    510k Summary Text (Full-text Search) :

    -------------|------------------------------------------------------|
    | Regulation Number: | 21 CFR 866.3130
    |
    | Regulation number: | 21 CFR §866.3130
    Class II Special Controls Guidelines appropriate for 21 CFR §866.3130.
    difficile" (issued on August 27, 2015) as appropriate for this Class II device under regulation 21 CFR §866.3130
    Infinity System),

    • Demonstrate compliance with the Class II Special Controls under regulation 21 CFR §866.3130
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xpert® C. difficile/Epi test is a qualitative in vitro diagnostic test for detection of toxin B gene sequences and for presumptive identification of 027/NAP1/BI strains of toxigenic Clostridioides difficile from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of 027/NAP1/BI strains of C. difficile is by detection of binary toxin (CDT) gene sequences and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the GeneXpert® Instrument Systems and utilizes automated real-time polymerase chain reaction (PCR) to detect toxin gene sequences associated with toxin producing C. difficile. The Xpert® C. difficile/Epi test is intended as an aid in the diagnosis of CDI. Detection of 027/NAP/Bl strains of C. difficile by the Xpert® C. difficile/Evi test is presumptive and is solely for enidemiological purposes and is not intended to guide or monitor treatment for C. difficile infections. Concomitant culture is necessary only if further typing or organism recovery is required.

    Device Description

    The Xpert C. difficile/Epi test is an automated in vitro diagnostic test for qualitative detection of toxin producing Clostridioides difficile (formerly known as Clostridium difficile) directly from unformed (liquid or soft) stool specimens of patients suspected of having Clostridioides difficile infection (CDI). The test detects the toxin B gene (tcdB), the binary toxin (CDT) gene, and the single base pair deletion at nucleotide 117 (tcdCΔ117) within the gene encoding TcdC, a negative regulator of toxin production. The combined presence of the genes encoding toxin B and binary toxin and the tcdCA117 deletion has been associated with a hypervirulent C. difficile strain known as 027/NAP1/BI, which has been associated with severe disease outbreaks in healthcare facilities worldwide.

    The test is performed on the GeneXpert® Instrument Systems (comprised of the GeneXpert® Dx System. GeneXpert® System with Touchscreen, and GeneXpert® Infinity System). In the GeneXpert® Instrument Systems, sample preparation, amplification, and real-time detection are all fully automated and completely integrated. The single-use disposable Xpert C. difficile/Evi cartridge, required by the platform for use, holds the PCR reagents and hosts the PCR process. Because the cartridge is self-contained, cross-contamination between samples is minimized.

    The Xpert C. difficile/Epi test includes reagents for the detection of toxigenic C. difficile and the detection of sequences for presumptive identification of 027/NAP 1/BI strains. In addition, the test reagents include an internal sample processing control (SPC) to ensure adequate processing of the target bacteria and to monitor the presence of inhibitor(s) in the PCR Assay. The SPC also ensures that the PCR conditions (temperature and time) are appropriate for the amplification reaction and that the PCR reagents are functional. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

    The Xpert C. difficile/Evi test system performs sample preparation and real-time, multiplex polymerase chain reaction (PCR) for detection of target-specific DNA. A swab is inserted into the stool specimen and then placed in a tube containing Sample Reagent. Following brief vortexing, the content of the Sample Reagent is transferred to the Sample Chamber of the disposable fluidic cartridge (the Xpert C. difficile/Epi cartridge). The user initiates a test from the system user interface of the GeneXpert® Instrument Systems and places the cartridge with sample into a GeneXpert® instrument system which performs hands-off real-time PCR for detection of C. difficile DNA.

    Depending on the specific instrument, a GeneXpert® instrument system may contain 1-80 modules, each of which are randomly accessible and capable of performing separate sample preparation and real-time PCR tests for the detection of gene sequences for C. difficile toxin B and binary toxin and the tcdCA117 deletion in less than 45 minutes. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

    Each instrument in the GeneXpert® instrument family is equipped with a Windows OS-based personal computer that is preloaded with software applications for running the tests and viewing the results, as described in Table 1.

    AI/ML Overview

    The provided document is a 510(k) summary for the Cepheid Xpert C. difficile/Epi test. It details the device's characteristics, intended use, and performance studies conducted to demonstrate substantial equivalence to a predicate device and compliance with special controls.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state formal acceptance criteria with numerical targets (e.g., minimum sensitivity/specificity percentages). Instead, it summarizes "verification studies" to demonstrate performance and compliance. The key performance indicators mentioned relate to agreement with expected results and accurate identification of strains.

    Acceptance Criteria (Implied)Reported Device Performance
    Maximum acceptable hold time for a prepared cartridgeVerified as 4 hours under all tested environments (ambient, elevated with high humidity, high temperatures).
    Equivalent performance across different GeneXpert instruments/softwareFunctional testing showed 100% agreement with expected results and no statistically significant differences in Ct values among test runs in all instrument/software combinations (GeneXpert Dx System, GeneXpert Infinity System running GeneXpert Dx v6.5 and GeneXpert Xpertise v6.8/v7.1).
    Accurate identification of C. difficile strains (Inclusivity)Accurately identified all 26 strains of toxigenic C. difficile tested, demonstrating an inclusivity rate of 100%.
    Compliance with Class II Special Controls GuidelinesThe device met the requirements of the Class II Special Controls and accurately identified all strains of toxigenic C. difficile tested.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set:
      • Prepared Cartridge Hold Time Study: Not explicitly stated how many "contrived positive" and "negative" samples were used, only that they were held for "various times under three (3) hold environments."
      • Functional Testing: Not explicitly stated how many "contrived positive" and "negative" samples were used.
      • Inclusivity Study: 26 C. difficile strains were tested (18 from previous study K110203 and 8 additional strains for K243730).
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. The studies mentioned (Hold Time, Functional Testing, Inclusivity) appear to be laboratory-based verification studies, rather than clinical trials with patient samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not mention the use of experts to establish ground truth for the test set. The performance studies appear to be analytical verification studies using contrived samples and known bacterial strains, implying that the ground truth for these studies was established by laboratory methods rather than expert consensus on clinical samples.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any adjudication method for establishing ground truth, as the studies are analytical verification studies using controlled samples with known statuses (contrived positive/negative, specific C. difficile strains).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is a diagnostic assay (Nucleic Acid Amplification Test) for infectious disease, not an AI-powered image analysis tool requiring human reader interpretation. No mention of AI assistance or human readers is present.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies described (Prepared Cartridge Hold Time, Functional Testing, Inclusivity) assess the performance of the Xpert C. difficile/Epi test in a standalone manner. The device is an automated in vitro diagnostic test that performs sample preparation, amplification, and real-time detection without human intervention in the result determination process. It's an "algorithm-only" performance in the sense that the instrument provides a final result based on its internal processes.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the verification studies was based on:

    • Known status of contrived samples: "contrived positive (C. difficile cells added to negative matrix) and negative (negative matrix only) samples" for hold time and functional testing.
    • Known bacterial strains: "C. difficile strains selected to broadly represent the majority of C. difficile toxinotypes encountered in practice" for the inclusivity study, where accurate identification of these known strains served as the ground truth.

    8. The sample size for the training set

    The document does not mention a "training set" in the context of machine learning or AI. This is a diagnostic assay, and its development likely involved traditional analytical validation and verification rather than an AI model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable, as no training set (in the AI/ML sense) is mentioned or implied for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232092

    Validate with FDA (Live)

    Date Cleared
    2023-11-14

    (124 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: Great Basin Toxigenic C. difficile Direct Test (CDF2) Regulation Number: 21 CFR 866.3130
    Requlation: 21 CFR 866.3130 - Clostridium difficile toxin gene amplification assay

    • b.
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Great Basin Toxigenic C. difficile Direct Test, performed on the Great Basin PA500 Analyzer, is a qualitative in vitro diagnostic test for the detection of toxigenic Clostridioides difficile in unformed (liquid or soft) stool samples collected from patients suspected of having C. difficile infection (CDI). The automated assay utilizes polymerase chain reaction (PCR) to detect a conserved region of the toxin gene (tcdB) associated with toxin-producing C. difficile.

    The Toxigenic C. difficile Direct Test is intended for use as an aid in the diagnosis of CDI in conjunction with clinical and epidemiological risk factors.

    Device Description

    The Great Basin Toxigenic C. difficile Direct Test (CDF2) performed on the PA500 Analyzer utilizes automated, hot-start PCR amplification technology to amplify specific nucleic acid sequences that are then detected using hybridization probes immobilized on a modified silicon chip surface, in a single-use, self-contained test cartridge.

    A swab volume of the specimen (raw stool) is first processed using the Sample Preparation Device (SPD). During processing through the SPD the specimen is infused with a synthetic bacterial sample processing control (SPC). An aliquot (250 µL) of the eluate obtained from the SPD, containing diluted specimen mixed with SPC, is loaded into the sample port of the CDF2 Test Cartridge.

    Genomic DNA is extracted from microbial cells and diluted to reduce potential inhibitors of PCR. During the PCR process, biotin-labeled primers direct the amplification of specific nucleic acid sequences within a conserved region of the C. difficile Toxin B (tcdB) gene.

