Re: K252295
Trade/Device Name: Ryurei
Regulation Number: 21 CFR 870.5100
coronary angioplasty, percutaneous
Classification Panel: Cardiovascular
Regulation: 21 CFR 870.5100
Ryurei (1.5 mm) is indicated for:
- Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
- Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
Ryurei (2.0–4.0 mm) is indicated for:
- Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
- Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
- Balloon post-deployment expansion of balloon expandable stents (bare metal and drug-eluting).
The tip of the catheter is equipped with a balloon inflatable to a specific diameter and length at recommended pressures.
One or two radiopaque markers inside the balloon.
On the shaft, there are two depth markers.
Double-lumen structure, with a guidewire lumen and an inflation lumen.
A hub is attached to the balloon for inflation/deflation.
The surface of catheter is coated with hydrophilic polymer coating.
N/A
Ask a specific question about this device
(205 days)
Name:** Sprinter Legend Rapid Exchange Balloon Dilatation Catheter
Regulation Number: 21 CFR 870.5100
Angioplasty (PTCA) Catheter
Classification Panel: Cardiovascular
Classification Regulation: 21 CFR 870.5100
The Sprinter Legend RX and OTW 1.25mm Balloon Dilatation Catheter is indicated as a pre dilatation catheter for enlarging coronary luminal diameters during PCI procedures.
The Sprinter Legend RX 1.5-4.0mm balloon dilatation catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter (balloon models 2.25 mm - 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents.
Sprinter Legend Rapid Exchange Balloon Dilatation Catheter is a percutaneous transluminal coronary angioplasty (PTCA) device. The proximal end comprises of a luer hub, strain relief and a hypotube and stiffening wire subassembly (proximal shaft assembly). The distal end comprises of an inflatable balloon, dual lumen tip (guidewire lumen and the inflation lumen) and radiopaque marker bands. The distal shaft includes a hydrophilic coating to aid in device advancement to the target lesion. The guidewire lumen at the distal end enables the use of a 0.014" guidewire to position the device during use. Radiopaque balloon markers enable accurate placement at the target lesion. Exit markers on the proximal shaft indicate the exit of the balloon tip out of the guide catheter for brachial and femoral approaches.
N/A
Ask a specific question about this device
(25 days)
Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter
Regulation Number: 21 CFR 870.1250, 21 CFR 870.5100
Catheter, transluminal coronary angioplasty, percutaneous
Product Code: LOX
Regulation Number: 870.5100
| Common name | PTCA Catheter | Identical |
| Classification | PTCA Catheter, LOX, Class II (21 CFR 870.5100
Catheter, transluminal coronary angioplasty, percutaneous
Product Code: LOX
Regulation Number: 870.5100
| Common name | PTCA Catheter | Identical |
| Classification | PTCA Catheter, LOX, Class II (21 CFR 870.5100
The Passeo-35 Xeo peripheral dilatation catheter is indicated to dilate stenosis in the iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Passeo-35 Xeo is also recommended for post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature.
The Passeo-18 peripheral dilatation catheter is indicated to dilate stenosis in the femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Passeo-14 is indicated for balloon dilatation of the stenotic portion of a lower limb artery for the purpose of improving perfusion.
The Oscar Peripheral Multifunctional Catheter system is indicated for percutaneous transluminal interventions in the peripheral vasculature to provide support during access into and to dilate stenoses in femoral, popliteal and infrapopliteal arteries. The product is also intended for injection of radiopaque contrast media for the purpose of angiography.
The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.
The Pantera Pro is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Pantera Pro (balloon diameter 2.0 – 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents.
BIOTRONIK's Passeo 35 Xeo Catheter is an over-the-wire (OTW) balloon dilatation catheter, indicated for dilatation of stenotic segments in peripheral vessels. The Passeo 35 Xeo Catheter is a dual lumen design with both lumens contained within one tube. The smaller lumen is the balloon inflation/deflation lumen. The larger lumen permits the use of guide wires with a maximum diameter of 0.035" to facilitate advancement of the Passeo 35 Xeo Catheter towards and through the lesion(s) to be dilated.
The Passeo-18 peripheral dilatation catheter is intended for dilatation of stenotic segments in peripheral vessels and arteriovenous dialysis fistulae. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. One radiopaque marker is located at each end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation catheter includes a soft tapered tip to facilitate advancement of the catheter. The dilatation catheter has two Luer-ports at the proximal end. One port (inflation port) serves for connecting an inflation device to inflate/deflate the balloon. The other port enables flushing of the guide wire lumen. The dilatation catheter has a hydrophobic silicone coating on the shaft outer surface and a hydrophobic patchwork coating on the balloon.
