    Prima™ MR Dual Vacuum-Assisted Breast Biopsy System Vacuum Canister

    Regulation Number: 21 CFR 876.1075

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mammotome Prima™ MR Vacuum-Assisted Breast Biopsy System is intended to acquire breast tissue from an imaged abnormality during MRI-guided breast biopsy procedures for histologic examination.

    The extent of a histologic abnormality cannot always be reliably determined from an imaged appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

    Device Description

    The Mammotome Prima™ MR Dual Vacuum-Assisted Breast Biopsy System is a battery operated, vacuum-assisted breast biopsy device designed to conduct minimally invasive procedures in the MRI environment. It is intended for use in an MR environment. The system is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The system, used with MR imaging modality, facilitates the diagnostic removal of tissue with fluid management through a combination of vacuum and radial cutting functions. The system is comprised of reusable and disposable components. Reusable components include: a battery-operated cordless control module for in-room use, two batteries and a battery charger, a foot pedal switch, and an optional tray. Disposable components include: an 8 gauge probe; a universal targeting set consisting of a targeting cube, targeting sleeve, obturator cap; and obturator with integrated insertion tip and removable obturator handle.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    ZeniCore Automatic Biopsy Device 16ga x 16cm with Coaxial Introducer Kit)

    Regulation Number: 21 CFR 876.1075
    Gastroenterology-urology biopsy instrument
    Device Classification: 2
    Product Code: KNW
    Regulation: 21 CFR 876.1075
    Same |
    | Device Classification Name | Instrument, biopsy | Same | Same | Same |
    | Regulation | 876.1075

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZeniCore™ Automatic Biopsy Instrument and optional Co-Axial Introducer Needle is a single-use device indicated for obtaining core samples of soft tissue masses via percutaneous needle biopsy. It is not intended for use in bone.

    Device Description

    The ZeniCore™ Automatic Biopsy Instrument and optional, coaxial Introducer needle is a single-use medical device intended for obtaining core soft tissue samples via percutaneous needle biopsy. It is not intended for use in bone lesions. The ZeniCore™ Automatic Biopsy Instrument is a sterile, single patient use device comprised of a spring-loaded biopsy needle fitted into a plastic handle permitting single handed specimen collection than can be used with an optional corresponding coaxial Introducer Needle. The position of the needle may be visualized by imaging techniques such as X-ray, fluoroscopy, and ultrasound. The needle has an echogenic tip with numerically ordered centimeter markings to facilitate precise depth placement. The throw length of the needle when collecting a specimen is 25mm. There are five available needle gauge (G) sizes (12G, 14G, 16G, 18G and 20G) which are identified by a color-coded actuator button. The device is sold either standalone, or as a kit, with an optional, corresponding coaxial needle (cleared under K980004).

    The optional Co-Axial Introducer Needle system includes an outer cannula and an inner stylet which terminate in a trocar tip and serves to support acquisition of tissue samples. Each Co-Axial Introducer Needle has an adjustable spring clip that serves as a locking mechanism to secure the position of the outer co-axial needle relative to the skin which ensures the biopsy needle does not accidentally slide forward or backward during the procedure, thereby maintaining the accuracy and safety of the biopsy. The Luer cap of the stylet is color coded to match the corresponding biopsy instrument it is sized to fit.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252681

    Validate with FDA (Live)

    Date Cleared
    2025-12-12

    (109 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: EnCor EnCompass™ Breast Biopsy and Tissue Removal System
    Regulation Number: 21 CFR 876.1075
    Classification Panel:** General & Plastic Surgery
    Regulatory Class: II
    Regulation Number: 876.1075

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EnCor EnCompass™ Breast Biopsy and Tissue Removal System is indicated to acquire breast tissue for histologic examination with partial or complete removal of the abnormality.

    Device Description

    The subject device, the EnCor EnCompass™ Breast Biopsy and Tissue Removal System, is a console-based vacuum-assisted breast biopsy device. It is intended to obtain breast tissue samples for diagnostic or histologic analysis and may be used under Ultrasound, Stereotactic, or MR imaging guidance. The system consists of a console, handheld driver, foot switch, biopsy probes, canister liner bags, MRI extension tubing, a stereotactic adapter, MRI introducer sets, and spare sample containers. Replacement parts and components of the console, driver, and foot switch that are also sold separately include a reusable vacuum canister, exhaust filter, driver extension cable, foot switch cable, and power cable.

