Search Results
Found 93 results
510(k) Data Aggregation
(108 days)
67000 Strasbourg France
Re: K253734
Trade/Device Name: Taurus Clip
Regulation Number: 21 CFR 876.4400
Ligator
Device Name: Hemostatic Metal Clip for the GI Tract
Classification Number: 21 CFR §876.4400
Manufacturer:** Boston Scientific
510(k) Number: K222503
Classification Number: 21 CFR §876.4400
Ligator
Device Name: Hemostatic Metal Clip for the GI Tract
Classification Number: 21 CFR §876.4400
Manufacturer:** Boston Scientific
510(k) Number: K222503
Classification Number: 21 CFR §876.4400
The Taurus Clip is indicated for placement within the Gastrointestinal (GI) tract of adult patients for the purpose of:
-
Endoscopic marking
-
Hemostasis for:
- Mucosal/sub-mucosal defects < 3 cm
- Bleeding ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the colon
- Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection
-
Anchoring to affix jejunal feeding tubes to the wall of the small bowel; and
Anchoring to affix fully covered esophageal self-expanding metal stents to the wall of the esophagus
-
As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively
The Taurus Clip is a sterile, single-use endoscopic clip designed for closure, hemostasis, endoscopic marking, and anchoring. The Taurus Clip is designed to be compatible with forward-viewing endoscopes with working channels equal to or greater than 2.8mm.
The Taurus Clip Device consists of two main components: one preloaded titanium Clip Implant and a flexible delivery system, referred to as the "Applier".
The Clip Implant can be used to manipulate the mucosal tissue of the GI tract as desired and then applied (deformed and released) with the handle once in place. The Applier is removed after clip deployment.
N/A
Ask a specific question about this device
(210 days)
Trade/Device Name:* OTSCneo System Set (100.03n-14n, 100.27n-31n)
Regulation Number: 21 CFR 876.4400
Metal Clip for the GI Tract |
| Classification Name | Hemorrhoidal Ligator |
| Regulation Number | 876.4400
- Instrument for flexible endoscopy and for compression of tissue in the gastrointestinal (GI) tract, for hemostasis or for treating gastrointestinal organ wall lesions. Marking of lesions.
- The OTSCneo clip is indicated for clip placement within the GI tract for the purpose of:
- Endoscopic marking
- Hemostasis for:
- Mucosal/submucosal defects < 3 cm
- Bleeding ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticuli in the colon
- Closure of GI tract luminal perforations < 20 mm that can be treated conservatively.
NOTE: The OTSCneo System Set is not indicated for stent fixation. For stent fixation, use the stentfix OTSC® System Set.
The subject device constitutes an over-the-scope clip application device used within the context of flexible endoscopy in the gastrointestinal tract. The clip application capabilities center around a memory-shape alloy clip mounted on a transparent plastic cap. Cap (and clip) are mounted on the distal end of an endoscope. Endoscopic instruments or suction are used via endoscope working channels for mobilization and retraction of target tissue into the cap lumen. With the target tissue inside the cap lumen, the memory-shape alloy clip is applied onto the target tissue using a manual thread mechanism.
Contents of the OTSCneo System Set are:
- Application cap (with spout, application ring, thread, clip mounted on the cap)
- Hand Wheel
- Thread Retriever
In line with cleared indications, clip application with the subject device can be used for hemostasis, treatment of lesions, or marking of lesions in the GI tract.
N/A
Ask a specific question about this device
(114 days)
Zhejiang 311103
China
Re: K252271
Trade/Device Name: Hemoclip
Regulation Number: 21 CFR 876.4400
Hemoclip |
| Common Name: | Hemoclip |
| Regulation class: | Class II |
| Regulation Number: | 876.4400
|
| 510(k) Number: | K211787 |
| Regulation class: | Class II |
| Regulation Number: | 876.4400
The hemoclip is indicated for endoscopic clip placement within the gastrointestinal tract for the purpose of:
- Endoscopic marking,
- Hemostasis for:
- Mucosal/sub-mucosal defects <3cm
- Bleeding ulcers
- Arteries<2mm
- Polyps<1.5cm in diameter
- Diverticula in the colon
- Anchoring to affix jejunal feeding tubes to the wall of the small bowel.
- As a supplementary method, closure of GI tract luminal perforations<20mm that can be treated conservatively.
Hemoclip consists of Release part and Clip part. Clip part consists of Clip and Frap Tube. Release part consists of Spring End, Plastic Coated Spring Tube / Spring Tube and Handle assembly. EO Sterilization and use for single use only.
