510(k) Data Aggregation

    K Number
    K253531

    Validate with FDA (Live)

    Date Cleared
    2026-02-23

    (102 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Alexis® Lighted Wound Protector-Retractor, Flexible, Medium (CL302)
    Regulation Number: 21 CFR 878.4370
    Protector-Retractor
    Common Name: Lighted Wound Protector and Retractor
    Product Classification: KGW (21CFR 878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alexis Lighted Wound Protector-Retractor is a device that allows the surgeon to access the abdominal cavity through an atraumatically retracted wound, providing maximum exposure with minimum incision size. In addition to incision retraction, it is intended to provide supplemental illumination and protect against wound contamination during both laparoscopic and open surgery. The Alexis Lighted Wound Protector-Retractor is capable of temporarily closing an incision to maintain pneumoperitoneum during laparoscopic surgery, or to serve as an additional trocar port site. In addition, the device may be used to access the thoracic cavity or other soft tissue retraction during cardiac and general surgical procedures.

    Device Description

    Applied Medical's Alexis Lighted Wound Protector-Retractor provides wound protection and retraction during laparoscopic and open surgery while providing supplemental illumination. The device is provided sterile, and the reusable power supply is provided non-sterile. The device and power supply are used in the clinical setting.

    The system includes five main components:

    • A sheath that protects the wound and provides retraction with tension applied.
    • An outer ring that is rolled to create tension on the sheath and anchor the device.
    • An inner ring with embedded LEDs that anchors the device and provides illumination.
    • A power cord that connects to the reusable power supply.
    • A reusable power supply that connects to a grounded receptable to provide power to the LEDs.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250735

    Validate with FDA (Live)

    Date Cleared
    2025-06-10

    (91 days)

    Product Code
    Regulation Number
    878.4370
    Panel
    Dental
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CHINA

    Re: K250735
    Trade/Device Name: Dental Barrier and Sleeves
    Regulation Number: 21 CFR 878.4370
    CHINA

    Re: K250735
    Trade/Device Name: Dental Barrier and Sleeves
    Regulation Number: 21 CFR 878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K250606

    Validate with FDA (Live)

    Date Cleared
    2025-05-29

    (90 days)

    Product Code
    Regulation Number
    878.4370
    Panel
    Dental
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    China

    Re: K250606
    Trade/Device Name: Dental Barrier and Sleeves
    Regulation Number: 21 CFR 878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K242435

    Validate with FDA (Live)

    Date Cleared
    2024-10-18

    (63 days)

    Product Code
    Regulation Number
    878.4370
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    200122 CHINA

    Re: K242435

    Trade/Device Name: Dental Barrier and Sleeves Regulation Number: 21 CFR 878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dental Barrier and Sleeves are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) clearance letter from the FDA for a device named "Dental Barrier and Sleeves." It explicitly states that the device is a physical barrier and not an AI/software device. Therefore, the questions related to AI/software performance acceptance criteria, test sets, ground truth, expert opinions, and training sets are not applicable to this document.

    The document discusses regulatory aspects of a Class II medical device, including:

    • Trade/Device Name: Dental Barrier and Sleeves
    • Regulation Number/Name: 21 CFR 878.4370, Surgical Drape And Drape Accessories
    • Regulatory Class: Class II
    • Product Code: PEM, PUI
    • Indications for Use: Intended to be used as a disposable barrier for dental instruments and equipment. Non-sterile and intended for single patient use only.

    Since the device is a physical product and not an AI or software-based system, the acceptance criteria would relate to its physical properties, material safety, effectiveness as a barrier, and manufacturing quality, rather than algorithms or data performance. Such tests would typically involve:

    • Material Biocompatibility: Testing to ensure the materials are not toxic or irritating when in contact with patients or users.
    • Barrier Integrity/Effectiveness: Physical testing to confirm its ability to prevent the passage of fluids, microorganisms, or other contaminants.
    • Physical Properties: Tensile strength, tear resistance, and dimensions to ensure it performs as intended under typical dental use conditions.
    • Sterility (if applicable, though this device is noted as non-sterile): If it were a sterile barrier, testing would confirm the absence of viable microorganisms.
    • Shelf Life: Stability testing to ensure the product remains effective and safe over its intended storage period.
    • Manufacturing Quality Control: Adherence to Good Manufacturing Practices (GMP) to ensure consistent quality.

