    510k Summary Text (Full-text Search) :

    Primary Predicate Device:**
    Technomed EEG / EMG needle electrodes(K990015)
    Product Code: 21 CFR § 882.1350
    Electrodes (K130136)
    Product Code: 21 CFR § 890.1385 IKT
    Subsequent Product Code21 CFR § 882.1350
    predicate device |
    | Regulation number | 21 CFR 890.1385 | 21 CFR 890.1385 | 21 CFR 890.1385 | 21 CFR 882.1350

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Concentric Needle Electrodes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalograph (EEG), electromyography (EMG) and nerve potential signal.

    Device Description

    The Disposable Concentric Needle Electrode (DCNE) is a recording electrode used for recording of bio-electric, electromyography (EMG) or electroencephalography (EEG) signals in patients undergoing clinical neurophysiological testing. It is inserted into muscle tissue of patient, acting as an interface between the EMG equipment and the patient. It is connected to the EMG cable (not in this submission), which is connected to the EMG equipment to detect bio-electric signals distally.
    The subject devices are invasive, single-use and supplied sterile. It is used by healthcare professionals, specifically trained and certified in electrophysiology and recording/monitoring techniques.
    The subject devices consist of 6 variants. The differences between the variants are the diameter, length of the needle and the color of the handle.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252842

    Validate with FDA (Live)

    Date Cleared
    2026-01-11

    (125 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    to predicate, SafeOp 3 System (K234092) |
    | Regulation Number (21 CFR) | §882.1870, §882.1320, §882.1350
    , §890.1375, §874.1820, §874.1820 | §882.1870, §882.1320, §882.1350, §890.1375, §874.1820, §874.1820

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, extracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.

    SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.

    Device Description

    The SafeOp™ 3: Neural Informatix System (SafeOp 3 System), consists of the SafeOp patient interface with power supply and IV pole mount, the Alpha Informatix Tablet with docking station and power supply and a data transfer USB cable. Associated disposable accessories consists of an electrode harness, surface and/or subdermal needle electrodes, MEP Activator, Cranial Hub, Delta Dilators and Stimulating Probe and Delta Clip contained in various kits.

    The subject device is intended for use by trained healthcare professionals, clinical neurophysiologists/technologists and appropriately trained non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by both clinical and trained non-clinical personnel.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the SafeOp 3: Neural Informatix System (K252842) primarily describe the device's substantial equivalence to a predicate device (K234092) and outline a Predetermined Change Control Plan (PCCP) for future modifications. However, it does not include a detailed study proving the device meets specific acceptance criteria in the context of clinical performance or AI algorithm validation for the initial clearance.

    The "Performance Data" section explicitly states: "Determination of substantial equivalence is not based on an assessment of clinical performance data." and refers to nonclinical performance testing (functional, system, software requirements) and usability testing.

    Therefore, the following information is extracted and synthesized based on the available document, but a comprehensive study proving acceptance criteria for all aspects of neural informatic performance is not present. The document focuses on the future PCCP for algorithmic changes, which will have performance criteria.


    Acceptance Criteria and Reported Device Performance

    The core of the documentation for this 510(k) clearance is about demonstrating substantial equivalence to an existing predicate device (SafeOp 3™: Neural Informatix System, K234092). Therefore, the "acceptance criteria" for the cleared device itself are primarily its ability to match the predicate's performance and specifications, which is shown through a direct comparison.

    The document does not report specific quantitative performance metrics for the SafeOp 3: Neural Informatix System for its initial clearance, as it relies on substantial equivalence.

    However, the Predetermined Change Control Plan (PCCP) outlines future modifications to the SSEP Baseline Classification Algorithm and defines acceptance criteria for these future algorithmic changes. These are the only explicitly stated performance acceptance criteria in the document.

    Table 1: Acceptance Criteria and Reported Device Performance (from PCCP for future algorithmic changes)

    Acceptance Criteria (for future SSEP algorithm modifications)Reported Device Performance (for future SSEP algorithm modifications)Notes
    Modification #1 & #4 (SSEP Baseline Classification Algorithm Retraining/Maintenance):No reported performance data for the current cleared device.The document states these are criteria for future modifications of the SSEP algorithm, not for the initial clearance of the SafeOp 3 System.
    Overall Positive Predictive Value (PPV) ≥ 90%To be demonstrated with verification testing of modified algorithm.These criteria are goals for future changes.
    Overall Accuracy ≥ 85%To be demonstrated with verification testing of modified algorithm.
    Overall F1-score ≥ 85%To be demonstrated with verification testing of modified algorithm.
    Performance of each stimulation site-specific classifier to be demonstrated as non-inferior to the current SafeOp 3 System at the 95% confidence level.To be demonstrated with verification testing of modified algorithm.
    Modification #2 (Removal of Recalculating Active Trace (RAT) State):No reported performance data for the current cleared device.These are criteria for a future modification involving signal processing.
    Average noise content for the reconstructed Ensemble Average (EA) following restart to be demonstrated as non-inferior to that of the preceding EA(s) with 95% confidence.To be demonstrated with internal testing and retrospective analysis of intraoperative data.These criteria are goals for future changes.

    Study Details (Based on the document, for the initial clearance)

    Because the 510(k) is based on substantial equivalence and explicitly states "Determination of substantial equivalence is not based on an assessment of clinical performance data," there is no detailed "study" proving the device meets acceptance criteria in the way a de novo or PMA submission might. The "study" mentioned below refers to the verification and validation activities conducted to demonstrate substantial equivalence and safety/effectiveness for the initial clearance, as described in the "Performance Data" section.

    1. Sample size used for the test set and the data provenance:

      • Not explicitly stated for clinical performance data for the initial clearance. The document mentions "nonclinical performance testing" for functional, system, and software requirements, and "usability testing." These typically involve controlled test cases and user interactions rather than a large clinical test set.
      • For the future PCCP modifications of the SSEP algorithm, it mentions "an independent, firewalled dataset" and "additional labeled data" for re-training and testing, but no sample size or provenance is specified within this document.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not explicitly stated for the initial clearance. Since the clearance is based on substantial equivalence and does not assess clinical performance data, detailed expert ground truth establishment for a clinical test set is not described.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not explicitly stated for the initial clearance. Not applicable given the reliance on substantial equivalence and lack of clinical performance data assessment.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. The document does not describe an MRMC study. The device is a neural informatic system, and while it involves signal processing, it's not positioned as an AI-assisted diagnostic tool that human readers would interpret with assistance in the traditional sense of an MRMC study. The SSEP algorithm mentioned in the PCCP appears to be a standalone classification, not a human-in-the-loop assistance tool subject to such a study.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • For the initial clearance, the document refers to "nonclinical performance testing" demonstrating the system meets functional, system, and software requirements. This implies standalone testing of device functions.
      • For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), the acceptance criteria (PPV, Accuracy, F1-score) clearly indicate standalone algorithm performance is intended to be evaluated. However, no such standalone performance data is provided for the initially cleared device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not explicitly stated for the initial clearance. For the "nonclinical performance testing" and "usability testing," ground truth would typically be established based on engineering specifications and validated operational procedures.
      • For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it refers to "labeled data," which implies a ground truth established by experts or other definitive methods, but the specific type is not detailed in this document.
    7. The sample size for the training set:

      • Not explicitly stated for the initial clearance. No AI algorithms are reported as being evaluated for the initial clearance.
      • For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it mentions the algorithm "will be re-trained using site-specific data" and "periodically re-trained as additional labeled data become available," but no sample size is specified within this document.
    8. How the ground truth for the training set was established:

      • Not explicitly stated for the initial clearance.
      • For the future SSEP Baseline Classification Algorithm (PCCP Modification #1 & #4), it refers to "labeled data," but the method of labeling (establishing ground truth) is not detailed in this document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241045

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-11

    (238 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K241045

    Trade/Device Name: Disposable Subdermal Needle Electrode, Corkscrew Regulation Number: 21 CFR 882.1350
    Common or Usual Name

    Disposable Subdermal Corkscrew Needle Electrode

    Classification

    21 C.F.R. § 882.1350
    K130136 Disposable Monopolar Needle Electrode and Disposable Subdermal Needle Electrode, 21 C.F.R. § 882.1350
    number | 21 CFR 882.1350

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Needle Electrodes for neurological purposes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalograph (EEG), electromyograph (EMG) and nerve potential signals, and are intended for stimulation/recording with stimulation/recording equipment for electromyograph (EMG) and nerve potential signals.

    Device Description

    Disposable Subdermal Needle Electrodes, Corkscrew are single-patient use electrodes, delivered Ethylene Oxide (EtO) sterilized. They include a needle at one end which is connected to a lead wire with a 1.5 mm female safety DIN 42802 connector. The needle, which is fabricated using stainless steel, is formed into a corkscrew shape, with a regular bevel tip. The connecting lead wire, which may be supplied in different lengths and colors is terminated with a DIN 42802 connector to enable connection to recording or monitoring / stimulation equipment.

    AI/ML Overview

    The document provided describes a 510(k) premarket notification for a medical device, specifically a "Disposable Subdermal Needle Electrode, Corkscrew." The purpose of this notification is to demonstrate substantial equivalence to an existing legally marketed predicate device. This type of submission relies on non-clinical testing to show that changes do not raise new questions of safety or effectiveness, rather than a clinical effectiveness study comparing the device to a gold standard.

    Therefore, the requested information regarding acceptance criteria and studies proving the device meets those criteria, particularly in the context of diagnostic accuracy, AI, and human reader performance, is not applicable to this submission. This document focuses on demonstrating that a modified version of an existing device is just as safe and effective as its predicate through various engineering and biocompatibility tests.

    Here's a breakdown of why many of your questions cannot be answered from this document:

    • Acceptance Criteria & Device Performance (Table): While the document lists "tensile strength" and "impedance" as performance metrics with target values (which can be considered acceptance criteria in an engineering sense), these are not diagnostic performance metrics (e.g., sensitivity, specificity, AUC) that would be relevant to an AI/diagnostic device study. The reported performance is that the device "passed the tests and meets the set requirements."

      Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Impedance<15 kΩPassed the tests / Meets requirements
    Tensile strength needle to cable>22N (Predicate) / >25N (New Device)Passed the tests / Meets requirements
    Tensile strength needle to hub>25N (New Device)Passed the tests / Meets requirements
    Maximum retraction force protection cap vs hubMax. 10N (New Device)Passed the tests / Meets requirements
    Torsion force Needle to hub> 15.5 Ncm (New Device)Passed the tests / Meets requirements
    BiocompatibilityIn accordance with ISO 10993 seriesEvaluated to be biocompatible
    Sterility Assurance Level10^-6Passed the tests / Meets requirements
    • Sample Size for Test Set & Data Provenance: Not applicable as this is not a diagnostic device study. The "test sets" here refer to batches of devices used for various engineering and biocompatibility tests.
    • Number of Experts & Qualifications for Ground Truth: Not applicable. Ground truth in this context would be engineering specifications and validated test methods, not expert consensus on diagnostic interpretations.
    • Adjudication Method: Not applicable.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. This is not an AI-assisted diagnostic device.
    • Standalone Performance: Not applicable. This is not an AI algorithm. Its "performance" relates to its physical and electrical properties, not diagnostic accuracy.
    • Type of Ground Truth Used: The ground truth for the non-clinical tests is established by recognized international standards (e.g., ISO, ASTM, IEC) and internal test methods with defined specifications. For sterility, it's a Sterility Assurance Level of 10^-6. For biocompatibility, it's compliance with the ISO 10993 series.
    • Sample Size for Training Set: Not applicable. This device does not involve a "training set" in the machine learning sense.
    • How Ground Truth for Training Set was Established: Not applicable.

    Summary of Non-Applicable Information and Reasons:

    Information RequestApplicabilityReason
    1. A table of acceptance criteria and the reported device performance (diagnostic metrics)Not ApplicableThe provided document is for a 510(k) submission of a physical medical device (needle electrode), focusing on demonstrating substantial equivalence through non-clinical engineering and biocompatibility testing. It addresses whether mechanical, electrical, and material properties meet specifications, not diagnostic accuracy or effectiveness in comparison to a gold standard. The table above reflects the engineering-focused criteria.
    2. Sample size used for the test set and the data provenanceNot ApplicableThe "test set" in this context refers to samples of the device used for various engineering and biocompatibility tests, not a dataset for evaluating diagnostic performance. The document doesn't specify sample sizes for these internal tests, nor is 'data provenance' (e.g., country of origin, retrospective/prospective) relevant to device manufacturing and testing.
    3. Number of experts used to establish the ground truth for the test set and their qualificationsNot ApplicableGround truth for this device is based on established engineering specifications, international standards (ISO, ASTM, IEC), and internal validated test methods, not expert clinical interpretation of images or patient data.
    4. Adjudication methodNot ApplicableAdjudication methods (e.g., 2+1) are used for resolving discrepancies in expert interpretation of diagnostic data, which is not relevant to the non-clinical testing of a needle electrode.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, and effect sizeNot ApplicableMRMC studies are used to assess the comparative diagnostic performance of readers, usually with and without AI assistance. This device is a physical electrode and does not involve AI or human interpretation of diagnostic results in that manner.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was doneNot ApplicableStandalone performance refers to the diagnostic accuracy of an AI algorithm on its own. This device is not an algorithm.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)Manufacturing/Engineering Specifications, Recognized StandardsThe ground truth for this device's evaluation is based on meeting predefined specifications for its physical, electrical, material, and sterility properties as determined by engineering standards, internationally recognized testing methodologies (e.g., ISO, ASTM, IEC), and internal validation protocols.
    8. The sample size for the training setNot ApplicableThis device does not utilize machine learning or AI algorithms; therefore, there is no "training set" in the computational sense.
    9. How the ground truth for the training set was establishedNot ApplicableAs there is no training set for an AI algorithm, this question is not applicable.

