Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K252435

    Validate with FDA (Live)

    Date Cleared
    2025-09-26

    (56 days)

    Product Code
    Regulation Number
    878.4815
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K252435
    Trade/Device Name: Magnetic Surgical System
    Regulation Number: 21 CFR 878.4815
    Re: K252435
    Trade/Device Name: Magnetic Surgical System
    Regulation Number: 21 CFR 878.4815
    instrument system |
    | Classification Name | Magnetic Surgical System |
    | Regulation Number | 878.4815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Magnetic Surgical System is indicated for use in patients within a BMI range of 20 to 60 kg/m2 and designed to grasp and retract soft tissues and organs to facilitate access and visualization of the surgical site for the following:

    • The body and fundus of the gallbladder in laparoscopic cholecystectomy procedures
    • The liver in bariatric procedures
    • The prostate and periprostatic tissue in prostatectomy procedures
    • The colon, rectum, and pericolorectal tissue in colorectal procedures

    The Grasper Tip 12.5 is indicated for use in patients within a BMI range of 20 to 60 kg/m2 and designed to grasp and retract soft tissues and organs to facilitate access and visualization of the surgical site for the following:

    • The pillar of the diaphragm and peripillar tissue in bariatric or hiatal hernia procedures

    Pediatric use:

    The Magnetic Surgical System is indicated for use in adolescents (12-21) within a BMI range of 20 to 34 kg/m2 and designed to grasp and retract soft tissues and organs to facilitate access and visualization of the surgical site for the following:

    • The body and fundus of the gallbladder in laparoscopic cholecystectomy procedures
    Device Description

    The Magnetic Surgical System comprises two hand-held instruments, the Magnetic Grasper and the external Magnetic Controller. The Magnetic Grasper, disposable and provided sterile for single use, comprises two main components: a detachable Grasper Tip (6.5cm overall length) and a shaft with Handle. An optional 12.5cm length Grasper Tip (also single use, disposable, and provided sterile) may be used as a replacement for the 6.5cm Grasper Tip and can be used interchangeably with the Shaft.

    The Magnetic Grasper is actuated via its pistol-grip handle with two distinct scissor-type motions to open and close the Grasper Tip jaws. Once the Magnetic Grasper is inserted through a compatible >= 10 mm laparoscopic port to the surgical site and the Grasper Tip is attached to the desired tissue, the Grasper Tip can be detached from the Shaft and controlled externally using the Magnetic Controller. Traction of the tissue is maintained through the magnetic field attraction between the Grasper Tip and the Magnetic Controller.

    The Magnetic Controller is a non-sterile, reusable unit with handles that is held external to the body and emits a magnetic field that attracts the detachable Grasper Tip. Once the Grasper Tip is attached to the desired tissue and detached from the Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Grasper Tip to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Grasper Tip will modulate the magnetic attraction used for tissue retraction/mobilization. If desired, the user can connect the Magnetic Controller to a commercially available surgical support arm that is compatible with its arm mount.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250746

    Validate with FDA (Live)

    Date Cleared
    2025-06-07

    (87 days)

    Product Code
    Regulation Number
    878.4815
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 94010

    Re: K250746
    Trade/Device Name: Magnetic Surgical System
    Regulation Number: 21 CFR 878.4815
    surgical instrument system |
    | Classification Name | Magnetic Surgical System |
    | Regulation Number | 878.4815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Levita Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures; the liver in bariatric procedures; the pillar of the diaphragm and peripillar tissue in bariatric or hiatal hernia procedures; the prostate and periprostatic tissue in prostatectomy procedures; and the colon, rectum, and pericolorectal tissue in colorectal procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients with a BMI range of 20-60-kg/m2.

    The Grasper Tip, 12.5 is designed to grasp and retract the pillar of the diaphragm and peripillar tissue in bariatric or hiatal hernia procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients with a BMI range of 20-60 kg/m2.

    Device Description

    The Magnetic Surgical System comprises two hand-held instruments, the Magnetic Grasper and the external Magnetic Controller. The Magnetic Grasper, disposable and provided sterile for single use, comprises two main components: a detachable Grasper Tip (6.5cm overall length) and a Shaft with Handle. An optional 12.5cm length Grasper Tip (also single use, disposable, and provided sterile) may be used as a replacement for the 6.5cm Grasper Tip and can be used interchangeably with the Shaft.

    The Magnetic Grasper is actuated via its pistol-grip handle with two distinct scissor-type motions to open and close the Grasper Tip jaws. Once the Magnetic Grasper is inserted through a compatible ≥ 10 mm laparoscopic port to the surgical site and the Grasper Tip is attached to the desired tissue, the Grasper Tip can be detached from the Shaft and controlled externally using the Magnetic Controller. Traction of the tissue is maintained through the magnetic field attraction between the Grasper Tip and the Magnetic Controller.

    The Magnetic Controller is a non-sterile, reusable unit with handles that is held external to the body and emits a magnetic field that attracts the detachable Grasper Tip. Once the Grasper Tip is attached to the desired tissue and detached from the Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Grasper Tip to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Grasper Tip will modulate the magnetic attraction used for tissue retraction/mobilization. If desired, the user can connect the Magnetic Controller to a commercially available surgical support arm that is compatible with its arm mount.

    AI/ML Overview

    The provided text describes the 510(k) clearance for the Levita Magnetic Surgical System. Here's a breakdown of the requested information.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document mentions acceptance criteria were met across various tests, but it does not specify the numerical criteria or detailed performance metrics for each. It generally states that the device "passed all tests and met all acceptance criteria."

    Test CategoryAcceptance Criteria (Not explicitly stated numerically, but implied as "met")Reported Device Performance
    Non-Clinical Mechanical PerformanceAll design requirements and performance specifications met for the modified Magnetic Grasper and Grasper Tip, 12.5 line extension for 12 months.The device "passed all tests and met all acceptance criteria." These results "demonstrate that for the entirety of the labeled shelf-life of 12 months, all design requirements and performance specifications have been met and the design conforms to user needs and meets the intended use."
    Human Factors Assessment (IEC 62366-1)No additional human factors validation testing required for the changes.Concluded "no additional human factors validation testing is required for the changes."
    Biocompatibility (ISO 10993-1)Compliance with standard.Biocompatibility testing of the Grasper Tip, 12.5 line extension passed.
    Sterilization Adoption (AAMI TIR35, ISO 11137-1, ISO 11137-2)Compliance with standards.Sterilization adoption testing for the line extension passed.
    Real-time Shelf-life Testing12-month shelf life maintained.Real-time shelf-life testing of the modified Magnetic Grasper passed, demonstrating a 12-month shelf life.
    Packaging Testing (ASTM F2096, ASTM F88)Compliance with standards.Packaging testing for the line extension passed.
    Risk Assessment (ISO 14971: 2019)No new or different risks identified."No new or different risks were identified."
    Clinical Study - Primary Safety EndpointNo severe or serious adverse events related to the device."The safety results met the criteria outlined for the primary safety endpoint for this study; there were no severe or serious adverse events related to the device."
    Clinical Study - Performance EndpointsMSS able to grasp the pillar of the diaphragm or peripillar tissue to adequately mobilize the liver, achieving effective exposure of the target tissue. No need for another liver retractor. No conversions to open surgery due to device malfunction. No reported device malfunctions leading to conversions or use of another retractor. No unanticipated adverse events related to the device."In all 30 cases, the MSS was able to grasp the pillar of the diaphragm or peripillar tissue to adequately mobilize the liver, achieving an effective exposure of the target tissue." "It was not necessary to use another liver retractor during any procedures." "No procedures required conversion to an open surgical approach." "There were no reported device malfunctions that led to conversions to open surgery or use of another liver retractor." "There were no unanticipated adverse events related to the device."

