Search Filters

Search Results

Found 330 results

510(k) Data Aggregation

    K Number
    K260147

    Validate with FDA (Live)

    Date Cleared
    2026-03-13

    (52 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: Electric nasal aspirator (BC026, BC025, BC023, BC028, BC029, BC036)
    Regulation Number: 21 CFR 878.4780
    suction pump
    Classification Name: Pump, portable, aspiration (powered)
    Regulation: 21 CFR § 878.4780
    nasal aspirator, model: BC026, BC025, BC023 | / |
    | Classification | Class II Device, BTA (21 CFR § 878.4780
    ) | Class II Device, BTA (21 CFR § 878.4780) | Identical |
    | Classification Panel | General & Plastic

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electric Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.

    Device Description

    Electric nasal aspirator (model: BC026, BC025, BC023, BC028, BC029, BC036) consists of main unit, and suction portion working together as one unit. The Electric nasal aspirator is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions.The motor pump is powered by a rechargeable battery. The rechargeable battery can be charged via an external power adapter(not included in this device) using the provided charging line. The user interface consists of buttons and LED display, and the user can control the air flow level through the button.

    The differences between BC026, BC025, BC023, BC028, BC029 and BC036 are as follows:

    1. The flow rate range of BC036 at different flow levels is different from other models.
    2. BC025 and BC036 do not have "Light" function(used to distract and pacify the child), while other models do.
    3. BC026, BC028, BC029 can pause suction during the suction process, while other models do not.
    4. BC026, BC028, BC029 and BC023 have 4 buttons, BC025 and BC036 have 2 buttons.
    5. In addition, BC026, BC028, BC029, BC025, BC036 and BC023 have different Display, Size, Weight, and Appearance.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253429

    Validate with FDA (Live)

    Date Cleared
    2026-03-12

    (162 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    18 - 150
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    11801

    Re: K253429
    Trade/Device Name: Genadyne ASTRA NPWT
    Regulation Number: 21 CFR 878.4780
    Prepared:** March 12, 2026

    Device Name/ Trade Name: Genadyne ASTRA NPWT
    Regulation Name: 21 CFR 878.4780

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Genadyne Astra NPWT System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device can be used for wound management by the removal of excess exudates, infectious material and tissue debris.

    Appropriate wound type include:

    • Chronic
    • Acute
    • Traumatic
    • Subacute and dehisced wounds
    • Partial thickness burns
    • Ulcers (such as diabetic or pressure)
    • Flaps and grafts

    For use by healthcare professional and home use. This NPWT System can only be used with approved dressing kits under K233614.

    Device Description

    The Genadyne ASTRA NPWT is a Negative Pressure Wound Therapy. The ASTRA NPWT features an interface touch panel which provides user selectable therapy modes (continuous mode and variable mode) and multi pressure options from 40mmHg to 200 mmHg). The ASTRA NPWT is packaged in a box that includes a universal charger, power cable, a user manual and a user carry bag. The dressing kits (approved under K233614) includes the transparent film dressings, the Genadyne foam dressing and the Genadyne Port. The Genadyne ASTRA NPWT is designed to be used with a canister collection system. The canister will have 3 sizes available (200cc, 400cc and 600cc).

    All the dressings and canisters are single use disposable items. To help ensure safe and effective use, the ASTRA NPWT are to be used only with the Genadyne supplied dressings (approved under K233614) and canisters. For our bench test, the ASTRA NPWT was tested with the Genadyne green thick foam dressing kit.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250586

    Validate with FDA (Live)

    Date Cleared
    2025-11-21

    (267 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Negative Pressure Wound Therapy Black Foam Kits and White Foam Accessory
    Regulation Number: 21 CFR 878.4780
    Classification Name** | Negative Pressure Wound Therapy Powered Suction Pump |
    | Regulation Number | 878.4780

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Prospera Spectruum™ Negative Pressure Wound Therapy System is a wound management system intended for use in acute, extended and home care settings.