    The studies performed were non-clinical tests to ensure the device's physical, electrical, and biological safety and performance:

    • Performance Testing: Visual, dimensional, electrical (impedance), and mechanical (torsion, pull force) aspects were tested using internal methods.
    • Shelf Life (Accelerated Aging): Tested according to ASTM F1980-21, including mechanical, dimensional, electrical, visual, label assessment, and pouch integrity.
    • Transit Testing: Electrical and mechanical assessment, label assessment, and pouch integrity after transit simulation (ISTA 3A).
    • Electrical, Mechanical and Thermal Safety: Compliance with IEC 60601-1 (Ed. 3.2), specifically section 8.5.2.3.
    • Biocompatibility: Extensive testing (Chemical characterization, cytotoxicity, intracutaneous reactivity, acute systemic toxicity, sensitization, pyrogenicity) performed by an external laboratory in accordance with the ISO 10993 series and related standards.
    • Sterilization: Validation leveraged from the predicate device (ISO 11135:2014) as the changes did not affect sterilization.
    • Pyrogenicity: LAL test performed after sterilization.
    • Usability: Applied the "User Interface of Unknown Provenance" (UOUP) principle according to IEC 62366-1, considering the user interface remained unchanged from the predicate.

    In conclusion, the document successfully argues for substantial equivalence based on rigorous engineering, material, and safety testing, but it does not contain the type of diagnostic performance data or AI study details you requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K234092

    Validate with FDA (Live)

    Date Cleared
    2024-04-19

    (115 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | §882.1870, §882.1320, §882.1350
    | §882.1870, §882.1320, §882.1350

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SafeOp 3: Neural Informatix System is intended for use in monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.

    SafeOp 3 Accessories: The SafeOp Accessories are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.

    Device Description

    The SafeOp™ 3: Neural Informatix System (SafeOp 3 System), consists of the SafeOp patient interface with power supply and IV pole mount, the Alpha Informatix Tablet with docking station and power supply and a data transfer USB cable. Associated disposable accessories consists of an electrode harness, surface and/or subdermal needle electrodes, MEP Activator, Cranial Hub, PMAP Dilators and stimulating probe or clip contained in various kits.

    The subject device is intended for use by trained healthcare professionals, clinical neurophysiologists/technologists and appropriately trained non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by both clinical and trained non-clinical personnel.

    The subject device records the following modalities:

    • Somatosensory evoked potentials (SSEP)
    • Motor evoked potentials (MEP),
    • . Train-of-four neuromuscular junction (TO4),
    • Triggered electromyography (tEMG) and ●
    • . Free run electromyography (sEMG)
    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria for the device's performance, nor does it describe a study that rigorously proves the device meets such criteria through a clinical validation or similar performance evaluation.

    The document is a 510(k) premarket notification summary for the "SafeOp 3: Neural Informatix System." Its primary purpose is to demonstrate substantial equivalence to a previously cleared predicate device (SafeOp2: Neural Informatix System, K213849, and reference device Cascade IOMAX Intraoperative Monitor, K162199), rather than to present a full clinical performance study with defined acceptance criteria and detailed results.

    Here's a breakdown of what the document does say, and what it lacks in relation to your request:

    What the document provides:

    • Device Name: SafeOp 3: Neural Informatix System
    • Intended Use/Indications for Use: Monitoring neurological status by recording transcranial motor evoked potentials (MEP), somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ) during various surgical procedures.
    • Technological Comparison: A table comparing the SafeOp 3 System to predicate and reference devices, focusing on technical specifications like monitoring modalities, amplifier channels, stimulation parameters (voltage, current, pulse duration, repetition rate), and filter ranges. This comparison primarily aims to establish that the differences in technology do not raise new questions of safety or effectiveness.
    • Performance Data (Non-clinical): Mentions that "Nonclinical performance testing demonstrates that the subject SafeOp 3 System meets the functional, system, and software requirements." It also states "EMC and Electrical Safety Testing... was performed to ensure all functions... are electrically safe, and comply with recognized electrical safety standards." Usability testing was also performed.
    • Clinical Information Disclaimer: Explicitly states, "Determination of substantial equivalence is not based on an assessment of clinical performance data."

    What the document lacks significantly for your request:

    • A table of acceptance criteria and reported device performance: This is the most significant omission for your request. The document details technical specifications and comparisons but does not provide quantitative performance metrics (e.g., accuracy, sensitivity, specificity, or specific error rates) against pre-defined acceptance thresholds for any of its functionalities (MEP, SSEP, EMG, NMJ). The performance data mentioned are non-clinical (functional, system, software, EMC, electrical safety, usability), not clinical performance metrics.
    • Sample size used for the test set and data provenance: Since specific clinical performance studies are not detailed, this information is not provided.
    • Number of experts used to establish ground truth and qualifications: Not applicable as a clinical ground truth establishment process for performance evaluation is not described.
    • Adjudication method for the test set: Not applicable.
    • MRMC comparative effectiveness study: No such study is mentioned or detailed.
    • Standalone (algorithm only) performance: While the device is an "algorithm only" in a sense (it processes physiological signals), its performance isn't quantified in a standalone clinical evaluation or comparative study.
    • Type of ground truth used: No clinical ground truth is described for performance evaluation.
    • Sample size for the training set: Not applicable, as this is related to AI/ML development and training, which is not described. The device is a neuromonitoring system, not explicitly stated to be an AI/ML device in the context of this submission.
    • How the ground truth for the training set was established: Not applicable.

    Why this information is missing:

    The FDA 510(k) pathway for "substantial equivalence" often relies on demonstrating that a new device is as safe and effective as a legally marketed predicate, without necessarily requiring new clinical trials or detailed performance studies if the technological differences are minor and well-understood. The focus is on showing that any differences do not introduce new safety or effectiveness concerns.

    In summary, based solely on the provided text, I cannot complete the table of acceptance criteria or describe a study that proves the device meets these criteria in a clinical performance context. The document focuses on demonstrating substantial equivalence through technical comparison and non-clinical testing, rather than presenting clinical performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232581

    Validate with FDA (Live)

    Date Cleared
    2023-12-04

    (101 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    : K232581

    Trade/Device Name: Medical Disposable Sterile Needle Electrode Regulation Number: 21 CFR 882.1350
    RegulationNumber
    21 CFR 882.1350
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended for use with recording, monitoring equipment for the recording of bioelectrical signals, such as electromyograph (EMG). The electrodes are sterile and for single patient use only.

