Search Results
Found 37 results
510(k) Data Aggregation
(80 days)
Light Chain Assay
Diazyme Human Lambda (λ) Free Light Chain Assay
Regulation Number: 21 CFR 866.5550
| s) | Classification | Regulation Section | Panel |
|---|---|---|---|
| DFH, DEH | Class II | 21 CFR 866.5550 |
The Diazyme Human Kappa (κ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS) in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
The Diazyme Human Lambda (λ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS) in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
No modification is made to the kit components for the Diazyme Human Kappa (κ) and Lambda (λ) Free Light Chain Assays cleared in K153394.
The Diazyme Human Kappa (κ) Free Light Chain Assay is comprised of the following reagents:
- Reagent R1: Tris buffer solution
- Reagent R2: Carboxyl Polystyrene particles coated with polyclonal rabbit and goat anti-Kappa FLC antibodies
- Calibrators: A five-level set in liquid form (5 x 1 mL)
Material required but not provided
- Controls Set: Two levels, serum based, liquid form
The Diazyme Human Lambda (λ) Free Light Chain Assay is comprised of the following reagents:
- Reagent R1: Tris buffer solution
- Reagent R2: Carboxyl Polystyrene particles coated with polyclonal rabbit anti-Lambda FLC antibodies
- Calibrators: A five-level set in liquid form (5 x 1 mL)
Material required but not provided
- Controls Set: Two levels, serum based, liquid form
N/A
Ask a specific question about this device
(87 days)
Optilite Freelite Mx Kappa Free Kit
Optilite Freelite Mx Lambda Free Kit
Regulation Number: 21 CFR 866.5550
Test:** Quantitative, immunoturbidimetry
Regulatory information:
Regulation section: 21 CFR 866.5550
Optilite® Freelite Mx Kappa Free Kit:
The Optilite Freelite Mx Kappa Free Kit is intended for the quantitative in vitro measurement of Kappa free light chains in serum and urine using the Binding Site Optilite analyser. Measurement of free light chains in serum and urine aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE); and in serum aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS). Results of the free light chain measurements should always be interpreted in conjunction with other laboratory and clinical findings.
Optilite® Freelite Mx Lambda Free Kit:
The Optilite Freelite Mx Lambda Free Kit is intended for the quantitative in vitro measurement of Lambda free light chains in serum and urine using the Binding Site Optilite analyser. Measurement of free light chains in serum and urine aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE); and in serum aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS). Results of the free light chain measurements should always be interpreted in conjunction with other laboratory and clinical findings.
The determination of soluble antigen concentration by turbidimetric methods involves the reaction with specific antiserum to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument.
The provided FDA 510(k) clearance letter and its associated 510(k) Summary pertains to the Optilite Freelite Mx Kappa Free Kit and Optilite Freelite Mx Lambda Free Kit. The submission sought to add a claim for the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS) to the intended use statement (in serum).
Here's an analysis of the acceptance criteria and the study proving the device meets these criteria, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly list acceptance criteria in terms of pre-defined numerical thresholds for sensitivity, specificity, or positive/negative rates that were required to be met for the MGUS claim. Instead, it describes clinical performance studies designed to demonstrate the utility of the device for MGUS evaluation. The conclusion states that "The studies generated successful results where all pre-defined acceptance criteria were met," implying these criteria were internal to the manufacturer's study design and not explicitly detailed in the public document as numerical targets.
However, we can infer the de facto "reported device performance" based on the results of Study 1 and Study 2.
Inferred Performance/Results:
| Performance Metric | Freelite Mx Kappa & Lambda Kits (for MGUS evaluation) | Commentary |
|---|---|---|
| Study 1: Determination of Positive Rate Ratio in MGUS | This study aimed to show the "positive rate" of the test in clinically confirmed MGUS patients. The "positive rate" here refers to the percentage of MGUS patients whose FLC ratio was abnormal based on the device's reference intervals. It is not a traditional sensitivity as it's not classifying all MGUS cases, but rather indicating how often the FLC ratio is abnormal within the MGUS population. | |
| All MGUS positive | 56.3% (n=129/N=229) | Implies that 56.3% of confirmed MGUS patients had an abnormal FLC ratio according to the device's criteria. |
| Light chain MGUS | 100.0% (n=10/N=10) | Shows that the device identified all LC-MGUS cases as "abnormal" (test-positive). This is a strong indicator for this specific subtype. |
| Non-light chain only MGUS | 54.3% (n=119/N=219) | Shows the positive rate for MGUS cases that are not LC-MGUS. |
| Study 1: Determination of Negative Rate Ratio in Disease Controls (non-MGUS) | This study aimed to show the "negative rate" of the test in non-MGUS patients with other conditions. The "negative rate" here refers to the percentage of non-MGUS patients whose FLC ratio was within the normal reference interval. It is not a traditional specificity as it's not classifying all non-MGUS cases as negative, but rather indicating how often the FLC ratio is normal in a disease control group. | |
| Disease Controls (Non-MGUS) | 91.9% (n=125/N=136) | Implies that 91.9% of patients with polyclonal hypergammaglobulinemia (non-MGUS disease controls) had a normal FLC ratio. |
| Study 2: Evaluation of MGUS Progression (Stable vs. Progressive) | This study retrospectively evaluated the ability of the device to track stable versus progressive MGUS based on FLC changes. The acceptance criteria here would likely revolve around the consistency of FLC results with the clinical status (stable or progressive). | |
| Stable MGUS test positive patients | 95.6% (n=43/N=45) | This indicates that 95.6% of patients with clinically stable MGUS showed FLC results consistent with stability (i.e., less than 25% increase in involved FLC). |
| Progressive MGUS test positive patients | 50.0% (n=2/N=4) | This indicates that 50.0% of patients with clinically progressive MGUS (converting to MM) showed FLC results consistent with progression (i.e., FLC ratio outside reference interval AND >= 25% increase in involved FLC from baseline). The small sample size (N=4) for progressive MGUS patients should be noted for this metric. |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Study 1 (Evaluation of MGUS and disease controls at single time points):
- MGUS Test Set: 229 samples from patients with clinically confirmed MGUS.
- Disease Control Test Set (non-MGUS): 136 samples from patients with polyclonal hypergammaglobulinemia.
- Provenance: Retrospective study. The country of origin is not explicitly stated, but the submitter (The Binding Site Ltd) is based in the United Kingdom.
-
Study 2 (Evaluation of MGUS progression):
- Total Patients: 49 MGUS patients.
- Stable MGUS Cohort: 45 patients.
- Progressive MGUS Cohort: 4 patients (who progressed from MGUS to MM).
- Total Samples: 185 samples (up to 4 individual draws for stable, up to 6 for progressive).
- Provenance: Retrospective study. Country of origin not explicitly stated.
- Total Patients: 49 MGUS patients.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- The document states that the clinical diagnostic criteria for MGUS and disease controls were "confirmed with each site" and "as practiced clinically."
- For the progressive MGUS study, changes were defined as clinical progression from MGUS to MM.