    When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts. When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.

    Device Description

    The Prospera Spectruum™ Negative Pressure Wound Therapy (NPWT) System is a medical device designed to allow for wound management through controlled negative pressure. The NPWT device and canister work in tandem to generate and maintain the specified negative pressure level, while the dressing kit provides fluid management and coverage over the wound site. This system provides flexibility and targeted therapy to manage the wound-healing environment in acute, subacute, chronic, and surgically closed wounds, suitable for use in diverse care settings, including hospital, extended, and home care.

    The Prospera Spectruum Negative Pressure Wound Therapy System consists of:

    • Prospera Spectruum Negative Pressure Wound Therapy Unit
    • A disposable canister which collects wound exudate
    • Prospera Spectruum Negative Pressure Wound Therapy Kit which includes a wound interface dressing, semi-occlusive wound drape, and dual lumen dressing connection.
    • Optional White Foam Accessory

    The dressing kit is available in multiple sizes, allowing for the use of either black foam alone or with an optional White Foam Accessory, based on wound characteristics and clinical needs. The black foam, as part of the primary dressing, is a hydrophobic polyurethane foam designed to assist in uniform pressure distribution across the wound bed. The optional white foam, made of polyvinyl alcohol (PVA) is a highly retentive material that supports fluid absorption and even pressure distribution. Both foams are secured with semi-occlusive wound drape and connected to the NPWT device via tubing, enabling removal of wound exudate and infectious materials.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251646

    Validate with FDA (Live)

    Date Cleared
    2025-10-03

    (127 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: extriCARE® 1000 Negative Pressure Wound Therapy System
    Regulation Number: 21 CFR 878.4780
    TRADE NAME: | extriCARE® 1000 NPWT System |
    | DEVICE CLASSIFICATION: | Class II per 21 CFR 878.4780

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The extriCARE® 1000 NPWT System is intended to generate negative pressure to remove wound exudates, infectious material, and tissue debris from the wound bed.

    The extriCARE® 1000 NPWT System is intended for use in wounds with low to moderate levels of exudate. Appropriate wound types include:

    • Chronic wounds
    • Acute wounds
    • Traumatic wounds
    • Subacute and dehisced wounds
    • Partial-thickness burns
    • Ulcer wounds (such as diabetic or pressure)
    • Flaps and grafts
    • Closed surgical incisions

    The extriCARE® 1000 NPWT System is intended for use in healthcare facilities. The extriCARE® 1000 NPWT System is intended for use with extriCARE® 1000 Bandage Dressings.

    Device Description

    The extriCARE® 1000 Negative Pressure Wound Therapy (NPWT) System consists of a portable pump unit intended to generate vacuum pressure to remove low to moderate level of wound exudates, infectious material, and tissue debris from the wound bed. Three preset vacuum pressures can be selected by the user – 80, 100, and 125mmHg. The extriCARE® 1000 NPWT System is packaged and provided with the following components:

    1. extriCARE® 1000 Negative Pressure Wound Therapy Pump
    2. extriCARE® 1000 50cc Collection Canister
    3. Adapter Power Cord(5VDC)
    4. extriCARE® 1000 Bandage Dressing/K140634 (this component is provided separately)
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251826

    Validate with FDA (Live)

    Date Cleared
    2025-09-11

    (90 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    /Device Name:** RENASYS WOUND+ Dressing Kit with AIRLOCK Technology
    Regulation Number: 21 CFR 878.4780

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RENASYS WOUND+ Dressing Kit with AIRLOCK Technology is intended to be used in conjunction with Smith+Nephew traditional Negative Pressure Wound Therapy (tNPWT) RENASYS systems. The Smith+Nephew RENASYS NPWT system is indicated for patients who would benefit from a suction pump (Negative Pressure Wound Therapy), as it allows for wound management via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Appropriate wound types for RENASYS WOUND+ Dressing include:

    • Chronic
    • Acute
    • Traumatic
    • Sub-Acute and dehisced wounds
    • Ulcers (such as pressure or diabetic)
    • Partial-thickness burns
    • Flaps
    • Grafts

    RENASYS WOUND+ Dressing without filler can be used on the following additional wound type:

    • Closed surgical incisions

    The RENASYS WOUND+ Dressing Kit with AIRLOCK Technology in Small, Medium, and Large sizes is intended to be used with Smith+Nephew RENASYS TOUCH, RENASYS GO, and RENASYS EDGE NPWT systems.

