510(k) Data Aggregation
(397 days)
South Korea
Re: K243449
Trade/Device Name: INFINITT DPS
Regulation Number: 21 CFR 864.3700
Classification Product Code: QKQ
Classification Regulation: 21 CFR 864.3700
**Classification
Classification Product Code: PSY
Regulation Number: 21 CFR 864.3700
Device Class: Class
devices. |
| Device Class | Class II | Class II | No difference |
| Regulation Number | 21 CFR 864.3700
| 21 CFR 864.3700 | No difference |
| Classification Name | Whole Slide Imaging System | Whole Slide
INFINITT DPS is a software device intended for viewing and management of whole slide digital images derived from scanned surgical pathology slides prepared from formalin-fixed, paraffin-embedded (FFPE) tissue. It serves as an aid for pathologists to review and interpret these digital images for the purpose of pathology primary diagnosis.
INFINITT DPS is intended for use with Hamamatsu NanoZoomer S360MD scanner and Barco MDPC-8127 display.
It is the responsibility of the pathologist to implement appropriate procedures and safeguards that assure the integrity and accuracy of image interpretation when utilizing the INFINITT DPS. The system is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.
INFINITT DPS software provides a more efficient and convenient alternative to reviewing slides with physical microscopes. The primary users of INFINITT DPS are pathologists in histopathology, as well as technologists who handle specimens. Users can access whole slide images directly through a web browser, with no need to install additional software on the client PC.
INFINITT DPS, version 1.1, is a web-based software device that enables qualified pathologists to review digital pathology images (WSIs) of formalin-fixed, paraffin-embedded (FFPE) tissue specimens for pathology primary diagnosis. The device operates through a supported web browser without installing any application locally.
The Operation and Workflow of INFINITT DPS are described below:
- Data Input
- The device accepts WSI files in the NDPI file format generated by a Hamamatsu NanoZoomer S360MD Slide Scanner. The WSI files are internally converted into the Digital Imaging and Communications in Medicine (DICOM) file format using the built-in INFINITT DPS Acquisition module.
- Data Management and Storage
- The converted images are stored on the INFINITT DPS server. The user is highly recommended to back up the original NDPI files. The device can be integrated with an external LIS (Laboratory Information System)/EMR (Electronic Medical Record) system to automate case registration and retrieval.
- Image Access and Navigation
- Users can log in to the device using the INFINITT DPS WebViewer running on a supported web browser (i.e., Chrome, Edge, or Firefox). The stored WSIs can be searched, selected, and opened directly within the INFINITT DPS WebViewer.
- Image Review and Interpretation
- INFINITT DPS WebViewer provides the following interactive features:
- Zoom, pan, and rotate for image navigation
- Measurement tools for length and area analysis
- INFINITT DPS WebViewer provides the following interactive features:
The interoperable components of INFINITT DPS and other system specifications are provided in tables 1 – 3 below.:
Table 1. Interoperable Components for Use with INFINITT DPS
| Components | Manufacturer | Model |
|---|---|---|
| Scanner | Hamamatsu | NanoZoomer S360MD Slide scanner |
| Display | Barco | MDPC-8127 |
Table 2. Computer Environment / System for Use with INFINITT DPS
| Component | Requirement |
|---|---|
| Operating System | Windows 11 |
| Memory | 32GB RAM |
| Processor | Intel Core i7-11800H |
| Supported Browsers | Google Chrome version 90.0 and above |
| Microsoft Edge version 90.0 and above | |
| Mozilla Firefox version 80.0 and above |
Table 3. Server System Requirements
| Component | Web Server | Database Server |
|---|---|---|
| CPU | 2.8GHz 12-core or higher | 2.2GHz 8-core or higher |
| Memory | 64GB or more | 16GB or more |
| Storage | 2TB or more | 4TB or more |
| Software Platforms | Microsoft IIS (Internet Information Services) 10 or higher | Not Found |
| Operating Systems | Microsoft Windows Server 2016 or higher | Not Found |
| Database | N/A | Oracle Database 19c or higher |
| Network | 1Gbps | 1Gbps |
N/A
Ask a specific question about this device
(88 days)
Albuquerque, NM 87114
Re: K252762
Trade/Device Name: HALO AP Dx
Regulation Number: 21 CFR 864.3700
Version Number: 2.4
Regulation Name: Whole Slide Imaging System
Regulation: 21 CFR §864.3700
HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the interoperable components specified in the Table below.
HALO AP Dx, version 2.4 (abbreviated as v2.4), is a browser-based software-only device intended to aid pathology professionals in viewing, manipulating, management, and interpretation of digital pathology whole slide images (WSI) of glass slides obtained from the Hamamatsu Photonics K.K. NanoZoomer S360MD scanner or the Leica Biosystems Imaging, Inc. Aperio GT 450 DX.
HALO AP Dx is typically operated as follows:
-
Image acquisition is performed using the predicate device. The operator performs quality control of the digital slides per the instructions of the predicate device and lab specifications to determine if re-scans are necessary.
-
Once image acquisition is complete, according to its Instructions for Use, the unaltered image is saved in an external image storage location. HALO AP Dx ingests the image, and a copy of image metadata is stored in the subject device's database to improve viewing response times.
-
Depending upon a laboratory's workflow, the scanned images may first be reviewed by histotechnicians to confirm image quality and initiate any re-scans. After review, the histotechnician may modify the case status and make it available to the pathologist.
-
The reading pathologist selects a patient case from a selected worklist within HALO AP Dx whereby the subject device fetches the associated images from external image storage.
-
The reading pathologist uses the subject device to view the images and can perform the following actions, as needed:
a. Zoom and pan the image.
b. Measure distances and areas in the image.
c. Annotate images.
d. View multiple images side by side in a synchronized fashion. -
The above steps are repeated as necessary.
After viewing all images belonging to a particular case (patient), the pathologist will make a diagnosis which is documented in another system, such as a Laboratory Information System.
HALO AP Dx operates and is validated for use with the FDA cleared components listed in the intended use statement table above.
N/A
Ask a specific question about this device
(238 days)
MA 02215
June 26, 2025
Re: K243391
Trade/Device Name: AISight Dx
Regulation Number: 21 CFR 864.3700
Version:** V 2.9
Classification Name: Whole Slide Imaging System
Regulation Number: 21 CFR 864.3700
scanner system | Aperio GT 450 DX | AISight Dx |
| Product Code | PSY | PSY | QKQ |
| Regulation | 21 CFR 864.3700
| 21 CFR 864.3700 | 21 CFR 864.3700 |
| Regulation Name | Whole Slide Imaging System | Whole Slide Imaging
AISight Dx is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret, and manage digital images of these slides for primary diagnosis. AISight Dx is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. AISight DX is intended to be used with interoperable displays, scanners and file formats, and web browsers that have been 510(k) cleared for use with the AISight Dx or 510(k)-cleared displays, 510(k)-cleared scanners and file formats, and web browsers that have been assessed in accordance with the Predetermined Change Control Plan (PCCP) for qualifying interoperable devices.
AISight Dx is a web-based, software-only device that is intended to aid pathology professionals in viewing, interpretation, and management of digital whole slide images (WSI) of scanned surgical pathology slides prepared from formalin-fixed, paraffin-embedded (FFPE) tissue obtained from Hamamatsu NanoZoomer S360MD Slide scanner or Leica Aperio GT 450 DX scanner (Table 1). It aids the pathologist in the review, interpretation, and management of pathology slide digital images used to generate a primary diagnosis.
