    510k Summary Text (Full-text Search) :

    Mini

    Primary Requlation Number for Device Classification: 21 CFR 866.3981

    Regulation Number: 21 CFR 866.2680
    Nucleic Acid Test Including SARS-CoV-2 And Other Microbial Agents

    Additional Regulation Numbers: 21 CFR 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel Mini is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis (Sore Throat menu).

    The following analytes are identified and differentiated using the SPOTFIRE R/ST Panel Mini:

    Respiratory Menu:
    Viruses
    Coronavirus SARS-CoV-2
    Human rhinovirus
    Influenza A virus
    Influenza B virus
    Respiratory syncytial virus

    Sore Throat Menu:
    Viruses
    Human rhinovirus
    Influenza A virus
    Influenza B virus
    Respiratory syncytial virus
    Bacteria
    Streptococcus pyogenes (group A Strep)

    Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS/TS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngitis is indicative of the presence of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS or TS specimen. Positive results do not rule out co-infection with other organisms. The agent(s) detected by the SPOTFIRE R/ST Panel Mini may not be the definite cause of disease.

    Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis.

    Device Description

    The SPOTFIRE R/ST Panel Mini simultaneously identifies 5 different respiratory viral pathogens in nasopharyngeal swabs (NPS) or 5 different viral and bacterial pharyngitis pathogens in throat swabs (TS) from individuals with signs and symptoms of respiratory tract infections or pharyngitis, respectively, (see Table 1) The SPOTFIRE R/ST Panel Mini is compatible with the BIOFIRE® System, a polymerase chain reaction (PCR)-based in vitro diagnostic system for infectious disease testing. The BIOFIRE System Sottware executes the SPOTFIRE R/ST Panel Mini test and interprets and reports the test results. The SPOTFIRE R/ST Panel Mini was designed to be used in CLIA-waived environments.

    A test is initiated by loading Hydration Solution injection solution injection port of the SPOTFIRE R/ST Panel Mini pouch and NPS or TS specimen, mixed with the provided Sample injection port of the SPOTFIRE R/ST Panel Mini pouch and placing it in the SPOTFIRE System. The reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the SPOTFIRE System Software guides the user through the steps of placing the pouch into the instrument, scanning the sample identification, and initiating the run.

    The SPOTFIRE System contains coordinated systems of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

    Nucleic acid extraction occurs within the SPOTFIRE R/ST Panel Mini pouch using mechanical Ivsis followed by purfication using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the SPOTFIRE System performs a nested multiplex PCR that is executed in two stage. During the first stage, the SPOTFIRE System performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent doublestranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

    The SPOTFIRE System Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the SPOTFIRE R/ST Panel Mini.

    AI/ML Overview

    This document describes the BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel Mini, a multiplex PCR test.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides extensive analytical performance data rather than a direct comparison of acceptance criteria to reported clinical performance metrics (like PPA and NPA). However, the "Summary of Performance Data" for clinical studies does present sensitivity/PPA and specificity/NPA, which can be interpreted as the reported device performance against implied clinical acceptance criteria.

    Clinical Performance Summary (NPS Specimens - Respiratory Menu)

    AnalytePerformance Metric (Prospective)%95% CI
    Coronavirus SARS-CoV-2 (PPA)71/7397.390.5-99.2%
    Coronavirus SARS-CoV-2 (NPA)1031/103799.498.7-99.7%
    Human rhinovirus (PPA)345/34899.197.5-99.7%
    Human rhinovirus (NPA)695/76790.688.3-92.5%
    Influenza A virus (PPA)0/0 (no positive cases identified)--
    Influenza A virus (NPA)1115/111510099.7-100%
    Influenza B virus (PPA)0/0 (no positive cases identified)--
    Influenza B virus (NPA)1110/111010099.7-100%
    Respiratory syncytial virus (PPA)26/2796.381.7-99.3%
    Respiratory syncytial virus (NPA)1086/108899.899.3-100%

    Clinical Performance Summary (TS Specimens - Sore Throat Menu)

    AnalytePerformance Metric (Prospective)%95% CI
    Human rhinovirus (Sensitivity/PPA)202/21394.891.0-97.1%
    Human rhinovirus (Specificity/NPA)619/66293.591.4-95.1%
    Influenza A virus (Sensitivity/PPA)35/3510090.1-100%
    Influenza A virus (Specificity/NPA)840/84010099.5-100%
    Influenza B virus (Sensitivity/PPA)4/410051.0-100%
    Influenza B virus (Specificity/NPA)872/87210099.6-100%
    Respiratory syncytial virus (Sensitivity/PPA)21/2487.569.0-95.7%
    Respiratory syncytial virus (Specificity/NPA)849/85199.899.1-99.9%
    Streptococcus pyogenes (PPA - PCR)209/21796.392.9-98.1%
    Streptococcus pyogenes (NPA - PCR)654/66099.198.0-99.6%
    Streptococcus pyogenes (Sensitivity - Culture)174/17798.395.1-99.4%
    Streptococcus pyogenes (Specificity - Culture)654/69294.592.6-96.0%

    Analytical Acceptance Criteria and Results for key studies:

    StudyAcceptance CriteriaReported Device Performance (Results)
    Sample Storage and Handling100% expected positive results in all samples tested for each organism. Crossing point (Cp) values evaluated and trended across conditions to assess analyte stability.Positive results were observed in 100% of all TSa samples tested at all conditions evaluated for all SPOTFIRE R/ST Panel Mini analytes.
    Limit of Detection (LoD)LoD confirmed when positive results were reported in at least 95% (≥19/20) of replicates tested at 1x LoD, and fewer than 95% (≤18/20) of replicates tested at 0.1x LoD. Equivalent detection in single and multi-analyte samples based on concordance of positive/negative results.The LoD concentrations for the SPOTFIRE R/ST Panel Mini analytes were confirmed in viable or infectious units and/or nucleic acid copies/mL. The panel accurately detected viruses and bacteria in samples contrived in either VTM or Amies media containing one or multiple organisms.
    Analytical Reactivity (Inclusivity)Assay reactivity of each isolate confirmed if positive results were reported for the appropriate analyte in 3/3 or 4/5 replicates tested within 10x LoD. If fewer than 4/5 replicates, additional testing at 100x LoD or higher. Isolates with reactivity limitations noted in product literature.Analytical reactivity testing demonstrated that the SPOTFIRE R/ST Panel Mini can detect and accurately report results for a diverse collection of isolates from a variety of strains, serotypes, and genotypes with few limitations. (Limitations noted in conclusion include rare S. pyogenes strains not detected).
    Analytical Specificity (Exclusivity)On-panel organisms expected positive for target analyte and negative for others. Off-panel organisms expected negative for all panel analytes, unless otherwise indicated.Three cross-reactivities were identified by empirical and/or in silico evaluations: SARS-CoV-2 with closely related sarbecoviruses, some Bordetella species with Human Rhinovirus (at high concentration), and some bovine/canine picornaviruses with Human Rhinovirus. These limitations are noted in the device labeling.
    InterferencePrimary results evaluated: pass/fail/invalid for internal controls, and analyte positive/negative results. If unexpected result/control failure for one replicate, retested in two additional pouches.Accurate results for the SPOTFIRE R/ST Panel Mini were reported in the presence of a variety of potentially interfering substances (endogenous, exogenous, technique-specific, microorganisms).
    Near-LoD/ReproducibilityMinimum of 90% agreement with expected positive results (≥95% desired) for all organisms. Minimum of 95% agreement with expected negative results.For positive samples, agreement with expected positive results (all systems/sites) was ≥98% for all analytes. Agreement with expected negative results was 100% for all analytes. Total positive agreement nearly identical between BioFire and clinical sites (99.8% vs. 99.0%).
    Matrix ValidationEquivalent performance between artificial and natural matrices based on agreement of positive and negative results at each test concentration. Considered equivalent if negative results observed at same or similar test concentration.Equivalent results achieved when samples prepared in natural and artificial NPS or natural and artificial TS matrices and tested with the SPOTFIRE R/ST Panel Mini.
    Transport Media ValidationPrimary metric: percent agreement between candidate medium and control medium (CDC VTM) for each spiked analyte at each test concentration. 100% agreement when testing above LoD and ≥95% at LoD for compatibility.Equivalent analyte detection observed for all representative analytes when samples were prepared in each of the candidate media types (BD™ Universal Viral Transport, and Remel MicroTest™ M4RT® Multi-Microbe Media) compared to the control medium (CDC VTM).
    Sample Carry OverPositive and negative analyte results evaluated. Positive samples expected positive for target and negative for others. Negative samples expected negative for all analytes.No unexpected positive results were observed in this study.

    2. Sample Sizes and Data Provenance

    • Clinical Performance (Test Set):
      • NPS Specimens (Respiratory Menu - Prospective): Total of 1115 specimens. The document doesn't explicitly state the country of origin but implies clinical sites (e.g., "as tested by intended users"). This is prospective data.
      • NPS Specimens (Respiratory Menu - Archived): Used for some analytes, e.g., Human Rhinovirus (30 positive, 454 negative), Influenza A (59 positive, 423 negative), Influenza B (30 positive, 28 negative), RSV (37 positive, 447 negative). This is retrospective data.
      • TS Specimens (Sore Throat Menu - Prospective): Total of 876 specimens for most viral targets. Streptococcus pyogenes had 217 positive (PCR) / 177 positive (Culture) and 660 negative (PCR) / 692 negative (Culture). This is prospective data.
      • TS Specimens (Sore Throat Menu - Archived): Used for some analytes, e.g., Human Rhinovirus (2 positive, 57 negative), Influenza A (11 positive, 44 negative), Influenza B (20 positive, 0 negative), RSV (2 positive, 57 negative), Streptococcus pyogenes (39 positive, 10 negative). This is retrospective data.
      • TS Specimens (Sore Throat Menu - Contrived): Used for some analytes, e.g., Influenza A (93 positive, 332 negative), Influenza B (49 positive, 333 negative), RSV (50 positive, 381 negative). This would be laboratory-generated data.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not specify the number or qualifications of experts used to establish the ground truth for the clinical test set. It mentions using "molecular assays or known specimen composition" as comparator methods for most analytes, and "culture" as the reference method for Streptococcus pyogenes.

    4. Adjudication Method

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing the ground truth or resolving discrepancies in the clinical test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned or implied, as this device is an in vitro diagnostic (IVD) PCR test for direct pathogen detection, not an AI-assisted diagnostic imaging or interpretation tool for human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply.

    6. Standalone Performance

    Yes, the studies described are for standalone performance. The BIOFIRE® SPOTFIRE® R/ST Panel Mini provides automated interpretation and reporting of test results based on the PCR assay. It is designed to be used independently to generate a qualitative detection and identification of microbial nucleic acids.

    7. Type of Ground Truth Used

    • Clinical Performance (Prospective/Archived): The ground truth for most analytes was established using molecular assays or, in some cases, known specimen composition. For Streptococcus pyogenes, culture was also used as a reference method for some comparisons.
    • Analytical Performance (LoD, Inclusivity, Exclusivity, Interference, Reproducibility, Matrix Validation, Transport Media Validation, Carry Over): The ground truth was established through known specimen composition (e.g., contrived samples with known concentrations of organisms, presence of interfering substances, specific transport media).

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI algorithm development. This device is a PCR-based test, and its performance is validated through analytical and clinical studies, not typically through a machine learning training phase with a distinct dataset. The "training" in this context refers to the development and optimization of the PCR primers, probes, and reaction conditions.

    9. How the Ground Truth for the Training Set Was Established

    Given that this is a PCR diagnostic device, not an AI algorithm in the typical sense of needing a "training set" for model learning, this question isn't directly applicable. The "ground truth" for developing and optimizing the PCR assays themselves would have been established through:

    • Careful selection and validation of synthetic nucleic acid targets.
    • Testing with characterized microbial isolates and clinical samples whose status was confirmed by established reference methods (e.g., sequencing, culture, validated molecular tests).
    • In silico analysis of genetic sequences to design primers and probes with high specificity and inclusivity.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232954

    Validate with FDA (Live)

    Date Cleared
    2024-03-26

    (187 days)

    Product Code
    Regulation Number
    866.3981
    Age Range
    2 - 65
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Nucleic Acid Test Including SARS-CoV-2 And Other Microbial Agents

    Additional Requlation Numbers: 21 CFR 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel is a multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE® System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis (Sore Throat menu).

