    510k Summary Text (Full-text Search) :

    57107

    Re: K252627
    Trade/Device Name: Lyme-ID IgG Test; Bio-ID800
    Regulation Number: 21 CFR 866.3830
    Information
    Product Code(s): LSR, NSU
    Reagent Classification: Class II
    Regulation Section: 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Inanovate Lyme-ID IgG Test is an in vitro qualitative microarray assay for the detection of IgG antibodies to Borrelia burgdorferi in human serum. The assay is intended for testing serum samples from symptomatic patients or those suspected of Lyme Disease. It is intended to detect antibodies to VlsE and multiple other B.burgdorferi antigens following a modified two-tier test methodology. Positive results from the Lyme-ID IgG Test are supportive evidence for the presence of antibodies and exposure to B. burgdorferi, the causative agent for Lyme disease. Negative results do not preclude infection with B. burgdorferi. Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures as an aid in diagnosis of Lyme disease.

    The Inanovate Lyme-ID IgG Test must be used with Inanovate's Bio-ID800 instrument and Lyme-ID Software.

    Device Description

    Designed as a modified solid-phase ELISA, the Lyme-ID IgG Test is a protein microarray assay. A whole cell lysate sample of B. burgdorferi, along with antigens Outer Surface Protein C (OspC) and Variable Lipoprotein Surface-Exposed (VlsE) are bound to the glass surface of the Lyme-ID IgG Test cartridge. The antigens are immobilized as individual spots onto the glass surface. Positions of the spots are exactly defined and can be assigned to each antigen reliably. A negative control spot, positive control spot, and six signal control spots are applied to each microarray to ensure that the assay performs properly and to verify the reliability of individual sample results.

    The Bio-ID800 Instrument is a bench-top analyzer which contains an optics system for detecting and reading fluorescence produced during the assay, pressure system for flowing reagents through the microfluidics channels in the cartridge, and a manifold for clamping the cartridge and aligning the cartridge over the optics scan area. The optics system excites the fluorescent reagent for analyte detection using a 532nm wavelength laser projected through a series of mirrors and expanders to a scan lens located directly under the stage for the cartridge. This enables the assay area of the cartridge to be scanned for fluorescence through an aperture and read as it returns down an optics path to a photomultiplier tube (PMT). The PMT converts the collected fluorescent light into a corresponding output voltage that is read by a high-speed data acquisition card. Reagent and sample flow rates are pulsed using high-speed valves connected to externally sourced compressed air through a series of regulators and sensors. Assay reaction temperature is controlled using filtered fans and vents to maintain the chamber at room temp with a negative pressure at the cartridge loading door.

    The cartridge contains up to six patient samples and two controls (positive and negative), detection reagent, and blocking buffer wells all connected by microfluidic channels to an assay array field. As reagents or samples flow through the field, they are collected in waste wells, so that no waste leaves the cartridge. Each cartridge consists of a glass plate (containing the protein and cell lysate antigens, along with positive and negative control spots), a PDMS polymer assay reservoir which contains the microfluidic flow paths molded into the bottom, and a protective frame that aids in the alignment.

    The Lyme-ID Software controls the Bio-ID800 and communicates with the instrument to collect data into a database, enabling both storage of the assay data and analysis. The software enables self-tests of the instrument and communication through the user interface with the user and instrument for warnings, errors, and prompts. During assay performance, the control of the valves for pressurizing cartridges is controlled by the software. Signal generated through the PMT is interpreted using a statistical average function and stored by the software as an image file. The software then analyzes the image files to ensure the test was performed without error and analyzes the data collected from the analytes to compute the positive or negative status of the sample. Results are then printed from the computer with a standalone printer.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242872

    Validate with FDA (Live)

    Date Cleared
    2025-06-12

    (262 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K242872**
    Trade/Device Name: iDart Lyme IgM ImmunoBlot Kit
    Regulation Number: 21 CFR 866.3830
    Reagents, Borrelia Serological Reagent
    Classification: Class II
    Regulation Section: 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iDart™ Lyme IgM ImmunoBlot Kit is an immunoblot assay intended for the in vitro qualitative detection of IgM antibodies to Borrelia burgdorferi in human serum. The iDart™ Lyme IgM ImmunoBlot Kit is intended to detect antibodies to Lyme Screen Antigen (LSA) and multiple other B. burgdorferi antigens following a modified two-tier test methodology. Positive results from the iDart™ Lyme IgM ImmunoBlot Kit are supportive evidence for the presence of antibodies and exposure to B. burgdorferi. Negative results do not preclude infection with B. burgdorferi. iDart™ Lyme IgM ImmunoBlot Kit is intended to aid in the diagnosis of Lyme disease and the test kit should only be used on samples from patients with clinical history, signs and symptoms consistent with Lyme disease. The iDart™ Lyme IgM Immunoblot Kit is not intended as a screen for asymptomatic patients.

    Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures.

    Device Description

    The iDart™ Lyme IgM ImmunoBlot tests are line immunoblot assays. Antigenic proteins specific for Borrelia species that cause Lyme Disease are produced by recombinant DNA technology in Escherichia coli. The purified proteins are then applied as discrete lines on a nitrocellulose membrane along with two control proteins. The iDart™ Lyme IgM ImmunoBlot Kit contains IgM ImmunoBlot strips and the proteins are applied in the following order: C1 (IgG/IgM – conjugate control), C2 (Protein L – calibrator/serum control), P93, P41 (2 antigen bands), P39 (2 antigen bands), P23 (9 antigen bands), P31 (9 antigen bands), P34, C10 and LSA (a chimeric VlsE peptide termed the Lyme Screen Antigen).

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided FDA 510(k) clearance letter for the iDart™ Lyme IgM ImmunoBlot Kit.

    Note: The provided text details the performance characteristics but does not explicitly state "acceptance criteria" in a separate section with numerical thresholds. Therefore, the "Acceptance Criteria" column in the table below is inferred from the achieved "Reported Device Performance" and common expectations for such devices (e.g., high agreement, specificity, and sensitivity). The "study that proves the device meets the acceptance criteria" refers to the clinical and analytical performance studies conducted.


    1. Table of Acceptance Criteria and Reported Device Performance

    Criteria CategoryAcceptance Criteria (Inferred)Reported Device Performance
    ReproducibilityHigh agreement across sites, operators, and runs (e.g., ≥95% agreement)100% agreement on all test results across runs, days, and sites for all panel samples (negative, high negative, low positive, moderate positives, high positive). (N=90 replicates per sample across 3 sites, 2 operators, 5 days, 2 runs/day).
    Analytical Specificity (Endemic)High Agreement (e.g., ≥95%) in endemic healthy population99.44% specificity (1 false positive out of 177 samples) on samples from apparently healthy individuals in endemic areas (US, CDC, Bay Area Foundation NY, MA, WI).
    Analytical Specificity (Non-Endemic)High Agreement (e.g., ≥95%) in non-endemic healthy population99.21% specificity (1 false positive out of 127 samples) on samples from apparently healthy individuals in non-endemic areas (US, CDC, IGeneX, Inc.).
    Cross-ReactivityHigh specificity (e.g., ≥95%) in samples with potentially interfering conditions97.94% specificity (5 false positives out of 243 samples) in cross-reactivity study (various bacterial/viral infections, autoimmune diseases). Note: 2 Leptospira samples and 1 Mononucleosis, 1 H. pylori, 1 Parvovirus-19 sample were positive with iDart Lyme IgM. The Leptospira samples were also positive by STTT.
    Interference (Endogenous Analytes)No interference from common endogenous substances at specified concentrationsNo interference observed from Bilirubin (1, 15mg/dL), Albumin (3.5, 5g/dL), Cholesterol (150, 250mg/dL), Triglycerides (150, 500mg/dL), Hemoglobin (10, 20g/dL) in positive, low positive, and negative Borrelia IgM samples.
    Method Comparison (PPA)Acceptable Positive Percent Agreement (PPA) with STTT (e.g., lower bound of 95% CI > 80-85%)90.91% PPA (95% CI: 81.55%– 95.77%) against STTT in a cohort of 997 serum samples.
    Method Comparison (NPA)Acceptable Negative Percent Agreement (NPA) with STTT (e.g., lower bound of 95% CI > 95%)98.07% NPA (95% CI: 96.96%– 98.77%) against STTT in a cohort of 997 serum samples.
    Clinical Sensitivity (CDC Panel)Acceptable sensitivity across disease stages compared to STTT (e.g., comparable or better performance)Overall Sensitivity (CDC Panel): 65.8% compared to STTT's 55.7%. Stage I: 66.0% (STTT 56.0%), Stage II: 88.9% (STTT 77.8%), Stage III: 55.0% (STTT 45.0%). The iDart kit showed higher sensitivity than STTT across all stages in this particular panel.
    Clinical Specificity (CDC Healthy Controls)High specificity (e.g., ≥95%) in healthy controls for CDC panel97.9% agreement (3.1% false positive rate) for healthy controls within the CDC panel (N=95). Note that the table reports "Agreement," which implies specificity here.
    Clinical Specificity (CDC Disease Controls)High specificity (e.g., ≥95%) in disease controls for CDC panel98.9% agreement (1.1% false positive rate) for disease controls within the CDC panel (N=84). Note that the table reports "Agreement," which implies specificity here.
    Fresh vs. Frozen SamplesEquivalent performance between fresh and frozen samplesPerformance is equivalent between fresh and frozen samples. (N=63 fresh, N=63 frozen; same positive/negative counts).
    Antibody Class SpecificitySpecific detection of IgM antibodies (i.e., minimal cross-reactivity with IgG)The kit specifically detected IgM; treatment with human IgM blocked reactivity, while treatment with human IgG did not affect reactivity in positive samples. Negative samples remained negative under all conditions.