Here's a breakdown of the acceptance criteria and the study details for the AISight Dx device, based on the provided FDA 510(k) Clearance Letter:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Pixel-wise Comparison | Identical image reproduction (max pixelwise difference < 1 CIEDE2000) | Maximum pixelwise difference was 0 CIEDE2000, indicating pixelwise identical output images. Meets criteria. |
| Non-inferiority to Glass Slide Reads (Major Discordance Rate - Hamamatsu Scanner) | Upper limit of 95% CI for difference in major discordance rate (MD vs. GT vs. MO vs. GT) less than 4%. | Upper limit of 95% CI was 1.16%. Meets criteria. |
| Non-inferiority to Glass Slide Reads (Major Discordance Rate - Leica Scanner) | Upper limit of 95% CI for difference in major discordance rate (MD vs. GT vs. MO vs. GT) less than 4%. | Upper limit of 95% CI was 2.52%. Meets criteria. |
| Turnaround Time | Adequate for intended use (image processing, loading, panning, zooming). | Test results showed these to be adequate for the intended use. Meets criteria. |
| Measurement Accuracy | Accurate distance and area measurements. | Tests verified that distances and areas measured in AISight Dx accurately reflected those on a calibrated slide. Meets criteria. |
| Human Factors | Safe and effective for intended users, uses, and use environments. | AISight Dx has been found to be safe and effective for the intended users, uses, and use environments. Meets criteria. |
Study Details
-
Sample size used for the test set and the data provenance:
- The document states that two separate clinical studies were conducted, one for each scanner (Hamamatsu NanoZoomer S360MD and Leica Aperio GT 450 DX).
- The sample sizes for these clinical studies are not explicitly stated in the provided text.
- Data Provenance: Not explicitly mentioned, but the study compares performance against "the original sign-out pathologic diagnosis using MO [ground truth, (GT)] rendered at the institution," suggesting the data is derived from clinical practice, likely retrospective or a mix, given the "original sign-out" aspect. The country of origin is not specified.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The study involved "3 reading pathologists" for assessing the differences in major discordance rates.
- Qualifications of experts: Not explicitly stated, but they are referred to as "reading pathologists," indicating they are qualified to make primary diagnoses.
-
Adjudication method for the test set:
- The "reference (main) diagnosis" was the "original sign-out pathologic diagnosis using MO [ground truth, (GT)] rendered at the institution."
- The document implies that this "original sign-out" acted as the ground truth. There's no explicit mention of an adjudication process (e.g., 2+1, 3+1 consensus) to establish this ground truth beyond the initial clinical diagnosis. The major discordance rate was calculated between MD (manual digital read), MO (manual optical read), and GT (original sign-out).
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Yes, an MRMC-like study was done as it involved "3 reading pathologists" evaluating cases using both manual digital (MD) and manual optical (MO) methods.
- Effect size of improvement with AI vs without AI assistance: This study did not measure the improvement of human readers with AI assistance. The AISight Dx is presented as a viewer and management software, not an AI-assisted diagnostic tool. The study aimed to demonstrate non-inferiority of digital viewing (MD) versus traditional optical viewing (MO) for primary diagnosis, where the software is simply the viewing platform, not an aid in interpretation itself. Therefore, no "effect size of human readers improving with AI vs without AI assistance" is reported because the device is not described as providing AI assistance for diagnostic tasks in this context.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. The AISight Dx is explicitly described as "an aid to the pathologist to review, interpret, and manage digital images." The clinical study evaluated "manual digital read (MD)" which is a human pathologist reading digital slides using the AISight Dx, compared to "manual optical (MO)" which is a human pathologist reading glass slides. The device is not an autonomous AI diagnostic algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth (GT) for the clinical study was the original sign-out pathologic diagnosis using manual optical microscopy (MO) rendered at the institution. This can be categorized as a form of expert (pathologist) ground truth based on clinical practice/standard of care.
-
The sample size for the training set:
- The document for AISight Dx does not mention a training set size. This is expected as AISight Dx is described as a viewing and management software, not an AI model that requires a training set for diagnostic capabilities. The performance data focuses on its function as a display and interpretation platform for human pathologists.
-
How the ground truth for the training set was established:
- As there's no mention of a training set for an AI model within the AISight Dx software (it's a viewer), there's no information on how a ground truth for such a set would be established.
Ask a specific question about this device
(81 days)
Re: K250968**
Trade/Device Name: PathPresenter Clinical Viewer
Regulation Number: 21 CFR 864.3700
Version:** V1.0.1
Classification Name: Whole Slide Imaging System
Regulation Number: 21 CFR 864.3700
Slide scanner system
Submission Number: K233027
Device Class: Class II
CFR Section: 864.3700
For In Vitro Diagnostic Use
The PathPresenter Clinical Viewer is a software intended for viewing and managing whole slide images of scanned glass sides derived from formalin fixed paraffin embedded (FFPE) tissue. It is an aid to pathologists to review and render a diagnosis using the digital images for the purposes of primary diagnosis. PathPresenter Clinical is not intended for use with frozen sections, cytology specimens, or non-FFPE specimens. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using PathPresenter Clinical software. PathPresenter Clinical Viewer is intended for use with Hamamatsu NanoZoomer S360MD Slide scanner NDPI image formats viewed on the Barco NV MDPC-8127 display device.
The PathPresenter Clinical Viewer (version V1.0.1) is a web-based software application designed for viewing and managing whole slide images generated from scanned glass slides of formalin-fixed, paraffin-embedded (FFPE) surgical pathology tissue. It serves as a diagnostic aid, enabling pathologists to review digital images and render a primary pathology diagnosis. Functions of the viewer include zooming and panning the image, annotating the image, measuring distances and areas in the image and retrieving multiple images from the slide tray including prior cases and deprecated slides.
Here's a breakdown of the acceptance criteria and study information for the PathPresenter Clinical Viewer based on the provided FDA 510(k) clearance letter:
Acceptance Criteria and Device Performance for PathPresenter Clinical Viewer
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Pixelwise Comparison | The 95th percentile of the pixel-wise color difference in any image pair is less than 3 CIEDE2000 (< 3 ΔE00) when comparing the PathPresenter Clinical Viewer (Subject Viewer) with the Hamamatsu NanoZoomer S360MD Slide scanner with NZViewMD viewer (Predicate Viewer). | The device demonstrates substantial equivalence, with the 95th percentile of the pixel-wise color difference being less than 3 CIEDE2000 (< 3 ΔE00) for all comparisons (PathPresenter Clinical Viewer on Microsoft Edge vs. Predicate, and PathPresenter Clinical Viewer on Google Chrome vs. Predicate). |
| Turnaround Time (TAT) Study - Image Loading | Loading of the first image visible to the user: ≤ 8 seconds | Actual: 2.72 seconds |
| Turnaround Time (TAT) Study - Panning | Loading of the whole field of view after panning: ≤ 2 seconds | Actual: 1.22 seconds |
| Turnaround Time (TAT) Study - Zooming | Loading of the whole field of view after zooming: ≤ 3 seconds | Actual: 0.60 seconds |
| Measurement Accuracy | All measured values match predetermined measurements relevant to the zoom level of the viewer, with no allowable deviation. | All measured values matched the reference values exactly, with zero observed error across multiple magnification settings for both length and area measurements. |
| Human Factors Study | Critical tasks required for operation are performed accurately and without any use-related errors that could result in patient harm or diagnostic inaccuracies. Device meets HFE/UE requirements and is acceptable for deployment use in clinical settings. | Validation results confirmed that critical tasks were performed accurately and without any use-related errors that could result in patient harm or diagnostic inaccuracies. Observed usability issues were considered easily mitigable through design improvements, training, and labeling, posing no unacceptable risk. |
2. Sample Size Used for the Test Set and Data Provenance
- Pixelwise Comparison:
- Sample Size: Images from 30 FFPE tissue glass slides, with 3 Regions of Interest (ROIs) selected per slide, and 2 zoom levels (20x and 40x) per ROI. This resulted in 180 total image pairs (30 slides x 3 ROIs x 3 Zoom Levels [implied, likely meaning 20x, 40x, and perhaps an overview] x 2 Browsers).