    Following PCR, biotin-labeled, amplified target DNA sequences are hybridized to sequence specific probes immobilized on the silicon chip surface and incubated with antibody conjugated to the horseradish peroxidase enzyme (HRP). The unbound conjugate is washed away, and then tetramethylbenzidine (TMB) is added to produce a colored precipitate at the location of the probe/target sequence complex.

    The resulting signal is detected by the automated PA500 Optical Reader within the PA500 Analyzer System. The SPC undergoes the same extraction, and detection steps as the sample in order to monitor for inhibitory substances, as well as process inefficiency due to instrument or reagent failure. No operator intervention is required once the sample is loaded into the sample port and the Toxigenic C. difficile Direct Test cartridge is loaded into the PA500 Analyzer.

    The Great Basin System is a fully automated in vitro diagnostic system that includes: the PA500 Analyzer, single-use SPDs, single-use Toxigenic C. difficile Direct Test Cartridges, and the PA500 Data Analysis Software Program. The PA500 Analyzer is designed to perform automated sample preparation, target amplification via PCR, chip-based (optical) target detection, and integrated data analysis in less than two hours.

    AI/ML Overview

    The Great Basin Toxigenic C. difficile Direct Test (CDF2) is a qualitative in vitro diagnostic test designed to detect toxigenic Clostridioides difficile in unformed stool samples from patients suspected of CDI. The device utilizes PCR to detect a conserved region of the toxin B gene (tcdB).

    The study that proves the device meets the acceptance criteria is a prospective multi-center method comparison study (section H. Performance Data - Clinical Studies, page 12).

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the provided text. However, the study aims to demonstrate "acceptable performance" and states that the "information submitted...supports substantial equivalence." Based on the clinical performance, the implicit acceptance criteria would involve high positive and negative percent agreements with an FDA-cleared molecular reference.

    MetricAcceptance Criteria (Implicit)Reported Device Performance (All Sites Combined)
    Positive Percent Agreement (PPA)High agreement with reference95.8% (91.2% - 98.5% CI)
    Negative Percent Agreement (NPA)High agreement with reference98.5% (97.3% - 99.3% CI)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Test Set): A total of 829 specimens were included in the final dataset (page 12).
    • Data Provenance: The study was conducted at three geographically diverse U.S. clinical study sites (Midwest, West). The data is prospective, as specimens were collected from symptomatic individuals suspected of CDI between June 2022 and February 2023 (page 12). The specimens were "excess remnants of unpreserved stool specimens...processed according to routine standard of care testing, and would have otherwise been discarded" (page 12).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The ground truth for the clinical test set was established using an FDA-cleared molecular reference (page 12), not human experts. Discrepant results were investigated by testing on a third molecular test (page 14). Therefore, no human experts were directly involved in establishing the primary ground truth for the clinical samples.

    4. Adjudication Method for the Test Set

    The adjudication method for discrepant results was comparison to a third FDA-cleared molecular C. difficile test (page 14). If the CDF2 test result differed from the initial molecular comparator, the specimen was re-tested using a third molecular test to resolve the discrepancy.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The study compares the device performance against a molecular reference, not human readers.

    6. If a Standalone Study (algorithm only without human-in-the-loop performance) was done

    Yes, this was a standalone study. The device is an automated in vitro diagnostic test where "No operator intervention is required once the sample is loaded into the sample port and the Toxigenic C. difficile Direct Test cartridge is loaded into the PA500 Analyzer" (page 4). The performance metrics (PPA, NPA) directly reflect the device's accuracy against a molecular comparator without human interpretation of the results.

    7. The Type of Ground Truth Used

    The primary ground truth for the clinical study was an FDA-cleared molecular reference (page 12). For discrepant samples, a third FDA-cleared molecular C. difficile test was used for resolution (page 14).

    8. The Sample Size for the Training Set

    The document does not explicitly state a separate training set or its size for the algorithm itself. The studies described are primarily analytical (LoD, inclusivity, exclusivity, interference, reproducibility) and a clinical validation study. These types of in vitro diagnostic device submissions typically focus on analytical and clinical performance rather than a distinct "training set" for an AI algorithm in the same sense as image recognition or similar machine learning applications. The device's "automated data analysis software program" (page 4) is part of its design, and its performance is evaluated through the described analytical and clinical studies.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" is described for an AI algorithm, the method for establishing its ground truth is not detailed. However, the performance is established through rigorous analytical studies using characterized microbial strains and spiked samples (e.g., cell cultures with known CFU/mL) and clinical specimens compared to a molecular reference.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212427

    Validate with FDA (Live)

    Date Cleared
    2021-10-20

    (77 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name: cobas Cdiff nucleic acid test for use on the cobas Liat System Regulation Number: 21 CFR 866.3130
    |
    | Regulation Number | 21 CFR 866.3130
    |
    | RegulationNumber | Same | 21 CFR 866.3130

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is an automated, qualitative in vitro diagnostic test, that uses real-time polymerase chain reaction (PCR), for the detection of the toxin B (tcdB) gene of toxigenic Clostridioides difficile (C.difficile) in unformed (liquid or soft) stool specimens obtained from patients suspected of having C. difficile Infection (CDI). The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is intended for use as an aid in the diagnosis of CDI in humans in conjunction with clinical and epidemiological risk factors.

    Device Description

    The cobas® Cdiff Nucleic Acid Test for use on the cobas® Liat® System (cobas® Cdiff) is a rapid, automated in vitro diagnostic test for qualitative detection and differentiation of C. difficile DNA in human stool specimens. The cobas® Liat® is for in vitro diagnostic use. The system is designed to identify and/or measure presence of genetic material in a biological sample. The system automates all nucleic acid amplification test (NAAT) processes, including reagent preparation, target enrichment, inhibitor removal, nucleic acid extraction, amplification, real-time detection, and result interpretation in a rapid manner. The cobas® Cdiff test uses silica magnetic particle-based nucleic acid extraction and TaqMan probe-based real-time PCR amplification and detection. The cobas® Liat® Analyzer automates and integrates sample purification, nucleic acid amplification and detection of the target sequence in biological samples. Other than adding the sample to the cobas® Cdiff assay tube, no reagent preparation or additional steps are required. The cobas® Cdiff assay tube that holds all of the sample purification and PCR reagents and hosts the sample preparation and PCR process specific for the Cdiff analyte. The test uses the assay tube as both the sample and reaction vessel. The assay tube comprises flexible tubing containing all required unit dose reagents pre-packed in tube segments, separated by pressure-sensitive seals, in the order of reagent use. During the testing process, multiple sample processing actuators of the analyzer compress the cobas® Cdiff assay tube to selectively release reagents from tube segments, move the sample from one segment to another, and control reaction conditions such as reaction volume, temperature, pressure, and incubation time. Precise control of all these parameters provides optimal conditions for assay reactions, allowing the test to achieve high performance similar to or better than that of currently available molecular assays. The cobas® Liat® Analyzer software controls and coordinated these actions to perform all required assay processes, including sample preparation, nucleic acid extraction, target enrichment, inhibitor removal, nucleic acid elution, and real-time PCR. All assay steps are performed within the closed and self-contained cobas® Cdiff assay tube, thereby eliminating the potential for cross-contamination between samples. The collected data are automatically analyzed and the result is displayed in the assay report on the integrated LCD touch screen of the cobas® Liat® Analyzer.

    AI/ML Overview

    The document provided does not contain the level of detail required to answer all aspects of your request. Specifically, it is a 510(k) summary for a change in shelf life for an existing device, the cobas® Cdiff Nucleic Acid Test. Therefore, it focuses on demonstrating that this change does not negatively impact performance, rather than providing all the original study details for the device's initial clearance.

    Here's an analysis of the provided text in relation to your questions:

    1. A table of acceptance criteria and the reported device performance.

    The document does not explicitly state specific numerical acceptance criteria for the initial device performance. It focuses on demonstrating that the performance of the device with the new shelf life is equivalent to the currently cleared device.

    However, it does mention a new stability study was performed with "modified stability testing approach and acceptance criteria." It then states the result:

    Acceptance Criteria (Implied)Reported Device Performance
    Maintenance of overall cobas® Cdiff assay performance and claims, substantially equivalent to the currently cleared device.Overall cobas® Cdiff assay performance and claims are substantially equivalent to the currently cleared device, with reagents up to ten (10) months after the date of manufacture, supporting a shelf life claim of 9 months.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective).

    The document mentions a "new stability study" was performed to verify the shelf life. It does not provide:

    • Sample size: No explicit number of samples or runs for this stability study is provided.
    • Data Provenance: Not mentioned (e.g., country of origin).
    • Retrospective/Prospective: Implied to be prospective for the shelf-life stability study, as it's a "new stability study." However, for the original clearance of the device itself, this information is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience).

    This device is an in vitro diagnostic (IVD) test for detecting a bacterial gene (Clostridioides difficile toxin B gene). The "ground truth" for such a test would typically be established by:

    • Reference methods: E.g., bacterial culture, toxigenic culture, or a highly sensitive and specific laboratory reference PCR assay.
    • Clinical diagnosis: Based on patient symptoms, other lab tests, and clinical assessment.

    This document does not mention the use of human experts (like radiologists) for ground truth establishment. For an IVD, the "experts" are the laboratory and clinical staff following established protocols to provide the comparator data. The document does not specify details about these "experts" or their qualifications.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set.