The Passeo-14 peripheral dilatation catheter is intended for the dilatation of stenotic segments in lower limb arteries. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. One radiopaque marker is located at each end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation catheter includes a soft tapered tip to facilitate advancement of the catheter. The dilatation catheter has two Luer-ports at the proximal end. One port (inflation port) serves for connecting an inflation device to inflate/deflate the balloon. The other port enables flushing of the guide wire lumen.
The Oscar Peripheral Multifunctional Catheter system is an intravascular balloon catheter system, supplied with a retractable sheath (Oscar Support Catheter), a flexible catheter (Oscar Dilator) and a PTA balloon (Oscar PTA balloon), allowing a variable guide wire support and injection of fluids, and adjustable length inflatable balloon up to 180 mm. The over the wire (OTW) catheter has a retractable sheath allowing the balloon to be inflated at various lengths as determined by the physician. The balloon lengths are graduated with evenly spaced radiopaque markers. The Oscar Peripheral Multifunctional Catheter system is a 4F and 6F catheter system with a shaft working length of 60 cm or 120 cm, compatible with 0.014" (Oscar 4F) and 0.018" (Oscar 6F) guide wires. The device uses a semi-compliant balloon with a size dependent rated burst pressure and an indicated clinical use range of 6 atm to 16 atm. The balloon expands to a set nominal diameter (2.0, 2.5, 3.0, 4.0, 5.0, 6.0 and 7.0 mm). If required, separate PTA balloon catheters in same size range are also available separately.
Pantera LEO is a PTCA rapid exchange system with a balloon at the distal end of the catheter. A Luer port at the proximal end enables the attachment of an inflation device for the inflation of the balloon. The catheter provides a lumen, which enables the use of a guide wire to position the catheter. Radiopaque balloon markers aid in the placement of the catheter's balloon segment under fluoroscopy.
Pantera Pro is a PTCA rapid exchange system with a balloon at the distal end of the catheter. A Luer port at the proximal end enables the attachment of an inflation device for the inflation of the balloon. The catheter provides a lumen, which enables the use of a guide wire to position the catheter. Radiopaque balloon markers aid in the placement of the catheter's balloon segment under fluoroscopy.
The provided FDA 510(k) clearance letter (K250706) is for BIOTRONIK's peripheral and coronary dilatation catheters. This document describes several devices: Passeo-35 Xeo, Passeo-18, Passeo-14, Oscar Peripheral Multifunctional Catheter System, Pantera LEO PTCA Catheter, and Pantera Pro PTCA Catheter.
It's important to note that this 510(k) Summary does not describe an AI/ML powered device. Instead, it focuses on the physical and performance characteristics of medical devices (catheters). Therefore, many of the requested criteria, such as "number of experts used to establish ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone algorithm performance," and details about training data, are not applicable to this type of submission.
The "performance testing" mentioned throughout the document refers to bench testing and engineering assessments to ensure the physical device meets its design specifications and performs safely and effectively. It does not involve AI model performance evaluation with clinical data.
Here's an attempt to answer the applicable parts of your request based on the provided text, with clear indications where the information is not present or not relevant to an AI/ML device:
Acceptance Criteria and Study to Prove Device Meets Criteria
The acceptance criteria for these devices are implicitly derived from the design specifications and are demonstrated through performance (bench) testing. The study proving these devices meet the acceptance criteria is a series of performance (bench) tests.
1. Table of Acceptance Criteria and Reported Device Performance
The document describes technological characteristics and states that "All necessary performance testing was conducted... to ensure that the devices conform to the design specification." The "Subject Devices" column in the tables explicitly states "Identical" for most characteristics, meaning they meet the same specifications as the predicate devices. The only explicit change mentioned is that "Components will be brought into compliance with ISO 80369-7:2021" for Luer connectors and manifolds. The full set of specific performance acceptance criteria values (e.g., specific burst pressure thresholds, flexibility requirements, etc.) are not explicitly detailed in this summary but are indirectly stated as being met for "consistent performance during its intended use."