    The EnCor EnCompass™ Breast Biopsy and Tissue Removal System consists of multiple devices that are intended to be used together to obtain breast tissue samples. This includes the EnCor EnCompass™ Console, EnCor EnCompass™ Driver, and EnCor EnCompass™ Foot Switch. The console, driver, and foot switch are designed to provide a user interface and to accept inputs used for specific procedures. The console contains a vacuum system and touchscreen monitor. The driver contains multiple buttons to control procedure functionality and is compatible for use with Ultrasound, Stereotactic, and Magnetic Resonance Imaging (MRI) procedures. The foot switch also contains buttons and a pedal to control procedure functionality and has both wired and wireless connectivity options. They are intended to be used with EnCor EnCompass™ Probes, Canister Liner Bags, MRI Extension Tubing, VisiLoc™ MRI Introducer Sets, and Stereotactic Adapter, which are sold separately. The Probes are the Applied Part of the EnCor EnCompass™ Breast Biopsy and Tissue Removal System.

    The EnCor EnCompass™ Probe contains a tubular cutter for tissue acquisition and either a sharp trocar tip at the distal end for insertion or a blunt tip at the distal end for use in areas with sensitive structures. A detachable tissue collection chamber is located on the proximal end of the device. The tissue collection chamber incorporates both a suction line for acquisition and transport of tissue samples and a saline line for optional lavage functionality. A Y-valve luer lock is provided for introducing anesthetic to the biopsy site. Compatible breast tissue markers which are sold and packaged separately may be inserted through the proximal end of the device.

    The integrated tubing cassette provides vacuum to an EnCor EnCompass™ Probe during a biopsy procedure.

    The EnCor EnCompass™ Probe sample container collects samples as they are acquired and transported during the biopsy procedure. The basket allows for drainage of fluid that is acquired during sampling, vacuuming, and lavaging of the cavity. The basket allows visualization of the samples during and after the procedure and is detachable for sample retrieval post procedure.

    The marker adapter allows for marker placement using a marker deployment device through the back of the probe.

    The EnCor EnCompass™ Canister Liner Bag collects fluid waste generated during the procedure.

    The EnCor EnCompass™ Stereotactic Adapter provides a mechanical interface between the combined EnCor EnCompass™ Driver and EnCor EnCompass™ Probe and the stereotactic biopsy table. The EnCor EnCompass™ Stereotactic Adapter is reusable and affixes directly to the EnCor EnCompass™ Probe with a T-shape fastener and two pins and the EnCor EnCompass™ Driver with rear sliding latches.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252646

    Validate with FDA (Live)

    Date Cleared
    2025-10-24

    (64 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name:** Single-Use Fine Needle Biopsy (FNB) device (NA-U210H)
    Regulation Number: 21 CFR 876.1075
    Classification Name: Gastroenterology-Urology Biopsy Instrument

    • Product Code: FCG
    • Regulation Number: 21 CFR 876.1075
      Gastroenterology-Urology Biopsy Instrument | Gastroenterology-Urology Biopsy Instrument | Same |
      | Regulation number | 876.1075
      | 876.1075 | Same |
      | Regulatory Class | II | II | Same |
      | Product Code | FCG | FCG | Same |
      | Classification
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single-use Fine Needle Biopsy (FNB) device NA-U210H has been designed to be used with an ultrasound endoscope for sampling of submucosal and extramural lesions within or adjacent the gastrointestinal tract.

    Device Description

    The Single-Use Fine Needle Biopsy (FNB) Device NA-U210H (SecureFlex) is a sterile, single-use fine needle biopsy device intended for use with an ultrasound endoscope to sample submucosal and extramural lesions within or adjacent to the gastrointestinal tract. It consists of:

    • Biopsy Needle Assembly: Handle, needle (19G, 22G, or 25G), and coiled sheath.
    • Aspiration System: VACLOK syringe and stopcock (FDA-cleared under K994253).

    The device features a bifurcated needle tip for enhanced tissue acquisition, echo-enhanced regions for ultrasound visibility, and is compatible with Olympus GF and TGF endoscopes with ≥2.8 mm channel diameter.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252183

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-24

    (75 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Pennsylvania 19103

    Re: K252183
    Trade/Device Name: Resitu Slider 09 (RESL09)
    Regulation Number: 21 CFR 876.1075

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Resitu Slider 09 is intended to provide breast tissue for diagnostic analysis of imaged abnormalities.

    Device Description

    Resitu Slider 09 (RESL09) is a minimally invasive, vacuum assisted device, intended to provide large intact breast tissue samples for histologic examination of imaged abnormalities. The device is provided sterile and is for single-use. Resitu Slider 09 is hand-held (vacuum is created in the handle) and is part of a monopolar high frequency electrosurgical system.

