Search Results
Found 526 results
510(k) Data Aggregation
(215 days)
Infusion System; High-Flo SubQ Needle Set; Precision Flow Rate Tubing
Regulation Number: 21 CFR 880.5725
Needle Sets™
Common or Usual Name: Infusion Pump
Classification Name: Pump, Infusion (21CFR 880.5725
K214045)
Common or Usual Name: Infusion Pump
Classification Name: Pump, Infusion (21CFR 880.5725
The FreedomEdge® Syringe Infusion System consists of the following components:
- FreedomEdge® Infusion Pump
- Precision Flow Rate Tubing™
- High-Flo SubQ Needle Sets™
The FreedomEdge® Syringe Infusion System is indicated for use with the BD® 20 ml syringe (US Reference number: 302830), BD 30 ml syringe (US Reference number: 302832) and Hizentra® 20 ml prefilled syringe (NDC 44206-458-96).
For Immunoglobulin Administration:
The FreedomEdge Infusion System is specifically indicated for the subcutaneous infusion of the following human plasma-derived immunoglobulins in the home, hospital, or ambulatory settings when administered according to the FDA approved biologic labeling:
- Cuvitru®, Immune Globulin Infusion (Human) 20% (manufactured by Takeda®); Gammagard Liquid®, Immune Globulin Infusion (Human) 10% (manufactured by Takeda®);
- Hizentra®, Immune Globulin Subcutaneous (Human) 20% Liquid (manufactured by CSL Behring®)
- Hizentra®, Immune Globulin Subcutaneous (Human) 20% Liquid 20ml Single-use pre-filled syringe (manufactured by CSL Behring®)
For EMPAVELI™ (pegcetacoplan) Administration:
The FreedomEdge Infusion System is specifically indicated for the subcutaneous infusion of EMPAVELI™ (pegcetacoplan) in the home, hospital, or ambulatory setting when administered according to the approved drug product labeling.
For RYSTIGGO® (rozanolixizumab-noli) Administration:
The FreedomEdge Infusion System is specifically indicated for the subcutaneous infusion of RYSTIGGO® (rozanolixizumab-noli) when administered according to the approved drug product labeling.
For Intravenous Antibiotic Administration:
The FreedomEdge Infusion Pump and Precision Flow Rate Tubing are specifically indicated for the intravenous infusion of the following antibiotics when used according to the FDA approved drug product labeling:
- Ertapenem, Meropenem, Oxacillin, and Tobramycin
The FreedomEdge Infusion System is a non-electrically powered, mechanical infusion system which is intended for subcutaneous or intravenous administration of drug products in a home, ambulatory, or clinical setting. The infusion system includes the FreedomEdge Infusion Pump, the High-Flo SubQ Needle Sets, and the Precision Flow Rate Tubing Sets. The infusion system is designed to be used with sterile, single use syringes (not manufactured or sold by Koru Medical Systems) as the drug reservoir. There have been no significant changes to the technological features (e.g., material, design, energy source, packaging, sterility, manufacturing method) since the previous clearance.
The fundamental operating principle of the FreedomEdge Infusion System is based on the Hagen-Poiseuille equation. The syringe containing the drug to be infused is connected to the flow rate control tubing and the subcutaneous needle set or venous access device and inserted into the pump. The syringe pusher of the pump applies a constant force to the plunger of the syringe. The infusion sets generate a defined amount of resistance which determines the flow rate. Flow rates are modeled using the Hagen-Poiseuille equation and verified at the boundary conditions for the system.
The design and operating principles of the system and its components have not been significantly modified since the device was first cleared. The same test methods were used as in prior submissions. Acceptance criteria were generated for Rystiggo using the same flow rate prediction model as used in prior submissions.
N/A
Ask a specific question about this device
(255 days)
95762
Re: K251105
Trade/Device Name: Tetra (TM-20); Tetra (TM-50)
Regulation Number: 21 CFR 880.5725
Infusion Pump
Classification Name: Immunoglobulin G (IgG) Infusion System
Regulation Number: 880.5725
The EMED Tetra Micro Infusion System is intended for subcutaneous infusion for prescription liquid medicines for use in the home and hospital environment by adults and children (2 years and older), when administered according to the approved biologic or drug product labeling. The Tetra Micro Infusion System is indicated for use with:
- Hizentra Immune Globulin Subcutaneous (Human) 20%, manufactured by CSL Behring.
The Tetra Micro Infuser is a reusable multi-patient use device, battery powered, prescription-only, designed to be used as a system for subcutaneous infusion using compatible syringes through EMED patient administration sets.
The syringe connects to the Tetra Micro Infuser, and the patient administration sets connects to the syringe. Patient administration sets are sterile and single use only.
There are 2 model options for the Tetra Micro Infuser: TM-20 (designed to connect to a 20mL syringe); and TM-50 (designed to connect a 50mL syringe). The TM-50 model can be used with 5 mL, 10 mL, and 20 mL syringe sizes when used with the optional Syringe Adapter (single-use, non-sterile; sold separately). Once attached to the syringe, the Syringe Adapter is not intended to be removed.
The patient administration sets are commercially available, previously marketed devices (K140131) and support the subject device without modification.
The Tetra Micro Infuser is designed for use in home and healthcare facility environments.
The Tetra Micro Infuser functions by applying controlled pressure to the syringe stopper (gasket) to deliver medication through the connected patient administration sets.
The user interface consists of two buttons used to power the device on/off and to select discrete delivery settings. The resulting flow rate depends on the selected setting and system factors such as syringe size, medication viscosity, and downstream resistance.
The device provides visual and audible notifications for status and certain conditions. The device battery is charged through a USB-C port, and infusion is disabled during charging. There is no user-accessible wireless or networking connectivity, and USB communication is restricted to factory/service access.
The Tetra Micro Infuser does not require calibration. Periodic cleaning of external surfaces is recommended between uses, and the infuser should be disinfected by healthcare or facility personnel between each patient when transferring to a new user.
Each Tetra Micro model is packaged in a labeled box containing: one infuser (either Tetra Micro TM-20 or TM-50); one USB-C to USB-A charging cable; one Instructions for Use document; and one Quick Reference Guide document.
N/A
Ask a specific question about this device
(90 days)
Netherlands
Re: K252636
Trade/Device Name: Fluido Compact System
Regulation Number: 21 CFR 880.5725
Compact SystemCommon or Usual Name: Infusion PumpClassification Name: Infusion Pump, 21 CFR 880.5725
System | Fluido® Compact Blood and Fluid Warming System | Same |
| CFR Regulation Number | 21 CFR 880.5725
| 21 CFR 880.5725 | Same |
| Product Code | LGZ | LGZ | Same |
| Classification | Class II |
The device can be used with the following products:
- Plasma
- Packed red blood cells
- Crystalloid intra venous (IV) fluids
- Synthetic Colloid intra venous (IV) fluids
The system is indicated to be used in a healthcare environment to help prevent hypothermia. The devices can be used in the full clinical environment. The Fluido Compact System can be used in adult and pediatric patients (weighing 3.5 kg or more) who require fluid warming prior to the administration of blood products or IV fluids to help prevent hypothermia. The system is intended to be used by healthcare professionals, e.g., nurses, medical specialists, and doctors.
The Fluido Compact System is a Blood and Fluid Warming System. The system is suitable for low to moderate flow applications. For high flow applications (higher than 100 ml/min for more than 5 minutes) the use of a high flow system is recommended.
The Fluido Compact System uses conduction heating technology to warm blood and fluids. Based on in-line sensors, the system calculates the energy required to safely warm the infusates.
The Fluido Compact System can be used with patients who require fluid warming prior to the administration of blood products or IV fluids, to help prevent hypothermia. The system is intended to be used in a healthcare environment by healthcare professionals, e.g. nurses, medical specialists, doctors. Only medical professionals shall interact with the system. The devices must only be used on the order of a physician.