    None of this information is detailed in the provided FDA clearance letter, as the letter primarily serves to confirm substantial equivalence to a predicate device and outlines regulatory obligations, not to present detailed study results for a non-AI/software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240573

    Validate with FDA (Live)

    Date Cleared
    2024-08-16

    (168 days)

    Product Code
    Regulation Number
    872.1745
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Lumina Pro wand barrier sleeve testing met requirements for dental barriers regulated under 21 C.F.R §878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iTero Lumina Pro NRI functionality is a diagnostic aid for the detection of interproximal carious lesions above the gingiva and for monitoring the progress of such lesions.

    Device Description

    The iTero Lumina Pro is an integrated intraoral imaging system capable of capturing 3D optical impressions for CAD/CAM of dental devices, 2D color images and Near Infrared (NIR) images. Scanning cycles are employed allowing the system to quickly alternate between the three modalities, creating a virtually simultaneous capture of the 3D optical impression, 2D color and NIR images.

    The iTero Lumina Pro consists of the following main components:

    • A. Scanning Wand used for intraoral scanning.
    • B. Wand Cradle used to hold the Scanning Wand when not in use.
    • C. Application Software that includes Graphical User Interface ("GUI").
    • D. Computer platform

    The scanning wand houses electronic, mechanical, and optical components, including projectors, LEDs and cameras which enables capturing 3D impression data as well as 2D color and NIRI images. The wand body is made of Aluminum, Copolyester and Glass. When not in use, it is stored in the Wand Cradle.

    The iTero Lumina Pro features proprietary software that can be installed on the various configurations of the computer platform. The software is designed to manage the GUI, control the scanning process, display the scanned images, and connect to the stored patient data, including scanned data for display or export.

    The iTero Lumina scanning wand is designed to be used with a single-use, non-sterile, disposable, wand barrier sleeve. The only patient-contacting materials are those contained within the wand barrier sleeve.

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria and a study that comprehensively proves the device meets these criteria in the format requested. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a full clinical trial report with specific performance metrics against pre-defined acceptance criteria.

    However, I can extract information related to performance data and testing methods mentioned in the "Performance Data" section to infer some aspects of device performance and the types of studies conducted.

    Here's an attempt to answer your questions based on the provided text, while acknowledging the limitations of the available information:


    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria for NIRI functionality or specific performance targets. Instead, it states that "All testing showed similar performance to the predicate system." This implies that the acceptance criterion was likely comparable performance to the predicate device across various technical attributes.

    Acceptance Criterion (Inferred from testing performed)Reported Device Performance (Summary)
    Field of View (NIRI)Similar performance to the predicate system
    Image Sharpness (NIRI)Similar performance to the predicate system
    Depth of Field (NIRI)Similar performance to the predicate system
    Illumination Power (NIRI)Similar performance to the predicate system
    Wavelength and Spectral Bandwidth (NIRI)Similar performance to the predicate system (uses 850nm, same as predicate)
    Specular Reflection Suppression (NIRI)Similar performance to the predicate system
    Signal to Noise Ratio (NIRI)Similar performance to the predicate system
    Wand barrier sleeve material propertiesMet requirements for dental barriers (e.g., tensile strength, tear resistance, puncture, penetration)
    Wand barrier sleeve finished assemblyMet requirements (Microbial Ingress Testing, Pull testing) and showed equivalent performance to predicate
    Basic Safety (IEC 60601-1, IEC 80601-2-60, etc.)Complies with relevant standards
    Electromagnetic Compatibility (IEC 60601-1-2, FDA Gdn)Complies with relevant standards
    Usability (IEC 60601-1-6, IEC 62366-1)Complies with relevant standards
    Software Life-Cycle (IEC 62304, FDA Gdn)Designed and developed according to standards
    Cybersecurity (UL ANSI 2900-1, UL ANSI 2900-2-1, FDA)Ensured by compliance with requirements
    Biocompatibility (ISO 10993-series)Patient contacting components shown to be biocompatible
    Cleaning/Reprocessing (ANSI/AAMI ST98:2022, FDA Gdn)Found to be reliably cleaned and validated

    2. Sample size used for the test set and the data provenance

    The document explicitly states that the company performed "bench testing" to support substantial equivalence. It does not provide details on sample sizes for the NIRI performance attributes (Field of View, Image Sharpness, etc.). It mentions ASTM standards for material testing of the wand barrier sleeve, which would imply defined sample sizes per those standards, but these are not enumerated in the text. There is no information provided about the origin or nature of the "test set" data in terms of patient population (e.g., country of origin, retrospective or prospective).