N/A
Ask a specific question about this device
(224 days)
Re: K250229
Trade/Device Name: Dual Action Tissue Closure Device
Regulation Number: 21 CFR 876.4400
Metal Clip For The GI Tract
Classification: 2
Product Code: PKL
Regulation Number: 876.4400
-----------------------------------|---------|
| Product Code | PKL | PKL | SE |
| Regulation No. | 876.4400
| 876.4400 | SE |
| Class | 2 | 2 | SE |
| Indications for Use | The Dual Action Tissue Closure Device
The Dual Action Tissue Closure Device is intended for use in flexible endoscopy for the compression of tissue in the gastrointestinal tract.
The Dual Action Tissue Closure Device is indicated for clip placement within the gastrointestinal tract for the purpose of:
- Endoscopic marking,
- Hemostasis for
- Mucosal/sub-mucosal defects < 5 cm,
- Bleeding ulcers,
- Polyps < 1.5 cm in diameter,
- Diverticula in the colon,
- Arteries < 2 mm,
- As a supplementary method, closure of GI tract luminal perforations < 20 mm that can be treated conservatively.
The proposed device Dual Action Tissue Closure Device is a sterile, single-use endoscopic clipping device. The Dual Action Tissue Closure Device is intended for use in flexible endoscopy for the compression of tissue in the gastrointestinal tract for adult patient only. It consists of two main components, delivery system and clip component. The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 10⁻⁶ and placed in a sterility maintenance package to ensure a shelf life of 1 year.
N/A
Ask a specific question about this device
(121 days)
accessories |
| Classification/ Regulation: | Class II per 21 CFR 876.5010 | Class II per 21CFR 876.4400
(K201808) |
|---|---|---|---|
| Regulation Number | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 876.4400
The EndoFix™Tissue Fixation System is an endoscopic tissue approximation system intended for anchoring the gallbladder to the stomach or duodenum to aid placement of a luminal apposing metal stent (LAMS) for EUS-guided gallbladder drainage.
The EndoFix™ Tissue Fixation System (TFS) is an endoscopic (through the scope) tissue approximation device consisting of two main components:
- EndoFix Tissue Fixation Device – Enables the approximation of soft tissue in the gastrointestinal (GI) tract using a suture-based implant delivered through an 18 gauge echogenic needle.
- Secure Suture Locking Device – Secures the implant assembly with an implant grade suture lock and cuts the suture once approximation is complete.
The EndoFix TFS is designed to place an implant for endoscopic tissue approximation to anchor the gallbladder to the stomach or duodenum to aid placement of a luminal metal apposing stent (LAMS). During use, the EndoFix Tissue Fixation Device is advanced through a ≥3.2 mm working channel of a commercially available echoendoscope (EUS). Under ultrasound guidance, the EndoFix TFS needle is advanced through the target tissues and the implant subassembly is deployed to approximate the luminal tissues (either transgastric or transduodenal based on clinical requirements). With applied suture tension, a Suture Lock is deployed to secure the tissues using the Secure Suture Locking Device. The implant subassembly and Suture Lock are considered permanent implants.
Each of the respective delivery systems used to deliver the implant assembly/Suture Lock are short term contact devices.
This FDA 510(k) clearance letter for the EndoFix™ Tissue Fixation System does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and a study proving those criteria are met. This document primarily focuses on establishing substantial equivalence to predicate devices for regulatory clearance.
Specifically, it does not include:
- A table of acceptance criteria with reported device performance metrics (e.g., success rates, tensile strength, specific measurements).
- Details about a "test set" for a performance study (sample size, data provenance, ground truth establishment, expert qualifications, or adjudication methods).
- Information on a multi-reader multi-case (MRMC) comparative effectiveness study or any effect size for human reader improvement.
- Distinction between standalone algorithm performance and human-in-the-loop performance, as this is a medical device, not an AI/software-only device in the context of typical AI performance criteria.
- Training set sample size or how its ground truth was established.
However, based on the provided text, I can infer and extract some relevant information regarding performance testing.
Inferable Acceptance Criteria and Reported Device Performance
The document lists categories of performance testing that were conducted with "acceptable results." While specific numerical acceptance criteria and reported performance values are not given, the implication is that the device met internal, pre-defined criteria for these aspects.
A table of acceptance criteria and reported performance cannot be fully constructed from the provided text because specific numerical values for criteria and performance are absent. However, based on the categories mentioned under "Performance Data," we can infer the types of criteria and the general statement of performance:
| Acceptance Criteria Category | General Reported Device Performance |
|---|---|
| Visual Inspection | Acceptable |
| Dimensional Verification | Acceptable |
| Endoscope Compatibility | Acceptable |
| Functionality (approximate/suture tissue) | Acceptable |
| Destructive Testing (Product Integrity, Tensile of joints, withstand minimum forces) | Acceptable |
| Packaging Verification | Acceptable |
| Biocompatibility | Acceptable (per ISO 10993-1) |
| Usability Evaluation | Acceptable |
| Sterilization Validation | Acceptable (per ISO 11135, SAL 10-6) |
| Safety & Performance in Animal Model | Safe and appropriately designed as a pre-stenting tissue fixation device for EUS-guided gallbladder drainage. |
Study Details (as far as extractable):
-
Sample size used for the test set and the data provenance:
- The document mentions "bench testing" and a "preclinical animal study." For bench testing, sample sizes are not specified.