    Device Description

    RENASYS WOUND+ Dressing kits with AIRLOCK Technology can be used without a wound filler, with foam or with gauze wound filler. The kits are accessories intended for use in conjunction with RENASYS TOUCH, RENASYS GO and RENASYS EDGE Negative Pressure Wound Therapy systems. RENASYS Negative Pressure Wound Therapy allows for wound management via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243187

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-27

    (270 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name:** Nisus ONE Negative Pressure Wound Therapy System
    Regulation Number: 21 CFR 878.4780
    Therapy System

    COMMON NAME: Nisus ONE NPWT System

    DEVICE CLASSIFICATION: Class II per 21 CFR 878.4780
    (k) Number:** K243187 | 510(k) Number: K140022 |
    | Regulation Number / Product Code: 21 CFR 878.4780
    / OMP | Regulation Number / Product Code: 21 CFR 878.4780 / OMP |

    Indications for Use:

    | *

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nisus ONE Negative Pressure Wound Therapy System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device allows wound management.

    The Nisus ONE Negative Pressure Wound Therapy System is only intended to be used with the Cork NPWT Wound Dressing Kit (K132004)

    The Nisus ONE Negative Pressure Wound Therapy System is suitable for use in both a professional healthcare facility and home use environment.

    Device Description

    Cork Medical has developed a negative pressure wound therapy (NPWT) system with the same intended use as the predicate device (K140022). The subject device features a modified user interface, and an adjusted feature set intended to support ease of use, while maintaining the same fundamental therapeutic function and performance characteristics as the predicate device. The Nisus ONE Negative Pressure Wound Therapy System interface utilizes a single membrane switch keypad to power the device and switch between continuous and intermittent therapy modes.

    The device display is limited to presenting critical operational parameters, including therapy mode, battery status, and target pressure settings. Visual and audible alarms are consistent with the predicate device (K140022) and alert critical battery, pressure leakage, system blockage, and full canister occur. While the Nisus ONE NPWT System's interface differentiates from the predicate Cork NPWT system, the mechanical components and therapy application are identical. The new model is intended to enhance accessibility by aligning with the needs of a broad and diverse patient population across various care settings.

    The components included within the Nisus ONE NPWT System are:

    • Nisus ONE Negative Pressure Wound Therapy Pump (OCMPP-100)
    • Nisus ONE Pump Battery Charger (NIS-CHRG-A-18)
    • Nisus NPWT Canister 500-mL (CPC-500)
    • Nisus NPWT Canister 250-mL (CPC-250, previously cleared in K140022)

    Accessory components are required to operate the device. Injection molded components are sonically welded to form a canister designed to mate with the pump and collect excess exudates, infectious material, and tissue debris. The Nisus NPWT Canister 250-mL design was previously cleared in the Nisus NPWT system 510k application (K140022). The Nisus NPWT Canister 500-mL is based on the same design as the Nisus NPWT Canister 250-mL. The dimensions were enlarged to provide a larger reservoir for patient convenience. Thus, the subject and predicate canisters have substantially equivalent characteristics.

    The Wound Kit comes in multiple iterations containing a minimum of one port pad, wound foam and peel and stick drape. Integrated with the pump mechanism, these components ensure the safe and efficient removal of wound exudate. NPWT Accessories were previously cleared in K132004. Our application for the Nisus ONE NPWT system includes no revisions to these previously cleared components.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Nisus ONE Negative Pressure Wound Therapy System does not contain information related to an AI/ML-based device performance study.