    The intended sample site is reached via a skin incision followed by blunt penetration under ultrasound guidance. The sample capture relies on a combination of vacuum that fixates the sample and cutting with a circular knife. While cutting around the sample area, the vacuum allows the whole tissue sample to enter into the tube of the device. An electrosurgical electrode, designed to mimic a mechanical cutting blade, severs the top base of the breast tissue sample from surrounding area. The device is withdrawn with the intact breast tissue sample inside the tube.

    The device functions as an active accessory by passing energy that is controlled by the electrosurgical generator to the tissue, similar to other electrosurgical tools.

    The Resitu Slider 09 electrosurgical system consists of the following:

    • Active Accessory (Monopolar)
    • Electrosurgical Unit
    • Active Accessory Cord
    • Neutral Electrode
    • Neutral Electrode Connection Cord
    • Foot Switch (Pedal)

    The device consists of a plastic handle, to which a transparent tube with a circular stainless steel knife at the end is attached. Inside the tube is a rod that creates a blunt end when in its outer position. The rod can be pulled back to expose the circular knife and create a vacuum, which will assist in excising the tissue, inside the tube.

    In the center of the tube is an electrode, a diathermic knife, which can be pushed forward, backwards, and also turn 360 degrees, thereby enabling a radial cutting from the center.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250994

    Validate with FDA (Live)

    Date Cleared
    2025-08-21

    (142 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    92612

    Re: K250994
    Trade/Device Name: ClearTip FNA and FNB Types
    Regulation Number: 21 CFR 876.1075
    Classification Name : Gastroenterology-Urology Biopsy Instrument

    • Product Code : FCG
    • Regulation Number : 876.1075
      Gastroenterology-urology Biopsy instrument |
      | Product Code | FCG | FCG | FCG |
      | Regulation Number | 21 CFR 876.1075
      | 21 CFR 876.1075 | 21 CFR 876.1075 |
      | Indications for Use | The ClearTip FNA and FNB Types are
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClearTip FNA and FNB Types are intended for Ultrasonically Guided Fine Needle Aspiration, (FNA) of submucosal and extraluminal lesions of the Gastrointestinal Tract, (e.g., lymph nodes, abnormal tissue in the mediastinum).

    Device Description

    The ClearTip is a manually operated endoscopic instrument intended to obtain tissue specimens of gastrointestinal tract (=digestive tract). The subject device mainly consists of a handle unit with an insertion part, a syringe, a stopcock, and a connector. ClearTip is offered in various needle gauges for insertion and aspiration. The available needle gauge sizes of 19, 22, and 25 gauge. The Needle which is dimpled for ultrasonic visualization is advanced into the target site for aspiration. The device allows for adjustment of the length of the coil sheath and the needle to enable the user to adjust for the working length of the endoscope and to control needle insertion depth. It is preloaded with a stylet to aid in inserting the Needle which is removed for injection and aspiration. This device passes through the working channel of endoscope, and the average contact time with mucosa of the human gastrointestinal tract is less than 1 hour. This device is supplied sterile for single-patient use and shall be not reused or re-sterilized.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a medical device, and as such, it does not detail acceptance criteria or a study that would demonstrate the device meets those criteria in the way typically seen for AI/software-as-medical-device (SaMD) products. This document describes a traditional medical device (a biopsy instrument), not an AI system.

    The "acceptance criteria" and "device performance" mentioned in your request are usually quantifiable metrics (e.g., sensitivity, specificity, AUC) for SaMD products, often against a ground truth for diagnostic accuracy. For a physical device like the ClearTip FNA and FNB Types, acceptance criteria typically relate to physical properties, functionality, safety, and compatibility, which are assessed through non-clinical (engineering, bench) tests rather than clinical performance studies against a diagnostic "ground truth."

    Therefore, I will extract the information provided based on the context of a physical medical device submission, focusing on the nearest equivalents to your requested categories.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a physical device, the "acceptance criteria" are implied by the performance standards for the non-clinical tests conducted. The reported "device performance" is typically that the device met these standards. The document does not provide specific numerical acceptance criteria or performance metrics for each test, but rather lists the types of tests performed.

    Acceptance Criteria Category (Implied)Reported Device Performance (Implied)
    Sterility ValidationValidated (device is sterilizable by EO)
    Shelf-Life3 Years (device maintains integrity for this period)
    AppearanceMet (device's visual characteristics are acceptable)
    DimensionsMet (device dimensions are within specified limits)
    OperabilityMet (device functions as intended)
    ElasticityMet (device's elastic properties are acceptable)
    Bending StrengthMet (device withstands specified bending forces)
    Pull-outMet (device components withstand specified pull-out forces)
    Tensile ForceMet (device components withstand specified tensile forces)
    BiocompatibilityMet (device materials are biocompatible)
    Endoscope CompatibilityMet (device is compatible with specified endoscope channels)

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not describe a "test set" in the context of clinical data for diagnostic performance. The "tests" mentioned are non-clinical (bench/engineering) tests conducted on the physical device itself. Therefore, information on sample size for a "test set" or "data provenance" (country of origin, retrospective/prospective) is not applicable or provided given the nature of the device and the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is not an AI/SaMD product requiring expert-established ground truth for diagnostic accuracy. The "ground truth" for a physical device's performance often relates to engineering specifications and regulatory standards.

