510(k) Data Aggregation

    K Number
    K173439

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-29

    (146 days)

    Product Code
    Regulation Number
    870.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Colorado 80108

    Re: K173439

    Trade/Device Name: Bloom2 Cardiac Stimulator System Regulation Number: 21 CFR 870.1750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bloom2 is intended for use in the electrophysiology laboratory during electrophysiology testing, under the direction of qualified personnel trained to administer electrophysiology procedures. The Bloom2 Cardiac Stimulator System is a programmable, electrical, pulse generator intended to provide diagnostic, electrical, cardiac stimulation. It is not intended for use, nor should it be used, as an external pacemaker for long-term support of the pacemaker dependent patient.

    Device Description

    The Bloom2™ Cardiac Stimulator System (Bloom²) is a programmable pulse generator used for electrophysiological testing of the human heart. The Bloom² has three main components, the Bloom2 Console, the Bloom2 Stimulator Unit and the Bloom2 Increment / Decrement Floater. The Console and the Floater contain the main user controls of the system. The Console consists of a touch screen and will allow the operator to setup the functions and parameters of the system. The user interface of the Console is based on the proven and familiar interface of the Bloom DTU-215B. The Floater provides additional user controls and will provide the operator with tactile feedback on those functions that the user is expected to use while going through their protocols. The Stimulator is the heart and brains of the system. The Stimulator has four, independent output channels that have a direct connection to the patient through third party catheters.

    AI/ML Overview

    The provided text is a 510(k) summary for the Bloom2 Cardiac Stimulator System. It describes the device, its intended use, and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets specific performance metrics in terms of diagnostic accuracy or clinical outcomes. Instead, it focuses on non-clinical performance and safety testing.

    Therefore, most of the requested information cannot be extracted from this document, particularly regarding studies related to AI performance, expert adjudication, or ground truth establishment relevant to an AI/diagnostic device.

    However, I can provide what is discussed regarding non-clinical performance:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Standard)Reported Device Performance
    IEC/EN 60601-1 3.1 Edition Safety Test ReportBloom2 device fulfills the referenced standard
    IEC 60601-1-1-2 Fourth Edition EMC Test ReportBloom2 device fulfills the referenced standard
    Software Life Cycle Processes Test ReportNot specified if it fulfills a named standard, but implicitly indicates software safety/functionality.
    FCC CFR 47, Part 15, subpart EMC Test ReportBloom2 device fulfills the referenced standard
    ICES-001 for Class A Digital EMC Test ReportBloom2 device fulfills the referenced standard
    Canada Differences to IEC 60601-1 Test ReportBloom2 device fulfills the referenced standard
    Withstand normal shipping and storage environments and meet labeling standards (21 CFR 801, EN 1041, ISO 15223)The packaged device will withstand normal shipping and storage and labeling shall meet the specified standards. (This is a statement of intent/design, not a report of specific test results within this summary).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    No information about sample size, test sets, or data provenance in the context of clinical or diagnostic performance is provided, as no such study was conducted. The non-clinical tests relate to the device itself and its components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    Not applicable, as no expert-adjudicated test set is detailed. The device is a stimulator, not an AI diagnostic tool requiring ground truth from medical experts in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This device is a cardiac stimulator, not an AI system for diagnostic image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    For the non-clinical tests, the "ground truth" would be the specifications and requirements of the safety and EMC standards (e.g., IEC 60601-1). The tests demonstrated that the device met these engineering and safety standards.

    8. The sample size for the training set:
    Not applicable, as no training set for an AI algorithm is mentioned.

    9. How the ground truth for the training set was established:
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092913

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-10-16

    (24 days)

    Product Code
    Regulation Number
    870.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    External Programmable Pacemaker Pulse Generator |
    | Classification Name | JOQ, Class II, 21 CFR 870.1750
    OCT 1 6 2009

    Re: K092913

    Trade/Device Name: EP-4TM Cardiac Stimulator Regulatory Number: 21 CFR 870.1750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EP-4 Cardiac Stimulator is indicated for use during clinical cardiac electrophysiology procedures. The EP-4 Cardiac Stimulator is intended to be used for diagnostic electrical stimulation of the heart for the purpose of refractory measurements, initiation and termination of tachyarrhythmias, and measurements of electrical conduction.

    Device Description

    The EP-4 Cardiac Stimulator consists of a touch-screen computer, an AC-powered stimulation module, which provides electrical stimulation to the heart during electrophysiology case studies. The EP-4 consists of a touch-screen computer that controls a two- or four-channel AC-powered stimulation module, and a keyboard. The stimulator is capable of performing single channel, simultaneous, or sequential stimulation at programmed output settings using existing/predicate programmed protocols. The device is independent and self-contained, with the exception of marker outputs and an external sync input connector. It can be configured as a stand-alone system or in conjunction with the EP-WorkMate™ Recording System. It may also be used with other EP recording systems.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the EP-4™ Cardiac Stimulator, which is an external programmable pacemaker pulse generator. The purpose of this submission is to demonstrate substantial equivalence to a predicate device, the EP-4™ Clinical Stimulator (K040207).

    The document does not contain acceptance criteria in terms of numerical performance metrics (e.g., sensitivity, specificity, accuracy, or specific physiological parameters with defined thresholds) or a formal study designed to "prove" the device meets such criteria through a statistical analysis against a ground truth.