    The document details the substantial equivalence of the Nisus ONE system to a predicate device, focusing on changes to the user interface, physical dimensions, and technical specifications of a powered suction pump. The "Nonclinical Tests" section describes bench performance testing of the pump's functionality (e.g., pressure, alarms, flow rate) using simulated wound exudate and a test bed fixture, as well as software verification and validation, electrical safety, EMC, and usability testing. These are standard engineering and regulatory tests for medical devices, not performance studies of an AI/ML algorithm.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and reported device performance for an AI/ML algorithm.
    • Sample size, data provenance, expert qualifications, and adjudication methods for a test set related to an AI/ML algorithm.
    • Details on MRMC studies or standalone AI performance.
    • Information on training set size or ground truth establishment for an AI/ML algorithm.

    The document explicitly states: "No Clinical Testing was required to support these 510(k) submissions. No clinical testing has been performed." This further indicates that no studies involving human data or expert review for AI/ML performance were conducted or reported in this submission.

    If you have a document describing an AI/ML device, please provide that, and I would be happy to analyze it according to your criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243638

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-13

    (200 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K243638**
    Trade/Device Name: MV Flow Aspiration Pump (MVFLOWPMP01)
    Regulation Number: 21 CFR 878.4780
    | Classification Name | Apparatus, Suction, Ward Use, Portable, Ac-Powered |
    | Regulation Number | 878.4780

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MV Flow Aspiration Pump is intended for general suction use in hospitals or clinics.

    Device Description

    The MV Flow Aspiration Pump is an electrically powered (100 VAC/50 Hz, 240 VAC/60 Hz) device designed to generate vacuum pressure and serve as the vacuum source intended for general suction use in hospitals or clinics. The MV Flow Aspiration Pump is not intended for use in transport or field settings. The pump delivers 29 inHg of maximum vacuum pressure with a maximum flow rate of 36 L/min. The power button and vacuum-relief button provide pressure control. The brushless DC, dual-diaphragm pump design includes reverse-flow protection. Compatible pump accessories, supplied separately, include the MV Flow Canister. The MV Flow Canister consists of a single-use collection canister with integrated bacteria filter, overflow protection, and a pre-installed intermediate tubing.

    AI/ML Overview

    The provided 510(k) clearance letter is for the MV Flow Aspiration Pump, which is a medical device, not an AI/ML-driven diagnostic or therapeutic device. As such, the information you've requested regarding acceptance criteria, study methodologies, and ground truth establishment (which are typically relevant for AI/ML device clearances) is not applicable to this particular device.

    The clearance for the MV Flow Aspiration Pump focuses on demonstrating substantial equivalence to a predicate device (Penumbra Pump MAX) based on non-clinical performance testing, electrical safety, EMC/EMI compliance, and device-specific pump standards. There are no mentions of AI/ML components, image analysis, human reader studies, or complex ground truth establishment processes as would be found in clearances for AI-powered diagnostic tools.

    Therefore, I cannot provide the requested information in the format of a table of acceptance criteria for AI performance, sample sizes for test/training sets, expert adjudication methods, or MRMC studies, as these types of studies were not conducted or required for the clearance of this specific device.

    If you have a document related to an AI/ML medical device, I would be happy to analyze it according to your requested criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251450

    Validate with FDA (Live)

    Date Cleared
    2025-06-06

    (28 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    (BN007); Baby Nasal Aspirator (BN008); Baby Nasal Aspirator (BN009)
    Regulation Number: 21 CFR 878.4780
    (BN007); Baby Nasal Aspirator (BN008); Baby Nasal Aspirator (BN009)

    Regulation Number: 21 CFR 878.4780

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Baby Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided FDA 510(k) Clearance Letter for the "Baby Nasal Aspirator" does not contain information about acceptance criteria and a study proving the device meets those criteria.