- For the animal study, the model used was a "porcine model," indicating an animal test subject. The number of animals or specific cases within that study is not provided.
- Data provenance: Porcine model (animal study), laboratory/bench for other tests. Retrospective/prospective is not specified, but animal studies are typically prospective experimental designs.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not specify the number or qualifications of experts for establishing ground truth, as this device's performance assessment relies on physical and functional tests, and an animal study, rather than expert interpretation of medical images or data.
-
Adjudication method for the test set:
- Not applicable/Not specified. The performance testing described (bench, animal study) does not involve adjudication processes typically used in clinical imaging or diagnostic studies.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a medical device, not an AI-assisted diagnostic tool subject to MRMC studies. The clearance is for the physical device itself.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm, so "standalone" performance in the context of AI software is not relevant here.
-
The type of ground truth used:
- For bench testing: Engineering specifications, physical measurements, functional demonstrations.
- For the animal study: Observation of safety and performance (e.g., successful tissue approximation, LAMS deployment, absence of adverse events) within the porcine model. This could be considered "outcomes data" in an experimental animal model context.
-
The sample size for the training set:
- Not applicable. This is a physical medical device. There is no "training set" in the context of machine learning algorithms. Its design would be informed by engineering principles and possibly prior animal/cadaver studies, but not a formal 'training set'.
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set mentioned for this type of device.
Ask a specific question about this device
(29 days)
Re: K250842
Trade/Device Name: PADLOCK CLIP EFTR Kit (00713229)
Regulation Number: 21 CFR 876.4400
Full-Thickness Resection Device
Classification Name: Hemorrhoidal Ligator
Classification Number: 21 CFR 876.4400
effective and performs at least as well as the legally cleared predicate device K240274, Class II (21 CFR 876.4400
The PADLOCK CLIP™ EFTR Device is an electrosurgical device that is designed for a flexible endoscope. It is intended for diagnostic tissue acquisition and full thickness resection through the removal of lesions in the colon and rectum.
The PADLOCK CLIP™ EFTR Device is indicated for the resection of lesions less than 3cm in the colon and rectum. Resection sizes may be less than 3cm due to targeted tissue thickness and degree of fibrosis.
The disposable RAPTOR Grasping Device is intended to be used to grasp tissue and/or retrieve foreign bodies, excised tissue and stents during endoscopic procedures.
The TOUCHSOFT Coagulator is used to provide monopolar therapeutic hemostasis and tissue ablation in the gastrointestinal tract through a flexible endoscope by a trained physician or clinician.
The PADLOCK CLIP Endoscopic Full Thickness Resection (EFTR) Device is an electrosurgical, single use tissue resection system. The system is part of a kit that consists of a preloaded, radiopaque, Super Elastic NiTi (Nitinol) single use clip within a mechanical delivery system along with a preloaded snare for tissue resection. The PADLOCK CLIP EFTR Device is mounted over the distal tip of a flexible endoscope. The linking cables run along the outside of the endoscope and are attached to the handle. The PADLOCK CLIP EFTR Device includes a tissue chamber that resides on the distal tip of the endoscope.
The kit's two additional devices, the TOUCHSOFT Coagulator Monopolar Single Use Device, and RAPTOR Grasping Device are used in conjunction with the PADLOCK CLIP EFTR Device through the flexible endoscope's accessory channel. The TOUCHSOFT Coagulator Monopolar Single Use Device (ablation device) will be used to mark a lesion by ablating the lesion's perimeter, thereby indicating the area for resection. Instructions for Use for the TOUCHSOFT Coagulator Monopolar Single Use Device are included within the kit. The RAPTOR Grasping Device will be used to grasp and retract the target lesion into the tissue chamber. Instructions for Use for the RAPTOR are included within the kit.
These devices are used for colorectal lesions that would otherwise not be resectable by standard polypectomy techniques.
The PADLOCK CLIP EFTR Device includes a resection clip made of super elastic nitinol and a snare made of stainless steel. The PADLOCK CLIP EFTR Device also includes handle components molded from ABS material, Stainless Steel drive wires, HDPE catheter, and a scope attachment assembly made of TPE and Polycarbonate.
This document is a 510(k) clearance letter for a medical device (PADLOCK CLIP EFTR Kit). It does not contain information about an AI/ML medical device, clinical study performance metrics, or the specific details typically requested for AI/ML device approval. The 510(k) is for a physical electrosurgical device and states that the proposed device is "the same device with an updated Instructions for Use" as its predicate.