    The following organism types and subtypes are identified and differentiated using the SPOTFIRE R/ST Panel:

    Respiratory Menu Viruses Adenovirus Coronavirus SARS-CoV-2 Coronavirus (seasonal) Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus Influenza A virus/H1-2009 Influenza A virus/ H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus

    • Bacteria Bordetella parapertussis Bordetella pertussis Chlamydia pneumoniae Mycoplasma pneumoniae
      Sore Throat Menu Viruses Adenovirus Coronavirus (seasonal) Human metapneumovirus Human rhinovirus/enterovirus Influenza A virus Influenza A virus/H1-2009 Influenza A virus/H3 Influenza B virus Parainfluenza virus Respiratory syncytial virus

    Bacteria Chlamydia pneumoniae Mycoplasma pneumoniae Streptococcus dysgalactiae (Group C/G Strep) Streptococcus pyogenes (Group A Strep)

    Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS/TS specimens during the acute phase of infection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngtis are indicative of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

    Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS or TS specimen. Positive results do not rule out coinfection with other organisms. The agent(s) detected by the SPOTFIRE R/ST Panel may not be the definite cause of disease.

    Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis.

    Device Description

    The SPOTFIRE R/ST Panel simultaneously identifies 15 different respiratory viral and bacterial pathogens in nasopharyngeal swabs (NPS) or 14 viral and bacterial pharyngitis pathogens in throat swabs (TS) from individuals with signs and symptoms of respiratory tract infections or pharyngitis, respectively, (see Intended Use:). The SPOTFIRE R/ST Panel is compatible with the SPOTFIRE System, a polymerase chain reaction (PCR)-based in vitro diagnostic system for infectious disease testing. The SPOTFIRE System Software executes the SPOTFIRE R/ST Panel test and interprets and reports the test results. The SPOTFIRE RIST Panel was designed to be used in CLIA-waived environments. A test is initiated by loading Hydration Solution into one port of the SPOTFIRE R/ST Panel pouch and NPS or TS specimen, mixed with the provided Sample Buffer, into the port of the SPOTFIRE R/ST Panel pouch and placing it in the SPOTFIRE System. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the SPOTFIRE System Software guides the user through the steps of placing the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

    The SPOTFIRE System contains coordinated systems of inflatable bladders and seal points, which act on the pouch to control the movement of liguid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liguid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

    Nucleic acid extraction occurs within the SPOTFIRE R/ST Panel pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the SPOTFIRE System performs a nested multiplex PCR that is executed in two stage, the SPOTFIRE System performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent doublestranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

    The SPOTFIRE System Software automatically interprets the results of each DNA met curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the SPOTFIRE R/ST Panel.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study proving the device meets those criteria, based on the provided FDA 510(k) summary for the BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel:

    The document describes the BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel, a multiplexed PCR test for simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids. The data presented primarily focuses on analytical performance and clinical performance to demonstrate substantial equivalence to a predicate device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the performance metrics reported, specifically the Positive Percent Agreement (PPA) / Sensitivity and Negative Percent Agreement (NPA) / Specificity. For analytical studies, acceptance criteria related to percentages of positive results and agreement levels are explicitly stated.

    Clinical Performance Acceptance Criteria (Implied by Data Reported):
    For each analyte, the acceptance criteria are generally to achieve high PPA and NPA. While specific numerical thresholds for PPA/NPA are not explicitly listed as "acceptance criteria" in the clinical tables themselves, the FDA's clearance implies that the presented performance met their internal requirements for substantial equivalence. For diagnostic tests, generally, PPA and NPA values in the high 90s are expected.

    Analytical Performance Acceptance Criteria and Reported Performance:

    StudyAcceptance CriteriaReported Device Performance and Conclusion
    Clinical Performance (NPS Specimens)High Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) for each analyte. (Specific numerical thresholds not explicitly stated as acceptance criteria in the table but are implied by the nature of regulatory submissions for diagnostic tests.)Table 2. R/ST Panel Performance Summary for NPS Specimens (Respiratory Menu):Across prospective and archived studies, PPAs for individual viruses ranged from 96.3% (Respiratory syncytial virus) to 100% (Adenovirus, Human metapneumovirus, Influenza A A/H3, Influenza B, Chlamydia pneumoniae, Mycoplasma pneumoniae). NPAs for individual viruses ranged from 90.6% (Human rhinovirus/enterovirus) to 100% (Human metapneumovirus, Influenza A, Influenza A A/H1-2009, Influenza A A/H3, Influenza B, Bordetella parapertussis, Bordetella pertussis, Chlamydia pneumoniae, Mycoplasma pneumoniae). Overall high performance observed.
    Clinical Performance (TS Specimens)High Sensitivity / PPA and Specificity / NPA for each analyte. (Specific numerical thresholds not explicitly stated as acceptance criteria in the table but are implied by the nature of regulatory submissions for diagnostic tests.)Table 3. R/ST Panel Performance Summary for TS Specimens (Sore Throat Menu):Across prospective, archived, and contrived studies, Sensitivities/PPAs for individual viruses ranged from 81.8% (Adenovirus, archived) to 100% (multiple analytes). Specificities/NPAs for individual viruses ranged from 91.7% (Adenovirus, archived) to 100% (multiple analytes). Overall high performance observed, with some variations based on study type (e.g., contrived samples often showing 100% agreement due to controlled conditions). Culture was used as the reference method for Streptococcus dysgalactiae and Streptococcus pyogenes.
    Sample Storage and Handling"For the storage condition to be considered acceptable for each organism, 100% expected positive results were required to be observed in all samples tested. In addition, crossing point (Cp) values were evaluated for each relevant assay and trended across the conditions to assess analyte stability over time.""Positive results were observed in 100% of all TSa samples tested at all conditions evaluated for all SPOTFIRE R/ST Panel analytes."Conclusion: "The SPOTFIRE R/ST Panel provides accurate results when TS specimens are stored in Amies media for up to 4 hours at ambient temperature (15-25 °C), up to 3 days at refrigerated temperature (2-8 °C), and up to 30 days at frozen temperature (≤ -15 °C). Similar results were previously observed with NPS specimens stored in transport media."
    Limit of Detection (LoD)"The LoD for each SPOTFIRE R/ST Panel analyte was confirmed when positive results were reported in at least 95% (≥19/20) of replicates tested at the LoD (1× LoD), and fewer than 95% (≤18/20) of replicates tested at a concentration 10-fold below LoD (0.1× LoD). Equivalent detection of representative analytes in single analyte and multi-analyte samples was determined primarily based on concordance of positive or negative results at each test concentration.""The LoD concentrations for the SPOTFIRE R/ST Panel analytes were confirmed...The panel accurately detected viruses and bacteria in samples contrived in either VTM or Amies media containing one or multiple organisms."Conclusion: "The SPOTFIRE R/ST panel provides accurate detection results for all analytes in single or polymicrobial specimens when present at or above the LoD. No adverse effect on the analytical sensitivity of the SPOTFIRE R/ST Panel was observed when evaluating multi-analyte specimens."
    Analytical Reactivity (Inclusivity)"The assay reactivity of each isolate was confirmed if positive results were reported for the appropriate analyte in 3/3 or 4/5 replicates tested within 10× LoD. If positive results were reported in fewer than 4/5 replicates, additional testing was performed at 100× LoD or higher.""Analytical reactivity testing demonstrated that the SPOTFIRE R/ST Panel can detect and accurately report results for a diverse collection of isolates from a variety of strains, serotypes, and genotypes of species collected over many years and from geographically distinct locations with few limitations." Specific limitations related to Streptococcus dysgalactiae and Streptococcus pyogenes were identified and noted in device labeling.
    Analytical Specificity (Exclusivity)"On-panel organisms were expected to have a positive result for the analyte being tested and negative results for all other analytes targeted by the panel. Off-panel organisms were expected to have negative results for all panel analytes, unless otherwise indicated.""Six cross-reactivities were identified by empirical and/or in silico evaluations that are predicted to cause inaccurate test results...Five of the identified cross-reactivities are either due to reactivity between phylogenetic near-neighbors that are rarely observed in human populations or were further evaluated and found to not impact the panel's specificity relevant to the intended use." Specific limitations related to SARS-CoV-2, B. bronchiseptica, Bordetella species, influenza A viruses of swine origin, bovine/canine picornaviruses, and Chlamydia gallinacea were identified and noted in device labeling.
    Interference (Interfering Substances)"If an unexpected result or control failure was observed for one replicate of a sample containing a potentially interfering substance, the affected sample was retested in two additional pouches to determine if the failure was reproducible." Implied: accurate results should be maintained in the presence of interfering substances."Accurate results for the SPOTFIRE R/ST Panel were reported in the presence of a variety of potentially interfering substances...".Conclusion: "The SPOTFIRE R/ST Panel provides accurate results in the presence of various potentially interfering substances." A warning about bleach was noted.
    Near-LoD/Reproducibility"For all organisms, a minimum of 90% agreement with the expected positive results (with 95% agreement desired) to demonstrate the reproducibility of positive results, and a minimum of 95% agreement with the expected negative results was required.""For positive samples, agreement with the expected positive results (all systems/sites) was $\ge$ 95% for all analytes. The agreement with the expected negative results was 100% for all analytes. The total positive agreement reported for testing completed at BioFire and at clinical sites was nearly identical (99.1% (2052/2070) and 98.9% (1365/1380), respectively)..."Conclusion: "The SPOTFIRE R/ST Panel provides accurate and reproducible analyte detection results over time and in actual use conditions when testing was performed over multiple days, by operators with differing skill levels, at different sites, using different SpotFire Systems and different reagent kit lots. ... support use of the SPOTFIRE R/ST Panel and SPOTFIRE System at sites that hold a CLIA Certificate of Waiver."
    Matrix Validation"Equivalent performance between the artificial and natural sample matrices was determined primarily based on agreement of positive and negative results at each test concentration. Artificial and natural matrices were considered equivalent if negative results were observed at the same or similar test concentration.""Performance of the SPOTFIRE R/ST Panel was determined to be equivalent in natural and artificial NPS (nNS and aNS) and in natural and artificial throat swab (nTS and aTS) matrices for five representative panel analytes. In all cases, negative results were observed in artificial and natural matrices at the same or similar test concentrations."Conclusion: "The results of this study demonstrated that the artificial NPS and artificial TS matrices were acceptable for use in analytical evaluation of SPOTFIRE R/ST Panel performance."
    Transport Media Validation"If the overall agreement was 100% when testing above the LoD and ≥95% when testing at the LoD, then the candidate medium was determined to be compatible. It was acceptable for the agreement to be less than 95% when testing below the LoD.""Equivalent analyte detection was observed for all representative analytes when samples were prepared in each of the candidate media types (BD™ Universal Viral Transport, and Remel MicroTest™ M4RT® Multi-Microbe Media) compared to the control medium (CDC VTM)."Conclusion: "The SPOTFIRE R/ST Panel demonstrated equivalent results when samples were prepared in Viral Transport Media, BD™ Universal Viral Transport, and Remel MicroTest™ M4RT® Multi-Microbe Media. These transport media are indicated in the product labeling as suitable for use with the SPOTFIRE R/ST Panel."
    Sample Carry Over"For positive samples, a positive result was expected for the analyte being tested and negative results were expected for all other analytes on the panel. Negative samples were expected to have a negative result for all analytes." Implied: No unexpected positive results due to carry-over."No unexpected positive results were observed in this study."Conclusion: "This study demonstrated that sample-to-sample carry-over between samples containing high concentrations of organism and negative samples is unlikely to occur and that carry-over poses an acceptable risk to the accuracy of the SPOTFIRE R/ST Panel test results when testing is performed according to the instructions for use."

    2. Sample Sizes Used for the Test Set and Data Provenance

    The sample sizes vary by analyte and study type (prospective, archived, contrived).