    This document is a formal clearance letter issued by the FDA, confirming that the device is substantially equivalent to legally marketed predicate devices. It focuses on regulatory aspects such as:

    • Device Identification: Trade name, regulation number, product code, regulatory class.
    • Substantial Equivalence Determination: The FDA's finding that the device is equivalent to existing legal devices.
    • Regulatory Requirements: Information about general controls (registration, listing, GMP, labeling, misbranding, adulteration), additional controls for Class II/III devices, Quality System (QS) regulation, medical device reporting, Unique Device Identification (UDI) system, and contact information for regulatory advice.
    • Indications for Use: The stated purpose and target user group for the device.

    The letter explicitly states: "We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..."

    The letter does not include details on:

    • Specific performance acceptance criteria for the device itself (e.g., suction pressure, battery life, noise levels).
    • Any studies or test results demonstrating the device meets such criteria.
    • Information about test set sample sizes, data provenance, expert ground truth establishment, or clinical study methodologies (like MRMC studies) as these are typically associated with performance data for more complex or AI/imaging devices, not simple mechanical devices like a nasal aspirator.

    For a device like a baby nasal aspirator, the "acceptance criteria" and "proof of meeting them" would typically involve:

    • Bench testing: Measuring suction pressure, flow rate, durability, material biocompatibility, and safety (e.g., absence of sharp edges, proper battery compartment sealing).
    • Biocompatibility testing: To ensure materials are safe for contact with human tissue.
    • Electrical safety testing: If powered, to ensure it meets relevant electrical safety standards.
    • Usability testing: To ensure it can be safely and effectively used by the intended users (parents or caregivers).

    However, none of these specific test results or their underlying methodologies are detailed in the provided FDA clearance letter. The letter signifies regulatory clearance based on a review of the submitted documentation, which would have included these types of test reports, but it does not reproduce them within the letter itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    K244033

    Validate with FDA (Live)

    Date Cleared
    2025-05-16

    (137 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Aspirator (NASA601,NASA602,NASA603,NASA605,NASA606,NASA608,NASA610)
    Regulation Number: 21 CFR 878.4780
    Powered Suction Pump
    Regulatory Class: II
    Product Code: BTA
    Regulation Number: 21 CRF 878.4780
    Desida Technology Co., Ltd. | Shenzhen Desida Technology Co., Ltd. | / |
    | Regulation number | 21 CFR § 878.4780
    | 21 CFR § 878.4780 | Same |
    | Product code | BTA | BTA | Same |
    | Device classification | Class II

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Electric Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2~12 years old). This device is used in a home environment.

    Device Description

    Electric Nasal Aspirator consists of silicone nozzle which has three models (conical nozzle, flat nozzle, angled nozzle), collection cup, air tube, tweezers, charging cable-USB and aspirator body. The Electric Nasal Aspirator is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. The motor pump operates on a rechargeable battery. The rechargeable battery can be charged from the external power adapter (not included in this device) through the provided charging line.

    The user interface consists of buttons and LED display, and the user can control the vacuum pressure through the button. Three different shapes of silicone nasal tips are provided to enable easier and more effective removal of the nasal mucus.

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for an Electric Nasal Aspirator. It details the device's characteristics and its comparison to a predicate device, along with the performance data submitted to demonstrate substantial equivalence.

    However, the nature of this medical device (a nasal aspirator, essentially a suction pump) means the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are focused on engineering performance parameters (e.g., vacuum pressure, flow rate, safety standards) rather than clinical efficacy studies with human subjects that measure the device's ability to diagnose or treat a condition, which is what is typically discussed in the context of "human-in-the-loop performance," "ground truth," "expert consensus," and "MRMC studies" for AI/software-as-a-medical-device (SaMD).