Therefore, many of the requested points related to AI/ML device study design, such as expert ground truth, MRMC studies, training set details, etc., are not applicable to this document.
However, based on the provided text, I can extract the following information about the device's acceptance criteria and the (non-AI/ML) study that proves it meets those criteria:
Acceptance Criteria and Device Performance for PADLOCK CLIP EFTR Kit
This document describes the 510(k) clearance for a physical electrosurgical device, not an AI/ML-driven device. Therefore, the "study" described is verification testing for a physical modification (updated Instruction for Use) rather than a clinical performance study with human readers or AI.
1. Table of Acceptance Criteria and Reported Device Performance
| Testing | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Verification Testing | Clip deploys, and the snare closes on the simulated tissue. | Pass. |
2. Sample Size Used for the Test Set and Data Provenance
The document describes "Verification Testing" on "simulated tissue." It does not specify a numerical sample size for this test. Since it's a physical device verification using simulated tissue and not clinical data, "data provenance" in terms of country of origin or retrospective/prospective is not applicable.
3. Number of Experts Used to Establish Ground Truth and Qualifications
Not Applicable. This is a physical device verification, not a study requiring expert ground truth for diagnostic or interpretative tasks. The "ground truth" here is the objective functional success of the device (clip deploys and snare closes on simulated tissue).
4. Adjudication Method for the Test Set
Not Applicable. This is a physical device verification test. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in image interpretation studies, which is not relevant here. The "Pass" result suggests direct observation of the device's functionality.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. An MRMC study is relevant for AI/ML devices or diagnostic tools where human interpretation is a key component. This document pertains to a physical electrosurgical device used for tissue resection.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Not Applicable. This is a physical device, not an algorithm.
7. The Type of Ground Truth Used
The ground truth used was the functional success of the device's mechanical actions on simulated tissue. Specifically, the acceptance criterion was whether the "Clip deploys, and the snare closes on the simulated tissue."
8. The Sample Size for the Training Set
Not Applicable. This is a physical medical device, not an AI/ML model that requires a training set. The "testing" referred to is product verification testing to ensure the device performs as intended with updated instructions, not the training of a predictive model.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As there is no AI/ML model or training set, there is no corresponding ground truth establishment process.
Ask a specific question about this device
(27 days)
Clip (M00522610); Resolution Clip (M00522611); Resolution Clip (M00522612) Regulation Number: 21 CFR 876.4400
|
| Regulation Number: | 876.4400
|
| Classification Name: | Hemostatic Metal Clip for the GI Tract |
| Regulation Number: | 876.4400
The Resolution™ Clip is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:
- Endoscopic marking
- Hemostasis for:
- Mucosal/sub-mucosal defects < 3 cm
- Bleeding ulcers · Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- · Diverticula in the colon
- Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection
- Anchoring to affix jejunal feeding tubes to the wall bowel; and Anchoring to affix fully covered esophageal self-expanding metal stents to the wall of the esophagus
- As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively.
The Resolution™ Clip is a sterile device consisting of a pre-loaded, radiopaque, single-use, endoscopic clipping device consisting of two main components: the delivery system and the clip.
The delivery system consists of a handle assembly and delivery catheter. The clip delivery system is offered in a 155cm and 235cm working length. The clip consists of a stainless-steel capsule and clip arms, a cobalt chrome yoke, and a styrene tension breaker. The clip is deployed from the delivery system during use. The clip jaws are engineered such that they can be opened and closed up to five times prior to deployment, aiding in repositioning of the clip at the lesion site. Re-opening and closing may be limited by clinical circumstances and patient anatomy, among other factors. The Resolution™ Clip is designed to be compatible with endoscopes with working channels equal to or greater than 2.8 mm. There are no associated accessories included with this device.
The provided text is a 510(k) summary for a medical device (Resolution Clip) and does not describe acceptance criteria or a study proving the device meets acceptance criteria for an AI/ML-driven medical device.
The document discusses a physical medical device, specifically a hemostatic clip for the GI tract. The 510(k) submission is for a modification to an existing device (removal of an outer sheath component). The "performance data" mentioned refers to non-clinical bench testing to confirm that the modified physical device still meets performance requirements (e.g., proper device removal from endoscope, passability within endoscope, absence of endoscope damage).
Therefore, I cannot extract the requested information about acceptance criteria or a study proving an **AI/ML-**driven device meets acceptance criteria from the provided text. The questions regarding sample size, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, training sets, and ground truth establishment for training sets are not applicable to this type of device modification submission.