Here's a summary derived from the comparison tables, focusing on the acceptance of "identical" characteristics:
| Acceptance Criteria Category | Specific Criteria (from Predicate Device) | Reported Device Performance (Subject Device) |
|---|---|---|
| Intended User | Physicians competent in PTA procedures | Identical |
| Method of Placement | Standard percutaneous access to site over a guide wire, with fluoroscopic visualization | Identical |
| Sterilization | EO gas, SAL 10⁻⁶ | Identical |
| Shelf Life | 3 years | Identical |
| Radiopaque Markers | Presence, material, number, length, and spacing as specified for each device type | Identical |
| Usable Length | Specified lengths (e.g., 90, 130, 170 cm for Passeo-35 Xeo) | Identical |
| Introducer Sheath Compatibility | Specified F sizes and balloon diameter/length compatibility | Identical |
| Crossing Profile | Maximum inches/mm as specified | Identical |
| Guide Wire Compatibility | Specified guide wire diameter (e.g., 0.035" for Passeo-35 Xeo) | Identical |
| Shaft Outer Diameter | Specified F size | Identical |
| Balloon Diameter | Specified range of diameters (e.g., 3.0-12.0 mm for Passeo-35 Xeo) | Identical |
| Balloon Length | Specified range of lengths (e.g., 20-250 mm for Passeo-35 Xeo) | Identical |
| Balloon Wrapping | Specified folds (e.g., 3-5 folds for Passeo-35 Xeo) | Identical |
| Balloon Nominal Pressure | Specified atm | Identical |
| Balloon RBP (Rated Burst Pressure) | Specified atm, often varying by balloon diameter/length | Identical |
| Guiding Catheter Compatibility (PTCA catheters) | Minimum F size/ID | Identical |
| Distal Outer Shaft/Balloon Coating (PTCA catheters) | Hydrophilic/Hydrophobic coating as specified | Identical |
| Luer Connectors and Manifolds | Previously Luer lock connectors (L2) | Will be brought into compliance with ISO 80369-7:2021 (This is the only explicitly noted difference/update) |
Overall Performance Conclusion: "The collective results of the performed testing demonstrated that the materials chosen, the manufacturing processes, and design of the components meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the proposed device does not introduce new issues of safety or effectiveness when compared to the predicate device."
2. Sample size used for the test set and the data provenance
The document specifies "performance testing" or "bench testing," which refers to laboratory-based evaluations of the physical device. It does not mention a test set with patient data or any sample size related to clinical data. The provenance is not applicable as this concerns physical device characteristics, not data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as the evaluation is not based on expert-labeled ground truth for an AI/ML algorithm but on physical measurements and engineering tests of the device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the evaluation is not based on expert consensus or adjudication of clinical cases for an AI/ML algorithm.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable as the device is a physical medical instrument (catheter), not an AI/ML diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable as the device is a physical medical instrument (catheter), not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for these devices is the established engineering specifications, material properties, and performance benchmarks that the predicate devices have successfully met. These are verified through various physical and mechanical bench tests (e.g., burst pressure, flexibility, lubricity, dimensional accuracy, etc.). The document indicates these tests confirm the new devices' conformity to these specifications.
8. The sample size for the training set
This information is not applicable as the device is a physical medical instrument (catheter) and does not involve AI/ML training.
9. How the ground truth for the training set was established
This information is not applicable as the device is a physical medical instrument (catheter) and does not involve AI/ML training or a training set with ground truth.
Ask a specific question about this device
(29 days)
Oregon 97035
Re: K242969
Trade/Device Name: Pantera Pro: Pantera LEO Regulation Number: 21 CFR 870.5100
|
| Regulation | 21 CFR 870.5100
| 21 CFR 870.5100
The Pantera Pro is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Pantera Pro (balloon diameter 2.0 - 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents.
The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.
Pantera Pro is a PTCA rapid exchange system with a balloon at the distal end of the catheter. A Luer port at the proximal end enables the attachment of an inflation device for the inflation of the balloon. The catheter provides a lumen, which enables the use of a guide wire to position the catheter. Radiopaque balloon markers aid in the placement of the catheter's balloon segment under fluoroscopy.
Pantera LEO is a PTCA rapid exchange system with a balloon at the distal end of the catheter. A Luer port at the proximal end enables the attachment of an inflation device for the inflation of the balloon. The catheter provides a lumen, which enables the use of a guide wire to position the catheter. Radiopaque balloon markers aid in the placement of the catheter's balloon segment under fluoroscopy.
The provided text is a 510(k) premarket notification for medical devices (Pantera Pro and Pantera LEO catheters). It focuses on demonstrating substantial equivalence to predicate devices, primarily through comparison of technological characteristics and performance data.
Crucially, this document does not contain information about studies related to Artificial Intelligence (AI) or machine learning (ML) models, nor does it provide details about human reader performance, expert adjudication, or ground truth establishment in the context of diagnostic or prognostic AI applications.
The "Performance Data" section solely mentions:
- Product performance - coating integrity and particulates
- Biocompatibility
- Microbiological - Bioburden
- Shelf-Life - coating integrity and particulates
These are standard performance metrics for physical medical devices like catheters, not for AI/ML software.