- Data Provenance: Not specified, but generally regulatory submissions imply real-world pathology samples. It is not explicitly stated whether the data was retrospective or prospective, nor the country of origin.
- Turnaround Time Study:
- Sample Size: Minimum of 20 slides.
- Data Provenance: Not specified.
- Measurement Accuracy:
- Sample Size: An image of a calibration scale slide with objects of known sizes.
- Data Provenance: Not specified.
- Human Factors Study:
- Sample Size: "Representative users" (board-certified pathologists). No specific number provided.
- Data Provenance: Not specified.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Pixelwise Comparison: One board-certified Pathologist was used to pre-identify the Regions of Interest (ROIs). Their specific years of experience are not mentioned beyond "board-certified Pathologist." This pathologist was blinded to the rest of the pixel-wise comparison study.
- Measurement Accuracy: The ground truth was established by using an image of a calibration scale slide with objects of known sizes. This does not involve human experts for ground truth establishment.
- Human Factors Study: "Board-certified pathologists" served as representative users, performing critical tasks. They were the subjects of the study, not necessarily establishing ground truth for device performance but rather testing usability and safety in a clinical context.
4. Adjudication Method for the Test Set
- The document does not explicitly describe an adjudication method for establishing ground truth for the pixelwise comparison or measurement accuracy tests.
- For the pixelwise comparison, the individual pathologist selected ROIs, but the automated comparison of pixel differences is a technical measurement against a predicate, not requiring multi-expert adjudication.
- For measurement accuracy, it was a direct comparison against known values from a calibration slide.
- The Human Factors study involved observations of user performance, not adjudication of diagnostic interpretations.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Reader Improvement with AI vs. Without AI Assistance
- No, an MRMC comparative effectiveness study was not performed, or at least not described in this summary. The PathPresenter Clinical Viewer is described as a "software intended for viewing and managing whole slide images" and an "aid to pathologists." It is a viewer, not an AI diagnostic algorithm that provides an independent reading or an assistive output that would typically be evaluated in an MRMC study comparing human performance with and without AI assistance. The studies focused on technical equivalence to a predicate viewer and usability.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- The document describes standalone performance tests for certain aspects of the software, particularly the pixelwise comparison and measurement accuracy. These tests evaluated the software's output (pixel rendering, measurement values) directly against a known reference (predicate viewer, calibration slide) without human interpretation as part of the core evaluation criteria.
- However, the overall device is an "aid to pathologists" implying human-in-the-loop use. There is no mention of the software making diagnostic recommendations or interpretations by itself.
7. The Type of Ground Truth Used
- Pixelwise Comparison: The ground truth for comparison was the image output from the predicate device (Hamamatsu NanoZoomer S360MD Slide scanner with NZViewMD viewer). The "ground truth" here is the established rendering quality of the predicate viewer.
- Turnaround Time Study: The ground truth involved established time targets for specific operations.
- Measurement Accuracy: The ground truth was the known measurements of objects on a calibration scale slide.
- Human Factors Study: The "ground truth" for success was the accurate and error-free performance of critical tasks by representative users, consistent with safety.
8. The Sample Size for the Training Set
- The provided document does not mention a training set for the PathPresenter Clinical Viewer. This is expected as the viewer is described as image management and viewing software, not an AI algorithm that learns from data.
9. How the Ground Truth for the Training Set Was Established
- As no training set is mentioned or implied for this device, information on how its ground truth was established is not applicable here.
Ask a specific question about this device
(269 days)
924-8566
Japan
Re: K242545
Trade/Device Name: RadiForce MX317W-PA
Regulation Number: 21 CFR 864.3700
Name: 30.5 inch (77.5 cm) Color LCD Monitor
- Classification Name: Whole Slide Imaging System (21 CFR 864.3700
or Usual Name: Digital Pathology Display - Classification Name: Whole Slide Imaging System (21 CFR 864.3700
RadiForce MX317W-PA is intended for in vitro diagnostic use to display digital images of histopathology slides acquired from IVD-labeled whole-slide imaging scanners and viewed using IVD-labeled digital pathology image viewing software that have been validated for use with this device.
RadiForce MX317W-PA is an aid to the pathologist and is used for review and interpretation of histopathology slides for the purposes of primary diagnosis. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using this product. The display is not intended for use with digital images from frozen section, cytology, or non- formalin-fixed, paraffin embedded (non-FFPE) hematopathology specimens.
RadiForce MX317W-PA is a color LCD monitor for viewing digital images of histopathology slides. The color LCD panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size is 4,096 x 2,160 pixels (8MP) with a pixel pitch of 0.1674 mm.
Since factory calibrated display modes, each of which is characterized by a specific tone curve, a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor. This helps ensure tone curves even if a display controller or workstation must be replaced or serviced.
"Patho" is for intended digital pathology use mode.
The provided FDA 510(k) clearance letter for the RadiForce MX317W-PA describes a display device for digital histopathology. It does not contain information about an AI/ML medical device. Therefore, a study proving the device meets acceptance criteria related to AI/ML performance (such as accuracy, sensitivity, specificity, MRMC studies, and ground truth establishment methods for large datasets) is not present in this document.
The document primarily focuses on the technical performance and equivalence of a display monitor to a predicate device. The "performance testing" section refers to bench tests validating display characteristics like spatial resolution, luminance, and color, not the clinical performance of an AI algorithm interpreting medical images.
Given the information provided, here's an analysis based on the actual content:
Based on the provided document, the RadiForce MX317W-PA is a display monitor, not an AI/ML medical device designed for image interpretation. Therefore, the acceptance criteria and study detailed below pertain to the display's technical performance and its equivalence to a predicate display, not to an AI algorithm's diagnostic accuracy.
1. Table of Acceptance Criteria and Reported Device Performance
The document states that "the display characteristics of the RadiForce MX317W-PA meet the pre-defined criteria when criteria are set." However, the exact numerical acceptance criteria for each bench test (e.g., minimum luminance, pixel defect limits) are not explicitly listed in the provided text. The document only lists the types of tests performed and states that the device "has display characteristics equivalent to those of the predicate device" and "meet the pre-defined criteria."
| Acceptance Criteria Category | Reported Device Performance Summary (as per document) |
|---|---|
| User controls (Modes & settings) | Performed, assumed met |
| Spatial resolution | Performed, assumed met, equivalent to predicate |
| Pixel defects | Performed, assumed met, equivalent to predicate |
| Artifacts | Performed, assumed met, equivalent to predicate |
| Temporal response | Performed, assumed met, equivalent to predicate |
| Maximum and minimum luminance | Performed, assumed met, equivalent to predicate |
| Grayscale | Performed, assumed met, equivalent to predicate |
| Luminance uniformity and Mura test | Performed, assumed met, equivalent to predicate |
| Stability of luminance and chromaticity response | Performed, assumed met, equivalent to predicate |
| Bidirectional reflection distribution function | Performed, assumed met, equivalent to predicate |
| Gray Tracking | Performed, assumed met, equivalent to predicate |
| Color scale | Performed, assumed met, equivalent to predicate |
| Color gamut volume | Performed, assumed met, equivalent to predicate |
Note: The document only states that these tests were performed and that the results show equivalence to the predicate device and that the device meets pre-defined criteria. It does not provide the specific numerical results or the exact numerical acceptance criteria for each test.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document describes bench tests performed on a single device, the RadiForce MX317W-PA (it's a physical monitor, not a software algorithm processing a dataset). There is no mention of a "test set" in the context of a dataset of medical images.