Ask a specific question about this device
(265 days)
55 Bergen, 5008 Norway
Re: K241005
Trade/Device Name: Haemostasis Clips Regulation Number: 21 CFR 876.4400
Subject Device
Trade name/common name: Haemostasis Clips Regulatory Class: II Regulation Number: 876.4400
Hangzhou AGS MedTech CO., Ltd Trade name: Hemoclip 510(K) Number: K172727 Regulation Number: 21 CFR 876.4400
---------------------------------|------------|
| RegulationNumber | 21 CFR 876.4400
| 21 CFR 876.4400
The The Haemostasis Clips are indicated for endoscopic clip placement within the gastrointestinal tract for the purpose of:
- Endoscopic marking.
- Hemostasis for:
Mucosal/sub-mucosal defects <3cm;
Bleeding ulcers;
Arteries<2mm;
Polyps<1.5cm in diameter;
Diverticula in the colon;
Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. - Anchoring to affix jejunal feeding tubes to the wall of the small bowel;
- As a supplementary method, closure for GI tract luminal perforations<20mm that can be treated conservatively.
The Haemostasis Clips are sterile devices consisting of a pre-loaded, radiopaque, single-use, endoscopic clipping device consisting of two main components: the delivery system and the clip part. The delivery system consists of a handle assembly and delivery catheter assembly. The delivery system is constructed using stainless steel and polyester materials. The delivery system will allow for the device to rotate at the distal end. The delivery system is offered in 1650mm, 1950mm and 2350mm working lengths. The clip part consists of a stainless steel tightening capsule, movable pin and clip arms. The clip is deployed from the delivery system during use. The hemoclip jaws are engineered such that they can be opened and closed up to five times prior to deployment, aiding in repositioning of the clip at the lesion site. Re-opening, closing, and rotation capability may be limited by clinical circumstances and patient anatomy. There are no associated accessories included with this device.
The provided text describes the 510(k) premarket notification for the "Haemostasis Clips" device. It outlines non-clinical testing performed to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets specific acceptance criteria based on clinical outcomes or a multi-reader, multi-case study in the traditional sense of AI/software device evaluation.
The device is a medical instrument (Haemostasis Clips), not a software device or AI system, so many of the requested categories for AI/software evaluation (like MRMC studies, training set details, or ground truth establishment by experts for performance metrics) are not applicable. The evaluations are focused on the physical and performance characteristics of the clips themselves.
Here's a breakdown of the information that can be extracted and which requested points are not applicable in this context:
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly list numerical acceptance criteria for each test. Instead, it states that the tests were performed and the results "Pass" or "meet the requirement". For biocompatibility, specific outcomes like "Non-cytotoxic" are listed.
| Test | Reported Device Performance |
|---|---|
| Appearance | Pass |
| Size | Pass |
| Open and Close | Pass |
| Rotation property | Pass |
| Separation and clamping properties | Pass |
| Clip release force testing | Pass |
| Clip hardness | Pass |
| Corrosion | Pass |
| Connection strength before clip released | Pass |
| Breaking force of the delivery part | Pass |
| Clamping force maintenance | Pass |
| Chemical properties | Pass |
| In vitro cytotoxicity | Non-cytotoxic |
| Skin sensitization | Non-sensitive |
| Intracutaneous reactivity | Non-irritation |
| Acute systemic toxicity | Non-acute systemic toxicity |
| Pyrogen | Non-pyrogenic |
| Hemolysis test | Non Hemolysis reaction |
| Sub chronic toxicity | Non-Sub chronic toxicity |
| Implantation effect | Non-Implantation effect |
| Sterility Assurance Level (SAL) | 10⁻⁶ demonstrated |
| EO residuals | Meets requirements |
| Shelf-life | 3 years (based on accelerated aging) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample sizes for each of the bench or biocompatibility tests. It also does not mention data provenance as these are non-clinical (laboratory/bench) tests on manufactured devices, not data-driven studies from human subjects.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. These are non-clinical, bench and biocompatibility tests for a physical medical device. "Ground truth" in the context of expert consensus on medical images or clinical outcomes is not relevant here. The "ground truth" for these tests are objective, measurable physical and chemical properties and biological reactions determined by standardized test methods.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This concept relates to human expert review and consensus, which is not part of the non-clinical bench or biocompatibility testing described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a physical medical instrument (Haemostasis Clip), not an AI/software device. Therefore, no MRMC study or evaluation of human reader improvement with AI assistance was performed or is relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is a physical medical instrument, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. The "ground truth" for these tests are the inherent physical, chemical, and biological properties measured against established standards (e.g., ISO standards, "Pass" criteria for physical performance).
8. The sample size for the training set
Not applicable. This information pertains to machine learning models, which are not relevant to this device.
9. How the ground truth for the training set was established
Not applicable. This information pertains to machine learning models, which are not relevant to this device.