    Device Description

    Medical disposable sterile needle electrodes are sterile, single use products that penetrate the skin and directly contact the extracellular fluid during use, forming a good electrolyte interface. The pointed part of the needle electrode makes contact with the nerves, muscles, or other tissues under examination to collect and transmit the corresponding bioelectrical signals (myoelectric currents) to the electromyography (EMG) recorder. Needle electrodes are used in conjunction with EMG diagnostic instruments and serve as accessories to the EMG machine for detecting myoelectric potentials. Medical disposable sterile needle electrodes are invasive since they are positioned subcutaneously and are used under the supervision of a licensed physician. The subject device, available in configurations with and wires, which encompasses subdermal, monopolar, spiral, and concentric designs. The subject device's configurations of monopolar and concentric needles correspond to those without lead wires, while the configurations involving spiral and subdermal designs include lead wires.

    AI/ML Overview

    This appears to be an FDA 510(k) clearance letter for a Medical Disposable Sterile Needle Electrode. The document describes the device, its intended use, and non-clinical testing performed to establish substantial equivalence to a predicate device.

    Crucially, this document is for a medical device (a physical needle electrode), not a software/AI-driven device that would involve a test set, ground truth experts, MRMC studies, or training sets in the way your request describes. Therefore, many of the requested criteria are not applicable.

    However, I can extract the information that is present and explain why other sections are not relevant.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for this type of device are primarily based on non-clinical performance (physical and biological safety) and demonstrating substantial equivalence to a predicate device. There isn't a "performance" metric in the sense of accuracy, sensitivity, or specificity as one would find for an AI algorithm.

    Test / AttributeAcceptance Criteria (Implied by standard and "Pass" result)Reported Device Performance
    Non-Clinical TestingMeets specified standards for functional performance.
    Surface qualityAcceptable surface finish; no defects impacting safety or performance.Pass
    Sharpness indexMeets specified sharpness requirements for penetration.Pass
    ToughnessWithstands expected stresses without breakage or loss of integrity.Pass
    Connection strengthConnections (e.g., needle to wire) maintain integrity during use.Pass
    Electrical conductivityConducts electrical signals effectively as intended.Pass
    Conductive terminals and needle body insulationInsulation prevents unintended electrical paths.Pass
    Sterility testingAchieves and maintains sterility (SAL of $10^{-6}$).Pass (and SAL $10^{-6}$ demonstrated)
    Biocompatibility TestingMeets biological safety requirements per ISO 10993-1 for subdermal communicating devices with muscle.
    In vitro cytotoxicityNon-cytotoxic (as per ISO 10993-5).Non-cytotoxic
    Skin sensitizationNon-sensitive (as per ISO 10993-10).Non-sensitive
    Intracutaneous reactivityNon-irritation (as per ISO 10993-23).Non-irritation
    Acute systemic toxicityNon-acute systemic toxicity (as per ISO 10993-11).Non-acute systemic toxicity
    PyrogenNon-pyrogenic (as per ISO 10993-11).Non-pyrogenic
    SterilizationValidated according to ISO 11135:2014, achieving SAL of $10^{-6}$ and meeting EO residuals per ISO 10993-7.Yes (meets ISO 11135, SAL $10^{-6}$, and ISO 10993-7 residuals)
    Shelf-LifeMaintains performance and safety for stated duration.3 years (based on accelerated and real-time aging)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable to this type of device. The "tests" mentioned are physical and chemical laboratory tests on manufactured samples of the device, not clinical data sets of patients. There is no concept of "data provenance" in the context of clinical images or patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. There is no "ground truth" established by experts in the context of an electromyography needle electrode. The performance is assessed through standardized laboratory tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. There is no "test set" requiring adjudication in the context of an electromyography needle electrode.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This is a physical medical device, not an AI software. MRMC studies are used for evaluating diagnostic imaging AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    As mentioned, "ground truth" in the AI/software sense is not applicable. For this device, the "truth" is established by adherence to recognized international standards for biocompatibility (ISO 10993 series), sterilization (ISO 11135), and general physical performance testing (e.g., surface quality, sharpness, electrical conductivity).

    8. The sample size for the training set

    This is not applicable. There is no "training set" for a physical medical device.

    9. How the ground truth for the training set was established

    This is not applicable. There is no "training set" or associated "ground truth" for a physical medical device. The "truth" is established through engineering design, material science, and adherence to established manufacturing and testing standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231691

    Validate with FDA (Live)

    Date Cleared
    2023-06-29

    (20 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Regulation | 21CFR 882.1350

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Phantom XL Insulated Dilators are indicated for use during surgery of the spine to deliver an electrical stimulus to the tissues and nerves at the operative site, to assist in locating those nerves at risk during the surgical procedure.

    Device Description

    TSI's Phantom XL Insulated Dilators are used as instruments to deliver electrical stimulation to tissue during intraoperative neurological monitoring. Phantom XL Insulated Dilators are available in monopolar configuration and four diameter sizes. They are supplied sterile, are non-pyrogenic, and are intended for single use only.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called "Phantom XL Insulated Dilators". This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with specific performance metrics like sensitivity, specificity, or AUC.

    Therefore, the information you're requesting regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of typical AI/diagnostic device performance studies (which would involve metrics like sensitivity, specificity, human reader performance, etc.) is not applicable to this submission.

    This submission focuses on engineering-level performance testing to demonstrate that changes made to the existing predicate device (Phantom XL Insulated Dilators, K140088) do not raise new questions of safety or effectiveness. The changes are minor:

    • Change in packaging: from backer card/Tyvek pouch/Shelf Box to preformed tray sealed in Tyvek Lid placed in Shelf Box.
    • Extended shelf life: from 2 years to 5 years.

    The acceptance criteria and studies described relate to these specific physical changes and their impact on the device's ability to maintain sterility, packaging integrity, and functionality over its extended shelf life.

    Here's a breakdown of the information provided, reframed to fit the context of this 510(k) submission:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a "table of acceptance criteria and reported device performance" in the way one might expect for a diagnostic or AI device (e.g., sensitivity > X%, specificity > Y%). Instead, it lists the types of changes made to the device and the standards/tests performed to ensure these changes do not negatively impact safety and effectiveness. The "acceptance criteria" are implied by compliance with the referenced ASTM, ISO, and AAMI standards. The "reported device performance" is that the device "demonstrates that the TSI Phantom XL Dilators are safe for use perform comparably to the predicate device that is currently marketed for the same intended use" after these tests.