- No specific number of experts or their qualifications (e.g., "Radiologist with 10 years of experience") are mentioned in the provided text. The ground truth relies on "clinical diagnosis" or "clinical truth," which typically implies the consensus opinion of treating clinicians or established diagnostic criteria within the medical community (e.g., IMWG guidelines for MGUS/MM).
4. Adjudication Method for the Test Set
- The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing the clinical ground truth. It relies on "clinical diagnosis" or "clinical truth" established by the sites providing the samples.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC study was done. This device is an in vitro diagnostic (IVD) kit for quantitative measurement of biomarkers, not an image-based AI device that would typically involve human readers interpreting images. The closest analog would be a clinical utility study comparing outcomes with and without the FLC test, but that is beyond the scope of a 510(k) for an IVD kit unless explicitly requested for a novel claim.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- This is a standalone study. The Optilite Freelite Mx Kappa and Lambda Free Kits are laboratory assays. The "performance data" presented (positive rates, negative rates, tracking of progression) are derived solely from the measurements made by the device (Optilite Analyser) on patient samples, compared against established clinical diagnoses (ground truth). There is no "human-in-the-loop" aspect to the device's direct measurement and result generation. The results are then interpreted by clinicians in conjunction with other findings, but the device's output itself is standalone.
7. Type of Ground Truth Used
- The ground truth used was clinical diagnosis/clinical truth.
- For Study 1: "clinically confirmed MGUS" and "polyclonal hypergammaglobulinemia confirmed by study testing... with supporting clinical information." The diagnostic criteria for MGUS were stated to align with or be similar to those outlined by the International Myeloma Working Group (IMWG).
- For Study 2: "clinically determined stable or progressive status" for MGUS patients, with progression defined as conversion from MGUS to Multiple Myeloma (MM) based on clinical diagnosis. FLC results criteria adapted from IMWG guidelines were used for evaluation of the device's performance, but the patient's status (stable/progressive) was the "clinical diagnosis."
8. Sample Size for the Training Set
- The document states, "The devices in this submission have not materially changed since originally cleared under K150658." This implies that the core analytical performance studies (e.g., analytical measuring range, precision, accuracy) were previously established and referenced from prior 510(k) submissions (K173732 and K150658).
- For the new MGUS claim, the performance data presented are for clinical validation on the test sets described in points 1 and 2.
- The document does not describe a distinct "training set" for the clinical claim of MGUS evaluation, nor is it typical for IVD kits to have a "training set" in the machine learning sense for their clinical claims. The performance studies demonstrate the device's ability to measure FLCs in patient populations relevant to MGUS, and these measures are compared against clinical ground truth. The assay itself (reagents, instrument) would have been "trained" (i.e., optimized and validated analytically) during its initial development and previous clearances.
9. How the Ground Truth for the Training Set Was Established
- As a "training set" for the clinical claim is not explicitly mentioned or relevant for this type of IVD 510(k), this question is not directly applicable. For the device itself, analytical validation and calibration would have established its operational parameters, but this isn't "ground truth" in the diagnostic performance sense. The clinical ground truth for the test sets was derived from established clinical diagnoses and diagnostic criteria.
Ask a specific question about this device
(267 days)
Norcross, Georgia 30093
Re: K231601
Trade/Device Name: FLC Kappa, FLC Lambda Regulation Number: 21 CFR 866.5550
| 21CFR sec. 866.5550 - Immunoglobulin (light chainspecific) immunological
Site Freelite®Human Kappa Free Kit for use onthe Siemens BN™ II | K031016 | 866.5550
Site Freelite®Human Lambda Free Kit for useon the Siemens BN™ II | K031016 | 866.5550
The FLC Kappa kit is intended for the quantification of Kappa free light chains in human serum from adults with an Enzyme-Linked Immunosorbent Assay (ELISA) procedure. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma and AL amyloidosis. It must be used in conjunction with other laboratory and clinical findings. For In Vitro Diagnostic Use only.
The FLC Lambda kit is intended for the quantification of Lambda free light chains in human serum from adults with an Enzyme-Linked Immunosorbent Assay (ELISA) procedure. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma and AL amyloidosis. It must be used in conjunction with other laboratory and clinical findings. For In Vitro Diagnostic Use only.
The FLC Kappa and FLC Lambda test kits are intended for the quantification of free light chains in human serum from adults with an Enzyme-Linked Immunosorbent Assay (ELISA) procedure utilizing specific antibodies targeting anti-Lambda free light chains.
It is carried out in 8 successive steps:
- Incubation of the previously diluted samples and calibrators, in the wells of the microplate, where specific free light chain antibodies are fixed.
- Washing of the wells to remove elements that have not been fixed by the anti-free light chain antiserum.
- Incubation with an anti- light chain antiserum (Kit specific) conjugated to peroxidase.
- Washing of the wells to remove the excess of antiserum conjugated to peroxidase.
- Incubation with peroxidase substrate.
- Stopping of the enzymatic reaction with an acidic solution.
- Reading of the optical density by absorbance spectrophotometry at 450 nm of the colored product.
- Calculation of the free light chain concentration of the sample using a calibration curve obtained with calibrators that have been analyzed on the same microplate.
The provided document is an FDA 510(k) clearance letter and associated summary for the Sebia FLC Kappa and FLC Lambda kits. These kits are in vitro diagnostic (IVD) devices used for quantifying free light chains in human serum. They are immunoassay-based tests, not AI/ML-driven devices.
Therefore, the requested information regarding acceptance criteria and study details for an AI/ML device cannot be extracted from this document. The concepts of "test set," "training set," "experts to establish ground truth," "adjudication methods," "MRMC studies," and "standalone performance" are relevant to AI/ML device evaluations, but they do not apply to the traditional IVD assay described here.
The document discusses analytical and clinical studies for the immunoassay, focusing on performance characteristics such as:
- Clinical Study: Evaluates the concordance of FLC Kappa and FLC Lambda kit results with clinical assessment and other tests for monitoring Multiple Myeloma and AL Amyloidosis.
- Sample size:
- Multiple Myeloma: 551 follow-up samples from 235 unique subjects.
- AL-amyloidosis: 190 follow-up samples from 87 unique subjects.
- Data Provenance: Subjects had "expanded racial and ethnic diversity (Caucasian, African American, Hispanic, Asian)". No specific country of origin is mentioned, but the manufacturer is based in France and the submission is to the US FDA. The studies appear to be retrospective analyses of follow-up samples.
- Ground Truth: Clinical assessment and other tests based on criteria from the IMWG (International Myeloma Working Group) for Multiple Myeloma and consensus guidelines (Comenzo et al., 2012; Palladini et al., 2012; Kumar et al., 2022 NCCN guidelines) for AL-Amyloidosis.
- No mention of experts establishing ground truth or adjudication methods in the context of human readers for an AI system, as this is a laboratory assay.
- No MRMC study, as this is not an image-based or AI-assisted diagnostic.
- No separate "standalone" algorithm performance because the device itself is the diagnostic assay.