    Instead, the submission focuses on demonstrating substantial equivalence to a predicate device. This is a regulatory pathway where a new device is shown to be as safe and effective as a legally marketed device that is not subject to premarket approval (PMA). The "study" described is primarily bench testing to confirm that design changes did not adversely affect safety or efficacy.

    Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is not:


    1. A table of acceptance criteria and the reported device performance

      The document does not provide a table of acceptance criteria with numerical performance metrics. The "acceptance criteria" in the context of this 510(k) submission are implicitly tied to demonstrating that the EP-4 Cardiac Stimulator is "substantially equivalent" to its predicate device (EP-4 Clinical Stimulator).

      The reported "performance" is described qualitatively:

      • "Bench testing was performed to confirm that the changes met design requirements and did not adversely affect the safety or efficacy of the product."
      • "Where design differences exist between the proposed device and the predicate device, performance testing demonstrated that these differences do not adversely affect the device's safety and effectiveness."
      • The device has the "same intended use and fundamental scientific technology" and "all technological characteristics... are substantially equivalent" to the predicate.

      No specific numerical performance values (e.g., stimulation accuracy within a certain mV or ms, battery life, etc.) that would typically be associated with explicit acceptance criteria are provided in this regulatory summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      Not applicable. The document describes "bench testing," which refers to laboratory testing of the device itself, not a clinical study involving patients or a dataset of patient data. Therefore, there is no "test set" in the context of patient data, nor is there information on data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      Not applicable. As this was bench testing for substantial equivalence, clinical ground truth establishment in the traditional sense (e.g., by medical experts for diagnostic accuracy) was not the focus. The "ground truth" for bench testing would be engineering specifications and predicate device performance, verified by engineers and quality control personnel. No information on specific expert qualifications for this type of verification is provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      Not applicable, as there is no specific "test set" of patient data requiring clinical adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      No. The EP-4 Cardiac Stimulator is a medical device for electrical stimulation, not an AI-assisted diagnostic tool. Therefore, an MRMC study and AI-related effectiveness are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      Not applicable. This device is a hardware system with embedded software for cardiac stimulation, not an algorithm being tested in isolation. The device's functionality is inherently "standalone" in operation once programmed, but this is distinct from testing an AI algorithm's performance without human interaction.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      For bench testing, the "ground truth" would be established against the design requirements, engineering specifications, and performance characteristics of the predicate device. The goal was to confirm that the changes "met design requirements" and did not "adversely affect the safety or efficacy" when compared to the predicate. This is verified through a combination of standard engineering tests, electrical measurements, and functional checks.

    8. The sample size for the training set

      Not applicable. The device's software is menu-driven and firmware-based, controlling stimulation functions. It's not described as a machine learning system that requires a "training set" in the AI sense. The software's development followed "St. Jude Medical's Quality System requirements" and "design controls requirements documented in 21 CFR 820.30," which dictate rigorous testing and verification, but not "training data" in the AI context.

    9. How the ground truth for the training set was established

      Not applicable, as there is no "training set" for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072200

    Validate with FDA (Live)

    Device Name
    STIMLAB
    Date Cleared
    2007-10-26

    (79 days)

    Product Code
    Regulation Number
    870.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K072200

    StimLab™ Programmable Cardiac Electrophysiological Stimulator Regulation Number: 21 CFR 870.1750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The StimLab Cardiac stimulator is intended to be used for diagnostic electrical stimulation of the heart for the purpose of initiation and termination of tachyarrhythmias, refractory measurements and measurements of electrical conduction.

    Device Description

    The StimLab™ is an EPS320 cardiac stimulator modified by the addition of a remote bedside touch monitor. The system consists of a manually controllable pulse generator issuing standard cardiac pacing pulses, normally controlled by a separate computer containing a graphic touch user interface and pre-programmed complex stimulation patterns suited to electrophysiology (EP) studies. In the StimLab™, a second remotely locatable beside control touch screen duplicates the video display and touch controls allowing verification and control of stimulator settings from two locations, one typically located in an electrophysiology laboratory central control room and the second at the operating table. The device is not a life support device and makes no diagnoses.

    AI/ML Overview

    The provided document is a 510(k) summary for the StimLab™ Dual Control Computerized Diagnostic Programmable Cardiac Stimulator. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not include detailed information regarding acceptance criteria, specific device performance metrics, or a study design with sample sizes, expert qualifications, or ground truth establishment.

    Therefore, many of the requested sections regarding acceptance criteria and the study that proves the device meets them cannot be filled from the provided text. The document focuses on regulatory clearance based on substantial equivalence, not on a detailed performance study with specific quantitative acceptance criteria.