    Not applicable or not mentioned. Adjudication methods like 2+1 or 3+1 are typically used in image analysis studies where human readers interpret images. For an IVD, the comparison is usually against a pre-defined "gold standard" or reference method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

    Not applicable. This is an IVD test, not an AI-powered diagnostic imaging device that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done.

    Yes, by its nature, an IVD test like the cobas® Cdiff Nucleic Acid Test on the cobas® Liat® System is a standalone device. The "algorithm" here refers to the pre-programmed steps of nucleic acid extraction, amplification, and real-time detection performed by the automated system, leading to a qualitative result (positive/negative). The document states: "The cobas® Liat® Analyzer automates and integrates sample purification, nucleic acid amplification and detection of the target sequence in biological samples... The collected data are automatically analyzed and the result is displayed in the assay report on the integrated LCD touch screen of the cobas® Liat® Analyzer." This indicates standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc).

    The document implies the ground truth for establishing initial performance would have been comparison to a reference method, likely a combination of toxigenic C. difficile culture and a highly reliable molecular method, or clinical diagnosis supported by confirmatory lab tests in the original clearance (K171770). For the current shelf-life stability study, the ground truth appears to be the expectation that the device continues to perform according to its established analytical specifications when tested at different time points after manufacturing. It refers to "assay performance and claims."

    8. The sample size for the training set.

    Not applicable/Not mentioned. As an IVD test, there isn't a "training set" in the machine learning sense. The device's performance is established through analytical and clinical validation studies with defined sample sizes. The document doesn't provide these sizes for the original device, nor for any "training" in the context of this specific 510(k) which is for a shelf-life change.

    9. How the ground truth for the training set was established.

    Not applicable/Not mentioned for the same reasons as #8. For an IVD, the "ground truth" for validation studies (analogous to a test set in ML) would be established by reference methods or clinical diagnosis.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210385

    Validate with FDA (Live)

    Date Cleared
    2021-09-09

    (212 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name: cobas Cdiff nucleic acid test for use on the cobas Liat System Regulation Number: 21 CFR 866.3130
    |
    | Regulation Number | 21 CFR 866.3130
    Number | Same | 21 CFR 866.3130

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is an automated, qualitative in vitro diagnostic test that uses real-time polymerase chain reaction (PCR) for the detection of the toxin B (tcdB) gene of toxigenic Clostridioides difficile (C.difficile) in unformed (liquid or soft) stool specimens obtained from patients suspected of having C. difficile infection (CDI). The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is intended for use as an aid in the diagnosis of CDI in humans in conjunction with clinical and epidemiological risk factors.

    Device Description

    The cobas® Cdiff Nucleic Acid Test for use on the cobas® Liat® System (cobas® Cdiff) is a rapid, automated in vitro diagnostic test for qualitative detection and differentiation of C. difficile DNA in human stool specimens. The cobas® Liat® is for in vitro diagnostic use. The system is designed to identify and/or measure presence of genetic material in a biological sample. The system automates all nucleic acid amplification test (NAAT) processes, including reagent preparation, target enrichment, inhibitor removal, nucleic acid extraction, amplification, real-time detection, and result interpretation in a rapid manner. The cobas® Cdiff test uses silica magnetic particle-based nucleic acid extraction and TaqMan probe-based real-time PCR amplification and detection. The cobas® Liat® Analyzer automates and integrates sample purification, nucleic acid amplification and detection of the target sequence in biological samples. Other than adding the sample to the cobas® Cdiff assay tube, no reagent preparation or additional steps are required. The cobas® Cdiff assay tube that holds all of the sample purification and PCR reagents and hosts the sample preparation and PCR process specific for the Cdiff analyte. The test uses the assay tube as both the sample and reaction vessel. The assay tube comprises flexible tubing containing all required unit dose reagents pre-packed in tube segments, separated by pressure-sensitive seals, in the order of reagent use. During the testing process, multiple sample processing actuators of the analyzer compress the cobas® Cdiff assay tube to selectively release reagents from tube segments, move the sample from one segment to another, and control reaction conditions such as reaction volume, temperature, pressure, and incubation time. Precise control of all these parameters provides optimal conditions for assay reactions, allowing the test to achieve high performance similar to or better than that of currently available molecular assays. The cobas® Liat® Analyzer software controls and coordinates these actions to perform all required assay processes, including sample preparation, nucleic acid extraction, target enrichment, inhibitor removal, nucleic acid elution, and real-time PCR. All assay steps are performed within the closed and self-contained cobas® Cdiff assay tube, thereby eliminating the potential for cross-contamination between samples. The collected data are automatically analyzed and the result is displayed in the assay report on the integrated LCD touch screen of the cobas® Liat® Analyzer.

    AI/ML Overview

    The provided text is a 510(k) summary for the cobas® Cdiff Nucleic Acid Test for use on the cobas® Liat® System. The purpose of this submission is to demonstrate substantial equivalence to a previously cleared device (K171770), specifically addressing a software update (Liat® Analyzer Software 3.3).

    Therefore, the primary study discussed is a performance evaluation to demonstrate that the software update does not negatively impact the device's performance, thereby maintaining substantial equivalence to the predicate device.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with numerically reported performance metrics for sensitivity, specificity, accuracy, etc., as typically seen in initial device clearance studies. Instead, it states that the performance of the modified device was evaluated and that the overall cobas® Cdiff assay performance and claims were not impacted by the software changes. This implies that the acceptance criterion was essentially "no significant change in performance" compared to the predicate device, which had already met its own performance criteria.

    Acceptance Criterion (Implied)Reported Device Performance
    No significant impact on overall cobas® Cdiff assay performance"The result of this evaluation determined that the overall cobas® Cdiff assay performance and claims were not impacted by changes implemented in cobas® Liat® Analyzer Software 3.3, when compared to the current commercially available core software version."
    Continued substantial equivalence to predicate device (K171770)"Equivalent performance of the modified device and the current commercial device has been demonstrated, and analytical or clinical performance has not changed. The modified device is substantially equivalent to the predicate device, as cleared through K171770."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide details on the sample size or data provenance (e.g., country of origin, retrospective/prospective) for the performance evaluation related to the software update. It only discusses the evaluation of the software's impact on the assay. The original predicate device clearance (K171770) would contain this information for the initial performance claims.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the context of this software update evaluation. The ground truth for the test set of the original device was likely established during its initial clearance (K171770), and this document focuses on confirming that the software change doesn't alter that established performance.

    4. Adjudication Method for the Test Set

    This information is not provided in this document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study is not applicable here. This device is an automated in vitro diagnostic test (nucleic acid amplification test), not an imaging device or a system that involves human readers interpreting results in a comparative effectiveness study setting. The device is designed to provide a qualitative result (detection of the toxin B gene).

    6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, implicitly. The evaluation focuses on the performance of the cobas® Cdiff assay with the updated software on the cobas® Liat® System. This is a standalone performance assessment of the automated diagnostic device (algorithm and hardware combined) without human interpretation impacting the primary result itself. The result is "automatically analyzed and...displayed" (page 5).

    7. The Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used for this specific software update evaluation. For the initial clearance of a diagnostic test like this, ground truth for clinical performance studies would typically involve:

    • Culture confirmation (e.g., toxigenic C. difficile culture)
    • An FDA-cleared reference method or composite reference method (CRM) utilizing multiple tests and/or expert clinical judgment.

    This document assumes that the established ground truth methodology for the predicate device remains valid and that the software update does not affect the device's ability to correlate with that ground truth.

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size for the training set for the software (algorithm) itself. Given that this is a software update for an already cleared diagnostic system, "training set" in the context of machine learning (if applicable for some internal functions) is not detailed. The test uses established PCR principles and validated oligonucleotide sequences.

    9. How the Ground Truth for the Training Set was Established

    This information is not provided and is likely not relevant as this is an update to an analytical testing platform rather than a de novo AI/ML algorithm requiring a specific "training set" with ground truth in the traditional sense. The assay's core principles (oligonucleotide sequences, PCR) are described as unchanged from the predicate, which would have undergone its own rigorous analytical and clinical validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172569

    Validate with FDA (Live)

    Device Name
    GenePOC CDiff
    Manufacturer
    Date Cleared
    2017-11-22

    (89 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Franquet Quebec, G1P 4N3 Ca

    Re: K172569

    Trade/Device Name: GenePOC CDiff Regulation Number: 21 CFR 866.3130
    Regulation: 21 CFR 866.3130 - Clostridium difficile toxin gene amplification assay

    • 3.
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GenePOC CDiff assay performed on the revogene instrument is a qualitative in vitro diagnostic test that utilizes automated sample processing and real-time polymerase chain reaction (PCR) to detect the toxin B (tcdB) gene of toxigenic Clostridium difficile( C. difficile) in unformed (liquid or soft) stool specimens obtained from patients suspected of having C. difficile infection (CDI). The GenePOC CDiff assay is intended to aid in the diagnosis of CDI.