    Study Proving Device Meets Acceptance Criteria

    The studies detailed in the "PERFORMANCE CHARACTERISTICS" section of the 510(k) summary are designed to prove that the iDart™ Lyme IgM ImmunoBlot Kit meets the necessary performance standards for its intended use. These include:

    • Reproducibility Study: Tested consistency across multiple sites, operators, and runs.
    • Analytical Specificity Studies: Evaluated performance in healthy endemic and non-endemic populations, and cross-reactivity with various non-Borrelia pathogens and autoantibodies.
    • Interference Study: Assessed impact of common endogenous substances.
    • Method Comparison with STTT: Compared the device's accuracy (PPA and NPA) against the Standard Two-Tier Test methodology, which is a recognized approach for Lyme disease diagnosis.
    • CDC Serum Panel Study: Evaluated sensitivity and specificity across different disease stages and in various control groups using a well-characterized reference panel.
    • Fresh and Frozen Samples Comparison Study: Demonstrated stability of samples under different storage conditions.
    • Antibody Class Specificity Study: Confirmed the specific detection of IgM antibodies.

    Additional Information:

    2. Sample size used for the test set and the data provenance:

    • Reproducibility: A panel of coded samples with different levels of anti-B. burgdorferi IgM (negative, high negative, low positive, moderate positives, and high positive samples). 90 replicates per sample were generated. The specific number of distinct samples in this panel is not explicitly stated beyond "a panel of coded samples".
    • Analytical Specificity (Endemic): 177 samples from apparently healthy individuals in the US from endemic areas (CDC and Bay Area Foundation NY, MA, WI).
    • Analytical Specificity (Non-Endemic): 127 samples from apparently healthy individuals in the US from non-endemic areas (CDC and IGeneX, Inc.).
    • Cross-Reactivity Study: 243 serum samples from patients with bacterial or viral infections, as well as sera from patients with diagnoses that could be confused with Lyme disease (CDC, IGeneX, Inc., New York Biologics (NY), Kamineni Life Sciences Pvt. Ltd, Hydrabad (India), Warde Medical Laboratory (MI)).
    • Interference Study: One positive, one low positive, and one negative Borrelia IgM samples (tested in singlicate with various spiked interfering agents).
    • Method Comparison with STTT: 997 prospectively collected serum samples procured from IGeneX, Inc.
    • CDC Serum Panel: 258 serum samples from the CDC panel. These included patients with Lyme disease at different stages and Lyme disease look-alike infections, and healthy controls.
    • Fresh and Frozen Samples Comparison Study: 63 fresh samples and 63 frozen samples.
    • Antibody Class Specificity: 8 previously tested patient samples (4 negatives, 4 positives).

    Data Provenance: The data appears to be a mix of retrospective and prospective. Samples from "apparently healthy individuals" and "patients with bacterial or viral infections" can be either. The "Method Comparison with STTT" explicitly states "Prospectively collected serum samples" from IGeneX, Inc. The CDC Serum Panel is a reference panel, usually well-characterized retrospectively. Locations include the US (CDC, Bay Area Foundation NY, MA, WI, IGeneX, Inc., New York Biologics (NY), Warde Medical Laboratory (MI)) and India (Kamineni Life Sciences Pvt. Ltd, Hydrabad).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    The document does not explicitly state the number of experts or their qualifications for establishing ground truth for the clinical study samples. For the "Method Comparison with comparators (STTT)", the comparison is against "FDA-cleared EIA and immunoblot as part of the standard two-tier test methodology (STTT)". This implies the ground truth for these samples was established by the results of the STTT, which is a standardized and accepted diagnostic algorithm, rather than by individual expert interpretation. For the CDC Panel, the samples are "from patients diagnosed with Lyme Disease at different stages" or "Lyme disease look-like infections", suggesting that the CDC provided established diagnoses based on their protocols, which would implicitly involve expert clinical and laboratory evaluation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    The document does not describe an explicit adjudication method (like 2+1 or 3+1) for establishing ground truth for the clinical test sets. The ground truth for the method comparison study was the result of the "FDA-cleared EIA and immunoblot as part of the standard two-tier test methodology (STTT)". For the CDC panel, it was based on the provided "diagnosis".

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic (IVD) kit for laboratory use (an ImmunoBlot assay, specifically evaluating IgM antibodies) rather than an AI-powered image analysis system or a device requiring human-in-the-loop interpretation that would typically necessitate such a study. The "reading" of the immunoblot strips is performed visually but the interpretation criteria are clearly defined, not requiring subjective interpretive discretion usually addressed by MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    This question is not directly applicable in the terms typically used for AI/software devices. The iDart™ Lyme IgM ImmunoBlot Kit is a lab-based immunoassay. Its performance characteristics (sensitivity, specificity, reproducibility) were evaluated as a standalone device (without external assistance like AI or human readers for interpretation, other than the standard visual reading of the bands per the kit's instructions), as demonstrated by the analytical and clinical studies. The results are generated by the chemical reaction on the strip and interpreted based on predefined criteria. It's an "algorithm only" in the sense of the assay's biochemical process and visual interpretation rules, operating without additional "human-in-the-loop" modifications to its fundamental mechanism or output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    The ground truth used primarily appears to be:

    • Standard Two-Tier Test (STTT) results: For the majority of the clinical method comparison study (N=997), the device was compared against the results from "FDA-cleared EIA and immunoblot as part of the standard two-tier test methodology".
    • Clinical Diagnosis from CDC: For the CDC reference panel, samples were from patients "diagnosed with Lyme Disease at different stages" or "Lyme disease look-like infections". This implies a ground truth established through a combination of clinical evaluation, laboratory tests, and expert judgment according to CDC protocols.

    8. The sample size for the training set:
    Not applicable/Not specified. This is an in vitro diagnostic kit (ImmunoBlot assay), not a machine learning or AI-based device that typically requires a distinct "training set" for model development. The antigens used are recombinant proteins, and the assay's interpretation criteria are rule-based, not learned from a dataset.

    9. How the ground truth for the training set was established:
    Not applicable. As stated above, this device does not utilize a training set in the context of machine learning. The design and validation of immunoblots are based on known antigen-antibody interactions and assay development methodologies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241427

    Validate with FDA (Live)

    Device Name
    Access Syphilis
    Date Cleared
    2024-09-06

    (109 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Chaska, Minnesota 55318

    Re: K241427

    Trade/Device Name: Access Syphilis Regulation Number: 21 CFR 866.3830
    Syphilis Common Name: Treponema pallidum treponemal test reagents Classification Requlation: 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Syphilis assay is a paramagnetic particle, chemiluminescent immunoassay for the qualitative detection of total antibodies to Treponema pallidum in human serum and plasma using the Access lmmunoassay Systems. It is intended to be used as an aid in the diagnosis of syphilis or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection. The Access Syphilis assay is not intended for blood and tissue donor screening.