Therefore, I cannot provide the requested information for an AI/ML-based device based on this document. The information you're asking for would typically be found in a submission for a software as a medical device (SaMD) or an AI-enabled device, which this document is not.
Ask a specific question about this device
(290 days)
Trade/Device Name: Sapphire NC ULTRA Coronary Dilatation Catheter Regulation Number: 21 CFR 870.5100
CoronaryAngioplasty (PTCA) CatheterTrade Name: Sapphire NC ULTRARegulation Number: 21 CFR 870.5100
The Sapphire NC ULTRA Coronary Dilatation Catheter is indicated for:
· Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in □ patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
· Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial □ infarction.
· In-stent restenosis.
· Post-delivery expansion of balloon expandable coronary stents.
The Sapphire NC ULTRA Coronary Dilatation Catheter is designed to allow easy exchange of the catheter using a standard length 0.014 inch guidewire. Balloon diameters range from 1.75mm to 5.0mm. The balloon material is made of a minimally compliant material, 1.75mm to 4.0mm balloons have a rated burst pressure of 20 atmospheres, and 4.5mm to 5.0mm balloons have a rated burst pressure of 18 atmospheres. The minimally compliant balloon material allows high pressure dilatation while maintaining precise control of the balloon diameter and length. The proximal shaft of the catheter is composed of a female luer connector connected to a PTFE coated stainless steel tube. The proximal shaft allows superior proximal pushability and trackability with a smooth transition to a distal shaft. Two radiopaque platinum/iridium marker bands are located within the balloon segment. The guidewire enters the catheter tip and advances coaxially out the distal Rx port, thereby allowing both coaxial guidance and rapid exchange of catheter with a single standard-length guidewire. Two marked sections of 5mm length each located on the proximal shaft indicate catheter position relative to the tip of either a brachial or femoral guiding catheter.
The provided text is a 510(k) summary for the Sapphire NC ULTRA Coronary Dilatation Catheter. It outlines various performance tests conducted to establish substantial equivalence to a predicate device. However, it does not contain the specific level of detail required to fully answer all aspects of your request, particularly regarding clinical study design, expert qualifications, or detailed performance metrics against acceptance criteria.
Here's an attempt to answer your questions based on the available information:
1. A table of acceptance criteria and the reported device performance
The document states: "The test results met all acceptance criteria, which are the same or similar to the predicate device and ensure that the Sapphire NC ULTRA Coronary Dilatation Catheter design and construction are suitable for their intended use."
While it lists the types of tests performed, it does not provide a table with specific acceptance criteria or quantitative performance results for each test. For example, it lists "Balloon Rated Burst Pressure (in-stent)" as a test, but doesn't state what the accepted pressure was or what the device achieved.
Here's a generalized table based on the types of tests mentioned, but without specific numerical criteria or performance data:
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| In vitro Performance: | |
| Visual Inspection | Met acceptance criteria |
| Particulate Evaluation | Met acceptance criteria |
| Dimensional Verification | Met acceptance criteria |
| Balloon Compliance | Met acceptance criteria |
| Hub Leakage Test | Met acceptance criteria |
| Balloon Prep, Deployment, Retraction | Met acceptance criteria |
| Balloon Inflation/Deflation Time | Met acceptance criteria |
| Coating Integrity | Met acceptance criteria |
| Balloon Fatigue (in-stent) | Met acceptance criteria |
| Balloon Fatigue | Met acceptance criteria |
| Balloon Rated Burst Pressure (in-stent) | Met acceptance criteria |
| Balloon Rated Burst Pressure | Met acceptance criteria |
| Shaft Burst | Met acceptance criteria |
| Catheter Bond Strength | Met acceptance criteria |
| Tip Pull Strength | Met acceptance criteria |
| Flexibility and Kink | Met acceptance criteria |
| Torque Strength | Met acceptance criteria |
| Marker Band Radiopacity | Met acceptance criteria |
| Biocompatibility: | |
| Cytotoxicity | Met acceptance criteria |
| Sensitization | Met acceptance criteria |
| Intracutaneous Reactivity | Met acceptance criteria |
| Acute Systemic Toxicity | Met acceptance criteria |
| Pyrogenicity | Met acceptance criteria |
| Hemolysis | Met acceptance criteria |
| Partial Thromboplastin Time | Met acceptance criteria |
| Platelet and Leukocyte Counts | Met acceptance criteria |
| Complement Activation | Met acceptance criteria |
| Toxicological Risk Assessment (TRA) of Extractable Chemicals | Met acceptance criteria |
| Other: | |
| Packaging and Sterilization Validation | Met acceptance criteria |
| Shelf Life | Met acceptance criteria |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document describes in vitro performance tests and biocompatibility tests. This means the "test set" refers to material samples (e.g., catheter units, material extracts) used for these laboratory tests, not a clinical patient dataset.
