demonstrate substantial equivalence to previously cleared predicate devices. The focus is on comparing the proposed device's technical characteristics, indications for use, and performance data (primarily non-clinical) to those of the predicate devices.

    Here's a breakdown of what is and isn't present, based on your request:

    What's missing that you asked for:

    • Acceptance Criteria for clinical performance: The document does not specify quantitative clinical acceptance criteria (e.g., minimum sensitivity, specificity, accuracy, or specific thresholds for clinical improvement).
    • Study Proving Device Meets Acceptance Criteria: There is no clinical study described that directly evaluates the device's performance against predefined clinical acceptance criteria.
    • Reported Device Performance (against clinical criteria): Because there are no clinical acceptance criteria or clinical studies described, there's no reported device performance in those terms.
    • Sample size for test set: No test set of patients/cases is described for clinical evaluation.
    • Data provenance (country, retrospective/prospective): Not applicable as no clinical study is presented.
    • Number of experts for ground truth & qualifications: Not applicable.
    • Adjudication method for test set: Not applicable.
    • Multi-reader multi-case (MRMC) comparative effectiveness study: Not mentioned or conducted.
    • Effect size of human readers with/without AI assistance: Not applicable as it's not an AI-assisted diagnostic device and no MRMC study was done.
    • Standalone (algorithm-only) performance: Not applicable as it's a physical electrical stimulator, not an algorithm-based diagnostic tool.
    • Type of ground truth used: Not applicable as no clinical study is presented.
    • Sample size for training set: Not applicable as it's not an AI/ML device that requires a training set in that context.
    • How ground truth for training set was established: Not applicable.

    What is provided regarding "Performance Data" in the 510(k) Summary (Section 6):

    The "Performance Data" section focuses on non-clinical testing to ensure the device is safe and performs as intended, and that any technological differences from predicates do not raise new questions of safety or effectiveness.

    1. Biocompatibility testing:

      • Acceptance Criteria (implicit): Compliance with ISO 10993-1.
      • Reported Performance: "The biocompatibility evaluation for the accessories (Electrode pad, Vaginal probe and Rectal probe) was conducted in accordance with the International Standard ISO 10993-1..."
      • Tests Conducted: Cytotoxicity, Sensitization, Irritation.
      • This is not a "study that proves" clinical effectiveness, but rather a demonstration of material safety.
    2. Electrical safety and electromagnetic compatibility (EMC):

      • Acceptance Criteria (implicit): Compliance with relevant IEC standards.
      • Reported Performance: "The system complies with the IEC 60601-1, IEC60601-1-11 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC."
      • This is technical performance testing, not clinical.
    3. Software Verification and Validation Testing:

      • Acceptance Criteria (implicit): Compliance with FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
      • Reported Performance: "Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance..." "The software for this device was considered as a 'Moderate' level of concern."
      • This ensures the software functions correctly and safely, not clinical effectiveness.
    4. Output waveform Testing:

      • Acceptance Criteria (implicit): Verification of output specifications according to IEC 60601-2-10.
      • Reported Performance: "oscilloscope tracing diagrams describing the electrical output waveform was provided to verify the output specifications of the device according to IEC 60601-2-10."
      • This confirms the device delivers the specified electrical output, which is a technical performance aspect.

    Conclusion from the document:

    The document concludes that the device is substantially equivalent to the predicate devices and that any technological differences do not raise new questions regarding safety and effectiveness. This is the premise of a 510(k) submission – to demonstrate equivalence to an already cleared device, often without new clinical studies if non-clinical data suffice.

    In summary, based only on the provided text, a comprehensive response to your structured request for acceptance criteria and clinical study details is not possible because the document focuses on non-clinical performance and substantial equivalence rather than a clinical trial demonstrating performance against specific clinical acceptance criteria.