- No specific training set or how its ground truth was established is mentioned, as this is a traditional laboratory assay, not a machine learning model.
- Sample size:
- Stability Studies: Determined shelf-life for the kits and controls.
In summary, this document describes a traditional immunoassay, not an AI/ML device. Thus, the requested AI/ML-specific information is not applicable and not present.
Ask a specific question about this device
(265 days)
Name: Optilite Freelite Kappa Free Kit Optilite Freelite Lambda Free Kit Regulation Number: 21 CFR 866.5550
|
Regulatory information:
| Regulation section: | 21 CFR 866.5550
The Optilite Freelite Kappa Free Kit is intended for the quantitative in vitro measurement of Kappa free light chains in serum using the Binding Site Optilite analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE), and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS). Results of the free light chain measurements should always be interpreted in conjunction with other laboratory and clinical findings.
The Optilite Freelite Lambda Free Kit is intended for the quantitative in vitro measurement of Lambda in serum using the Binding Site Optilite analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE), and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS). Results of the free light chain measurements should always be interpreted in conjunction with other laboratory and clinical findings.
The determination of soluble antigen concentration by turbidimetric methods involves the reaction with specific antiserum to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Optilite Freelite Kappa Free Kit and Optilite Freelite Lambda Free Kit:
Context: This submission is for a modification to a previously cleared device (K150658) to extend its indications for use to "aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS)". The core device technology and its principles (immunoturbidimetry for quantitative measurement of free light chains) remain unchanged. Therefore, the performance data provided specifically addresses the new MGUS claim.
Acceptance Criteria and Reported Device Performance
Device: Optilite® Freelite® Kappa Free Kit and Optilite® Freelite® Lambda Free Kit
Intended Use Extension: Aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS).
| Acceptance Criteria (Pre-defined) | Reported Device Performance |
|---|---|
| Study 1: Clinical Performance - Sensitivity | |
| Sensitivity for all MGUS samples: at least 30% | 59.4% (95% CI: 53.0 - 65.5%) |
| Sensitivity for Light Chain MGUS (LC-MGUS): Not explicitly listed, but the study highlighted this subgroup. | 100% (95% CI: 75.8 - 100%) |
| Sensitivity for non-LC-MGUS samples: Not explicitly listed, but the study highlighted this subgroup. | 57.2% (95% CI: 50.6 - 63.5%) |
| Study 1: Clinical Performance - Specificity | |
| Specificity for non-MGUS disease controls: at least 85% | 86.4% (95% CI: 79.8 - 91.1%) |
| Study 2: Evaluation of MGUS Progression (Stable) | |
| Correct categorization of clinically stable MGUS by FLC testing: at least 80% | 93.3% of stable patients were categorized as "test positive" (meaning FLC stable based on criteria) |
| Study 2: Evaluation of MGUS Progression (Progressive) | |
| Correct categorization of clinically progressive MGUS by FLC testing: at least 30% | 50.0% of progressive patients were categorized as "test positive" (meaning FLC progressive based on criteria) |
Study Details:
1. Sample Sizes and Data Provenance:
- Study 1 (Sensitivity and Specificity):
- MGUS Samples (Sensitivity): 234 samples from patients with clinically confirmed MGUS.
- Disease Controls (Specificity): 140 samples from patients with polyclonal hypergammaglobulinemia (non-MGUS).
- Data Provenance: Retrospective testing of residual samples. The country of origin is not explicitly stated, but the company (The Binding Site Ltd.) is based in the United Kingdom.
- Study 2 (MGUS Progression):
- Total Samples: 185 samples from 49 MGUS patients (45 stable, 4 progressive). Up to 4 individual sample draws for stable, up to 6 for progressive per patient.
- Data Provenance: Retrospective testing of residual samples. Country of origin not explicitly stated.
2. Number of Experts and Qualifications for Ground Truth:
- The document states that the clinical diagnostic criteria that clinicians used to establish the "clinical truth" of MGUS positive samples were confirmed with each site.
- The ground truth for non-MGUS samples was based on "clinically confirmed polyclonal hypergammaglobulinemia" and "supporting clinical information."
- For the progression study, "clinically determined stable or progressive status" was the basis.
- The number of experts and their specific qualifications (e.g., radiologist with 10 years of experience) are not explicitly provided in the summary. It generally refers to "clinicians" and "clinical diagnosis."
3. Adjudication Method for the Test Set:
- Not explicitly mentioned. Given that the ground truth relies on "clinical diagnosis" and "clinically confirmed MGUS," it implies that the diagnostic criteria were applied by the treating clinicians at the respective sites prior to the retrospective study. There is no indication of an independent multi-expert adjudication process specifically for this study's test set.
4. MRMC Comparative Effectiveness Study:
- No. This submission is for an in-vitro diagnostic (IVD) device (a laboratory test), not an AI-assisted diagnostic tool that would directly assist human readers in image interpretation. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study to measure improvement in human readers with AI assistance is not applicable and was not performed.
5. Standalone Performance (Algorithm Only without Human-in-the-Loop):
- Yes. The study evaluates the performance of the device (Optilite Freelite Kappa and Lambda Free Kits on the Optilite Analyser) in measuring FLC levels and deriving the kappa:lambda ratio. The performance metrics (sensitivity, specificity, categorization of stable/progressive) are reported for the device's output itself, compared to the clinical ground truth. This is a standalone performance assessment of the IVD test.
6. Type of Ground Truth Used:
- Clinical Diagnosis / Clinical Findings:
- For MGUS positive samples: "clinically confirmed MGUS" based on diagnostic criteria, including those outlined by the 'International Myeloma Working Group (IMWG)' consensus.
- For non-MGUS samples: "polyclonal hypergammaglobulinemia confirmed by study testing (total IgG/lgA/lgM and serum IFE), with supporting clinical information."
- For stable/progressive MGUS: "clinically determined stable or progressive status" based on patient follow-up and conversion to MM or stable status over time. This includes evaluation criteria based on IMWG guidelines for multiple myeloma regarding FLC changes.
7. Sample Size for the Training Set:
- Not applicable. This submission focuses on the performance of a lab-based immunoassay kit, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The test method is immunoturbidimetry, which relies on chemical reactions and optical detection, not an algorithm that learns from data.
8. How Ground Truth for the Training Set Was Established:
- Not applicable, as there is no "training set" for this type of device. The method is a validated laboratory assay.
Ask a specific question about this device
(626 days)
Norcross, Georgia 30093
Re: K210623
Trade/Device Name: FLC Kappa, FLC Lambda Regulation Number: 21 CFR 866.5550
| 21CFR sec. 866.5550 - Immunoglobulin (light chainspecific) immunological
Freelite®Human Kappa Free Kit for use onthe Siemens BN™ II | K031016 | 866.5550
Freelite®Human Lambda Free Kit for useon the Siemens BN™ II | K031016 | 866.5550
The FLC Kappa kit is intended for the quantification of Kappa free light chains in human serum from adults with an Enzyme-Linked Immunosorbent Assay (ELISA) procedure. Measurement of free light chains aids in the diagnosis of multiple myeloma and AL amyloidosis. It must be used in conjunction with other laboratory and clinical findings. For In Vitro Diagnostic Use.