- Sample Size: The document does not specify the sample size for the in vitro or biocompatibility tests.
- Data Provenance: Not applicable in the context of in vitro and biocompatibility testing. These are laboratory tests conducted on device components or finished products.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This question is geared towards clinical studies involving human interpretation or pathology. Since the provided text only details in vitro and biocompatibility testing, there is no information about experts or their qualifications for establishing ground truth as would be required in a clinical setting. These tests typically follow standardized protocols and are evaluated by lab personnel, not medical experts establishing clinical ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Again, this question is relevant for clinical studies with human assessors. For the in vitro and biocompatibility testing described, adjudication methods like N+1 are not applicable. The results are typically quantitative measurements or observations against predefined pass/fail criteria according to established standards (e.g., ISO, FDA guidance).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. The device is an angioplasty catheter, not an AI software or a device that assists human readers in interpreting images. Therefore, this question is not applicable to the information provided.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is also not applicable. The Sapphire NC ULTRA Coronary Dilatation Catheter is a physical medical device, not an algorithm or software requiring standalone performance testing in that context.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the in vitro performance tests, the "ground truth" is established by adherence to predefined engineering specifications, material properties, and performance benchmarks derived from industry standards (e.g., those in the "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters").
For biocompatibility testing, the "ground truth" is established by compliance with international standards like ISO 10993-1, which define acceptable biological responses and safety profiles for medical device materials.
8. The sample size for the training set
This question is applicable to machine learning or AI models. Since the device is a physical medical catheter and the testing described is primarily in vitro and biocompatibility, there is no concept of a "training set" in this context.
9. How the ground truth for the training set was established
As there is no training set for a physical medical device, this question is not applicable.
Ask a specific question about this device
(289 days)
: K233505
Trade/Device Name: Sapphire ULTRA Coronary Dilatation Catheter Regulation Number: 21 CFR 870.5100
|
| Regulation Number: | 21 CFR 870.5100
The Sapphire ULTRA Coronary Dilatation Catheter (1.0-1.25mm configurations) is indicated for: balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis) for the purpose of improving myocardial perfusion.
The Sapphire ULTRA Coronary Dilatation Catheter (1.5-4.0mm configurations) is indicated for: balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in . patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
The Sapphire ULTRA Coronary Dilatation Catheter is designed to allow easy exchange of the catheter using a standard length 0.014 inch guidewire. Balloon diameters range from 1.0mm to 4.0mm. The balloon material is made of the semi-compliant material with a rated burst pressure of 14 atmospheres. The proximal shaft of the catheter is composed of a female luer connector connected to a PTFE coated stainless steel tube. The proximal shaft allows superior proximal pushability with a smooth transition to a distal shaft composed of an outer tube and an inner tube with a balloon laser welded to both tubes at the distal tip. Two radiopaque platinum marker bands are located within the balloon segment (there is only one centrally located marker band for Ø1.0-1.5mm configurations). The guidewire enters the catheter tip and advances coaxially out the distal Rx port, thereby allowing both coaxial guidance and rapid exchange of catheter with a single standard length guidewire. Two marked sections, 5mm length each, located on the proximal shaft indicate catheter position relative to the tip of either a brachial or femoral guiding catheter.
The provided text is a 510(k) summary for the Sapphire ULTRA Coronary Dilatation Catheter. It outlines the device's description, indications for use, technological characteristics, and performance tests conducted to demonstrate substantial equivalence to predicate devices.
However, the provided document DOES NOT contain information regarding:
- Acceptance criteria for an AI/ML device. This document describes a medical device (a catheter), not an AI/ML diagnostic or prognostic tool. Therefore, there are no AI/ML specific acceptance criteria, study designs (like MRMC), or details about ground truth establishment as would be typically required for such devices.
- Study that proves the device meets AI/ML acceptance criteria. The performance tests listed are for the physical properties and biocompatibility of a medical catheter, not for the performance of an AI/ML algorithm.
Therefore, I cannot provide the requested information regarding AI/ML acceptance criteria and proof of performance. The document focuses on the physical and biological safety and efficacy of a coronary dilatation catheter, which is a hardware medical device.
Ask a specific question about this device
(192 days)
Trade Name Ringer perfusion balloon catheter Common / Usual Name Catheter Classification Name 21 CFR 870.5100
The Ringer Perfusion Balloon Catheter is indicated for balloon dilatation of coronary artery or coronary bypass graft stenoses where the physician desires distal blood perfusion during balloon inflation for the purpose of improving myocardial perfusion.
