K Number
K191762

Validate with FDA (Live)

Date Cleared
2020-04-01

(275 days)

Product Code
Regulation Number
878.4815
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Levita Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures; the liver in bariatric procedures; the prostate and periprostatic tissue in prostatectomy procedures; and the colon, rectum, and pericolorectal procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients with a BMI range of 20-60-kg/m².

Device Description

The Levita Magnetic Surgical System is composed of two hand-held instruments, the Magnetic Grasper and external Magnetic Controller. The Magnetic Grasper (sterile, single use), comprised of a distal detachable Grasper Tip attached to a Shaft with handle, is actuated via its pistol-grip handle with two distinct scissor-type motions to open and close the detachable Grasper Tip jaws. Once the Magnetic Grasper is inserted through a compatible ≥ 10 mm laparoscopic port and the Grasper Tip is attached to the desired tissue, the Grasper Tip can be detached from the Shaft and controlled externally using the Magnetic Controller. Traction of the tissue is maintained through the magnetic field attraction between the Grasper Tip and the Magnetic Controller. The Magnetic Controller (non-sterile, reusable) is a single unit with handles that is held external to the body and emits a magnetic field that attracts the Grasper Tip. Once the Grasper Tip is attached to the desired tissue and detached from the Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Grasper Tip to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Grasper Tip will modulate the magnetic attraction used for tissue retraction/mobilization. If desired, the user can connect the Magnetic Controller to a commercially available surgical support arm that is compatible with its arm mount.

AI/ML Overview

The Levita Magnetic Surgical System is a medical device designed to grasp and retract soft tissues and organs during surgical procedures. The device, which consists of a Magnetic Grasper and an external Magnetic Controller, received an administrative change related to its previous substantial equivalence (SE) determination from the FDA on April 15, 2020. The change was an update to its indications for use, expanding its application to colorectal procedures.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Special Control - 21 CFR 878.4815)Reported Device Performance
(1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize or manipulate soft tissue and organs. (Specifically, for the new indication: colorectal procedures, including grasping and retracting the colon, pericolorectal tissue and adjacent organs)Met: An animal study in a porcine model demonstrated the safety and effectiveness of the Levita Magnetic Surgical System (MSS) in simulated colorectal procedures. The results met all predetermined acceptance criteria: - The MSS provided adequate retraction of the colon, peri-colorectal tissues, and adjacent organs to achieve effective access and visualization of the target tissue. - No device-related serious or severe adverse events were reported. - There was no post-excision or intraoperative tissue trauma. - Surgeons reported that the MSS was the same or better than conventional laparoscopic graspers for use in the subject procedures. The study concluded that the MSS is safe and effective in colorectal procedures while using one fewer access port (trocar), supporting the risk assessment that there are no new or different risks.
(2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use. (a) Magnetic field strength testing characterization; (b) Ability of the internal surgical instrument(s) to be coupled, de-coupled, and re-coupled.Met (by previous submission): The modifications for the current submission did not change the magnetic field strength or the coupling mechanism. Therefore, previously conducted non-clinical performance testing, which demonstrated that the device performs as intended under anticipated conditions, was deemed applicable.
(3) The patient-contacting components of the device must be demonstrated to be biocompatible.Met (by previous submission): The modifications did not change any aspect of the device relevant to biocompatibility. Therefore, previously conducted testing demonstrating biocompatibility was deemed applicable.
(4) Performance data must demonstrate the sterility of the device components that are patient-contacting.Met (by previous submission): The modifications did not change any aspect of the device relevant to sterility. Therefore, previously conducted testing demonstrating sterility was deemed applicable.
(5) Methods and instructions for reprocessing reusable components must be validated.Met (by previous submission): The modifications did not change any aspect of the device relevant to reprocessing. Therefore, previously conducted testing validating reprocessing methods and instructions was deemed applicable.
(6) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components and device functionality over the labeled shelf life.Met (by previous submission): The modifications did not change any aspect of the device relevant to shelf life. Therefore, previously conducted testing validating shelf life was deemed applicable.
(7) Training must be developed and validated by human factors testing and analysis to ensure users can follow the instructions for use to allow safe use of the device.Met (by previous submission): A usability study to validate the training program was previously conducted for the predicate device. Changes to the training program for the updated Indications for Use did not require a repeat of the previous usability study.
(8) Labeling must include: (a) Magnetic field safezones; (b) Instructions for proper device use; (c) Screening checklist; (d) Reprocessing instructions; (e) Shelf life; (f) Use life.Met: The labeling complies with all the stated special controls.