    Here's an attempt to answer based only on the provided text, indicating where information is missing:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified in the document. The document states: "Appropriate risk-analysis driven product testing was conducted to evaluate conformance to product specification and substantial equivalence to the predicate device." However, the specific "product specifications" and the criteria for "conformance" are not detailed.The device was found to be "substantially equivalent" to the predicate device (Micropace EPS320 Cardiac Stimulator - K011826). The summary states: "The StimLab™ Cardiac Stimulator is equivalent to the predicate product. The indications for use, basic overall function, and materials used are equivalent." This implies that its performance met the standards of the predicate device.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified in the document.
    • Data Provenance: Not specified in the document. It mentions "product testing," but details about the origin or type of data involved are absent.
    • Retrospective/Prospective: Not specified in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not specified. The document does not describe a study involving expert-established ground truth for a test set. The review is based on substantial equivalence to a predicate device, which implies that the predicate device's performance (established through prior testing/clinical use) serves as the benchmark.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not specified. No adjudication method is mentioned as there's no described test set requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. The StimLab™ is described as a "dual control computerized diagnostic programmable cardiac stimulator," not an AI-assisted diagnostic tool for human readers. It's a stimulator for diagnostic electrical stimulation of the heart. Therefore, an MRMC study comparing human readers with/without AI assistance would not be relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. The device is a "stimulator" with "manually controllable pulse generator" and "graphic touch user interface." It is not an algorithm working independently to make diagnoses. The document explicitly states: "The device is not a life support device and makes no diagnoses." Its purpose is to facilitate diagnostic procedures by stimulating the heart.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not specified. Given the nature of a cardiac stimulator, "ground truth" would typically relate to the accuracy of the stimulation parameters, electrical output, timing, and safety, as compared to established physiological responses or engineering specifications. The document only states "conformance to product specification and substantial equivalence to the predicate device" as the basis for testing.

    8. The sample size for the training set

    • Not applicable/Not specified. This device is hardware with software control for stimulation, not a machine learning or AI algorithm that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established

    • Not applicable/Not specified. As mentioned above, there is no indication of a "training set" for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071348

    Validate with FDA (Live)

    Date Cleared
    2007-09-07

    (116 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Number: | 870.1425 computer, diagnostic, programmable |
    | | 870.1750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EMS-XL is indicated to acquire, filter, digitize, amplify, display and record electrical signals obtained during electrophysiological studies and related procedures conducted in an electrophysiological laboratory. Signal types acquired include ECG signals, direct cardiac signals and pressure recordings. Physiological parameters such as the diastolic, systolic and mean blood pressure, heart rate and cycle length are derived from the signal data, displayed and recorded.

    The system allows the user to monitor, display and record the signal data.

    The system allows the user to monitor the acquisition data, review the data, store the data, perform elementary caliper-type measurements of the data, and generate reports on the data. The system may display and record data received from other medical devices typically used during these procedures, such as Ablation RF generators. The system incorporates a stimulator intended to be used for diagnostic cardiac stimulation during electrophysiological testing of the human heart.

    Device Description

    The EMS-XL Cardiac Electrophysiology System consists of a software driven, multichannel amplifier and stimulator that are connected to a computer.

    Input signals include 12 lead surfaces ECG, two Blood pressure channels and 18 or 50 channel intra-cardiac ECG.

    The signals are digitized and sent to the computer for analysis, display and storage.

    Signals are displayed with sweep speeds of 25 to 300mm/sec in either predefined or customized configurations.

    The system consists of a computer, Front end amplifiers with integrated stimulator.

    Two display screens, one used for Real Time signal display and the other for Non real Time display for analysis and review and playback of stored signals.

    A Laser printer is used to print surface and intra-cardiac ECG and a CD or DVD is used for archive of signals.

    The two display screens are used to show the signals. The real time display, RT, shows the real time signal waveforms. Each waveform channel is color-coded for easy identification. The non real time display, NRT, serves as a Review monitor allowing the manipulation and processing of data, including caliper measurements, event marking, snapshot storage and final report processing.

    The stimulator is computer controlled and provides basic stimulation rate plus up to 4 stimulations delayed after the basic stimuli. Stimulation can be either synchronized to the cardiac electrical activity or non synchronized. Pacing protocols with automatic increment and decrement functions provides refractory period measurements.

    A variety of stimulation sequences can be created and stored for future use.

    The EMS-XL can be used for all types of electrophysiological procedures, including His bundle recording, Sinus Node Recovery Time - SNRT, Overdrive, Wenckebach, and tachy-arrhythmias. An interface to ablation generators provide Ablation Start and End information with data on the ablation parameters.

    Each waveform is color-coded for easy identification.

    The review monitor allows the manipulation and processing of data, including caliper measurements, event marking, snapshot storage and final report processing.

    Special display options support the signal analysis

    This includes Trigger mode and template comparison. Event marking. Holter mode: quick. minute-by-minute scrolls and jumps to event, On screen calipers: both real-time and review screen with auto measurement , Auto Tachycardia detection. Free text labeling, and Customized reports in Word™.

    The input to the amplifiers is via a patient connection box, to which the intra-cardiac catheter electrodes (not manufactured and not supplied by Mennen Medical) are connected.

    AI/ML Overview

    This 510(k) summary for the Mennen Medical EMS-XL Cardiac Electrophysiology System focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving the device meets specific performance acceptance criteria through clinical trials. The provided document is a predicate device comparison, which is a common approach for 510(k) submissions.

    Therefore, the acceptance criteria and study detailed below are interpreted from the provided predicate comparison tables and the accompanying "Rational for claims of equivalence".