The Ringer perfusion balloon catheter is a rapid exchange catheter with a spiral-shaped inflatable balloon on the distal end and a wire lumen for delivery over a ≤ 0.014" guidewire. The inflated spiral balloon approximates a hollow cylinder. Radiopaque markers at the proximal and distal end of the balloon facilitate visualization and intravascular placement of the catheter prior to inflation. The catheter shaft has positioning marks located at 95cm (single mark) and 105cm (double mark) from the distal tip. The Ringer perfusion balloon catheter has been sterilized with ethylene oxide.
The Ringer perfusion balloon catheter is indicated for balloon dilatation of coronary bypass graft stenoses where the physician desires distal blood perfusion during balloon inflation for the purpose of improving myocardial perfusion.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Primary Effectiveness Endpoint: Successful delivery, inflation, deflation, and removal of the study device; no vessel perforation, flow-limiting vessel dissection, no reduction in TIMI flow from baseline, and no clinically significant arrhythmias requiring medical treatment or device intervention following dilatation with the study device; achievement of final TIMI flow grade of 3 at the conclusion of the PCI procedure for the target lesion. | Met in 57/60 (95%) of the subjects. |
| Primary Safety Endpoint: Occurrence of clinically relevant events during prolonged balloon inflations and freedom from MACE at hospital discharge. | Met in 95% (57/60) of the subjects. |
| Secondary Efficacy Endpoint 1: Successful PCI defined as final residual stenosis ≤20% diameter stenosis (in-stent) or ≤ 50% diameter stenosis (stand-alone balloon) by visual angiographic estimation and freedom from MACE at hospital discharge. | Met in 93% (56/60) of the subjects. |
| Secondary Efficacy Endpoint 2: Maintenance of TIMI-2 or -3 flow into distal coronary bed during study device inflation. | Achieved in 100% of cases. |
| Secondary Safety Endpoint: Tolerance of at least one study device inflation equal to or greater than one minute (prolonged inflation) at target site. | Met in 50/60 (83.3%) of subjects. |
| Procedural Success | 95% of cases |
| No Serious Adverse Events attributable to the subject device | Achieved (implicitly, as 95% primary safety endpoint was met) |
2. Sample size used for the test set and the data provenance
- Sample Size: 60 subjects.
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. However, it refers to the study as "G200321, 'Ringer PTCA Study'," which suggests a dedicated clinical study, typically prospective in nature for device clearance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not provide specific details about the number of experts used or their qualifications for establishing ground truth within the clinical study. Angiographic estimations (e.g., final residual stenosis, TIMI flow grade) are typically performed by interventional cardiologists or trained core lab personnel, but this is not explicitly stated.
4. Adjudication method for the test set
The document does not describe a specific adjudication method (e.g., 2+1, 3+1). Clinical endpoints for device studies are often determined by the treating physician or a study's Clinical Events Committee (CEC), but this information is not elaborated upon.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study with AI assistance was not done. This study focuses on the performance of a medical device (a balloon catheter) rather than an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No, this is a study evaluating a physical medical device (balloon catheter) in human subjects, not a standalone algorithm.
7. The type of ground truth used
The ground truth for the clinical study was established through a combination of:
- Angiographic evaluation: Visual assessment of TIMI flow grade and residual stenosis.
- Clinical outcomes: Occurrence of clinically relevant events, freedom from MACE (Major Adverse Cardiac Events) at hospital discharge, and tolerance of prolonged inflations.
- Procedural observations: Successful delivery, inflation, deflation, removal, and absence of complications like vessel perforation or dissection.
8. The sample size for the training set
This information is not applicable as this is a clinical study for a physical medical device, not a machine learning model that requires a "training set."
9. How the ground truth for the training set was established
This information is not applicable, as there is no "training set" in the context of this device's clinical evaluation.
Ask a specific question about this device
(282 days)
Re: K223189
Trade/Device Name: Firefighter™ NC Pro PTCA Balloon Catheter Regulation Number: 21 CFR 870.5100
|
| Classification: | Class II, 21 CFR Part 870.5100
Firefighter™ NC Pro PTCA Balloon Catheter is indicated to be used for:
Balloon dilatation of the stenotic portion of a native coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion;
Balloon dilatation of post-delivery expansion of balloon expandable stents (models 2.00-5.00mm).