N/A
Ask a specific question about this device
(349 days)
Physiological Monitors
Secondary Classification Regulations / Product Code, and Name
21 CFR §880.5725
§870.1025 | MSX, 21 CFR §870.1025 | Same. |
| Secondary Product Code, Regulation | PHC, 21 CFR §880.5725
| PHC, 21 CFR §880.5725 | Same. |
| Serves as secondary means of annunciating patient events | Yes
The intended use of TigerConnect Alarm Management is to provide an interface with clinical systems to forward information associated to the particular event to the designated display device(s).
For medical alarms, TigerConnect Alarm Management is intended to serve as a parallel, redundant, forwarding mechanism to inform healthcare professionals of particular medical related events TigerConnect Alarm Management does not alter the behavior of the primary medical devices and associated alarm annunciations. The display device provides a visual, and/or audible and/or vibrating mechanism upon receipt of the alert.
TigerConnect Alarm Management is intended for use as a secondary alert. It does not replace the alarm function on the primary device.
TigerConnect (TC) Alarm Management is a software as a medical device (SaMD) installed on the Amazon Web Services (AWS) TigerConnect cloud capable of acquiring alarms, events, and parameters from healthcare systems, and intelligently forwarding this information as secondary notifications to designated commercial endpoint devices. The endpoint devices act as non-regulated medical device data system (MDDS) devices that display the information from TC Alarm Management.
TC Alarm Management is intended for use as a secondary alarm; it does not replace the alarm function on the primary device.
Users receive interactive, time-critical information from clinical systems directly via their endpoint devices as text (visual) or alarms (audible) or data. Received attributes related to the presentation of alerts include color and quantity of tones (beeps) in addition to, and in coordination with, event priorities. TC Alarm Management allows users to be aware of their patients' status and alarm conditions when they are away from the patient and patient monitoring system.
A portion of the software, the Health Services Agent, is deployed to the customers' premises using ECS Anywhere, an extension of Amazon ECS that allows customers to run native ECS tasks on customer-managed infrastructure without compromising on the simplicity and control of the cloud. ECS Anywhere must be deployed in a customer provided virtual machine prior to implementing the Health Systems Agent. This allows TC Alarm Management to securely send, receive, process, and appropriately encrypt data from systems that are typically unencrypted.
TC Alarm Management connects to the information sources through wired ethernet connections which are part of the customer's infrastructure and acquires patient data from primary medical device sources and clinical systems. TigerConnect implementation engineers configure TC Alarm Management, in collaboration with the customer, to determine which information, including alarm notifications, is delivered to which users. TC Alarm Management then formats the data for wireless delivery to the endpoint devices through TigerConnect's proprietary Clinical Collaboration Platform (CCP) non-regulated MDDS messaging service.
TC Alarm Management is designed to accept inputs from a variety of healthcare systems utilizing the HL7 Minimum Lower Layer Protocol (MLLP) communication protocol. These clinical systems include the following:
- Medical device gateways providing HL7 MLLP communications, which includes, but is not limited to, the following parameters and events surrounding them, e.g. alarms:
- Patient monitors such as electrocardiograms (ECGs), respiratory rate, pulse oximetry (SpO2 and heart rate), end-tidal carbon dioxide (EtCO2), non-invasive blood pressure, cardiac output, temperature and associated derived outputs.
- Lung ventilators.
- Respiratory gas monitors.
- IV infusion (including patient-controlled analgesia) pumps.
- Nurse call systems.
- Electronic health record (EHRs) systems.
TC Alarm Management provides a wireless communications system platform for delivery of secondary notifications to endpoint devices with Android or iOS operating systems (OSs) through the CCP messaging service.
N/A
Ask a specific question about this device
(90 days)
Netherlands
Re: K251733
Trade/Device Name: Fluido® AirGuard System
Regulation Number: 21 CFR 880.5725
AirGuard System
Common or Usual Name: Infusion Pump
Classification Name: Infusion Pump, 21 CFR 880.5725
| Infusion Pump | Infusion Pump | Infusion Pump | Not applicable |
| Regulation Number | 21 CFR 880.5725
| 21 CFR 880.5725 | 21 CFR 880.5725 | Not applicable |
| Product Code | LGZ | FRN, LGZ | LGZ | Not
• Infusion of crystalloid, colloid or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery.
• Infusion of warmed fluid to patients with the ability to prevent hypothermia.
The Fluido® AirGuard System (FAS) is a pressurized infusion system intended for warming blood and fluids prior to administration to help prevent perioperative hypothermia. The FAS uses infrared heating technology to safely warm blood and fluids; the system detects air to prevent it from entering the patient.
The FAS consists of the Fluido Pressure Chambers, Fluido Blood and Fluid Warmer, Fluido AirGuard, Fluido Compressor, Fluido IV Pole, and Fluido Single-Use (Disposables): Standard Set and Trauma Set.
N/A
Ask a specific question about this device
(60 days)
Device Name:** SIGMA Spectrum Infusion Pump with Master Drug Library
Regulation Number: 21 CFR 880.5725
Classification Panel:** 80 General Hospital
Regulation Name: Pump, Infusion
Regulation Number: 21 CFR 880.5725
The pump provides delivery of fluids into a patient in a controlled manner, as identified in 21 CFR 880.5725
The pump provides delivery of fluids into a patient in a controlled manner, as identified in 21 CFR 880.5725
The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, or epidural. The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user.
The SIGMA Spectrum Infusion Pump with Master Drug Library is suitable for a variety of patient care environments such as, but not limited to, hospitals and outpatient care areas.
The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors. The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used by trained healthcare professionals.
SIGMA Spectrum is a large volume pump within the SIGMA Spectrum infusion system used by clinicians at the patient bedside to control the delivery of medications from a bag. The pump moves fluid from the bag to the patient via specified administration sets using a peristaltic pumping action. The pump offers various programmable delivery modes to address specific patient care needs. The delivery modes available to support the patient are determined by how the pump and its associated drug library are configured.
The pump provides delivery of fluids into a patient in a controlled manner, as identified in 21 CFR 880.5725. The system includes a software controlled, electromechanical pump used for the controlled administration of fluids including pharmaceutical drugs, blood, blood products, and mixtures of required patient therapy through administration sets at clinician's selectable rates and volumes.
The pump is intended for the controlled administration of fluids through the following clinically accepted routes of administration: intravenous, arterial, subcutaneous, and epidural. The pump is intended to be used in conjunction with legally marketed and compatible administration sets, as indicated in the device labeling, and medications provided by the user. The subject device is suitable for patient care in hospitals and outpatient health care facilities.
The Master Drug Library (MDL) is a stand-alone (not embedded in the pump) software application installed on a hospital-provided computing platform and used to create a drug library file. MDL facilitates the generation, configuration, and management of a facility-specific drug library file for dedicated infusion pumps. The drug library file is intended to be distributed to all compatible infusion pumps in the hospital.
This submission includes software design and labeling changes to address the issues leading to recalls Z-0530-2022 and Z-2103-2023.
This FDA 510(k) clearance letter pertains to an infusion pump, not an AI/ML powered medical device. Therefore, many of the requested categories in your prompt (such as "Number of experts used to establish the ground truth," "Adjudication method," "MRMC study," "Standalone performance," "Type of ground truth," and "Training set sample size/ground truth establishment") are not applicable to this type of medical device submission.
The document primarily focuses on demonstrating substantial equivalence to a predicate device through a comparison of technical characteristics and verification of performance against established requirements.
Here's an analysis based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" for the overall device in a quantifiable manner that would typically be found in an AI/ML context (e.g., specific sensitivity, specificity, or AUC targets). Instead, it demonstrates compliance with a range of technical specifications, which inherently act as acceptance criteria for the design and performance. The "Reported Device Performance" is implied by the statement that "Non-clinical testing met all acceptance criteria."