    Device Description

    The GenePOC™ CDiff assay is a single-use test for the qualitative detection of the toxin B (tcdB) gene of toxigenic Clostridium difficile (C. difficile) in unformed (liquid or soft) stool specimens. The GenePOC™ CDiff assay kit is comprised of the disposable CDiff microfluidic cartridges (PIE), Disposable Transfer Loops (DTL), Sample Buffer Tubes (SBT), and Disposable Transfer Tools (DTT; pipette). These components are used to suspend the sample, extract, amplify, and detect C. difficile nucleic acid. A Process Control (PrC) is also incorporated into each PIE to verify sample processing and amplification steps. The PrC allows for the verification of potential inhibitor substances as well as microfluidic, instrument or reagent failure. The GenePOC™ CDiff assay is designed to be used on the revogene™. The revogene™ is an instrument that automates sample homogenization, sample dilution, cells lysis, DNA amplification and detection of the amplified PCR products.

    Each GenePOC™ CDiff assay kit provides components for 24 tests. User intervention is required for sample preparation, transferring the stool specimen with the DTL into the SBT, using the DTT to transfer the sample into the PIE, and loading/unloading the PIE into the revogene™ carousel. Each PIE is a completely integrated closed device in which a sample is dispensed and processed through different microfluidic chambers and channels that allow for the sample processing and subsequent real-time PCR steps.

    Upon completion of a run, the results are computed by the revogene™ from measured fluorescent signals and embedded calculation algorithms. The output results include positive, negative, indeterminate, and unresolved. Upon completion of a run, the user removes the used cartridges and disposes of them in normal biological waste. Results may be viewed, printed, transferred, and/or stored by the user.

    AI/ML Overview

    The provided document describes the regulatory submission for the GenePOC CDiff assay, a diagnostic test for Clostridium difficile infection (CDI). The document focuses on the analytical and clinical performance of the device to demonstrate its substantial equivalence to a predicate device.

    Here's a breakdown of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in a formal table. However, the performance characteristics sections imply the criteria for various analytical and clinical aspects. The reported performance is presented in the tables and text within the document.

    Let's infer the acceptance criteria from the reported results, especially concerning agreement percentages and confidence intervals. Typically, for in vitro diagnostics, high sensitivity and specificity are desired, often above certain thresholds (e.g., 85% or 90% for sensitivity, and 95% or more for specificity). For reproducibility, high agreement is also expected (e.g., >95%).

    MetricAcceptance Criterion (Implied)Reported Device Performance (GenePOC CDiff)
    Analytical Performance
    Reproducibility
    Between-Laboratory Reproducibility:
    Low Positive Samples (Overall Agreement)High agreement (e.g., >90%)97.2% (95% CI: 93.6-99.1%)
    Moderate Positive Samples (Overall Agreement)High agreement (e.g., >90%)98.3% (95% CI: 95.2-99.7%)
    True Negative Samples (Overall Agreement)Very high agreement (e.g., >95%)100% (95% CI: 97.5-100%)
    CVs for Ct ValuesLow variability (e.g., <10%)Consistently below 7.54% (Overall)
    Between-Lot Reproducibility:
    Low Positive Samples (Overall Agreement)High agreement (e.g., >90%)95.0% (95% CI: 90.7-97.7%)
    Moderate Positive Samples (Overall Agreement)High agreement (e.g., >90%)98.3% (95% CI: 95.2-99.7%)
    True Negative Samples (Overall Agreement)Very high agreement (e.g., >95%)100% (95% CI: 97.5-100%)
    Within-Laboratory Precision:
    Low Positive Samples (Overall Agreement)High agreement (e.g., >90%)94.4% (95% CI: 89.3-97.6%)
    Moderate Positive Samples (Overall Agreement)High agreement (e.g., >90%)96.5% (95% CI: 92.1-98.9%)
    True Negative Samples (Overall Agreement)Very high agreement (e.g., >95%)100% (95% CI: 96.9-100%)
    Detection Limit (LoD)Lowest concentration for 95%+ detection1,500 CFU/mL of SB for C. difficile strains (95% or greater detection)
    Analytical InclusivityAll strains detected at 2-3xLoDAll 20 toxigenic C. difficile strains detected at 3,750 CFU/mL SB (2-3xLoD)
    Analytical SpecificityMinimal to no false positivesLimited specific false positives at high concentrations for some Clostridium species. No reactivity at lower concentrations. One case of false positive for Enterococcus faecalis at high concentration.
    InterferenceNo interferenceNo interference from 30 tested organisms; interference from Tums and Stomaax at high concentrations; no interference from 5 endogenous agents.
    Carry-Over & Cross-ContaminationAbsence of carry-over/cross-contaminationDemonstrated absence.
    Clinical Performance
    Sensitivity (Fresh Specimens)High sensitivity (e.g., >80%)80.5% (95% CI: 72.0-87.4%)
    Specificity (Fresh Specimens)High specificity (e.g., >95%)97.1% (95% CI: 95.5-98.2%)
    Sensitivity (Frozen Specimens)High sensitivity (e.g., >80%)87.3% (95% CI: 82.1-91.4%)
    Specificity (Frozen Specimens)High specificity (e.g., >95%)97.3% (95% CI: 96.3-98.1%)
    Unresolved Rate (After Repeat Testing)Low rate (e.g., <1%)Fresh: 0.1% (1/798); Frozen: 0.0% (0/1665)
    Indeterminate Rate (After Repeat Testing)Low rate (e.g., <1%)Fresh: 0.0% (0/798); Frozen: 0.1% (1/1665)

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Study Test Set (Patients):

      • Total Samples: 2,461 fully compliant samples (797 fresh + 1,664 frozen).
      • Data Provenance: Prospective multicenter trial at 7 geographically diverse clinical trial sites (US and Canada).
      • Retrospective/Prospective: The study design included testing fresh samples (prospectively collected and tested immediately) and retrospectively collected samples that were stored frozen (prospectively collected but tested after freezing).
    • Analytical Performance Test Sets:

      • Precision/Reproducibility: 1,022 samples (383 low positive, 384 moderate positive, and 255 negative samples) were tested across different studies (Between-Laboratory, Between-Lot, Within-Laboratory). These were spiked stool samples. The provenance of the negative stool pool for preparing these panels is not specified by country, but the panel strains are ATCC (American Type Culture Collection), indicating standardized reference strains.
      • Detection Limit (LoD): Not specified in terms of sample size, but involved testing replicates (multiple runs, instruments, operators) of specific C. difficile strains.
      • Analytical Inclusivity: 20 strains of toxigenic C. difficile from various geographic origins.
      • Analytical Specificity: 58 various analytes (1 yeast, 6 viruses, 50 bacteria, and human DNA).
      • Interference (Non-Target Organisms): Tests involved 30 organisms.
      • Interference (Exogenous/Endogenous Substances): 21 potentially interfering substances.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • This information is not explicitly stated in the provided text.
    • For the clinical study, the reference method (ground truth) was "Combined Direct and Enriched Culture" followed by cytotoxicity testing (CCNA) on isolated C. difficile colonies. While this is a laboratory-based gold standard, the document does not specify the number of laboratory experts/technicians involved in performing and interpreting these reference methods or their specific qualifications. It also doesn't mention expert clinical adjudication for the patient diagnosis itself, as the device is for in vitro diagnostic testing based on lab results.

    4. Adjudication Method for the Test Set

    • For the clinical study, the "ground truth" (reference method) was established as follows:
      • A specimen was considered positive for toxigenic C. difficile if C. difficile was recovered by either direct or enriched culture AND if bacterial isolates tested positive by cytotoxicity testing (CCNA).
      • A specimen was considered negative only if it tested negative by both direct and combined culture (i.e., direct and enriched culture).
    • This is a laboratory-based algorithm for establishing ground truth, not a multi-reader clinical adjudication process. The document does not describe an adjudication method involving multiple human readers for the final clinical diagnosis or the interpretation of the reference method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done.
    • This device is an in vitro diagnostic (IVD) assay that automates sample processing, DNA amplification, and detection of a specific gene. It provides a "positive," "negative," "indeterminate," or "unresolved" result. It is not an AI-based image analysis tool or a system that assists human readers in interpreting complex data where reader improvement could be measured. Therefore, the concept of human readers improving with AI assistance is not applicable to this type of device. The study evaluates the performance of the automated assay itself against a laboratory reference method.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, this was a standalone performance study. The GenePOC CDiff assay operates on the revogene instrument, which automates the entire process from sample loading to result computation. The results are "computed by the revogene™ from measured fluorescent signals and embedded calculation algorithms." The performance metrics (sensitivity, specificity, reproducibility) are reported for the device as a whole, essentially as an algorithm-only (automated instrument) performance, without human "in-the-loop" interpretation of the primary signals. Human intervention is limited to sample preparation and loading/unloading.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • The ground truth used for the clinical study was a laboratory-based reference method: "Combined Direct and Enriched Culture" for C. difficile recovery, followed by cytotoxicity testing (CCNA) on the isolated bacterial strain to confirm toxin production. This is a robust and widely accepted method for confirming toxigenic C. difficile in stool samples.

    8. The Sample Size for the Training Set

    • The document does not mention a separate "training set" for an algorithm. This suggests that the device's algorithms and parameters were likely developed and validated internally by the manufacturer during product development, possibly using a series of development and internal validation studies. The clinical and analytical studies described are primarily for performance demonstration and likely served as a "test set" for regulatory submission, rather than a training set for a machine learning model.
    • For an IVD like this, the "training" (calibration, optimization of thresholds) would typically be done during the engineering and design phases and locked down before clinical validation.