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the specifications of the predicate devices. The reported device performance (EMS-XL) is presented in direct comparison to these predicate specifications. The table below synthesizes information from the various comparison tables provided in the document (Amplifier Spec comparison, Stimulator Spec Comparison, Comparison of Systems).

    FeatureAcceptance Criteria (Predicate)Reported Device Performance (EMS-XL)Deviations & Rationale (Effect on Safety/Intended Use/Performance)
    Amplifier SpecificationsCardiolab EP System (GE Prucka)EMS-XL Amplifier
    Physical (HxWxD, 32/64 ch)9.5x 14.0x 14.0 in (64/32 ch)4.5x 11.5x 8 in (32 ch); 9.0x 11.5x 8 in (64 ch)No (Smaller physical size, no effect)
    Temperature Operating0°C to +35°CSameSame
    Temperature Transport/Storage-15°C to +50°C-30°C to +65°CNo (Wider range, no effect on safety/intended use)
    Humidity Operating<95% RH at 35°C non-condensing30% - 75% non-condensingNo (Within predicate range, no effect)
    Humidity Transport/Storage<95% RH at 35°C non-condensing5% - 95% non-condensingNo (Wider range, no effect)
    Power Requirements100-240 V AC; 50-60 HzSameSame
    Sampling Rate1K, 2K and 4K1KHzNo (Meets clinical requirements, Note 2)
    CMMR100 dB minSameSame
    Input Impedance>1 Billion OhmsTypical 20 MΩ (Above 2.5 MOhm req by ANSI/AAMI EC11 par 3.2.9)No (Meets standard, no effect)
    Leakage Current (Patient Source)<10 uASame - Per ISO 60601-1Same
    Leakage Current (Patient Sink)<10 uASame - Per ISO 60601-1Same
    Leakage Current (Chassis)<100 uASame - Per ISO 60601-1Same
    ECG Outputs12 lead ECG producedSameSame
    ECG High Pass Filter0.05 Hz, 0.5 Hz, 5 Hz0.05, 0.2, 40, 80 HzNo (Provides necessary filtration, Note 3)
    ECG Low Pass Filter100 HzSameSame
    Gain50-10,000 in 8 settingsBetween 0 and 255 mm/mVolt – ContinuousNo (Equivalent gain, different terms/control, Note 4)
    Saturation RecoveryLess than 1 secSameSame
    Notch FilterPower line (50/60 Hz)SameSame
    Intracardiac Inputs (32ch)32 intracardiac, 4 pressure, 10 ECG36 intracardiac (18 ch), 2 pressure, 10 ECG (12 ch)No (Covers wide range of clinical applications, Note 5)
    Channel SwitchingAny input can be switched to any output within a BlockEach channel can be Bipolar or Unipolar with Manual switchingNo (Difference overcome by manual switching, Note 6)
    IECG High Pass FilterDC, 0.05 Hz, 0.5 Hz, 5.0 Hz, 30 Hz, 100 Hz0.05, 0.2, 40, 80 HzNo (Provides necessary filtration, Note 7)
    IECG Low Pass Filter500 Hz, 2,000 Hz500 HzNo (500Hz same, 2000Hz not relevant for 1KHz sampling, Note 8)
    Pressure InputsCompatible with all standard external pressure transducers. 5uV/V/mmHgSameSame
    Excitation Voltage5V DC10 V DC (+5 to -5V)No (Sensitivity is important parameter, which is same, Note 9)
    Pressure Channels OutputsUp to 4 pressure channels2 pressure channelsNo (Covers wide range of clinical applications, Note 5)
    Stimulator SpecificationsMicropace model EPS 320EMS-XL (integrated)
    Isolated Stimulus channels22Same
    Pulse Amplitude Range0.1 to 25 mA into 800 Ω load0.1 to 25 mA into 1500 Ω loadNo (Same current range, higher load impedance source, Note 10)
    Pulse Amplitude Increment0.1 mA up to 1 mA Amplitude0.1 mA within whole rangeNo
    Pulse Amplitude Accuracy+/- 2% or +/- 0.2 mA (whichever is larger)SameSame
    Pulse Duration Range0.5 to 10 msec0.1 to 9.9 msecNo
    Pulse Duration Increment0.5 - 10 msec, increment of 1-10 msec0.1 msecNo (Smaller increments, no effect on performance)
    ISI Range (Pace)180 msec to 9990 msecSameSame
    ISI Range (Burst)30 - 9900 msecSameSame
    ISI Increment1 msec10 msecNo (Easier control, Note 11)
    Sequential Delay (AV) Range10-1000+/- 1 msec (maximum ISI - 50 msec)11 - 250 msecNo (No clinical significance for longer intervals, Note 12)
    Number of Extra-stimuli6 (S2-S7)4 (S2-S5)No (4 extra sufficient for most clinical apps, Note 13)
    Sensing (ECG sync)External: 50 – 2000 mV; Internal: Pacing catheter tipInternal from any surface or intra-cardiac channelNo
    Sync OutputMultiple sync outputs for external devicesNoNo (Integrated system doesn't need external sync, Note 14)
    Power Source (Stimulator)Mains 220/110 to 14.5 VDC, 750 mA low voltage power supply. Backup batteryIntegrated with Amplifier power (+5V, -+12V, -12V)Yes (Integrated, but safety covered by other precautions, Note 15)
    Programmed Protocols (Auto TP/Ectopic/Stop Tachycardia)YesNoYes (Physician decision used instead due to safety burden, Note 16)
    Backup StimulatorIntegratedNo backup stimulator (Requires external backup)Yes (User manual requires external backup, Note 17)
    System SpecificationsCarto® XP System (GE Prucka)EMS-XL
    ProcessorIntel® 2.66GHz Pentium IV Xeon or greaterIntel® 2.4GHz Pentium IV or greaterNo
    RAM512 MBSameSame
    Hard drives2 x 40 GB80 GBNo
    Monitors1 or 22No
    Holter ModeYesIntegrated Holter mode with full disclosureNo
    Ablation ConnectivityConnectivity to RF Ablation generator(s)SameSame
    Statistics and InventoryXi2 Data Management SystemNoNo (Administrative package, no direct use on EP procedure, Note 18)
    MappingIsochronal and Isopotential MapsNoNo (External devices can be used for mapping, Note 19)
    Vital SignsIntegrated Vitals Monitoring with Audible IndicatorsNot integrated (Integrated tachycardia message; recommends external monitor, Note 20)No