    2. Sample Size Used for the Test Set and Data Provenance

    For the clinical study:

    • Sample Size: Thirty (30) subjects.
    • Data Provenance: Prospective, single-arm, open-label clinical study conducted at three (3) sites in Santiago, Chile.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts used to establish ground truth or their specific qualifications for the clinical study. It mentions that "five (5) surgeons" performed the surgeries and evaluated the device's performance, implying their clinical judgment served as the ground truth. However, their specific qualifications (e.g., years of experience, specialization) are not detailed beyond being "surgeons."

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method like 2+1 or 3+1. Given it was a single-arm, open-label study where surgeons performed the procedures and evaluated outcomes, the assessment of safety and performance endpoints likely relied on the treating physicians' observations and reporting, aligned with the study protocol. There is no mention of an independent adjudication committee for specific endpoints.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done for this submission. The clinical study was a single-arm study evaluating the device's performance directly, not comparing human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This device is a physical surgical system (magnetic grasper and controller), not an algorithm or AI software. Therefore, the concept of "standalone (algorithm only)" performance is not applicable. The device's performance inherently involves human surgeons operating it.

    7. The Type of Ground Truth Used

    For the clinical study, the ground truth was established by clinical observation and assessment by the operating surgeons based on predefined safety and performance endpoints. This can be categorized as expert consensus/clinical judgment from the treating surgeons. The document states:

    • "The study results met all predefined safety and performance endpoints."
    • "The MSS was able to grasp the pillar of the diaphragm or peripillar tissue to adequately mobilize the liver, achieving an effective exposure of the target tissue."

    8. The Sample Size for the Training Set

    The document describes a physical medical device and primarily focuses on its hardware components and clinical performance. There is no mention of a "training set" in the context of an algorithm or AI model development. Therefore, this information is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a "training set" for an algorithm, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223673

    Validate with FDA (Live)

    Date Cleared
    2023-08-04

    (240 days)

    Product Code
    Regulation Number
    878.4815
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    : K223673

    Trade/Device Name: Levita Magnetics Dual Robotic Arm Accessory Regulation Number: 21 CFR 878.4815
    System |
    | Classification Name: | Magnetic Surgical Instrument System (21CFR 878.4815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surgeon Controlled Arm is indicated for use to hold and position a rigid laparoscope/endoscope and the Magnetic Controller of the Levita Magnetic Surgical System in minimally invasive interventions where the Levita Magnetic Surgical System is indicated for use.

    Device Description

    The Surgeon Controlled Arm for the Levita Magnetic Surgical System (MSS) comprises two collaborative surgeon-controlled arms, which along with their controllers, control software, and custom attachments, are intended to hold and control the Magnetic Controller of the commercially available MSS (most recently cleared under K191762) using an arm and to hold and control a commercially available endoscope with camera using the second arm. The design of the surgeon-controlled arms, referred to as the Scope Arm and the Magnet Arm, includes several redundant safety features that make them well suited for use with the MSS in the operating room environment, including hardware and software safety controls.

    When mounted on the Scope Arm, an endoscope can be pivoted and translated as controlled by the user, while maintaining minimal motion at the point where the endoscope enters the patient's body via a standard trocar. Similarly, when mounted on the Magnet Arm, the Magnetic Controller can be moved to different positions on or away from the abdominal wall.

    The Surgeon Controlled Arm includes the following components:

    • (1) Two Carts, which are the Scope Cart and the Magnet Cart. Each Cart includes a surgeoncontrolled arm mounted on top of a moveable cart with custom attachments. One arm functions as a Scope Arm to hold a commercially available endoscope with endoscopic camera. The other arm functions as a Magnet Arm and includes a Magnet Holder to hold the Magnetic Controller of the MSS;
    • (2) a Scope Holder for attachment and detachment of a compatible commercially-available endoscope to the end of the Scope Arm:
    • (3) a Foot Controller for foot-actuated selection of the active arm and foot-actuated remote control for one of the arms at a time;
    • (4) custom-programmed software, the Levita Custom Control Software Program.

    The Surgeon Controlled Arm provides for computer control of a dual- arm device with motorized axes to hold and control the movements of a rigid endoscope and the Magnetic Controller of the Levita MSS during the minimally-invasive procedures cleared for the MSS (most recently cleared under K191762). The user selects the active arm and controls its movement using a Foot Controller. Only one arm is enabled for computer-controlled movement at any given time; while one arm is being moved, the other arm remains stationary. The user may also manually move the arms (i.e. hand guiding) at any time without the use of the Foot Controller.

    For the Scope Arm, the reference point for scope movement is established as the Trocar Point and is set by the user, who positions the endoscope near the trocar inserted in the body opening (e.g. the abdominal fascial layer) and presses the Teach Trocar button on the Scope Adaptor. The Custom Control Software Program then calculates the required individual movements of the Scope Arm motorized axes relative to the Trocar Point in order to achieve the desired total movement of the endoscope with respect to the endoscope view. Once the Custom Control Software Program completes the Teach Trocar procedure, the Foot Controller is enabled, allowing the user in a standing position by the operating table to direct endoscope movement by using their foot to press on switches corresponding to the allowed movement directions.

    For the Magnet Arm, the same arm hardware, software, and Foot Controller are employed to move the Magnetic Controller also in the same directions.

    AI/ML Overview

    Based on the provided text, the device in question is the "Levita Magnetics Dual Robotic Arm Accessory" (also referred to as "Surgeon Controlled Arm"). The document is a 510(k) summary for FDA clearance. It describes the device, its intended use, comparison to a predicate device, and performance data submitted to demonstrate safety and effectiveness.

    Here's an analysis of the acceptance criteria and study information, addressing your points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with specific quantitative targets for each test. Instead, it describes various performance tests conducted and generally states that the device "meets functional requirements" or that tests were "performed successfully."

    I will infer "acceptance criteria" as the general successful completion of the tests listed, and "reported device performance" as the documented successful outcome of these tests.