    • Clinical Performance (Test Set):

      • NPS Specimens (Respiratory Menu):
        • Prospective Data: The total number of NPS specimens tested across all viruses and bacteria is not explicitly stated as a single number but can be aggregated from the "Positive" and "Negative" counts for each analyte. For example, for Adenovirus, there were 33 positives and 1082 negatives, totaling 1115 prospective NPS specimens where Adenovirus was assessed. This appears to be the total number of clinical samples evaluated in the prospective study.
        • Archived Data: Similarly, for Adenovirus, there were 31 positives and 439 negatives, totaling 470 archived NPS specimens.
        • Contrived Data: Indicated as 0/0 for all NPS analytes, suggesting contrived samples were not used for clinical performance evaluation of NPS specimens.
      • TS Specimens (Sore Throat Menu):
        • Prospective Data: For Adenovirus, 65 positives and 810 negatives (total 875). This seems to be the total number of clinical throat swab specimens assessed.
        • Archived Data: For Adenovirus, 11 positives and 48 negatives (total 59).
        • Contrived Data: For Adenovirus, 50 positives and 381 negatives (total 431). Contrived samples were used for TS clinical performance studies.
      • Data Provenance: The document states "as tested by intended users," implying clinical sites. No specific country of origin is mentioned, but typically, these studies for FDA clearance involve sites within the US. The "Prospective" studies indicate prospective collection, while "Archived" refers to retrospective samples. "Contrived" samples are laboratory-prepared.
    • Analytical Performance (Test Set):

      • Limit of Detection (LoD): At least 20 replicates (e.g., 19/20) tested at LoD and 10-fold below LoD for each analyte.
      • Analytical Reactivity (Inclusivity): 3/3 or 4/5 replicates tested within 10x LoD for each isolate.
      • Near-LoD/Reproducibility: Not explicitly stated but mentions "multiple days," "multiple operators," "three unique SPOTFIRE Systems," "three distinct clinical sites holding a CLIA waiver." The reported positive and negative agreement totals are 2070 (BioFire) and 1380 (clinical sites) tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not specify the number of experts or their qualifications (e.g., radiologist with 10 years of experience) used to establish the ground truth. This is a common characteristic of in vitro diagnostic (IVD) submissions focusing on molecular diagnostic tests.

    • For molecular tests, the "ground truth" is typically established by comparator methods, often laboratory-developed tests (LDTs) or other cleared/validated molecular diagnostic assays (e.g., PCR followed by sequencing, or highly sensitive and specific reference PCR methods).
    • For bacterial culture (used for Streptococcus dysgalactiae and Streptococcus pyogenes), the ground truth is established by standard microbiological culture and identification techniques, performed by trained laboratory personnel.

    4. Adjudication Method for the Test Set

    The document does not describe a formal adjudication method (e.g., 2+1, 3+1) for the clinical test sets. This is expected given that the ground truth is established by laboratory reference methods (molecular or culture), which typically do not involve human reader adjudication in the same way imaging studies might. Any discrepancies between the investigational device and the reference method would be investigated by the manufacturer, rather than through an expert consensus adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is primarily relevant for imaging-based AI devices where human readers interpret medical images, and the AI assists or augments their performance. The BIOFIRE SPOTFIRE R/ST Panel is a molecular diagnostic test where the output is an automated detection of nucleic acids, not an interpretation by a human reader that could be augmented by AI.


    6. If a Standalone (i.e. Algorithm Only Without Human-in-the Loop Performance) Was Done

    Yes, the performance data presented (Tables 2 and 3 for Clinical Performance, and Table 4 for Analytical Performance) represents the standalone performance of the BIOFIRE SPOTFIRE R/ST Panel, which includes the instrument and its automated software interpretation. The device itself is designed to provide automated results without human interpretive input for the detection of targets. The "as tested by intended users" for clinical performance indicates that the device was used in a realistic setting, but the performance metrics reflect the direct output of the system.


    7. The Type of Ground Truth Used

    The ground truth for the clinical performance evaluation was established using:

    • Comparator molecular assays or known specimen composition for most analytes (referred to as "molecular assays or known specimen composition were used as comparator methods" for PPA/NPA). This suggests a combination of validated PCR assays, and potentially sequencing for confirmation.
    • Culture for Streptococcus dysgalactiae (Group C/G Strep) and Streptococcus pyogenes (Group A Strep). This is explicitly stated: "Performance measures of sensitivity and specificity refer to the prospective and archived Streptococcus and archived Streptococcus analytes for which culture was used as the reference method."

    For analytical studies, the ground truth was established by:

    • Known concentrations of organisms (e.g., viable or infectious units, nucleic acid copies/mL) for LoD and Inclusivity studies.
    • Defined panels of organisms or substances for Exclusivity and Interference studies.

    8. The Sample Size for the Training Set

    This document does not provide information on the training set size directly. The presented studies are for validation/testing of the device's performance. For molecular diagnostic assays like this, the 'training' of the algorithms (e.g., primer design, melt curve analysis interpretation) typically happens during the assay development phase, often using synthetic targets, cultured organisms, and preliminary clinical samples. However, this is not detailed in a 510(k) summary, which focuses on validation data against a defined product. The "software was verified and validated" statement implies that developmental data was used, but details on sample size for that specific phase are not provided in this regulatory summary.


    9. How the Ground Truth for the Training Set Was Established

    Since the training set details are not provided, the method for establishing its ground truth is also not explicitly stated in this document. However, based on typical IVD development practices:

    • Ground truth for assay development (which informs the 'training' of a molecular diagnostic system) would involve well-characterized positive and negative controls, reference strains, and potentially sequenced clinical isolates.
    • Molecular target sequences (in silico analysis)
    • Analytical dilution series where the exact concentration of the pathogen is known.
    • Samples confirmed by multiple, orthogonal laboratory methods.

    The rigorous analytical and clinical studies described in the results section serve as the validation of the final trained/developed system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220801

    Validate with FDA (Live)

    Date Cleared
    2022-06-24

    (98 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Multiplex Nucleic Acid System (per 21 CFR 866.3980) Streptococcus spp Nucleic Acid-Based Assay (per 21 CFR 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ID NOW™ Influenza A & B 2 assay performed on the ID NOW™ Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in direct nasal or nasopharyngeal swabs and nasal or nasopharyngeal swabs eluted in viral transport media from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus.

    Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

    Performance characteristics for influenza A were established during the 2016-2017 influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.

    If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

    ID NOW™ Strep A 2 is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections.

    Device Description

    ID NOW™ Influenza A & B 2 is a rapid. instrument-based isothermal test for the qualitative detection and differentiation of influenza A and influenza B from nasal swab or nasopharyngeal swabs tested directly or after elution in viral transport media collected from patients presenting with signs and symptoms of respiratory infection.

    ID NOW™ Strep A 2 is a rapid, instrument-based isothermal test for the qualitative detection of Group A Strep from throat swab specimens.

    All ID NOW™ assays utilize isothermal nucleic acid amplification technology and are comprised of:

    • Sample Receiver single use, disposable containing the elution buffer
    • Test Base single use, disposable comprising two sealed reaction tubes, each containing a lyophilized pellet
    • . Transfer Cartridge - single use, disposable for transfer of the eluted sample to the Test Base, and
    • ID NOW™ Instrument repeat use reader

    The reaction tubes in the ID NOW™ Influenza A & B 2 Test Base contain the reagents required for amplification of the target nucleic acid and an internal control. ID NOW™ Influenza A & B 2 utilizes a pair of templates (similar to primers) for the specific amplification of RNA from influenza A and B and a fluorescently labeled molecular beacon designed to specifically identify the amplified RNA targets.

    The reaction tubes in the ID NOW™ Strep A 2 Test Base contain the reagents required for Group A Strep bacterial lysis and the subsequent amplification of the target nucleic acid and an internal control. ID NOW™ Strep A 2 utilizes a pair of templates (similar to primers) for the specific amplification of DNA from Group A Strep and a fluorescently labeled molecular beacon designed to specifically identify the amplified nucleic acid target.

    All ID NOW™ assays are performed within the confinement of the Test Base, and no other part of the ID NOW™ Instrument has contact with the sample during the amplification process. This reduces the risk of instrument contamination and sample carry-over between measurements.

    To perform the assay, the Sample Receiver and Test Base are inserted into the ID NOW™ Instrument and the elution buffer is automatically heated by the instrument. The sample is added to the Sample Receiver and transferred via the Transfer Cartridge to the Test Base, resuspending the lyophilized pellets contained within the Test Base and initiating bacterial lysis (for ID NOW™ Strep A 2) and target amplification. Heating, mixing and detection by fluorescence is provided by the instrument, with results automatically reported.

    Results are displayed by the ID NOW™ Instrument and are also stored in an on-board archive and are assigned to a sample ID that has been entered into the ID NOW™ Instrument by the operator, and the date/time the test was performed. Data can be retrieved and downloaded by the operator at any time after testing. An external Universal Printer can be attached via USB to the ID NOW™ Instrument to print test results.

    AI/ML Overview

    The document describes the modified software for the ID Now Instrument, encompassing ID NOW Influenza A & B 2 and ID NOW Strep A 2 assays. The modification specifically addresses false invalid results caused by baseline values being lower than allowed by the original algorithm, leading to incorrect identification as "Empty Tube Values." This is an algorithm update only, with no changes made to the chemistry of the assays.

    Here's the breakdown of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the modified software address the reduction of false invalid results. The document implies that the "performance" here relates to the analytical performance characteristics of the assays (e.g., sensitivity, specificity) remaining equivalent to the predicate devices despite the software change. While explicit numerical acceptance criteria for reduction in false invalid rate are not provided in this excerpt, the study aims to demonstrate that the new algorithm resolves the "false invalid" issue without compromising the core analytical performance.

    For ID NOW™ Influenza A & B 2 (with software modification):

    ParameterAcceptance Criteria (Implied: Equivalent to Predicate)Reported Device Performance (with software modification)
    FDA Product CodeOCC, OZE, OOIOCC, OZE, OOI
    Assay TargetInfluenza A, Influenza BInfluenza A, Influenza B
    Intended UseQualitative detection and discrimination of influenza A and B viral RNA in direct nasal or nasopharyngeal swabs and nasal or nasopharyngeal swabs eluted in viral transport media from patients with signs and symptoms of respiratory infection, as an aid in differential diagnosis. Not for Influenza C. Negative results do not preclude infection. Performance characteristics for influenza A established during 2016-2017 influenza season (A/H3 and A/H1N1). Precautions for novel influenza A viruses.Same as predicate
    Intended Environment for UseProfessional use, in a medical laboratory or point of careProfessional use, in a medical laboratory or point of care
    InstrumentationID NOW™ InstrumentID NOW™ Instrument
    Sample TypeNasopharyngeal Swab, Nasal Swab and Nasal or Nasopharyngeal Swabs Eluted in Viral Transport MediaNasopharyngeal Swab, Nasal Swab and Nasal or Nasopharyngeal Swabs Eluted in Viral Transport Media
    Influenza A Viral TargetPB2 segmentPB2 segment
    Influenza B Viral TargetPA segmentPA segment
    TechnologyIsothermal nucleic acid amplificationIsothermal nucleic acid amplification
    Internal ControlYesYes
    Result InterpretationAutomatedAutomated
    Assay ResultQualitativeQualitative
    Time to Result< 15 minutes< 15 minutes
    False Invalid RateReduction of false invalid results associated with low baseline values.The study (though details not fully in this excerpt) would have demonstrated that the software modification successfully mitigates the false invalid issue without negatively affecting the accuracy or other performance characteristics of the assay.

    For ID NOW™ Strep A 2 (with software modification):

    ParameterAcceptance Criteria (Implied: Equivalent to Predicate)Reported Device Performance (with software modification)
    FDA Product CodePGX, OOIPGX, OOI
    Assay TargetStreptococcus pyogenesStreptococcus pyogenes
    Intended UseRapid, instrument-based, molecular in vitro diagnostic test for qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens from patients with signs and symptoms of pharyngitis, to aid in rapid diagnosis.Same as predicate
    Intended Environment for UseProfessional use, in a medical laboratory or point of careProfessional use, in a medical laboratory or point of care
    InstrumentationID NOW™ InstrumentID NOW™ Instrument
    Sample TypeThroat SwabThroat Swab
    Target AnalyteGroup A Streptococcus (Streptococcus pyogenes)Group A Streptococcus (Streptococcus pyogenes)
    TechnologyIsothermal nucleic acid amplificationIsothermal nucleic acid amplification
    Internal ControlYesYes
    Result InterpretationAutomatedAutomated
    Assay ResultQualitativeQualitative
    Time to Result< 8 minutes< 8 minutes
    False Invalid RateReduction of false invalid results associated with low baseline values.The study would have demonstrated that the software modification successfully mitigates the false invalid issue without negatively affecting the accuracy or other performance characteristics of the assay.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a Special 510(k) submission which is typically used for well-defined modifications to a legally marketed device where the modification does not affect the fundamental scientific technology of the device or its intended use. In such cases, extensive new clinical studies with large, prospectively collected sample sets might not be required if analytical performance remains equivalent and the specific issue (false invalids) is addressed.