    The testing described appears to be primarily non-clinical bench testing rather than clinical study data involving patient samples or diagnostic accuracy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not describe a process for establishing ground truth using human experts for diagnostic accuracy. The performance data presented focuses on technical and engineering performance characteristics of the device itself (e.g., optical properties, material strength, electrical safety).

    4. Adjudication method for the test set

    Not applicable, as there is no description of a test set involving human interpretation and adjudication of diagnostic results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document does not mention any MRMC study or AI assistance. The device is described as a "diagnostic aid" providing NIR images for a dentist's review. The NIRI functionality itself is the diagnostic aid, not an AI component assisting human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not describe any standalone algorithm performance testing. The device's NIRI functionality generates images that are then reviewed by a dentist.

    7. The type of ground truth used

    For the technical and engineering tests described, the ground truth would be established by the specifications and performance characteristics of the predicate device and relevant international standards (e.g., for electrical safety, biocompatibility, material properties). For the NIRI imagery itself, the comparison is made against the predicate device's image output, not against a clinical "ground truth" like pathology for caries detection in this summary.

    8. The sample size for the training set

    Not applicable. The document describes a medical device generating images, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a training set for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233918

    Validate with FDA (Live)

    Date Cleared
    2024-05-09

    (148 days)

    Product Code
    Regulation Number
    878.4370
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    200122 CHINA

    Re: K233918

    Trade/Device Name: Dental Barrier and Sleeves Regulation Number: 21 CFR 878.4370
    Surgical Drape And Drape Accessories Classification: Class II Product code: PEM Regulation Number: 21 CFR 878.4370
    | / |
    | Classification | Class II Device (21 CFR 878.4370
    | Class II Device (21 CFR 878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dental Barrier and Sleeves are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

    Device Description

    Dental Barrier and Sleeves consist of various sizes and shapes of polyethylene covers which are positioned on various small hand-held dental instruments such as hand pieces, curing lights, air/water syringes and similar hand instruments. In other forms, they are used to cover various devices such as dental chairs, dental instrument trays, x-ray heads, and others. The products are sold non-sterile, prepackaged, and are disposable, single use only.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called "Dental Barrier and Sleeves." It outlines the device's characteristics and compares them to a legally marketed predicate device to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Important Note: This submission is for a Class II medical device, "Dental Barrier and Sleeves," which are physical barriers (polyethylene covers) for dental instruments and equipment. The document does not describe an AI/Software as a Medical Device (SaMD) and therefore doesn't contain information related to AI model performance, expert consensus on image interpretation, or multi-reader multi-case studies. The performance criteria and studies described are for the physical properties and biocompatibility of the barrier sleeves themselves.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on demonstrating equivalence in physical properties and biocompatibility with the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:
    Acceptance Criteria CategorySpecific Acceptance Criteria (Method)Reported Device Performance and Outcome
    Intended UseSame intended use as predicate deviceSame: "disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only."
    ClassificationClass II (21 CFR 878.4370)Same: Class II (21 CFR 878.4370)
    Product CodePEMSame: PEM
    MaterialPolyethylene film (same composition as predicate or justified)Same: Polyethylene film (LLDPE 80%, LDPE 20%)
    Specifications & TolerancesFilm Thickness: 0.02-0.06mm, Tolerance: <0.01mm (similar to predicate)Film Thickness: 0.02-0.06mm, Tolerance: <0.01mm (Deemed "Similar" by sponsor)
    SterilityNon-sterileNon-sterile
    Performance PropertiesSynthetic Blood Penetration test (ASTM F1670) - PassPass
    Viral Penetration test (ASTM F1671) - PassPass
    Tear Resistance (ASTM D1004)Performed (Result implied to be acceptable for equivalence)
    Tensile Strength (ASTM D882)Performed (Result implied to be acceptable for equivalence)
    Puncture Resistance (ASTM F1342)Performed (Result implied to be acceptable for equivalence)
    Effectiveness on X-ray devices (no impact on function)Performed to demonstrate no impact on dental x-ray devices' function.
    BiocompatibilityCompliance with ISO 10993-5 (Cytotoxicity)Test results indicate non-cytotoxic.
    Compliance with ISO 10993-10 (Sensitization)Test results indicate non-sensitizing.
    Compliance with ISO 10993-23 (Irritation)Test results indicate non-irritating.
    LabelingSingle use deviceSame: Single use device