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a predicate comparison for a 510(k) submission, not a direct clinical study with a specified "test set" as one might find for an AI/ML device.

    • Sample Size: Not applicable in the context of a prospective or retrospective clinical study of user data. The "sample" here refers implicitly to the specifications and functionalities of the two predicate devices being compared.
    • Data Provenance: The data provenance is the technical specifications and reported features of the two predicate devices:
      • K993414 Cardiolab EP system (Prucka Engineering, Inc)
      • K011826 Micropace model EPS 320 (Micropace Pty Ltd.)
        The submission itself is from Mennen Medical Ltd., Israel. The clinical experience mentioned for EMS-XL notes use by clinics "out of the USA" since December 2003, implying international (likely European) usage.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable in this type of submission. The "ground truth" for this submission is the established and legally marketed performance and specifications of the predicate devices. The review and interpretation of these specifications for equivalence determination would be performed by the manufacturer's regulatory affairs team and then by FDA reviewers.

    4. Adjudication Method for the Test Set

    Not applicable. This is a technical specification comparison, not a clinical trial requiring adjudication of patient outcomes or expert reads. The "adjudication" is essentially the manufacturer's justification (Rational for claims of equivalence) and the FDA's acceptance of those justifications for "substantial equivalence."

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No, an MRMC study was not done. This device is a cardiac electrophysiology system, which includes signal acquisition, display, recording, and stimulation capabilities. It is not an AI/ML diagnostic interpretation tool that would typically undergo an MRMC study to assess reader performance with and without AI assistance. The submission confirms that the EMS-XL system is a software-driven, multichannel amplifier and stimulator connected to a computer, focused on acquiring and analyzing physiological signals.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, implicitly. The comparison tables are a standalone comparison of the technical specifications and functions of the EMS-XL device against the technical specifications and functions of its predicate devices. The "performance" being evaluated is the device's inherent functional capabilities as designed and manufactured, not an AI algorithm's interpretive ability.

    7. The Type of Ground Truth Used

    The "ground truth" used is the published and legally established technical specifications and performance characteristics of the predicate devices (Cardiolab EP system and Micropace model EPS 320) as detailed in their 510(k) clearances and product documentation. This is a regulatory "truth" rather than a clinical "truth" (like pathology or outcomes data).

    8. The Sample Size for the Training Set

    Not applicable. This document does not describe the development of an AI/ML algorithm that would require a "training set." The EMS-XL is a hardware and software system for electrophysiological studies.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for an AI/ML algorithm is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070743

    Validate with FDA (Live)

    Device Name
    Z6 STIMULATOR
    Date Cleared
    2007-07-02

    (108 days)

    Product Code
    Regulation Number
    870.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Circle Broomfield, CO 80020

    Re: K070743

    Trade Name: Z6 Stimulator System Regulation Number: 21 CFR 870.1750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Z6 Stimulator System is an electrical stimulus generator for diagnostic cardiac stimulation electrophysiological testing of the human heart.

    Device Description

    The Z6 Stimulator is a cardiac pulse generator designed for use in an Electrophysiology Laboratory during electrophysiological studies. The system operates in conjunction with ECC Recording devices to deliver electrical stimulus to the patient's heart through up to four simultaneously operable output channels. The Z6 Stimulator is controlled by a CPU based Workstation with Windows XP. This provides programmability and flexibility with regard to the stimulus program and allows access to system users. Additionally, the system features a graphical interface with a Windows XP look and feel.

    Each of the four independent output channels of the Z6 Stimulator is capable of generating between 0.1 and 25 mAs of current to the patient, with voltages up to 20 Volts. The System is capable of delivering pulses on all channels simultaneously. The duration of the pulses can be programmed to be between 0.5 and 10 ms, maintaining timing accuracy (> +/- 0.15 ms).

    AI/ML Overview

    The provided document, K070743, is a 510(k) premarket notification for the Z6 Stimulator, a programmable cardiac electrophysiological stimulator. The purpose of this notification is to demonstrate that the Z6 Stimulator is substantially equivalent to a legally marketed predicate device, the Micropace Pty Ltd. EPS320 Clinical Stimulator.