Ask a specific question about this device
(230 days)
Ohio 44060
Re: K240274
Trade/Device Name: PADLOCK CLIP EFTR Kit (00713229) Regulation Number: 21 CFR 876.4400
Name: | Hemorrhoidal Ligator |
| Regulation Number: | 21 CFR 876.4400
effective, and performs at least as well as the legally marketed predicate device K170867, Class II (21 CFR 876.4400
The PADLOCK CLIP™ EFTR Device is a device that is designed for a flexible endoscope. It is intended for diagnostic tissue acquisition and full-thickness resection through the removal of lesions in the colon and rectum.
The PADLOCK CLIP™ EFTR Device is indicated for the resection of lesions less than 3cm in the colon and rectum. Resection sizes may be less than 3cm due to targeted tissue thickness and degree of fibrosis.
The disposable RAPTOR Grasping Device is intended to be used to grasp tissue and/or retrieve foreign bodies, excised tissue and stents during endoscopic procedures.
The TOUCHSOFT Coagulator is used to provide monopolar therapeutic hemostasis and tissue ablation in the gastrointestinal tract through a flexible endoscope by a trained physician or clinician.
The PADLOCK CLIP Endoscopic Full Thickness Resection (EFTR) Device is an electrosurgical, single use tissue resection system. The system is part of a kit that consists of a preloaded, radiopaque, Super Elastic NiTi (Nitinol) single use clip within a mechanical delivery system along with a preloaded snare for tissue resection. The PADLOCK CLIP EFTR Device is mounted over the distal tip of a flexible endoscope. The linking cables run along the outside of the endoscope and are attached to the handle. The PADLOCK CLIP EFTR Device includes a tissue chamber that resides on the distal tip of the endoscope.
The kit's two additional devices, the TOUCHSOFT Coagulator Monopolar Single Use Device, and RAPTOR Grasping Device are used in conjunction with the PADLOCK CLIP EFTR Device through the flexible endoscope's accessory channel. The TOUCHSOFT Coagulator Monopolar Single Use Device (ablation device) will be used to mark a lesion by ablating the lesson's perimeter, thereby indicating the area for resection. Instructions for Use for the TOUCHSOFT Coagulator Monopolar Single Use Device are included within the kit. The RAPTOR Grasping Device will be used to grasp and retract the target lesion into the tissue chamber. Instructions for Use for the RAPTOR are included within the kit. These devices are used for colorectal lesions that would otherwise not be resectable by standard polypectomy techniques.
The PADLOCK CLIP EFTR Device includes a resection clip made of super elastic nitinol and a snare made of stainless steel. The PADLOCK CLIP EFTR Device also includes handle components molded from ABS material, Stainless Steel drive wires, HDPE catheter, and a scope attachment assembly made of TPE and Polycarbonate.
The provided document is a 510(k) summary for the PADLOCK CLIP EFTR Kit, which is a medical device for endoscopic full-thickness resection. The document outlines verification, validation, animal, and other testing to demonstrate substantial equivalence to a predicate device.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance:
| Category | Testing | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Simulated Use | Endoscope Detachment Force | Meets Design Requirements | Pass |
| Transparent Housing Verification | Meets Design Requirements | Pass | |
| Clip and Snare Deployment | Meets Design Requirements | Pass | |
| Over the Scope Attachment Verification | Meets Design Requirements | Pass | |
| Safety Cap Verification | Meets Design Requirements | Pass | |
| Handle Stay Visual Verification | Meets Design Requirements | Pass | |
| Safety Cap Visual Verification | Meets Design Requirements | Pass | |
| Snare Proximal to Clip Visual Verification | Meets Design Requirements | Pass | |
| Packaging Requirements | Peel Seal Strength Verification | Meets Design Requirements | Pass |
| Clip Deployment Force | Meets Design Requirements | Pass | |
| MAX Handle Snare Retraction Force | Meets Design Requirements | Pass | |
| Snare Continuity Test | Meets Design Requirements | Pass | |
| Pusher Compression Force | Meets Design Requirements | Pass | |
| Spool Compression Force | Meets Design Requirements | Pass | |
| Clip Pusher OD Measurement | Meets Design Requirements | Pass | |
| Device Working Length Measurement | Meets Design Requirements | Pass | |
| Tissue Chamber Depth Measurement | Meets Design Requirements | Pass | |
| Tissue Chamber Distal ID Measurement | Meets Design Requirements | Pass | |
| Component Force and Functional Requirements | Active Cord Detachment Force | Meets Design Requirements | Pass |
| Catheter (110302) to Housing Force | Meets Design Requirements | Pass | |
| Catheter (110052) to Housing Force | Meets Design Requirements | Pass | |
| Catheter (110303) to Pusher Force | Meets Design Requirements | Pass | |
| Drive Cable to Snare Loop Force | Meets Design Requirements | Pass | |
| Tissue Retention Force | Meets Design Requirements | Pass | |