    1. Sample Sizes Used for the Test Set and Data Provenance:

      • Test Set Sample Size: The document does not specify a specific "sample size" in terms of number of units for the performance tests (e.g., blood penetration, viral penetration, physical properties). It implies that standard testing procedures typical for medical devices of this type were followed.
      • Data Provenance: The studies are "Non-Clinical Testing" which means they were likely conducted in a laboratory setting. The country of origin for the tests or data is not explicitly stated. It is considered retrospective as the tests were performed to support the 510(k) submission.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • This question is not applicable to this device. As a physical barrier, there is no "ground truth" to be established by experts in the context of image interpretation or diagnostic accuracy. The performance is based on measurable physical and biological properties.
    3. Adjudication Method for the Test Set:

      • This question is not applicable to this device. Adjudication methods like "2+1" or "3+1" are relevant for expert review of medical images or clinical outcomes, which is not the nature of the testing for this device.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size:

      • No. This is a physical device, not an AI/SaMD. MRMC studies are not relevant to its evaluation.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is a physical device, not an algorithm. Standalone performance is not relevant.
    6. The Type of Ground Truth Used:

      • The "ground truth" for this device's performance is based on standardized test methods and material properties.
        • For performance properties (blood/viral penetration, tear, tensile, puncture), the ground truth is defined by the passing criteria of the respective ASTM standards (ASTM F1670, ASTM F1671, ASTM D882, ASTM F1342, ASTM D1004).
        • For biocompatibility, the ground truth is defined by the "non-cytotoxic, non-sensitizing, and non-irritating" outcomes determined by the ISO 10993 suite of standards (ISO 10993-5, ISO 10993-10, ISO 10993-23).
        • For X-ray device effectiveness, the ground truth is the continued proper function of dental X-ray devices when covered by the barriers.
    7. The Sample Size for the Training Set:

      • Not applicable. This is a manufactured physical device, not an AI model. There is no "training set" in the context of machine learning.
    8. How the Ground Truth for the Training Set was Established:

      • Not applicable. As there is no training set, there is no ground truth to be established for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230645

    Validate with FDA (Live)

    Date Cleared
    2023-10-25

    (231 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: BeneHold™ Surgical Incise Drape with CHG antimicrobial II Regulation Number: 21 CFR 878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BeneHold™ Surgical Incise Drape with CHG antimicrobial II is indicated for use as an incise drape with continuous antimicrobial activity (in vitro) from 30 minutes through 4 hours of use against the following organisms: S. aureus, S. epidermidis, E. coli, P. aeruginosa, E. aerogenes, and C. albicans. In vitro testing on the subject drape has also shown continuous antimicrobial activity from 30 minutes through 2.5 hours against E. faecalis and F. oxysporum. It is intended for external use only. Correlation between in vitro results and any clinical event has not been tested.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) clearance letter from the FDA for a surgical incise drape. It outlines the regulatory clearance of the device, its intended use, and relevant FDA regulations.

    This document does not contain any information about acceptance criteria for a study proving the device meets acceptance criteria, nor does it detail any specific study design, sample sizes, expert involvement, or adjudication methods. The letter focuses solely on the regulatory approval process for the medical device.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets those criteria based on the provided text. The information required (e.g., performance metrics, test set details, ground truth establishment) is not present in this regulatory clearance document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223669

    Validate with FDA (Live)

    Date Cleared
    2023-06-12

    (187 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K223669

    Trade/Device Name: SecurePortIV Advanced Catheter Securement Adhesive Regulation Number: 21 CFR 878.4370
    and Plastic Surgery |
    | Regulation: | Class II, under 21 CFR 878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SECUREPORTIV® ADVANCED Catheter Securement Adhesive is to be applied as a film forming securement and sealant at the point of vascular access catheter skin entry. The film holds the catheter to the skin to reduce catheter movement, migration, and/or dislodgment. It is used to protect the catheter skin entry site by creating a sealant that immobilizes surface bacteria, preventing them from entering into the catheter skin entry site while a moisture barrier. SECUREPORTIV® ADVANCED is intended to be used with a transparent film dressing for the securement of short-term and long-term vascular access catheters including peripheral IVs, PICCs, and CVCs.