    The provided excerpt does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). For a software modification intended to fix a specific issue like "false invalid results due to baseline values," the testing typically involves:

    • Retrospective analysis of stored clinical samples that previously yielded false invalid results.
    • Prospective testing of both contrived and potentially clinical samples designed to challenge the new algorithm's ability to correctly classify results, especially those with low baseline values.
    • Analytical studies (e.g., limit of detection, inclusivity/exclusivity) to confirm that the change did not negatively impact the assay's performance characteristics.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Given this is a molecular diagnostic test for direct pathogen detection, the ground truth for such devices is typically established through definitive laboratory methods (e.g., highly sensitive PCR, culture, sequencing) rather than expert human interpretation of images or other subjective data. Therefore, the concept of "experts" in the sense of radiologists providing ground truth is not applicable here.

    4. Adjudication Method for the Test Set

    Not applicable in the context of a molecular diagnostic assay where ground truth is established by objective laboratory methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a molecular diagnostic assay, not an AI-assisted diagnostic imaging device that involves human reader interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, in principle, the "performance" of the algorithm is assessed in a standalone manner. The device outputs a qualitative result ("Positive", "Negative", or "Invalid"). The study demonstrating the effectiveness of the software modification would specifically evaluate the algorithm's ability to correctly interpret the reaction (i.e., generate "Positive" or "Negative" results) and to correctly resolve cases that were previously generating "false invalid" results. There is no human-in-the-loop during the interpretation phase of these tests.

    7. The Type of Ground Truth Used

    The ground truth for molecular diagnostic tests like ID NOW Influenza A & B 2 and ID NOW Strep A 2 is generally established using reference laboratory methods that reliably detect the target nucleic acid or pathogen. This could include:

    • Culture: For bacterial targets like Streptococcus pyogenes.
    • PCR (Polymerase Chain Reaction) or RT-PCR: Highly sensitive and specific molecular methods, often considered the gold standard for detecting viral or bacterial nucleic acids.
    • Sequencing: To confirm the presence and identity of specific pathogen strains.

    The purpose of the software modification was to improve the algorithm's interpretation of internal controls and baseline signals to prevent erroneous invalid results, while ensuring the accuracy (sensitivity and specificity) of positive/negative calls remained consistent with the gold standard.

    8. The Sample Size for the Training Set

    The document does not specify the sample size for the training set. For a software modification to an existing algorithm like this, the "training" (or development and refinement) might involve:

    • Analyzing a dataset of historical internal control and baseline signal data, especially from cases that yielded false invalid results.
    • Developing and testing new thresholds or algorithms on this historical data.

    9. How the Ground Truth for the Training Set was Established

    For the purpose of training an algorithm to address "false invalid" results, the "ground truth" would be established by:

    • Investigating the source of the invalid signal: This might involve re-running the samples on a reference method to determine the true positive/negative status and then analyzing the raw signal data from the initial ID NOW test to understand why the invalid result occurred.
    • Characterizing "Empty Tube Values": Understanding the expected signal profiles when no reaction occurs or when the internal controls fail, versus when the assay is truly inhibited, or the baseline is genuinely low but still allows for a valid result. The ground truth for these investigations would be meticulously verified analytical data and, if applicable, the true clinical status (positive/negative by a gold standard method).
    Ask a Question

    Ask a specific question about this device

    K Number
    K201269

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-09

    (181 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Diego, California 92121

    Re: K201269

    Trade/Device Name: Accula Strep A Test Regulation Number: 21 CFR 866.2680
    Trade Name: Accula™ Strep A Test Classification of Device: 21 CFR 866.2680, Streptococcus spp.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Accula™ Strep A Test performed on the Accula Dock is a molecular in vitro diagnostic test utilizing polymerase chain reaction (PCR) and lateral flow technologies for the qualitative, visual detection of Streptococcus pyogenes (Group A B-hemolytic Streptococcus, Strep A) bacterial nucleic acid. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections from throat swabs of patients with signs and symptoms of pharyngitis.

    All negative test results should be confirmed by bacterial culture because negative results do not preclude infection with Group A Streptococcus and should not be used as the sole basis for treatment.

    Device Description

    The Accula™ Strep A Test is a semi-automated, colorimetric polymerase chain reaction (PCR) nucleic acid amplification test to qualitatively detect Streptococcus pyogenes (Group A Bhemolytic Streptococcus, Strep A) bacterial nucleic acid from unprocessed throat swabs that have not undergone prior nucleic acid extraction. The system integrates nucleic acid extraction, a novel Mesa Biotech PCR nucleic acid amplification technology named OscAR™, and hybridization-based visual detection into a completely self-contained and automated system. The Accula Strep A system consists of a small reusable Dock to drive the automated testing process, and a single-use disposable test cassette that contains all the enzymes and reagents.

    AI/ML Overview

    The Mesa Biotech Accula Strep A Test is an in vitro diagnostic test for the qualitative, visual detection of Streptococcus pyogenes (Group A B-hemolytic Streptococcus, Strep A) bacterial nucleic acid from throat swabs. The device integrates nucleic acid extraction, OscAR™ PCR amplification technology, and hybridization-based visual detection.

    Acceptance Criteria and Device Performance:

    The primary performance metrics for the Accula Strep A Test were Sensitivity, Specificity, Positive Percent Agreement (PPA), and Negative Percent Agreement (NPA). These were evaluated against a reference bacterial culture and an FDA-cleared molecular comparator.

    MetricAcceptance Criteria (Implied)Reported Device Performance (vs. Blood Agar Culture)Reported Device Performance (vs. Molecular Comparator)
    SensitivityHigh, expected to be comparable to or better than predicate96.2% (126/131) (95% CI: 91.4%-98.4%)N/A (PPA used for molecular comparison)
    SpecificityHigh, expected to be comparable to or better than predicate97.5% (510/523) (95% CI: 95.8%-98.5%)N/A (NPA used for molecular comparison)
    Positive Percent AgreementHigh, expected to be comparable to or better than predicateN/A93.8% (137/146) (95% CI: 88.7%-96.7%)
    Negative Percent AgreementHigh, expected to be comparable to or better than predicateN/A99.8% (501/502) (95% CI: 98.9%-100%)
    Reproducibility (Low Positive)High agreement (e.g., >95%) across sites, operators, and days98.9% (89/90) (95% CI: 94.0%-99.8%)N/A
    Reproducibility (Moderate Positive)High agreement (e.g., >95%) across sites, operators, and days97.8% (87/89) (95% CI: 92.2%-99.4%)N/A
    Reproducibility (Negative)High agreement (e.g., >95%) across sites, operators, and days97.8% (88/90) (95% CI: 92.3%-99.4%)N/A
    Limit of DetectionExpected to detect Strep A at low concentrationsBAA-946: 75 CFU/mL, ATCC 19615: 10 CFU/mLN/A
    Analytical Reactivity100% detection of tested Strep A strains at appropriate levels100% detection for 3/4 strains at 1.5x LoD, 100% for all at 3.0x LoDN/A
    Analytical SpecificityNo cross-reactivity with common respiratory pathogens and floraAll 47 tested organisms showed 0/3 positive when Strep A absent; 3/3 positive when Strep A present for all but two cases subsequently resolved by lower concentrationN/A
    Interfering SubstancesNo interference from common substances found in throat samples100% agreement with expected results for most tested substances at specified concentrationsN/A

    Study Details:

    1. Sample Size and Data Provenance:

      • Test Set (Clinical Study):
        • Evaluable for Accula vs. Culture: 654 samples from 669 enrolled subjects.
        • Evaluable for Accula vs. Molecular Comparator: 648 samples from 669 enrolled subjects.
        • Provenance: Prospective clinical study conducted at nine U.S. sites from May 2019 to January 2020.
      • Reproducibility/Near-Cutoff Study: 90 samples per condition (Low Positive, Moderate Positive, Negative) tested across three CLIA-waived sites. These were contrived throat swabs.
      • Limit of Detection (LoD): Replicates of 20 for confirmatory testing of two Strep A strains.
      • Analytical Reactivity: 3 replicates per strain (4 strains total) at two concentrations.
      • Analytical Specificity (Cross-Reactivity): 3 replicates per organism (47 organisms total), both in presence and absence of Strep A.
      • Interfering Substances: 3 replicates per substance, positive and negative Strep A samples.
    2. Number of Experts and Qualifications for Test Set Ground Truth:

      • The document does not explicitly state the "number of experts" used to establish the ground truth for the clinical test set. However, for the reference methods:
        • Bacterial Culture (Blood Agar Culture): Performed at a "central laboratory" according to "instructions from the reference laboratory." This implies trained laboratory personnel, but specific qualifications are not detailed.
        • FDA-cleared molecular test (comparator): Performed at the central laboratory.
        • Second FDA-cleared molecular test (discrepant analysis): Used for all discrepant results.
    3. Adjudication Method for the Test Set:

      • For the clinical study, a discrepant analysis method was used. "All specimens generating discrepant results between the Accula Strep A Test and Blood Agar Culture, or between Accula and the molecular comparator test, were tested with a second FDA-cleared molecular test." This effectively acts as a "tie-breaker" or confirmatory method for unusual results.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No MRMC comparative effectiveness study was specifically described in terms of human readers' improvement with AI vs. without AI assistance.
      • However, the Reproducibility/Near-Cutoff Study and the CLIA Waiver Studies involved "non-laboratory personnel" at "CLIA-waived sites" (Point of Care sites) to demonstrate that the device could be accurately used by intended users in the intended environment. This indirectly assesses the effectiveness of the device in the hands of typical users, rather than an AI assistance to human readers.
    5. Standalone (Algorithm Only) Performance:

      • Yes, the clinical performance described (Sensitivity, Specificity, PPA, NPA) represents the standalone performance of the Accula Strep A Test against established reference methods (Blood Agar Culture and FDA-cleared molecular tests). The device itself is a semi-automated system; these metrics evaluate its diagnostic accuracy independent of a human interpretation layer (beyond reading the visual result in the lateral flow).
    6. Type of Ground Truth Used:

      • For the clinical study, the primary ground truth was bacterial culture (Blood Agar Culture) for Streptococcus pyogenes.
      • A second FDA-cleared molecular test was used as a confirmatory ground truth for discrepant results.
      • Additionally, an FDA-cleared molecular comparator method served as another reference standard for direct comparison.
      • For analytical studies (LoD, Reactivity, Specificity, Interfering Substances), the ground truth was established by contriving samples with known concentrations of specific organisms or substances.
    7. Sample Size for the Training Set:

      • The document describes performance studies (validation). It does not provide information on a "training set" in the context of machine learning, as this is a nucleic acid amplification test, not an AI-driven image analysis or algorithm that would typically require a training set. The device's components (enzymes, reagents, PCR technology) are developed and optimized rather than "trained."
    8. How Ground Truth for the Training Set Was Established:

      • Not applicable as described in item 7. The device operates on molecular principles and does not involve a machine learning training phase with a labeled dataset in the traditional sense.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183366

    Validate with FDA (Live)

    Device Name
    GenePOC Strep A
    Manufacturer
    Date Cleared
    2019-03-06

    (92 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    ---|
    | Classification: | Class II |
    | Regulation: | 21 CFR 866.2680
    |
    | Regulation | 21 CFR 866.2680
    Franquet Québec G1P 4N3 Ca

    Re: K183366

    Trade/Device Name: GenePOC Strep A Regulation Number: 21 CFR 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GenePOC"M Strep A assay, performed on the revogene"14 instrument, is an automated, qualitative in vitro diagnostic test that utilizes real-time polymerase chain reaction (PCR) for the direct detection of Streptococcus pyogenes (Group A B-hemolytic Streptococcus) nucleic acids from throat swab specimens obtained from patients with signs and symptoms of pharyngitis. The GenePOC™ Strep A assay is intended for use as an aid in the diagnosis of Group A Streptococcus infection.