The FLC Lambda kit is intended for the quantification of Lambda free light chains in human serum from adults with an Enzyme-Linked Immunosorbent Assay (ELISA) procedure. Measurement of free light chains aids in the diagnosis of multiple myeloma and AL amyloidosis. It must be used in conjunction with other laboratory and clinical findings. For In Vitro Diagnostic Use.
The FLC Kappa and FLC Lambda test kits are intended for the quantification of free light chains in human serum from adults with an Enzyme-Linked Immunosorbent Assay (ELISA) procedure utilizing specific antibodies targeting anti-Lambda free light chains. It is carried out in 8 successive steps: Incubation of the previously diluted samples and calibrators, in the wells of the microplate, where specific free light chain antibodies are fixed. Washing of the wells to remove elements that have not been fixed by the anti-free light chain antiserum. Incubation with an anti- light chain antiserum (Kit specific) conjugated to peroxidase. Washing of the wells to remove the excess of antiserum conjugated to peroxidase. Incubation with peroxidase substrate. Stopping of the enzymatic reaction with an acidic solution. Reading of the optical density by absorbance spectrophotometry at 450 nm of the colored product. Calculation of the free light chain concentration of the sample using a calibration curve obtained with calibrators that have been analyzed on the same microplate.
Here's a breakdown of the acceptance criteria and study details for the Sebia FLC Kappa and FLC Lambda kits, based on the provided FDA 510(k) summary.
Note: This document describes an in vitro diagnostic (IVD) device, which measures specific analytes (free light chains) in human serum using a laboratory assay (ELISA). The concepts of "AI assistance," "human readers," "radiologist," and "image" are not relevant to this type of device. Therefore, sections pertaining to these concepts (like MRMC studies) are not applicable and will be marked as such. The "ground truth" for an IVD kit is established through various analytical performance studies demonstrating its accuracy and reliability compared to established methods, and clinical studies correlating its results with disease states.
Acceptance Criteria and Device Performance for Sebia FLC Kappa and FLC Lambda Kits
1. Table of Acceptance Criteria & Reported Device Performance
The FDA 510(k) summary does not explicitly present a "table of acceptance criteria" in the format of pass/fail thresholds. Instead, it details various performance studies and their results, which implicitly form the basis for acceptance. The reported device performance is indicated by the successful execution of these studies and the obtained values meeting recognized clinical laboratory standards (e.g., CLSI guidelines).
General Performance Categories and Key Findings:
| Performance Category | Acceptance Criteria (Implicit from CLSI Guidelines/Industry Norms) | Reported Device Performance (Key Findings) |
|---|---|---|
| Precision (Reproducibility) | Low Coefficient of Variation (CV%) across various conditions (within-run, between-runs, within-day, between-days, operators, lots, sites). Generally, CV% should be within acceptable limits for diagnostic assays. | FLC Kappa: Total %CVs for single-site, multi-operator, multi-lot, and multi-site reproducibility studies were generally below 15% across various sample concentrations, which is typical for acceptable immunoassay precision. FLC Lambda: Similar to Kappa, total %CVs across various reproducibility studies were generally below 15%, demonstrating acceptable precision. (e.g., Multi-site total reproducibility: 11.4% to 15.2%) |
| Linearity/Assay Range | Demonstrated linearity across the claimed analytical measuring range, with slopes near 1 and intercepts near 0 for linear regression. | FLC Kappa: Linear between 4.5 mg/L and 76.2 mg/L. Three panels showed linear regressions with slopes close to 1 (e.g., Y=1.009x -0.2967, Y=1.016x -0.2014, Y=0.9332x + 0.4952) and small Y-intercepts. FLC Lambda: Linear between 3.8 mg/L and 66.8 mg/L. Three panels showed linear regressions with slopes close to 1 (e.g., Y=1.046x -1.537, Y=0.9561x -0.62, Y=0.9611x -0.31) and small Y-intercepts. |
| Limits (LOB, LOD, LOQ) | Limits of Blank (LOB), Detection (LOD), and Quantitation (LOQ) must be established and clinically relevant. LOQ typically related to imprecision (e.g., <20% CV). | FLC Kappa: LOB = 0 mg/L, LOD = 0.8 mg/L, LOQ = 4.5 mg/L. FLC Lambda: LOB = 0 mg/L, LOD = 1.1 mg/L, LOQ = 3.8 mg/L. LOQ defined as lowest concentration with <20% within-laboratory imprecision. |
| Hook Effect | No significant Hook effect (antigen excess) within the expected clinical range. | No interference driven by the Hook effect was observed. |
| Analytical Specificity (Interferences) | No significant interference from common biological substances or drugs at clinically relevant concentrations. | No interference detected from conjugated/unconjugated bilirubin, total protein, hemoglobin, triglycerides, rheumatoid factor, or specified drugs (melphalan, dexamethasone, daratumumab, bortezomib, lenalidomide, pomalidomide, carfilzomib, isatuximab) at tested levels. |
| Stability | Kits and controls must maintain stated performance over their claimed shelf-life and in-use periods. | FLC Kappa & Lambda Kits: Shelf life: 9 months at 2-8°C. In-use: 5 uses within 1 month and 15 cumulative hours maximum at room temperature (19-25°C). All results met stability acceptance criteria. |
| Method Comparison (Quantitative) | Demonstrated agreement with a legally marketed predicate device (Siemens BN™ II Freelite® assays), often assessed by regression analysis (slope, intercept, R2). | FLC Kappa (vs. predicate): N=216, Sample Range 1.0-1947.0 mg/L. Slope: 0.557 (95% CI: 0.5125-0.6021), Y-Intercept: 0.912 (95% CI: 0.752-1.072), R2: 0.917. FLC Lambda (vs. predicate): N=221, Sample Range 1.6-860.4 mg/L. Slope: 0.608 (95% CI: 0.5224-0.6937), Y-Intercept: 2.243 (95% CI: 1.149-3.337), R2: 0.749. Note: Slopes deviate significantly from 1, indicating a proportional bias. This is common when comparing different assay methods and platforms, and is acceptable as long as the correlation is strong and clinical utility is demonstrated. |
| Method Comparison (Qualitative - Ratio) | Acceptable Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) with the predicate for the KFLC/LFLC ratio. | Kappa FLC / Lambda FLC Ratio (vs. predicate): PPA = 82.3%, NPA = 89.3%. (N=216 serum samples). |
| Clinical Performance (Sensitivity/Specificity) | Acceptable clinical sensitivity and specificity for indicated conditions (multiple myeloma, AL amyloidosis), demonstrated by correlation with clinical diagnosis. | Multiple Myeloma (N=366): Clinical sensitivity: 96.6% (95% CI: 94.0%-99.3%), Clinical specificity: 85.1% (95% CI: 79.4%-89.5%).AL Amyloidosis (N=333): Clinical sensitivity: 91.0% (95% CI: 86.3%-95.7%), Clinical specificity: 85.1% (95% CI: 79.4%-89.5%). |
| Reference Range | Established reference intervals for healthy populations. | US-population (N=238 healthy adults): Kappa FLC: 6.4 - 17.4 mg/L. Lambda FLC: 8.4 - 21.8 mg/L. Kappa FLC / Lambda FLC ratio: 0.46 - 1.51. |
2. Sample Sizes and Data Provenance
- Precision (Reproducibility):
- Single-site: 6 samples per kit (4 serum, 2 controls), 3 replicates/sample, 20 days, 2 runs/day = 120 results per sample.