    Device Description

    SecurePortIV Advanced Catheter Adhesive is a sterile, professional liquid cyanoacrylatebased adhesive containing a monomeric formulation (2-octyl cyanoacrylate) and the colorant D&C Violet #2. The device is an applicator with the formulation incorporated in an ampoule housed in a tapered plastic tube. The SecurePortIV Advanced liquid is applied as a film forming securement and sealant at the point of catheter skin entry, polymerizing in minutes.

    AI/ML Overview

    The SecurePortIV Advanced Catheter Securement Adhesive underwent various tests to demonstrate its safety and effectiveness, leading to its clearance based on substantial equivalence to a predicate device.

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Catheter Adhesion to SkinHold BD Autoguard Catheter 1, 3, and 7 days; Hold Nexiva catheter 1 to 7 daysPass, same as predicate device
    Sealant of the Cannulation SitePrevent dye penetration at 1, 4, and 7 daysPass, same as predicate device
    Immobilization of Surface BacteriaPrevent bacteria from penetrating cannulation sitePass, same as predicate device
    Removal TimeSame removal time as predicate devicePass, same as predicate device
    Clinical (Securement Time)Securement time obtainedPass (14 seconds, compared to predicate's 143 seconds)
    CytotoxicityISO 10993-5: Tests for in vitro cytotoxicityPass
    SensitizationISO 10993-10: Tests for irritation and skin sensitizationPass
    IrritationISO 10993-10: Tests for irritation and skin sensitizationPass
    PyrogenicityLAL Limit Screen; Current USP <85>Pass
    Acute Systemic ToxicityISO 10993-10: Tests for irritation and skin sensitizationPass
    Subacute Systemic ToxicityISO 10993-11: Tests for systemic toxicityPass
    ImplantationISO 10993-6: Tests for local effects after implantationPass
    IntracutaneousISO 10993-10: Tests for irritation and skin sensitizationPass
    Hydrolytic Degradation<10µg/mLPass
    Heat of PolymerizationNo greater than predicate devicePass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each individual test related to catheter adhesion, sealant effectiveness, or bacterial immobilization. The tests are described generally, indicating they were qualitative "Pass" results compared to the predicate device.

    For the Clinical (Securement Time), the result for the subject device is 14 seconds compared to the predicate's 143 seconds. However, the sample size or methodology for obtaining these times is not detailed in the provided text.

    The data provenance is internal to Adhezion Biomedical, LLC, and appears to be from non-clinical (laboratory/bench) studies. There is no mention of specific country of origin or whether these studies were retrospective or prospective, but given they are performance tests against acceptance criteria, they are inherently prospective for this device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the given document. The document describes non-clinical performance and biocompatibility testing, which typically rely on standardized methods and laboratory assessments rather than expert consensus on a test set.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies when multiple reviewers assess subjective data (e.g., medical images). Since the provided data focuses on non-clinical performance and biocompatibility parameters, an adjudication method is not applicable and therefore not mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of an MRMC comparative effectiveness study being conducted or its results in the provided document. The study focuses on direct performance comparisons and biocompatibility rather than human reader performance with or without AI assistance.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    This device is not an AI algorithm; it's a physical medical device (an adhesive). Therefore, the concept of "standalone performance (algorithm only without human-in-the-loop performance)" is not applicable. The performance criteria are for the device itself.

    7. Type of Ground Truth Used

    The ground truth for the non-clinical tests appears to be based on:

    • Established laboratory performance standards: For adhesion, sealant penetration, and bacterial immobilization, the device's performance was compared to that of the legally marketed predicate device.
    • Standardized ISO and USP guidelines: For biocompatibility testing (e.g., ISO 10993 series for cytotoxicity, sensitization, irritation, systemic toxicity, implantation; USP <85> for pyrogenicity).
    • Quantitative measurements: For securement time, a specific time (14 seconds) was measured and reported.

    8. Sample Size for the Training Set

    This question is not applicable as the device is a physical medical product, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as the device is a physical medical product, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222578

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-18

    (266 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    3M™ Ioban™ CHG Chlorhexidine Gluconate Incise Drape (contains 2% w/w CHG) Regulation Number: 21 CFR 878.4370
    |
    | Classification Name: | Surgical drape and drape accessories[21 CFR § 878.4370
    |
    | Regulation | 21 CFR 878.4370
    | 21 CFR 878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    3M™ Ioban™ CHG Chlorhexidine Gluconate Incise Drape (contains 2% w/w CHG) is indicated for use as an incise drape with continuous antibacterial activity based on in vitro time kill data out to 90 minutes. It is intended for external use only.