- Data Provenance: Not applicable. The "data" here refers to the measured performance characteristics of the physical display device itself during bench testing, not patient data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not applicable. The ground truth for a display monitor's technical performance is established by standardized measurement equipment and protocols, not by expert interpretation of images. The device itself is the object under test for its physical characteristics.
4. Adjudication Method for the Test Set
- Not applicable. This concept applies to human or AI interpretation of medical images, where discrepancies among readers or algorithms might need resolution. For physical device performance, measurements are generally objective.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not performed/Applicable. An MRMC study is designed to assess the performance of a diagnostic aid (like AI) on image interpretation by human readers. This device is a display monitor, not an AI algorithm. Its function is to display images, not to interpret them or assist human interpreters in a diagnostic decision-making process that would warrant an MRMC study.
6. Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Study
- Not applicable. As stated, this is a display monitor, not an algorithm.
7. Type of Ground Truth Used:
- The ground truth for the display's performance tests would be metrology-based standards and calibration references (e.g., standard luminance values, colorimetry standards) against which the display's output is measured. It is not expert consensus, pathology, or outcomes data, as these relate to diagnostic accuracy studies.
8. The Sample Size for the Training Set
- Not applicable. This device is hardware; it does not involve training data or machine learning algorithms.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. No training set exists for this device.
Ask a specific question about this device
(90 days)
Vaugirard
Paris, 75015
France
Re: K250414
Trade/Device Name: CaloPix
Regulation Number: 21 CFR 864.3700
Subject Device
Trade/Device name: CaloPix
Version: 6.1.0 IVDUS
Regulation number: 21 CFR 864.3700
| K232202 |
| Clearance date | September 27, 2022 | April 16, 2024 |
| Regulation number: | 21 CFR 864.3700
Slide scanner system | Aperio GT 450 DX |
| Product code | QKQ | PSY | PSY |
| Regulation | 21 CFR 864.3700
| 21 CFR 864.3700 | 21 CFR 864.3700 |
| Regulation name | Whole Slide Imaging System | Whole Slide Imaging
For In Vitro Diagnostic Use Only
CaloPix is a software only device for viewing and management of digital images of scanned surgical pathology slides prepared from Formalin-Fixed Paraffin Embedded (FFPE) tissue.
CaloPix is intended for in vitro diagnostic use as an aid to the pathologist to review, interpret and manage these digital slide images for the purpose of primary diagnosis.
CaloPix is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and the validity of the interpretation of images using CaloPix.
CaloPix is intended to be used with the interoperable components specified in the below Table:
| Scanner Hardware | Scanner Output File Format | Interoperable Displays |
|---|---|---|
| Leica Aperio GT 450 DX scanner | SVS | Dell U3223QE |
| Hamamatsu NanoZoomer S360MD Slide scanner | NDPI | JVC Kenwood JD-C240BN01A |
CaloPix, version 6.1.0 IVDUS, is a web-based software-only device that is intended to aid pathology professionals in viewing, interpreting and managing digital Whole Slide Images (WSI) of glass slides obtained from the Hamamatsu NanoZoomer S360MD slide scanner (NDPI file format) and viewed on the JVC Kenwood JD-C240BN01A display, as well as those obtained from the Leica Aperio GT 450 DX scanner (SVS file format) and viewed on the Dell U3223QE display.
CaloPix does not include any automated Image Analysis Applications that would constitute computer aided detection or diagnosis.
CaloPix is for viewing digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy.
As a whole, CaloPix is a pathology Image Management System (IMS) which brings case-centric digital pathology image management, collaboration, and image processing. CaloPix consists of:
-
Integration with Laboratory Information Systems (LIS): Allows to obtain automatically from the LIS patient data associated with the cases, scanned whole slide images and other related medical images to be analyzed. The data stored in the database is automatically updated according to the interface protocol with the LIS.
-
DataBase: After ingestion, scanned WSI can be organized in the CaloPix database consisting of folders (cases) containing patient identification data and examination results from a LIS.
Ingestion of the slides is performed through an integrated module that allows their automatic indexation based on patient data retrieved from the LIS. After their ingestion, image files are stored in a CaloPix-specific file storage environment, that can be on premises or in the cloud.
- The CaloPix viewer component to process scanned whole slide images, that includes functions for panning, zooming, screen capture, annotations, distance and surface measurement, and image registration. This viewer relies on image servers (IMGSRV) which extract image tiles from the whole slide image file and send these tiles to the CaloPix viewer for smooth and fast viewing.
The FDA 510(k) clearance letter for CaloPix indicates that the device's performance was evaluated through a series of tests to demonstrate its safety and effectiveness. The primary study described in the provided document focuses on technical performance testing rather than a clinical multi-reader multi-case (MRMC) study.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Pixel-wise comparison (Image Reproduction Accuracy) | The 95th percentile of the pixel-wise color differences (CIEDE2000, ΔE00) in any image pair between CaloPix and the predicate device's IRMS must be less than 3 (ΔE00 < 3). | Test results showed that the 95th percentile of pixel-wise differences between CaloPix and the comparators (NZViewMD and Aperio WebViewer DX) were less than 3 CIEDE2000. The output images were considered pixel-wise identical and visually adequate. |
| Measurements of Area and Distance | Accuracy of measurements made in CaloPix reviewer confirmed by comparing to predicate device viewers. (Specific numerical criteria not provided). | Tests verified that the distance and area measurements made in the CaloPix viewer accurately reflected the measurements made in the NZViewMD viewer and the Aperio WebViewer DX viewer. |
| Turnaround Time | - Image loading: ≤ 10 seconds- Panning: ≤ 7 seconds- Zooming: ≤ 7 seconds | Turnaround times for opening an image and panning or zooming have been determined and found to be adequate for the intended use of the subject device. (Specific measured times not provided, but stated to meet criteria). |
| Human Factors Validation Study | CaloPix to be safe and effective for the intended users, uses, and use environments (per FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices (2016)"). | CaloPix has been found to be safe and effective for the intended users, uses and use environments. |
2. Sample Size and Data Provenance for Test Set
- Sample Size for Pixel-wise Comparison: 30 whole slide images (WSIs). Specifically:
- 25 H&E-stained (Formalin-Fixed Paraffin Embedded - FFPE) tissue glass slides.
- 5 IHC-stained (Masson's trichrome stain, CD8, CD3, or CD20) FFPE tissue glass slides.
- Data Provenance: Not explicitly stated regarding country of origin. The study appears to be a technical performance test using pre-scanned slides. It is unclear if these were from a retrospective or prospective collection for the purpose of this specific test. The mention of "FFPE tissue glass slides" implies clinical relevance.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: One pathologist.
- Qualifications of Experts: The text states, "as verified by a pathologist," but does not specify the pathologist's experience level (e.g., years of experience, board certification). This was for identifying "relevant pathological features" for ROIs.
4. Adjudication Method for Test Set
- No "adjudication method" in the sense of multiple readers coming to a consensus for diagnostic ground truth is described. The ground truth for the pixel-wise comparison was based on pre-identified Regions of Interest (ROIs) verified by a single pathologist. For the measurement and turnaround time tests, the comparison was against predicate device software performance.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done comparing human reader performance with and without AI assistance. The described tests are technical performance evaluations of the viewing and management software's rendering accuracy, measurement functions, and speed.