    Test CategoryImplied Acceptance CriteriaReported Device Performance
    Electrical Safety & EMCCompliance with IEC 60601-1 and IEC 60601-1-2.Performed to applicable requirements.
    Software Verification & ValidationCompliance with FDA guidance "Software Contained in Medical Devices" (Moderate level of concern); Compliance with IEC 62366 and IEC 62304.Conducted and documentation provided as recommended; DRAA complies with IEC 62366 and IEC 62304.
    Bench/Mechanical Testing:Device meets functional requirements for specific parameters.Conducted to verify the device meets functional requirements.
    - Cart stabilityStable operation.(Implied success, no specific data provided)
    - Strength of Magnet Holder attachmentSecure attachment.(Implied success, no specific data provided)
    - Release force Magnet Holder latchingAppropriate release force.(Implied success, no specific data provided)
    - Magnetic field strengthWithin acceptable limits.(Implied success, no specific data provided)
    - Force to engage/disengage Scope HolderAppropriate force.(Implied success, no specific data provided)
    - Strength of Scope Holder engagementSecure engagement.(Implied success, no specific data provided)
    - Actuation force Scope Holder lockingAppropriate actuation force.(Implied success, no specific data provided)
    - Strength of Scope Holder lockingSecure locking.(Implied success, no specific data provided)
    - Durability of attachmentsDurable performance.(Implied success, no specific data provided)
    - Forces to activate switches & pedalsAppropriate activation forces.(Implied success, no specific data provided)
    - Tensile strength Foot Controller cableSufficient strength.(Implied success, no specific data provided)
    - Integrity Foot Controller after dropMaintains integrity.(Implied success, no specific data provided)
    - Max linear & angular velocitiesWithin specified limits.(Implied success, no specific data provided)
    - Max linear & angular accelerationsWithin specified limits.(Implied success, no specific data provided)
    - Latency of arm motionWithin specified limits.(Implied success, no specific data provided)
    - Range of arm motionWithin specified limits.(Implied success, no specific data provided)
    - Safety limits of arm motionAdherence to safety limits.(Implied success, no specific data provided)
    Simulated Use/Design ValidationDevice specifications conform to intended use and user needs.Conducted by three surgeons, representing intended users, implying successful conformity.
    Cybersecurity Risk AssessmentCompliance with FDA guidance "Management of Cybersecurity in Medical Devices."Conducted in accordance with FDA guidance document.
    Sterilization, Disinfection, CleaningValidated processes in accordance with FDA guidance.Completed in accordance with FDA guidance document.
    Human Factors TestingSuccessful testing with intended users; no unexpected device-related events.Conducted with 30 testers representing intended users; performed successfully; no unexpected device-related events.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • For Simulated Use/Design Validation Testing: "three surgeons who perform laparoscopic surgery, representing the intended user." This implies a sample size of 3 for this specific validation.
      • For Human Factors Testing: "30 testers representing intended users." This specifies a sample size of 30 for human factors validation.
      • For other tests (Electrical Safety, Software V&V, Bench/Mechanical, Cybersecurity, Sterilization), the sample size would typically refer to the number of devices or software units tested, which is not specified but implicitly assumed to be sufficient for regulatory compliance.
    • Data Provenance: The document does not explicitly state the country of origin for the data or whether the studies were retrospective or prospective. Given it's an FDA 510(k) submission, the studies would typically be prospective for validation purposes, conducted by the manufacturer or a contracted lab, likely in the US or a country with comparable regulatory standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • For Simulated Use/Design Validation Testing: "three surgeons who perform laparoscopic surgery." Their qualification is that they "perform laparoscopic surgery." No specific experience (e.g., years) is mentioned.
    • For Human Factors Testing: "30 testers representing intended users." The specific qualifications (e.g., medical professionals, years of experience) beyond "intended users" are not detailed.
    • For other tests, ground truth is established via established engineering standards, software testing methodologies, or laboratory validations rather than expert medical consensus.

    4. Adjudication Method for the Test Set

    The document does not describe any formal "adjudication method" (like 2+1 or 3+1 consensus) for the "test set" in the context of expert review.

    • For the Simulated Use/Design Validation Testing with three surgeons, it's implied that their collective assessment determined if the device specifications conformed to intended use and user needs. The document does not specify a disagreement resolution process or voting mechanism for these three surgeons.
    • For Human Factors Testing, the assessment of "successful performance" and "no unexpected device-related events" would be based on the outcomes observed during the testing with the 30 participants, likely against predefined pass/fail criteria from a human factors study protocol. No expert adjudication is mentioned here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described in this 510(k) summary. This type of study is more common for diagnostic AI devices where human performance (e.g., radiologists interpreting images) is directly compared with and without AI assistance on a set of cases. The Levita Magnetics Dual Robotic Arm Accessory is a surgical instrument system, not a diagnostic imaging AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The concept of "standalone" performance, typically applied to AI algorithms, doesn't directly apply here in the same way. The software (Levita Custom Control Software Program) is an integral part of the robotic arm's control system. Its "performance" is evaluated as part of the overall system's functional and safety testing (e.g., software V&V, latency of arm motion, adherence to safety limits), rather than as a standalone diagnostic algorithm. The device, by its nature, is human-in-the-loop (surgeon-controlled).

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluations is based on:

    • Engineering Standards and Specifications: For electrical safety, EMC, and most bench/mechanical tests (e.g., strength, durability, velocities, accelerations, latency, range of motion, safety limits). The "ground truth" is compliance with these predefined technical and safety specifications.
    • FDA Guidance Documents and Regulatory Standards: For software V&V, cybersecurity, sterilization, and human factors. The "ground truth" is adherence to these regulatory requirements and best practices.
    • User Needs and Intended Use: For the simulated use/design validation and human factors testing, the ground truth is whether the device effectively meets user needs and functions as intended in a simulated clinical environment. This is established by observing and evaluating the performance of the device by representative users. There's no "pathology" or "outcomes data" ground truth as this is not a diagnostic device.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" or "training data." This is because the device is a hardware-based robotic surgical assist system with integrated control software, not a machine learning or AI algorithm that requires a distinct training phase. The software is custom-programmed, not "trained" in the machine learning sense from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" in the machine learning context, this question is not applicable. The software's functionality is based on its design specifications, algorithms, and logical programming, verified through traditional software engineering V&V processes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191762

    Validate with FDA (Live)

    Date Cleared
    2020-04-01

    (275 days)

    Product Code
    Regulation Number
    878.4815
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 94010

    Re: K191762

    Trade/Device Name: Magnetic Surgical System Regulation Number: 21 CFR 878.4815
    California 94010

    Re: K191762

    Trade/Device Name: Magnetic Surgical System Regulation Number: 21 CFR 878.4815
    Information:

    Trade Name: Levita Magnetic Surgical System Common Name: Magnetic Surgical System 21 CFR 878.4815
    Controls:

    The Levita Magnetic Surgical System is subject to the special controls described in §21 CFR 878.4815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Levita Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures; the liver in bariatric procedures; the prostate and periprostatic tissue in prostatectomy procedures; and the colon, rectum, and pericolorectal procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients with a BMI range of 20-60-kg/m².

    Device Description

    The Levita Magnetic Surgical System is composed of two hand-held instruments, the Magnetic Grasper and external Magnetic Controller. The Magnetic Grasper (sterile, single use), comprised of a distal detachable Grasper Tip attached to a Shaft with handle, is actuated via its pistol-grip handle with two distinct scissor-type motions to open and close the detachable Grasper Tip jaws. Once the Magnetic Grasper is inserted through a compatible ≥ 10 mm laparoscopic port and the Grasper Tip is attached to the desired tissue, the Grasper Tip can be detached from the Shaft and controlled externally using the Magnetic Controller. Traction of the tissue is maintained through the magnetic field attraction between the Grasper Tip and the Magnetic Controller. The Magnetic Controller (non-sterile, reusable) is a single unit with handles that is held external to the body and emits a magnetic field that attracts the Grasper Tip. Once the Grasper Tip is attached to the desired tissue and detached from the Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Grasper Tip to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Grasper Tip will modulate the magnetic attraction used for tissue retraction/mobilization. If desired, the user can connect the Magnetic Controller to a commercially available surgical support arm that is compatible with its arm mount.