Below is a table summarizing key technical characteristics that function as performance criteria for the infusion pump. Since the subject device is deemed "substantially equivalent" to the predicate, and no new performance claims are made that deviate from the predicate, their performance characteristics are identical as presented.
| Characteristic | Acceptance Criteria (Subject Device & Predicate K230022) | Reported Device Performance (Subject Device) |
|---|---|---|
| Pumping Mechanism | Linear peristaltic design | Linear peristaltic design |
| Modes of Delivery | Continuous, Intermittent | Continuous, Intermittent |
| Routes of Admin. | Intravenous, Arterial, Subcutaneous, Epidural | Intravenous, Arterial, Subcutaneous, Epidural |
| User Interface Display | Color LCD | Color LCD |
| AC Power Input | 115 VAC ±15%, 50 - 60 Hz / 300 mA Max | 115 VAC ±15%, 50 - 60 Hz / 300 mA Max |
| AC Power Output | 9 VDC/1200 mA, short circuit protected | 9 VDC/1200 mA, short circuit protected |
| Operating Temp (Std/WBM) | 15.6 to 32.2°C (60 to 90°F), 20-90% RH non-condensing | 15.6 to 32.2°C (60 to 90°F), 20-90% RH non-condensing |
| Operating Temp (802.11b/g) | 15.6 to 26.7°C (60 to 80°F), 20-90% RH non-condensing | 15.6 to 26.7°C (60 to 80°F), 20-90% RH non-condensing |
| Storage Temp. | -10 to +49°C (14 to 120°F), 10-90% RH non-condensing | -10 to +49°C (14 to 120°F), 10-90% RH non-condensing |
| Single Fault Bolus | Max 0.56 mL | Max 0.56 mL |
| Anti-Free Flow System | Set-based, utilizing IV set slide clamp | Set-based, utilizing IV set slide clamp |
| Low Battery Alarm | ≤15 minutes of battery power remaining | ≤15 minutes of battery power remaining |
| Air-In-Line Detection | >2.5 cm air bubbles (140 μL in Baxter sets); >1 mL accumulated air over 15 min (room temp); >1.5 mL accumulated air over 15 min (15.5°C) | Meets criteria |
| Downstream Occlusion Alarms | User adjustable Low (41 kPa ±27 kPa), Medium (89 kPa ±41 kPa), High (131 kPa ±62 kPa) | User adjustable, meets specified values |
| Max Downstream Occlusion Press. | 207 kPa (30 psi) | 207 kPa (30 psi) |
| Flow Rate Range | 0.5 to 999 mL/hr | 0.5 to 999 mL/hr |
| Low-Flow Continuity | Max period of no-flow is 90 seconds at 0.5 mL/hr | Max period of no-flow is 90 seconds at 0.5 mL/hr |
| Volumetric Accuracy (DEHP sets) - 0.5-1.9 mL/hr | ±0.1 mL/hr (over 1 hr, up to 96 hrs) | ±0.1 mL/hr (over 1 hr, up to 96 hrs) |
| Volumetric Accuracy (DEHP sets) - 2.0-999 mL/hr | ±5% (over 1 hr, up to 96 hrs) | ±5% (over 1 hr, up to 96 hrs) |
| Volumetric Accuracy (Non-DEHP sets) - 10-125 mL/hr | ±10% (< 36 hrs tubing use) | ±10% (< 36 hrs tubing use) |
| Volumetric Accuracy (Non-DEHP sets) - 126-250 mL/hr | ±10% (< 4 hrs tubing use) | ±10% (< 4 hrs tubing use) |
| Drug Library Capacity - Care Areas | 32 | 32 |
| Drug Library Capacity - Drugs | 5000 | 5000 |
| Drug Library Capacity - Clinical Advisories | 400 | 400 |
| Drug Library Capacity - Modifiers | 500 | 500 |
| All Rates/Drug Library Limits | Within 0.5 to 999 mL/hr operational range | Within 0.5 to 999 mL/hr operational range |
2. Sample Size Used for the Test Set and Data Provenance
The document indicates that "Non-clinical testing has been executed against requirements for performance and safety." However, it does not specify the sample sizes for this testing.
The "data provenance" is implied to be from internal lab testing ("Non-clinical testing") conducted by Baxter Healthcare Corporation for verification of the device's functional and safety requirements. Details regarding country of origin or whether it's retrospective/prospective are not provided, but typically such testing for physical devices like infusion pumps would be prospective, controlled lab experiments.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable and therefore not provided in the document. For an infusion pump, "ground truth" is typically defined by physical and engineering measurements against established standards and specifications, rather than expert interpretation of data.
4. Adjudication Method for the Test Set
This information is not applicable and therefore not provided. Adjudication is a concept typically relevant to human review of (often ambiguous) data in clinical studies or AI performance evaluations, which is not the primary mechanism for verifying an infusion pump's compliance with engineering specifications.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
This information is not applicable and therefore not provided. The device is an infusion pump, not an AI-assisted diagnostic or therapeutic device used by human readers for interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This information is not applicable and therefore not provided. While the infusion pump contains software (algorithms), the "standalone performance" concept as typically applied to AI refers to the algorithm's performance in interpreting data directly without human intervention. The pump's performance is intrinsically linked to its electromechanical components and its interaction with fluids, administration sets, and clinical settings.
7. The Type of Ground Truth Used
For this infusion pump, the "ground truth" is established through design specifications, engineering standards, and direct physical/electrical measurements. For example, flow rate accuracy is measured against known volumes over time, alarm thresholds are tested against specific pressure or bubble sizes, and battery life is measured under defined operating conditions. This is not "expert consensus," "pathology," or "outcomes data" in the AI/ML sense.
8. The Sample Size for the Training Set
This information is not applicable and therefore not provided. Infusion pumps are not typically "trained" using data sets in the way AI/ML models are. Their functionality is programmed based on engineering principles and validated through testing.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable and therefore not provided for the reasons outlined in point 8. The "ground truth" for developing an infusion pump's functionality is its detailed engineering design and the physical laws and medical requirements it must adhere to.
Ask a specific question about this device
(60 days)
Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)
Regulation Number: 21 CFR 880.5725
Classification Panel:** 80 General Hospital
Regulation Name: Pump, Infusion
Regulation Number: 21 CFR 880.5725
The pump provides delivery of fluids into a patient in a controlled manner, as identified in 21 CFR 880.5725
The pump provides delivery of fluids into a patient in a controlled manner, as identified in 21 CFR 880.5725
The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, and blood products. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, or epidural. The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user.
The Spectrum IQ Infusion System with Dose IQ Safety Software is suitable for a variety of patient care environments such as, but not limited to, hospitals and outpatient care areas.
The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to reduce operator interaction through guided programming, including a way to automate the programming of infusion parameters and documentation of infusion therapies. This automation is intended to reduce pump programming errors.
The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to be used by trained healthcare professionals.
Spectrum IQ is a large volume pump within the Spectrum IQ infusion system used by clinicians at the patient bedside to control the delivery of medications from a bag. The pump moves fluid from the bag to the patient via specified administration sets using a peristaltic pumping action. The pump offers various programmable delivery modes to address specific patient care needs. The delivery modes available to support the patient are determined by how the pump and its associated drug library are configured.
The pump provides delivery of fluids into a patient in a controlled manner, as identified in 21 CFR 880.5725. The system includes a software controlled, electromechanical pump used for the controlled administration of fluids including pharmaceutical drugs, blood and blood products through administration sets at clinician's selectable rates and volumes.
The pump is intended for the controlled administration of fluids through the following clinically accepted routes of administration: intravenous, arterial, subcutaneous, and epidural. The pump is intended to be used in conjunction with legally marketed and compatible administration sets, as indicated in the device labeling, and medications provided by the user. The subject device is suitable for patient care in hospitals and outpatient health care facilities.
Dose IQ is a standalone software application installed on a hospital-provided computing platform and used to create a drug library file. Dose IQ facilitates the generation, configuration, and management of a facility-specific drug library file for dedicated infusion pumps. The drug library file includes customers' dosing limits and additional pump configuration settings. The drug library file is intended to be distributed to all compatible infusion pumps in the hospital.
This submission includes software design and labeling changes to address the issues leading to recalls Z-0529-2022 and Z-2104-2023.
The provided FDA 510(k) clearance letter for the Spectrum IQ Infusion System with Dose IQ Safety Software (K251636) does NOT describe a study involving an AI/Machine Learning algorithm for diagnostic or prognostic purposes, which would typically involve the criteria you've outlined.