- Multi-operator: 6 samples per kit (4 serum, 2 controls), 5 replicates/sample, 5 days, 3 operators = 75 results per sample.
- Multi-lot: 6 samples per kit (4 serum, 2 controls), 5 replicates/sample, 5 days, 3 lots = 75 results per sample.
- Multi-site: 6 samples per kit (4 serum, 2 controls), 5 replicates/sample, 5 days, 3 sites (3 operators).
- Linearity: 3 patient sample pools for each FLC assay (Kappa, Lambda). Number of individual samples/aliquots per pool not specified, but typically involves serial dilutions.
- LOD/LOQ: 5 low-concentration samples, tested in 2 runs of 4 replicates over 7 days using 2 reagent lots (56 replicates/sample/lot).
- Method Comparison (Quantitative): 216 serum samples (Kappa), 221 serum samples (Lambda).
- Method Comparison (Qualitative - Ratio): 216 serum samples.
- Reference Range: 238 apparently healthy adults (US-population).
- Clinical Study: 510 samples total (177 Multiple Myeloma, 144 AL Amyloidosis, 189 non-myeloma/non-amyloidosis subjects with various clinical conditions).
- Data Provenance: The document explicitly states the reference range study was conducted on a "US-population." The clinical study involved samples from various clinical conditions, implying real-world patient data, likely retrospective given the disease categories. The other analytical studies were performed in a controlled laboratory setting. The exact countries of origin for all samples beyond the US reference range study are not specified, but the manufacturing site is in France, suggesting international collaboration or testing. All studies appear to be prospective in their design for generating the performance data for the submission.
3. Number of Experts & Qualifications for Ground Truth
This is an IVD device, not an AI image analysis device. As such, there are no "experts" in the sense of radiologists providing interpretations for AI model ground truth. The "ground truth" for the performance studies is established by:
- Reference Methods: Comparison studies use a predicate device (The Binding Site Freelite® Human Kappa/Lambda Free kits on Siemens BN™ II) as a reference.
- Established CLSI Guidelines: All analytical studies (precision, linearity, limits, interference) follow rigorous Clinical and Laboratory Standards Institute (CLSI) guidelines (e.g., EP05-A3, EP06-ed2, EP17-A2, EP7-A2, EP09-A3, EP28-A3c). Compliance with these guidelines, developed by consensus among clinical laboratory experts, forms the basis for acceptable performance.
- Clinical Diagnosis: For the clinical study, the "ground truth" for multiple myeloma and AL amyloidosis was the clinical diagnosis of the patients. This diagnosis would be established by medical specialists (e.g., hematologists, oncologists) based on a comprehensive evaluation of laboratory findings, biopsy results, imaging, and clinical presentation. No specific number of such adjudicating clinicians is provided.
4. Adjudication Method for the Test Set
Not applicable in the context of an IVD device measuring analytes. The ground truth for analytical performance is adherence to validated laboratory procedures and statistical analysis per CLSI guidelines, and for clinical performance, it's the established clinical diagnosis for the patient samples.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is an in vitro diagnostic assay, not an AI-assisted diagnostic imaging device that involves human readers interpreting images with or without AI.
6. Standalone Performance
Yes, the entire performance data provided (precision, linearity, limits, interference, stability, quantitative and qualitative comparisons) represents the standalone performance of the Sebia FLC Kappa and FLC Lambda ELISA kits. The results are generated directly by the assay without human interpretation of complex outputs like images. The only "human-in-the-loop" aspect is the performance of the laboratory testing, which is standardized by the kit's instructions for use and quality control.
7. Type of Ground Truth Used
- Analytical Performance: Ground truth derived from established laboratory reference methods (e.g., predicate device for comparison), consensus guidelines (CLSI), and defined metrological properties (e.g., purified standards for linearity, known low-concentration samples for LOD/LOQ).
- Clinical Performance: "Ground truth" for the clinical study was the clinical diagnosis (Multiple Myeloma and AL Amyloidosis) of the patients from whom the serum samples were collected. This is based on comprehensive clinical and laboratory findings, not a single test.
8. Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device that requires a training set. This is a traditional IVD ELISA assay.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" for this type of IVD device.
Ask a specific question about this device
(468 days)
Kappa Free Light Chain Assay Diazyme Human Lambda Free Light Chain Assay Regulation Number: 21 CFR 866.5550
The Diazyme Human Kappa Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
The Diazyme Human Lambda Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple mycloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
Not Found
I'm sorry, but the provided text does not contain the information requested in your prompt regarding acceptance criteria and a study proving device performance for an AI/ML medical device.
The document is an FDA 510(k) clearance letter for Diazyme Laboratories Inc.'s Human Kappa Free Light Chain Assay and Human Lambda Free Light Chain Assay. These are in-vitro diagnostic assays (laboratory tests), not AI/ML-based devices. The document primarily discusses:
- The FDA's review and determination of substantial equivalence for these specific assays.
- General regulatory requirements for medical devices (e.g., registration, labeling, good manufacturing practices).
- Indications for use for each assay (quantitative determination of Kappa/Lambda FLC concentrations in serum to aid in diagnosis and monitoring of multiple myeloma).
Therefore, I cannot extract the following information because it is not present in the provided text:
- A table of acceptance criteria and reported device performance (for an AI/ML device): The document does not describe performance metrics like sensitivity, specificity, AUC, etc., in the context of an AI/ML model for image or other data interpretation.
- Sample size used for the test set and data provenance: No mention of test sets for an AI/ML model.
- Number of experts used to establish ground truth and qualifications: This is irrelevant for a chemical assay.
- Adjudication method: Not applicable.
- Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable as there is no AI assistance to human readers.
- Standalone (algorithm only) performance: Not applicable.
- Type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for chemical assays is typically established through reference methods and clinical correlations, not "expert consensus" in the way it's used for AI image analysis.
- Sample size for the training set: Not applicable for an IVD assay.
- How ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(212 days)
Kappa Free Light Chain Assay Diazyme Human Lambda Free Light Chain Assay Regulation Number: 21 CFR 866.5550
The Diazyme Human Kappa Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
The Diazyme Human Lambda Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple mycloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
Not Found
This document is an FDA 510(k) clearance letter for Diazyme Human Kappa Free Light Chain Assay and Diazyme Human Lambda Free Light Chain Assay. It describes the indications for use of the devices but does not contain any information regarding the acceptance criteria or a study proving the device meets the acceptance criteria.