| Clip Closure Force | Meets Design Requirements | Pass | |
| Actuator to Catheter (110302) Force | Meets Design Requirements | Pass | |
| Actuator to Catheter (110052) Force | Meets Design Requirements | Pass | |
| Drive Cable to Connector Force | Meets Design Requirements | Pass | |
| Scope Attachment to Housing Force | Meets Design Requirements | Pass | |
| Legibility of Markings | Meets Design Requirements | Pass | |
| Validation Testing | Clip Deployment and Snare Resection Validation | Meets Design Requirements | Pass |
| Straight and Articulated Endoscope Position Use Validation | Meets Design Requirements | Pass | |
| Visualization in the Lower GI Tract Validation | Meets Design Requirements | Pass | |
| Endoscope Compatibility Validation | Meets Design Requirements | Pass | |
| Safety Feature for Deployment of Clip Validation | Meets Design Requirements | Pass | |
| Animal Test - Design Validation | The PADLOCK CLIP™ EFTR Device Clip must facilitate healing after the removal of lesions up to 3cm in the GI tract | Meets Design Requirements | Pass |
| Additional Testing | MRI Testing | Meets the Requirements of ASTM F2052-21, ASTM F2182-19, ASTM F2119-2013 (17) | Pass |
| Electrical testing | Meets the Requirements of IEC 60601-1:2005 Ed.3 + A1; A2, IEC 60601-2-2:2017 Ed.6 and, IEC 60601-2-18:2009 Ed.3 | Pass | |
| Biocompatibility | Meets the requirements of ISO 10993-5:2009 Ed.3, 10993-12:2021 Ed.5, 10993-11:2017 Ed.3 & 10993-23:2021 Ed.1, 10993-1:2018, 10993-6 Ed. 3, 10993-18:2020 | Pass | |
| Human Factors | Meet Predetermined Critical Tasks | Pass | |
| EMC Testing | Meets the Requirements of 60601-1-2:2014+A1:2020 and 60601-4-2:2016; Meets the Requirements of EMC clauses 201.17/202 of ANSI AAMI IEC 60601-2-2:2017; Meets the Requirements of EMC clauses of 201.17/202 of IEC 60601-2-18 Ed. 3 2009-08 | Pass |
2. Sample Size Used for the Test Set and the Data Provenance:
The document summarizes the types of tests performed (Simulated Use, Packaging Requirements, Component Force and Functional Requirements, Validation Testing, Animal Test - Design Validation, MRI Testing, Electrical Testing, Biocompatibility, Human Factors, EMC Testing) and their outcomes ("Pass"). However, it does not provide details on the specific sample sizes used for each test set within these categories.
Similarly, the document does not mention the data provenance (e.g., country of origin, retrospective or prospective nature) for any of the non-clinical tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
The document does not specify the number of experts used or their qualifications for establishing ground truth for any of the tests. The acceptance criteria are generally phrased as "Meets Design Requirements" or conformity to international standards, implying compliance with engineering specifications and regulatory benchmarks rather than a human-expert-derived ground truth.
4. Adjudication Method for the Test Set:
The document does not describe any adjudication method like 2+1 or 3+1. The tests listed are primarily performance-based and compliance-based, with results indicated as "Pass," suggesting a direct comparison to pre-defined specifications rather than expert consensus requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:
The document does not mention or describe a multi-reader multi-case (MRMC) comparative effectiveness study. The PADLOCK CLIP EFTR Kit is a mechanical and electrosurgical device, not an AI-powered diagnostic or assistive tool, so such a study would not be relevant in this context.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done:
This question is not applicable as the PADLOCK CLIP EFTR Kit is a physical medical device, not an algorithm or software requiring standalone performance testing in the typical sense of AI/software. Its performance is evaluated through a series of engineering, functional, and biological tests.
7. The Type of Ground Truth Used:
The ground truth for most of the tests appears to be engineering design requirements, predefined specifications, and compliance with established international standards (e.g., ASTM, IEC, ISO). For the animal study, the "ground truth" for success was the device's ability to "facilitate healing after the removal of lesions up to 3cm in the GI tract," which would be assessed by veterinary or pathology evaluation of the animal tissue post-procedure.
8. The Sample Size for the Training Set:
The concept of a "training set" is typically associated with machine learning or AI models. Since the PADLOCK CLIP EFTR Kit is a physical medical device, there is no mention of a training set in this context. The non-clinical tests are verification and validation efforts, not machine learning model training.
9. How the Ground Truth for the Training Set Was Established:
Again, as there is no "training set" in the context of this device, this question is not applicable. The ground truth for the device's performance is established through adherence to design specifications and regulatory standards.