Instead, this device is an infusion pump with safety software, intended to reduce operator interaction through guided programming to minimize "pump programming errors." The summary explicitly states: "This submission includes software design and labeling changes to address the issues leading to recalls Z-0529-2022 and Z-2104-2023." This suggests that the changes are primarily bug fixes, usability improvements, and potentially enhanced safety features rather than the introduction of a novel AI-driven diagnostic tool.
Therefore, many of the requested details about acceptance criteria and study design (e.g., sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth) are not applicable to this type of device and submission. These criteria are typically relevant for AI/ML-enabled devices that perform interpretation of medical images, signals, or patient data for diagnostic or prognostic purposes where "ground truth" and human reader performance are central to validation.
The 510(k) summary focuses on demonstrating substantial equivalence to an existing predicate device (K230041) through a comparison of technological characteristics and non-clinical performance and safety testing.
Here's how to interpret the provided information in the context of your request:
Acceptance Criteria and Device Performance (Based on the provided document)
The document doesn't present specific quantitative acceptance criteria in a table format for a "study" as you might expect for an AI/ML diagnostic device. Instead, "acceptance criteria" for an infusion pump typically refer to meeting defined performance specifications (e.g., volumetric accuracy, alarm thresholds, power consumption, environmental limits) and safety standards (e.g., IEC 60601 series).
The document states that "Non-clinical testing met all acceptance criteria, demonstrating that the device is safe and effective for its intended use." However, it does not provide the specific quantitative criteria or the numerical reported device performance for each. It assumes these were adequately documented in the full submission to FDA.
The "performance" described relates to the functional aspects of the pump and software:
| Characteristic | Acceptance Criteria (Implicit from "Subject Device" values) | Reported Device Performance (Same as Subject Device values) |
|---|---|---|
| Spectrum IQ Infusion System | ||
| Pumping Mechanism | Linear peristaltic design | Linear peristaltic design |
| Modes of Delivery | Continuous, Intermittent | Continuous, Intermittent |
| Routes of Administration | IV, Arterial, Subcutaneous, Epidural | IV, Arterial, Subcutaneous, Epidural |
| User Interface Display | Color LCD | Color LCD |
| AC Power (Input) | 100-240 VAC, 50 / 60 Hz / 300 mA Max | 100-240 VAC, 50 / 60 Hz / 300 mA Max |
| AC Power (Output) | 9 VDC/1000 mA, short circuit protected | 9 VDC/1000 mA, short circuit protected |
| Operating Temperature | 15.6 to 32.2°C (60 to 90°F), 20 to 90% RH non-condensing | 15.6 to 32.2°C (60 to 90°F), 20 to 90% RH non-condensing |
| Atmospheric Pressure | 66kPa to 102kPa | 66kPa to 102kPa |
| Storage Temperature | -10 to +35°C (14 to 95°F), 10 to 90% RH non-condensing | -10 to +35°C (14 to 95°F), 10 to 90% RH non-condensing |
| Single Fault Condition Bolus | Max 0.56 mL | Max 0.56 mL |
| Air-In-Line Detection (> 2.5 cm) | Detects >2.5 cm (approx 140 μL) bubbles | Detects >2.5 cm (approx 140 μL) bubbles |
| Air-In-Line Accumulated Air (Room Temp) | Detects >1 mL over 15 min, excluding <10μL | Detects >1 mL over 15 min, excluding <10μL |
| Air-In-Line Accumulated Air (15.5°C) | Detects >1.5 mL over 15 min, excluding <10μL | Detects >1.5 mL over 15 min, excluding <10μL |
| Low Battery Alarm | ≤15 minutes of battery power remaining | ≤15 minutes of battery power remaining |
| Flow Rate Range | 0.5 to 999 mL/hr | 0.5 to 999 mL/hr |
| Low-Flow Continuity (No-Flow Period) | Max 90 seconds at 0.5 mL/hr | Max 90 seconds at 0.5 mL/hr |
| Low-Flow Continuity (Bolus Volume) | Not exceed 15 μL over 5 sec at 0.5-1.0 mL/hr | Not exceed 15 μL over 5 sec at 0.5-1.0 mL/hr |
| Volumetric Accuracy (Non-DEHP, 0.5-1.9 mL/hr) | ±0.1 mL/hr over ≤ 4 hrs | ±0.1 mL/hr over ≤ 4 hrs |
| Volumetric Accuracy (Non-DEHP, 2.0-999 mL/hr) | ±5% over ≤ 4 hrs | ±5% over ≤ 4 hrs |
| Dose IQ Safety Software | ||
| Drug Library Capacity (Drugs) | 5000 drugs | 5000 drugs |
| Drug Library Capacity (Care Areas) | 32 care areas | 32 care areas |
| Drug Library Capacity (Clinical Advisories) | 400 clinical advisories | 400 clinical advisories |
| Drug Library Capacity (Modifiers) | 500 modifiers | 500 modifiers |
| Rate/Dose Limits Fall within Pump Range | 0.5 to 999 mL/hr | 0.5 to 999 mL/hr |
Regarding the other specific questions:
-
Sample sizes used for the test set and data provenance:
This information is not provided in a way that aligns with AI/ML model validation. The "non-clinical testing" would refer to engineering tests (e.g., bench testing for flow accuracy, alarm functionality, battery life) rather than a dataset for an AI model. Therefore, specific "test set" sample sizes or data provenance (e.g., country, retrospective/prospective) are not detailed because it's not a data-driven AI/ML diagnostic product. -
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This device does not use human experts to establish "ground truth" for a diagnostic "test set" in the way an AI/ML imaging device would. Ground truth for an infusion pump's performance is typically established by engineering measurement standards and physical verification methods, not expert consensus on medical data. -
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable for this type of device. -
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is an infusion pump with safety software, not an AI-assisted diagnostic or interpretation tool that augments human readers. Its purpose is to "reduce operator interaction through guided programming... to reduce pump programming errors," which implies improved safety and efficiency, but not through a "reader" study. -
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. The "Dose IQ Safety Software" is part of the "Spectrum IQ Infusion System," which is a physical medical device (an infusion pump). It's inherently a human-in-the-loop device where clinicians interact with it for programming and administration. Its software functionality is integral to the pump's operation. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Ground truth for an infusion pump's performance is based on engineering specifications, physical measurements (e.g., volume delivered over time, pressure readings, temperature, electrical parameters), and adherence to recognized industry standards (e.g., IEC 60601-1-8 for alarms). It's not based on expert consensus on clinical findings, pathology, or patient outcomes for diagnostic purposes. -
The sample size for the training set:
Not applicable. This is not an AI/ML device trained on a "dataset." The software is designed and validated against functional requirements and safety standards, not "trained" on a dataset in the machine learning sense. The "training set" for software development would typically be the iterative development and testing cycles against specifications. -
How the ground truth for the training set was established:
Not applicable, as there isn't a "training set" in the AI/ML context. Ground truth in this engineering context is established by design specifications derived from medical necessity, regulatory requirements, and established physical principles.
In conclusion, the provided FDA clearance letter details the regulatory approval of an infusion pump. While it includes "Dose IQ Safety Software," this software is focused on guided programming and error reduction (likely via a drug library and limits), not on AI/ML-driven diagnostics where the detailed validation criteria you've asked for would apply. The "study" referenced is non-clinical testing against performance and safety requirements, not a clinical trial or AI model validation study.
Ask a specific question about this device
(266 days)
The BD Intelliport™ System is an automated record keeping system that incorporates patient safety features that are aligned with hospital patient records and protocols. The system is comprised of an injection port and software that enables the identification, measurement, alerting and documentation of the administration of medications to patients.
The BD Intelliport™ System allows the clinician to record anesthesia-related medication administration events during pre-procedure, intra-procedure and recovery phase. The system is indicated for use by healthcare professionals in a hospital or medical center setting with patients who are receiving manually administered bolus intravenous injections as part of their care to facilitate documentation of the medications.