    Device Description

    The Access Syphilis assay is a two-step enzyme immunoassay. A sample is added to a reaction vessel with buffer, paramagnetic particles coated with recombinant Treponema pallidum antigens Tp17 and Tp47, and Tp47, and biotinylated Treponema Tp17 & Tp47 antigens. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Alkaline phosphatase conjugates are added, and the conjugates bind to the immunoglobulin captured on the particles. A chemilyminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is proportional to the amount of Treponema pallidum antibodies in the sample. The light quantity measured for a sample allows a determination of the presence of the analyte by comparison with a cut-off value defined during the assay calibration on the instrument. The Access Syphilis reagents are provided in liquid ready-to-use format designed for optimal performance on the Beckman Coulter Access Immunoassay Systems. Each reagent kit contains two reagent packs.

    AI/ML Overview

    The Access Syphilis assay is a qualitative immunoassay for detecting total antibodies to Treponema pallidum in human serum and plasma, used as an aid in diagnosing syphilis. The device was evaluated for its clinical performance on two systems: the Access 2 Immunoassay System and the Dxl 9000 Access Immunoassay Analyzer. The acceptance criteria and performance data are primarily based on percent agreement with a composite reference method.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state pre-defined acceptance criteria values for Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). However, the results presented are implicitly the "performance" that aims to demonstrate substantial equivalence. The following table summarizes the reported performance in key clinical evaluation cohorts:

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Access 2 & Dxl 9000)
    PPA (Overall Intended Use Population)High agreement expected, generally >95%100% (184/184) [95% CI: 98.0 - 100%]
    NPA (Overall Intended Use Population)High agreement expected, generally >95%96.7% (890/920) [95% CI: 95.4 - 97.7%]
    PPA (Retrospective Specimens)High agreement expected, generally >95%100% (398/398) [95% CI: 99.0 - 100%]
    NPA (Retrospective Specimens)High agreement expected, generally >95%25.0% (1/4) [95% CI: 4.6 - 69.9%] *
    PPA (High-Risk Individuals)High agreement expected, generally >95%100% (20/20) [95% CI: 83.9 - 100%]
    NPA (High-Risk Individuals)High agreement expected, generally >95%80.0% (24/30) [95% CI: 62.7 - 90.5%]

    *Note on NPA for Retrospective Specimens: The low NPA for retrospective specimens is due to a very small number of non-reactive specimens in the comparator (only 4), making the percentage highly sensitive to a single misclassification. The 31 reactive results from Access Syphilis where the comparator was nonreactive need further investigation, which the document attributes to "3 specimens were reactive by treponemal immunoassay and nonreactive by RPR and TPPA". This suggests potential discordance with the composite comparator definition for these few cases rather than a broad failing of the device's negative detection.

    2. Sample Size for the Test Set and Data Provenance:

    The study involved a total of 1,910 specimens for clinical performance evaluation. These specimens were broadly characterized as:

    • 1,104 prospectively collected specimens from the intended use population. The age range was 12 to >89 years, with 63.8% female and 36.2% male. This cohort included:
      • 399 patients sent for syphilis testing
      • 405 pregnant women
      • 300 HIV positive patients
    • 402 retrospective specimens from patients (including 22 from pregnant females).
    • 204 prospectively collected specimens from apparently healthy individuals.
    • 150 retrospective specimens from patients with medically diagnosed syphilis (primary, secondary, and latent stages).
    • 50 retrospective specimens from individuals at high-risk of sexually transmitted disease.

    The provenance of the data is multicenter, meaning specimens were collected from multiple clinical sites. Both retrospective and prospective data were used. The document does not explicitly state the country of origin, but the submission to the U.S. FDA suggests a U.S. or international clinical setting adhering to FDA standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    The document does not mention the use of human experts to establish the ground truth for the test set in the traditional sense (e.g., radiologists interpreting images). Instead, the "final comparator result" (ground truth) was established using a composite testing algorithm of multiple FDA-approved laboratory assays, which is standard practice for in vitro diagnostic devices.

    4. Adjudication Method for the Test Set:

    The adjudication method used a composite testing algorithm as the "final comparator result." This algorithm consisted of:

    • An FDA-approved predicate treponemal immunoassay.
    • A non-treponemal assay (RPR - Rapid Plasma Reagin).
    • A second treponemal assay (TPPA - Treponema Pallidum Particle Agglutination).

    The document does not detail a specific "2+1" or "3+1" adjudication process involving human review for discordant results beyond the internal algorithmic comparison. However, for discordant results in the HIV positive patient subpopulation (where the Access Syphilis assay showed lower NPA), an additional FDA-cleared electrochemiluminescent immunoassay was used to further evaluate 28 discordant specimens.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic assay, not an imaging AI device that involves human readers interpreting images with or without AI assistance. The performance is assessed by comparing the device's output to established laboratory reference methods, not human interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the clinical performance study is a standalone performance evaluation. The Access Syphilis assay is an automated immunoassay system that provides results independently. Its performance (reactive/non-reactive) is directly compared against the composite reference standard without human interpretation of the assay's raw output.

    7. The type of ground truth used:

    The ground truth was established using a composite testing algorithm consisting of results from:

    • An FDA-approved predicate treponemal immunoassay.
    • A non-treponemal assay (RPR).
    • A second treponemal assay (TPPA).

    This acts as a "reference standard" or "expert consensus" in the context of laboratory diagnostics, where consensus among multiple established tests determines the final disease status. In some cases, "medically diagnosed syphilis" for specific retrospective cohorts also contributed to the understanding of the samples.

    8. The sample size for the training set:

    The document does not specify a separate training set size for the Access Syphilis assay. As a chemiluminescent immunoassay, it is a biochemical test, not an AI/machine learning algorithm that typically requires a discrete "training set" in the same way. The assay's parameters would have been optimized during its development phase using internal studies, but these are not referred to as a "training set" in the context of typical AI device submissions.

    9. How the ground truth for the training set was established:

    Since a distinct "training set" in the AI/ML sense is not described, the method for establishing its ground truth is not applicable in the document. The development of an immunoassay involves extensive analytical and clinical validation, which ensures the reagents and system accurately detect the target antibodies. This process implicitly refines the assay's performance against known positive and negative samples, similar to how a training set might function for an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241534

    Validate with FDA (Live)

    Date Cleared
    2024-08-28

    (90 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: VITROS Immunodiagnostic Products Syphilis Reagent Pack Regulation Number: 21 CFR 866.3830
    ---------------------------------------------------|--------------|
    | LIP | II | 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the qualitative determination of total (IgG and IgM) antibodies to Treponema pallidum (TP) specific antigens in human serum and plasma using the VITROS 5600 Integrated System.

    The presence of antibodies to Treponema pallidum (TP) specific antigens, in conjunction with non-treponemal laboratory tests and clinical findings may aid in the diagnosis of syphilis infection.

    The VITROS Syphilis test is not intended for blood and tissue donor screening.

    Device Description

    The VITROS Immunodiagnostic Products Syphilis test is performed using the VITROS Immunodiagnostic Products Syphilis Reagent Pack and VITROS Immunodiagnostic Products Syphilis Calibrator on the VITROS 5600 Integrated System.

    An immunometric technique is used; this involves a two-stage reaction. In the first stage antibodies to Syphilis TP specific antigens present in the sample bind with biotinylated recombinant Syphilis TP antigens immobilized on streptavidin coated wells. Unbound sample is removed by washing. In the second stage conjugate reagent containing horseradish peroxidase (HRP)-labeled recombinant Syphilis TP antigens is added. The conjugate binds specifically to any antibody to Syphilis TP specific antigens captured on the well in the first stage. Unbound conjugate is removed by washing.