6. Standalone (Algorithm Only) Performance
- This device, CaloPix, is described as a "software only device for viewing and management of digital images." It "does not include any automated Image Analysis Applications that would constitute computer aided detection or diagnosis." Therefore, no standalone (algorithm only) performance, as typically understood for an AI/CADe/CADx device, was conducted or is applicable since the device's function is solely as a viewing platform, not an analytical algorithm.
7. Type of Ground Truth Used
- Technical Ground Truth: For the pixel-wise comparison, the "ground truth" was established by comparing CaloPix's image rendering output against the image rendering of the predicate device's Image Review Manipulation Software (IRMS) (i.e., NZViewMD and Aperio WebViewer DX), with the acceptance criteria based on CIEDE2000 values. The relevant pathological features within the slides were identified by a pathologist for ROI selection.
- For measurements, the ground truth was the measurements obtained from the predicate device's IRMS.
- For turnaround time, the ground truth was the pre-defined time limits.
- For human factors, the ground truth was the satisfaction of usability and safety criteria.
8. Sample Size for the Training Set
- Not Applicable / Not Provided: As CaloPix is a viewing and management software without AI/CAD algorithms, there is no mention of a "training set" in the context of machine learning model development. The tests described are functional and technical performance evaluations, not AI model validation.
9. How Ground Truth for Training Set was Established
- Not Applicable / Not Provided: See point 8.
Ask a specific question about this device
(226 days)
UT 84043
Re: K242244
Trade/Device Name: Viewer+ Regulation Number: 21 CFR 864.3700 Regulation Name:
|
| Regulation Number: | 21 CFR 864.3700 |
| Product Code:
For In Vitro Diagnostic Use
Viewer+ is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides for primary diagnosis. Viewer+ is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. Viewer+ is intended for use with Hamamatsu NanoZoomer S360MD Slide scanner and BARCO MDPC-8127 display.
Viewer+, version 1.0.1, is a web-based software device that facilitates the viewing and navigating of digitized pathology images of slides prepared from FFPE-tissue specimens acquired from Hamamatsu NanoZoomer S360MD Slide scanner and viewed on BARCO MDPC-8127 display. Viewer+ renders these digitized pathology images for review, management, and navigation for pathology primary diagnosis.
Viewer+ is operated as follows:
-
- Image acquisition is performed using the NanoZoomer S360MD Slide scanner according to its Instructions for Use. The operator performs quality control of the digital slides per the instructions of the NanoZoomer and lab specifications to determine if re-scans are necessary.
-
- Once image acquisition is complete and the image becomes available in the scanner's database file system, a separate medical image communications software (not part of the device) automatically uploads the image and its corresponding metadata to persistent cloud storage. Image and data integrity checks are performed during the upload to ensure data accuracy.
-
- The subject device enables the reading pathologist to open a patient case, view the images, and perform actions such as zooming, panning, measuring distances and areas, and annotating images as needed. After reviewing all images for a case, the pathologist will render a diagnosis.
Here's a breakdown of the acceptance criteria and the study details for the Viewer+ device, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Pixel-wise comparison (of images reproduced by Viewer+ and NZViewMD for the same file generated from NanoZoomer S360md Slide Scanner) | The 95th percentile of pixel-wise differences between Viewer+ and NZViewMD was less than 3 CIEDE2000, indicating their output images are pixel-wise identical and visually adequate. |
| Turnaround time (for opening, panning, and zooming an image) | Found to be adequate for the intended use of the device. |
| Measurement accuracy (using scanned images of biological slides) | Viewer+ was found to perform accurate measurements with respect to its intended use. |
| Usability testing | Demonstrated that the subject device is safe and effective for the intended users, uses, and use environments. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample size of images or cases used for the "Test Set" in the performance studies. It mentions "scanned images of the biological slides" for measurement accuracy and "images reproduced by Viewer+ and NZViewMD for the same file" for pixel-wise comparison.
The data provenance (country of origin, retrospective/prospective) is also not specified in the provided text.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not specify the number of experts or their qualifications used to establish ground truth for the test set. It mentions that the device is "an aid to the pathologist" and that "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision." However, this relates to the intended use and not a specific part of the performance testing described.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1) used for establishing ground truth or evaluating the test set results. The pixel-wise comparison relies on quantitative color differences, and usability is assessed according to FDA guidance.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance is mentioned or implied in the provided text. The device is a "viewer" and not an AI-assisted diagnostic tool that would typically involve such a study.
6. Standalone Performance (Algorithm Only without Human-in-the-Loop)
The performance tests described (pixel-wise comparison, turnaround time, measurements) primarily relate to the technical functionality of the Viewer+ software itself, which is a viewing and management tool. These tests can be interpreted as standalone assessments of the software's performance in rendering images and providing basic functions like measurements. However, it's crucial to note that Viewer+ is an "aid to the pathologist" and not intended to provide automated diagnoses without human intervention. The "standalone" performance here refers to its core functionalities as a viewer, not as an autonomous diagnostic algorithm.
7. Type of Ground Truth Used
- Pixel-wise comparison: The ground truth for this test was the image reproduced by the predicate device's software (NZViewMD) for the same scanned file. The comparison was quantitative (CIEDE2000).
- Measurements: The ground truth would likely be established by known physical dimensions on the biological slides, verified by other means, or through precise calibration. The document states "Measurement accuracy has been verified using scanned images of the biological slides."
- Usability testing: The ground truth here is the fulfillment of usability requirements and user satisfaction/safety criteria, as assessed against FDA guidance.
8. Sample Size for the Training Set
The document does not mention the existence of a "training set" in the context of the Viewer+ device. This is a software-only device for viewing and managing images, not an AI/ML algorithm that typically requires a training set for model development.
9. How the Ground Truth for the Training Set Was Established
As no training set is mentioned for this device, information on how its ground truth was established is not applicable.
Ask a specific question about this device
(79 days)
Re: K243871
Trade/Device Name: Philips IntelliSite Pathology Solution 5.1 Regulation Number: 21 CFR 864.3700
IntelliSite Pathology Solution 5.1 510(k) number: K243871 Device Class: Class II Classification regulation: 864.3700
Philips IntelliSite Pathology Solution 5.1 510(k) number: K242848 Device Class: Class II CFR section: 864.3700
The Philips IntelliSite Pathology Solution (PIPS) 5.1 is an automated digital slide creation, viewing, and management system. The PIPS 5.1 is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The PIPS 5.1 is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
The PIPS 5.1 comprises the Imagement System (IMS) 4.2, Ultra Fast Scanner (UFS), Pathology Scanner SG20. Pathology Scanner SG60, Pathology Scanner SG300 and Philips PP27QHD display, a Beacon C411W display or a Barco MDCC-4430 display. The PIPS 5.1 is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using PIPS 5.1.
The Philips IntelliSite Pathology Solution (PIPS) 5.1 is an automated digital slide creation, viewing, and management system. PIPS 5.1 consists of two subsystems and a display component:
-
- A scanner in any combination of the following scanner models
- . Ultra Fast Scanner (UFS)
- Pathology Scanner SG with different versions for varying slide capacity . Pathology Scanner SG20, Pathology Scanner SG60, Pathology Scanner SG300
-
- Image Management System (IMS) 4.2
-
- Clinical display
- PP27QHD or C411W or MDCC-4430 .