2. Sample Size and Data Provenance for the Test Set

  • Sample Size: The document specifies an "animal study... in the porcine model." It does not explicitly state the number of animals used in the test set.
  • Data Provenance: The study was an "animal study," specifically in a "porcine model." The country of origin of the data is not specified, nor is whether it was prospective or retrospective, though animal studies for device approval are typically prospective.

3. Number of Experts and Qualifications to Establish Ground Truth for the Test Set

  • Number of Experts: The document states, "Surgeons reported that the MSS was the same or better than conventional laparoscopic graspers." It does not specify the exact number of surgeons (experts) involved in this assessment.
  • Qualifications of Experts: The experts are referred to as "Surgeons." No further details on their specific qualifications (e.g., years of experience, subspecialty) are provided in the provided text.

4. Adjudication Method for the Test Set

  • The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The assessment seems to be based on the reports of "surgeons."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study is mentioned in the provided text. The study described is an animal study evaluating the device's performance, not a comparison of human readers with and without AI assistance.

6. Standalone (Algorithm Only) Performance Study

  • The Levita Magnetic Surgical System is a physical surgical instrument, not an AI algorithm. Therefore, a standalone (algorithm only) performance study is not applicable and was not performed. The performance evaluation focuses on the instrument's ability to manipulate tissue.

7. Type of Ground Truth Used

  • For the in vivo performance, the ground truth was established by direct observation and assessment of the device's performance during simulated surgical procedures in live animals (porcine model). This included:
    • Direct visualization of access and visualization of the target tissue.
    • Monitoring for device-related serious or severe adverse events.
    • Pathological assessment for post-excision or intraoperative tissue trauma.
    • Subjective feedback from "Surgeons" regarding the device's effectiveness compared to conventional instruments.

8. Sample Size for the Training Set

  • The supplied text does not refer to a "training set" in the context of machine learning or AI models, as the Levita Magnetic Surgical System is a physical surgical device. Therefore, this question is not applicable.

9. How Ground Truth for the Training Set Was Established

  • As the device is a physical surgical system and not an AI/ML product, the concept of a "training set" and establishing ground truth for it does not apply.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA. On the right is the logo for the FDA U.S. Food & Drug Administration. The FDA logo has the letters FDA in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font in blue.

April 15, 2020

Levita Magnetics International Corp. % Cindy Domecus Principal, Domecus Consulting Services LLC Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, California 94010

Re: K191762

Trade/Device Name: Magnetic Surgical System Regulation Number: 21 CFR 878.4815 Regulation Name: Magnetic surgical instrument system Regulatory Class: Class II Product Code: PNL

Dear Cindy Domecus:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 1, 2020. Specifically, FDA is updating this SE Letter as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Long Chen, Ph.D., OHT4: Office of Surgical and Infection Control Devices, Assistant Director at 301-796-6389 or Long.Chen@fda.hhs.gov.

Sincerely,

Digitally signed by Long H. Chen
Long H. Chen -S -s
Date: 2020.04.15 13:23:36 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

April 1, 2020

Levita Magnetics International Corp. % Cindy Domecus Regulatory Consultant to Levita Magnetics Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, California 94010

Re: K191762

Trade/Device Name: Magnetic Surgical System Regulation Number: 21 CFR 878.4815 Regulation Name: Magnetic Surgical Instrument System Regulatory Class: Class II Product Code: PNL Dated: February 26, 2020 Received: February 28, 2020

Dear Cindy Domecus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Long H. Chen
Long H. Chen -S-s
Date: 2020.04.01 07:33:30 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K191762

Device Name Levita Magnetic Surgical System

Indications for Use (Describe)

The Levita Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures; the liver in bariatric procedures; the prostate and periprostatic tissue in prostatectomy procedures; and the colon, rectum, and pericolorectal procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients with a BMI range of 20-60-kg/m².

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary is being submitted in accordance with the requirements of 21 CFR 807.92.

510(k) Owner Information:

Levita Magnetics International Corp. 1430 S. Amphlett Blvd, Suite 240 San Mateo, CA 94402 Phone: 650-241-0320 Fax: 650-241-1825

Submission Correspondent:

Cindy Domecus, R.A.C. (US & EU) Principal, Domecus Consulting Services LLC Regulatory Consultant to Levita Magnetics Email: domecusconsulting@comcast.net Phone: (650) 343-4813

Device Information:

Trade Name: Levita Magnetic Surgical System Common Name: Magnetic Surgical System 21 CFR 878.4815 Regulation: Classification Panel: General & Plastic Surgery Device Type: Magnetic Surgical Instrument System Device Class: ll Product Code: PNL

Predicate Devices:

Levita Magnetic Surgical System, K190006

Date Prepared:

April 8, 2020

Device Description:

The Levita Magnetic Surgical System is composed of two hand-held instruments, the Magnetic Grasper and external Magnetic Controller.