    The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent (a substituted acetanilide) increases the level of light produced and prolongs its emission. The light signals are read by the system.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study proving the device meets these criteria, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the clinical performance results meeting high percentage agreements. While explicit numeric acceptance criteria (e.g., "PPA must be >X%") are not directly stated in a dedicated section, the sponsor presents the results and concludes substantial equivalence, indicating that these results were considered acceptable.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    All Prospective Subgroups
    Positive Percent Agreement (PPA)High agreement with comparator99.5% (218/219)
    Negative Percent Agreement (NPA)High agreement with comparator97.6% (688/705)
    Routine Syphilis (Prospective)
    PPAHigh agreement with comparator98.8% (83/84)
    NPAHigh agreement with comparator99.2% (527/531)
    Pregnant Women (Prospective)
    PPAHigh agreement with comparator100.0% (1/1)
    NPAHigh agreement with comparator97.8% (45/46)
    HIV Positive (Prospective)
    PPAHigh agreement with comparator100.0% (134/134)
    NPAHigh agreement with comparator90.6% (116/128)
    All Retrospective Specimens
    PPAHigh agreement with comparator100.0% (213/213)
    NPAHigh agreement with comparator98.8% (330/334)
    Pregnant Women (Retrospective)
    PPAHigh agreement with comparator100.0% (31/31)
    NPAHigh agreement with comparator100.0% (212/212)
    HIV Positive (Retrospective)
    PPAHigh agreement with comparator100.0% (30/30)
    NPAHigh agreement with comparator96.7% (118/122)
    Pre-selected Positive (Retrospective)
    PPAHigh agreement with comparator100.0% (152/152)
    NPAN/A (denominator is zero)N/A
    Medically Diagnosed (Syphilis) ReactivityConsistent detection100% reactive (151/151) across all stages and treatment statuses
    Apparently Healthy Individuals ReactivityLow false positive rate2.0% reactive (4/201), with 3 of these confirmed positive by final comparator

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size:
      • Prospective Specimens: 924 total specimens.
        • 615 subjects for routine syphilis testing (from 6 sites in the United States).
        • 47 pregnant women.
        • 262 HIV positive subjects.
      • Retrospective Specimens: 547 total specimens.
        • 243 samples from pregnant women.
        • 152 HIV positive samples.
        • 152 pre-selected positive samples.
      • Medically Diagnosed Individuals: 151 samples.
      • Apparently Healthy Individuals: 201 prospective specimens (from 3 sites in the United States).
      • Total Clinical Samples: 924 (prospective) + 547 (retrospective) + 201 (apparently healthy) = 1672 specimens for primary clinical evaluation. The medically diagnosed group is a subset likely included within the prospective/retrospective groups or as additional targeted samples.
    • Data Provenance:
      • Country of Origin: United States (for prospective and apparently healthy specimens). Retrospective samples are "purchased," implying they may originate from various sources but were tested at sites.
      • Retrospective or Prospective: Both retrospective and prospective clinical studies were conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts or their qualifications for establishing the ground truth. Instead, the ground truth was established using a "composite testing algorithm."

    4. Adjudication method for the test set

    The adjudication method relied on a composite testing algorithm using:

    • A treponemal electrochemiluminescence immunoassay (TP-ECLIA)
    • A Rapid Plasma Reagin (RPR) non-treponemal assay
    • A Treponema pallidum Particle Agglutination (TP-PA) Treponema-specific assay

    Cases were categorized as "Positive for Syphilis" or "Negative for Syphilis" based on the results of these FDA-cleared comparator assays. The document describes specific combinations of results from these three assays that led to a final comparator result. For example, "Non-reactive TP-ECLIA + Non-reactive RPR" was deemed "Negative," while "Reactive TP-ECLIA + Reactive RPR (and N/A TP-PA)" was "Positive." Discordant results with the final comparator were further analyzed (e.g., 14 discordant prospective samples were found reactive by another FDA-cleared TP antibody test).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving human readers and AI assistance was not conducted or reported. This device is an in vitro diagnostic test (IVD) for laboratory use, not an AI-powered image analysis or diagnostic aid for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the studies presented are standalone performance evaluations of the VITROS Immunodiagnostic Products Syphilis test (an algorithm-driven instrument/reagent system) without human-in-the-loop performance being a variable in the reported clinical accuracy. The assay itself performs the determination.

    7. The type of ground truth used

    The ground truth used was a composite testing algorithm based on multiple FDA-cleared comparator assays (TP-ECLIA, RPR, and TP-PA). This is a form of clinical surrogate ground truth derived from established diagnostic tests, rather than direct pathology, biopsy, or long-term outcomes data, which are typically used for definitive diagnosis in some other medical fields.

    8. The sample size for the training set

    The document does not provide information on the sample size for a training set. This is because the device described is an immunoassay, a biochemical test, not a machine learning or AI algorithm that typically requires a separate training set. The reported studies evaluate the locked-down analytical and clinical performance of the manufactured diagnostic kit.

    9. How the ground truth for the training set was established

    Since there is no training set mentioned for an AI/machine learning algorithm, the concept of establishing ground truth for a training set is not applicable here. The assay's internal calibration and optimization would have been performed during product development, but this is distinct from "training" an AI model in the context of imaging or predictive analytics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233367

    Validate with FDA (Live)

    Date Cleared
    2024-08-12

    (315 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    August 12, 2024

    Re: K233367

    Trade/Device Name: iDart Lyme IgG ImmunoBlot Kit Regulation Number: 21 CFR 866.3830
    ): LSR - Reagents, Borrelia Serological Reagent Classification: Class II Regulation Section: 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iDart™ Lyme IgG ImmunoBlot Kit is an immunoblot assay intended for the in vitro qualitative detection of IgG antibodies to Borrelia burgdorferi in human serum. The iDart Lyme IgG ImmunoBlot Kit is intended to detect antibodies to LSA and multiple other B. burgdorferi antigens following a modified two-tier test methodology. Positive results from the iDart Lyme IgG ImmunoBlot Kit are supportive evidence for the presence of antibodies and exposure to B. burgdorferi. Negative results do not preclude infection with B. burgdorferi. iDart™ Lyme IgG ImmunoBlot Kit is intended to aid in the diagnosis of Lyme disease and the test kit should only be used on samples from patients with clinical history, signs and symptoms consistent with Lyme disease. The iDart Lyme IgG Immunoblot Kit is not intended as a screen for asymptomatic patients.

    Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures.

    For in vitro diagnostic use only
    For professional use only
    For prescription use only

    Device Description

    The iDart™ Lyme IgG ImmunoBlot tests are line immunoblot assays. Antigenic proteins specific for Borrelia species that cause Lyme Disease are produced by recombinant DNA technology in Escherichia coli. The purified proteins are then applied as discrete lines on a nitrocellulose membrane along with two control proteins.

    The iDart™ Lyme IgG ImmunoBlot Kit contains IgG ImmunoBlot strips and the proteins are applied in the following order: C1 (lgG/lgM - conjugate control), C2 (Protein L - calibrator/serum control), P93, P41, P39, P23, P31, P66, P58, P45, P34, P30, P28, P18 and LSA (a chimeric VISE peptide termed the Lyme Screen Antigen).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the iDart™ Lyme IgG ImmunoBlot Kit are primarily demonstrated through its analytical performance (reproducibility, analytical specificity, cross-reactivity, interference) and clinical performance (method comparison with STTT, clinical sensitivity/specificity using a CDC panel).

    Acceptance Criteria CategorySpecific Metric / EvaluationAcceptance Threshold (Implied/Explicit)Reported Device Performance (as stated)
    Analytical Performance
    ReproducibilityAgreement across sites, operators, runs, days100% agreement expected100% agreement of all bands among all runs, all days and across 3 sites for negative, moderate negative, high negative, moderate positive, and high positive samples (Table 1).
    Analytical Specificity (Endemic)Specificity in healthy individuals from endemic areasHigh specificity99.36% (2 false positives out of 313 samples from CDC and Bay Area Lyme Foundation) (Table 2).
    Analytical Specificity (Non-Endemic)Specificity in healthy individuals from non-endemic areasHigh specificity100% (0 false positives out of 112 samples from CDC and CA) (Table 3).
    Cross-ReactivityFalse positivity with various conditionsLow/no cross-reactivity100% specificity for LSA, 98.67% specificity for ≥2 bands, and 100% specificity for IgG Positive across 376 potentially cross-reactive samples from various disease states and infections (Table 4). Minor false positives (5 for ≥2 bands) were noted but resulted in 0% IgG positive or only a single band out of the two required for positivity.
    InterferenceEffect of endogenous analytesNo interferenceNo interference observed for bilirubin, albumin, cholesterol, triglycerides, and hemoglobin at specified low and high concentrations on positive, low positive, and negative Borrelia IgG samples (Table 5).
    Clinical Performance
    Method Comparison (STTT)Positive Percent Agreement (PPA) with STTTHigh PPA and NPABay Area Lyme Foundation (n=290): PPA: 95.00% (95% CI: 76.39% – 99.11%), NPA: 86.67% (95% CI: 82.09% – 90.21%) (Table 7) IGeneX Inc. Cohort 2 (n=248): PPA: 95.00% (95% CI: 89.52% – 97.69%), NPA: 90.63% (95% CI: 84.33% - 94.56%) (Table 8) IGeneX Inc. Cohort 3 (n=230): PPA: 90.91% (95% CI: 62.27% – 98.38%), NPA: 96.80% (95% CI: 93.55% – 98.44%) (Table 9)
    Clinical SensitivityPerformance against CDC Reference PanelHigh sensitivity for later stagesStage I: 58.33% (higher than STTT at 30.00%) Stage II: 90.00% (equal to STTT) Stage III: 100% (equal to STTT) Overall: 71.11% (higher than STTT at 52.22%) (Table 10).
    Clinical SpecificityPerformance against CDC Reference PanelHigh specificityHealthy controls: 100% (equal to STTT) Disease Controls: 100% (equal to STTT) (Table 10).
    Fresh and Frozen Sample ComparabilityConsistent results between fresh and frozen samplesConsistent resultsAll IgG positive samples remained positive and all negative samples remained negative after freezing (Table 11).
    Antibody Class SpecificitySpecificity of anti-human IgG conjugateSpecific to IgGAll positive samples tested without treatment or with human IgM remained positive, and all negative samples remained negative. When treated with human IgG, all positive samples became negative, confirming specificity (Table 12).