    AI/ML Overview

    The Levita Magnetic Surgical System is a medical device designed to grasp and retract soft tissues and organs during surgical procedures. The device, which consists of a Magnetic Grasper and an external Magnetic Controller, received an administrative change related to its previous substantial equivalence (SE) determination from the FDA on April 15, 2020. The change was an update to its indications for use, expanding its application to colorectal procedures.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Special Control - 21 CFR 878.4815)Reported Device Performance
    (1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize or manipulate soft tissue and organs. (Specifically, for the new indication: colorectal procedures, including grasping and retracting the colon, pericolorectal tissue and adjacent organs)Met: An animal study in a porcine model demonstrated the safety and effectiveness of the Levita Magnetic Surgical System (MSS) in simulated colorectal procedures. The results met all predetermined acceptance criteria: - The MSS provided adequate retraction of the colon, peri-colorectal tissues, and adjacent organs to achieve effective access and visualization of the target tissue. - No device-related serious or severe adverse events were reported. - There was no post-excision or intraoperative tissue trauma. - Surgeons reported that the MSS was the same or better than conventional laparoscopic graspers for use in the subject procedures. The study concluded that the MSS is safe and effective in colorectal procedures while using one fewer access port (trocar), supporting the risk assessment that there are no new or different risks.
    (2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use. (a) Magnetic field strength testing characterization; (b) Ability of the internal surgical instrument(s) to be coupled, de-coupled, and re-coupled.Met (by previous submission): The modifications for the current submission did not change the magnetic field strength or the coupling mechanism. Therefore, previously conducted non-clinical performance testing, which demonstrated that the device performs as intended under anticipated conditions, was deemed applicable.
    (3) The patient-contacting components of the device must be demonstrated to be biocompatible.Met (by previous submission): The modifications did not change any aspect of the device relevant to biocompatibility. Therefore, previously conducted testing demonstrating biocompatibility was deemed applicable.
    (4) Performance data must demonstrate the sterility of the device components that are patient-contacting.Met (by previous submission): The modifications did not change any aspect of the device relevant to sterility. Therefore, previously conducted testing demonstrating sterility was deemed applicable.
    (5) Methods and instructions for reprocessing reusable components must be validated.Met (by previous submission): The modifications did not change any aspect of the device relevant to reprocessing. Therefore, previously conducted testing validating reprocessing methods and instructions was deemed applicable.
    (6) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components and device functionality over the labeled shelf life.Met (by previous submission): The modifications did not change any aspect of the device relevant to shelf life. Therefore, previously conducted testing validating shelf life was deemed applicable.
    (7) Training must be developed and validated by human factors testing and analysis to ensure users can follow the instructions for use to allow safe use of the device.Met (by previous submission): A usability study to validate the training program was previously conducted for the predicate device. Changes to the training program for the updated Indications for Use did not require a repeat of the previous usability study.
    (8) Labeling must include: (a) Magnetic field safezones; (b) Instructions for proper device use; (c) Screening checklist; (d) Reprocessing instructions; (e) Shelf life; (f) Use life.Met: The labeling complies with all the stated special controls.

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: The document specifies an "animal study... in the porcine model." It does not explicitly state the number of animals used in the test set.
    • Data Provenance: The study was an "animal study," specifically in a "porcine model." The country of origin of the data is not specified, nor is whether it was prospective or retrospective, though animal studies for device approval are typically prospective.

    3. Number of Experts and Qualifications to Establish Ground Truth for the Test Set

    • Number of Experts: The document states, "Surgeons reported that the MSS was the same or better than conventional laparoscopic graspers." It does not specify the exact number of surgeons (experts) involved in this assessment.
    • Qualifications of Experts: The experts are referred to as "Surgeons." No further details on their specific qualifications (e.g., years of experience, subspecialty) are provided in the provided text.

    4. Adjudication Method for the Test Set

    • The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The assessment seems to be based on the reports of "surgeons."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study is mentioned in the provided text. The study described is an animal study evaluating the device's performance, not a comparison of human readers with and without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • The Levita Magnetic Surgical System is a physical surgical instrument, not an AI algorithm. Therefore, a standalone (algorithm only) performance study is not applicable and was not performed. The performance evaluation focuses on the instrument's ability to manipulate tissue.

    7. Type of Ground Truth Used

    • For the in vivo performance, the ground truth was established by direct observation and assessment of the device's performance during simulated surgical procedures in live animals (porcine model). This included:
      • Direct visualization of access and visualization of the target tissue.
      • Monitoring for device-related serious or severe adverse events.
      • Pathological assessment for post-excision or intraoperative tissue trauma.
      • Subjective feedback from "Surgeons" regarding the device's effectiveness compared to conventional instruments.

    8. Sample Size for the Training Set

    • The supplied text does not refer to a "training set" in the context of machine learning or AI models, as the Levita Magnetic Surgical System is a physical surgical device. Therefore, this question is not applicable.

    9. How Ground Truth for the Training Set Was Established

    • As the device is a physical surgical system and not an AI/ML product, the concept of a "training set" and establishing ground truth for it does not apply.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190006

    Validate with FDA (Live)

    Date Cleared
    2019-04-29

    (117 days)

    Product Code
    Regulation Number
    878.4815
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    94010

    Re: K190006

    Trade/Device Name: Levita Magnetic Surgical System Regulation Number: 21 CFR§ 878.4815
    -----------|
    | Common Name: | Magnetic Surgical System |
    | Regulation: | 21 CFR 878.4815
    Special Controls:

    The Magnetic Surgical System is subject to the special controls described in §21 CFR 878.4815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Levita Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures, the liver in bariatric procedures, and the prostatic tissue in prostatectomy procedures to facilitate access and visualization of the surgical site. The in patients with a BMI range of 20-60 kg/m2.