The BD Intelliport™ System is intended for patients with body weights >20 kg.
Do not use the BD Intelliport™ System with blood, blood products, biologics, or chemotherapeutics.
BD Intelliport™ System integrates into an intravenous line and automatically captures information about the anesthesia medications administered to the patient. It wirelessly transmits anesthesia medication administration information to the patient's Electronic Medical Record (EMR) via hospital server applications (Gateway software). The BD Intelliport™ System provides core technologies that enable key functions of the system:
• Medication Identification: Informs clinician of medication and concentration along with any informational notifications such as patient allergy and expired medication reminders. This occurs when syringes with the correct type of RFID encoded label are attached.
• Dose Measurement: Measures volume of drug administered to the patient through the system, then calculates dose weight.
• Automatic Charting: Wirelessly transmits measured doses to the EMR.
The following are the main system components:
• BD Intelliport™ Injection Site which is comprised of the following two components:
- BD Intelliport™ Sensor
- BD Intelliport™ Reader
• BD Intelliport™ Mount (optional accessory)
• BD Intelliport™ 2-Bay Charger (accessory)
• BD Intelliport™ Gateway
The provided FDA 510(k) clearance letter and summary for the BD Intelliport™ System (K243062) describes the performance testing conducted to demonstrate substantial equivalence to its predicate device (K182092). However, it does not provide specific acceptance criteria or reported device performance values in a quantifiable table format, nor does it detail a standalone study with quantitative results, or a multi-reader multi-case (MRMC) comparative effectiveness study.
The document primarily focuses on demonstrating substantial equivalence through a comparison of technological characteristics and a list of performance tests completed.
Here's an attempt to structure the answer based on the available information, with caveats where data is missing:
Acceptance Criteria and Device Performance
The 510(k) summary states that "The subject device, the BD Intelliport™ System, has met all predetermined acceptance criteria for the non-clinical and human factors testing conducted in accordance with relevant FDA guidance, recognized consensus standards, and internal requirements." However, the specific numerical acceptance criteria and the quantitative reported device performance values for most tests are not explicitly stated in the provided document.
The only quantitative performance criteria and reported values mentioned are for "Volume measurement accuracy" and "Volume Measurement Resolution."
| Acceptance Criteria | Reported Device Performance (Subject Device) |
|---|---|
| Volume measurement accuracy: | |
| For volumes >1.0 mL | ± 10% |
| For volumes 0.4 – 1.0 mL | ± 0.2mL |
| Volume Measurement Resolution | Uniform increments of 0.5 mL |
Note: The document states these are "Identical" to the predicate device, implying the reported performance matches the specified criteria.
Study Information
Due to the nature of a 510(k) summary, detailed study reports with specific quantitative results (beyond volume measurement accuracy) are not included. The document generally refers to "performance testing" and "human factors evaluation."
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated for each test. For "Volume Measurement Performance Window," the volume range tested was 0.5ml to 30ml, and average push speed 10ml/min to 400 ml/min. This implies tests were conducted across this range, but the number of injections or trials is not provided.
- Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be non-clinical (laboratory-based) and human factors studies. The human factors testing likely involved simulated clinical environments.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not provided within the scope of this 510(k) summary. These types of details are usually found in full study reports, not the summary itself, especially for a device that primarily automates record-keeping and measurement, rather than making diagnostic assessments that require expert ground truth. The human factors testing involved "intended device users," but their specific qualifications or roles in establishing "ground truth" (in a diagnostic sense) are not outlined.
-
Adjudication method for the test set:
- Not applicable/Not provided. Adjudication methods (like 2+1, 3+1) are typically used in studies where there is disagreement in expert interpretation of diagnostic data. This device automates measurements and record-keeping, so such a method is not relevant.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not explicitly reported or appears not to have been the primary method for demonstrating substantial equivalence. The device's primary function is "automated record keeping" and facilitating "documentation of the medications." It improves efficiency and accuracy of recording medication administration, rather than assisting human readers in interpreting medical images or data. The human factors evaluation assessed "critical tasks completed by intended device users," implying usability and user performance with the system, not a comparison of expert diagnostic accuracy with and without AI.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The document implies that standalone (algorithm-only) performance testing was conducted for various technical attributes. For example, "Bolus volume measurement accuracy," "Sensor flow rate," "Decoding response time," "Wifi functionality," and "Dose transmission time" are intrinsic functions of the system and its algorithms, which would have been tested independently of a human operator to verify their technical specifications. The "Flow Algorithm" itself was updated and "qualified through verification testing." However, explicit, detailed results from a standalone study with acceptance criteria are not presented in a formal table like format for all algorithm-driven functions.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the technical performance tests (e.g., volume measurement), the "ground truth" would have been established using calibrated instruments and reference standards (e.g., known volumes, known flow rates) in a laboratory setting.
- For the human factors evaluation, the "ground truth" would be defined optimal task performance and safety outcomes as determined by medical device standards and clinical best practices.
-
The sample size for the training set:
- The document does not provide information on a training set size. The BD Intelliport System involves embedded software and algorithms (e.g., flow algorithm, RFID reading, EMR communication). While such systems often involve development and testing cycles, the summary does not detail a specific "training set" like one would find for a machine learning or AI algorithm that learns from data in a traditional sense. The "Flow Algorithm" was updated and "qualified through verification testing," which implies validation against known physical models or experimental data, rather than a statistical "training set" in the context of deep learning.
-
How the ground truth for the training set was established:
- As no "training set" is explicitly mentioned for a machine learning model, this question is not applicable in the context of the provided information. The "ground truth" for verifying the updated flow algorithm would have been established through physical experiments and engineering measurements with known parameters (e.g., precise drug volumes, flow rates) to ensure the algorithm accurately processes the sensor data.
Ask a specific question about this device
(258 days)
Trade/Device Name: MRidium 3870 MRI Infusion Pump System (3870)
Regulation Number: 21 CFR 880.5725
Common name:** Infusion Pump
Classification name: Pump, Infusion
Regulation number: 21 CFR 880.5725
System | Same |
| Classification Name | Pump, Infusion | Pump, Infusion | Same |
| Regulation | 21 CFR 880.5725
| 21 CFR 880.5725 | Same |
| Class | II | II | Same |
| Product Code | FRN | FRN | Same |
| High Level
The Iradimed Corporation's MRidium 3870 MRI Infusion Pump System is indicated for general hospital or clinical use by medical professionals whenever it is required to infuse patients with intravenous fluids where a large magnetic field could be present such as Magnetic Resonance Imaging (MRI). The MRidium 3870 MRI Infusion pump can also be used during transport within the clinical environment as part of the periprocedural care before or after the MRI exam. The MRidium 3870 MRI Infusion Pump can be used with up to a 3.0 Tesla MRI system when placed as close as clinically possible to the MR system, without being placed within the magnet's bore, up to a 15,000 Gauss magnetic field line.
The pump and infusion sets are designed to deliver saline and IV medication. The system is useful in the administration of fluids requiring precisely controlled infusion rates. The system can operate in primary, loading, bolus, or KVO delivery mode.
The dedicated Iradimed 1056, 1058, and 1059 Infusion Sets for the MRidium 3870 MRI Infusion Pump are intended for single patient use up to 6 hours.
The device is intended for Adult and Pediatric patients, including Neonates. Careful clinical considerations involving drug and dosing should always be applied especially with regard to neonatal subpopulations (i.e., low and extremely low birth weight).
This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications.
The MRidium 3870 MRI Infusion Pump System is indicated for general hospital or clinical use by medical professionals whenever it is required to infuse patients with intravenous fluids where a large magnetic field could be present such as Magnetic Resonance Imaging (MRI). The MRidium 3870 MRI Infusion Pump can also be used during transport within the clinical environment as part of the periprocedural care before or after the MRI exam. The MRidium 3870 MRI Infusion Pump can be used with up to a 3.0 Tesla MRI system allowing placement as close as physically possible to the MRI scanner without being placed within the magnet's bore, up to a 15,000 Gauss magnetic field line. The magnetic content of the Pump is such that there is no hazard of magnetic attraction, up to 3T (30,000 gauss).