    Therefore, the requested information elements related to AI/SaMD (e.g., training set, test set, expert adjudication, MRMC studies, standalone performance, ground truth for image analysis) are not applicable to this specific device submission. The "study" here is a series of bench tests and electrical/biocompatibility assessments.

    Below is a breakdown of the applicable acceptance criteria and the "study" (bench testing) as described in the provided text, while explicitly stating when an element is not applicable.


    Acceptance Criteria and Device Performance for Electric Nasal Aspirator (K244033)

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Predicate Device Performance)Reported Device Performance (Subject Device)Compliance/Remark
    Intended UseIntermittent removal of nasal secretions and mucus from children (age 2~12 years old) in a home environment.Intermittent removal of nasal secretions and mucus from children (age 2~12 years old) in a home environment.Same, Meets.
    Patient PopulationAge 2-12 years oldAge 2-12 years oldSame, Meets.
    Vacuum Pressure52-60 KPa52-60 KPaSame, Meets.
    Music FunctionYesYesSame, Meets.
    Light FunctionYesYesSame, Meets.
    Motor Type3.7V DC3.7V DCSame, Meets.
    Operating Condition5℃(41℉) to 40℃(104℉); 15% to 93% R.H.5℃(41℉) to 40℃(104℉); 15% to 93% R.H.Same, Meets.
    Storage Condition-10°C(-23℉) to 70°C (158℉); 10% to 95% R.H.-10°C(-23℉) to 70°C (158℉); 10% to 95% R.H.Same, Meets.
    Expected Service Life2 years2 yearsSame, Meets.
    Type BF Applied PartType BF applied partType BF applied partSame, Meets.
    Water-ResistanceIP22IP22Same, Meets.
    BiocompatibilityTested to ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Skin Sensitization), ISO 10993-23 (Skin Irritation) for silicone nasal aspiration tips (contact < 24 hrs).Passed ISO 10993-5, ISO 10993-10, ISO 10993-23 for silicone nasal aspiration tips.Passed tests, Meets.
    Electrical SafetyTested to IEC 60601-1, IEC 60601-1-11.Passed IEC 60601-1, IEC 60601-1-11.Passed tests, Meets.
    Electromagnetic Compatibility (EMC)Tested to IEC 60601-1-2.Passed IEC 60601-1-2.Passed tests, Meets.
    Battery SafetyTested to IEC 62133-2.Passed IEC 62133-2.Passed tests, Meets.
    Software Verification & ValidationConsistent with basic documentation level; all software requirements met, hazards mitigated.System validation testing completed, requirements met, hazards mitigated.Meets.
    Appearance (Color, Dimension, Weight)- (Compared to predicate, but specific criteria not given as numerical range)Different from predicate, but product performance test met requirements.Met requirements through performance testing.
    Flow Rate2.5-2.7 L/min (Predicate)3.3 - 6.0 L/min (Subject Device)Higher flow rate in subject device, but same vacuum pressure (52-60 kPa). Performance testing verified requirements. Does not raise issues.
    Power Consumption2W (max) (Predicate)5W (Subject Device)Different, but complies with safety standards (IEC 62133-2, IEC 60601-1, IEC 60601-1-2) so it does not raise safety/effectiveness issues.
    Power SourceDC 3.7 V / 700mAh Rechargeable Li-ion battery (Predicate)DC 3.7 V / 2500mAh Rechargeable Li-ion battery (Subject Device)Different, but complies with safety standards (IEC 62133-2, IEC 60601-1, IEC 60601-1-2) so it does not raise safety/effectiveness issues.
    Structural CompositionSilicone nozzle, collection cup, charging cable-USB (Predicate)Silicone nozzle, collection cup, air tube, tweezers, charging cable-USB and aspirator main body.Different (subject device has air tube and tweezers). Performance tests support safety/effectiveness of air tube. Tweezers are an accessory and do not raise safety/effectiveness concerns.
    Tubing Bending Resistance(Implied standard for performance test)Assured performance (Product Performance Test).Met.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated as a number of devices tested. The studies described are type testing (verification and validation) against engineering standards (e.g., IEC, ISO). This implies a representative sample of manufactured devices were tested to ensure design compliance. For biocompatibility, electrical safety, EMC, battery safety, and software V&V, these are typically conducted on a limited number of devices or components to demonstrate compliance with the standards' requirements for design verification.
    • Data Provenance: The device manufacturer, Shenzhen Desida Technology Co., Ltd. (China), performed these tests. The data provenance is internal testing to support the 510(k) submission. These are bench tests, not clinical data.
      • Retrospective/Prospective: Not applicable as these are bench tests and not clinical studies involving patient data collected over time.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable. For this type of medical device (Electric Nasal Aspirator), "ground truth" as typically applied to clinical or diagnostic AI/SaMD studies (e.g., for image interpretation) is not established through expert consensus on a test set. The validation relies on meeting pre-defined engineering and safety standards. The "experts" involved would be accredited test houses and engineers specializing in the relevant standards (e.g., electrical safety, biocompatibility).