Ask a specific question about this device
(60 days)
Tuebingen, 72076 Germany
Re: K241858
Trade/Device Name: BARS Set (100.60) Regulation Number: 21 CFR 876.4400
GI Tract |
| Classification Name | Hemorrhoidal ligator |
| Regulation Number | 876.4400
The BARS Set is indicated for clip placement within the context of flexible endoscopy in the GI tract for the purpose of:
- Hemostasis for:
- Mucosal/submucosal defects < 3 cm
- Arteries < 2 mm
- Treatment of luminal perforations (< 20 mm) and / or anastomoses (< 50 mm)
- Lumen reduction
NOTE: The BARS Set is not indicated for stent fixation, For stent fixation, use the stentfix OTSC System Set.
The subject device constitutes an over-the-scope clip application device used within the context of flexible endoscopy in the gastrointestinal tract. The over-the-scope clip application capabilities center around a memory-shape alloy clip mounted on a transparent plastic cap. Cap (and clip) are mounted on the distal end of an endoscope. Endoscopic instruments are provided in the BARS Set to facilitate insertion and for mobilization of target tissue into the cap lumen. With the target tissue inside the cap lumen, the memory-shape alloy clip is applied onto the target tissue using a thread mechanism. Contents of the BARS Set are: - BARS (readily assembled unit of memory-shape alloy clip, transparent plastic cap, thread deployment components, working channels, and guide elements)
- Endoscopic instruments (BARS Anchor silver, BARS Anchor black, Insertion Balloon, Space Keeper Balloon, Guide Wire)
The provided text is a 510(k) Premarket Notification from the FDA for a device called the BARS Set. It describes the device, its indications for use, and a summary of non-clinical tests. However, the document explicitly states "Not applicable - no clinical testing submitted for demonstration of substantial equivalence."
Therefore, I cannot provide the detailed information requested regarding acceptance criteria and a study that proves the device meets those criteria from this document alone. The questions posed in your prompt (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set details) are all related to clinical or performance studies with human data/expert review, which were not submitted for this 510(k).
The document focuses on demonstrating substantial equivalence through bench testing and comparison to predicate and reference devices, rather than a new clinical study.
Here's what I can extract based on the provided text, primarily regarding the non-clinical bench testing:
Non-Clinical Bench Testing Summary (based on the device's substantial equivalence pathway):
-
Acceptance Criteria for Non-Clinical Bench Testing & Reported Device Performance: This information is not presented in a table with specific criteria and results. The document states:
- "Performance bench testing settings investigated key aspects of device safety and effectiveness."
- "The established precedent from these devices were met as evidenced by application of acceptance criteria and / or comparison to their respective performance bench testing substantial equivalence."
- The tests mentioned for the subject device are:
- Functional capability and robustness
- Endoscope compatibility
- Clip application ring strength
- Space keeper balloon capability and robustness
- Usability
- MRI suitability
- However, specific quantitative acceptance criteria (e.g., "Clip application force > X N") and the numerical results the BARS Set achieved are not detailed in this public summary. The document only confirms that the device "met" these criteria or "the established precedent" set by reference devices.
-
Sample Size Used for the Test Set and Data Provenance:
- Not applicable as no clinical test set (i.e., human subject data for performance claims) was submitted. The tests mentioned were bench tests.
- The provenance of any data used for bench testing (e.g., whether conducted in the US, Germany, etc.) is not specified.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Experts:
- Not applicable as no clinical test set or expert-adjudicated ground truth was submitted for device performance evaluation. The "ground truth" for the non-clinical tests would be defined by engineering specifications and physical measurements.
-
Adjudication Method for the Test Set:
- Not applicable as no clinical test set requiring expert adjudication was submitted.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, explicitly stated: "Not applicable - no clinical testing submitted for demonstration of substantial equivalence."
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a mechanical medical device, not an AI/algorithm.
-
The Type of Ground Truth Used:
- For the non-clinical bench tests, the "ground truth" would be the engineering specifications and physical properties of the device and its components, and comparison to the performance of legally marketed predicate/reference devices. This is not clinical ground truth (e.g., pathology, outcomes data).
-
The Sample Size for the Training Set:
- Not applicable. This is not an AI/ML device that requires a training set.
-
How the Ground Truth for the Training Set was Established:
- Not applicable.
In summary: The provided FDA 510(k) summary focuses on demonstrating "substantial equivalence" of the BARS Set to existing predicate and reference devices through non-clinical bench testing, without the need for new clinical data or human-in-the-loop performance studies like those typically conducted for AI/software devices. The document confirms that the device "met" acceptance criteria related to its physical and functional characteristics based on these bench tests.
Ask a specific question about this device
Page 1 of 10