PIPS 5.1 is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. The PIPS does not include any automated image analysis applications that would constitute computer aided detection or diagnosis. The pathologists only view the scanned images and utilize the image review manipulation software in the PIPS 5.1.
This document is a 510(k) summary for the Philips IntelliSite Pathology Solution (PIPS) 5.1. It describes the device, its intended use, and compares it to a legally marketed predicate device (also PIPS 5.1, K242848). The key change in the subject device is the introduction of a new clinical display, Barco MDCC-4430.
Here's the breakdown of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The submission focuses on demonstrating substantial equivalence of the new display (Barco MDCC-4430) to the predicate's display (Philips PP27QHD). The acceptance criteria are largely derived from the FDA's "Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices" (TPA Guidance) and compliance with international consensus standards. The performance is reported as successful verification showing equivalence.
| Acceptance Criteria (TPA Guidance 항목) | Reported Device Performance (Subject Device with Barco MDCC-4430) | Conclusion on Substantial Equivalence |
|---|---|---|
| Display type | Color LCD | Substantially equivalent: Minor difference in physical display size is a minor change and does not raise any questions of safety or effectiveness. |
| Manufacturer | Barco N.V. | Same as above. |
| Technology | IPS technology with a-Si Thin Film Transistor (unchanged from predicate) | Substantially equivalent: Proposed and predicate device are considered substantially equivalent. |
| Physical display size | 714 mm x 478 mm x 74 mm | Substantially equivalent: Minor change, does not raise safety/effectiveness questions. |
| Active display area | 655 mm x 410 mm (30.4 inch diagonal) | Substantially equivalent: Slightly higher viewable area is a minor change. Verification testing confirms image quality is equivalent to the predicate device. |
| Aspect ratio | 16:10 | Substantially equivalent: This change does not raise any new concerns on safety and effectiveness. Proposed and predicate device are considered substantially equivalent. |
| Resolution | 2560 x 1600 pixels | Substantially equivalent: Slightly higher resolution and pixel size is a minor change. Verification testing confirms image quality is equivalent to the predicate device. Conclusion: This change does not raise any new concerns on safety and effectiveness. Proposed and predicate device are considered substantially equivalent. |
| Pixel Pitch | 0.256 mm x 0.256 mm | Same as above. |
| Color calibration tools (software) | QAWeb Enterprise version 2.14.0 installed on the workstation | Substantially equivalent: New display uses different calibration software, but calibration method (built-in front sensor), calibration targets, and frequency of quality control tests remain unchanged. Conclusion: This change does not raise new safety/effectiveness concerns. |
| Color calibration tools (hardware) | Built-in front sensor (same as predicate) | Same as above. |
| Additional Non-clinical Performance Tests (TPA Guidance) | Verification that technological characteristics of the display were not affected by the new panel, including: Spatial resolution, Pixel defects, Artifacts, Temporal response, Maximum and minimum luminance, Grayscale, Luminance uniformity, Stability of luminance and chromaticity, Bidirectional reflection distribution function, Grav tracking, Color scale response, Color gamut volume. | Conclusion: Verification for the new display showed that the proposed device has similar technological characteristics compared to the predicate device following the TPA guidance. In compliance with international/FDA-recognized consensus standards (IEC 60601-1, IEC 60601-1-6, IEC 62471, ISO 14971). Safe and effective, conforms to intended use. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a "sample size" in terms of cases or images for the non-clinical performance tests. The tests were performed on "the display of the proposed device" to verify its technological characteristics. This implies testing on representative units of the Barco MDCC-4430 display.
The data provenance is not specified in terms of country of origin or retrospective/prospective, as the tests were bench testing (laboratory-based performance evaluation of the display hardware) rather than clinical studies with patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and their Qualifications
This information is not applicable to this submission. The tests performed were technical performance evaluations of hardware (the display), not clinical evaluations requiring expert interpretation of medical images. Ground truth for these technical tests would be established by objective measurements against specified technical standards and parameters.
4. Adjudication Method for the Test Set
This information is not applicable to this submission. As the tests were technical performance evaluations of hardware, there would not be an adjudication process involving multiple human observers interpreting results in the same way there would be for a clinical trial.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done.
The submission explicitly states: "The proposed device with the new display did not require clinical performance data since substantial equivalence to the currently marketed predicate device was demonstrated with the following attributes: Intended Use / Indications for Use, Technological characteristics, Non-clinical performance testing, and Safety and effectiveness."
Therefore, there is no effect size reported for human readers with and without AI assistance, as AI functionality for diagnostic interpretation is not the subject of this 510(k) (the PIPS 5.1 "does not include any automated image analysis applications that would constitute computer aided detection or diagnosis").
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. The PIPS 5.1 is a digital slide creation, viewing, and management system, not an AI algorithm for diagnostic interpretation. The focus of this 510(k) is the display component. The device itself is designed for human-in-the-loop use by a pathologist.
7. The Type of Ground Truth Used
For the non-clinical performance data, the "ground truth" was based on:
- International and FDA-recognized consensus standards: This includes IEC 60601-1, IEC 60601-1-6, IEC 62471, and ISO 14971.
- TPA Guidance: The "Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices" guidance document, which specifies technical parameters for displays.
- Predicate device characteristics: Demonstrating that the new display's performance matches or is equivalent to the legally marketed predicate device's display across various technical parameters.
In essence, the ground truth was established by engineering specifications, technical performance targets, and regulatory standards for display devices.
8. The Sample Size for the Training Set
This information is not applicable. The PIPS 5.1, as described, is a system for digital pathology, not an AI algorithm that requires a training set of data. The 510(k) specifically mentions: "The PIPS does not include any automated image analysis applications that would constitute computer aided detection or diagnosis." Therefore, there is no AI training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable, as there is no AI training set.
Ask a specific question about this device
(259 days)
: K241717
Trade/Device Name: Epredia E1000 Dx Digital Pathology Solution Regulation Number: 21 CFR 864.3700
Classification Name: | Whole Slide Imaging System |
| Regulation Number: | 21 CFR 864.3700
of the E1000 Dx Digital Pathology Solution in accordance with the special controls listed in 21 CFR 864.3700
The Epredia E1000 Dx Digital Pathology Solution is an automated digital slide creation, viewing, and management system. The Epredia E1000 Dx Digital Pathology Solution is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Epredia E1000 Dx Digital Pathology Solution is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
The Epredia E1000 Dx Digital Pathology Solution consists of a Scanner (E1000 Dx Digital Pathology Scanner), which generates in MRXS image file format, E1000 Dx Scanner Software, Image Management System (E1000 Dx IMS), E1000 Dx Viewer Software, and Display (Barco MDPC-8127). The Epredia E1000 Dx Digital Pathology Solution is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using Epredia E1000 Dx Digital Pathology Solution.
The E1000 Dx Digital Pathology Solution is a high-capacity, automated whole slide imaging system for the creation, viewing, and management of digital images of surgical pathology slides. It allows whole slide digital images to be viewed on a display monitor that would otherwise be appropriate for manual visualization by conventional brightfield microscopy.
The E1000 Dx Digital Pathology Solution consists of the following three components: Scanner component:
- . E1000 Dx Digital Pathology Scanner with E1000 firmware version 2.0.3
- . E1000 Dx Scanner Software version 2.0.3
Viewer component:
- E1000 Dx Image Management System (IMS) Server version 2.3.2 ●
- . E1000 Dx Viewer Software version 2.7.2
Display component:
- . Barco MDPC-8127
The E1000 Dx Digital Pathology Solution automatically creates digital whole slide images by scanning formalin-fixed, paraffin-embedded (FFPE) tissue slides, with a capacity to process up to 1,000 slides. The E1000 Dx Scanner Software (EDSS), which runs on the scanner workstation, controls the operation of the E1000 Dx Digital Pathology Scanner. The scanner workstation, provided with the E1000 Dx Digital Pathology Solution, includes a PC, monitor, kevboard, and mouse. The solution uses a proprietary MRXS format to store and transmit images between the E1000 Dx Digital Pathology Scanner and the E1000 Dx Image Management System (IMS).