The pump unit is designed with an integral single peristaltic pump channel utilizing an ultrasonic (non-magnetic) motor. The integral infusion set mounting channel is horizontally oriented in the front of the pump with position detection sensors and graphical user feedback to aid the user in the correct infusion set loading process. The main assembly of the pump unit contains the controls, display, power supply, battery, and processor/memory functions suitable to meet all system requirements.
A remote display (3875) is also available as an option for independent viewing and control from the adjacent MRI Control Room area (Zone III).
The Dose Rate Calculator feature allows the user to set up a patient's infusion rate based upon user selected parameters, including volume to be infused, dose, concentration, weight, and/or time. The Dose Rate Calculator feature also provides a Drug Library, allowing the user to program a patient's infusion protocol from selected parameters, including volume to be infused, dose, concentration, weight and/or time with soft and hard limits for each drug. These Drug Library inputs can only be created or modified by the Iradimed in-house clinical team based on input and validation from the healthcare facility. Once programmed, the library is digitally signed requiring proper verification by a software 'key' within the 3870 MR IV pump upon loading. Loading of the drug library requires access to the Password protected Service Menu.
The Dose Error Reduction System (DERS) feature allows user-facilities to provide input to custom names, and doses, with hard and soft limits for use in the Drug Library. The Drug Library supports care area specific infusion protocols for primary, bolus, loading dose, and KVO delivery modes, retrievable by drug/protocol name. The library, as prescribed by the healthcare facility, is programmed with nominal starting values for: Dose, Concentration, and Time. Also, hard limits (maximum and minimum allowable) and soft limits (high and low limits that require a user confirmation to exceed) for Dose, Concentration, Time and Patient Weight can be programmed.
The provided FDA 510(k) clearance letter for the MRidium 3870 MRI Infusion Pump System focuses on demonstrating substantial equivalence to a predicate device (MRidium 3860+ MRI Infusion Pump/Monitoring System) through non-clinical performance testing. It does not describe a study involving human readers or the establishment of ground truth by expert consensus, as might be typical for AI/ML-driven medical device submissions.
Therefore, I will extract information related to the device's functional performance, which serves as its "acceptance criteria," and the "study" (non-clinical testing) that proves it meets these criteria. Many of the requested points (e.g., number of experts, adjudication methods, MRMC studies, sample sizes for training sets) are not applicable to this type of device and submission.
Here's the breakdown based on the provided document:
Acceptance Criteria and Study Proving Device Performance for MRidium 3870 MRI Infusion Pump System
1. Table of Acceptance Criteria and Reported Device Performance
The document describes performance specifications and indicates that the device meets them, primarily by demonstrating substantial equivalence to the predicate and through various verification tests. The "acceptance criteria" are implied by the specifications themselves, and the "reported device performance" is that it complies with these specifications and has been tested to perform as intended.
| Characteristic / Acceptance Criteria | Reported Device Performance (as stated in comparison or test summary) |
|---|---|
| Flow Rate Range | 0.4 to 1000 mL/hr in 0.1 mL/hr increments |
| Flow Rate Accuracy | +/- 5% 1.0 mL/hr to 1000 mL/hr; +/- 10% 0.4 ml/hr to <1.0 mL/hr (during 6-hour Infusion Set life) |
| Primary Volume to Be Infused (VTBI) Range | 0.1 to 1000 mL |
| Total Volume Infused (VI) Range | 0.1 to 99999 mL |
| Keep Vein Open (KVO) Rate Range | Programmed in Drug Library – OFF, Rate (0.4 mL/hr to 20 mL/hr), Continue Primary |
| Patient Line (downstream) Back-Pressure Range | +300 to -100 mmHg |
| Downstream Occlusion Detection Range | 1 to 10 PSI (6.9 to 68.8 kPa), user-adjustable |
| Occlusion Detection (no flow) Time | Variable, detailed in Operation Manual (REF 1234) table based on IV Set, Flow Rate, and Occlusion Pressure Alarm Limit Setting. (Improved labeling over simple predicate statement). |
| Maximum Post-Occlusion Bolus Volume | 0.7 mL max |
| Air-in-Line Detector Threshold(s) | > 100 uL (+20%) |
| Audible Alarm Range Pump (at 1 Meter) | Minimum: 65 dBA; Maximum: > 85 dBA @ 1 meter |
| Pump Battery Operating Time | > 8 Hours at 125 mL/hr Rate |
| Pump Battery Charge Time | < 5 hours to 95% capacity |
| Pump MRI Conditions: B0 Field / Gauss-Line Limit | ≤ 15,000 gauss |
| Operating Temperature Range | +15○C to +30○ C (+59○F to +86○ F) |
| Operating Relative Humidity Range | 30% to 80% RH, non-condensing |
| Ingress Protection | IPX2 |
| Wireless Encryption | Encrypted for security (Improvement over predicate) |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Sample Size for Test Set: The document indicates that non-clinical bench testing was performed. Specific numerical sample sizes (e.g., how many pumps were tested, how many infusion sets) are not explicitly stated in the provided text. The testing categories include:
- Verification Testing of Product Requirements (e.g., flow rate accuracy, essential performance, reliability, air-in-line, occlusion, alarms, software V&V, cybersecurity, battery, EMC/EMI, environmental, shock/vibration, electrical safety, mechanical, biocompatibility, ingress protection, wireless co-existence, standards testing).
- Human Factors Validation Testing.
- MRI Environment Conditions of Use Testing (magnetic displacement, spurious emissions, essential performance during MRI scans including flow accuracy, alarms, unintended bolus, wireless communication).
-
Data Provenance: The data is from non-clinical bench testing performed by Iradimed Corporation (the manufacturer) as part of their 510(k) submission. No information about country of origin of "data" in the sense of patient data is relevant here, as it's not a clinical study on patients. It is implicitly prospective testing, designed to demonstrate compliance for the submission.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This section is not applicable. The device is an infusion pump, and its performance is evaluated against engineering and safety standards, not against human expert interpretations of medical images or patient outcomes. The "ground truth" for this device's performance verification comes from established engineering principles, metrology, and consensus standards (e.g., AAMI TIR101 for flow accuracy, IEC 60601 series for safety).
4. Adjudication Method for the Test Set
This section is not applicable. As the testing is non-clinical and pertains to physical and electrical performance, there is no need for expert adjudication in the traditional sense of clinical or image-based studies. Performance is measured against predefined specifications.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This section is not applicable. MRMC studies are typically performed for AI/ML-driven diagnostic devices where human readers provide interpretations (e.g., radiologists reading medical images) and their performance with and without AI assistance is compared. This device is an infusion pump, not a diagnostic AI tool.
6. Standalone Performance Study (i.e., algorithm only without human-in-the-loop performance)
This is largely applicable to the extent that the device's functional performance (e.g., flow rate, alarm detection, battery life) is evaluated independently of human interaction for its core functions in a controlled bench setting. The "Summary of Performance Testing (Non-Clinical)" details various tests (e.g., flow rate accuracy, essential performance, reliability, air-in-line, occlusion performance, alarms, cybersecurity) that represent the device's standalone operation.
7. Type of Ground Truth Used
The "ground truth" for this device's performance is established through:
- Engineering specifications and design requirements: The device is designed to meet specific measurable performance parameters (e.g., flow rate accuracy, pressure detection thresholds).
- Consensus standards: Performance is verified against recognized industry standards (e.g., AAMI TIR101 for flow accuracy, IEC 60601 series for safety and alarms).
- Bench testing measurements: Physical measurements using calibrated equipment are the "ground truth" to confirm the device's output against its specifications.
8. The Sample Size for the Training Set
This section is not applicable. This is a hardware medical device (an infusion pump) and not an AI/ML device that undergoes "training" on a "training set" of data. Its "training" is in its design, engineering, and manufacturing processes, culminating in verification and validation testing.
9. How the Ground Truth for the Training Set Was Established
This section is not applicable for the same reason as point 8.