Therefore, I cannot provide the requested information based on the provided text. To answer your questions, I would need a different document, such as a summary of safety and effectiveness data, a clinical study report, or a 510(k) summary that details the performance studies and acceptance criteria for these specific diagnostic assays.
Ask a specific question about this device
(212 days)
Germany
Re: K212379
Trade/Device Name: N Latex FLC kappa, N Latex FLC lambda Regulation Number: 21 CFR 866.5550
Classification Name: | Immunoglobulin (light chain specific)immunological test system per 21CFR866.5550
Immunoglobulin(light chain specific)immunological testsystem | Class II per21CFR866.5550
Immunoglobulin(light chain specific)immunological testsystem | Class II per21CFR866.5550
N Latex FLC kappa and lambda are in-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA-plasma. N Latex FLC kappa and lambda assays are used: • as an aid in the diagnosis and monitoring of multiple myeloma (MM) on the BN Systems and Atellica® CH Analyzer. • as an aid in the diagnosis of immunoglobulin light-chain amyloidosis (AL) on the BN Systems and Atellica® CH Analyzer. • as an aid in the monitoring of immunoglobulin light-chain amyloidosis (AL) on the BN Systems. · as an aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS) on the BN Systems and Atellica® CH Analyzer. Results of FLC measurements should always be interpreted in conjunction with other laboratory and clinical findings.
The N Latex FLC (free light chain) assays are in vitro diagnostic reagents for the quantitative determination of free light chains, type kappa or type lambda, in human serum and EDTA plasma by means of particle-enhanced immunoassay determination. Used in conjunction with other clinical and laboratory findings, FLC measurements are used as an aid in the diagnosis and monitoring of multiple myeloma (MM), as an aid in the diagnosis of amyloidosis (AL) on the BN Systems and Atellica® CH Analyzer; as an aid in the monitoring of amyloidosis (AL) on the BN Systems and as an aid in the evaluation of MGUS on the BN Systems. The FLC test systems on the Atellica® CH Analyzer are based upon the principles of particle-enhanced turbidimetry. Polystyrene particles coated with antibodies to human free light chains, type kappa or lambda, respectively, are agglutinated with samples containing FLC. Monitoring the agglutination by measuring the increase in turbidity, a concentration curve is obtained. The actual change in absorbance is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.
The provided text describes the acceptance criteria and performance study for the Siemens Healthcare Diagnostics Products GmbH N Latex FLC kappa and N Latex FLC lambda assays when applied to the Atellica® CH Analyzer, specifically for the added indication as an aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS).
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The device performance is assessed through a method comparison study between the N Latex FLC kappa and N Latex FLC lambda assays on the Atellica® CH Analyzer (candidate device) and the N Latex FLC kappa and N Latex FLC lambda assays on the BN ProSpec System (predicate device).
| Acceptance Criteria | N Latex FLC Kappa Performance | N Latex FLC Lambda Performance | Acceptance Criteria Met? (Overall) |
|---|---|---|---|
| Pearson correlation coefficient: r ≥ 0.95 | 0.983 (Overall) 0.977 (MGUS samples only) | 0.970 (Overall) 0.970 (MGUS samples only) | Yes |
| Slope: 0.9 - 1.1 | 0.995 (Overall) 0.966 (MGUS samples only) | 0.914 (Overall) 0.841 (MGUS samples only) | Yes (Overall), No (Lambda MGUS) |
| Predicted bias: ≤ +/- 10% | |||
| Lower limit reference interval | 0.93% (Overall) 4.33% (MGUS samples only) | -7.60% (Overall) -15.7% (MGUS samples only) | Yes (Overall), No (Lambda MGUS) |
| Upper limit reference interval | -0.10% (Overall) -1.18% (MGUS samples only) | -8.56% (Overall) -16.9% (MGUS samples only) | Yes (Overall), No (Lambda MGUS) |
| Number of samples: n ≥ 160 | 212 (Overall) 38 (MGUS samples only) | 202 (Overall) 35 (MGUS samples only) | Yes (Overall), No (MGUS samples only meets criteria for specific MGUS analysis) |
Note on MGUS Specific Results: While the overall study meets the acceptance criteria for slope and predicted bias, the separate analysis for MGUS patients only for N Latex FLC lambda shows predicted biases of -15.7% and -16.9% at the lower and upper limit reference intervals respectively, and a slope of 0.841. These specific MGUS-only results for lambda do not meet the stated acceptance criteria of ≤ +/- 10% bias and a slope of 0.9-1.1. However, the document's general conclusion states "Acceptance criteria fulfilled" based on the overall results (Table 7.3.1-2 and 7.3.1-3). The separate MGUS-only table (7.3.1-4 and 7.3.1-5) is presented as a sub-analysis, and the acceptance criteria for this sub-analysis are not explicitly re-stated or modified from the general criteria. The FDA’s acceptance of this submission implies that the overall performance was considered sufficient.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Overall):
- N Latex FLC kappa: 212 samples
- N Latex FLC lambda: 202 samples
- Sample Size (MGUS-specific analysis):
- N Latex FLC kappa: 38 MGUS samples
- N Latex FLC lambda: 35 MGUS samples
- Data Provenance: The method comparison study was conducted internally at the Siemens Healthcare Diagnostics Products GmbH site in Marburg, Germany. It incorporated "samples derived from MGUS patients." The text for the clinical performance study for MGUS evaluation mentions "121 MGUS samples (89 Non-IgM, 21 IqM and 11 LC MGUS) and 102 polyclonal immunostimulation samples," but it's important to differentiate these from the test set numbers used in the method comparison study (212/202 and 38/35). The specific type (retrospective/prospective) is not explicitly stated for the source of these samples, but the mention of "derived from MGUS patients" suggests pre-existing samples, which would typically be retrospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This information is not provided in the given text. The device is an in-vitro diagnostic reagent for quantitative determination of free light chains, and the "ground truth" for the method comparison study is established by comparison to a legally marketed predicate device (N Latex FLC assays on the BN ProSpec System).
4. Adjudication Method for the Test Set
This information is not applicable as the ground truth is established by a predicate diagnostic device, not by expert consensus or adjudication. The study is a method comparison, not a reader study.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic assay, not an imaging AI device that relies on human reader interpretation. The study evaluates the analytical performance of the assay compared to a predicate, not how human readers improve with AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, in essence. This is a standalone performance study of the diagnostic assay on a new instrument platform (Atellica® CH Analyzer) compared to its performance on a predicate instrument (BN ProSpec System). The "algorithm" here refers to the analytical process of the in-vitro diagnostic test.
7. The Type of Ground Truth Used
The ground truth for the method comparison study is the performance of the predicate device (N Latex FLC kappa and N Latex FLC lambda assays on the BN ProSpec System). This is a comparative study aiming to show equivalence to an already legally marketed and accepted diagnostic method.