    Device Description

    3M™ Ioban™ CHG Chlorhexidine Gluconate Incise Drape (contains 2% w/w CHG) is a sterile adhesive film that is applied pre-operatively to skin. It adheres and conforms to body contours, providing a sterile surface on top of the skin prior to surgery. The surgeon incises through the adhesive film and skin, ensuring a sterile film barrier all the way to the edge of the incision. The Chlorhexidine Gluconate (CHG) contained in the adhesive provides antibacterial activity to help reduce the risk of bacterial contamination in the surgical wound.

    The drape consists of a transparent, conformable, barrier film coated with an acrylic adhesive containing 2% w/w CHG. The adhesive drape is delivered on a release liner that is discarded after application.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results from non-clinical testing performed for the 3M™ Ioban™ CHG Chlorhexidine Gluconate Incise Drape.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Barrier PerformanceLevel 4PASS
    Force at 25% Elongation< 425 g/inPASS
    Assessment of Tear ResistanceRationalization ProvidedPASS
    FlammabilityClass I (Normal)PASS
    Moisture Vapor Transmission Rate> 400 g/m²/24 hoursPASS
    in vitro Direct Time Kill and Minimum Effective Concentration> 4 log reduction at 90 minutesPASS
    CHG Release Kinetics> 0.0% after 90 minute extractionPASS
    Ethylene Oxide ResidualsPer ISO 10993-7: 2008; Amd 1: 2019PASS
    Biocompatibility (Chemical Characterization, Hemolysis, Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity)Not cytotoxic, slight irritant, not a potential skin sensitizer, not pyrogenic, not acutely systemically toxic, non-hemolytic (for predicate, but additional endpoints tested for submission device)All tests PASS (e.g., Not cytotoxic, slight irritant, not a potential skin sensitizer, not pyrogenic, not acutely systemically toxic, non-hemolytic)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document primarily details non-clinical bench testing and biocompatibility testing. It does not specify sample sizes for each individual test. The data provenance is not explicitly stated beyond being "bench testing" and "biocompatibility testing," implying laboratory-based data rather than patient data from a specific geographical location. The nature of these tests suggests they are prospective, as they were conducted specifically to evaluate the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the document discusses non-clinical bench and biocompatibility testing. These types of tests typically rely on standardized methods and laboratory analysis rather than expert human interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the document discusses non-clinical bench and biocompatibility testing. Adjudication methods are typically used in clinical studies or studies involving subjective human interpretation of outputs, which is not the case here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is an incise drape with antibacterial properties, not an AI-assisted diagnostic or interpretive device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a physical medical device (incise drape), not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the non-clinical tests is established by objective measurements against predefined scientific and regulatory standards/methods. For example:

    • Barrier Performance: Measured against AAMI PB70:2012 using ASTM F1670.
    • Antibacterial Activity: Measured by in vitro direct time kill based on ASTM E2315, aiming for a >4 log reduction.
    • Biocompatibility: Evaluated against ISO 10993-1 and FDA Guidance for biological evaluation.

    8. The sample size for the training set

    This information is not applicable. The document describes the testing of a physical medical device, not a machine learning model that would require a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221249

    Validate with FDA (Live)

    Device Name
    L1P-1F (TRIOS 5)
    Manufacturer
    Date Cleared
    2022-09-16

    (137 days)

    Product Code
    Regulation Number
    872.1745
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    to ensure that requirements for clearance of dental barriers and sleeves requlated under 21 C.F.R §878.4370

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The L1P-1F intraoral scanner (IOS) system is intended for aid in diagnosis of caries.

    Device Description

    The L1P-1F intraoral scanner (IOS) system is intended for aid in diagnosis of caries. The L1P-1F system consists of a scanner and a detachable scanner tip which enables scanning using white and blue (fluorescence) Light Emitting Diodes (LEDs). A single-use scanner body sleeve is mounted on the scanner before being introduced into the mouth of a patient for scanning of the teeth.

    The interface to the computer is handled by ScanSuite TRIOS which functions as a Hardware Abstraction Layer (HAL). The scan control of, and information from the scanner is handled by the TRIOS module software, which holds the interface to the user.