    2. Sample Size Used for the Test Set and Data Provenance

    • Reproducibility: 90 samples for each of the 6 sample types (High Positive, Moderate Positive, Negative-1, Negative-2, Negative-3, Low Positive). Total of 540 tests performed across 3 sites, 2 operators, 5 days, 2 runs/day. Data provenance is not explicitly stated beyond "blinded and coded samples."
    • Analytical Specificity (Healthy Individuals):
      • 313 samples from endemic areas (CDC, Bay Area Lyme Foundation - NY, MA, WI).
      • 112 samples from non-endemic areas (CDC, CA).
    • Cross-Reactivity Study: 376 samples from various disease states/infections (CDC, IGeneX (CA), New York Biological (NY), BEI, Kamineni Life Sciences Pvt. Ltd, Hydrabad (India), Warde Medical Laboratory (MI)).
    • Interference Study: One positive, one low positive, and one negative Borrelia IgG sample were used for each interference agent and concentration, tested in singlicate.
    • Clinical Studies (Method Comparison with STTT): A total of 768 serum samples.
      • Site 1: 290 clinical serum samples from Bay Area Lyme Foundation.
      • Site 2: 37 clinical serum samples (Cohort 2) + 230 clinical serum samples (Cohort 3) from IGeneX Inc.
      • Site 3: 211 clinical serum samples (Cohort 2) from IGeneX Inc.
      • Data provenance: "procured from two vendors" (Bay Area Lyme Foundation and IGeneX Inc.). Samples were "prospective banked samples" or "clinical serum samples."
    • Clinical Sensitivity/Specificity (CDC Serum Panel): 280 serum samples from CDC (patients with Lyme disease at different stages, look-alike conditions, healthy controls from endemic and non-endemic regions).
    • Fresh and Frozen Samples Comparison Study: 72 decoded left-over patient serum samples.
    • Antibody Class Specificity: 10 previously tested patient samples (6 negatives, 4 positives).

    All clinical samples were "blinded, re-coded."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state that experts were used to establish ground truth for the test set derived from clinical samples. Instead, the ground truth for these clinical performance studies appears to be based on:

    • STTT (Standard Two-Tier Test Methodology): This is a laboratory-based diagnostic algorithm involving an EIA/IFA screen followed by an immunoblot. Its results serve as the comparator (ground truth) for the method comparison study.
    • CDC Reference Panel: For the clinical sensitivity/specificity, the CDC reference panel implicitly has established diagnoses for Lyme disease stages and other conditions. The process by which CDC established these diagnoses (e.g., expert consensus, other gold standards) is not detailed here.

    For the reproducibility study, the samples were "blinded and coded" with "expected result" (e.g., High Positive, Negative). It's not specified how these initial categorizations were established.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method involving multiple experts resolving discrepancies for the test set results. The evaluations are primarily against established comparators like STTT or reference panels (CDC). For the reproducibility study, the agreement was 100%, so no adjudication would have been required for discrepancies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The performance studies focus on the device's standalone accuracy against existing diagnostic methods/reference panels, rather than how human readers' performance might improve with or without AI assistance. The device is an ImmunoBlot Kit, not an AI-assisted diagnostic imaging or interpretation system.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies reflect the standalone performance of the iDart™ Lyme IgG ImmunoBlot Kit. The "Principle of procedures" describes manual interpretation ("A strip reading guide included in each test kit shows the location of specific antigens in the test strip. ... Any band found having a visual intensity equal to or greater than the C2 control band intensity is considered as a significant (positive) band. Depending on the observed bands pattern, one can interpret the presence or absence of Lyme specific IgG antibodies in the patient serum."). The "Result Generation" for the device is listed as "Manual reading" in the comparison table. This indicates the studies assess the kit's performance as a laboratory test, interpreted by a human, but without a "human-in-the-loop" AI system.

    7. The Type of Ground Truth Used

    • Clinical Performance (Method Comparison): The "ground truth" was established by Standard Two-Tier Test Methodology (STTT), which involves FDA-cleared EIA and immunoblot tests, performed by laboratory personnel.
    • Clinical Sensitivity/Specificity: The "ground truth" was established by the CDC Reference Panel, which represents diagnosed cases of Lyme disease at various stages, look-alike conditions, and healthy controls. The methods for establishing these CDC diagnoses are not detailed but likely involve a combination of clinical assessment and established laboratory criteria.
    • Analytical Specificity / Cross-reactivity: Ground truth for these samples was "known to contain potentially cross-reactive antibodies to Lyme infection" or "healthy individuals" or specific disease states, implying prior clinical diagnoses or sample characterization.
    • Reproducibility: Samples were initially characterized as "High Positive," "Negative," etc., which served as the expected result for comparison. The method for this initial characterization is not specified.
    • Fresh/Frozen & Antibody Class Specificity: Ground truth was based on "previously tested patient samples" with known IgG status.

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of machine learning or AI. The iDart™ Lyme IgG ImmunoBlot Kit is an immunoassay kit, not an AI-based device that requires model training. Therefore, this question is not applicable to the information provided.

    9. How the Ground Truth for the Training Set Was Established

    As the device is not an AI/ML product, there is no "training set" or corresponding ground truth establishment process described in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230863

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-07-13

    (106 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    IgG/IgM Test System: ZEUS Solinas Borrelia VlsE1/pepC10 IgG/IgM Control Kit Regulation Number: 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZEUS Solinas Borrelia VIsE1/pepC10 IgG/IgM Test System uses chemiluminescent immunoassay (CLIA) technology for the qualitative detection of IgGlgM antibodies to Borrelia burgdorferi in human serum. This assay is intended for use on samples from patients with signs and symptoms consistent with or patients suspected of having Lyme disease to assess the presence of IgG/IgM antibodies.

    Positive results with the ZEUS Solinas Borrelia VIsE1/pepC10 IgG/IgM Test System should be supplemented with additional testing with a Standard two-tier test (STTT) methodology using an IgG and/or IgM Borrelia burgdorferi immunoblot assay following current guidelines.

    Positive supplemental results are supportive evidence of antibodies and exposure to Borrelia burgdorferi and may be used along with patient history, symptoms and other laboratory data to support a clinical diagnosis of Lyme disease.

    Negative results by the ZEUS Solinas Borrelia VlsE1/pepC10 IgG/IgM Test System should not be used to exclude Lyme disease. The test must be performed on the ZEUS Solinas instrument.

    The ZEUS Solinas Borrelia VIsE1/pepC10 IgG/IgM Control Kit is intended for use as assayed quality control samples to monitor the performance of the ZEUS Solinas Borrelia VlsE1/pepC10 IgG/IgM Test System. The performance characteristics of the ZEUS Solinas Borrelia VIsE1/pepC10 IgG/IgM Control Kit have not been established for any other assays or instrument platforms.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for the ZEUS Solinas Borrelia VlsE1/pepC10 IgG/IgM Test System. It describes a diagnostic test for Lyme disease. The information provided does not contain details about acceptance criteria, study data, sample sizes, expert qualifications, or ground truth establishment typically found in a clinical study report for an AI/ML medical device.

    The questions you've asked are most relevant to AI/ML device approval processes, where the device "learns" from data. This document describes a Chemiluminescent Immunoassay (CLIA) technology, which is a laboratory-based diagnostic test, not an AI/ML algorithm. Therefore, the concepts of training sets, test sets, human readers assisting AI, adjudication methods, and expert consensus for AI ground truth do not directly apply to this type of device.