    Here's an analysis of the acceptance criteria and the study presented in the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document uses the specifications of the predicate device, the Micropace EP320 Stimulator, as the acceptance criteria for the Z6 Stimulator's performance. The "Non-clinical Testing" section explicitly states that the Z6 Stimulator's output pulses were compared to these specifications and "found to be within the specifications."

    Acceptance Criteria (Based on Micropace EP320)Z6 Stimulator Reported Performance
    Output Pulse Timing AccuracyWithin specifications
    Output Pulse Amplitude AccuracyWithin specifications
    Isolated Stimulus Channels: 24 (Improved, exceeding criteria)
    Pulse Amplitude: 0.1-25 mA0.1-25 mA
    Pulse Duration: 0.5-10 ms0.5 to 10 ms
    Interstimulus Interval: 30-9900 ms30-10000 ms (Improved)
    Number of Extra Stimuli: 68 (Improved, exceeding criteria)

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document does not specify a sample size for the test set. The non-clinical testing appears to be a technical verification comparing the Z6 Stimulator's output characteristics against the published specifications of the predicate device.
    • The data provenance is not explicitly stated as country of origin or retrospective/prospective. As it's a technical verification of the device itself rather than patient data, these categories might not be applicable in the traditional sense. It's an internal validation of the Z6 Stimulator's hardware and software performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This is not applicable to this type of submission. The ground truth for this technical verification is the published specifications of the predicate device, which are objective engineering parameters. Clinical expert consensus is not used for this.

    4. Adjudication Method for the Test Set:

    • This is not applicable. The "test set" here refers to the device's output parameters which are objectively measured and compared to numerical specifications. There is no need for expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC comparative effectiveness study was not done. This submission focuses on the substantial equivalence of device specifications, not on clinical effectiveness in human readers. The Z6 Stimulator is a tool used by clinicians, not an AI diagnostic tool that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, in spirit. The non-clinical testing for the Z6 Stimulator is inherently a "standalone" evaluation of the device's functional performance (timing, amplitude accuracy) against established specifications. It's the performance of the device itself, independent of a human operator's cognitive input in interpreting results.

    7. The Type of Ground Truth Used:

    • Engineering specifications and published data of the predicate device. The Z6 Stimulator's output pulses (timing and amplitude accuracy) were compared against the "specifications of the EP320 Stimulator output pulses."

    8. The Sample Size for the Training Set:

    • Not applicable. This submission is for a medical device (a stimulator), not an AI algorithm that requires a "training set" in the machine learning sense. The device is designed based on engineering principles and validated against specifications.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As stated above, there is no "training set" in the context of this device and submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K041442

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-09-14

    (105 days)

    Product Code
    Regulation Number
    870.1750
    Age Range
    18 - 120
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Berlin, NJ 08091-9293

    Re: K041442

    Trade Name: EP-4 Clinical Stimulator Regulation Number: 21 CFR 870.1750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EP-4 Clinical Stimulator is intended to be used for diagnostic electrical Stimulation of refractory measurements, initiation and termination of tachyarrhymias, and measurements of electrical conductions.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry. I cannot answer your question. The document does not describe the acceptance criteria, the study details, or the performance of the device. It is a 510(k) clearance letter and an "Indications for Use" statement for a medical device called the "EP-4 Clinical Stimulator." It confirms that the device is substantially equivalent to legally marketed predicate devices and outlines its intended uses.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040207

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-02-27

    (28 days)

    Product Code
    Regulation Number
    870.1750
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Berlin, NJ 08091-9293

    K040207 Re:

    Trade Name: EP-4 Clinical Stimulator Regulation Number: 21 CFR 870.1750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EP-4 Clinical Stimulator is intended to be used for diagnostic electrical Stimulation of the heart for the purpose of refractory measurements, initiation and termination of tachyarrhymias, and measurements of electrical conductions,

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) clearance letter from the FDA for a device called the "EP-4 Clinical Stimulator." This document confirms that the FDA has reviewed the premarket notification and found the device to be substantially equivalent to legally marketed predicate devices.

    However, this document does not contain any information about acceptance criteria, device performance results, sample sizes, ground truth establishment, or study designs (like MRMC or standalone studies). This letter is a regulatory approval document, not a clinical study report or a technical performance specification.

    Therefore, I cannot fulfill your request for the detailed information about acceptance criteria and the study that proves the device meets them based on the provided text. The requested information would typically be found in a separate submission, such as an engineering report, validation study report, or clinical trial documentation, which is not part of this 510(k) clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011826

    Validate with FDA (Live)

    Date Cleared
    2002-01-24

    (227 days)

    Product Code
    Regulation Number
    870.1750
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Electrophysiological Stimulator External programmable pacemaker pulse generator (Per 21 CFR section 870.1750
    Canterbury Road Campsie, NSW 2194 Australia

    Re: K011826 EPS320 Cardiac Stimulator Regulation Number: 870.1750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EPS320 Stimulator System is an electrical stimulus generator for diagnostic cardiac stimulation during electrophysiological testing of the human heart.