Ask a specific question about this device
(130 days)
Calculation Services (CS) (N/A); BD Alaris Systems Manager (SM) (9601)
Regulation Number: 21 CFR 880.5725
Description | Components/Module |
|----------------|--------------|-------------|-------------------|
| 880.5725
The BD Alaris Infusion System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. See Pediatric*, Neonate**, and Adult Patient Population Tables 2 and 3 for the module-specific variations. Administered fluids include pharmaceutical drugs, red blood cells, and other blood components (platelets and fresh frozen plasma) as required for patient therapy. The BD Alaris Infusion System with Guardrails Suite MX is an interoperable system capable of communicating and exchanging data with compatible information technology systems.
The BD Alaris Infusion System with Guardrails Suite MX includes the PC Unit (PCU) and one or more of the following: Pump Module, Syringe Module, end-tidal CO2 (EtCO2) Module, Auto-ID Module, Patient-Controlled Analgesia (PCA) Module, and associated software applications. EtCO2 Module is a capnograph that continuously monitors end-tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).
BD Alaris Pump Module and Syringe Module and the Alaris PCA Module are indicated for varying patient populations, routes of administration, and infusates.
The BD Alaris Infusion System with Guardrails Suite MX is a modular infusion and monitoring system designed to provide controlled delivery of drugs and fluids, and to provide monitoring of respiratory parameters. The BD Alaris Infusion System with Guardrails Suite MX has three (3) major components:
-
System Hardware: A core hardware unit with user interface (BD Alaris PC Unit or PCU) and attachable modules each with a distinct function.
- BD Alaris Pump Module (LVP)
- BD Alaris Syringe Module (SYR)
- Alaris PCA Module (PCA)
- BD Alaris EtCO2 Module (EtCO2)
- Alaris Auto-ID Module (Auto-ID)
-
Guardrails Suite MX Software: Software applications for support and interaction with the system hardware
- BD Guardrails Editor (GRE)
- BD Alaris Systems Manager (SM)
- BD Alaris Systems Maintenance (ASM)
-
Interoperability Software: Software applications for facilitating bi-directional communication between the PCU and attached LVP and SYR modules, and an electronic medical records (EMR) system via BD Alaris Infusion Systems Manager (SM) and Care Coordination Engine (CCE), a non-medical device Medical Device Data System (MDDS).
- Calculation Services
- Infusion Adapter (IA)
The PCU is the core of the BD Alaris Infusion System with Guardrails Suite MX and necessary for its operation. It provides a common user interface for programming, and powering and monitoring attached modules. Modules must be physically connected to the PCU to operate. The connection is made by direct attachment to a PCU or through attachment to a module that is attached to a PCU. The attachment is made using inter-unit interface (IUI) connectors built into both sides of the PCU and modules, which also serve to provide power to the modules and communication between the PCU and attached modules. The PCU is powered by AC power and has a rechargeable battery to allow for continued therapy during power interruptions.
The attachable modules are dedicated to controlled delivery of fluids, pharmaceutical drugs, parenteral nutrition, and blood and blood products into patients, patient-controlled administration of analgesics, monitoring of end-tidal carbon dioxide, and scanning identifications of patient, physician, and infusates into the system.
-
The BD Alaris Pump Module (LVP) delivers fluids accurately over programmed times and can detect and notify the user of situations that could impact patient safety, such as improper set loading, occlusion, and air-in-line. It can deliver fluids continuously or intermittently from any compatible container using a dedicated BD Alaris infusion set. Flow rates range from 0.1 to 999 mL/h and bolus doses can be programmed at the start or during continuous infusion.
-
BD Alaris Syringe Module (SYR) is designed for injecting fluids from compatible syringes and can detect and notify the user of situations that could impact patient safety, such as an improperly loaded syringe and occlusion. It can deliver continuous or intermittent volumes from 1 to 50 mL syringes at flow rates of 0.01 to 999 mL/hr.
-
The Alaris PCA Module (PCA) is designed for patient-controlled analgesia. It shares core components and functionality with the BD Alaris Syringe Module but includes additional features such as: a dose request cord for self-administering pain medication, software with a dose lockout interval, and a locking syringe enclosure door with a key. When configured for use with the BD Alaris EtCO2 Module, it can also trigger a pause of the PCA infusion if the respiratory rate of a patient falls outside the limits.
-
The BD Alaris EtCO2 Module (EtCO2) a capnograph used for continuous, non-invasive monitoring of end-tidal CO2, fractional inspired CO2, and respiratory rate. It can be used to monitor respiratory depression in patients when using the Alaris PCA Module.
-
The Alaris Auto-ID Module (Auto-ID) features an internal barcode image scanner and an optional handheld scanner. Scanning a clinician ID unlocks the PCU panel in authorized user mode and links clinical event logs with the clinician. Scanning a patient ID band associates logs with the patient, while scanning IV fluid or medication barcodes selects the specific item from the drug library for infusion modules.
The PCU and attachable modules have multiple processors running embedded software. The embedded software provides various functions, such as: bootloader, user interface, networking, sensor monitoring, motor control, data processing, power control, keypad processing, and communication.
The PCU with its attached modules is designed to be configured to communicate and interact with the Guardrails Suite MX software applications including software for interoperability with Electronic Medical Record (EMR) systems. Communication between the PCU and the software applications is accomplished through either a direct serial connection or through a wireless connection utilize the respective Guardrails Suite MX Software applications.
-
The BD Guardrails Editor (GRE) allows for the creation of drug and fluid libraries and guidelines, called 'profiles,' for specific patient populations. GRE also provides a transfer tool to transfer a profile to PCU via serial cable.
-
The BD Alaris Systems Manager (SM) manages connectivity and includes a web application, communications server, and database software for managing data, creating reports, connecting with a healthcare facility's network, and storing system configuration, user permissions, and historical data. Use of SM also supports transferring wireless software updates to the PCU during system servicing.
-
BD Alaris Systems Maintenance (ASM) is used for standard maintenance tasks, including module calibration and network configurations
-
The BD Alaris Interoperable software includes the Infusion Adapter (IA) and Calculation Services to support bi-directional communication between the BD Alaris Infusion System with Guardrails Suite MX and the healthcare facility's EMR. The Infusion Adapter facilitates data exchange ensuring correct message formats and content. Calculation Services performs pre-defined rule-based calculations to obtain infusion duration, body surface area (BSA), and weight-based dose.
It is important to note that interoperability does not include remote control of the BD Alaris Infusion System with Guardrails Suite MX. The PCU and attached modules cannot be programmed remotely. Only infusion parameters can be prepopulated on the pump using interoperability and these parameters must be manually confirmed by the clinician at the bedside before they are activated.
The provided FDA 510(k) clearance letter and summary for the BD Alaris Infusion System with Guardrails Suite MX (K243855) do not contain detailed information about specific acceptance criteria and a study proving the device meets those criteria in the context of an AI/algorithm performance evaluation. Instead, the document focuses on the substantial equivalence of an infusion pump system to a predicate device, with an emphasis on its hardware, software (including safety management and interoperability features), and general electrical and functional safety.
The text does not describe an AI/algorithm in the sense of a diagnostic or assistive AI that requires expert-driven ground truth, MRMC studies, or standalone performance metrics typically associated with AI/ML-based medical devices. The "Guardrails Suite MX" and "Calculation Services" mentioned are primarily about drug library management, dose error reduction, and rule-based calculations, which are more akin to conventional software functionalities rather than adaptive AI algorithms that learn from data.
Therefore, many of the requested categories for describing an AI/algorithm acceptance study are not applicable to the information provided in this document. Given the nature of the device (an infusion pump system), the "acceptance criteria" discussed are related to its functional performance, safety, and compliance with regulations and standards.
However, I can extract the relevant "acceptance criteria" and "performance" data that are presented in the document, framed within the context of a traditional medical device's non-clinical testing.