    4. Adjudication Method for the Test Set

    • Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are used in clinical studies, particularly for establishing ground truth in diagnostic accuracy assessments involving human interpretation. This device's validation focuses on engineering performance, not interpretative accuracy.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI/SaMD device, and no MRMC study was conducted. The device is a physical product for suction, not a diagnostic or interpretive tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm-based device. Performance testing for this device is "standalone" in the sense that the device itself is tested against its specifications and relevant safety standards, without continuous human operation being part of the assessment of device performance (though a human operates it to perform the tests).

    7. The Type of Ground Truth Used

    • Engineering and Safety Standards: The "ground truth" for this device's acceptance is adherence to established international and national engineering standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133-2, ISO 10993 series) and the manufacturer's own verified product requirements derived from these standards and the predicate device's performance. For biocompatibility, this typically involves laboratory assays confirming the absence of cytotoxicity, skin sensitization, and irritation based on ISO 10993 protocols. For physical properties, it involves measurements against specified ranges (e.g., 52-60 KPa vacuum pressure).
    • Not clinical outcomes data, pathology, or expert consensus (in the traditional sense for diagnostic/AI devices).

    8. The Sample Size for the Training Set

    • Not Applicable. This device is a physical product, not an AI/Software device requiring a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no "training set" for this physical device, there is no ground truth established for it in the context of machine learning. The device's design and manufacturing processes are validated against engineering principles and standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K251111

    Validate with FDA (Live)

    Device Name
    Cary
    Manufacturer
    Date Cleared
    2025-05-07

    (26 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Blvd
    Warren, New Jersey 07059

    Re: K251111
    Trade/Device Name: Cary
    Regulation Number: 21 CFR 878.4780
    Classification Name:* Apparatus, Suction, Ward Use, Portable, Ac-Powered
    Regulation Number: 878.4780

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cary is indicated for removal of fluids from the upper airway. The device creates a negative pressure (vacuum) that draws fluids through disposable suction tube that is connected to a disposable collection canister. The fluids are captured in the collection canister for proper disposal. Cary is for use only on the order of a physician or other licensed practitioner (e.g. Emergency Medical Technician in field use).

    Device Description

    [Cary] is a handheld, electrically powered medical suction device used to clear the upper airway during emergency care. It consists of a reusable drive unit and disposable collection container (pumpister) and tubing. The drive unit powers a piston pump, creating negative pressure (vacuum) in the disposable collection container. The vacuum draws fluids and other matter through a disposable tubing connected to the container, where they are collected for disposal. The size and shape of the device enable use by one hand, and a rechargeable lithium-ion battery allows field and transport use of the device.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the "Cary" device ([K251111](https://510k.innolitics.com/device/K251111)) primarily discuss non-clinical performance testing to demonstrate substantial equivalence to a predicate device. This document does not describe a clinical study (e.g., MRMC or standalone human-in-the-loop study) involving human readers or expert panels for establishing ground truth, as would typically be seen for AI/ML-driven diagnostic devices. Instead, the testing focuses on the physical and functional performance of the suction pump itself against recognized consensus standards.