The Magnetic Grasper (sterile, single use), comprised of a distal detachable Grasper Tip attached to a Shaft with handle, is actuated via its pistol-grip handle with two distinct scissor-type motions to open and close the detachable Grasper Tip jaws. Once the Magnetic Grasper is inserted through a compatible

Page 1/5

{5}------------------------------------------------

Page 2/5

≥ 10 mm laparoscopic port and the Grasper Tip is attached to the desired tissue, the Grasper Tip can be detached from the Shaft and controlled externally using the Magnetic Controller. Traction of the tissue is maintained through the magnetic field attraction between the Grasper Tip and the Magnetic Controller.

The Magnetic Controller (non-sterile, reusable) is a single unit with handles that is held external to the body and emits a magnetic field that attracts the Grasper Tip. Once the Grasper Tip is attached to the desired tissue and detached from the Shaft, the Magnetic Controller is placed external to the body to magnetically attract the Grasper Tip to manipulate the target tissue. Adjusting the distance between the Magnetic Controller and the Grasper Tip will modulate the magnetic attraction used for tissue retraction/mobilization. If desired, the user can connect the Magnetic Controller to a commercially available surgical support arm that is compatible with its arm mount.

Indications for Use:

The Levita Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures; the liver in bariatric procedures; the prostate and periprostatic tissue in prostatectomy procedures; and the colon, rectum, and pericolorectal tissue in colorectal procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients with a BMI range of 20-60 kg/m².

Comparison of Intended Use, Indications for Use and Technological Characteristics with the Predicate Device:

The subject and the predicate devices are identical in technological characteristics. The subject device differs from the predicate device by only a modifications for Use and the associated changes to the labeling. Results of animal testing demonstrated that the differences in the indications for use do not raise different questions of safety and effectiveness, and, therefore, the subject device has the same intended use as the predicate device.

Subject DevicePredicate Device
Intended UseTo grasp, hold, retract, or manipulate soft tissue and organs.Same
Indications for useThe Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures; the liver in bariatric procedures; the prostate and periprostatic tissue in prostatectomy procedures; and the colon, pericolorectal tissue and adjacent organs in colorectal procedures to facilitate access and visualization of theThe Magnetic Surgical System is designed to grasp and retract the body and the fundus of the gallbladder in laparoscopic cholecystectomy procedures, the liver in bariatric procedures, and the prostate and periprostatic tissue in prostatectomy procedures to facilitate access and visualization of the surgical site. The device is indicated for use in patients with a

Comparison of Intended Use and Indications for Use

{6}------------------------------------------------

Subject DevicePredicate Device
surgical site. The device is indicated foruse in patients with a BMI range of 20-60 kg/m².BMI range of 20-60 kg/m².

The subject and predicate device share identical technological characteristics:

  • . Both devices are composed of two hand-held instruments: a Magnetic Grasper comprised of a detachable Grasper Tip and a Shaft, and an external Magnetic Controller.
  • . The principles of operation remain unchanged. For both devices, the Magnetic Grasper is actuated manually using the handle of the Shaft to grasp and release tissue; and the Grasper Tip is controlled manually by the handle or the external Magnetic Controller to retract tissue.
  • The design, features and materials remain unchanged.

As such, the design, materials and function of the subject Levita Magnetic Surgical System are substantially equivalent to the predicate Levita Magnetic Surgical System.

Performance Data:

There were no technological changes to the subject device, thus no bench, electromagnetic compatibility testing, sterilization testing or biocompatibility testing was required.

Animal performance data demonstrated the subject device is safe and effective in grasping and retracting the colon, pericolorectal tissue and adjacent organs in colorectal procedures.

Special Controls:

The Levita Magnetic Surgical System is subject to the special controls described in §21 CFR 878.4815. These special controls are stated below. Compliance with these requirements has been met as noted in the italicized text below each requirement.

  • (1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize or manipulate soft tissue and organs.
    This special control is addressed by in vivo performance data from an animal study conducted to evaluate the safety and effectiveness of the Levita Magnetic Surgical System (MSS) in simulated colorectal procedures in the porcine model.