    Type of ChangeDescription of Tests Performed (Implied Acceptance Criteria: Compliance with these standards)Reported Device Performance (Summary)
    Packaging- ASTM F1886:2016 (Visual Inspection for Seal Integrity)The new packaging design maintains integrity and sterility, demonstrating satisfactory verification and validation in compliance with Design Control Process.
    - ASTM F2096-11 (Gross Leaks by Internal Pressurization)
    - ASTM F88/F88M-21, 2021 (Seal Strength of Flexible Barrier Materials)
    - Dupont Medical Packaging (Compliance of Tyvek to ISO 11607)
    - ASTM D4169:2022 (Performance Testing of Shipping Containers and Systems)
    - ISTA 3A: 2018 (Packaged-Products for Parcel Delivery System Shipments)
    - ISO 11607-2:2019 (Validation requirements for sealing and assembly processes)
    - ISO 16269-6:2014 (Statistical tolerance intervals)
    Expiration Date- ASTM F1980; 2021 (Accelerated Aging of Sterile Medical Device Packages)Extended shelf life from 2 to 5 years is supported by accelerated aging studies, showing safety and effectiveness are maintained over the longer period.
    Sterilization- ISO 11607-1:2019 (Requirements for materials, sterile barrier systems, packaging)Sterility (maintaining SAL) with the new packaging and extended shelf life is confirmed through bench testing and compliance with relevant standards.
    - AAMI/ISO TIR16775: 2014 (Guidance on application of ISO 11607-1 & 11607-2)
    - AAMI TIR28:2016 (Product adoption and process equivalency for EtO sterilization)
    - ANSI/AAMI ST67:2019 (Selecting a sterility assurance level (SAL))
    - ANSI/AAMI/ISO 11135:2014 (Ethylene oxide sterilization process validation)
    - ANSI/AAMI ST72; 2019 (Bacterial Endotoxins-Test methods)
    General- Bench testing for sterility (as per standards above)No new issues of safety or effectiveness were identified. All risk mitigations were satisfactorily verified and validated.
    - Bench testing for new expiration dating (as per standards above)
    - Bench testing for packaging integrity (as per standards above)
    - Compliance with company's Design Control process (21 CFR Part 820.30)
    - Risk analysis per ISO 14971 Standard

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each test listed (e.g., how many packages were tested for seal strength). The studies mentioned are engineering/bench tests on the device and its packaging components, not human subject studies. Therefore, "data provenance" in terms of country of origin or retrospective/prospective is not relevant here.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable. The "ground truth" for this type of submission is defined by the technical specifications of the device and its packaging, and ensuring compliance with recognized industry standards (ASTM, ISO, AAMI). There is no "ground truth" to be established by medical experts (e.g., radiologists interpreting images) as this is not a diagnostic device with performance metrics relying on human interpretation. The testing is performed by engineers and technicians following standard protocols.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies when there's ambiguity in establishing a ground truth (e.g., for disease diagnosis from images) or for resolving disagreements among multiple human readers. This submission describes physical and performance tests against standards, not diagnostic performance.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a surgical instrument (an insulated dilator used for nerve stimulation/localization during spine surgery), not an AI-powered diagnostic tool. Therefore, MRMC studies and AI assistance comparisons are irrelevant.


    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    This is not applicable. As stated above, this is a surgical instrument, not an algorithm or AI device.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is the compliance with established engineering and medical device standards (ASTM, ISO, AAMI). For example, for packaging, the "ground truth" is that the seal strength meets the requirements specified by ASTM F88, or that the sterility assurance level (SAL) meets the requirements of ANSI/AAMI ST67 after ethylene oxide sterilization. It's about meeting predefined technical specifications and standards, not a medical condition.


    8. The sample size for the training set

    This is not applicable. This is a physical medical device, not an AI or machine learning model. There is no "training set."


    9. How the ground truth for the training set was established

    This is not applicable. As there is no training set for an AI or ML model, this question does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213849

    Validate with FDA (Live)

    Date Cleared
    2022-03-09

    (89 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Regulation Number(21 CFR) | §882.1870, §882.1320, §882.1350
    | §882.1870, §882.1320, §882.1350

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SafeOp 2: Neural Informatix System is intended for use in monitoring neurological status by recording somatosensory evoked potentials (SSEP), electromyography (EMG), or assessing the neuromuscular junction (NMJ). Neuromonitoring procedures include intracranial, intratemporal, extratemporal, neck dissections, upper and lower extremities, spinal degenerative treatments, pedicle screw fixation, intervertebral fusion cages, rhizotomy, orthopedic surgery, open/percutaneous, lumbar, thoracic, and cervical surgical procedures.

    Device Description

    The SafeOp™ 2: Neural Informatix System (SafeOp 2 System), formerly known as EPAD 2 (K182542), consists of the SafeOp head unit with power supply and IV pole mount, the Alpha Informatix Tablet with docking station and power supply and a data transfer USB cable. Associated disposable accessories consists of an electrode harness, surface and/or subdermal needle electrodes, and stimulating probe or clip contained in various kits. The SafeOp 2 System head unit contains a complete data acquisition system that has builtin stimulators, amplifiers, relays, A/D Converters, Digital Signal Processors, CPUs, and storage devices. The head unit interfaces with other equipment through communication ports and serves as the patient-contacting portion of the system where it is close to the surgical field. The head unit hardware contains an eight acquisition (input) channel and six-output channel Evoked Potential Stimulator that is used in the operating room to display nerve and muscle responses. The user can use these responses to diagnose insults to the peripheral or central nerves and to determine relative nerve location, proximity, and integrity data. The SafeOp 2 System application provides the primary graphical user interface and controls for the SafeOp 2 System. The application runs on a touchscreen tablet mobile device which connects to the head unit either via wired USB cable or wireless via Wi-Fi, enabling both user input (e.g., patient and procedure information, adjustment of stimulus and acquisition parameters) and display of output (e.g., display of acquired waveforms, data, messages and alerts to the clinician).

    AI/ML Overview

    The provided text is a 510(k) summary for the SafeOp 2: Neural Informatix System. It primarily focuses on demonstrating substantial equivalence to a predicate device (EPAD™ 2 System, K182542) rather than presenting a detailed study proving the device meets specific acceptance criteria based on clinical performance.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative "acceptance criteria" and "reported device performance" in the typical sense of a clinical study measuring diagnostic accuracy or treatment efficacy. Instead, it demonstrates "substantial equivalence" to a predicate device by comparing various specifications. The "acceptance criteria" for this submission would broadly be that the device's specifications and performance are comparable to the predicate, with any differences not introducing new safety or effectiveness concerns.