    Therefore, many of the requested points regarding clinical study design, expert panels, ground truth, and MRMC studies are not applicable based on the provided text.

    Here's an analysis based on the information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes several non-clinical tests with implied or explicit acceptance criteria based on industry standards and comparison to the predicate device.

    Test CategoryAcceptance Criteria (Implied/Explicit from Standards/Predicate)Reported Device Performance
    Operating PositionAbility to operate in any spatial orientation.Confirmed.
    Protection DevicesCompliance with overfill protection and reverse flow protection.Results within acceptance criteria.
    Noise LevelUnspecified standard, implied safe/acceptable level.70.5 dB. "Within the acceptance criteria apart from 0.5dB excess when the suction tip was blocked. This is considered as an acceptable deviation from the standard requirement as it does not decrease the safety or effectiveness of the subject device when compared to the predicate device."
    Air LeakageAs per standard requirements for general use suction equipment.Results within acceptance criteria.
    Vacuum Levels & Free Air FlowWithin acceptance criteria for High vacuum/high flow and Low vacuum/low flow (similar to predicate: High vacuum <600 mmHg vs <550 mmHg; Low vacuum ≤100 mmHg vs 50 mmHg; High flow ≥ 20 l/min vs 30 l/m).Results within acceptance criteria. Suction performance "similar to LCSU 4".
    Pharyngeal Suction EquipmentAbility to evacuate ≥ 200 ml of simulated vomitus in ≤ 10 s.Results within acceptance criteria.
    Reprocessing ValidationAppropriateness of manual reprocessing process; user ability to carry out instructions.Outcome within acceptance criteria.
    BiocompatibilityCompliance with ISO 10993 standards for cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity.Confirmed for direct-contact materials.
    SoftwareCompliance with IEC 62304 and FDA guidance for software functions, classified as Class B.Verification and validation conducted, documentation provided as recommended. (Specific performance metrics are not applicable here as it's about compliance, not a clinical outcome).
    EMC and Electrical SafetyCompliance with IEC 60601-1-2 and IEC 60601-1 series.Confirmed.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical, laboratory-based testing of the device's physical and functional performance, not a clinical study involving a dataset of patient cases. Therefore, "sample size for the test set" and "data provenance" in the context of patient data are not applicable. The tests were performed on the device prototypes/production samples in controlled laboratory environments.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth, in the context of device performance testing for a suction pump, refers to established engineering specifications, physical laws, and recognized standards (e.g., ISO 10079-1). There are no "experts" in the sense of medical professionals interpreting patient data to establish ground truth for this type of device.

    4. Adjudication Method for the Test Set

    Not applicable. As this is non-clinical performance testing against engineering standards, an adjudication method for conflicting interpretations of medical data is not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly describes non-clinical testing and comparison to a predicate device based on device performance metrics (e.g., vacuum levels, flow rates, noise), not on human reader performance with or without AI assistance. This device is a powered suction pump, not an AI diagnostic/imaging device that would typically undergo MRMC studies.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. The device is a mechanical suction pump, not an AI algorithm. Its performance is inherent to its physical design and operation, not an output from an algorithm in the typical AI sense.

    7. The Type of Ground Truth Used

    The ground truth used for these non-clinical tests is based on:

    • Recognized Consensus Standards: e.g., ISO 10079-1, ISO 10079-4, IEC 62304, ISO 10993, IEC 60601 series.
    • Engineering Specifications: Explicit performance requirements (e.g., ≥ 20 l/min flow rate, < 1 kPa air leakage).
    • Comparison to Predicate Device Performance: Establishing equivalence or superiority in key performance areas.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not an AI/ML device that requires machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 33