The results met all predetermined acceptance criteria. The MSS provided adequate retraction of the colon, peri-colorectal tissues, and adjacent organs to achieve effective access and visualization of the target tissue. No device-related serious or severe adverse events were reported. There was no post-excision or intraoperative tissue trauma. Surgeons reported that the MSS was the same or better than conventional laparoscopic graspers for use in the subject

{7}------------------------------------------------

procedures. The study demonstrated that the MSS is safe and effective in colorectal procedures while using one fewer access port (trocar), supporting the risk assessment that there are no new or different risks. The study demonstrates that the MSS performs as intended under anticipated conditions of use during colorectal procedures and that the MSS is able to grasp and retract the colon, pericolorectal tissue and adjacent organs.

  • (2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
    • (a) Magnetic field strength testing characterization to identify the distances from the magnet that are safe for patients and users with ferromagnetic implants, devices or objects.
    • (b) Ability of the internal surgical instrument(s) to be coupled, de-coupled, and recoupled with the external magnet over the external magnet use life.

The modifications that are the subject of this submission do not change the magnetic field strength or coupling of the internal surgical instrument with the external magnet. Therefore, previously conducted nonclinical performance testing that demonstrated that the device performs as intended under the anticipated conditions of use applies here.

  • (3) The patient-contacting components of the device must be demonstrated to be biocompatible.
    The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that demonstrated that the device is biocompatible applies here.

  • (4) Performance data must demonstrate the sterility of the device components that are patientcontacting.
    The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that demonstrated the sterility of the device applies here.

(5) Methods and instructions for reprocessing reusable components must be validated.

The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that validated the methods and instructions for reprocessing reusable components applies here.

(6) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components and device functionality over the labeled shelf life.

The modifications that are the subject of this submission do not change any aspect of the device. Therefore, previously conducted testing that validated the shelf life of the device applies here.

  • (7) Training must be developed and validated by human factors testing and analysis to ensure users can follow the instructions for use to allow safe use of the device.

{8}------------------------------------------------

A usability study to validate the training program was previously conducted and submitted in support of clearance of the predicate device. None of the changes made to the training program to reflect the change in the subject Indications for Use require a repeat of the previous usability study.

  • (8) Labeling must include:
    • (a) Magnetic field safezones.
    • (b) Instructions for proper device use.
    • (c) A screening checklist to ensure that all patients and operating staff are screened from bringing ferromagnetic implants, devices or objects near the external magnet.
    • (d) Reprocessing instructions for any reusable components.
    • (e) Shelf life.
    • (f) Use life.

The labeling complies with the special controls stated above.

Conclusion:

The Levita Magnetic Surgical System has the same intended use and technological characteristics as the predicate device. Therefore, the Levita Magnetic Surgical System is substantially equivalent to the cleared predicate device.

§ 878.4815 Magnetic surgical instrument system.

(a)
Identification. A magnetic surgical instrument system is a prescription device used in laparoscopic surgical procedures consisting of several components, such as surgical instruments, and a magnetic controller. The magnetic controller is provided separately from the surgical instrument and is used outside the patient. The external magnetic controller is magnetically coupled with the internal surgical instrument(s) at the surgical site to grasp, hold, retract, mobilize, or manipulate soft tissue and organs.(b)
Classification. Class II (special controls). The special controls for this device are:(1) In vivo performance data must demonstrate that the device performs as intended under anticipated conditions of use. Testing must demonstrate the ability of the device to grasp, hold, retract, mobilize, or manipulate soft tissue and organs.
(2) Non-clinical performance data must demonstrate that the system performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Magnetic field strength testing characterization to identify the distances from the magnet that are safe for patients and users with ferromagnetic implants, devices, or objects.
(ii) Ability of the internal surgical instrument(s) to be coupled, de-coupled, and re-coupled with the external magnet over the external magnet use life.
(3) The patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of the device components that are patient-contacting.
(5) Methods and instructions for reprocessing reusable components must be validated.
(6) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components and device functionality over the labeled shelf life.
(7) Training must be developed and validated by human factors testing and analysis to ensure users can follow the instructions for use to allow safe use of the device.
(8) Labeling must include:
(i) Magnetic field safe zones.
(ii) Instructions for proper device use.
(iii) A screening checklist to ensure that all patients and operating staff are screened from bringing ferromagnetic implants, devices, or objects near the external magnet.
(iv) Reprocessing instructions for any reusable components.
(v) Shelf life.
(vi) Use life.