    Device Description

    The GenePOC™ Strep A assay is a single-use test for qualitative detection of Streptococcus pyogenes (group A Streptococcus - GAS) nucleic acids from throat swab specimens obtained from patients with signs and symptoms of pharyngitis. The GenePOCTM Strep A assay kit is comprised of the disposable Strep A microfluidic cartridge (PIE), Sample Buffer Tube (SBT), and Disposable Transfer Tool (DTT). These components are used to suspend the sample, extract, amplify, and detect Streptococcus pyogenes (S. pyogenes) nucleic acid.

    A Process Control (PrC) is also incorporated into each PIE to verify sample processing and amplification steps. The PrC allows for the verification of potential inhibitor substances as well as microfluidic, instrument or reagent failure. The GenePOC™ Strep A assay is designed to be used on the revogene™. The revogene™ is an instrument that automates sample homogenization, sample dilution, cell lysis, DNA amplification and detection of the amplified PCR products.

    Each GenePOC™ Strep A assay kit provides components for twenty-four (24) tests. User intervention is required for sample preparation, transferring throat swab specimen into the SBT, using the DTT to transfer the sample into the PIE, and loading the PIE into the revogene™ carousel. Each PIE is a completely integrated closed device in which a sample is dispensed and processed through different microfluidic chambers and channels that allow for the sample processing and subsequent real-time PCR steps.

    During the run and at run completion, the results are computed by the revogene™ from measured fluorescent signals and embedded calculation algorithms. The output results include positive, negative, indeterminate, and unresolved. An Early Positive Result Outcome (E-PRO) feature provides positive result if the signal from the target DNA reaches a predetermined threshold before the full PCR cycles have been completion of a run, the user removes the used cartridges and disposes of them in normal biological waste. Results may be viewed, printed, transferred, and/or stored by the user.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study proving the GenePOC™ Strep A device meets these criteria. Since the document is a 510(k) summary for an in vitro diagnostic (IVD) assay, the acceptance criteria are generally related to analytical performance (e.g., precision, detection limit, inclusivity, specificity, interference) and clinical performance (sensitivity and specificity compared to a reference method). The study design is focused on demonstrating the reliability and accuracy of the diagnostic test in detecting Streptococcus pyogenes.

    Here's the breakdown of the information requested based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    For an IVD such as this, the acceptance criteria are typically implicit in the "Performance Characteristics" section, where the manufacturer demonstrates that the device performs reliably and accurately for its intended use. There are no explicit pass/fail acceptance values stated for each measured characteristic, but the reported performance values are the data presented to demonstrate adequacy.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Precision/ReproducibilityOverall Qualitative Agreement
    Within-Laboratory PrecisionHigh agreement for negative, low positive, and moderate positive samplesNegative: 97.5% (95% CI: [91.3%-99.7%]) (78/80 detected negative)Low Positive: 100% (95% CI: [96.3%-100%]) (80/80 detected positive)Moderate Positive: 100% (95% CI: [96.3%-100%]) (80/80 detected positive)
    Between-Laboratory ReproducibilityHigh agreement across multiple sites for low positive, moderate positive, and negative samplesLow Positive: 100% (95% CI: [96.7%-100%]) (90/90 detected positive)Moderate Positive: 100% (95% CI: [96.7%-100%]) (90/90 detected positive)Negative: 100% (95% CI: [97.5%-100%]) (120/120 detected negative)
    Between-Lot ReproducibilityHigh agreement across multiple reagent lots for low positive, moderate positive, and negative samplesLow Positive: 97.8% (95% CI: [92.2-99.7%]) (88/90 detected positive)Moderate Positive: 100% (95% CI: [96.7-100%]) (90/90 detected positive)Negative: 100% (95% CI: [97.5-100%]) (120/120 detected negative)
    Limit of Detection (LoD)To establish the lowest detectable concentration of S. pyogenesRanges from 333 to 1,333 CFU/mL of Sample Buffer (SB) depending on the strain. Confirmed with various swab types.
    Analytical Reactivity/InclusivityTo detect a broad range of clinically relevant Streptococcus pyogenes strains with high positivity8 out of 9 strains detected with 100% positivity at 999 CFU/mL SB. One strain (ATCC® 49399™) detected with 100% positivity at 1667 CFU/mL SB. (9/9 replicates for each)
    Analytical Specificity/Cross-ReactivityNo cross-reactivity with common throat/mouth microorganisms, phylogenetically related species, or human gDNANo cross-reactivity observed among 50 non-specific analytes tested (concentrations up to ≥10^6 CFU/mL or cp/mL). Bioinformatic analysis also showed no significant homology with primers/probe.
    Carry-over and Cross-ContaminationNo false positive results due to carry-over or cross-contaminationNo false positive results detected (n=80 in cross-contamination, n=80 in carry-over).
    Assay InterferenceMinimal to no interference from endogenous/exogenous substances or other microorganisms at relevant concentrationsSome substances (Analgesic/Antipyretic, NSAID, Bronchodilator, Whole Blood, Mucin) showed inhibitory effect at high concentrations (4.3% w/v or v/v) but no reportable interference at lower, clinically relevant concentrations (0.1-0.4% w/v or v/v). A combinatory effect of specific Streptococcus species or S. dysgalactiae was noted as potentially inhibitory.
    Clinical PerformanceHigh sensitivity and specificity for direct detection of S. pyogenes in throat swab specimens compared to the reference method (culture).Sensitivity: 98.1% (151/154), 95% CI: [94.4% - 99.3%]Specificity: 94.7% (426/450), 95% CI: [92.2% - 96.4%]

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Clinical Performance Test Set (Clinical Studies):

      • Sample Size: 604 fully compliant prospective specimens.
      • Data Provenance:
        • Country of Origin: Geographically diverse clinical trial sites; two (2) in Canada and six (6) in the United States.
        • Retrospective or Prospective: Prospective multicenter trial. Throat swab specimens were collected from patients with signs and symptoms of pharyngitis.
    • Analytical Performance Test Sets (Examples):

      • Within-Laboratory Precision: 240 samples (80 low positive, 80 moderate positive, 80 negative samples) tested over 20 days.
      • Between-Laboratory Reproducibility: 300 samples (90 low positive, 90 moderate positive, 120 negative samples) tested across 3 sites over 5 days.
      • Between-Lot Reproducibility: 300 samples (90 low positive, 90 moderate positive, 120 negative samples) tested across 3 reagent lots over 15 days.
      • Limit of Detection: 24 replicates per concentration per strain (3 strains) with 3 kit lots; confirmation with 20 replicates per strain.
      • Analytical Reactivity/Inclusivity: 9 replicates per strain (12 strains total).
      • Analytical Specificity: 50 analytes, concentrations tested for each.
      • Carry-over and Cross-Contamination: 80 samples for cross-contamination, 80 samples for carry-over.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Clinical Performance Test Set: The ground truth for the clinical study was established by a "Reference Method" which is explicitly stated as culture. The text does not mention the number or qualifications of human experts (e.g., microbiologists, lab technicians) involved in performing or interpreting the reference culture method, as culture is considered the gold standard and its results are objective.
      • For discrepancies in the clinical study, an "alternative PCR with bi-directional sequencing" was performed on the discordant samples. This suggests a secondary method for adjudication rather than expert consensus on the primary ground truth.

    4. Adjudication Method for the Test Set

    • Clinical Performance Test Set:
      • The primary ground truth was established by the Reference Method (culture).
      • For discordant results between the GenePOC™ Strep A assay and the culture, an alternative PCR with bi-directional sequencing was used for further investigation. This is a form of discrepancy resolution or adjudication for the clinical performance data. The results of this alternative PCR are footnoted in the clinical performance table (e.g., "17 of 24 were Strep A Positive" for samples positive by GenePOC but negative by culture).
      • The text does not indicate a system like 2+1 or 3+1 involving human experts directly adjudicating case outcomes for the primary study endpoint; rather, it uses a technical discrepancy resolution method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • Not applicable. This device is an automated, qualitative in vitro diagnostic test for direct detection of nucleic acids. It does not involve human "readers" or image interpretation tasks where AI assistance would directly improve human reader performance. The device provides a direct positive/negative/indeterminate result.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Yes, implicitly. The device, the GenePOC™ Strep A assay on the revogene™ instrument, is described as an "automated, qualitative in vitro diagnostic test." Once the sample is loaded, "No operator intervention is necessary once the PIE is loaded onto the revogene™." The results are "computed by the revogene™ from measured fluorescent signals and embedded calculation algorithms."
    • The clinical performance study directly evaluates this standalone performance (assay results vs. culture ground truth). The human element is in sample preparation and loading, but the detection and result interpretation are automated by the device.

    7. The Type of Ground Truth Used

    • Clinical Performance Test Set: Culture (for Streptococcus pyogenes). This is the traditional "gold standard" for bacterial identification in clinical microbiology.
    • Analytical Performance Test Sets: Defined scientific standards, such as known concentrations of specific bacterial strains (CFU/mL), absence of target (negative matrix), and specific interfering substances, or bioinformatic analysis for sequence specificity.

    8. The Sample Size for the Training Set

    • This document is a 510(k) summary for an IVD test, not a machine learning or AI algorithm summary. The "device" here refers to a diagnostic assay kit used on an automated instrument.
    • The assay's "embedded calculation algorithms" determine the results based on fluorescent signals and pre-set thresholds/cut-offs for RNA/DNA detection. There is no mention of a "training set" in the context of machine learning model training. The cut-offs were determined by testing n=509 native and contrived samples, which could be considered an "assay development" or "validation" dataset rather than a "training set" for a continually learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • As noted above, there isn't a traditional "training set" in the machine learning sense. The assay cut-offs were established using a mix of "native (throat swab specimens) and contrived samples" (n=509). The ground truth for these would logically be established by the same reliable reference method (culture) or by known spiked concentrations for contrived samples.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173653

    Validate with FDA (Live)

    Date Cleared
    2018-05-02

    (155 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Alere i Strep A 2 Control Swab Kit Regulation Number: 21 CFR 866.2680 Regulation Name: Streptococcus
    Strep A 2 Control Swab Kit

    COMMON NAME

    Alere™ i Strep 2, Alere™ i

    CLASSIFICATION NAME

    21 CFR 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Alere i Strep A 2 is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections.

    Device Description

    Alere™ i Strep A 2 is a rapid, instrument-based isothermal test for the qualitative detection of Streptococus pyogenes Group A Strep from throat swab specimens. The Alere™ i Strep A 2 System utilizes isothermal nucleic acid amplification technology and is comprised of:

    • . Sample Receiver - single use, disposable containing the elution buffer
    • Test Base – single use, disposable comprising two sealed reaction tubes, each containing a lyophilized pellet
    • Transfer Cartridge single use, disposable for transfer of the eluted sample to the Test Base, and ●
    • . Alere™ i Instrument – repeat use reader

    The reaction tubes in the Test Base contain the reagents required for Streptococcus pyogenes Group A Strep bacterial lysis and the subsequent amplification of the target nucleic acid and an internal control. Alere™ i Strep A 2 utilizes a pair of templates (similar to primers) for the specific amplification of DNA from Streptococcus pyggenes, Group A Strep and fluorescently labeled molecular beacons designed to specifically identify the amplified nucleic acid targets. Alere™ i Strep A 2 is performed within the confinement of the Test Base, and no other part of the Alere™ i Instrument has contact with the sample during the amplification process. This reduces the risk of instrument contamination and sample carry-over between measurements.

    To perform the assay, the Sample Receiver and Test Base are inserted into the Alere™ i Instrument. The sample is added to the Sample Receiver and transfer Cartridge to the Test Base, resuspending the lyophilized pellets contained within the Test Base and initiating bacterial lysis and target amplification. Heating, mixing and detection by fluorescence is provided by the instrument, with results automatically reported.

    Results are displayed by the Alere™ i Instrument and are also stored in an on-board archive and are assigned to a sample ID that has been entered into the Alere™ i Instrument by the operator, and the date/time the test was performed. Data can be retrieved and downloaded by the operator at any time after testing. An external Alere™ Universal Printer can be attached via USB to the Alere™ i Instrument to print test results.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the Alere™ i Strep A 2 device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance document doesn't explicitly state all "acceptance criteria" as clear numerical thresholds for performance metrics. However, it presents the clinical performance of the device against a comparator method (bacterial culture), and these reported values implicitly demonstrate that the device met the necessary performance expectations for clearance.