    4. Adjudication Method for the Test Set

    Not applicable. No "test set" requiring adjudication in a clinical context is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. An MRMC study is typically for evaluating diagnostic performance of imaging or AI systems with human readers. This clearance is for a physical biopsy instrument.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This is not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's acceptance is based on engineering specifications, material standards, and regulatory requirements (e.g., for sterility, biocompatibility, mechanical strength). It is not pathology, outcomes data, or expert consensus in the diagnostic sense.

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve a "training set" as it is not an AI/machine learning system.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243407

    Validate with FDA (Live)

    Date Cleared
    2025-07-29

    (270 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K243407**
    Trade/Device Name: Forcyte Autograft Harvest Kit
    Regulation Number: 21 CFR 876.1075
    Gastroenterology-urology biopsy instrument

    Classification Name: Instrument, Biopsy

    Regulation Number: 876.1075

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FORCYTE Autograft Harvest Kit is intended to harvest cancellous bone and bone marrow for combination with commercially available allograft or bone void fillers for filling bony voids or gaps that are not intrinsic to the stability of the bony structure.

    Device Description

    The Subject device is a kit comprised of a bone and marrow harvesting needle, collection jar, suction tubing, drill-tip guide pin, and bone void filler hydration chamber. The FORCYTE system harvests autologous bone and marrow for use as a bone void filler.

    AI/ML Overview

    Based on the provided FDA 510(k) clearance letter for the Forcyte Autograft Harvest Kit (K243407), the device is a Gastroenterology-Urology Biopsy Instrument. The letter explicitly states that "No clinical evaluations were conducted" for this device. This means that the 510(k) clearance relies solely on non-clinical (bench) testing, and therefore, it is not possible to provide acceptance criteria or a study description related to AI algorithm performance or diagnostic accuracy, human reader performance, or multi-reader multi-case (MRMC) comparative effectiveness studies from this document.

    The information requested in the prompt (AI algorithm performance data, expert ground truth, MRMC studies, etc.) is typically associated with software as a medical device (SaMD) or AI/ML-enabled devices that perform diagnostic or prognostic functions based on image analysis, pattern recognition, etc. The Forcyte Autograft Harvest Kit, however, is a physical instrument for harvesting bone and bone marrow.

    Therefore, many of the requested items are not applicable to the information provided in this 510(k) clearance letter. I will address the relevant points based on the document's content and explicitly state when information is not available.


    Analysis of the Forcyte Autograft Harvest Kit (K243407) based on the Provided 510(k) Clearance Letter:

    The Forcyte Autograft Harvest Kit (K243407) is a physical medical device, not an AI/ML-enabled device or software medical device. The 510(k) clearance for this device was based on non-clinical (bench) testing and substantial equivalence to a predicate device, not on clinical performance studies involving human readers or AI algorithms.

    Therefore, most of the requested information regarding AI acceptance criteria, clinical study design (sample size, expert ground truth, MRMC), and training data is not applicable to this specific device and the provided document.

    Here's an attempt to address each point based on the provided text, clarifying what information is present and what is absent:


    1. A table of acceptance criteria and the reported device performance

    Based on the document, the acceptance criteria and performance data are related to the physical characteristics and functionality of the harvesting kit, not diagnostic accuracy.

    Acceptance Criteria and Reported Device Performance (Non-Clinical)

    Performance Metric (Type of Test)Acceptance Criteria (Implicit)Reported Device Performance (Implicit)
    Simulated Use TestingSatisfactory OperationAll testing had satisfactory results.
    Tensile TestingWithin Engineering SpecificationsAll testing had satisfactory results.
    Compression TestingWithin Engineering SpecificationsAll testing had satisfactory results.
    Torque TestingWithin Engineering SpecificationsAll testing had satisfactory results.
    Flexural TestingWithin Engineering SpecificationsAll testing had satisfactory results.
    Vacuum TestingAchieve/Maintain Required VacuumAll testing had satisfactory results.
    BiocompatibilityMeet ISO 10993 StandardsIndicating safety and effectiveness.
    Sterilization/Shelf LifeMaintain Sterility & IntegrityAll testing had satisfactory results.