    Device Description

    The Levita Magnetic Surgical System is composed of two hand-held instruments, the Magnetic Grasper and external Magnetic Controller. The Magnetic Grasper (sterile, single use), comprised of a distal detachable Grasper Tip attached to a Shaft with handle, is actuated via its pistol-grip handle with two distinct scissor-type motions to open and close the detachable Grasper Tip jaws. Once the Magnetic Grasper is inserted through a compatible The Magnetic Controller (non-sterile, reusable) is a single unit with handles that is held external to the body and emits a magnetic field that attracts the Grasper Tip. Once the Grasper Tip is attached to the desired tissue and detached from the Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Grasper Tip to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Grasper Tip will modulate the magnetic attraction used for tissue retraction/mobilization. If desired, the user can connect the Magnetic Controller to a commercially available surgical support arm that is compatible with its arm mount.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the Levita Magnetic Surgical System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Special Controls - §21 CFR 878.4815)Reported Device Performance
    (1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize or manipulate soft tissue and organs.Effectiveness: The Magnetic Surgical System (MSS) was able to adequately retract the prostate to achieve an effective exposure of target tissue. It was not necessary to replace the MSS with another instrument to retract the prostate during the procedure. Safety: No cases of severe or serious adverse events and no device-related adverse events were reported. Eleven (11) adverse events occurred, all resolved with no clinical sequelae and were categorized as not related to the subject device. Procedure Times: Mean overall procedure time was 3 hours 23 minutes.
    (2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (a) Magnetic field strength testing characterization to identify the distances from the magnet that are safe for patients and users with ferromagnetic implants, devices or objects. (b) Ability of the internal surgical instrument(s) to be coupled, de-coupled, and recoupled with the external magnet over the external magnet use life.No new testing was required. The modifications did not change magnetic field strength or coupling. Previously conducted nonclinical performance testing for the predicate device applied.
    (3) The patient-contacting components of the device must be demonstrated to be biocompatible.No new testing was required. The modifications did not change any aspect of the device. Previously conducted biocompatibility testing for the predicate device applied.
    (4) Performance data must demonstrate the sterility of the device components that are patient-contacting.No new testing was required. The modifications did not change any aspect of the device. Previously conducted sterility testing for the predicate device applied.
    (5) Methods and instructions for reprocessing reusable components must be validated.No new testing was required. The modifications did not change any aspect of the device. Previously conducted testing validating reprocessing methods for the predicate device applied.
    (6) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components and device functionality over the labeled shelf life.No new testing was required. The modifications did not change any aspect of the device. Previously conducted testing validating shelf life for the predicate device applied.
    (7) Training must be developed and validated by human factors testing and analysis to ensure users can follow the instructions for use to allow safe use of the device.A training program was developed, and a usability study was completed to validate it.
    (8) Labeling must include: (a) Magnetic field safezones. (b) Instructions for proper device use. (c) A screening checklist to ensure that all patients and operating staff are screened from bringing ferromagnetic implants, devices or objects near the external magnet. (d) Reprocessing instructions for any reusable components. (e) Shelf life. (f) Use life.The labeling complies with these special controls.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size (Test Set): 30 subjects
    • Data Provenance:
      • Country of Origin: Santiago, Chile (Fundacion Arturo Lopez Perez)
      • Retrospective or Prospective: Prospective clinical study

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Number of Experts: 5 investigators (surgeons)
    • Qualifications: The document identifies them as "surgeons" and "investigators" participating in a clinical study for prostatectomy procedures. Specific experience levels (e.g., years of experience or specialization) are not detailed in this document.

    4. Adjudication Method for the Test Set:

    • The document states that "Investigators reported that adequate retraction was achieved since it was not necessary to replace the MSS with another instrument to retract the prostate during the procedure." and "All eleven events resolved with no clinical sequelae and were categorized as not related to the subject device."
    • This suggests that the surgeons themselves were the primary adjudicators of effectiveness and relatedness of adverse events during the procedure. There is no mention of an independent adjudication panel or a numerical adjudication rule (e.g., 2+1, 3+1). The study describes itself as a "single-arm, open label study."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a MRMC comparative effectiveness study was not done.
    • This was a single-arm study evaluating the device's performance in a new indication (prostatectomy), primarily to demonstrate safety and effectiveness for that specific use extension compared to the predicate device, not to show human reader improvement with AI assistance. The device itself is a surgical instrument, not an AI diagnostic tool.

    6. Standalone (Algorithm Only) Performance:

    • Not applicable. This device is a surgical instrument, not an algorithm, and thus does not have "standalone" algorithm performance in the typical sense of AI/diagnostic devices. Its performance is intrinsically linked to its use by a human surgeon.

    7. Type of Ground Truth Used:

    • Expert Consensus/Clinical Observation during Surgery: The "ground truth" for effectiveness was established by the clinical judgment of the participating surgeons during the procedure. Specifically, effectiveness was defined as the "ability to adequately retract the prostate to achieve an effective exposure of the target tissue," and "Adequate retraction was deemed to be achieved if it is not necessary to replace the Levita Magnetic Surgical System with another instrument to retract the prostate during the procedure." Safety was based on the absence of severe or serious adverse events reported by the investigators.

    8. Sample Size for the Training Set:

    • Not applicable / No specific training set mentioned. This is a medical device, not an AI model, so there isn't a "training set" in the machine learning sense. The device's design and previous versions (predicate devices) would have undergone development and testing, but a distinct "training set" for an algorithm is not relevant here.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As a surgical device and not an AI model, the concept of establishing ground truth for a training set does not apply. The development would have involved engineering, material science, and prior clinical experience with similar devices, leading to verification and validation activities. The clinical study described in the document is for validation of its performance in an extended indication, not for training an algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180894

    Validate with FDA (Live)

    Date Cleared
    2018-09-19

    (167 days)

    Product Code
    Regulation Number
    878.4815
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 94010

    Re: K180894

    Trade/Device Name: Levita Magnetic Surgical System Regulation Number: 21 CFR 878.4815
    Common Name:
    Regulation:
    Special Controls:

    The Magnetic Surgical System is subject to the special controls described in §21 CFR 878.4815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures and the liver in bariatric procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients with a BMI range of 20-60 kg/m².

    Device Description

    The Levita Magnetic Surgical System is composed of two hand-held instruments, the Magnetic Grasper Device and external Magnetic Controller. The Magnetic Grasper Device (sterile, single use), comprised of a distal Detachable Grasper attached to a Delivery/Retrieval shaft with Handle, is actuated via its pistol-grip Handle with two distinct scissor-type motions to open and close the Detachable Grasper jaws. Once the Magnetic Grasper Device is inserted through a compatible ≥ 10 mm laparoscopic port and the Detachable Grasper is attached to the desired tissue, the Detachable Grasper can be detached from the Delivery/Retrieval Shaft and controlled externally using the Magnetic Controller. Traction of the tissue is maintained through the magnetic field attraction between the Detachable Grasper and the Magnetic Controller. The Magnetic Controller (non-sterile, reusable) is a single unit with handles that is held external to the body and emits a magnetic field that attracts the Detachable Grasper. Once the Detachable Grasper is attached to the desired tissue and detached from the Delivery/Retrieval Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Detachable Grasper to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Detachable Grasper will modulate the magnetic attraction used for tissue retraction/mobilization. If desired, the user can connect the Magnetic Controller to a commercially available surgical support arm that is compatible with its arm mount.

    AI/ML Overview

    This document is a 510(k) summary for the Levita Magnetic Surgical System, focusing on a modification to its Indications for Use. The core of the submission demonstrates that the modified device remains substantially equivalent to its predicate.

    Here's an analysis of the provided information regarding acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the "Special Controls" for the Magnetic Surgical System (21 CFR 878.4815), which mandate specific performance data. The device performance is then reported against these special controls.