The E1000 Dx IMS is a software component intended for use with the Barco MDPC-8127 display monitor and runs on a separate, customer-provided pathologist viewing workstation PC. The E1000 Dx Viewer, an application managed through the E1000 Dx IMS, allows the obtained digital whole slide images to be annotated, stored, accessed, and examined on Barco MDPC-8127 video display monitor. This functionality aids pathologists in interpreting digital images as an alternative to conventional brightfield microscopy.
Here's a breakdown of the acceptance criteria and study proving the device meets them, based on the provided text:
Important Note: The provided text describes a Whole Slide Imaging System for digital pathology, which aids pathologists in reviewing and interpreting digital images of traditional glass slides. It does not describe an AI device for automated diagnosis or detection. Therefore, concepts like "effect size of how much human readers improve with AI vs without AI assistance" or "standalone (algorithm only without human-in-the-loop performance)" are not directly applicable to this device's proven capabilities as per the provided information.
Acceptance Criteria and Reported Device Performance
The core acceptance criterion for this device appears to be non-inferiority to optical microscopy in terms of major discordance rates when comparing digital review to a main sign-out diagnosis. Additionally, precision (intra-system, inter-system repeatability, and inter-site reproducibility) is a key performance metric.
Table 1: Overall Major Discordance Rate for MD and MO
| Metric | Acceptance Criteria (Implied Non-inferiority) | Reported Device Performance (Epredia E1000 Dx) |
|---|---|---|
| MD Major Discordance Rate | N/A (Compared to MO's performance) | 2.51% (95% CI: 2.26%; 2.79%) |
| MO Major Discordance Rate | N/A (Baseline for comparison) | 2.59% (95% CI: 2.29%; 2.82%) |
| Difference MD - MO | Within an acceptable non-inferiority margin | -0.15% (95% CI: -0.40%, 0.41%) |
| Study Met Acceptance Criteria | Yes, as defined in the protocol | Met |
Precision Study Acceptance Criteria and Reported Performance
| Metric | Acceptance Criteria (Lower limit of 95% CI) | Reported Device Performance (Epredia E1000 Dx) |
|---|---|---|
| Intra-System Repeatability (Average Positive Agreement) | > 85% | 96.9% (Lower limit of 96.1%) |
| Inter-System Repeatability (Average Positive Agreement) | > 85% | 95.1% (Lower limit of 94.1%) |
| Inter-Site Reproducibility (Average Positive Agreement) | > 85% | 95.4% (Lower limit of 93.6%) |
| All Precision Studies Met Acceptance Criteria | Yes | Met |
Study Details
2. Sample Size and Data Provenance:
-
Clinical Accuracy Study (Non-inferiority):
- Test Set Sample Size: 3897 digital image reviews (MD) and 3881 optical microscope reviews (MO). The dataset comprises surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue.
- Data Provenance: Not explicitly stated, but clinical studies for FDA clearance typically involve multiple institutions, often within the US or compliant with international standards, and are prospective in nature for device validation. The "multi-centered" description suggests multiple sites, implying diverse data. It is a "blinded, and randomized study," which are characteristics of a prospective study.
-
Precision Studies (Intra-system, Inter-system, Inter-site):
- Test Set Sample Size: A "comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems" was used. Specific slide numbers or FOV counts are mentioned as pairwise agreements (e.g., 2,511 comparison pairs for Intra-system, Inter-system; 837 comparison pairs for Inter-site) rather than raw slide counts.
- Data Provenance: Clinical specimens. Not specified directly, but likely from multiple sites for the reproducibility studies, suggesting a diverse, possibly prospective, collection.
3. Number of Experts and Qualifications:
- Clinical Accuracy Study: The study involved multiple pathologists who performed both digital and optical reviews. The exact number of pathologists is not specified beyond "pathologist" and "qualified pathologist." Their qualifications are generally implied by "qualified pathologist" and the context of a clinical study for an FDA-cleared device.
- Precision Studies:
- Intra-System Repeatability: "three different reading pathologists (RPs)."
- Inter-System Repeatability: "Three reading pathologists."
- Inter-Site Reproducibility: "three different reading pathologists, each located at one of three different sites."
- Qualifications: Referred to as "reading pathologists," implying trained and qualified professionals experienced in interpreting pathology slides.
4. Adjudication Method for the Test Set:
- Clinical Accuracy Study: The ground truth was established by a "main sign-out diagnosis (SD)." This implies a definitive diagnosis made by a primary pathologist, which served as the reference standard. It's not specified if this "main sign-out diagnosis" itself involved an adjudication process, but it is presented as the final reference.
- Precision Studies: For the precision studies, agreement rates were calculated based on the pathologists' readings of predetermined features on "fields of view (FOVs)." While individual "original assessment" seems to be the baseline for agreement in the intra-system study, the method to establish a single ground truth for all FOVs prior to the study (if any, beyond the initial "defined, clinically relevant histologic features") or an adjudication process during the study is not explicitly detailed. The agreement rates are pairwise comparisons between observers or system readings, not necessarily against a single adjudicated ground truth for each FOV.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- A comparative effectiveness study was indeed done, comparing human performance with the E1000 Dx Digital Pathology Solution (MD) to human performance with an optical microscope (MO).
- Effect Size: The study demonstrated non-inferiority of digital review to optical microscopy. The "effect size" is captured by the difference in major discordance rates:
- The estimated difference (MD - MO) was -0.15% (95% CI: -0.40%, 0.41%). This narrow confidence interval, inclusive of zero and generally close to zero, supports the non-inferiority claim, indicating no significant practical difference in major discordance rates between the two modalities when used by human readers.
6. Standalone (Algorithm Only) Performance:
- No, a standalone (algorithm only) performance study was not conducted or described. This device is a Whole Slide Imaging System intended as an aid to the pathologist for human review and interpretation, not an AI for automated diagnosis.
7. Type of Ground Truth Used:
- Clinical Accuracy Study: The ground truth used was the "main sign-out diagnosis (SD)." This is a form of expert consensus or definitive clinical diagnosis, widely accepted as the reference standard in pathology.
- Precision Study: For the precision studies, "defined, clinically relevant histologic features" were used, and pathologists recorded the presence of these features. While not explicitly stated as a "ground truth" in the same way as the sign-out diagnosis, the 'original assessment' or 'presumed correct' feature presence often serves as a practical ground truth for repeatability and reproducibility calculations.
8. Sample Size for the Training Set:
- The document does not mention a training set as this device is not an AI/ML algorithm that learns from data. It's a hardware and software system designed to digitize and display images for human review. The "development processes" mentioned are for the hardware and software functionality, not for training a model.
9. How the Ground Truth for the Training Set Was Established:
- This question is not applicable as there is no training set for this device as described. Ground truth establishment mentioned in the document relates to clinical validation and precision, not AI model training.
Ask a specific question about this device
(248 days)
Sq, Floor 37 New York, NY 10036
Re: K241273
Trade/Device Name: FullFocus Regulation Number: 21 CFR 864.3700
Classification Name: | Whole Slide Imaging System |
| Regulation Number: | 21 CFR 864.3700
Device Class: Class II CFR section: 864.3700 Product code: PSY Classification name: Whole Slide Imaging
For In Vitro Diagnostic Use
FullFocus is a software intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides for primary diagnosis. FullFocus is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. FullFocus is intended to be used with the interoperable components specified in the below Table.