Reported Device Performance and "Acceptance Criteria" (based on functional and safety requirements):
The document describes non-clinical testing to verify essential performance requirements. These requirements serve as the de facto "acceptance criteria" for the device's main functions.
| Acceptance Criterion (Implicitly Derived from "Essential Performance") | Reported Device Performance (Subject Device) | Notes/Comments |
|---|---|---|
| LVP Flow Rate Accuracy (Standard Operating Conditions) | ±5% system flow rate accuracy for 1 to 999 mL/hr -8 % to + 5.5% system flow rate accuracy for 0.1 to 1 mL/hr | This is explicitly stated as the updated claim for the LVP module at Standard Operating Conditions, reflecting no change in actual performance requirements from the predicate. The full range of accuracy at non-SOC is presented in the User Manual. |
| SYR Flow Rate Accuracy | ± 7% system flow rate accuracy for > 10% of syringe volume/hr ± 10% system flow rate accuracy for > 0.1 mL/hr (Syringe sizes < 12 mL) / > 1 mL/hr (Syringe sizes > 12 mL) ± 20% system flow rate accuracy for < 0.1 mL/hr (Syringe sizes < 12 mL) / < 1 mL/hr (Syringe sizes > 12 mL) | Explicitly stated performance. "SAME" as predicate. |
| PCA Flow Rate Accuracy | ± 7% system flow rate accuracy for > 10% of syringe volume/hr ± 10% system flow rate accuracy for > 1 mL/hr ± 20% system flow rate accuracy for < 1 mL/hr | Explicitly stated performance. "SAME" as predicate. |
| EtCO2 Module Accuracy | ± 2 mmHg CO2 Conc for 0–38 mmHg 5% of reading + 8% per mmHg (above 38 mmHg) for 39–99 mmHg | Explicitly stated performance. "SAME" as predicate. |
| Maximum Infusion Pressure (Pump Module) | 525 mmHg | Explicitly stated performance. "SAME" as predicate. |
| Maximum Infusion Pressure (Syringe Module) | Approx. 800 mmHg (without pressure sensing disc) 1060 mmHg (with pressure sensing disc) | Explicitly stated performance. "SAME" as predicate. Actual value varies by syringe. |
| LVP Bolus Accuracy | Without Rapid Bolus Feature: > 0.2 mL: ±10%; < 0.2 mL: 0.2 mL: ±0.02 mL, 0.1 mL: ±0.025 mL With Rapid Bolus Feature: > 1 mL: ±10%; > 0.6 mL and < 1 mL: ±15%; < 0.6 mL: 0.6 mL: ±0.06 mL, 0.1 mL: 0 mL to +0.055 mL | Explicitly stated performance. "SAME" as predicate. |
| SYR Bolus Accuracy | > 0.2 mL: ±10%; < 0.2 mL: ±20% (Recommended min bolus > 0.2 mL) | Explicitly stated performance. "SAME" as predicate. |
| PCA Bolus Accuracy | > 0.2mL: ±10%; < 0.2mL: ±20% (Recommended min bolus > 0.2 mL) | Explicitly stated performance. "SAME" as predicate. |
| Post-occlusion Bolus Volume (Pump Module) | ≤ 0.3 mL for all pressure settings (standard operating conditions) | Explicitly stated performance. "SAME" as predicate. |
| Post-occlusion Bolus Volume (Syringe and PCA Module) | ≤ 1.0 mL for all pressure settings (standard operating conditions) | Explicitly stated performance. "SAME" as predicate. |
| Protection against Inadvertent Delivery | Tested for critical volume, free flow, bolus during set loading, post-occlusion bolus, and means to pause infusion. | Confirmed as verified in non-clinical testing. |
| Alarm Detection/Notification | Tested for conditions like interrupted delivery/occlusions, air in line, battery status, device malfunction. | Confirmed as verified in non-clinical testing. |
| Software Requirements | Verified via code review, static analysis, unit testing, integration testing, and regression testing. | Confirmed as verified in non-clinical testing. |
| Hardware Requirements | Verified. | Confirmed as verified in non-clinical testing. |
| Hardware/Software Compatibility | Verified. | Confirmed as verified in non-clinical testing. |
| System Operational Requirements | Verified. | Confirmed as verified in non-clinical testing. |
| Medical Device Interoperability | Verified (BD Alaris Interoperable software facilitates bi-directional communication with EMR). | Confirmed as verified in non-clinical testing. |
| Biocompatibility | Verified as biocompatible. | Confirmed as verified in non-clinical testing. |
| System Reliability | Verified via testing and statistical analysis at system, device subsystem, and subsystem/component levels. | Confirmed as verified in non-clinical testing. |
| Electrical Safety & EMC Compliance | Successfully completed testing to ANSI/AAMI ES 60601-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-24, UL 1642, IEC 62133-1, ISO 80601-2-55, IEC 60601-1-8, ANSI/IEEE USEMCSC C63.27. | Confirmed as compliant with relevant standards. |
| Cleaning and Disinfection Validation | Validated according to FDA Guidance. | Confirmed as verified. |
| Human Factors/Usability | Design validation performed via clinical assessment, simulated testing, biomedical engineering use, use-related risk analysis, and IEC 62366-1. | Confirmed as safe and effective for intended use, users, and environments. |
| Cybersecurity Controls | Assessment and verification performed according to FDA guidance. | Confirmed as performed. |
Since the provided document is a 510(k) clearance letter for an infusion pump system, not an AI/ML-based diagnostic or assistive device, the following points are largely not applicable or not explicitly detailed in the text. I will state if the information is unavailable or implies "None" for the context of this specific device's clearance.
-
Sample sizes used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated for all performance tests. The document refers to "testing" and "statistical methods in sample size determination and data analysis" but does not provide specific numbers for each test (e.g., how many pumps were tested for flow rate accuracy). This is common for 510(k) summaries where detailed test reports are typically referenced but not fully included.
- Data Provenance: Not specified regarding country of origin. The testing would generally be conducted by the manufacturer (BD/CareFusion) or their approved test labs. It describes "non-clinical testing" and "simulated clinical conditions," which indicates a prospective validation within a controlled environment.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable in the AI/ML sense. Ground truth for an infusion pump's performance (e.g., flow rate accuracy, alarm function) is established through engineering measurements and adherence to international standards (like AAMI TIR 101, ISO 80601-2-55). It doesn't involve expert consensus on medical images or clinical outcomes in the way an AI diagnostic would. The "Human Factors evaluation" mentions "clinical assessment" and "biomedical engineering use," implying input from relevant experts, but not for "ground truth" labeling of data.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not Applicable. This is a method for resolving discrepancies in expert labeling of data, which is not relevant for the type of objective functional performance testing described for an infusion pump.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This device is an infusion pump system, not an AI for human reader assistance in diagnostic tasks.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Partially Applicable / Different Context. The document details extensive "essential performance" testing of the device (hardware and embedded software) in a standalone capacity, demonstrating its accuracy, safety mechanisms, and compliance with standards. This constitutes "algorithm only" performance in the sense of the pump's control algorithms (e.g., for flow rate, pressure detection). However, it's not an "AI algorithm" in the typical understanding of machine learning where a "human-in-the-loop" interaction for clinical decision-making is assessed.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Engineering Measurements and Compliance with Standards. The "ground truth" for this device's performance is derived from precise engineering measurements, calibration standards, and adherence to established medical device performance standards (e.g., AAMI TIR 101 for flow rate accuracy requires specific test methods and reference measurements). For the EtCO2 module, it's based on accuracy against known gas concentrations.
-
The sample size for the training set:
- Not Applicable (in the AI/ML sense). This device is not described as utilizing machine learning that requires a "training set" of data for algorithm development. Its software functionalities (e.g., Guardrails Suite MX) are rule-based systems or deterministic algorithms, developed through traditional software engineering and verification processes.
-
How the ground truth for the training set was established:
- Not Applicable. As no AI/ML training set is indicated, this question is not relevant.
In summary, the provided document meticulously outlines the non-clinical validation of an infusion pump system, demonstrating its safety and effectiveness through adherence to performance specifications and regulatory standards. It does not, however, pertain to the clearance of an AI/ML diagnostic or assistive algorithm, which would involve the specific types of studies and ground truth methodologies requested in the prompt.