    9. How the Ground Truth for the Training Set Was Established

    • As a dedicated "training set" is not explicitly mentioned or described, the method for establishing its ground truth is also not provided. If an internal training or development set was used, it would have likely relied on similar laboratory reference methods to establish the true positive/negative status of samples.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171770

    Validate with FDA (Live)

    Date Cleared
    2017-09-12

    (90 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name: cobas Cdiff Nucleic acid test for use on the cobas Liat System Regulation Number: 21 CFR 866.3130
    |
    | Classification Name | 21 CFR 866.3130

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is an automated, qualitative in vitro diagnostic test, that utilizes real-time polymerase chain reaction (PCR), for the detection of the toxin B (tcdB) gene of toxigenic Clostridium difficile in unformed (liquid or soft) stool specimens suspected of having C. difficile infection (CDI). The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System is intended for use as an aid in the diagnosis of CDI in humans in conjunction with clinical and epidemiological risk factors.

    Device Description

    The cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System (cobas® Cdiff ) is a rapid, automated in vitro diagnostic test for the qualitative detection of C. difficile DNA in human stool specimens. The cobas® Liat® System is for in vitro diagnostic use. The system is designed to identify and/or measure the presence of genetic material in a biological sample. The system automates all nucleic acid amplification test (NAAT) processes, including reagent preparation, target enrichment, inhibitor removal, nucleic acid extraction, amplification, real-time detection, and result interpretation in a rapid manner. The cobas® Cdiff test uses silica magnetic particle-based nucleic acid extraction and TaqMan probe-based real-time PCR amplification and detection. The cobas® Liat® Analyzer automates and integrates sample purification, nucleic acid amplification and detection of the target sequence in biological samples. Other than adding the sample to the cobas® Cdiff assay tube, no reagent preparation or additional steps are required. The cobas® Liat® System consists of an cobas® Liat® Analyzer with integrated software for running tests and analyzing the results, and a single-use disposable cobas® Cdiff assay tube that holds all of the sample purification and PCR reagents and hosts the sample preparation and PCR processes specific for the Cdiff analyte. The test uses the assay tube as both the sample and reaction vessel. The assay tube comprises flexible tubing containing all required unit dose reagents pre-packed in tube segments, separated by pressure-sensitive seals, in the order of reagent use. During the testing process, multiple sample processing actuators of the analyzer compress the cobas® Cdiff assay tube to selectively release reagents from tube segments, move the sample from one segment to another, and control reaction conditions such as reaction volume, temperature, pressure, and incubation time. Precise control of all these parameters provides optimal conditions for assay reactions, allowing the test to achieve high performance similar to or better than that of currently available molecular assays. The cobas® Liat® Analyzer software controls and coordinated these actions to perform all required assay processes, including sample preparation, nucleic acid extraction, target enrichment, inhibitor removal, nucleic acid elution, and real-time PCR. All assay steps are performed within the closed and self-contained cobas® Cdiff assay tube, thereby eliminating the potential for cross-contamination between samples. The collected data are automatically analyzed and the result is displayed in the assay report on the integrated LCD touch screen of the cobas® Liat® Analyzer.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details based on the provided text:

    Acceptance Criteria and Device Performance for cobas® Cdiff Nucleic acid test for use on the cobas® Liat® System

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" in a separate section with numerical targets like a typical clinical trial report for an AI device. Instead, the performance metrics are presented as the results of the studies conducted. I will infer the acceptance criteria from the reported performance, implying that the reported values met the internal standards set by the manufacturer for substantial equivalence.

    Performance MetricAcceptance Criteria (Inferred from Predicate/Study Goals)Reported Device Performance (cobas® Cdiff Test)
    Analytical Sensitivity (LOD)To detect C. difficile at a concentration to yield a Hit Rate of at least 95%, where all higher concentrations also have a Hit Rate of ≥ 95%.ATCC 43255: 90 CFU/swab R12087 (CD196): 45 CFU/swab
    InclusivityTo detect 100% of tested toxigenic strains at 3xLOD.All 37 toxigenic strains detected with 100% hit rate.
    Analytical Specificity (Cross-reactivity)No cross-reactivity with non-toxigenic C. difficile strains, human genomic DNA, or other common stool microorganisms, and no interference with toxigenic C. difficile detection.No cross-reactivity with 149 tested non-toxigenic Cdiff strains, human epithelial cells, or other microorganisms. No interference with detection of toxigenic Cdiff at 3x LOD.
    InterferenceNo interference from common medications and fecal substances at specified concentrations.Fecal Fat: No interference up to 39% (w/v) Whole blood: No interference up to 100% (v/v) Mucin: No interference up to 50% (w/v) Various other medications and substances tested and confirmed for no interference at specified concentrations (see Table 5).
    Reproducibility (~1xLOD)Overall percent agreement for low positive samples (~1xLOD) to demonstrate consistency across sites, operators, and lots.98.5% (262/266) with a 95% CI of (96.2%, 99.4%)
    Reproducibility (~3xLOD)Overall percent agreement for moderate positive samples (~3xLOD) to demonstrate consistency across sites, operators, and lots.99.3% (268/270) with a 95% CI of (97.3%, 99.8%)
    Reproducibility (Negative)Overall percent agreement for negative samples to demonstrate consistency across sites, operators, and lots.100.0% (262/262) with a 95% CI of (98.6%, 100.0%)
    Clinical Sensitivity (vs. Combined Culture)High sensitivity for detecting toxigenic C. difficile in clinical samples.87.2% (156/179; 95% CI: 81.5% to 91.3%)
    Clinical Specificity (vs. Combined Culture)High specificity for detecting toxigenic C. difficile in clinical samples.98.1% (818/834; 95% CI: 96.9% to 98.8%)
    Clinical PPV (vs. Combined Culture)High Positive Predictive Value.90.7% (156/172; 95% CI: 85.4% to 94.2%)
    Clinical NPV (vs. Combined Culture)High Negative Predictive Value.97.3% (818/841; 95% CI: 95.9% to 98.2%)
    Initial Invalid RateLow initial invalid rate.1.4%
    Initial Failed RateLow initial failed rate.0.2%
    Final Invalid Rate (after retest)Very low final invalid rate.0.1%
    Final Failed Rate (after retest)Zero final failed rate.0%

    2. Sample size used for the test set and the data provenance

    • Analytical Sensitivity (LOD): Not specified as a per-sample-size, but replicates of each 5-member test panel were tested with 2 different lots of assay tubes. Additional replicates were tested for confirmation.
    • Inclusivity: 3 replicates per strain (37 toxigenic strains) were tested.
    • Analytical Specificity (Cross-reactivity): Not specified as a per-sample-size, but 149 non-toxigenic organisms/cells were tested.
    • Interference: Not specified as a per-sample-size, but 38 commonly used medications, fecal fat, whole blood, and mucin were tested.
    • Reproducibility: 818 total tests were performed, with 798 valid for analysis across 3 panel members (Negative, ~1xLOD, ~3xLOD). 3 replicates per panel member, run on 5 nonconsecutive days by 2 operators at 3 sites for 3 reagent lots.
    • Clinical Performance: 1,013 evaluable fresh remnant stool specimens were prospectively collected.
      • Data Provenance:
        • Country of Origin: Not explicitly stated, but "Nine (9) study sites from geographically diverse locations participated in this study." This suggests a multi-center study, likely within a single country (such as the US, given the FDA submission) or potentially across a few countries, but a specific list is not provided.
        • Retrospective/Prospective: The clinical performance study was prospective. "Evaluable fresh remnant specimens were prospectively collected from 1,013 patients."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Analytical (LOD, Inclusivity, Specificity, Interference): Not applicable for these studies as the "ground truth" is based on engineered samples (known concentrations of specific strains, or presence/absence of interfering substances).
    • Reproducibility: Not applicable, as this assessed the consistency of the device's own results, not against an external ground truth.
    • Clinical Performance:
      • Number of experts: Not explicitly stated.
      • Qualifications of experts: Not explicitly stated. The "ground truth" was established by "combined results of direct and enriched toxigenic culture using leftover, deidentified, unformed stool samples" performed at a "reference laboratory." The document mentions "Nineteen (19) operators distributed across these study sites performed testing with the cobas® Cdiff test," but these are operators, not necessarily the adjudicating experts for the ground truth. For microbiological culture, laboratory personnel with specific training in microbiology would establish the ground truth.

    4. Adjudication method for the test set

    • Analytical, Reproducibility: Not applicable.
    • Clinical Performance:
      • Initial Ground Truth: "Combined results of direct and enriched toxigenic culture" served as the primary comparative method.
      • Discrepant Analysis: "All evaluable discordant specimens and an equal number of randomly selected and representative concordant samples were used for discrepant testing by an FDA-cleared comparator NAAT. Testing was performed at an external lab that was preselected and qualified by RMS." This implies a form of 2-method consensus with a third method for discrepancies. If direct and enrichment culture disagreed, or if the cobas test disagreed with the combined culture, a third (comparator NAAT) test was used for "discrepant analysis." However, the final clinical performance tables still directly compare the cobas test to the combined culture results, with the discrepant analysis results discussed qualitatively ("Of the 23 specimens with false-negative cobas® Cdiff test results relative to combined direct culture and enrichment culture, 19 were negative. 3 were positive and 1 was invalid by the second NAAT method.").