    Note: The document states "All testing had satisfactory results" but does not provide specific numerical acceptance criteria or detailed numerical results for each test. The acceptance criteria are implicitly that the device performs as intended and meets safety requirements.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in terms of number of units or test runs for the bench tests. The document only states "bench testing including simulated use testing, tensile, compression, torque, flexural, and vacuum testing were conducted."
    • Data Provenance (Country of Origin, Retrospective/Prospective): Not specified. As these are bench tests, the concept of "retrospective" or "prospective" data collection from patients is not applicable. The tests would have been performed in a laboratory setting. No geographical origin of the testing data is mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This device is a biopsy instrument, not an imaging or diagnostic AI device requiring expert interpretation for ground truth establishment. Bench testing does not typically involve "experts" establishing a "ground truth" in the diagnostic sense; rather, it involves engineers and quality control personnel assessing physical performance against specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As no expert interpretation or diagnostic "ground truth" was established, there was no need for an adjudication method. Bench testing results are typically assessed against pre-defined engineering and material specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states: "No clinical evaluations were conducted." Therefore, no MRMC study, human reader improvement, or AI assistance was involved in the clearance. This type of study is irrelevant for a physical bone harvesting instrument.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm or software. There is no "standalone" algorithm performance to evaluate.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable in the diagnostic sense. For a physical device, "ground truth" is established by manufacturing specifications, engineering design, and material properties. The device's performance is measured against these established physical and functional parameters (e.g., tensile strength, proper vacuum seal, successful harvesting of material, biocompatibility).

    8. The sample size for the training set

    • Not Applicable. This is not an AI/ML device that requires a "training set" of data.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for an AI/ML algorithm, no ground truth was established for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K244018

    Validate with FDA (Live)

    Date Cleared
    2025-07-22

    (207 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    China

    Re: K244018
    Trade/Device Name: Disposable Biopsy Needle
    Regulation Number: 21 CFR 876.1075
    Instrument, Biopsy |
    | Review panel | Gastroenterology/Urology |
    | Regulation number | 21 CFR 876.1075
    ® Aspirating Needle (Chiba) | Promisemed Fine Biopsy Needle | - |
    | Classification Regulation | 876.1075
    | 876.1075 | 876.1075 | Same |
    | Product Code | KNW | KNW | KNW | Same |
    | 510(k) Number | K244018

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Biopsy Needle is intended for use in soft tissue biopsies such as liver, kidney, breast, thyroid, prostate, spleen, pancreas, lymph nodes, lung and various soft tissue tumors. It is not intended for use in bone.

    Device Description

    Disposable Biopsy Needle is intended for use in obtaining an aspirate for cytology. Disposable Biopsy Needle has five models: BN-MAR-1, BN-MAR-2, BN-MAR-3, BN-MAR-4, BN-MAR-8. The Disposable Biopsy Needle consists of operating handle of Inner stylet, operating handle of cutting cannula, cutting cannula, inner stylet, depth stopper and protective sheath. The models differ as follows:

    • Cutting cannula tip type: The cannula tip type of BN-MAR-1, BN-MAR-2, BN-MAR-3 is chiba. The cannula tip type of BN-MAR-4, BN-MAR-8 is greene.
    • Operating handle designs:
      • BN-MAR-1, BN-MAR-3, BN-MAR-4 Models:
        a. BN-MAR-1, BN-MAR-3, and BN-MAR-4 have the same operating handle, including the operating handle of inner stylet and operating handle of cutting cannula.
        b. The Operating handle of Inner stylet connector connects to the Operating handle of cutting cannula connector via a straight snap-fit design, allowing for easy disconnection with one hand to remove the Inner stylet.
      • BN-MAR-2, BN-MAR-8 Models:
        a. The Operating handle of Inner stylet of BN-MAR-2 connects to the Operating handle of cutting cannula via a straight snap-fit design connector, allowing easy disconnection with one hand to remove the Inner stylet.
        b. The Operating handle of inner stylet of BN-MAR-8 connects to the Operating handle of cutting cannula via a female Luer taper and male Luer thread taper, which can meet the sealing performances in clinical applications.
    • Bevel Angle of the Cutting cannula tip: There are three kinds of bevel angle of the Cutting cannula tip, which are α, β, δ. The bevel angle of the cutting cannula tip of BN-MAR-1 and BN-MAR-2 is α. The bevel angle of the needle tip of BN-MAR-3 is β. The bevel angle of the Cutting cannula tip of BN-MAR-4 and BN-MAR-8 is δ.