    Acceptance Criterion (Special Control)Reported Device Performance
    (1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize or manipulate soft tissue and organs.Met: Clinical performance data from two separate studies totaling 103 subjects (73 retrospective, 30 prospective) demonstrated successful use for liver retraction in all cases. - No severe or serious adverse events related to the device. - No device failure or malfunctions. - Adequately grasped and retracted the liver for effective exposure. - Not necessary to use another instrument for retraction. - Overall procedure times similar to published averages. - All cases performed with a reduced port technique, avoiding an incision for a liver retractor.
    (2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use, specifically: (a) Magnetic field strength characterization and (b) Ability of internal surgical instrument(s) to couple, de-couple, and re-couple.Met: No technological changes to the device that would affect magnetic field strength or coupling. Previously conducted non-clinical performance testing applies.
    (3) Patient-contacting components must be demonstrated to be biocompatible.Met: No changes to the device's patient-contacting components. Previously conducted biocompatibility testing applies.
    (4) Performance data must demonstrate the sterility of patient-contacting device components.Met: No changes to the device that would affect sterility. Previously conducted sterility testing applies.
    (5) Methods and instructions for reprocessing reusable components must be validated.Met: No changes to the device that would affect reprocessing. Previously conducted validation of reprocessing methods and instructions applies.
    (6) Performance data must support shelf life by demonstrating continued sterility and functionality.Met: No changes to the device that would affect shelf life. Previously conducted shelf life testing applies.
    (7) Training must be developed and validated by human factors testing and analysis to ensure users can follow instructions for safe use.Met: The training program was updated for the revised Indications for Use (liver retraction in bariatric patients with higher BMI). A new human factors validation was deemed unnecessary because: 1) device design is unchanged, 2) no new use problems identified, 3) use-related hazards and mitigations are unchanged, and 4) critical tasks and errors remain the same. This was supported by clinical performance data from the two studies where investigators had used the updated training program, and the studies met safety and effectiveness endpoints. Previously conducted human factors testing applies.
    (8) Labeling Requirements (Magnetic field safezones, Instructions for proper use, Screening checklist, Reprocessing instructions, Shelf life, Use life).Met: The labeling complies with the stated special controls.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Total Test Set Sample Size: 103 subjects
      • Retrospective Study: 73 subjects
        • Country of Origin: Not explicitly stated, but the study was conducted at "Duke Center for Metabolic and Weight Loss Surgery" in the US.
        • Data Provenance: Retrospective (patients between October 2016 and August 2017).
      • Prospective Study: 30 subjects
        • Country of Origin: Santiago, Chile (Hospital La Florida).
        • Data Provenance: Prospective, single-arm, open-label study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The ground truth in this context is the successful clinical outcome and safe use of the device in surgical procedures. This was directly observed and reported by operating surgeons.

    • Retrospective Study: 10 participating surgeons. No specific qualifications beyond "surgeon" are provided.
    • Prospective Study: 3 participating surgeons. No specific qualifications beyond "surgeon" are provided.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method for the test set in the traditional sense of multiple independent readers evaluating images or data points and then adjudicating discrepancies. The "ground truth" was established by the performing surgeons themselves, who assessed the device's ability to adequately grasp and retract tissue and reported on safety outcomes (adverse events, device failures). Clinical outcome was the primary measure, not subjective reader interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The study evaluated the device's safety and effectiveness in a clinical setting, not how human readers' diagnostic performance might improve with or without AI assistance. This device is a surgical instrument, not an AI diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The Levita Magnetic Surgical System is a physical surgical instrument, not a software algorithm or AI. Its performance is intrinsically linked to its use by a human surgeon. The "standalone" concept applies to AI algorithms that can process data and make inferences without direct human intervention in the interpretation phase.

    7. The Type of Ground Truth Used

    The ground truth used was clinical outcome data and observed intraoperative performance, as documented by treating physicians (surgeons). This included:

    • Successful completion of the bariatric surgery using the device.
    • Adequate liver retraction and visualization of the surgical site.
    • Absence of the need to use another instrument for liver retraction.
    • Occurrence and type of adverse events (e.g., petechiae, liver capsule abrasions) and their resolution.
    • Device failure or malfunctions.
    • Procedure times.

    8. The Sample Size for the Training Set

    No "training set" in the context of machine learning is mentioned or applicable here. This device is a physical medical instrument, not an AI/ML algorithm that requires a data training set. The clinical studies represent performance validation, not data input for model training.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there was no AI/ML training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171429

    Validate with FDA (Live)

    Date Cleared
    2017-06-06

    (22 days)

    Product Code
    Regulation Number
    878.4815
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 94010

    Re: K171429

    Trade/Device Name: Magnetic Surgical System Regulation Number: 21 CFR 21 CFR §878.4815
    Magnetic Surgical Instrument System |
    | | (21 CFR §878.4815
    SPECIAL CONTROLS

    The Magnetic Surgical System is subject to the special controls described in §21 CFR 878.4815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients within a BMI range of 20 to 34 kg/m².

    Device Description

    The Magnetic Surgical System is composed of two hand-held instruments: a Magnetic Grasper Device and an external Magnetic Controller. The Magnetic Grasper Device is comprised of two main components: a Detachable Grasper and a Delivery/Retrieval Shaft. Once the Magnetic Grasper Device is inserted and the Grasper is attached to the desired tissue, the Detachable Grasper can be detached from the Delivery/Retrieval Shaft and controlled externally using the Magnetic Controller. Traction of the tissue is maintained through the magnetic field attraction between the Grasper and the Magnetic Controller. The Magnetic Grasper Device is compatible with a ≥ 10 mm laparoscopic port. The Magnetic Controller is a single unit with handles that is held external to the body and emits a magnetic field that attracts the Detachable Grasper. Once the Detachable Grasper is attached to the desired tissue and detached from the Delivery/Retrieval Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Grasper to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Detachable Grasper will modulate the magnetic attraction used for tissue retraction/manipulation.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Levita Magnetics International Corp. Magnetic Surgical System (K171429). This submission is a Special 510(k), meaning it addresses a modification to an already cleared device (predicate device DEN150007). In this specific case, the modification is a labeling change to the Instructions for Use regarding the use life statement.

    Therefore, the document explicitly states that no new performance testing was conducted for this submission, as the modification does not change the performance characteristics of the device. The substantial equivalence is based on the previously conducted testing for the predicate device.

    Given this, I cannot provide information on the acceptance criteria, reported device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, or standalone performance for this specific 510(k) submission, as no such new studies were part of this filing. The document refers to prior testing for the original clearance of the predicate device.

    However, the document lists the Special Controls that were applicable to the original device and implicitly served as general acceptance criteria for its performance. These controls and the claim that the device meets them (based on prior testing) are outlined below.