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    • No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) test, not an imaging AI device that relies on human-in-the-loop interpretation. The clinical study compares the device's output to laboratory culture results, not to human readers' interpretations with or without AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, the performance presented is purely standalone (algorithm/device only). The cobas® Cdiff Nucleic acid test is an automated, qualitative in vitro diagnostic test. The device provides "result interpretation" directly on its integrated LCD touch screen (page 5). The clinical performance data reflects the device's ability to detect the target gene independently, compared against culture methods.

    7. The type of ground truth used

    • Analytical Studies (LOD, Inclusivity, Specificity, Interference): Lab-prepared spiked samples with known concentrations/compositions.
    • Clinical Performance: Microbiological culture results. Specifically, "combined results of direct and enriched toxigenic culture." A "second NAAT method" was used for discrepant analysis.

    8. The sample size for the training set

    • Not applicable / Not explicitly stated. This is a molecular diagnostic test for Clostridium difficile, not a machine learning or AI algorithm in the typical sense that requires explicit "training data" for a model to learn from. The assay uses predefined primers and probes for real-time PCR detection (page 4). While the assay design and optimization would involve extensive internal development and testing, this is not characterized as a "training set" in the context of typical AI algorithm development.

    9. How the ground truth for the training set was established

    • Not applicable / Not explicitly stated. As mentioned above, this is not an AI/ML device that uses a "training set" in the conventional sense. The "ground truth" for developing the test was based on well-characterized isolates of C. difficile and other microorganisms to design and validate the specificity and sensitivity of the primers and probes.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171441

    Validate with FDA (Live)

    Date Cleared
    2017-07-21

    (66 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    ARIES C. difficile Assay Kit (24 cassettes), ARIES Stool Resuspension Kit Regulation Number: 21 CFR 866.3130
    -------------------------------|----------------------|
    | OZN | II | 21 CFR 866.3130

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARIES® C. difficile Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of toxigenic Clostridium difficile (C. difficile) nucleic acid in unpreserved, unformed (liquid or soft) stool specimens obtained from patients suspected of having Clostridium difficile infection (CDI). The test targets the C. difficile toxin A gene (tcdA) and toxin B gene (tcdB) and is indicated for use as an aid in the diagnosis of C. difficile infection (CDI). The ARIES® C. difficile Assay is indicated for use with ARIES® Systems.

    Device Description

    The ARIES® C. difficile Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system which consists of the ARIES® System or the ARIES® M1 System with their included ARIES® Software, a stool resuspension kit, an assay-specific cassette, and an assay-specific protocol file. The ARIES® C. difficile Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control (SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES® C. difficile Assay cassette directly detects toxigenic Clostridium difficile (C. difficile) from unformed stool specimens obtained from patients suspected of having Clostridium difficile infection. Specifically, the ARIES® C. difficile Assay cassette detects the C. difficile toxin A gene (tcdA) and toxin B gene (tcdB) and a DNA Sample Processing Control. Unpreserved raw stool is processed using the ARIES® Stool Resuspension Kit. The ARIES® Stool Resuspension Kit includes a flocked swab, a tube containing preprocessing beads, and Stool Resuspension Buffer. The preprocessing method involves transfer of a swab of stool specimen into a tube containing preprocessing beads and Stool Resuspension Buffer. The swab is mixed in the tube by vortexing and then centrifuged. Finally, the preprocessed sample is added to the ARIES® C. difficile Assay cassette. The specimen is lysed and nucleic acid is extracted using an ARIES® instrument. An extractable sample processing control (SPC) target is also present in the ARIES® C. difficile Assay cassette and is processed with the specimen. The SPC controls for recovery of extracted nucleic acid, for inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument. The Tm value of the SPC is used as a reference for determining the Tm of the tcdA and tcdB targets (if present). The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES® C. difficile Assay lyophilized PCR reagents in the PCR tube that contain primer pairs specific to tcdA, tcdB, and the SPC sequence. Each of the primer pairs is labeled with a distinct fluorophore and detected in distinct channels of an ARIES® System. PCR amplification is performed and assay fluorescence is monitored. Incorporation of a quencher-labeled nucleotide results in a decrease in fluorescence for the associated primer pair. Following amplification, the reaction is slowly heated to separate the fluorescent-labeled strand from the quencher-labeled strand, a process that results in an increase in the fluorescence signal. The reaction fluorescence is measured during this process and the temperature at which the change in fluorescence is the maximum Tm of the amplicon. The strands of the amplicons will separate at a specific melting temperature (Tm) and an increase in fluorescence is observed. The instrument fluorescence output is analyzed and test results are determined using the ARIES® System software and the ARIES® C. difficile Assay protocol and run files. ARIES® C. difficile Assay results may be reported from the ARIES Software or from the optional SYNCT® Software.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for the ARIES® C. difficile Assay:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary does not explicitly state pre-defined acceptance criteria values for clinical performance as a single table. Instead, it presents the assay's performance metrics against a reference method and implicitly indicates that these results were considered "acceptable" for substantial equivalence. Based on the clinical performance section, we can infer the key performance metrics reported.

    MetricAcceptance Criteria (Inferred from reported performance deemed acceptable)Reported Device Performance (ARIES® C. difficile Assay)
    Positive Percent Agreement (PPA) vs. Direct Toxigenic CultureHigh (e.g., typically >90%)98.1% (Lower Bound 95% CI: 93.3%)
    Negative Percent Agreement (NPA) vs. Direct Toxigenic CultureHigh (e.g., typically >90%)92.6% (Lower Bound 95% CI: 90.6%)
    Sensitivity vs. Direct & Enriched Toxigenic CultureHigh (e.g., typically >90%)90.5% (Lower Bound 95% CI: 84.6%)
    Specificity vs. Direct & Enriched Toxigenic CultureHigh (e.g., typically >90%)95.8% (Lower Bound 95% CI: 94.2%)
    Overall Invalid RateLow0.5% (after re-test)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:

      • Clinical Study: 984 unique specimens were included in the data analysis. After exclusions and re-tests, 979 eligible specimens were used for performance calculations.
      • Analytical Precision/Reproducibility:
        • Within-Laboratory: 252 samples (36 for each of 7 panel members).
        • Reproducibility (3 sites x 5 days x 6 panel members x 3 replicates): 270 samples (90 for each of 7 panel members, assuming negative also had 90 replicates).
      • Limit of Detection (LoD): Not specified precisely, but implied to be multiple replicates at various dilutions.
      • Analytical Reactivity: 15 distinct strains of toxigenic C. difficile (in addition to those in LoD).
      • Analytical Specificity (Cross-Reactivity/Interference): 61 microorganisms/viruses + human DNA, each tested with three replicates of two C. difficile strains (BAA-1870 and BAA-1871) and three negative replicates.
      • Carry-Over/Cross-Contamination: 30 high positive and 30 negative samples.
    • Data Provenance:

      • Clinical Study: Prospective. The specimens were collected from 31-October-2016 to 21-February-2017 at four geographically distinct clinical sites within the United States. The specimens consisted of excess leftover de-identified, unpreserved, unformed stool specimens from patients suspected of having Clostridium difficile infection (CDI).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    The document does not explicitly state the "number of experts" or their specific "qualifications" for establishing the ground truth using the reference method.

    • Ground Truth Method: The reference method for the clinical study was "direct and enriched toxigenic culture." This is a laboratory-based method.
    • Location of Ground Truth Establishment: "Reference method testing was performed at a centralized testing facility." It is implied that qualified laboratory personnel perform these culture tests, but specific number or qualifications are not provided.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (like 2+1, 3+1, none) for discordant results between the ARIES® C. difficile Assay and the reference method.

    However, for discordant cases, further characterization was performed:

    • For specimens negative by direct toxigenic culture but positive by ARIES® C. difficile Assay (False Positives based on direct culture), enriched toxigenic culture was used, and for some, bi-directional sequencing with analytically validated primers was performed.
    • For specimens positive by direct toxigenic culture but negative by ARIES® C. difficile Assay (False Negatives based on direct culture), bi-directional sequencing with analytically validated primers was used.
    • Similar follow-up testing (bi-directional sequencing) was done for discords when comparing to "Direct and Enriched Toxigenic Culture."

    This suggests a form of further characterization for discordant results rather than an adjudication panel of experts re-reading primary data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is an in vitro diagnostic (IVD) assay, not an AI imaging or diagnostic algorithm designed to assist human readers. Its performance is evaluated stand-alone against a reference laboratory method.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the clinical performance study evaluated the ARIES® C. difficile Assay as a standalone algorithm/device. The assay generates a qualitative result (POSITIVE, NEGATIVE, or INVALID) based on its internal processing and software analysis, without human interpretation of raw data for diagnosis. The ARIES® System software automatically determines and reports results.

    7. The Type of Ground Truth Used

    The primary ground truth used for the clinical performance study was direct and enriched toxigenic culture.

    • Direct Toxigenic Culture: This involves culturing the C. difficile from the stool sample directly to identify toxigenic strains.
    • Enriched Toxigenic Culture: This involves an enrichment step before culturing to increase the yield of C. difficile, potentially improving sensitivity.