    Device Description

    The Micropace EPS320 Cardiac Stimulator (Clinical Stimulator) is an external programmable computerised diagnostic cardiac stimulator. It is used in specialised hospital electrophysiological diagnostic laboratories by specialist cardiologists to electrically stimulate the heart to initiate and terminate tachyarrhythmias and allow measurement of refractory and conduction properties of the heart by third party equipment. The EPS320 system consists of a Stimulus Generator Unit (SGU) comprising of a manufactured metal instrument case and a standard IBM clone type Personal Computer (PC), the two interconnected by a serial data connection. The device has two independent fully programmable and isolated constant current pulse generator channels intended for temporary programmacro and locativa third party transvenous intracardiac electrodes. The pacing stimulus is output via a Stimulus Connection Box and may typically be routed to the intravenous pacing electrodes via third party electronic switching equipment. During normal operation of the EPS320, custom software on the PC provides keyboard input device and a graphical user interface which the operator uses to control the stimulation process in real time. The custom PC software interprets the user instructions and sends specific real-time commands to the I he ousehi I O software and and to control its function and generate appropriate pacing stimulus pulses. The two isolated outputs can deliver stimulus pulse widths from 0.5ms to 10ms, adjustable in 0.5 to 1ms steps with current amplitudes adjustable from 0.1mA to 25mA in steps of 0.1mA. Current delivery is limited within the specified range by a maximum available output voltage of 26V. The EPS320 is not a life support device and may be used only in the presence of a backup cardiac THE LE SES IS .In case of failure of mains power supply, malfunction of the PC or its software, or loss of the data link, the SGU can operate independently of the PC in the Manual Backup mode, powered by an internal backup battery. This operational mode can provides regular pacing stimuli at intervals and pulse current amplitudes programmable from and displayed on the SGU's front panel. It may be used to briefly support a bradycardic patient until temporary external pacing is established using an approved temporary pacemaker. A secondary electrically self-contained Emergency Fixed Pace approved willing the front of the SGU may be used for the same purpose in case of loss of normal SGU function.

    AI/ML Overview

    The provided text describes the Micropace Model EPS320 Clinical Stimulator and seeks 510(k) clearance based on its substantial equivalence to the predicate device, the DCI Model EP-2 Clinical Stimulator. The document primarily focuses on technical specifications and functional comparisons rather than a study demonstrating performance against specific acceptance criteria in a clinical setting.

    Here's an analysis based on the provided input:

    Acceptance Criteria and Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of a performance study with numerical thresholds. Instead, it aims to demonstrate that the EPS320 is substantially equivalent to the predicate device (DCI Model EP-2) and is as safe and effective. The "acceptance criteria" are implicitly met if the FDA agrees that the EPS320's specifications and functionality are comparable or improved, and its differences do not raise new questions of safety or effectiveness.

    The "reported device performance" is presented through a direct comparison with the predicate device's specifications in Tables 5-1 and 5-2.