Table: Interoperable components of FullFocus
| Scanner Hardware | Scanner Output file format | Interoperable Displays |
|---|---|---|
| Leica Aperio GT 450 DX scanner | DICOM, SVS | Dell UP3017Dell U3023E |
| Hamamatsu NanoZoomer S360MD Slide Scanner | NDPI | Dell U3223QEJVC-Kenwood JD-C240BN01A |
FullFocus, version 2.29, is a web-based software-only device that facilitates the viewing and navigating of digitized pathology images of slides prepared from FFPE-tissue specimens acquired from FDA cleared digital pathology scanners on FDA cleared displays. FullFocus renders these digitized pathology images for review, management and navigation for pathology primary diagnosis.
Image acquisition is performed using the intended scanner (s), with the operator conducting quality control on the digital WSI images according to the scanner's instructions for use and lab specifications to determine if re-scans are needed. Please see the Intended Use section and below tables for specifics on scanners and respective displays for clinical use.
Once a whole slide image is acquired using the intended scanner and becomes available in the scanner's database file system, a separate medical image communications software (not part of the device), automatically uploads the image and corresponding metadata to persistent cloud storage. Integrity checks are performed during the upload to ensure data accuracy.
The subject device enables the reading pathologist to open a patient case, view the images, and perform actions such as zooming, panning, measuring distances and annotating images as needed. After reviewing all images for a case, the pathologist will render a diagnosis.
FullFocus operates with and is validated for use with the FDA cleared components specified in the tables below:
| Scanner Hardware | Scanner Output file format | Interoperable Displays |
|---|---|---|
| Leica Aperio GT 450 DX scanner | DICOM, SVS | Dell UP3017Dell U3023E |
| Hamamatsu NanoZoomer S360MD Slide Scanner | NDPI | Dell U3223QEJVC-Kenwood JD-C240BN01A |
Table 1: Interoperable Components Intended for Use with FullFocus
FullFocus version 2.29 was not validated for the use with images generated with Philips Ultra Fast Scanner.
Table 2: Computer Environment/System Requirements for during the use of FullFocus
| Environment | Component | Minimum Requirements |
|---|---|---|
| Hardware | Processor | 1 CPU, 2 cores, 1.6GHz |
| Memory | 4 GB RAM | |
| Network | Bandwidth of 10Mbps | |
| Software | Operating System | • Windows• macOS |
| Browser | • Google Chrome (129.0.6668.90 or higher)• Microsoft Edge (129.0.2792.79 or higher) |
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Pixel-wise comparison: The 95th percentile of pixel-wise color differences in any image pair across all required screenshots must be less than 3.0 ΔE00 when compared to comparator (predicate device's Image Review Manipulation Software - IRMS) for identical image reproduction. This indicates visual adequacy for human readers. | The 95th percentile of pixel-wise differences between FullFocus and the comparators were less than 3 CIEDE2000, indicating that their output images can be considered to be pixel-wise identical. FullFocus has been found to visually adequately reproduce digital pathology images to human readers with respect to its intended use. |
| Turnaround time (Case selection): It should not take longer than 10 seconds until the image is fully loaded when selecting a case. | System requirements fulfilled: Not longer than 10 seconds until the image is fully loaded. |
| Turnaround time (Panning/Zooming): It shall not take longer than 7 seconds until the image is fully loaded when panning and zooming the image. | System requirements fulfilled: Not longer than 7 seconds until the image is fully loaded. |
| Measurement Accuracy (Straight Line): The 1mm measured line should match the reference value exactly 1mm ± 0mm. | All straight-line measurements compared to the reference were exactly 1mm, with no error. |
| Measurement Accuracy (Area): The measured area must match the reference area exactly 0.2 x 0.2 mm for a total of 0.04 mm² ± 0 mm². | All area measurements compared to the reference value were exactly 0.04mm², with no error. |
| Measurement Accuracy (Scalebar): 2mm scalebar is accurate. | All Tests Passed. |
| Human Factors Testing: (Implied from previous clearance) Safe and effective use by representative users for critical user tasks and use scenarios. | Human factors study designed around critical user tasks and use scenarios performed by representative users were conducted for previously cleared FullFocus, version 1.2.1, in K201005, per FDA guidance “Applying Human Factors and Usability Engineering to Medical Devices (2016)". Human factors validation testing is not necessary as the user interface hasn't changed. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Pixel-wise Comparison: 30 formalin-fixed paraffin-embedded (FFPE) tissue glass slides, representing a range of human anatomical sites.
- Sample Size for Turnaround Time & Measurements: Not explicitly stated as a number of distinct cases or images beyond the 30 slides used for pixel-wise comparison. For measurements, a "1 Calibration Slide" was used per test.
- Data Provenance: The text does not explicitly state the country of origin. The slides are described as "representing a range of human anatomical sites," implying a diverse set of real-world pathology samples. It is a retrospective study as it states "30 formalin-fixed paraffin-embedded (FFPE) tissue glass slides... were scanned".
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Pixel-wise Comparison: "For each WSI, three regions of interest (ROIs) were identified to highlight relevant pathological features, as verified by a pathologist."
- Number of Experts: At least one pathologist.
- Qualifications: "A pathologist" (specific qualifications like years of experience are not provided).
- Measurements: No expert was explicitly mentioned for establishing ground truth for measurements; it relies on a "test image containing objects with known sizes" (calibration slide) and "reference value."
4. Adjudication Method for the Test Set
- The text does not mention an explicit adjudication method (like 2+1 or 3+1 consensus) for the pixel-wise comparison or measurement accuracy. For the pixel-wise comparison, ROIs were "verified by a pathologist," suggesting a single-expert verification rather than a consensus process.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, an MRMC comparative effectiveness study was not done in this context. The study focused on demonstrating identical image reproduction (pixel-wise comparison) and technical performance (turnaround time, measurement accuracy) of the FullFocus viewer against predicate devices' viewing components. It did not directly assess the improvement in human reader performance (e.g., diagnostic accuracy or efficiency) with or without AI assistance. The device is a "viewer and management software," not an AI diagnostic aid in the sense of providing specific findings or interpretations.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, a standalone "algorithm only" performance was effectively done for the technical aspects. The pixel-wise comparison directly compares the image rendering of FullFocus with the predicate viewer's rendering without human intervention in the comparison process itself (though a pathologist verified ROIs). Similarly, turnaround times and measurement accuracy are intrinsic technical performances of the software.
7. The Type of Ground Truth Used
- Pixel-wise Comparison: The ground truth for this test was the digital image data as rendered by the predicate device's IRMS. The goal was to show that FullFocus reproduces the same image data. The "relevant pathological features" within ROIs were "verified by a pathologist" which served as a reference for what areas to test, not necessarily a diagnostic ground truth for the device's output.
- Measurements: The ground truth was based on known physical dimensions within a calibration slide and corresponding "reference values."
8. The Sample Size for the Training Set
- The provided text does not mention a training set. This is expected because FullFocus is a viewer and management software for digital pathology images, not an AI or machine learning algorithm that is "trained" on data to make predictions or assist in diagnosis directly. Its core function is to display existing image data accurately and efficiently.
9. How the Ground Truth for the Training Set Was Established
- As no training set is mentioned (since it's a viewer software), this question is not applicable based on the provided text.
Ask a specific question about this device