    For discordant results, further analysis using bi-directional sequencing with analytically validated primers was employed.

    8. The Sample Size for the Training Set

    The document does not specify a distinct training set sample size. For IVD submissions like this one, it's common that assay parameters (like Ct cut-offs and Tm windows) are established during "internal optimization studies" and "internal verification studies." The document explicitly states:

    • "Initial Assay Protocol File parameters were set during internal optimization studies."
    • "The final Assay Protocol File parameters were then established during internal verification studies."

    The sizes of these internal optimization and verification studies are not provided but would have involved various analytical samples (LoD, inclusivity, exclusivity, etc.) rather than a dedicated "training set" of clinical samples in the same way an AI model might be trained. The clinical trial data (979 samples) served as a test set for validation, not for training the assay's parameters.

    9. How the Ground Truth for the Training Set Was Established

    Since no distinct "training set" of clinical samples is explicitly described for algorithm development in the AI sense, the ground truth establishment for the internal optimization and verification studies likely involved:

    • Carefully characterized reference strains of C. difficile (toxigenic and non-toxigenic).
    • Spiked stool matrixes with known concentrations of C. difficile to determine detection limits and reactivity.
    • Potentially, a smaller set of pre-characterized clinical samples to refine initial assay parameters.

    Again, these studies would rely on established microbiological and molecular biology techniques (culture, PCR, sequencing) to define the presence/absence and characteristics of C. difficile in the samples used for internal development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170491

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-11

    (83 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Ohio 45701

    Re: K170491

    Trade/Device Name: Solana C. difficile Assay Regulation Number: 21 CFR 866.3130
    genes fromstool specimens ofsymptomaticpatients | Class II (SpecialControls) | 21 CFR 866.3130

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solana® C. difficile Assay is an in vitro diagnostic test for the direct, qualitative detection of the Clostridium difficile Toxin A gene (tcdA) in unformed stool specients of patients suspected of having Clostridium difficile-infection (CDI). The Solana C. difficile Assay is intended for use as an aid in diagnosis of CDI. The assay utilizes helicase-dependent amplification (HDA) for the amplification of a highly conserved fragment of the Toxin A gene sequence. The Solana C. difficile Assay is intended for use only with the Solana® instrument.

    Device Description

    The Solana C. difficile Assay combines simple sample processing and Helicase-Dependent Amplification (HDA) performed in Solana for the detection of toxigenic Clostridium difficile directly from CDI-suspected diarrheal specimens.

    A small amount of specimen is transferred to a Lysis Tube using a swab. The Lysis Tube is then subjected to heat-treatment at 95°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to a Reaction Tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of target sequence. In Solana, the target sequence is amplified by specific primers and detected by a specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC is amplified by the target-specific primers and detected by a PRC specific fluorescence probe.

    The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. Upon annealing to target or PRC amplicons, the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then report the test results to the user on its display screen, and it can print out the results via a printer.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Solana C. difficile Assay, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" as a separate section with predefined thresholds. However, the "Performance in Comparison to Enhanced toxigenic bacterial culture" section for sensitivity and specificity can be considered as the primary clinical performance evaluation against a recognized gold standard. The "Performance in Comparison to FDA-cleared molecular device" provides comparative performance.

    Given this, I will present the clinical performance metrics as the reported device performance against what would typically be considered implicit acceptance targets for diagnostic devices of this type.

    Table 1: Reported Device Performance Against Ground Truth (Enhanced Toxigenic Culture)

    Performance MetricReported Performance (95% CI)
    Sensitivity93.0% (86.9% - 96.4%)
    Specificity99.2% (98.2% - 99.6%)

    Table 2: Reported Device Performance Compared to an FDA-Cleared Molecular Device

    Performance MetricReported Performance (95% CI)
    Positive Percent Agreement97.0% (91.59% - 99.0%)
    Negative Percent Agreement97.9% (96.6% - 98.7%)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 854 raw specimens were collected. After removing 2 invalid specimens, 852 specimens were used for analysis.
    • Data Provenance:
      • Country of Origin: United States. The study was conducted at "three distinct geographical sites across the United States."
      • Retrospective or Prospective: Prospective. The study documents state, "Performance characteristics of the Solana C. difficile Assay were established during a prospective study conducted November 2016 to February 2017."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not specify the number of experts or their qualifications for establishing the ground truth (enhanced toxigenic culture). It describes the methodology for the enhanced toxigenic culture, which is a laboratory-based process, rather than an expert human interpretation.

    4. Adjudication Method for the Test Set

    The document does not explicitly detail an adjudication method beyond the methodology for the enhanced toxigenic culture. It mentions, "Three (3) of the six (6) specimens were positive for C. difficile toxin gene DNA by an alternate FDA cleared molecular device, three (3) were negative." and "Six (6) of eight (8) specimens were found positive for C. difficile toxin gene DNA by an alternate FDA cleared molecular device., and two (2) were negative", which suggests some level of discordant analysis against a second method, but no formal expert adjudication process is described for the ground truth establishment itself.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic test that provides a qualitative result (positive/negative) directly from the instrument. It is not designed to be interpreted by human readers, nor does it assist human readers in making a diagnosis. Therefore, the concept of human readers improving with or without AI assistance does not apply here.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The entire clinical performance evaluation described compares the "Solana C. difficile Assay" (the algorithm/device) directly against the enhanced toxigenic culture (ground truth) and an FDA-cleared molecular device. The results reported are for the device's performance without human intervention in the interpretation of the C. difficile Toxin A gene (tcdA) detection.

    7. Type of Ground Truth Used

    The primary ground truth used for evaluating the Solana C. difficile Assay was enhanced toxigenic bacterial culture.

    An "alternate FDA cleared molecular device" was also used for comparison in cases of discordance, but the initial sensitivity and specificity calculations used the enhanced toxigenic culture as the reference.

    8. Sample Size for the Training Set

    The document does not provide information regarding a specific training set size or methodology for the Solana C. difficile Assay. The information provided focuses on the analytical and clinical validation of the final product. For an IVD like this, the "training" may happen during internal development and optimization phases, which are not typically detailed in 510(k) summaries.

    9. How the Ground Truth for the Training Set Was Established

    As no specific training set or its sample size is mentioned, the method for establishing ground truth for a training set (if one existed separately from analytical validation) is not described in the provided document. The ground truth for the clinical test set was established by enhanced toxigenic bacterial culture.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163085

    Validate with FDA (Live)

    Date Cleared
    2017-02-14

    (103 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Simplexa C. difficile Direct: Simplexa C. difficile Positive Control Pack Regulation Number: 21 CFR 866.3130
    |
    | Classification | 21 CFR 866.3130

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Simplexa C. difficile Direct: The Focus Diagnostics Simplexa™ C. difficile Direct assay is intended for use on the Integrated Cycler instrument for the detection of Clostridium difficile toxin B gene (tcdB) present in liquid or unformed stool samples from individuals suspected of C. difficile infection (CDI). This test aids in the diagnosis of illness resulting from CDI. The assay is for professional and prescription use only.

    Simplexa C. difficile Positive Control Pack: The Simplexa™ C. difficile Positive Control Pack is intended to be used as a control with the Simplexa™ C. difficile Direct kit. This control is not intended for use with other assays or systems.

    Device Description

    The Simplexa™ C. difficile Direct assay system is a real-time PCR system that enables the direct amplification and detection of toxigenic C. difficile DNA from unprocessed liquid or unformed stool specimens that have not undergone nucleic acid extraction. The system consists of the Simplexa™ C. difficile Direct assay, the Integrated Cycler (with Integrated Cycler Studio Software), the Direct Amplification Disc and associated accessories. In the Simplexa™ C. difficile Direct assay, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify C. difficile and DNA internal control targets. The assay targets a sequence which is in a well conserved region of C. difficile toxin B gene (todB). A DNA internal control is used to detect PCR failure and/or inhibition

    AI/ML Overview

    Acceptance Criteria and Device Performance for Simplexa C. difficile Direct

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes the performance of the Simplexa C. difficile Direct assay in comparison to various gold standards and other NAATs. The acceptance criteria are implicit in the presented tables.

    Acceptance Criteria for Simplexa C. difficile Direct (versus Combined Culture + Toxin Assay):

    MetricAcceptance Criteria (Implied)Reported Device Performance95% Confidence Interval (CI)
    SensitivityNot explicitly stated85.9% (336/391)82.1% to 89.0%
    SpecificityNot explicitly stated95.1% (1849/1945)94.0% to 95.9%
    PPVNot explicitly stated77.8% (336/432)73.6% to 81.4%
    NPVNot explicitly stated97.1% (1849/1904)96.3% to 97.8%

    Additional Performance (versus other FDA-cleared NAATs for reference):

    ComparisonMetricReported Device Performance95% Confidence Interval (CI)
    Simplexa vs NAAT-1Positive Agreement93.4% (114/122)87.6% to 96.6%
    Negative Agreement96.6% (683/707)95.0% to 97.7%
    Simplexa vs NAAT-2Positive Agreement93.9% (138/147)88.8% to 96.7%
    Negative Agreement94.0% (591/629)91.8% to 95.6%
    Simplexa vs NAAT-3Positive Agreement84.8% (112/132)77.8% to 90.0%
    Negative Agreement99.2% (584/589)98.0% to 99.6%

    2. Sample Size Used for the Test Set and Data Provenance

    The primary clinical performance study (method comparison) used 2351 samples prospectively collected.