    Performance MetricAcceptance Criteria (Implied by Clearance)Reported Device Performance (Alere™ i Strep A 2 vs. Culture)
    Clinical SensitivityHigh (specific threshold not stated)98.5% (95% CI = 95.6%, 99.5%)
    Clinical SpecificityHigh (specific threshold not stated)93.4% (95% CI = 91.4%, 94.9%)
    Positive Predictive ValueHigh (specific threshold not stated)78.9% (95% CI = 74.3%, 83.6%)
    Negative Predictive ValueHigh (specific threshold not stated)99.6% (95% CI = 98.3%, 99.9%)
    Initial Invalid RateLow (specific threshold not stated)0.9%
    Invalid Rate (after retest)Very Low (specific threshold not stated)0.4%
    Analytical Sensitivity (LOD)Specific concentrations for each strainATCC 12344: 147 cells/mL (100% Detected) ATCC 19615: 25 cells/mL (95% Detected)
    ReactivityPositive results for tested strainsAll listed strains produced positive reactions
    Analytical Specificity (Cross-Reactivity)Negative results for tested microorganismsAll listed microorganisms and yeast produced negative results
    Interfering SubstancesNo effect on test performanceMost substances showed no effect; few instances of false-negative/positive at higher concentrations
    Reproducibility (low/moderate positive)100% agreement with expected results100% (90/90) agreement
    Reproducibility (negative)100% negative results100% (90) negative results

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study): 981 evaluable throat swab specimens.
    • Data Provenance: Multi-center, prospective clinical study conducted at nine (9) US trial sites in 2017.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • The document states that the Alere™ i Strep A 2 performance was evaluated "in comparison to bacterial culture." Bacterial culture is typically performed in a laboratory by trained microbiologists using established protocols. The document does not specify the number of experts, nor their specific qualifications (e.g., years of experience), but implies standard laboratory practices for culture results.
    • For discrepancies, a "laboratory developed real-time PCR assay" was used for confirmation. This also implies expert analysis within a laboratory setting, but specific expert details are not provided.

    4. Adjudication Method for the Test Set

    • The primary ground truth for the clinical study was bacterial culture. In cases of discordance between Alere™ i Strep A 2 and bacterial culture, a "laboratory developed real-time PCR assay" was used for further investigation:
      • 38 of 52 samples positive by Alere™ i Strep A 2 and negative by bacterial culture were positive by PCR.
      • 1 of 3 samples negative by Alere™ i Strep A 2 and positive by bacterial culture was negative by PCR.
    • This suggests an implicit adjudication based on a tertiary, highly sensitive method (PCR) for resolving some discrepancies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not conducted. This device is an in-vitro diagnostic (IVD) molecular test, not an imaging or diagnostic assistant used by human readers in the traditional sense. The output is an automated positive/negative result.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Yes, the clinical performance study directly reflects the standalone performance of the device (Alere™ i Strep A 2) without human interpretation affecting the result. The device is "instrument-based" and provides "automated" results. The operator's role is to perform the assay steps, but the diagnostic determination is made by the instrument/algorithm.

    7. The Type of Ground Truth Used

    • The primary ground truth used for the clinical performance study was bacterial culture of throat swab specimens.
    • For discordant results, a laboratory-developed real-time PCR assay was used as a confirmatory method.

    8. The Sample Size for the Training Set

    • The document does not provide information regarding a specific training set size. For IVD devices, especially molecular diagnostic kits, the "training" (development and optimization) process typically involves internal analytical studies rather than a distinct, prospectively collected "training set" of clinical samples with established ground truth in the same way an AI/ML algorithm might. Clinical studies are primarily for validation.

    9. How the Ground Truth for the Training Set Was Established

    • As a molecular diagnostic test, the "ground truth" for its development (analogous to a training set for AI) would primarily rely on well-characterized clinical samples and reference strains with known Streptococcus pyogenes status confirmed by methods like culture and sequencing during the assay development and optimization phases. However, the document does not detail this. The provided clinical study serves as the primary validation of the device against bacterial culture as the ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173398

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-04-26

    (177 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Sunnyvale, California 94089

    Re: K173398

    Trade/Device Name: Xpert Xpress Strep A Regulation Number: 21 CFR 866.2680
    | 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xpert Xpress Strep A test, performed on the GeneXpert Xpress System, is a rapid, qualitative in vitro diagnostic test for the detection of Streptococcus pyogenes (Group A B-hemolytic Strep A) in throat swab specimens from patients with signs and symptoms of pharyngitis. The Xpert Xpress Strep A test can be used as an aid in the diagnosis of Group A Streptococcal pharvngitis. The assay is not intended to monitor treatment for Group A Streptococus infections.

    Device Description

    The Xpert Xpress Strep A test is an automated real-time polymerase chain reaction (PCR) in vitro diagnostic test for qualitative detection of Streptococcus pyogenes from throat swab specimens from patients with signs and symptoms of pharyngitis.

    The Xpert Xpress Strep A test is performed on the Cepheid GeneXpert® Xpress System. The GeneXpert Xpress System platform automates sample preparation, amplification and real-time detection.

    The GeneXpert Xpress System requires the use of single-use, disposable cartridges (the Xpert Xpress Strep A cartridges) that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized.

    The Xpress Strep A test includes primers and probes for the detection of a targeted sequence of the S. pyogenes genome allowing detection of Strep A directly from throat swab specimens collected from patients with signs and symptoms of pharyngitis. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are internal controls utilized by the GeneXpert Xpress System platform. The SPC is present to control for adequate processing of the target bacteria and to monitor for the presence of inhibitor(s) in the PCR assay to avoid false-negative results. The Probe Check Control verifies reagent rehydration, real-time PCR tube filling in the cartridge, probe integrity, and dye stability.

    The single-use, multi-chambered fluidic cartridges are designed to complete sample preparation and real-time PCR for the detection of genomic DNA S. pyogenes in ~24 minutes or less. The GeneXpert Xpress System, comprised of the GeneXpert Xpress II and GeneXpert Xpress IV, is capable of performing separate sample preparation and realtime PCR and RT-PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection.

    Throat swab specimens are collected using the ESwab collection device and transported to the GeneXpert area and prepared according to package insert instructions. After mixing the specimen, the liquid sample is transferred to the Xpert Xpress Strep A cartridge. The user initiates a test from the system user interface and places the cartridge into the GeneXpert Xpress instrument platform, which performs hands-off real-time, multiplex PCR for detection of DNA. The results are automatically generated at the end of the process in a report that can be viewed and printed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Cepheid Xpert Xpress Strep A device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined "acceptance criteria" in a bulleted or numbered list with pass/fail thresholds. Instead, it presents the results of various studies and concludes substantial equivalence to a predicate device. For the purpose of this request, I will infer the acceptance criteria from the reported performance of the predicate device and the general regulatory expectations for such assays, then juxtapose it with the Xpert Xpress Strep A's performance.

    Acceptance Criterion (Inferred)Xpert Xpress Strep A Reported Performance
    Clinical Performance:
    Sensitivity (relative to culture)99.4% (95% CI: 96.5-99.9)
    Specificity (relative to culture)94.1% (95% CI: 91.6-95.9)
    Positive Predictive Value (PPV)85.3% (95% CI: 79.5-89.7)
    Negative Predictive Value (NPV)99.8% (95% CI: 98.7-100.0)
    Accuracy95.5% (95% CI: 93.5-96.8)
    Indeterminate Rate (initial)5.3% (33/623)
    Indeterminate Rate (after retest)0.8% (5/623)
    Analytical Performance:
    Limit of Detection (LoD) - ATCC BAA-9469 CFU/mL (3 CFU/test)
    Limit of Detection (LoD) - ATCC 1961518 CFU/mL (6 CFU/test)
    Analytical Reactivity (Inclusivity)100% (24 S. pyogenes strains correctly detected at 3X LoD)
    Analytical Specificity (Exclusivity)100% (70 potentially cross-reactive microorganisms reported as "Strep A NOT DETECTED")
    Microbial InterferenceNo interference observed from 27 commensal microorganisms (with Strep A at 3X LoD)
    Interfering SubstancesNo assay interference from 10 potentially interfering substances (e.g., blood, mucus, saliva)
    Carry-Over Contamination0% (All 42 negative samples correctly reported as "Strep A NOT DETECTED" after high positive samples)
    Reproducibility:
    Total Agreement (Negative)100% (90/90)
    Total Agreement (Strep A High Neg)91% (82/90)
    Total Agreement (Strep A Low Pos)97% (87/90)
    Total Agreement (Strep A Moderate Pos)100% (90/90)
    Initial Indeterminate Rate (reproducibility study)3.6% (13/360) (all resolved upon retesting)

    2. Sample Size and Data Provenance

    • Clinical Study Test Set Sample Size: 618 specimens were included in the final analysis (initially 666, with 43 excluded).
    • Data Provenance: The clinical study was a prospective, multi-center investigational study conducted at nine clinical sites in geographically diverse regions within the United States between January 2017 and May 2017.

    3. Number of Experts and Qualifications for Ground Truth (Clinical Study)

    The document specifies that the Xpert Xpress Strep A clinical performance was established relative to culture and latex agglutination for Strep A typing. It also mentions "an alternative PCR/bidirectional sequencing assay" used to investigate discordant results.

    • Number of Experts: Not explicitly stated for establishing the primary ground truth (culture and latex agglutination). These methods are standard laboratory procedures typically performed by trained medical technologists or microbiologists.
    • Qualifications of Experts: Not explicitly described. However, the use of standard microbiology lab techniques implies performance by qualified laboratory personnel. The "alternative PCR/bidirectional sequencing assay" would also be performed by trained molecular diagnosticians or researchers.

    4. Adjudication Method (Clinical Study)

    • Primary Adjudication: The primary ground truth for the clinical study was established by culture and latex agglutination for Strep A typing. This acts as the "gold standard" against which the device was compared.
    • Discordant Analysis: For specimens where the Xpert Xpress Strep A result differed from the culture result, an "alternative PCR/bidirectional sequencing assay" was used to investigate the discrepancy. This serves as a secondary adjudication method to verify the true status of discordant samples.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. The Xpert Xpress Strep A is an automated in vitro diagnostic (IVD) test, not an AI system designed to assist human readers in image interpretation or similar tasks. Its performance is evaluated independently against a reference method (culture).

    6. Standalone Performance

    • Yes, a standalone performance study was done. The entire clinical study, analytical sensitivity, specificity, interference, and reproducibility studies assess the performance of the algorithm and instrument (Xpert Xpress Strep A test on the GeneXpert Xpress System) directly against various reference standards and controlled conditions, without human interpretation of the final result. The device provides a "Strep A DETECTED" or "Strep A NOT DETECTED" result automatically.

    7. Type of Ground Truth Used (Clinical Study)

    • The primary ground truth used for the clinical study was bacterial culture and latex agglutination for Strep A typing.
    • For discordant sample resolution, an alternative PCR/bidirectional sequencing assay was used.

    8. Sample Size for the Training Set

    The document describes pre-market validation studies typically conducted for IVD devices, not machine learning model development. Therefore, there is no explicit "training set" sample size mentioned as would be the case for an AI/ML device. The device's underlying PCR assay design and algorithm are developed based on established scientific principles and analytical verification, rather than being "trained" on a large dataset in the AI sense.

    9. How Ground Truth for the Training Set Was Established

    As mentioned above, the concept of a "training set" and its "ground truth" in the AI/ML context is not directly applicable here. The device is a PCR assay with a defined molecular target and detection algorithm. The closest analogue to "ground truth establishment" during development would be:

    • Analytical Validation: Extensive analytical studies (e.g., Limit of Detection, Inclusivity, Exclusivity, Interference) using well-characterized bacterial strains, spiked samples, and clinical matrices to ensure the assay correctly identifies the target organism and differentiates it from non-targets under various conditions. These studies confirm the assay's fundamental ability to detect S. pyogenes DNA.
    • Assay Design and Optimization: The design of the primers and probes for the S. pyogenes genome target would be based on known genetic sequences, and optimized empirically to ensure specificity and sensitivity.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173932

    Validate with FDA (Live)

    Date Cleared
    2018-01-26

    (31 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Multiplex Nucleic Acid System (per 21 CFR 866.3980) Streptococcus spp Nucleic Acid-Based Assay (per 21 CFR 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alere™ i Influenza A & B assay performed on the Alere™ i Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in direct nasal or nasopharyngeal swabs eluted in viral transport media from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus.