    This clearance is based on substantial equivalence to legally marketed predicate devices, meaning it has similar indications for use and technological characteristics to devices already on the market. The performance of such a device is typically assessed through analytical and clinical performance studies, comparing its results to a reference method or clinical diagnosis, rather than through AI-specific metrics like those you've inquired about.

    To directly answer your request based on the provided document, I cannot fill in the table or provide the requested details because the document describes a traditional diagnostic assay, not an AI/ML device.

    Here's why each of your points cannot be addressed with the provided text:

    • 1. A table of acceptance criteria and the reported device performance: This document is the clearance letter, not the study report. It states the device is substantially equivalent but does not present the raw performance data or acceptance criteria that were met.
    • 2. Sample sizes used for the test set and the data provenance: Not present in this document.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable to a CLIA assay in the context typically asked for AI/ML. Ground truth for a diagnostic test usually refers to clinical diagnosis, a gold standard lab test, or patient outcomes, not expert image annotation.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable to a CLIA assay.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted device for human interpretation.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a laboratory test, not an algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): While not detailed, for a diagnostic test like this, ground truth would typically be established based on a combination of clinical diagnosis, other established laboratory tests, and possibly follow-up outcomes, not expert consensus in the way it's used for AI image interpretation.
    • 8. The sample size for the training set: Not applicable, as this is not an AI/ML device that requires a "training set."
    • 9. How the ground truth for the training set was established: Not applicable for the same reason as above.

    In summary, the provided document is an FDA 510(k) clearance letter for a traditional in-vitro diagnostic test, not an AI/ML medical device submission. Therefore, the specific questions related to AI/ML study design and performance metrics cannot be answered from this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220016

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-19

    (226 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K220016

    Trade/Device Name: Viramed Borrela All-In-One ViraChip Test Kit Regulation Number: 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Viramed Biotech AG Borrelia All-In-One ViraChip is an in vitro qualitative microarray assay for the detection of IgM and IgG antibodies to Borrelia burgdorferi in human serum. The assay is intended for testing serum samples from symptomatic patients or those suspected of Lyme Disease. It is intended to detect antibodies to VIsE and multiple other B. burgdorferi antigens following a modified two-tier test methodology. Positive results from the Viramed Biotech AG Borrelia All-In-One ViraChip are supportive evidence for the presence of antibodies and exposure to B. burgdorferi, the causative agent for Lyme disease. Negative results do not preclude infection with B. burgdorferi. Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures as an aid in diagnosis of Lyme disease.

    The Viramed Biotech AG Borrelia All-In-One ViraChip Test must be used with a ViraChip Reader and the ViraChip Software.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for a medical device. It does not contain the detailed performance study results, acceptance criteria, or ground truth establishment methods as requested. The letter only states that the device is substantially equivalent to a predicate device and provides its indications for use.

    Therefore, I cannot extract the information to complete the table and answer your questions directly from the provided text. The document clearly states "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent...". This implies that the detailed study information would be in the 510(k) submission itself, not in the clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211302

    Validate with FDA (Live)

    Device Name
    Elecsys Syphilis
    Manufacturer
    Date Cleared
    2021-07-20

    (82 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Indianapolis, Indiana 46250

    Re: K211302

    Trade/Device Name: Elecsys Syphilis Regulation Number: 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro qualitative detection of total antibodies (IgG and IgM) to Treponema pallidum in human serum and plasma. The test is intended as an aid in the diagnosis of syphilis infection in conjunction with clinical signs and symptoms.

    The Elecsys Syphilis immunoasay is not in screening blood or tissue donors. The effectiveness of this assay in testing blood or tissue donors has not been established.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.

    Device Description

    The Elecsys Syphilis immunoassay is a fully automated, qualitative assay that uses a double antigen sandwich format for the detection of IgM and IgG antibodies to T. pallidum.

    Recombinant T. pallidum antigens labeled with either biotin or a ruthenium complex bind to T. pallidum-specific IgG or IgM to form a double antigen sandwich complex. The sandwich complex binds to streptavidin-coated microparticles which can be immobilized magnetically to the surface of an electrode. Unbound substances are removed during a wash step using ProCell. A chemiluminescent substrate is then added to the reaction tube. Application of a voltage to the electrode induces a chemiluminescent emission which is measured by a photomultiplier.

    The presence or absence of anti-TP antibodies in the specimen is determined by comparing the chemiluminescent signal in the reaction to the cutoff index (COI) determined from an active calibration. The strength of the signal generated is proportional to the amount of bound conjugate and thus the amount of anti-T. pallidum antibodies present in the specimen. If the chemiluminescent signal in the reaction is greater than or equal to the cutoff signal, the specimen is considered reactive for anti-TP antibodies. If the chemiluminescent signal is below the cutoff signal, the specimen is considered nonreactive for the anti-TP antibodies.

    AI/ML Overview

    The provided text describes the Elecsys Syphilis immunoassay, an in vitro diagnostic device, and its evaluation for substantial equivalence to a predicate device (Elecsys Syphilis K160910) following an update to improve biotin tolerance.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this 510(k) submission revolve around demonstrating that the updated Elecsys Syphilis assay is as safe and effective as the predicate device, with the added benefit of improved biotin tolerance. The document states that "All performance specifications were met." While specific numerical acceptance criteria for each study (e.g., precision coefficients of variation, acceptable bias for method comparison) are not explicitly detailed as a formal table with pass/fail values, the reported device performance indicates that these criteria were satisfied.

    Study TypeAcceptance Criteria (Implied/General)Reported Device Performance
    Precision (CLSI EP5-A3)Demonstrate acceptable repeatability and intermediate imprecision for controls and human serum samples.Repeatability (CV%): HS, negative 1: 1.5; HS, negative 2: 2.0; HS, positive 1: 1.6; HS, positive 2: 2.4; HS, positive 3: 2.9; HS, positive 4: 2.5; HS, positive 5: 2.4; PC Syphilis 1: 1.1; PC Syphilis 2: 2.1 Intermediate Precision (CV%): HS, negative 1: 1.7; HS, negative 2: 3.0; HS, positive 1: 2.4; HS, positive 2: 3.1; HS, positive 3: 3.6; HS, positive 4: 3.6; HS, positive 5: 3.3; PC Syphilis 1: 1.4; PC Syphilis 2: 2.6 (These values are generally considered excellent for immunoassays, implying they met the internal acceptance limits.)
    Biotin Interference (CLSI EP07-A3)No significant interference with analyte quantitation up to a specified biotin concentration.Maximum value with no interference observed was 2400 ng/mL. The claim in the Instruction for Use will be < 1200 ng/mL. (This indicates a significant improvement over the predicate device's tolerance of ≤ 60 ng/mL, and the device met the new, higher biotin tolerance.)
    Method Comparison with PredicateDemonstrate substantial equivalence (e.g., high positive and negative agreement) between the updated and predicate device across the measuring range using samples from the intended use population.Positive Agreement and Negative Agreement were calculated between the current (predicate) and updated assay. The resulting data "support the equivalence of the current non-biotin and biotin-updated assay." (Specific percentages for agreement are not provided, but the statement indicates they met the substantial equivalence criteria.)
    Reagent and Calibration StabilityDemonstrate stability over time under specified storage conditions to support labeled claims.Stability studies and acceptance criteria were reviewed and found acceptable. The stability data "supports Roche Diagnostic's claims as reported in the package labeling."
    Lot Calibration StabilityDemonstrate consistent calibration performance across different manufacturing lots.(Implied to be met as part of the overall stability acceptance.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision Study:
      • Sample Size: Two aliquots of each of two levels of controls and 7 human serum samples were tested per run. This was done for 2 runs per day for 21 days. (Total of 9 distinct samples tested repeatedly).
      • Data Provenance: Not explicitly stated, but "human serum samples" suggests clinical samples. The text does not specify country of origin or if they were retrospective or prospective, but for precision studies, these factors are usually less critical than for clinical performance studies.
    • Biotin Interference Study:
      • Sample Size: Three serum samples (negative, close to cut-off, positive) were used. These were typically pooled samples.
      • Data Provenance: Not explicitly stated, but implies human serum.
    • Method Comparison Study:
      • Sample Size: A total of 232 samples from the intended use population.
      • Data Provenance: Not explicitly stated, but "samples from the intended use population" implies clinical samples. The text does not specify country of origin or if they were retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The Elecsys Syphilis assay is an in vitro diagnostic device that detects antibodies. Its performance is typically assessed by comparing its results to established reference methods (e.g., other FDA-cleared syphilis tests, Western Blot) or by using samples with a known status (e.g., from syphilis patients or healthy controls). The ground truth for this type of device is usually based on the results of these reference methods or known clinical status, rather than expert adjudication of images or clinical cases.