    • Data Provenance: Samples were prospectively collected from five (5) geographically diverse sites in the USA between December 3, 2015, and June 10, 2016.
    • Evaluable Samples: 2330 samples were evaluable on Simplexa C. difficile Direct & the Direct Culture method, and 2336 were evaluable on Simplexa C. difficile Direct & the Combination of Direct/Enriched Culture methods.
    • Invalid Rate: The initial invalid rate was 0.64% (15/2351), reduced to a final invalid rate of 0.13% (3/2351) after repeat testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the "number of experts" or their "qualifications" involved in establishing the ground truth for the clinical sample test set. However, the ground truth was based on Direct Culture + Toxin Assay and Combined Culture (Direct & Enriched) + Toxin Assay, with confirmatory testing using FDA cleared nucleic acid amplification test (NAAT) results for discrepant analysis. This implies that the ground truth was established by laboratory methods and not by individual expert interpretation.

    4. Adjudication Method for the Test Set

    For discrepant samples in the clinical study, discrepant analysis was performed using FDA cleared nucleic acid amplification test (NAAT) results provided by the sites.

    • For the "Simplexa C. difficile Direct versus Direct Toxigenic Culture Method" table:
      • 116/163 discrepant Simplexa-positive/Culture-negative samples were positive, and 46/163 were indeterminate for C. difficile Toxin B gene DNA when tested using an FDA cleared molecular test.
      • 8/12 discrepant Simplexa-negative/Culture-positive samples were negative, and 4/12 were positive for C. difficile Toxin B gene DNA when tested using an FDA cleared molecular test.
    • For the "Simplexa C. difficile Direct versus Combined Direct and Enriched Toxigenic Culture Methods" table:
      • 59/96 discrepant Simplexa-positive/Culture-negative samples were positive, and 37/96 were negative for C. difficile Toxin B gene DNA when tested using an FDA cleared molecular test.
      • 43/55 discrepant Simplexa-negative/Culture-positive samples were negative for C. difficile Toxin B gene DNA when tested using an FDA cleared molecular test.

    This indicates an adjudication process involving a third, independent, FDA-cleared NAAT for resolving discrepancies between the investigational device and the culture methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not explicitly described in the provided text. The study focuses on the performance of the analytical device rather than human reader interpretation with or without AI assistance.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-loop Performance) Was Done

    Yes, the studies presented are standalone performance evaluations of the Simplexa C. difficile Direct assay. The assay is an automated real-time PCR system designed for direct detection of C. difficile toxin B gene, meaning its performance is measured independently of human interpretation of its results, although trained lab personnel would operate the instrument.

    7. The Type of Ground Truth Used

    The ground truth for the clinical performance study was established using:

    • Direct Culture + Toxin Assay
    • Combined Culture (Direct & Enriched) + Toxin Assay
    • FDA cleared molecular tests (NAATs) for discrepant analysis.

    This represents a composite reference standard based on microbiological culture techniques and molecular detection for toxigenic C. difficile, with additional molecular confirmation for discordant results.

    8. The Sample Size for the Training Set

    The document does not specify a separate "training set" sample size. The studies described are performance evaluations of the device, implying that the device's development and internal optimization would have occurred prior to these validation studies. Therefore, information regarding a training set is not provided as this is a diagnostic assay, not a machine learning algorithm that requires a separate training and test set in the same way.

    9. How the Ground Truth for the Training Set Was Established

    As no specific training set is mentioned for the device (which is a molecular diagnostic assay), the method for establishing ground truth for such a set is not applicable or described in this document. The focus is on the analytical and clinical validation of the completed product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142422

    Validate with FDA (Live)

    Device Name
    cobas Cdiff Test
    Date Cleared
    2015-05-20

    (265 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    PLEASANTON CA 94588-2722

    Re: K142422

    Trade/Device Name: cobas Cdiff Test Regulation Number: 21 CFR 866.3130
    |
    | Classification Name | 21 CFR 866.3130

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas® Cdiff Test on the cobas® 4800 system is an automated, qualitative in vitro diagnostic test, that utilizes realtime polymerase chain reaction (PCR), for the direct detection of the toxin B (tcdB) gene of toxigenic Clostridium difficile (C. difficile) in unformed (liquid or soft) stool specimens obtained from patients suspected of having C. difficile infection (CDI). The cobas® Cdiff Test is intended for use as an aid in the diagnosis of CDI in humans in conjunction with clinical and epidemiological risk factors.

    Device Description

    The Roche Molecular Systems (RMS) cobas® Cdiff Test utilizes real-time polymerase chain reaction (PCR) for the detection of Clostridium difficile (C. difficile) DNA from unformed (liquid or soft) stool specimens to aid in the diagnosis of Clostridium difficile infections in humans. The cobas Cdiff Test contains two major processes: (1) automated sample preparation to extract nucleic acids from the unformed stool specimens; (2) PCR amplification of target DNA sequences using C. difficile specific primers, and real-time detection of cleaved fluorescentlabeled C. difficile specific oligonucleotide detection probes. An Internal Control (IC), containing unrelated randomized DNA sequence, is added to all samples prior to automated sample preparation and is amplified and detected simultaneously with each sample to monitor the entire process. The cobas® Cdiff Test utilizes six reagent kits and is used with the cobas® 4800 System, which includes the cobas x 480 Instrument for sample preparation and the cobas z 480 Analyzer for amplification and detection, controlled by the cobas® 4800 System Software.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the cobas® Cdiff Test, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria in a dedicated section. However, performance targets are implicitly demonstrated and met by the reported study results. The clinical performance section compares the cobas® Cdiff Test to a "combined direct and enrichment culture" which serves as the reference standard. The reported performance metrics are:

    Performance MetricAcceptance Criteria (Implied by Study Results)Reported Device Performance (95% CI)
    SensitivityHigh (e.g., >85-90%)92.9% (131/141; 87.4% to 96.1%)
    SpecificityHigh (e.g., >95%)98.7% (534/541; 97.4% to 99.4%)
    Positive Predictive Value (PPV)High (e.g., >90%)94.9% (89.9% to 97.5%)
    Negative Predictive Value (NPV)High (e.g., >95%)98.2% (96.6% to 99.0%)
    Analytical Sensitivity (LOD)Low concentration for 95% positive rate225 CFU/swab (highest LOD among 7 strains)
    Inclusivity≥ 95% positive rate at specified concentrationAchieved for 28 additional toxigenic strains
    Analytical Specificity (Cross-reactivity)No cross-reactivity with common organisms0% (expected negative results for 131 organisms)
    Cross-ContaminationVery low rate0.24% (1/423 negative samples became positive)
    Carry-over Contamination0%0.00% (0/282 negative samples became positive)
    Reproducibility (Positive Agreement)High at 1xLOD and 3xLOD100.0%
    Reproducibility (Negative Agreement)High for negative samples100.0%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Performance (Test Set): 682 specimens were included in the statistical analysis for comparison with combined direct and enrichment culture. 683 subjects were initially collected from.
    • Data Provenance: Prospective, multi-site investigation across five geographically diverse sites in the US. The samples were "leftover, de-identified, unformed stool samples from subjects suspected of having CDI."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth. However, the ground truth for the clinical study was based on toxigenic culture, which involved laboratory procedures conducted at a "single reference laboratory." This suggests trained laboratory personnel, but no specific professional qualifications (e.g., "radiologist with 10 years of experience") are provided.

    4. Adjudication Method for the Test Set

    The primary ground truth for the clinical performance was established by toxigenic culture (direct and enrichment culture followed by cytotoxicity testing). Discrepant analysis was performed on all samples with discordant results and a random subset of concordant results between the cobas® Cdiff Test and toxigenic culture using a second FDA-cleared nucleic acid amplification test (NAAT). This indicates an adjudication method involving a third, independent test to resolve discrepancies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. This study evaluates the impact of AI on human reader performance. The cobas® Cdiff Test is an automated diagnostic test that doesn't involve human interpretation of images or complex data in a way that an MRMC study would be applicable.

    6. If a Standalone Study Was Done

    Yes, a standalone study was done. The entire clinical performance evaluation directly assesses the performance of the cobas® Cdiff Test (the algorithm/device) against the established ground truth (toxigenic culture) without human intervention in the interpretation of the device's results.

    7. The Type of Ground Truth Used

    The ground truth used for the clinical performance evaluation was expert consensus informed by culture and cytotoxicity testing. Specifically, a specimen was considered positive for toxigenic C. difficile if C. difficile was recovered from stool by either direct or enriched toxigenic culture and if isolates recovered tested positive by cytotoxicity testing. Specimens were negative only if they tested negative by both direct, repeat direct, and enrichment culture. Additionally, a second FDA-cleared NAAT was used for discrepant analysis, further supporting the ground truth adjudication.

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size for the training set. This K142422 submission focuses on the performance of a developed device, not on the details of its developmental training data.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established. The submission focuses on the clinical validation of the device, not its initial development or training process.