    Each model has various specifications depending on different sizes such as needle lengths, needle diameters, etc. The cutting cannula has numerically ordered centimeter markings on it to provide reference for depth placement. In the distal area of the cutting cannula, an ultrasound enhancement is available to promote accurate placement under ultrasound or X-ray guidance.

    An adjustable depth stopper allows the user to restrict forward movement, localizing the needle tip to the biopsy site.

    The Operating handle of Inner stylet is color-coded, which indicates gauge size of the needle. It is available in several needle gauge sizes and lengths.

    The Disposable Biopsy Needle is a single-use device, supplied in a sterile state sterilized by EO gas. Re-sterilization by users is forbidden. The shelf life is defined for 3 years, which is verified.

    AI/ML Overview

    The provided document is a 510(k) clearance letter for a Disposable Biopsy Needle. It is a physical device, not an AI/software device. Therefore, the questions regarding acceptance criteria and studies related to AI performance, such as human-in-the-loop, standalone performance, ground truth establishment for training sets, multi-reader multi-case studies, and effect size of AI assistance, are not applicable.

    The document discusses non-clinical tests to demonstrate the safety and effectiveness of the device as a physical product.

    Here's a breakdown of the relevant information provided:

    1. Table of Acceptance Criteria and Reported Device Performance
    The document does not present a formal table of acceptance criteria with reported performance values in a side-by-side comparison. Instead, it describes various performance tests conducted. The acceptance criteria are implied by successful completion of these tests as per relevant standards.

    Acceptance Criteria (Implied from Tests)Reported Device Performance
    Packaging and Shelf-LifeMet requirements per ASTM F1980; Shelf life defined for 3 years, verified.
    BiocompatibilityMet requirements per ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, and ISO 10993-23.
    Sterilization EfficacyMet requirements per ISO 11135:2014; Device is sterile (sterilized by EO gas).
    EO ResidualsMet requirements per ISO 10993-7:2008.
    Physical and Functional Performance (General)Met requirements for: - Appearance - Basic dimensions - Sample collection space and accessibility - Luer connection - Connection firmness - Stiffness - Toughness - Corrosion resistance - Leakage - Ultrasound detectability - X-ray detectability - Protective sheath - Puncture force - Chemical Characteristics - EO and ECH Residual - Sterility test - Bacterial endotoxin
    Comparative Performance (Sampling)No significant difference in sampling weight or cell integrity compared to predicate devices across various tissues (liver, kidney, breast, thyroid, prostate, spleen, pancreas, lymph nodes, lungs).

    2. Sample size used for the test set and the data provenance
    The document details testing conducted on the device itself and does not refer to a "test set" in the context of diagnostic data. The tests performed are for physical and functional properties of the biopsy needle.

    • Sample size: Not explicitly stated for each test, but standard testing practices involve a sufficient number of samples to ensure representativeness and statistical validity.
    • Data provenance: Not applicable in the context of patient data; the data is generated directly from testing manufactured devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. This is a physical device. Ground truth refers to diagnostic accuracy in AI/imaging studies.

    4. Adjudication method for the test set
    Not applicable. Adjudication methods are used to establish ground truth in diagnostic studies, typically involving disagreements between expert readers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is a physical device, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This is a physical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    For the "Comparative study" mentioned in section 4 (Non-Clinical Tests), the ground truth for evaluating sampling performance would inherently be pathology-based assessment of the collected tissue samples (e.g., measuring sampling weight and assessing cell integrity). While not explicitly stated as "ground truth," this is the objective measure used to compare the device's performance against predicates.

    8. The sample size for the training set
    Not applicable. This is a physical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established
    Not applicable. This is a physical device, not an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243296

    Validate with FDA (Live)

    Date Cleared
    2025-07-11

    (266 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    China

    Re: K243296
    Trade/Device Name: Disposable Biopsy Needle
    Regulation Number: 21 CFR 876.1075
    Device Classification Name:** Gastroenterology-urology biopsy instrument
    Regulation Number: 21 CFR 876.1075
    ® Max-Core® Disposable Core Biopsy Instrument
    510(k) Number: K133948
    Regulation number: 876.1075
    Bard, Inc. | -- |
    | Regulation number | 21 CFR 876.1075 | 21 CFR 876.1075 | Same |
    | Product Code | KNW

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate. It is not intended for use in bone.

    Device Description

    The disposable biopsy needle is a sterile, spring loaded, disposable percutaneous soft tissue biopsy instrument. The disposable biopsy needle consists of a cannula, an inner stylet, a mechanical power device and a protective cover. It is used to obtain tissue samples for tissue biopsy, suitable for tissue biopsy of various organs such as the kidney, liver, prostate etc. The needle need to be inserted by a qualified physician under ultrasound guidance. The button of the mechanical power unit comply with the color coding requirements of ISO 6009, i.e. 18 gauge = Pink. The disposable biopsy needle includes three models: CMBNA/1810, CMBNA/1815, CMBNA/1820.