Firefighter™ NC Pro PTCA Balloon Catheter is a rapid exchange catheter, with a non-compliant balloon near the distal tip. Two radiopaque marker bands are located underneath the balloon and fixed on the inner lumen to position the balloon. The hydrophilic coating is located on the whole distal section to reduce friction and enhance the crossing capability. The effective length of the catheter is 145cm, the inner lumen of catheter permits the use of guide wire ≤ 0.014 inch (0.36mm). Firefighter™ NC Pro is available with balloon diameters of 2.00-5.50mm and balloon lengths of 6, 10, 15, 20, 25 and 30mm. 5F Guiding catheter is used for the delivery of Firefighter™ NC Pro with 2.00-4.00mm balloon diameter. 6F Guiding catheter is used for the delivery of Firefighter™ NC Pro with 4.50-5.50mm balloon diameter.
The provided document does not describe an AI/ML powered device. It details the 510(k) premarket notification for the "Firefighter™ NC Pro PTCA Balloon Catheter", which is a traditional medical device (a balloon catheter for angioplasty).
Therefore, I cannot provide information on acceptance criteria and study details related to an AI/ML device, as this document is not about such a device. The questions about test sets, data provenance, expert adjudication, MRMC studies, standalone performance, training sets, and ground truth establishment for a machine learning model are not applicable to the information contained within these pages.
Ask a specific question about this device
(136 days)
Catheter (VesTraveler®), Coronary Dilatation Balloon Catheter (VexPander®) Regulation Number: 21 CFR 870.5100
|
| Regulation Number: | 21 CFR 870.5100
The VesPenetrator®, VesTraveler® Coronary Dilatation Balloon Catheters are indicated for:
· balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion
· balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction
The VexPander® Coronary Dilatation Balloon Catheters are indicated for:
· balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion
· balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction
· in-stent restenosis
· post-delivery expansion of balloon expandable coronary stents
The Coronary Dilatation Balloon Catheter (VesPenetrator®, VesTraveler® and VexPander®) is a percutaneous transluminal coronary angioplasty (PTCA) is indicated for balloon dilatation of the stenotic portion of a coronary artery. Balloon diameters ranged from 1.5mm to 4.0mm. The balloon is made of Nylon material with a rated burst pressure of 16 atmospheres (atm) for both VesPenetrator and VesTraveler, and 18atm for VexPander. The proximal shaft of the catheter is composed of a female luer connector bonded to a PTFE coated stainless steel tube. The proximal shaft joins with a smooth transition to a distal shaft composed of an outer tube of nylon or Pebax and a tri-extrusion inner tube with a balloon welded to both tubes at the distal tip. Two radiopaque platinum marker bands are positioned within the balloon shoulders. The inner tube accepts a standard 0.014 inch guidewire. The guide wire enters the catheter's tip and advances coaxially out the distal Rx port, thereby allowing both coaxial guidance and rapid exchange of catheter with a single standard length guide wire. Two marked sections of 5mm length each located on the proximal shaft indicate catheter position relative to the tip of the brachial artery or the guide tube of the femoral artery.
The provided text is a 510(k) Summary for a medical device (Coronary Dilatation Balloon Catheters). It describes the device, its intended use, comparison to predicate devices, and the performance data that supports its substantial equivalence. However, this document does not contain information about an AI-based device or a study that involves human readers, ground truth establishment by experts, or MRMC studies.
The "Performance Data" section (Page 5 and 6) lists various non-clinical engineering tests performed on the balloon catheters (e.g., balloon inflation/deflation time, bond strength, flexibility, etc.) and biocompatibility testing. These are physical tests to ensure the device performs as expected and is safe for use in the human body.
Therefore, I cannot provide the requested information regarding AI device acceptance criteria, expert ground truth, MRMC studies, or training/test set details because this document does not pertain to an AI/ML medical device.
Here's why the requested information cannot be extracted from the provided text:
- No AI Device: The device described (Coronary Dilatation Balloon Catheter) is a physical medical instrument, not a software or AI-driven diagnostic/assistive tool.
- No AI-Specific Acceptance Criteria: The acceptance criteria mentioned relate to the physical and functional performance of the catheter (e.g., burst pressure, flexibility), not AI performance metrics like sensitivity, specificity, AUC, or concordance with human readers.
- No Human Readers/Experts: There is no mention of a study involving human readers (e.g., radiologists, cardiologists) or experts establishing ground truth for diagnostic purposes. The "experts" in this context would likely be engineers or quality control personnel evaluating the physical device.
- No MRMC Study: An MRMC (Multi-Reader Multi-Case) study is specific to evaluating the impact of AI on human diagnostic performance. This is not applicable here.
- No Training/Test Sets: The concepts of training sets and test sets are relevant to machine learning model development and evaluation, which is not described. The "testing" mentioned is laboratory testing of the physical catheter.