    4. Adjudication Method (for the test set)

    Adjudication methods (like 2+1, 3+1) are typically used in studies where human readers interpret complex data (e.g., medical images) to establish a consensus ground truth. This is not applicable to this type of in vitro diagnostic device study, as the ground truth is established through laboratory reference methods or known sample status.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This information is not applicable to this device. The Elecsys Syphilis is a fully automated immunoassay, not an AI-powered diagnostic device that assists human readers. Therefore, an MRMC study and analysis of human reader improvement with AI assistance are not relevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    The Elecsys Syphilis immunoassay is a standalone device in the sense that it is a fully automated assay that produces a result (Reactive or Nonreactive) without human interpretation steps in the analytical measurement process. Its performance (precision, biotin tolerance, method comparison) reflects its standalone analytical accuracy and reliability. However, its stated "Indications for Use" mention that it is an "aid in the diagnosis... in conjunction with clinical signs and symptoms" and that results "should always be interpreted in conjunction with additional treponemal or non-treponemal serologic test results." This implies that while the assay itself operates standalone, the final diagnosis based on its results involves a human clinician and other diagnostic information. Nevertheless, the performance studies described (precision, biotin tolerance, method comparison) represent the standalone analytical performance of the device.

    7. The Type of Ground Truth Used

    The ground truth for the test set samples used in the method comparison and potentially the precision studies is implied to be:

    • Clinical Status/Reference Method: Samples "from the intended use population" were used, suggesting samples from individuals with known or suspected syphilis status, likely confirmed by a reference method or a panel of tests. The comparison to the predicate device (another immunoassay) also helps establish agreement against a cleared standard.
    • Known Concentration/Status: For precision and biotin interference studies, samples might include negative, low positive, and high positive samples, or samples close to the cutoff, where their status or concentration is well-characterized.

    The document states that results "should always be interpreted in conjunction with additional treponemal or non-treponemal serologic test results (as appropriate), the patient's clinical symptoms, medical history, and other clinical and/or laboratory findings to produce a diagnosis of syphilis by disease stage." This holistic approach suggests the ultimate ground truth for a diagnosis is a comprehensive clinical assessment, but for assay performance evaluation, it relies on reference methods and known sample characteristics.

    8. The Sample Size for the Training Set

    This information is not provided. This device is an immunoassay, not a machine learning or AI algorithm in the context of typical "training sets." Immunoassays are developed based on chemical and biological principles (e.g., antibody-antigen binding, chemiluminescence) and calibrated, rather than "trained" with a dataset in the AI sense. Calibration involves using a small set of calibrator materials (Syphilis Cal1 and Cal2 mentioned).

    9. How the Ground Truth for the Training Set Was Established

    As explained above, the concept of a "training set" and "ground truth for a training set" in the AI sense does not directly apply to this immunoassay. The device is calibrated using specific calibrator materials (Syphilis Cal1 and Cal2), which are human serum preparations with known (negative or reactive) anti-TP antibody status, as stated in the device description. These calibrators establish the signal cutoff for reactive vs. nonreactive results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203289

    Validate with FDA (Live)

    Date Cleared
    2021-03-22

    (133 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test Kit Regulation Number: 21 CFR 866.3830
    Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test Kit Regulation Number: 21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gold Standard Diagnostics Borrelia burgdorferi VISE-OspC IgG/IgM ELISA Test Kit is intended as a qualitative test for the detection of IgG and IgM class antibodies to VIsE and OspC antigens from Borrelia burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of having Lyme disease. When used as the first-tier screening test, positive and equivocal results must be confirmed through additional testing by one of the following methods:

    • Standard two-tier test methodology (STTT) using an IgG and/or IgM blot testing following current interpretation guidelines. OR

    • Modified two-tier test methodology (MTTT) using one or more of the following three ELISA based assays:

    a. Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test

    b. Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test

    c. Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test

    The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with one or more of the following three ELISA based assays:

    a. Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test

    b. Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test

    c. Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test

    Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies. history, symptoms, and other laboratory findings.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter for the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test Kit. It states that the device is substantially equivalent to legally marketed predicate devices. However, this document does not contain the acceptance criteria for the device, nor does it provide details of a study proving the device meets acceptance criteria, or information regarding sample sizes, ground truth establishment, or expert involvement.

    The letter explicitly states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976..."

    Therefore, I cannot extract the requested information from the provided text. To answer your questions, the actual 510(k) summary or the full submission (which is usually publicly available on the FDA website) would be required.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203296

    Validate with FDA (Live)

    Date Cleared
    2021-03-22

    (133 days)

    Product Code
    Regulation Number
    866.3830
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name: Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit Regulation Number: 21 CFR 866.3830
    Borrelia burgdorferi IgG ELISA Test Kit
    Common Name: Lyme ELISA Test

    Regulation Section:

    (21 CFR 866.3830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit is intended as a qualitative test for the detection of IgG antibodies to B. burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. When used as the first-tier screening test, positive and equivocal results must be supplemented through additional testing by one of the following methods:
    · Standard two-tier test methodology (STTT) using an IgG blot test following current interpretation guidelines, OR
    • Modified two-tier test methodology (MTT) using the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/ IgM ELISA Test.
    The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test as the first-tier screening test.
    Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies, history, symptoms, and other laboratory findings.

    Device Description

    The kit includes 12 x 8 well Antigen Coated strips. Conjugate. Substrate, Stop Solution, Wash Buffer, Diluent, Negative Control, Positive Control, and Cutoff Control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations.
    During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgG antibodies conjugated with horseradish peroxidase is then added, which binds to the antigenantibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450nm.
    The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test kit is a combination of B. burgdorferi sensu stricto strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VlsE from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography. The purity of each antigen is assayed by SDS-PAGE followed by Coomassie staining and/or Western blotting.

    AI/ML Overview

    This document describes the validation of the Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit, including a new proposed Modified Two-tier Testing (MTTT) methodology. The provided text primarily focuses on comparative studies against predicate devices and existing standard methods, rather than defining explicit acceptance criteria in a quantitative table format. However, implied acceptance criteria can be derived from the performance percentages presented.

    Based on the provided information, here's a structured response addressing the requested points:

    1. Table of Acceptance Criteria (Implied) and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical cutoffs in the document. However, they can be inferred from the reported "Percent Agreement" values in the comparison studies. The studies aim to demonstrate substantial equivalence to established predicate devices and methodologies.