    Specification/PropertyPredicate Device (EPAD™ 2 System, K182542)Subject Device (SafeOp™ 2 System)Performance/Rationale
    Intended Use/Indications for UseIdenticalIdenticalIdentical
    Device ClassIIIIIdentical
    Product CodeGWF, GXY, GXZ, IKN, PDQ, ETNGWF, GXY, GXZ, IKN, PDQ, ETNIdentical
    Regulation Number§882.1870, §882.1320, §882.1350, §890.1375, §874.1820, §874.1820§882.1870, §882.1320, §882.1350, §890.1375, §874.1820, §874.1820Identical
    Device Classification NameStimulator, Electrical, Evoked ResponseStimulator, Electrical, Evoked ResponseIdentical
    Monitoring ModalitiesEMG, SSEP, NMJEMG, SSEP, NMJIdentical
    Head Unit Power Supply100 to 240 VAC, 50-60 Hz (input); 15 VDC, 2.5A (output)100 to 240 VAC, 50-60 Hz (input); 12 VDC, 2.5A (output)Head unit power supply does not impact device performance when compared to the predicate. The change in power does not introduce new risks, or impact existing risks. Therefore, this difference does not affect device safety or effectiveness.
    Mode of OperationContinuousContinuousIdentical
    Dimensions12"W x 8"H x 2"D12"W x 8"H x 2"DIdentical
    Weight< 2 lbs< 2 lbsIdentical
    Operating ModesTriggered EMG, Free run EMG, SSEP, NMJTriggered EMG, Free run EMG, SSEP, NMJIdentical
    Total Amplifier ChannelsUp to 8Up to 8Identical
    WaveformMonophasic, RectangularMonophasic, RectangularIdentical
    Pulse Duration50 to 300 µsec300 µsec (SSEP), 200 µsec (EMG)Identical (The provided predicate range encompasses the subject device's specific values, which are typical for each modality, implying no significant difference in capability or safety).
    Frequency (Pulse Rate)0.1 to 50 Hz0.1 to 50 HzIdentical
    Current Range0 to 100 mA0 to 100 mAIdentical
    Input Impedance> 50 MΩ (at DC)> 50 MΩ (at DC)Identical
    Low Frequency Filter10 Hz (SSEP), 30 Hz (EMG)10 Hz (SSEP), 30 Hz (EMG)Identical
    High Frequency Filter2.7 kHz (SSEP & EMG)2.7 kHz (SSEP & EMG)Identical
    Notch Filter50 or 60 Hz50 or 60 HzIdentical
    AlphaInformatix (AIX) Tablet OSAndroid powered tabletWindows 10 powered tabletCompleted V&V testing successfully demonstrates that the differences in the Operating Systems (OS) have no impact on device performance when compared to the predicate. The OS change does not introduce new risks, or impact existing risks. Therefore, this difference does not affect device safety or effectiveness.
    Remote AccessNoNoIdentical
    Surface Electrode Anatomical SitesSSEP: Upper/lower limbs and head/neckSSEP: Upper/lower limbs and head/neckIdentical
    Surface Electrode TypeCustomer cutaneous electrodes for use with SafeOp only. Single, double and triple electrodes.Customer cutaneous electrodes for use with SafeOp only. Single, double and triple electrodes.Identical
    Conductive Surface Area20x25mm25.4x25.4mm (1x1in.)Completed V&V testing successfully demonstrates that the differences in the surface area has no impact on device performance when compared to the predicate. The change does not introduce new risks, or impact existing risks. Therefore, this difference does not affect device safety or effectiveness.
    Conductive GelWet gelSolid gelCompleted V&V testing successfully demonstrates that the differences in the conductive gel has no impact on device performance when compared to the predicate. The change does not introduce new risks, or impact existing risks. Therefore, this difference does not affect device safety or effectiveness.
    ConnectorsNicomatic three pinNicomatic three pinIdentical
    Current Density< 0.75 mArms/cm2< 0.75 mArms/cm2Identical
    SterilityNon-sterile, single patient use, disposableNon-sterile, single patient use, disposableIdentical
    Surface Contact Time≤ 24 hours≤ 24 hoursIdentical
    Other AccessoriesNeedle Electrodes, Stimulating Probes, Cable, ElectrodeNeedle Electrodes, Stimulating Probes, Cable, ElectrodeIdentical
    Biocompatibility (ISO 10993-1)Yes (Tissue/bone/dentin for ≤ 24 hours)Yes (Tissue/bone/dentin for ≤ 24 hours)Identical

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states: "Determination of substantial equivalence is not based on an assessment of clinical performance data." This indicates that there was no human test set or clinical data involved in proving substantial equivalence for this submission. The evaluation was based on nonclinical performance testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no human test set or clinical data was used to establish substantial equivalence.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no human test set or clinical data was used to establish substantial equivalence.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is for intraoperative neuromonitoring, not an AI-assisted diagnostic tool that would typically involve a multi-reader multi-case study for diagnostic improvement. The document explicitly states: "Determination of substantial equivalence is not based on an assessment of clinical performance data."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document states that "Nonclinical performance testing demonstrates that the subject SafeOp 2 System meets the functional, system, and software requirements." This indicates that standalone performance testing was conducted for the device's functions (e.g., EMC, electrical safety, usability), but this refers to the system as a whole, not a specific algorithm's performance in isolation from a human user in a diagnostic or interpretive capacity. It does not provide details on the specific performance metrics (e.g., accuracy, sensitivity, specificity) for standalone algorithm performance in interpreting neurological signals.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the nonclinical performance testing (EMC, Electrical Safety, Usability), the ground truth would be based on engineering standards and predetermined functional specifications. For example, for electrical safety, the ground truth is compliance with recognized electrical safety standards (e.g., 60601-1). For usability, the ground truth is that the device "presents no adverse effect within the intended environment."

    8. The sample size for the training set

    Not applicable. The submission is for a neuromonitoring system and does not describe an AI/machine learning model that would require a "training set" in the conventional sense. The "training" here would refer to the development and testing of the engineering system.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a training set for an AI/machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200984

    Validate with FDA (Live)

    Date Cleared
    2020-05-27

    (43 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Carolina 29209

    Re: K200984

    Trade/Device Name: Guardian Needle Electrode Regulation Number: 21 CFR 882.1350
    | Guardian Needle™ ElectrodeNeedle ElectrodeGXZ21 CFR 882.1350
    | Subdermal Needle ElectrodesNeedle ElectrodeK022914GXZ21 CFR 882.1350
    |
    | Regulation # | 21 CFR 882.1350
    | 21 CFR 882.1350

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rhythmlink International Subdermal Needle Electrodes are intended for use with recording, monitoring and stimulation equipment for the purpose of recording of biopotential signals. Examples include: Electromyography (EMG), Electroencephalography (EEG), and Nerve potential signals. The electrodes are sterile and for single patient use.