- No Ground Truth for AI: The "ground truth" for a physical device is its engineered specifications and performance under defined conditions, not "expert consensus" or "pathology" in the context of an AI-driven diagnosis.
In summary, the provided document relates to the regulatory clearance of a physical medical device (coronary balloon catheter) through substantial equivalence, based on engineering and biocompatibility testing, not an AI/ML device.
Ask a specific question about this device
(28 days)
China
Re: K230705
Trade/Device Name: POT PTCA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100
China
Re: K230705
Trade/Device Name: POT PTCA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100
Name: Catheters, Transluminal Coronary Angioplasty, Percutaneous Classification Regulation: 21 CFR 870.5100
The POT PTCA balloon dilatation catheter is indicated for:
- balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion
- balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction
- balloon dilatation of a stent after implantation (balloon models Ø2.25 mm - Ø5.00 mm only)
Note: Bench testing was conducted with the POT PTCA Balloon Dilatation Catheter and marketed balloon expandable stents. Consideration should be taken when this device is used with different manufacturers' stents due to difference in stent design.
The POT PTCA balloon dilatation catheter is a sterile, flexible tube designed to be used in percutaneous transluminal coronary angioplasty (PTCA) to dilate a stenotic coronary artery by controlled inflation of a distensible Nylon balloon at its distal tip. It is typically a rapid exchange (Rx) type with a single-lumen catheter. It is a single-use device and available in various sizes.
Based on the provided text, the document is a 510(k) summary for the "POT PTCA Balloon Dilatation Catheter." This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed clinical study report with specific acceptance criteria, sample sizes, and expert adjudication for a novel AI/software device. As such, most of the requested information regarding an AI/ML device's acceptance criteria and study proving its performance is not present in this document.
However, I can extract the information related to performance testing for this physical medical device (a balloon catheter) and note the absence of data pertinent to AI/ML device evaluations.
Here's a breakdown based on the provided text:
Acceptance Criteria and Device Performance (for the physical device, not AI/ML):
The document mentions that bench testing was performed and met all acceptance criteria. However, it does not explicitly list the specific numerical acceptance criteria for each test. It only states that the tests demonstrated the device performed as intended and did not impact functionality.
| Acceptance Criteria (Description) | Reported Device Performance |
|---|---|
| Dimensional verification | Met all acceptance criteria (specific values not provided in document) |
| Simulated Use | Met all acceptance criteria (specific outcomes not provided in document) |
| Balloon Rated Burst Pressure | 20-22 atm (Identical to predicate) |
| Balloon Fatigue | Met all acceptance criteria (specific outcomes not provided in document) |
| Balloon Compliance | Met all acceptance criteria (specific outcomes not provided in document) |
| Balloon Inflation and Deflation Time | Met all acceptance criteria (specific outcomes not provided in document) |
| Balloon Burst (in stent) | Met all acceptance criteria (specific outcomes not provided in document) |
| Balloon Fatigue (in stent) | Met all acceptance criteria (specific outcomes not provided in document) |
Information NOT available in the document, as it pertains to AI/ML device studies, not physical medical devices like this one:
- Sample sizes used for the test set and the data provenance: Not applicable to this type of device submission. The "test set" here refers to physical bench tests, not a dataset for an algorithm.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's performance is typically established through engineering specifications and standardized testing, not expert consensus as in image interpretation.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For a physical device, ground truth relates to its physical properties and mechanical performance, established by engineering standards and measurements.
- The sample size for the training set: Not applicable (no AI/ML training involved).
- How the ground truth for the training set was established: Not applicable.
Summary of Relevant Information from the Document (for a physical medical device):
- Device Name: POT PTCA Balloon Dilatation Catheter
- Purpose: To demonstrate substantial equivalence to a predicate device (Apollo Balloon Dilatation Catheter K133852 & K153742). This is a 510(k) submission for a physical medical device, not an AI/ML software.
- Study Type: Bench testing (in vitro) to assess physical characteristics and performance.
- Tests Performed:
- Dimensional verification
- Simulated Use
- Balloon Rated Burst Pressure
- Balloon Fatigue
- Balloon Compliance
- Balloon Inflation and Deflation Time
- Balloon Burst (in stent)
- Balloon Fatigue (in stent)
- Result of Tests: "In vitro bench testing met all acceptance criteria and demonstrated that the subject device performed as intended and did not impact the functionality of the device."
- Ground Truth (for this device): The ground truth for performance of this physical device is based on engineering specifications, recognized standards, and comparison to the predicate device's established performance.
Ask a specific question about this device