    The L1P-1F system also consists of the TRIOS Patient Monitoring software which aids in diagnosis of occlusal and surface caries.

    The functional principle of L1P-1F (TRIOS 5) is to collect images of the object scanned by means of focus scanning, where light from the LEDs travels through the optical system to the object being scanned and back to an image sensor. 3D true colored images of the teeth combined with 2D green and red image as an additional texture are generated by emitting blue and white light in turn.

    AI/ML Overview

    The provided text describes the 3Shape TRIOS 5 (L1P-1F) intraoral scanner system, which is intended to aid in the diagnosis of caries. The documentation focuses on demonstrating substantial equivalence to a predicate device (DÜRR DENTAL AG – VistaCam iX "Proof" – K150672). However, it does not explicitly provide a table of acceptance criteria and reported device performance in the format requested, nor does it detail a specific study proving the device meets acceptance criteria with all the requested information.

    The document discusses key performance attributes tested and compared, which serve as implicit acceptance criteria for establishing substantial equivalence. These include:

    1. Fluorescence scanning
    2. Color Separation
    3. In vitro caries detection study
    4. Spatial resolution

    Based on the available information, here's an attempt to structure the response, acknowledging where specific details for the requested categories are not explicitly provided in the document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Fluorescence Scanning PerformanceThe L1P-1F uses two blue LEDs (405 nm) for fluorescence detection, yielding "sufficient performance." The device detects green to red visible light (>435 nm), similar to the predicate device. The blue-violet light causes tooth structures and bacterial metabolites to fluoresce, similar to the predicate.
    Color Separation/Categorization for Caries AidThe L1P-1F presents two severity categories: "Initial" (covering early-stage enamel caries, enamel caries up to enamel/dentine junction, and dentin junction already exceeded from the predicate's classification) and "Moderate/Extensive" (for deep dentin caries). This aligns with ICDAS merged codes.
    In vitro Caries Detection PerformancePerformance testing demonstrates that L1P-1F is "as safe and effective as a legally marketed device" and "does not raise different questions of safety and effectiveness than the predicate device."
    Spatial Resolution PerformanceA qualitative comparison shows "the resolution of L1P-1F must be at least as good or better than the predicate device."
    Electrical Safety / Electromagnetic Compatibility (EMC)Complies with IEC 60601-1:2005/AMD1:2012, IEC 60601-1-2:2014, and IEC 62366-1:2015+A1:2020.
    BiocompatibilityEvaluated according to relevant ISO 10993-series standards. No biocompatibility concerns were found for materials.
    Reprocessing (Scanner tip sterilization, microbial barrier)Scanner tip is autoclavable. Body Sleeve evaluated for tear resistance, tensile strength, puncture resistance, viral penetration, and synthetic blood penetration. Microbial barrier tested against viral ingress (ASTM F1671/F1671M-22). Reprocessing methods align with AAMI TIR12:2010 and AAMI TIR30:2011+R2016.
    Software Verification and ValidationVerification and validation conducted as per FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and EN 62304:2006+A1:2015.

    2. Sample size used for the test set and the data provenance

    The document mentions an "In vitro caries detection study" and performance testing, but does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide information regarding the number or qualifications of experts used to establish ground truth for any test sets.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    The document does not specify any adjudication method used for establishing ground truth in a test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    The document does not indicate that an MRMC comparative effectiveness study was done involving human readers with and without AI assistance for the L1P-1F device. The focus is on the device's standalone performance compared to a predicate device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The document describes "Performance Testing" including fluorescence scanning, color separation, in vitro caries detection, and spatial resolution. These tests would inherently involve the device's standalone performance to characterize its technical capabilities, which is then concluded to be "as safe and effective as a legally marketed device." While not explicitly termed "standalone performance study," the non-clinical data and performance testing sections describe the device's intrinsic capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document mentions an "in vitro caries detection study," which suggests that the ground truth would likely be established by a recognized method for in vitro caries assessment, such as histology, micro-CT, or a standardized artificial caries model assessed by experts, but this is not explicitly stated.

    8. The sample size for the training set

    The document does not specify the sample size for any training set. Given the nature of the device (an intraoral scanner with integrated fluorescence technology), it's possible that machine learning or AI components might be involved in the "TRIOS Patient Monitoring software," but details on specific training sets are absent.

    9. How the ground truth for the training set was established

    Since no training set details are provided, how its ground truth was established is not mentioned.