    AI/ML Overview

    The provided text is a 510(k) clearance letter for a "Disposable Biopsy Needle." It primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It does not include information about acceptance criteria, device performance from a clinical study, sample sizes for test/training sets, expert ground truth establishment, adjudication methods, or MRMC studies.

    Therefore, many of the requested elements cannot be extracted from this document. However, I can provide what is available, noting where information is absent.


    Acceptance Criteria and Device Performance (Based on Non-Clinical Testing)

    The document does not explicitly present a table of "acceptance criteria" against which a clinical device performance study was measured. Instead, it details that non-clinical testing was performed to demonstrate substantial equivalence to the predicate device. The performance data provided is primarily related to the physical and material characteristics of the device, rather than diagnostic accuracy or clinical outcomes.

    Summary of Reported Performance (Non-Clinical):

    Performance Aspect (Acceptance Criteria Implicitly Met by Testing)Reported Device Performance (Non-Clinical)
    BiocompatibilityComplies with ISO 10993-5 (cytotoxicity), -10 (sensitization), -23 (irritation), -11 (systemic toxicity), -4 (blood interactions), ASTM F756-17 (hemolytic properties), USP-NF <151> (pyrogen test).
    Packaging VerificationComplies with ASTM F1886/F1886M-16 (seal integrity), ASTM F1929-23 (dye penetration), ASTM F88/F88M-2021 (seal strength), DIN 58953-6:2023 (microbial barrier), USP-NF <71> (sterility).
    Sterilization & Shelf LifeEthylene oxide sterilization complies with ISO 11135 (process), ISO 10993-7 (ETO residuals), USP-NF <85> (bacterial endotoxins), ANSI AAMI ST72:2019 (bacterial endotoxins), ISO 11737-2 (sterility tests). Shelf life demonstrated for 2 years (compared to 3 years for predicate) through ASTM F1980-21 (accelerated aging), ASTM D4169-23 (shipping stability).
    Mechanical Performance (Needle Tubing)Complies with ISO 9626 (stainless steel needle tubing), ISO 7864 (hypodermic needles for single use), ISO 6009 (color coding).
    Sampling AbilitySamples collected from in vitro porcine liver, porcine kidney, and canine prostate were comparable in mass and quality to those collected using the predicate device.
    Functional ReliabilityCompleted 24 repeated sampling steps on in vitro tissues and pork; device fired correctly, and safety lock functioned normally, meeting acceptance criteria (implicitly, for smooth operation).
    Visibility (Ultrasound)B-ultrasound image display test report provided to prove visibility.

    Additional Requested Information:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: Not specified for any "test set" in the context of clinical performance. For non-clinical functional testing, "24 repeated sampling steps" were conducted for functional reliability.
      • Data Provenance: Not specified for any clinical data. Non-clinical bench testing was conducted.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable/Not provided. No clinical test set requiring expert ground truth was described.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable/Not provided. No clinical test set requiring adjudication was described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was mentioned. This device is a biopsy needle, not an AI-powered diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device (biopsy needle), not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not applicable for clinical ground truth. For non-clinical sampling verification, the "mass and quality" of collected tissue samples were compared.
    7. The sample size for the training set:

      • Not applicable/Not provided. This device is a physical instrument, not an AI model requiring a training set.
    8. How the ground truth for the training set was established:

      • Not applicable/Not provided. This device is a physical instrument, not an AI model requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250032

    Validate with FDA (Live)

    Date Cleared
    2025-07-10

    (184 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    MQK1620, MQK1810, MQK1816, MQK1820, MQK1825, MQK2010, MQK2016, MQK2020)
    Regulation Number: 21 CFR 876.1075
    Gastroenterology-Urology Biopsy Instrument
    Review Panel: Gastroenterology / Urology
    Regulation Number: 21 CFR 876.1075

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Marquee™ Disposable Core Biopsy Instrument and Kit are intended for use in obtaining biopsies from soft tissues such as breast, liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone.

    Device Description

    The Marquee™ Disposable Core Biopsy Instrument and Instrument Kit is a single use core biopsy device. It is available in several needle gauge sizes and lengths. The coaxial cannula release is color coded according to the various gauge sizes, e.g., yellow = 20 gauge, pink = 18 gauge, purple = 16 gauge, green = 14 gauge, and light blue = 12 gauge.