    Acceptance Criteria for the Magnetic Surgical System (Based on Special Controls for the original device DEN150007 and claimed continued compliance for K171429)

    Since this 510(k) is for a labeling change and no new performance testing was conducted, there are no new acceptance criteria or reported performance unique to this submission. The Special Controls described in 21 CFR §878.4815 for the Magnetic Surgical System serve as the foundational acceptance criteria that the device (including its substantial equivalent K171429) must meet. The submission claims that previous testing (for DEN150007) demonstrated compliance with these controls.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from 21 CFR §878.4815 Special Controls)Reported Device Performance (as stated in K171429, referring to prior testing for predicate DEN150007)
    (1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize or manipulate soft tissue and organs."The modifications that are the subject of this submission do not change the performance characteristics of the device. Therefore, previously conducted in vivo performance testing that demonstrated that the device performs as intended under the anticipated conditions of use applies here."
    (2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (a) Magnetic field strength testing characterization to identify the distances from the magnet that are safe for patients and users with ferromagnetic implants, devices or objects. (b) Ability of the internal surgical instrument(s) to be coupled, and recoupled with the external magnet over the external magnet use life."The modifications that are the subject of this submission do not change the performance characteristics of the device. Therefore, previously conducted nonclinical performance testing that demonstrated that the device performs as intended under the anticipated conditions of use applies here."
    (3) The patient-contacting components of the device must be demonstrated to be biocompatible."The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that demonstrated that the device is biocompatible applies here."
    (4) Performance data must demonstrate the sterility of the device components that are patient contacting."The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that demonstrated the sterility of the device applies here."
    (5) Methods and instructions for reprocessing reusable components must be validated."The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that validated the methods and instructions for reprocessing reusable components applies here."
    (6) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components and device functionality over the labeled shelf life."The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that validated the shelf life of the device applies here."
    (7) Training must be developed and validated by human factors testing and analysis to ensure users can follow the instructions for use to allow safe use of the device."The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted human factors testing which validated that users can follow the instructions to allow safe use of the device applies here."
    (8) Labeling must include: (a) Magnetic field safe zones. (b) Instructions for proper device use. (c) A screening checklist to ensure that all patients and operating staff are screened from bringing ferromagnetic implants, devices or objects near the external magnet. (d) Reprocessing instructions for any reusable components. (e) Shelf life. (f) Use life."The labeling complies with the special controls stated above." (Note: This specific 510(k) is about a labeling change related to use life, affirming compliance with this aspect.)

    For the following points, as this 510(k) exclusively concerns a labeling change and refers to prior testing, the specific details are not available in this document. The information below reflects that no new studies were conducted for K171429.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable to this 510(k) submission, as no new test set was used. The submission relies on prior testing for the predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable to this 510(k) submission, as no new test set requiring expert ground truth was used.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable to this 510(k) submission, as no new test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable; this device is a magnetic surgical system, not an AI-assisted diagnostic tool involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable; this device is a physical surgical system, not an algorithm. Its performance is inherent to its mechanical and magnetic properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the original predicate device's in vivo testing, presumed ground truth would likely have been direct surgical observation and potentially histological assessment of tissue integrity or adverse events. For non-clinical tests, ground truth would be established by physical measurements and engineering standards. However, these details are not provided in this document.

    8. The sample size for the training set

    Not applicable; this physical device does not utilize a training set in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not applicable; this physical device does not utilize a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN150007

    Validate with FDA (Live)

    Date Cleared
    2016-06-13

    (489 days)

    Product Code
    Regulation Number
    878.4815
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    NEW REGULATION NUMBER: 21 CFR 878.4815

    CLASSIFICATION: II

    PRODUCT CODE: PNL

    BACKGROUND

    DEVICE
    following:

    Product Code: PNL Device Type: Magnetic Surgical Instrument System Class: II Regulation: 21 CFR 878.4815

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Levita Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients within a BMI range of 20 to 34 kg/m2.

    Device Description

    The Levita Magnetic Surgical System is composed of two hand-held instruments, the Magnetic Grasper Device and external Magnetic Controller, which are intended to facilitate tissue grasping, retraction and mobilization during laparoscopic cholecystectomy procedures.

    1. Magnetic Grasper Device (sterile, single use) is actuated via its pistol-grip handle with two distinct scissor-type motions to open and close the Detachable Grasper jaws as shown in Figure 1. At the distal end is the Detachable Grasper, which is attached to the Delivery/Retrieval Shaft of the full Magnetic Grasper Device. This device requires a >10mm access port to introduce the device into the abdominal cavity. The Detachable Grasper is then coupled (held in place within the abdomen) by the external Magnetic Controller. If the surgeon wishes to use a single 10-12 mm port at the umbilicus such as a Hasson port and that is the port shared by the optic, applying and adjusting the Detachable Grasper may require repetitive placement of the optic in an alternative port. The Detachable Grasper is shown in Figure 2.

    2. An External Magnetic Controller (non-sterile, reusable) 3-inch diameter by 2-inch thick b(4) disk magnet, shown in Figure 3. This component holds the Detachable Grasper (magnetically) and is placed on the external abdomen wall.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device: Levita Magnetic Surgical System
    Indications for Use: To grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures to facilitate access and visualization of the surgical site, for patients with a BMI range of 20 to 34 kg/m2.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document outlines various tests with acceptance criteria and results. Here's a consolidated table:

    CategorySpecific Test / Performance CharacteristicAcceptance CriteriaReported Device Performance / Results
    BiocompatibilityCytotoxicity (MEM Elution)Test Sample Reactivity ≤ 2PASS
    Sensitization (0.9% NaCl, sesame oil)Test samples < 1 grade dermal reactionPASS
    Irritation/Intracutaneous Reactivity (0.9% NaCl, sesame oil)Control extract overall mean score ≤ 1.0PASS
    Shelf Life/SterilityDetachable Grasper Sterility (Gamma Sterilization)Sterility Assurance Level (SAL) of 10^-6Achieved a SAL of 10^-6 in accordance with ISO 11737-1:2006, ISO 11137-2:2012, and AAMI TIR33:2005.
    Detachable Grasper Shelf-life visual inspection: outer box damageNo visible damagePASS
    Detachable Grasper Shelf-life visual inspection: outer box labelLabel is legiblePASS
    Detachable Grasper Shelf-life visual inspection: pouch damageNo visible damagePASS
    Detachable Grasper Shelf-life visual inspection: pouch labelLabel is legiblePASS (6-month shelf life determined based on accelerated aging)
    EMI/Magnetic FieldSafe Working Distances (Based on ACGIH, ICNIRP, NRC guidelines)Defined safe zones based on measured magnetic field strengthMagnetic field strength characterized at various distances with and without carrying case; safe working zones defined and detailed in labeling (e.g., 50 cm for pacemakers, 20-50 cm for CRT monitors/magnetic data storage).
    Bench PerformanceGrasper Pull-off force (T=0 & T=6 months)b(4) CCI/TS (specific value omitted)PASS for both T=0 and T=6 months.
    Grasper Pull-off force when near Magnetic Controller (T=0 & T=6 months)b(4) CCI/TS (specific value omitted)PASS for both T=0 and T=6 months.
    Grasper Jaw Pinch Force (T=0 & T=6 months)b(4) CCI/TS (specific value omitted)PASS for both T=0 and T=6 months.
    Handle force to open grasper jaws (T=0 & T=6 months)b(4) CCI/TS (specific value omitted)PASS for both T=0 and T=6 months.
    Minimum grasper jaw opening (T=0 & T=6 months)b(4) CCI/TS (specific value omitted)PASS for both T=0 and T=6 months.
    Handle force to release grasper jaws (T=0 & T=6 months)b(4) CCI/TS (specific value omitted)PASS for both T=0 and T=6 months.
    Attraction force to external magnet at a distance of 4 cm (T=0 & T=6 months)b(4) CCI/TS (specific value omitted)PASS for both T=0 and T=6 months.
    Lifetime Cycling Performance (all above tests) after simulated use and actuations over lifetimeSame as all abovePASS
    Magnetic Controller Use Life Testing - Tensile testEncasement can be suspended by the upper edge b(4) without breakagePASS (use life of 25 uses determined)
    Magnetic Controller Use Life Testing - Visual InspectionAfter cycling and suspension, no defects that would compromise integrityPASS
    Animal Study (Safety & Performance)Safety: Absence of clinically relevant trauma to abdominal wall/tissue (gross & microscopic)No evidence of clinically relevant trauma. Minimal observations, defects from 5mm trocar more severe than test site. No significant increases in trauma compared to 5mm trocar.Met. Gross and histopathological findings showed no evidence of clinically relevant trauma.
    Performance: Ability to manipulate gallbladder & compatibility with instrumentation (single-incision cholecystectomy)Magnetic Surgical System usable to manipulate gallbladder, able to complete single-incision laparoscopic cholecystectomy, compatible with equipment/instrumentation.Met. Successfully manipulated gallbladder for several cycles without adverse events; completed a single-incision laparoscopic cholecystectomy; compatible with multiple types of electrosurgical equipment (monopolar, bi-polar, ultrasonic harmonic).
    Human FactorsNo critical failures (Critical Task Errors) and no more than two non-critical failures per subjectNo critical failures and no more than two non-critical failures per subject after training.All 15 surgeons passed all tests without any critical or non-critical errors. (Critical task errors include: user fails to inspect Magnetic Controller, fails to wipe down/cover Magnetic Controller, inserts Grasper without laparoscopic visualization, attempts to retrieve lost Detachable Grasper using Magnetic Controller, fails to retrieve Detachable Grasper, fails to clean/disinfect Magnetic Controller.)
    Clinical Study (Safety)Absence of any damage or side effect to the patient directly produced by the device during surgery (Device Failure and Serious Adverse Events)1. No evidence of Device Failure (breakage/malfunction requiring additional surgical intervention, reoperation, device removal). 2. No Serious Adverse Event probably or definitely related to grasping/tissue manipulation or device (revision/removal, permanent organ damage, death). Overall morbidity rate (any complications related to investigational device within 30 days post-op)Met. No cases of device failure. Three (3) serious adverse events and thirty-five (35) non-serious AEs reported; none were device-related. Twenty (20) device-related events were mild petechiae, considered not clinically significant and resolved. Overall morbidity rate not explicitly quantified but stated as met by absence of serious/device-related AEs.
    Clinical Study (Performance)Ability to adequately mobilize the gallbladder to achieve an effective exposure of the target tissue; feasibility will be a "failure" if an additional trocar is needed to insert another instrument to mobilize the gallbladder.Ability to adequately mobilize the gallbladder to achieve an effective exposure. No additional trocars placed to specifically mobilize the gallbladder.Met. In all cases, the device was able to adequately mobilize the gallbladder. One additional trocar was placed in one procedure, but after Magnetic Grasper use was complete and not to mobilize the gallbladder.

    Study Details:

    2. Sample Size and Data Provenance:

    • Test Set (Clinical Study):

      • Feasibility Study: 21 subjects underwent surgery.
      • Pivotal Study: 50 subjects underwent surgery.
      • Total Subjects for Test Set: 71
      • Data Provenance: Prospective, single arm, open-label studies conducted in Chile (three sites).
    • Animal Study Test Set:

      • 2 canines.
    • Human Factors Test Set:

      • 15 surgeons.

    3. Number of Experts and Qualifications for Ground Truth (Test Set):

    • Clinical Study: Not explicitly stated as "experts for ground truth determination" in the context of imaging review. The study involved surgeons performing the procedures and evaluating performance and safety, and there were also assessments of safety results by independent groups (e.g., AEs reviewed for device relatedness not explicitly stated if by experts). "Clinical evaluator experienced in single-incision laparoscopy" conducted evaluation in the animal study.
    • Animal Study: Safety evaluated grossly by a pathologist and microscopically by an independent histopathologist. Performance evaluated by a clinical evaluator experienced in single-incision laparoscopy.
    • Human Factors Study: The "ground truth" for task completion and errors was established by experienced proctors/observers during the simulated use environment, implicitly serving as experts.

    4. Adjudication Method for the Test Set:

    • Clinical Study: Not explicitly mentioned as a formal adjudication method (e.g., 2+1, 3+1 for imaging). Safety endpoints (device failure, serious adverse events) were defined and assessed. Device-related events (mild petechiae) were "not considered clinically significant by the study investigators."
    • Animal Study: Gross pathological findings by one pathologist, microscopic by an independent histopathologist, and clinical evaluation by an experienced laparoscopist. This implies a form of expert review, though not a multi-reader, consensus-based adjudication as typically seen in image-based AI studies.
    • Human Factors Study: Based on observation of 15 surgeons completing tasks against predefined critical and non-critical error lists during simulated use. The "acceptance criteria were no critical failures... and no more than two non-critical failures per subject." This is a pass/fail assessment based on observed performance against a checklist.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done.
    • The clinical study was a single-arm study, meaning there was no control group of traditional laparoscopic cholecystectomy without the device for direct comparison of outcomes or human reader (surgeon) performance improvement.
    • The document explicitly states: "Since the studies presented were single armed, comparisons could not be made due to lack of a traditional laparoscopic cholecystectomy control." Therefore, no effect size of human readers improving with AI vs. without AI assistance can be reported. The device is a surgical instrument, not an AI for image interpretation.

    6. Standalone (Algorithm only without human-in-the-loop performance):

    • This question is not applicable as the Levita Magnetic Surgical System is a medical device (a surgical instrument), not an AI algorithm. Its performance is intrinsically linked to human use (the surgeon). The "performance testing" described in the document (bench, animal, clinical) evaluates the device's functional integrity and its performance when operated by a human user.

    7. Type of Ground Truth Used:

    • Clinical Study:
      • Safety: Clinical endpoints observed during surgery and post-operative follow-up, assessed by study investigators (e.g., device failures, serious adverse events, morbidity, petechiae).
      • Performance: Surgeon assessment of the ability to mobilize the gallbladder and completion of the cholecystectomy procedure.
      • Patient Outcomes: Patient surveys, operative time, PACU time, length of stay, pain scores, blood loss, etc.
    • Animal Study: Gross pathological findings, microscopic histopathological findings, and clinical evaluation by an experienced surgeon.
    • Human Factors Study: Direct observation of task completion and identification of predefined critical/non-critical errors by trained proctors in a simulated environment.

    8. Sample Size for the Training Set:

    • Not explicitly stated for an AI model. As this is a medical device (surgical instrument), not an AI, there isn't a "training set" in the context of machine learning model development.
    • However, if "training set" refers to the subjects or data used to train the human users of the device:
      • Human Factors Study: 15 surgeons received training materials (Instructions for Use, labels, screening checklist, didactic presentation, hands-on demonstration) before being tested. This served as their "training."

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable in the context of an AI training set for model development.
    • For the human factors training, the "ground truth" for proper use was established by the manufacturer through the development of the Instructions for Use, training presentation, and a pre-defined list of correct (and incorrect) operational steps/critical tasks. This represents the manufacturer's validated understanding of safe and effective device operation.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1