    Device Description

    Guardian Needle™ Electrodes are a sterile, single-use device. The electrodes are applied during the study of biopotentials such as electromyography (EMG), electroencephalography (EEG), nerve conduction and stimulation/response. The electrodes are invasive as they are placed subcutaneously or in contact with nerve or muscle tissue. The needle is housed inside a sheath until application and secured to the patient during use with adhesive tape.

    AI/ML Overview

    The Rhythmlink International Guardian Needle™ Electrode (subject device) is compared to the predicate device, Subdermal Needle Electrodes (K022914), to demonstrate substantial equivalence. The document primarily focuses on non-clinical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaSubject Device (Guardian Needle™ Electrode) Performance
    Sterilization & Residuals- Compliance with TIR 28:2009 guidance.Subject device determined to fall under previously validated EtO sterilization cycle.
    - EtO and Ethylene Chlorohydrin (ECH) levels after 24-hour aeration below maximum limits described in ANSI/AAMI/ISO 10993-7:2008/(R) 2012.Reevaluation confirmed low levels of EtO and ECH.
    Biocompatibility- Biological safety for intact skin contact (≤24 hours) as per ISO 10993 standards and risk management framework.Biocompatibility evaluation establishes biological safety for intact skin contact for ≤24 hours.
    Functional Performance- Electrical ContinuityPassed predetermined acceptance criteria.
    - Adhesion Testing (for adhesive tape)Passed predetermined acceptance criteria.
    Technological Equivalence to Predicate- Intended UseIdentical to predicate.
    - Anatomical Site(s)Identical to predicate.
    - Environment UsageIdentical to predicate.
    - Electrode MaterialIdentical to predicate.
    - Electrode LengthIdentical to predicate.
    - Electrode DiameterIdentical to predicate.
    - Leadwire MaterialIdentical to predicate.
    - Leadwire LengthIdentical to predicate.
    - ConnectorIdentical to predicate.
    - Operation of Applying the DeviceIdentical to predicate.
    - Prescription UseIdentical to predicate.
    - Targeted ProceduresIdentical to predicate.
    - Compatibility with other devicesIdentical to predicate.
    - PackagingIdentical to predicate.
    - Sterilization MethodIdentical to predicate.
    - Single Patient UseIdentical to predicate.
    - Electrical SafetyIdentical to predicate (Connectors comply with IEC 60601-1 (1988) sub clause 56.3(c) per CFR 898.12).
    - Mechanical SafetyIdentical to predicate (Leadwires soldered to electrode using Tin/Silver solder and covered with heat shrink).
    - Duration of useIdentical to predicate (≤24 hours).
    - StandardsIdentical to predicate (IEC 60601-1-1: 1988/a1: 1991/A2 :1995 § 56.3(c) per CFR 898.12, DIN 42802, ISO 14971, ISO 15223-1).

    Note on Differences: The subject device includes a needle sheath (thermoplastic polyester elastomer sheath) and adhesive tape, which differ from the predicate device (PVC tube for sheath, no adhesive tape). These differences were assessed through testing and determined not to adversely impact safety or effectiveness.

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for the functional performance benchtop tests (Electrical Continuity and Adhesion Testing). However, it states that "The test methods were identical to those used to assess the predicate device."

    For sterilization, reevaluation of EtO residuals was performed, but the sample size is not explicitly stated.

    The data provenance is from non-clinical benchtop testing conducted by Rhythmlink International, LLC. The document does not indicate any country of origin for the data that would suggest external sources. All testing appears to be retrospective relative to the submission date, as it's part of a 510(k) premarket notification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The document describes non-clinical benchtop tests and compliance with recognized standards. There isn't an explicit "test set" in the context of expert review for establishing ground truth, as the criteria are based on engineering and biocompatibility standards.

    4. Adjudication method for the test set

    This information is not applicable as the document does not describe a test set requiring expert adjudication for ground truth (e.g., image interpretation). The evaluation relies on established engineering and biocompatibility testing methodologies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission explicitly states "No Clinical Tests were conducted as referenced in 21 CFR 807.92(b)(2)" and "No Clinical Tests were conducted as referenced in 21 CFR 807.92(b)(3)."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a physical invasive medical electrode, not a software algorithm or AI system.

    7. The type of ground truth used

    The "ground truth" for demonstrating the device meets acceptance criteria is primarily based on:

    • Compliance with recognized industry standards: ANSI/AAMI/ISO 10993-7:2008/(R) 2012 for EtO residuals, ISO 10993 for biocompatibility, IEC 60601-1-1: 1988/a1: 1991/A2 :1995 § 56.3(c) per CFR 898.12, DIN 42802, ISO 14971, ISO 15223-1, and TIR 28:2009 Guidance for sterilization.
    • Predicate device characteristics: Substantial equivalence is established by demonstrating the subject device has identical technological characteristics, intended use, and performance to a legally marketed predicate device, with differences (needle sheath, adhesive tape) assessed not to adversely impact safety or effectiveness.
    • Benchtop test results: Electrical continuity and adhesion testing against predetermined acceptance criteria.

    8. The sample size for the training set

    This question is not applicable as the device is not an AI/ML product that undergoes training.

    9. How the ground truth for the training set was established

    This question is not applicable as the device is not an AI/ML product that undergoes training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192603

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-22

    (63 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K192603

    Trade/Device Name: Spes Medica Subdermal Needle Electrodes Regulation Number: 21 CFR 882.1350
    |
    | Classification Regulation: | 882.1350
    | The same of the predicate device |
    | Regulation Number | 882.1350
    | 882.1350

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spes Medica Subdermal Needle Electrodes are intended for use with recording, monitoring and stimulation equipment for the purpose of recording of biopotential signals. Examples include: Electromyography (EMG), Electroencephalography (EEG) and Nerve potential signals. The electrodes are sterile and for single patient use only.

    Device Description

    Spes Medica Subdermal Needle Electrodes are monopolar needles intented to use for Electromyography (EMG), Electroencephalograph (EEG) and Nerve potential signals in example and they are intended for use with recording, monitoring and stimulation equipment for the purpose of recording of biopotential signals. The Subdermal Needle Electrodes are used for recording and stimulation (micro stimulation), this stimulation has no therapeutic purpose but has the purpose of activating an electrical response.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance testing of the Spes Medica Subdermal Needle Electrodes, but it largely focuses on biocompatibility, sterilization/packaging, and electrical safety standards rather than a comparative effectiveness study involving human readers or AI.

    Therefore, many of the requested fields related to AI performance, human reader studies, and large-scale clinical data will not be present in this document.