    Negative results do not preciude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

    Performance characteristics for influenza A were established during the 2012-2013 and the 2014-2015 influenza seasons when influenza A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.

    If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

    Alere™ i Strep A is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus progeness, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections.

    All negative test results should be confirmed by bacterial culture because negative results do not preclude infection with Group A Streptococcus and should not be used as the sole basis for treatment.

    The Alere™ i RSV assay performed on the Alere™ i Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection of respiratory syncytial virus (RSV) virus (RSV) virus (RSV) virus (RSV) virus (RSV) viral RNA in direct nasopharyngeal swabs and nasopharyngeal swabs cluted in viral transport media from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the diagnosis of RSV in children < 18 years and adults ≥60 years in conjunction with clinical and epidemiological risk factors.

    The Alere i Influenza A & B 2 assay performed on the Alere i Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in direct nasal or nasopharyngeal swabs and nasal or nasopharyngeal swabs eluted in viral transport media from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus.

    Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

    Performance characteristics for influenza A were established during the 2016-2017 influenza season when influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.

    If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in cases unless a BSL 3+ facility is available to receive and culture specimens.

    Device Description

    Alere™ i Influenza A & B is a rapid, instrument-based isothermal test for the qualitative detection and differentiation of influenza A and influenza B from nasal or nasopharyngeal swab specimens eluted in viral transport media collected from patients presenting with signs and symptoms of respiratory infection.

    Alere™ i Strep A is a rapid, instrument-based isothermal test for the qualitative detection of Group A Strep from throat swab specimens.

    Alere™ i RSV is a rapid, instrument-based isothermal test for the qualitative detection and differentiation of respiratory syncytial virus (RSV) viral RNA from direct nasopharyngeal swab (NPS) and NPS eluted in viral transport media from patients with signs and symptoms of respiratory infection.

    Alere™ i Influenza A & B 2 is a rapid, instrument-based isothermal test for the qualitative detection and differentiation of influenza A and influenza B from nasal swab or nasopharyngeal swabs tested directly or after elution in viral transport media collected from patients presenting with signs and symptoms of respiratory infection.

    All Alere™ i assays utilize isothermal nucleic acid amplification technology and are comprised of:

    • Sample Receiver single use, disposable containing the elution buffer .
    • . Test Base - single use, disposable comprising two sealed reaction tubes, each containing a lyophilized pellet
    • Transfer Cartridge sinqle use, disposable for transfer of the eluted sample to the Test Base, and ●
    • Alere™ i Instrument repeat use reader .

    The reaction tubes in the Alere™ i Influenza A & B and Alere™ i Influenza A & B 2 Test Base contain the reaqents required for amplification of the target nucleic acid and an internal control. Alere™ i Influenza A & B utilizes a pair of templates (similar to primers) for the specific amplification of RNA from influenza A and B and a fluorescently labeled molecular beacon designed to specifically identify the amplified RNA targets.

    The reaction tubes in the Alere™ i Strep A Test Base contain the reagents required for Group A Strep bacterial lysis and the subsequent amplification of the target nucleic acid and an internal control. Alere™ i Strep A utilizes a pair of templates (similar to primers) for the specific amplification of DNA from Group A Strep and a fluorescently labeled molecular beacon designed to specifically identify the amplified nucleic acid target.

    The reaction tubes in the Alere™ i RSV Test Base contain the reagents required for amplification of the target nucleic acid and an internal control. Alere™ i RSV utilizes a pair of templates (similar to primers) for the specific amplification of RNA from RSV A and B and a fluorescently labeled molecular beacon designed to specifically identify the amplified RNA targets.

    All Alere™ i assays are performed within the confinement of the Test Base, and no other part of the Alere™ i Instrument has contact with the sample during the amplification process. This reduces the risk of instrument contamination and sample carry-over between measurements.

    To perform the assay, the Sample Receiver and Test Base are inserted into the Alere™ i Instrument and the elution buffer is automatically heated by the instrument. The sample is added to the Sample Receiver and transferred via the Transfer Cartridge to the Test Base, resuspending the Iyophilized pellets contained within the Test Base and initiating bacterial lysis (for Alere™ i Strep A) and target amplification. Heating, mixing and detection by fluorescence is provided by the instrument, with results automatically reported.

    Results are displayed by the Alere™ i Instrument and are also stored in an on-board archive and are assigned to a sample ID that has been entered into the Alere™ i Instrument by the operator, and the date/time the test was performed. Data can be retrieved and downloaded by the operator at any time after testing. An external Alere™ Universal Printer can be attached via USB to the Alere™ i Instrument to print test results.

    AI/ML Overview

    This document describes a Special 510(k) submission for the Alere™ i Instrument and its associated assays (Influenza A & B, Strep A, RSV, and Influenza A & B 2). The purpose of this submission is to address a software modification to the Alere™ i Instrument's algorithm to mitigate issues with false invalid results due to baseline values being lower than allowed, which were incorrectly identified as "Empty Tube Values." The submission states that there have been no changes made to the chemistry of the assays.

    The document highlights the substantial equivalence of the modified devices to their legally marketed predicate devices. However, it does not contain specific acceptance criteria or detailed study data to prove the device meets these criteria. Instead, it focuses on comparing the modified devices with their predicates, stating that all parameters (FDA Product Code, Assay Target, Intended Use, Intended Environment for Use, Instrumentation, Sample Type, Viral/Bacterial Target, Technology, Internal Control, Result Interpretation, Assay Result, and Time to Result) are "Same" as the predicate devices.

    Given the information provided, it is not possible to complete a table of acceptance criteria and reported device performance, nor can we detail the study that proves the device meets the acceptance criteria, as this specific information is not present in the provided text. The document asserts "substantial equivalence" based on the described software modification and the consistency of other parameters with the predicate devices.

    Therefore, the following information can be extracted or derived based on the provided text, with many fields remaining unascertainable:

    1. A table of acceptance criteria and the reported device performance:

    This information is not provided in the document. The document states that the modified devices are "substantially equivalent" to their predicate devices and lists various parameters which are "Same" as the predicates. No specific numerical acceptance criteria or performance metrics (like sensitivity, specificity, accuracy) are detailed for the modified device or its predicate.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This information is not provided in the document. The submission focuses on the software modification and states that no changes were made to the assay chemistry. It does not describe any new clinical studies or test sets with sample sizes for the modified device to demonstrate its performance against new acceptance criteria. It refers to historical performance characteristics for influenza A established during certain influenza seasons for the Alere™ i Influenza A & B and Alere™ i Influenza A & B 2 assays.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided in the document. As there is no description of a new clinical study with a test set and ground truth establishment, these details are absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is an in vitro diagnostic test for detecting viral/bacterial RNA/DNA, not an AI-assisted diagnostic imaging device that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    The device is described as an "instrument-based, molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology" with "automated" result interpretation. This implies a standalone (algorithm only) performance, i.e., without human-in-the-loop performance influencing the assay result. However, specific standalone performance study details are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the original performance characteristics mentioned for the influenza assays, it can be inferred that the ground truth would have been established by a reference method for detecting viral RNA, likely PCR or viral culture, rather than expert consensus or pathology in the context of an IVD. However, the document does not explicitly state the ground truth used for performance characteristics, nor does it detail ground truth for any new studies related to the software modification.

    8. The sample size for the training set:

    This information is not provided in the document. Since this is a software modification to an existing algorithm for an IVD, it's possible the "training set" concept as used in AI/ML might not directly apply, or details about algorithm development data are not included.

    9. How the ground truth for the training set was established:

    This information is not provided in the document.

    In summary, the provided document is a 510(k) summary for a software modification to existing IVD devices, asserting substantial equivalence to predicates without detailing new clinical studies, acceptance criteria, or performance data for the modified device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172402

    Validate with FDA (Live)

    Date Cleared
    2017-10-30

    (82 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: ARIES Group A Strep Assay, ARIES Group A Strep Assay Protocol File Kit Regulation Number: 21 CFR 866.2680
    -------------------------------|----------------------|
    | PGX | Class II | 21 CFR 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARIES® Group A Strep Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of Streptococus pyogenes (Group A beta-hemolytic Streptococcus) in throat swab specimens from patients with signs and symptoms of pharyngitis.

    The ARIES® Group A Strep Assay can be used as an aid in the diagnosis of Group A Streptococcal pharyngitis. The assay is not intended to monitor treatment for Group A Streptococcus infections.

    The ARIES® Group A Strep Assay is indicated for use with ARIES® Systems.

    Device Description

    The ARIES® Group A Strep Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system that consists of the ARIES® System or the ARIES® M1 System with their included ARIES® Software, an assay-specific cassette, and an assay-specific protocol file. The ARIES® Group A Strep Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control (SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES® Group A Strep Assay cassette directly detects Streptococcus pyoqenes (Group A ß-hemolytic Streptococcus) in throat swab specimens collected from the surface of human tonsils and posterior pharyngeal wall.

    Throat swab specimens are collected from patients using a commercially available Liquid Amies based transport system (Nylon Flocked Swab with 1 mL modified Liquid Amies (ESwab™). The specimen is then transported to the laboratory for testing.

    The specimen is lysed and nucleic acid is extracted using an ARIES® System. An extractable sample processing control (SPC) target is present in the ARIES® Group A Strep Assay cassette and is processed with the specimen. The SPC controls for recovery of extracted nucleic acid, the presence of inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument. The Tm value of the SPC is used as a reference for determining the target Tm.

    The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES® Group A Strep Assay lyophilized PCR reagents in the PCR tube that contains primer pairs specific to the S. pyogenes DNaseB (sdaB) gene and the SPC sequence. Each of the primer pairs is labeled with a distinct fluorophore and detected in distinct channels of the ARIES® Systems. PCR amplification is performed and assay fluorescence is monitored. Incorporation of a quencher-labeled nucleotide results in a decrease in fluorescence for the associated primer pair. Following amplification, the reaction is slowly heated to separate the fluorescent-labeled strand from the quencher-labeled strand, a process that results in an increase in the fluorescence signal. The reaction fluorescence is measured during this process and the temperature at which the change in fluorescence is the maximum is the T ,, of the amplicon. The instrument fluorescence output is analyzed and test results are determined using the ARIES® System software and the ARIES® Group A Strep Assay protocol and run files. ARIES® Group A Strep Assay results may be reported from the ARIES® Software or from the optional SYNCT® Software.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the ARIES® Group A Strep Assay, based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (ARIES Group A Strep Assay)
    Clinical SensitivityHigh, to ensure detection of true positives97.5% (156/160), 95% CI: 93.7% - 99.0%
    Clinical SpecificityHigh, to minimize false positives97.8% (448/458), 95% CI: 96.0% - 98.8%
    Positive Predictive Value (PPV)High, indicating reliability of positive results94.0% (156/166), 95% CI: 89.3% - 96.7%
    Negative Predictive Value (NPV)High, indicating reliability of negative results99.1% (448/452), 95% CI: 97.7% - 99.7%
    Reproducibility (Overall Moderate Positive - 3X LoD)Consistent detection across sites98.9% (89/90)
    Reproducibility (Overall Low Positive - 1X LoD)Consistent detection across sites for low concentrations96.7% (87/90)
    Reproducibility (Overall Negative)Consistent negative results across sites1.1% (1/90) false positive, indicating high specificity
    Lot-to-Lot Reproducibility (Overall Moderate Positive - 3X LoD)Consistent detection across lots100% (45/45)
    Lot-to-Lot Reproducibility (Overall Low Positive - 1X LoD)Consistent detection across lots for low concentrations93.3% (42/45)*
    Lot-to-Lot Reproducibility (Overall Negative)Consistent negative results across lots0.0% (0/45)
    Within-Laboratory Precision/Repeatability (Moderate Positive - 3X LoD)Consistent detection within lab100% (30/30)
    Within-Laboratory Precision/Repeatability (Low Positive - 1X LoD)Consistent detection within lab for low concentrations93.3% (28/30)*
    Within-Laboratory Precision/Repeatability (Negative)Consistent negative results within lab0.0% (0/30)
    Limit of Detection (LoD) - Bruno strainLowest concentration detected with ≥ 95% positivity2.58E+03 CFU/mL (95%)
    Limit of Detection (LoD) - SF370 strainLowest concentration detected with ≥ 95% positivity4.13E+03 CFU/mL (100%)
    Analytical Reactivity (Inclusivity)Detection of various S. pyogenes strains8/9 strains detected with 100% positivity at 3x LoD, 1 strain at 5x LoD
    Interfering SubstancesNo inhibition/false results for tested substancesExpected results generally obtained, except for NyQuil (false negatives) and Mucin (invalid results)
    Cross-Reactivity/Microbial InterferenceNo cross-reactivity with common throat microorganisms34/35 tested microorganisms showed no cross-reactivity or interference with GAS positivity (1 Treponema denticola interference at high concentrations)
    Carry-Over/Cross-ContaminationNo carry-over between samples100% agreement with expected results for high positive and negative samples

    Note: For Lot-to-Lot Reproducibility (Low Positive) and Within-Laboratory Precision/Repeatability (Low Positive), additional replicates were tested to achieve an overall positivity >94%, indicating acceptable performance at the LoD.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Performance (Test Set): A total of 618 unique specimens were available for analysis in the clinical performance study. Initially, 735 specimens were collected, but 112 were excluded due to various reasons (protocol non-compliance, insufficient volume, incorrect device, etc.).
    • Data Provenance: The data was prospectively collected from patients suspected of pharyngitis in four geographically distinct clinical sites within the United States.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the "number of experts" or their specific "qualifications" (e.g., "Radiologist with 10 years of experience") for establishing the ground truth of the clinical test set.