    Study Type / Performance MetricImplied Acceptance Criteria (Goal: High Agreement/Sensitivity/Specificity)Reported Device Performance (Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit)
    Comparison with Predicate Device (STTT)
    Positive Percent Agreement (PPA) with Predicate IgG ELISAHigh (> 90% typically desired for equivalence)90.2% (55/61) [95% CI: 79.8% - 96.3%]
    Negative Percent Agreement (NPA) with Predicate IgG ELISAHigh (> 95% typically desired for equivalence)99.6% (460/462) [95% CI: 98.5% - 99.9%]
    Second-Tier Testing (STTT) - Agreement with FDA Cleared IgG Blot
    2nd Tier PPA (Tier 1 Pos/Equiv samples)100.0% (as reported)100.0% (37/37) [95% CI: 92.2% - 100.0%]
    Clinical Sensitivity (STTT)
    Early LymeImprovement over predicate46.6% (27/58) vs. Predicate: 27.6% (16/58)
    Disseminated LymeImprovement over predicate82.4% (14/17) vs. Predicate: 52.9% (9/17)
    Late LymeHigh agreement with predicate97.4% (38/39) vs. Predicate: 97.4% (38/39)
    CDC Panel (STTT)
    HealthyHigh (> 99% agreement, implying high specificity)99.0% (1/100 positive/equivocal)
    Early LymeImprovement over predicate68.3% (41/60 positive/equivocal) vs. Predicate: 35.0% (21/60)
    Cardiac LymeComparable/Improvement66.7% (2/3 positive/equivocal) vs. Predicate: 100.0% (3/3)
    Neurological LymeImprovement over predicate85.7% (6/7 positive/equivocal) vs. Predicate: 42.9% (3/7)
    Late LymeHigh agreement with predicate100.0% (20/20 positive/equivocal) vs. Predicate: 100.0% (20/20)
    Look-alike DiseaseHigh (> 85% agreement, implying high specificity)87.8% (11/90 positive/equivocal) vs. Predicate: 88.9% (10/90)
    Method Comparison MTTT – IgG (vs. STTT)
    PPA (second-tier test of MTTT vs. STTT)100.0% (as reported)100.0% (23/23) [95% CI: 85.2% - 100.0%]
    NPA (second-tier test of MTTT vs. STTT)High (implied > 95%)0.0% (0/15) [95% CI: 0.0% - 21.8%] - Note: This NPA is for a subset of samples that were negative by IgG Immunoblot, not overall. The overall NPA is 96.7%
    PPA (Overall MTTT vs. STTT)100.0% (as reported)100.0% (23/23) [95% CI: 85.2% - 100.0%]
    NPA (Overall MTTT vs. STTT)High (implied > 95%)96.7% (443/458) [95% CI: 94.7% - 98.2%]
    Clinical Sensitivity (MTTT)
    Early LymeImprovement over predicate STTT48.4% (30/62) vs. Predicate STTT: 8.1% (5/62)
    Disseminated LymeImprovement over predicate STTT81.8% (18/22) vs. Predicate STTT: 22.7% (5/22)
    Late LymeComparable/Improvement97.6% (40/41) vs. Predicate STTT: 95.1% (39/41)
    CDC Reference Panel (MTTT)
    Healthy100.0% (as reported)100.0% (0/100 positive/equivocal)
    Early LymeImprovement over predicate STTT60.0% (36/60 positive/equivocal) vs. Predicate STTT: 33.3% (12/60)
    Cardiac LymeImprovement over predicate STTT66.7% (2/3 positive/equivocal) vs. Predicate STTT: 33.3% (1/3)
    Neurological LymeImprovement over predicate STTT85.7% (6/7 positive/equivocal) vs. Predicate STTT: 14.3% (1/7)
    Late Lyme100.0% (as reported)100.0% (20/20 positive/equivocal) vs. Predicate STTT: 100.0% (20/20)
    Look-alike Disease100.0% (as reported)100.0% (0/90 positive/equivocal) vs. Predicate STTT: 100.0% (0/90)

    2. Sample Size Used for the Test Set and Data Provenance

    • Determination of Assay Cutoff (Nonclinical):

      • Sample Size: 210 normal sera (105 endemic, 105 non-endemic) + 194 samples (114 Lyme disease stages, 8 healthy negative, 72 negative Lyme disease with other conditions). Total = 404 samples.
      • Data Provenance: Not explicitly stated, but "endemic region of Lyme disease" and "non-endemic region of Lyme disease" suggest geographical diversity. Retrospective, as these are "tested" samples for cutoff determination.
    • Precision (Nonclinical):

      • Sample Size: 48 runs for each of 4 panel members (negative, high negative, low positive, moderate positive) + kit controls. This is a repetitive testing design, not individual patient samples.
    • Reproducibility (Nonclinical):

      • Sample Size: 90 runs (3 sites x 2 runs/day x 5 days x 3 replicates) for each of 4 panel members (negative, high negative, low positive, moderate positive) + kit controls (30 runs for Pos/Neg control, 60 runs for Cutoff control).
    • Analytical Specificity:

      • Sample Size: 208 asymptomatic individuals (103 from endemic regions, 105 from non-endemic regions).
      • Data Provenance: Endemic and non-endemic regions. Retrospective (asymptomatic individuals' samples).
    • Cross Reactivity:

      • Sample Size: 377 samples.
      • Data Provenance: Samples obtained from "serum vendors who confirmed their positivity for each respective marker," implying retrospective, controlled samples.
    • Comparison/Prospective Study (Clinical):

      • STTT Comparison: 523 serum samples.
      • MTTT Comparison: 481 serum samples for the initial screening. 38 positive/equivocal samples were carried forward for second-tier testing comparison.
      • Data Provenance: "Prospective samples submitted for Lyme serology testing" from "three sites (one internal and two external reference laboratories)." This indicates a prospective study design with samples from potentially diverse geographical locations (clinical labs).
    • Sensitivity Study (Clinical):

      • STTT: 114 clinically characterized samples.
      • MTTT: 125 clinically characterized samples.
      • Data Provenance: "Clinically characterized samples" from unspecified sources; likely retrospective collection of known patient cohorts.
    • CDC Panel (Clinical):

      • Sample Size: 280 positive and negative specimens.
      • Data Provenance: From the "Center of Disease Control (CDC)." These are reference panels, typically collected and characterized over time, so retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test sets.

    • For "clinically characterized samples" and "CDC panel" samples, the implication is that these samples have a pre-defined and reliable "clinical diagnosis" or "reference characterization" based on established medical criteria. This often involves consensus from clinical experts (e.g., infectious disease specialists) and/or a combination of clinical presentation, symptoms, and other laboratory findings, but the specifics are not detailed here.
    • For the "Second-Tier Testing" comparison, the "FDA cleared IgG blot assay" and "predicate B. burgdorferi IgG blot test" serve as the ground truth. These are approved diagnostic methods, and their interpretation would follow established guidelines.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for resolving discrepancies in the test results or for establishing the initial ground truth diagnoses. The comparisons are based on the results of the different assays and against established clinical characterizations or CDC panels.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was performed or described. This device is an ELISA test kit, a laboratory diagnostic assay for detecting antibodies, not an AI or imaging diagnostic tool that involves human readers interpreting images. Therefore, the concept of "human readers improving with AI assistance" is not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the primary performance evaluation is standalone. The Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit operates as a standalone diagnostic assay. Its performance (sensitivity, specificity, agreement) is measured as the output of the kit itself (optical density readings converted to units and qualitative results) without direct human interpretation of complex visual patterns or AI assistance. The results are quantitative and then interpreted qualitatively (positive, equivocal, negative) based on predefined cutoffs.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used varies by study:

    • Predicate Devices/Methodologies: For the STTT and MTTT comparison studies, the ground truth for the device's performance is a comparison against the results from legally marketed predicate devices (e.g., Trinity Biotech MarDx Borrelia burgdorferi EIA IgG Test Kit) and/or a FDA cleared IgG blot assay. This is a form of comparative ground truth against established methods.
    • Clinical Diagnosis/Characterization: For the Sensitivity Study and CDC Panel, the ground truth is "clinically characterized samples" and "positive and negative specimens from the Centers of Disease Control (CDC)." This implies a clinical diagnosis or consensus diagnosis based on comprehensive patient data (history, symptoms, other laboratory findings) or a reference panel with established disease status.
    • Analytical Ground Truth: For non-clinical studies like analytical specificity and cross-reactivity, the ground truth is often the confirmed absence or presence of specific conditions in the tested samples, usually based on prior laboratory testing or sample vendor claims.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning. The term "training set" is usually associated with the development of AI algorithms. For an ELISA kit, development involves:

    • Antigen Selection and Optimization: Which antigens to use (e.g., B31 lysate, 2591 lysate, recombinant VlsE).
    • Reagent Formulation: Optimizing conjugate, substrate, buffers, etc.
    • Cutoff Determination: "The cutoff was determined by testing a total of 210 normal sera..." and "An additional 194 samples..." This process of determining the cutoff could be considered analogous to a 'calibration' or 'training' phase for a traditional diagnostic test, where a dataset is used to define the diagnostic thresholds. In this sense, a total of 404 samples (210 normal + 194 other known samples) were used for cutoff determination.

    9. How the Ground Truth for the Training Set Was Established

    As discussed above, for establishing the cutoff (analogous to training/calibration):

    • Normal Sera: The 210 normal sera were likely verified as "normal" through standard clinical practice or donor screening, indicating the absence of target antibodies/disease in a healthy population.
    • Known Positive/Negative/Other Disease Samples: The additional 194 samples included "different phases of Lyme disease," "negative healthy samples," and "negative Lyme disease samples but do have other diseases that may cause serologic cross-reactivity." The "ground truth" for these samples would be their known clinical status or disease classification, typically established through comprehensive clinical evaluation, follow-up, and/or panels from disease banks. The ROC analysis was used to confirm the chosen cutoff provides the best compromise between sensitivity and specificity, indicating an analytical optimization process based on these known samples.