    Here's a breakdown of the available information based on your request:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test Description/Standard)Reported Device Performance (Results)
    Biocompatibility
    Cytotoxicity (ISO 10993-5)No reactivity
    Skin Irritation (ISO 10993-10)Negligible irritation response
    Hemolysis test (ISO 10993-4)Does not cause hemolysis
    Skin Sensitization (ISO 10993-10)Not sensitizing
    Systemic Toxicity (ISO 10993-11)No toxic symptoms, satisfies requirements
    Pyrogenicity test (ISO 10993-11)Meets requirements for absence of pyrogens
    Sterilization/Packaging
    Packaging Validation Protocol (5 years shelf-life)No substantial alterations after aging process, devices remain sterile (Tested against ASTM D4332, ASTM D4169, ISO 11135-1, 11737-1, 11737-2)
    Packaging validation test (Sterility test)All samples met seal strength requirements per ASTM F88
    Electrical Safety
    Electrical Safety - Isolation (IEC 60601-1: 3rd Ed + CORR 2:2007 + A1:2012 clause 8.5.2.3)Isolated and in compliance with requirements for electrically safety
    Impedance test and dielectric strengthStimulation doesn't affect or reduce the needle and its performance
    Physical Performance
    Cable tensile testingVery high strength force (average 40.1 N) during pull-out test, ensuring manufacturing process safety and strength of use
    Shipment Tests (Handling, Vehicle Stacking, Loose Load Vibration, Vehicle Vibration, Concentrated Impact)No damage to the product observed
    Climatic Tests (Changing temperature and relative humidity)No damage to the product observed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify general "test sets" or data provenance in the way one would for an AI or clinical study. The performance tests are laboratory-based and conducted on the device components or prototypes. The sample sizes for each specific test (e.g., number of units tested for tensile strength, number of samples for biocompatibility) are not explicitly stated, nor is the country of origin or whether they were retrospective/prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This document describes performance testing for a medical device to establish its safety and effectiveness based on international standards, not the establishment of ground truth by clinical experts for diagnostic accuracy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a diagnostic study requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document does not describe an AI device or an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the various performance tests, the "ground truth" is defined by the requirements of the international standards (e.g., ISO, ASTM, IEC) that the device must meet. For biocompatibility tests, it's the biological response against established thresholds; for electrical safety, it's adherence to voltage and isolation limits; for mechanical tests, it's resistance to forces.

    8. The sample size for the training set

    Not applicable. This is not an AI device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190801

    Validate with FDA (Live)

    Date Cleared
    2019-07-10

    (103 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Carolina 29209

    Re: K190801

    Trade/Device Name: PressOn Electrode Headset Regulation Number: 21 CFR 882.1350
    | PressOn™ Electrode HeadsetNeedle ElectrodeGXZ21 CFR 882.1350
    |
    | Regulation # | 21 CFR 882.1350
    | 21 CFR 882.1350

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PressOn™ Electrode Headset is intended for use in the recording of electroencephalogram (EEG), evoked potential (EP), or as a ground and reference in an EEG or EP recording. The device is provided sterile for single patient use only.

    Device Description

    The PressOn™ Electrode Headset is intended to be used in medical environments where quick EEG electrode application is required. The device provides a workflow solution where the availability of technologists or other specialty-trained EEG staff is limited for applying EEG electrodes.

    The headset comprises between 2 to 48 PressOn™ Electrodes, each loaded into individual button applicators, which are positioned in predetermined locations and interconnected by elastic netting, altogether forming the headset. The predetermined electrode positions are arranged by generally referencing the 10-20 Positioning System, but with flexibility to account for various head shapes and sizes and to avoid interference with intracranial pressure monitoring, ventricular drainage, and other separate devices. Like the predicate, the subject device is minimally invasive and does not directly contact neural tissues.

    The PressOn™ Electrode Headset is placed on the patient's head and is oriented and secured using a nasion marker and chinstrap. The distal end of the device contains the array of PressOn™ electrodes in predetermined locations that are placed on the scalp by minimal insertion of the micro needles into the epidermis layer of the skin for use during monitoring procedures. The PressOn™ electrodes in the subject device are inserted in an identical fashion as the predicate device (K130220). Minor material and dimensional changes have been made to the original applicator to allow for multiple applicators to be connected to form the headset.

    At the proximal end of the subject device is a multipin connector where the electrode leadwires terminate. The multipin connector interfaces with color-coded extension cables (identical to those used in another previously cleared Rhythmlink device, reference device K172503) that terminate into single pin connectors, which are then connected to monitoring equipment.

    The leadwires are constructed of ribbon cable arrays with electrode attachment points at various distances based on the predetermined locations. These features are designed with the intent for quick application by non-neurological based medical personnel with signal quality equivalent to predicate device [K130220].

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a structured format that directly maps to "AI performance metrics" commonly seen in such tables (e.g., sensitivity, specificity, AUC). Instead, it discusses functional performance equivalency to a predicate device based on benchtop tests.

    However, we can infer the acceptance criteria and reported performance from the "Summary of Non-Clinical Tests" section:

    Acceptance Criteria (Inferred)Reported Device Performance
    Electrical Continuity (Pass/Fail)Passed (implied as "all benchtop performance testing passed predetermined acceptance criteria")
    Deployment force to eject electrode from applicator and apply to synthetic skin (Within range)Passed (implied as "all benchtop performance testing passed predetermined acceptance criteria")
    Pull force required to remove the electrode from synthetic skin (Within range)Passed (implied as "all benchtop performance testing passed predetermined acceptance criteria")
    Visual assessment of electrode insertion uniformity into synthetic skin (Acceptable)Passed (implied as "all benchtop performance testing passed predetermined acceptance criteria")
    Dimensional assessment of electrode placement on head phantom (Within tolerance)Passed (implied as "all benchtop performance testing passed predetermined acceptance criteria")

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No Clinical Tests were conducted as referenced in 21 CFR 807.92(b)(2)."

    Therefore:

    • Sample size for test set: Not applicable (no clinical test set).
    • Data Provenance: Not applicable (benchtop testing on synthetic materials, not human data).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable, as no clinical tests were conducted and the "ground truth" was established based on engineering specifications and benchtop performance against predetermined criteria.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical tests were conducted involving expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document explicitly states: "No Clinical Tests were conducted." Therefore, no MRMC study was performed, and no effect size for human readers with and without AI assistance is reported. This device is a hardware electrode headset, not an AI algorithm.

    6. Standalone (Algorithm Only) Performance Study

    No. The device is a physical electrode headset, not an AI algorithm. Therefore, a standalone algorithm-only performance study is not relevant and not reported.

    7. Type of Ground Truth Used

    The "ground truth" for evaluating the device's performance was established via:

    • Engineering specifications and predetermined acceptance criteria for mechanical and electrical properties.
    • Comparison to the predicate device (K130220) for functional equivalency, implying the predicate's established performance served as a baseline.
    • Benchtop measurements and visual assessments on synthetic skin and head phantoms.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical device (electrode headset) and not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.