8. The Sample Size for the Training Set
This information is not provided in the text. The document describes a study to support a 510(k) submission for a diagnostic assay, which typically involves validation studies rather than "training sets" in the context of machine learning (unless the underlying technology uses machine learning, which is not indicated here). If there was any internal development or calibration process that could be considered "training," the sample size used for that is not disclosed.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable/not provided as there is no mention of a "training set" in the context of machine learning. If the question refers to how the "predicate device" performance (which acts as the reference for this study) was originally established, that information is outside the scope of this particular 510(k) submission document. The predicate device itself has been cleared under previous 510(k)s (K171742, K182098, K193047, K201496).
Ask a specific question about this device
(511 days)
N Latex FLC lambda Regulation Number: 21 CFR 866.5550 Regulation Name: Immunoglobulin (Light Chain Specific
Classification Name: | Immunoglobulin (light chain specific) immunologicaltest system per 21CFR 866.5550
N Latex FLC kappa and lambda are in-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA plasma. N Latex FLC kappa and lambda assays are used:
• as an aid in the diagnosis and monitoring of multiple myeloma (MM) on the BN Systems and Atellica® CH Analyzer.
• as an aid in the diagnosis of immunoglobulin light-chain amyloidosis (AL) on the BN Systems and Atellica® CH Analyzer.
· as an aid in the monitoring of immunoglobulin light-chain amyloidosis (AL) on the BN Systems.
· as an aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS) on the BN Systems.
Results of FLC measurements should always be interpreted in conjunction with other laboratory and clinical findings.
N Latex FLC kappa and lambda are in-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTAplasma. N Latex FLC kappa and lambda assays are used as an aid in the diagnosis and monitoring of multiple myeloma (MM) and immunoglobulin light-chain amyloidosis (AL) and as an aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS). Monitoring of immunoglobulin light-chain amyloidosis (AL) and evaluation of MGUS are cleared for use only on the BN Systems.
The N Latex FLC test systems are based upon the principles of particle-enhanced immunonephelometry. Polystyrene particles coated with monoclonal antibodies to human free light chains, type kappa or lambda, respectively, are agglutinated when mixed with samples containing free light chains. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.
This document describes the N Latex FLC kappa and N Latex FLC lambda assays, in vitro diagnostic reagents used for the quantitative determination of free light chains (FLC), type kappa or type lambda, in human serum and EDTA plasma. This submission seeks to add "monitoring of immunoglobulin light-chain amyloidosis (AL) on the BN Systems" to the device's indications for use.
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly derived from the comparative studies presented, aiming to demonstrate substantial equivalence to the predicate device and acceptable concordance with clinical status. The reported performance is summarized in the tables below, using Evaluation Mode 2 as favored by Siemens.
Comparison of N Latex FLC versus Freelite (Predicate Device) - Evaluation Mode 2
| Response Criteria | Acceptance Criteria (Implicit) | Reported Performance (Agreement Rate) | 95% CI Bootstrap |
|---|---|---|---|
| Complete Response | High agreement | 68.1% | 53.6 – 86.3% |
| VGPR | High agreement | 81.8% | 67.4 – 91.1% |
| Partial Response | High agreement | 45.5% | 15.0 – 60.0% |
| Stable Disease | High agreement | 84.4% | 76.5 – 95.4% |
| Progressive Disease | High agreement | 88.2% | 70.0 – 100.0% |
| PPA (Progressive Disease) | High agreement | 88.2% (15/17) | 70.0 – 100.0% |
| NPA (Progressive Disease) | High agreement | 97.8% (219/224) | 95.9 - 99.6% |
Concordance of N Latex FLC versus Clinical Status - Evaluation Mode 2
| Response Criteria | Acceptance Criteria (Implicit) | Reported Performance (Concordance Rate) | 95% CI Bootstrap |
|---|---|---|---|
| Complete Response | High concordance | 52.9% | 25.0 – 74.3% |
| VGPR | High concordance | 71.4% | 56.6 – 88.9% |
| Partial Response | High concordance | 26.8% | 8.0 – 30.3% |
| Stable Disease | High concordance | 57.6% | 38.2 – 72.9% |
| Progressive Disease | High concordance | 70.6% | 47.1 – 88.2% |
| Sensitivity (Progressive Disease) | High sensitivity | 70.6% (12/17) | 47.1 – 88.2% |
| Specificity (Progressive Disease) | High specificity | 96.1% (197/205) | 94.8 - 99.0% |
2. Sample Size Used for the Test Set and Data Provenance
The test set for the comparative effectiveness study included data from a multi-center study on immunoglobulin light-chain amyloidosis (AL) patients.
- Sample Size: The N Latex FLC versus Freelite comparison involves a total of 241 observations based on initial and consecutive blood draws. The comparison to clinical status involves 222 observations.
- Data Provenance: The document does not explicitly state the country of origin. The study is retrospective, comparing changes between an initial sample draw and each consecutive blood draw, and then comparing these results to both a predicate device and clinical response.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The "clinical response" used as ground truth was "provided by the physician taking into account all available clinical and laboratory information."
- Number of Experts: Not explicitly stated, but it implies individual physicians for each patient.
- Qualifications: "Physician" is the qualification mentioned. No specific experience levels (e.g., years of experience or specialization in oncology/hematology) are provided for these physicians.
4. Adjudication Method for the Test Set
The document mentions that clinical response was "provided by the physician taking into account all available clinical and laboratory information." There is no mention of a formal adjudication panel or process (e.g., 2+1, 3+1). The response evaluation criteria are derived from NCCN Response Criteria for serum M protein/IFE and FLC levels/ratios.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC study was conducted. This device is an in-vitro diagnostic assay, not an imaging AI device that would typically involve human readers. The comparative effectiveness assessment focuses on the agreement between the new device and a predicate device, and the concordance of both devices with physician-determined clinical status.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, the studies presented are standalone performance assessments of the N Latex FLC kappa and lambda assays. The device itself is an automated assay, and its performance is evaluated directly (algorithm only). The comparison to clinical status is essentially an evaluation of the algorithm's output versus an established clinical truth.
7. Type of Ground Truth Used
The ground truth for the clinical comparison was based on clinical status/physician assessment, which encompasses "all available clinical and laboratory information" including NCCN Response Criteria. This is closer to an "outcomes data" or "expert consensus" type of ground truth in a clinical context.
8. Sample Size for the Training Set
The document does not provide information on the training set size. This submission focuses on performance data for an extended indication for monitoring AL. It refers to previous submissions (K171742, K182098, K193047) for established analytical and clinical studies, implying that the device was already developed and validated. The current submission's study used a distinct set of AL patients for evaluating the new monitoring claim.
9. How the Ground Truth for the Training Set Was Established
Not applicable/provided. As this submission is for an extended indication and refers to previous clearances for the device, details on the ground truth establishment for the initial training set (if any for algorithm development) are not included in this document. The focus of the provided text is on demonstrating performance for the new monitoring claim.
Ask a specific question about this device
(621 days)
De
Re: K193047
Trade/Device Name: N Latex FLC kappa, N Latex FLC lambda Regulation Number: 21 CFR 866.5550
system Classification Name: lmmunoglobulin specific) (light chain immunological test system per 21CFR 866.5550
N Latex FLC kappa and lambda are in-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA-plasma. N Latex FLC kappa and lambda assays are used:
• as an aid in the diagnosis and monitoring of multiple myeloma (MM) on the BN Systems and Atellica® CH Analyzer,
• as an aid in the diagnosis of amyloidosis (AL) on the BN Systems and Atellica® CH Analyzer,
• as an aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS) on the BN Systems.