    Table of Acceptance Criteria (Implicit) and Reported Device Performance

    Feature/ParameterImplicit Acceptance Criteria (Comparable to Predicate Device)Micropace Model EPS320 Performance (Reported Values)Predicate Device (DCI Model EP-2) Performance (Reported Values)Rationale for Equivalence (as per submission)
    1. Isolated Stimulus ChannelsSame number of channels.22Same number of Channels as the EP-2.
    2. Pulse AmplitudeRange sufficient for therapeutic purposes; comparable to EP-2.0.1 to 25mA into 800Ω load Increment: 0.1 mA up to 1mA Amplitude Accuracy: ± 2% or ± 0.2 mA (whichever is greater)0.1-20 mA (2500Ω load) Accuracy: 5% or 0.2 mA (whichever is larger)EPS320 output current is sufficient for therapeutic purposes and substantially the same as the EP-2.
    3. Pulse DurationWithin accepted clinical requirements; comparable to EP-2.0.5 to 10 msec Increment: 0.5 – 10msec, increments of 1 – 10 msec Accuracy: ± 0.15ms0.5-10 msec (maximum 10% of ISI) Increment: 0.1 msec Accuracy: 0.05 msec (Pulse amplitude > 2 mA)Pulse Duration is within the profession's accepted requirements as attested by 4 years of post-market surveillance of the EPS320.
    4. Interstimulus Interval (ISI)Substantially the same as EP-2.180 msec to 9990 msec ± 1msec or 0.1% (whichever is greater) (Range) 30 - 9900 msec ± 1msec or 0.1% (whichever is greater) (Burst) Increment: 1 msec100-10000 ± 1 msec (Range) 10-1000 ± 1 msec (Burst) Increment: 1 msecSubstantially the same as the EP-2.
    5. Sequential (AV) DelaySubstantially the same as EP-2, with enhanced safety.Min: 25 - 980msec (maximum ISI – 120 msec) Increment: 1 msec10-1000 ± 1 msec (maximum ISI - 50 msec) Increment: 1 msecA-V delay is substantially the same as the EP-2. EPS320 is limited to ISI less 120 msec which is considered an enhanced safety feature whilst retaining equivalent operational effectiveness.
    6. Reprogrammed ProtocolsEquivalent functionality to EP-2, covering common clinical needs.Equivalent protocol to all but 1 protocol in the EP-2 (Refractory Threshold omitted as no longer in common clinical use). Contains additional protocols like Multi-Sx, Nodal ERP / RSync_S2, Overdrive Burst / Load ATP.Threshold, Refractory study, Wenckebach block, SA Node recovery, Pace, Burst, Overdrive, Underdrive, Arrhythmia Induction, Refractory Threshold.EPS320 contains an equivalent protocol to all but 1 protocol in the EP-2, providing substantially equivalent functionality. One protocol, the Refractory Threshold protocol present in the EP-2 is no longer in common clinical use in the experience of the expert designer and was not included in the EPS320.
    9. Sensing (ECG synchronization)External ECG input range appropriate; enhanced safety features.External: 50 - 2000mV Internal: Pacing catheter tip Trigger lockout (refractory time): 50 - 5000msec ECG delay: 50-5000msecAutomatic or manual trigger setting - Sensitivity: 1-500 mV Automatic or manual trigger setting - Trigger lockout (refractory time): 10-1000 msec (100 Hz max) Automatic or manual trigger setting - ECG delay: 100-5000 msecExternal ECG input range (mV) for the EPS320 is appropriate for common external ECG sources. Presence of pacing catheter tip sensing in EPS320 is considered an enhanced safety feature. Longer minimum lockout (50 vs. 10msec) is an enhanced safety feature reducing susceptibility to noise, without significant loss of functionality. Lockout period and ECG delay are substantially equivalent to the EP-2.
    Hardware PlatformSimilar architecture; safe and effective.Standard "PC" Personal Computer / custom "Stimulus Generator Unit ""Control Terminal" custom computer / custom "Stimulus Module"Both devices have similar architectures. The EPS320's 'open PC architecture' uses verification of software environment and resource integrity and is based on extensive risk analysis. The absolute safety and efficacy of the EPS320 hardware platform has been demonstrated in clinical practice and is thus at least equivalent to that of the predicate EP-2 device.
    Backup Pacing CircuitsAs safe or more safe than EP-2.Yes (2) – Manual Backup operation mode and Emergency Fixed Pace circuit.NoThese functions are safety enhancements over the predicate device EP-2.
    Device Feedback to UserAs safe or more safe than EP-2.Open Pacing Circuit Warning: Yes Stimulation impedance display: Yes Graphical display of ECG Trigger: YesNoThese features are safety enhancements over the predicate device EP-2.
    Power SourceComparable reliability and safety.Mains operated with 2 backup battery systems (12V lead acid, 9V Lithium).Internal rechargeable 6V battery.Mains source of power was considered safer than a battery as it removes the possibility of lack of device availability due to depleted batteries. Batteries in EPS320 are for fault conditions.

    Breakdown of Study Information:

    1. Sample Size Used for the Test Set and Data Provenance:

      • The submission does not describe a specific "test set" or a formal clinical study with a defined sample size to prove the device meets acceptance criteria in a quantitative manner.
      • The provided information states that the EPS320 is an "expert-designed product, based upon the electrophysiological expertise of the cardiologist Managing Director and on extensive post market experience with an earlier EPS220 model in service in eight (8) Australian hospitals since 1995." It also mentions that "The current and earlier EPS320 models have been in clinical use in more than 40 clinical centres predominantly in Australia, Europe and India, as well as in Turkey, Vietnam and China for up to 4 years without significant adverse effects." This "post-market experience" serves as informal evidence of performance and safety rather than a structured clinical test set.
      • Data Provenance: Australia, Europe, India, Turkey, Vietnam, and China.
      • Retrospective/Prospective: The "post-market experience" suggests a retrospective review of observational data from clinical use, not a prospective study designed to evaluate the EPS320 against specific endpoints.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Since there's no formal "test set" or ground truth establishment process described for a specific study, this information is not directly applicable.
      • However, the device design itself is based on "electrophysiological expertise of the cardiologist Managing Director." This individual serves as a key "expert" influencing the device's design and functionality. The submission does not detail specific qualifications of this Managing Director beyond "cardiologist."
    3. Adjudication Method for the Test Set:

      • Not applicable, as no formal test set and ground truth establishment with adjudication is described.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a cardiac stimulator, not an AI-driven diagnostic imaging device that uses multi-reader studies.
    5. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not applicable. The EPS320 is a medical device that requires human operation by specialist cardiologists to apply electrical stimulation. It is not an autonomous algorithm.
    6. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

      • No formal ground truth for a test set is described. The closest equivalent to "ground truth" for the device's design and effectiveness claims comes from:
        • Expert experience: "electrophysiological expertise of the cardiologist Managing Director."
        • Clinical outcomes/safety data: "4 years of post-market surveillance of the EPS320" and "clinical use in more than 40 clinical centres... for up to 4 years without significant adverse effects." This implicitly refers to positive clinical outcomes and lack of adverse events as evidence of safety and effectiveness in real-world scenarios.
        • Comparison to predicate: The EP-2's established safety and effectiveness serve as the baseline "ground truth" for the EPS320's substantial equivalence claim.
    7. The Sample Size for the Training Set:

      • Not applicable. This submission does not describe an AI or machine learning algorithm that requires a "training set." The extensive clinical use of "current and earlier EPS320 models" in "more than 40 clinical centres" acts more as an extended real-world validation of the device's design and safety.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable, as there is no training set described for an AI/ML algorithm.