    However, the ground truth for the clinical test set was established by a "reference method (bacterial culture followed by organism identification by Matrix-Assisted Laser Desorption/Ionization - Time-of-Flight Mass Spectrometry (MALDI-TOF MS))". This work was "performed at a centralized testing facility." It is implied that trained laboratory personnel and microbiologists conducted these reference tests, as is standard practice for such methods, but their specific experience levels are not detailed.

    4. Adjudication Method for the Test Set

    The document describes a robust "ground truth" establishment process using a reference laboratory technique (bacterial culture + MALDI-TOF MS). For specimens where the ARIES assay and the reference method disagreed, an adjudication method was employed:

    • Bidirectional sequencing analysis using analytically validated primers that targeted genomic regions distinct from the ARIES Group A Strep Assay was used.
      • For 2 false negative ARIES results (ARIES negative, culture positive), sequencing showed them to be GAS negative.
      • For 7 false positive ARIES results (ARIES positive, culture negative), sequencing showed them to be GAS positive.
        This indicates a method similar to "Truth by Consensus" with an expert-level tie-breaker (sequencing), effectively refining the ground truth against discrepancies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) assay that machine-interprets results (qualitative PCR). It does not involve human "readers" assessing images or data in a way that would be "assisted" by AI, hence the concept of a human reader improvement effect size does not apply.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was done. The ARIES® Group A Strep Assay is an automated qualitative in vitro diagnostic test system. The clinical performance study directly assessed the agreement of the ARIES® Group A Strep Assay (algorithm only within the ARIES® Systems) against the established reference method, without human interpretation of the assay's raw output. The results (Positive, Negative, Invalid) are determined by the ARIES® System software.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for the clinical performance study was primarily bacterial culture followed by organism identification by Matrix-Assisted Laser Desorption/Ionization - Time-of-Flight Mass Spectrometry (MALDI-TOF MS). In cases of discordance with the assay, bidirectional sequencing analysis was used as a confirmatory "higher truth" method. This can be considered a form of expert consensus/gold standard laboratory method.

    8. The Sample Size for the Training Set

    The document does not explicitly state a sample size for a training set. The ARIES® Group A Strep Assay is a PCR-based diagnostic test, where the "training" typically involves empirical optimization of assay parameters (primers, probes, reaction conditions, and cut-offs) during product development and internal verification studies, rather than machine learning on a distinct "training set" of patient data. The "Initial Assay Protocol File parameters were set during internal optimization studies" and "The final Assay Protocol File parameters were then established during internal verification studies." So, the optimization and verification process served a similar function to "training" for machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, a traditional "training set" with separate ground truth establishment isn't described in the context of a PCR assay. Instead, the "ground truth" for optimizing the assay protocol file parameters (Ct cut-off, Tm window, Tm Peak Threshold) was established through internal optimization and verification studies. These studies likely involved testing known positive and negative S. pyogenes samples (e.g., well-characterized strains, spiked samples) to define the optimal performance characteristics of the assay before clinical validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172126

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-09-25

    (73 days)

    Product Code
    Regulation Number
    866.2680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    SUNNYVALE CA 94089

    Re: K172126

    Trade/Device Name: Xpert Xpress Strep A Regulation Number: 21 CFR 866.2680
    | 866.2680

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xpert Xpress Strep A Assay, performed on the GeneXpert Instrument Systems, is a rapid, qualitative in vitro diagnostic test for the detection of Streptoccus pyogenes (Group A beta-hemolytic Strep A) in throat swab specimens from patients with signs and symptoms of pharyngitis.

    The Xpert Xpress Strep A Assay utilizes an automated real-time polymerase chain reaction (PCR) to detect Streptococcus pyogenes DNA.

    Device Description

    The Xpert Xpress Strep A Assay is an automated real-time polymerase chain reaction (PCR) in vitro diagnostic test for qualitative detection of Streptococcus pyogenes from throat swab specimens from patients with signs and symptoms of pharyngitis.

    The Xpert Xpress Strep A Assay is performed on the Cepheid GeneXpert® Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s, and GeneXpert Infinity-80 systems). The GeneXpert Instrument System platform automates sample preparation, amplification and real-time detection.

    The GeneXpert Instrument Systems require the use of single-use, disposable cartridges (the Xpert Xpress Strep A cartridges) that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized.

    The Xpress Strep A Assay includes primers and probes for the simultaneous detection and differentiation of a targeted sequence of the S. pyogenes genome allowing detection of Strep A directly from throat swab specimens collected from patients with signs and symptoms of pharyngitis. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are internal controls utilized by the GeneXpert Instrument System platform. The SPC is present to control for adequate processing of the target bacteria and to monitor for the presence of inhibitor(s) in the PCR assay to avoid false-negative results. The Probe Check Control verifies reagent rehydration, real-time PCR tube filling in the cartridge, probe integrity, and dye stability.

    The single-use, multi-chambered fluidic cartridges are designed to complete sample preparation and real-time PCR for the detection of genomic DNA S. pyogenes in ~24 minutes or less. The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems and the GeneXpert Infinity Systems, have 1 to 80 randomly accessible modules. depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR and RT-PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection.

    Throat swab specimens are collected using the ESwab collection device and transported to the GeneXpert area and prepared according to package insert instructions. After mixing the specimen, the liquid sample is transferred to the Xpert Xpress Strep A Assay cartridge. The user initiates a test from the system user interface and places the cartridge into the GeneXpert instrument platform, which performs hands-off real-time, multiplex PCR for detection of DNA. The results are automatically generated at the end of the process in a report that can be viewed and printed.

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for performance are not explicitly stated in a dedicated section with pre-defined numerical targets. However, based on the clinical study results and comparisons to the predicate, we can infer the demonstrated performance. The key performance metrics are Sensitivity, Specificity, Positive Predictive Value (PPV), and Negative Predictive Value (NPV) relative to culture and latex agglutination.

    Inferred Acceptance Criteria (Based on demonstrating substantial equivalence to predicate) and Reported Device Performance (Combined First and Second Swab Data):

    Performance MetricImplied Acceptance Criterion (Likely for Substantial Equivalence)Reported Device PerformanceComments
    SensitivityHigh (e.g., comparable to or better than predicate)100.0% (95% CI: 97.3-100.0)Excellent sensitivity.
    SpecificityHigh (e.g., comparable to or better than predicate)94.1% (95% CI: 91.5-95.9)Good specificity.
    PPVHigh (e.g., comparable to or better than predicate)84.1% (95% CI: 77.8-88.9)
    NPVHigh (e.g., comparable to or better than predicate)100.0% (95% CI: 99.1-100.0)Excellent negative predictive value.
    Indeterminate RateLow (e.g., <5%)0.9% (6/583) after retestAcceptable low rate.
    Analytical Sensitivity (LoD)Lowest concentration reproducibly distinguished 95% of the time9 CFU/mL (ATCC BAA-946) & 18 CFU/mL (ATCC 19615)
    Analytical Reactivity (Inclusivity)100% detection of tested Streptococcus pyogenes strains100% detection of 24 Strep A strainsAll 24 strains correctly reported as DETECTED.
    Analytical Specificity (Exclusivity)100% non-detection of non-target organisms100% non-detection of 70 potentially cross-reactive microorganismsAll 70 organisms correctly reported as NOT DETECTED.
    Microbial InterferenceNo interference with Strep A detectionNo interference from 27 tested microorganisms
    Potentially Interfering SubstancesNo assay interferenceNo interference from 10 tested substances
    Carry-Over ContaminationNo evidence of carry-overAll 42 negative samples correctly reported as NOT DETECTED. All 40 positive samples correctly reported as DETECTED.
    Reproducibility (Negative)100% agreement expected100% (144/144) agreement
    Reproducibility (Low Positive)High agreement expected (e.g., >95%)98.6% (142/144) agreement
    Reproducibility (Moderate Positive)100% agreement expected100% (144/144) agreement

    Study Proving Acceptance Criteria (Clinical Performance):

    • Study Design: A multi-site clinical study that collected throat ESwab specimens from patients with signs and symptoms of pharyngitis. The study combined data from two approaches:
      • One study collected a second prospective throat swab after a standard of care (SOC) swab.
      • Another study used leftover excess SOC throat swab specimens.

    2. Sample Sizes and Data Provenance for the Test Set:

    • Initial Enrolled Specimens: 844
    • Excluded Specimens: 261 (due to inclusion criteria failure, reference culture procedural error, delay in reference culture inoculation, delay in shipment, or labeling error).
    • Specimens Included in Performance Analysis (Test Set): 583
      • Successful on initial test: 565/583 (96.9%)
      • Valid results after retest (overall): 577/583 (99.0%)
    • Data Provenance: Geographically diverse regions within the United States. The study was conducted between December 2016 and March 2017, suggesting it was a prospective or mixed (prospective and retrospective for leftover samples) collection. The text states "one study enrolled consented subjects from whom a second prospective throat swab specimen was collected" and "another study tested specimens from subjects for which leftover excess standard of care (SOC) throat swab specimens were available."

    3. Number of Experts and Qualifications for Ground Truth for the Test Set:

    • The document does not specify the number of experts or their qualifications for establishing the initial ground truth (culture and latex agglutination). These are standard laboratory procedures, but details about expert reviewers for discordant results are mentioned.

    4. Adjudication Method for the Test Set:

    • Discordant results between the Xpert Xpress Strep A Assay and the reference method (culture) were investigated.
    • Method: An alternative PCR/bidirectional sequencing assay was used for adjudication. The results of this alternative PCR were footnoted in the performance tables (e.g., for 26 discrepant samples in Table 8-7, 21 were confirmed positive by alternative PCR, 4 negative, and 1 not tested). This indicates an independent molecular method was used to resolve discrepancies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) assay, not an imaging device or one that involves human "readers" interpreting results in a variable way that MRMC studies are designed for. Its performance is evaluated biochemically against a reference standard.

    6. Standalone Performance Study (Algorithm only without human-in-the-loop):

    • Yes, a standalone study was performed. The clinical performance data presented (Sensitivity, Specificity, PPV, NPV) represents the performance of the Xpert Xpress Strep A Assay (the "algorithm/device") functioning independently relative to a microbiological reference method (culture). The results are automatically generated by the GeneXpert Instrument Systems.

    7. Type of Ground Truth Used:

    • For the clinical performance study (test set), the primary ground truth reference method was culture and latex agglutination for Strep A typing.
    • For resolving discordant results, an alternative PCR/bidirectional sequencing assay was used.

    8. Sample Size for the Training Set:

    • The document does not specify the sample size for a "training set" in the context of an algorithm. For IVDs, the development process typically involves various internal testing and optimization (which could be considered analogous to training) but not usually a distinct "training set" of patient specimens in the same way an AI model would have. The document focuses on analytical and clinical validation studies.

    9. How Ground Truth for the Training Set Was Established:

    • As a training set is not explicitly defined in the context of this IVD device's approval process in this document, the method for establishing its ground truth is not applicable/not provided. The analytical and clinical validation studies use established reference methods as ground truth.