Results of FLC measurements should always be interpreted in conjunction with other laboratory and clinical findings.
The N Latex FLC (free light chain) assays are in vitro diagnostic reagents for the quantitative determination of free light chains, type kappa or type lambda, in human serum and EDTA plasma by means of particle-enhanced immunoassay determination. Used in conjunction with other clinical and laboratory findings, FLC measurements are used as an aid in the diagnosis and monitoring of multiple myeloma (MM), as an aid in the diagnosis of amyloidosis (AL) and on the BN Systems, as an aid in the evaluation of MGUS.
Polystyrene particles coated with antibodies to human free light chains, type kappa or lambda, are agglutinated when mixed with samples containing FLC. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the concordance percentages achieved in various patient cohorts. The document does not explicitly state numerical thresholds as "acceptance criteria" but rather presents the performance results.
| Study Endpoint / Criteria | Reported Device Performance |
|---|---|
| MGUS Concordance (Overall): The ability of the device to correctly identify patients with MGUS based on FLC measurements. | 50.4% Concordance across 121 MGUS samples (89 Non-IgM, 21 IgM, 11 LC MGUS). |
| LC MGUS Concordance: The ability of the device to correctly identify patients with Light Chain MGUS (LC MGUS) based on an abnormal FLC-Ratio and elevated iFLC. | 90.9% Concordance across 11 LC MGUS samples when considering both criteria (abnormal FLC-Ratio and positive for iFLC elevation). 10 samples were positive for both criteria, and 1 for one criterion. |
| Polyclonal Immunostimulation Concordance: The ability of the device to correctly differentiate polyclonal immunostimulation from MGUS, yielding a low concordance for MGUS. | 90.2% Concordance across 102 polyclonal immunostimulation samples. (This implies a high agreement that these samples are not MGUS, aligning with the expected performance of a device aiding in MGUS evaluation by not falsely classifying non-MGUS as MGUS). |
| Agreement with Clinical Diagnosis for Stable MGUS: The device's ability to show stability in FLC measurements consistent with clinically stable MGUS. | 98.4% Agreement across 61 stable MGUS patients. Agreement was defined by: 1) κ/λ ratio within the reference interval of 0.53 - 1.51 at the last draw, AND 2) two consecutive assessments not showing a relative change of ≥ 25% for the involved free light chain (iFLC). |
| Agreement with Clinical Diagnosis for Progressive MGUS: The device's ability to show changes in FLC measurements consistent with progression from MGUS to MM. | 75.0% Agreement across 4 progressive MGUS patients. Agreement was defined by: 1) κ/λ ratio outside the reference interval of 0.53 - 1.51 at the time of clinical MM diagnosis, AND 2) two consecutive assessments showing a relative change of ≥ 25% for the involved light chain (iFLC). (Note: The small sample size here for progressive cases should be considered when interpreting this percentage). |
2. Sample Sizes Used for the Test Set and Data Provenance
- MGUS Concordance:
- Test Set Sample Size: 121 MGUS samples (89 Non-IgM, 21 IgM, 11 LC MGUS)
- Data Provenance: Not explicitly stated, but the study was described as a "multi-center study," suggesting samples were collected from various clinical sites. It is implied to be retrospective as they are described as "clinically defined samples" and patients with "diagnosed" MGUS or polyclonal immunostimulation.
- LC MGUS Concordance:
- Test Set Sample Size: 11 LC MGUS samples (subset of the 121 MGUS samples).
- Data Provenance: Same as above (multi-center, implied retrospective).
- Polyclonal Immunostimulation Concordance:
- Test Set Sample Size: 102 specimens.
- Data Provenance: Same as above (multi-center, implied retrospective).
- Evaluation of MGUS Patients (Stable Cohort):
- Test Set Sample Size: 61 patients.
- Data Provenance: Not explicitly stated, but patients were "initially diagnosed for MGUS" and "evaluated over different time periods" with "at least 4 sample draws at various time intervals," indicating a prospective or longitudinal study design over time.
- Evaluation of MGUS Patients (Progressive Cohort):
- Test Set Sample Size: 4 patients.
- Data Provenance: Same as above (prospective/longitudinal).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The document does not specify the number or qualifications of experts used to establish the ground truth for the test set. It refers to "clinically defined samples" and "clinical diagnosis," implying that the ground truth was based on a comprehensive clinical assessment, likely by medical professionals (e.g., oncologists, hematologists), but the details are not provided.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method for establishing the ground truth of the test set (e.g., 2+1, 3+1). The ground truth appears to be based on pre-existing "clinical diagnosis" or "clinically defined samples."
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic reagent, which provides quantitative measurements, not an imaging device that human readers interpret. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable in this context.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Yes, a standalone performance study was done. The studies described (concordance and evaluation of stable/progressive patients) assessed the performance of the N Latex FLC assays (the "device" or "algorithm" in this context) directly against clinical diagnoses and criteria. The results are reported as the "N Latex FLC κ/λ Ratio Concordance" and "N Latex FLC results," indicating an assessment of the device's output without direct human-in-the-loop interpretation of the assay results during the diagnostic process itself (though human interpretation of the FLC results in conjunction with other clinical findings is part of the intended use).
7. Type of Ground Truth Used
The ground truth used was expert consensus / clinical diagnosis / outcomes data.
- "Clinically defined samples" with diagnoses of MGUS or polyclonal immunostimulation.
- "Clinical diagnosis of MGUS" and "change in clinical diagnosis of MGUS to MM" for the stable and progressive cohorts, respectively. This implies that the ground truth was established by medical professionals using a combination of clinical findings, established diagnostic criteria, and other laboratory results, which could be considered a form of expert consensus or clinical outcomes.
8. Sample Size for the Training Set
The document does not report on a separate training set or its sample size. The performance data provided is for the evaluation of the device, implying the device was already developed. For an in-vitro diagnostic assay like this (which is a reagent test and analytical system), the "training" typically refers to the development and optimization of the assay's chemical and biological components, calibration, and establishment of analytical performance characteristics, rather than machine learning model training on a distinct labeled dataset. The document refers to prior submissions (K171742 and K182098) for "previously documented analytical and clinical studies," which likely covered the initial development and analytical validation.
9. How the Ground Truth for the Training Set Was Established
As no explicit "training set" for an AI/ML model is mentioned, this question is not directly applicable. For a traditional diagnostic assay, the "ground truth" during development involves rigorous analytical validation (e.g., using reference materials, spiked samples, and comparison to established methods) and initial clinical studies to define normal ranges and characteristic responses in different disease states. These would have been established through standard laboratory and clinical practices, likely involving certified reference materials, reference methods, and clinical expert assessment of patient samples.
Ask a specific question about this device
Page 1 of 4