    The Marquee™ Disposable Core Biopsy Instrument is available as a biopsy instrument only and as a kit, which includes the Marquee™ Disposable Core Biopsy Instrument and compatible Disposable Coaxial Biopsy Needle.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Marquee™ Disposable Core Biopsy Instruments and Kits does not contain the detailed acceptance criteria and specific study results traditionally associated with the performance of AI/ML-driven medical devices.

    This document is for a physical medical device (biopsy instruments and kits) undergoing a "Special 510(k): Device Modification" clearance. As such, the performance data presented focuses on physical characteristics and material compatibility rather than AI/ML model performance metrics like accuracy, sensitivity, or specificity.

    Therefore, I cannot fulfill all parts of your request as the information is not present in the provided text. I will provide a summary based on the information that is available about the device's testing and a table of the acceptance criteria and reported "performance" based on the provided document's context, noting where information is not applicable or not provided.


    A. Acceptance Criteria and Reported Device "Performance" (for a physical medical device)

    Given that this is a physical biopsy instrument, the "performance" relates to its mechanical and material properties rather than diagnostic accuracy.

    Acceptance Criterion (Category)Reported Device "Performance" / Test Outcome
    BiocompatibilityConsidered biocompatible for its intended use.
    CytotoxicitySuccessfully performed
    SensitizationSuccessfully performed
    Intracutaneous ReactivitySuccessfully performed
    Acute System ToxicitySuccessfully performed
    Material Mediated PyrogenicitySuccessfully performed
    In Vitro Physical PerformanceEvaluated (details of acceptable ranges not provided, but tests were successful)
    Number of SamplesPerformed
    Penetration DepthsPerformed
    Stylet / Cannula to Handle Tensile StrengthPerformed
    Coaxial Stylet to Hub Tensile StrengthPerformed
    Corresponding Working Needle Length and Cutting Cannula/Coaxial Cannula ODPerformed
    SafetyPerformed
    ReliabilityPerformed
    Integrity of the Sterile BarrierPerformed
    Performance After Ship TestingPerformed
    Needle Protection After Shipping and StoragePerformed
    Shelf LifePerformed
    Ex-vivo Testing (Tissue Biopsy Quality)Determined sample quality using 12G, 14G, and 20G models on porcine tissue. (Specific metrics or acceptance thresholds for "sample quality" not detailed in the provided text, but implies satisfactory results for clearance).
    Substantial EquivalenceDemonstrated to be substantially equivalent to legally marketed predicate devices for the requested indications for use.

    B. Study Details (as inferable from the document for a physical device)

    1. Sample sizes used for the test set and the data provenance:

      • Sample Size: Not explicitly stated for specific tests (e.g., how many instruments were tested for tensile strength or how many biopsies were taken in ex-vivo tests). The document states "ex-vivo testing was conducted on porcine tissue" using 12G, 14G, and 20G models.
      • Data Provenance: The biocompatibility testing followed ISO 10993-1:2018. The ex-vivo testing was done on "porcine tissue." The context suggests these are laboratory-controlled, prospective tests conducted by the manufacturer, Bard Peripheral Vascular, Inc. No country of origin for the data is specified, but typically, these studies are conducted in the country of the manufacturer or approved testing facilities.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable / Not Provided for this device. For physical devices like biopsy instruments, the "ground truth" is measured against engineering specifications and material properties (e.g., tensile strength, successful sterile barrier). Expert consensus (like in image interpretation) is not directly relevant to these types of performance tests. The ex-vivo tissue quality assessment would likely involve pathologists or relevant subject matter experts, but their number and qualifications are not specified.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable / Not Provided for this device. Adjudication methods like 2+1 are typically used in clinical studies or studies involving subjective human interpretation (e.g., reading medical images). For the mechanical and material tests described, results are objective measurements against predefined specifications.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not an AI/ML device. An MRMC study is completely irrelevant to this type of medical device (physical biopsy instrument).
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. This is not an AI/ML device. Standalone performance testing refers to the performance of an algorithm without human intervention, which does not apply here.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For Biocompatibility: Compliance with ISO 10993-1:2018 standards and specific test results (cytotoxicity, sensitization, etc.).
      • For In Vitro Physical Performance: Engineering specifications, design tolerances, and established industry standards for mechanical properties.
      • For Ex-vivo Testing: The "sample quality" on porcine tissue would likely be assessed against histological standards (e.g., tissue integrity, architecture, presence of target cells). Whether this involved expert pathology consensus or objective criteria is not specified, but it would fall under a form of "pathology" ground truth for the sample quality.
    7. The sample size for the training set:

      • Not applicable. This device is a physical instrument, not an AI/ML model that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. This device is a physical instrument, not an AI/ML model that requires a training set.