510(k) Data Aggregation

    K Number
    K251058

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-11-26

    (236 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Nitrogen test system | 862.1770 | CDN / Class 2 |
    | Centrifugal chemistry analyzer for clinical use | 862.2140

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tru Kidney Health Test Panel is part of the TruWellness Panel™ and is intended for use on the Tru Analyzer. The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Creatinine (CRE) and Blood Urea Nitrogen (BUN) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From the CRE determination, Estimated Glomerular Filtration Rate (eGFR) is calculated by the analyzer.

    The Tru Kidney Health Test Panel (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in diagnosing the following disorders in adults 18 years of age or older:

    • Creatinine (CRE): Renal disease and monitoring of renal dialysis.
    • Blood Urea Nitrogen (BUN): Certain renal and metabolic diseases.

    The Tru Analyzer is an automated, multi-assay integrated system, containing an absorbance reader, a bead scanner, and a cell imager for in vitro diagnostic use. This analyzer is intended for the quantitative determination of various analyte concentrations found in lithium-heparinized venous whole blood. It is for clinical laboratory or point-of-care use.

    Use only TruWellness Panel™ test kits with the Tru Analyzer.

    Device Description

    The TruSystem is an automated, integrated in vitro diagnostic platform consisting of the Tru Analyzer and the Tru Kidney Health Test Panel (part of the TruWellness Panel™), a Single-Use Consumable Kit that includes a Disc and a Support Pack. Designed for point-of-care and clinical laboratory use, the system enables the simultaneous measurement of clinical chemistry, immunoassay, and hematology parameters from a lithium-heparinized venous whole blood sample in a single run. The TruSystem delivers quantitative results for routine clinical chemistry and immunoassay analytes as well as a complete blood count (CBC) with a 3-part differential, all without the need for specialized operating skills, external calibration, or complex infrastructure.

    The Tru Analyzer is a benchtop instrument that fully automates sample processing, assay execution, and result reporting. Its touchscreen interface allows operators to control workflows, initiate tests, and review results with minimal training. The Single-Use Consumable Kit and blood sample are loaded into the analyzer drawer, where all necessary processing takes place. Internally, the Tru Analyzer integrates:

    • A high precision pipettor for automated sample and reagent handling,
    • An onboard centrifuge to separate whole blood into plasma,
    • A closed-loop thermal control system to maintain precise assay temperatures,
    • A motion control system to fully automate sample processing,
    • Dedicated detection modules for clinical chemistry, immunoassay, and hematology testing,
    • A high-definition camera used to collect assay readings, image capture, and instrument quality control checks,
    • An electronics board to manage individual module boards and associated firmware,
    • An integrated computer running the instrument software, accessible via the touchscreen user interface.

    The Tru Analyzer is factory-calibrated and continuously monitors its performance using optical sensing and electronic feedback mechanisms. Every time the analyzer is powered on or a Single-Use Consumable Kit (TruWellness Panel™) is loaded, the system runs an automated self-test to verify that it remains within calibration and is functioning properly. Internal quality control (QC) checks occur throughout the testing process to ensure the integrity of the analyzer, Disc, and Support Pack.

    If an internal QC check does not meet specifications, the instrument will automatically suppress result reporting to prevent the release of potentially inaccurate results. Any impacted tests will display associated codes in the footnotes of the patient report. The Tru Analyzer also applies result flags to highlight conditions that may indicate abnormalities in the patient sample or require attention. These flags appear in-line with impacted assay results for clear visibility. If a system operation failure occurs, the Tru Analyzer will halt the run, and no results will be generated. In these cases, the instrument displays an on-screen error message with a unique code and step-by-step instructions to guide resolution. These measures ensure the system provides accurate results, while requiring no special user knowledge, training, or on-site calibration.

    The Single-Use Consumable Kit (TruWellness Panel™) houses all the components needed to process and analyze samples on the Tru Analyzer, including dried reagents, internal process control solutions, barcodes that manage the identity of the kit lot (e.g., Disc and Support Pack ID), calibration information, dilution buffers, and single-use plastic pipette tips. It also serves as a waste container which the user discards at the end of the run.

    The Tru Kidney Health Test Panel is part of the TruWellness Panel™ and is intended for use on the Tru Analyzer. The Tru Kidney Health Test Panel (part of the TruWellness Panel™) provides in vitro quantitative determination of Creatinine (CRE), Blood Urea Nitrogen (BUN), and calculated estimated Glomerular Filtration Rate (eGFR) in lithium-heparinized venous whole blood samples.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K171971

    Validate with FDA (Live)

    Date Cleared
    2018-01-25

    (209 days)

    Product Code
    Regulation Number
    862.1050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    ----------------------|
    | Analyzer, chemistry, centrifugal, for clinical use | JJG | 862.2140

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Comprehensive Metabolic Panel is intended to be used for the quantitative determination of Alkaline Phosphate (ALP), Alanine Aminotransferase (ALP/GPT), Aspartate Aminotransferase (AST/GOT), Blood Urea Nitrogen (BUN) and Creatinine (CREA) in concentrations in lithium-heparinized venous whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.

    • Alkaline phosphatase or its isoenzymes measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

    • Alanine aminotransferase measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases.

    • Aspartate aminotransferase measurements are used in the diagnosis and treatment of certain types of liver and heart disease.

    • Blood urea nitrogen measurements are used in the diagnosis and treatment of certain types of renal and metabolic diseases.

    • Creatinine measurements are used in the diagnosis and treatment of renal dialysis, and as a calculation basis for measuring other urine analytes.

    The skyla Clinical Chemistry Analyzer is an in-vitro diagnostic device for the quantitative determination of clinical chemistry analytes in lithium-heparinized venous whole blood, heparinized plasma, or serum. It is for clinical laboratory and point-of-care use.

    The Minicare C300 Clinical Analyzer is an in-vitro diagnostic devices for the quantitative determination of clinical chemistry analytes in lithium-heparinized venous whole blood, heparinized plasma, or serum. It is for clinical laboratory and point-of-care use.

    Device Description

    The skyla Clinical Chemistry Analyzer, Minicare C300 Clinical Chemistry Analyzer (private label) and Comprehensive Metabolic Panel is an automatic chemistry system intended for use in clinical laboratories or point-of-care locations. The system consists of a portable analyzer and single-use disposable reagent panel discs.

    The analyzer utilizes precision photometric measurement technology, combined with the use of specific reagent panel disc, to measure the amount of substance in blood. The analyzer measures absorbance change of each reaction well in reagent panel disc and covert it to a concentration value for each analyte included on the panel.

    The skyla and Minicare Comprehensive Metabolic Panel reagent disc (which contains the Alkaline Phosphatase, Alanine Aminotransferase, Aspartate Aminotransferase, Blood Urea Nitrogen and Creatinine test systems) is designed to separate a heparinized venous whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer.

    AI/ML Overview

    The Lite-On Technology Corp.'s Comprehensive Metabolic Panel, skyla Clinical Chemistry Analyzer, and Minicare C300 Clinical Chemistry Analyzer (K171971) were evaluated for substantial equivalence. The acceptance criteria and performance data are primarily based on precision, matrix comparison, detection limits, linearity, and interference testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the predicate device's performance and the established clinical laboratory standards (e.g., CLSI guidelines, recovery within 90-110% for interference). The reported device performance aligns with these expectations.

    Performance CharacteristicAcceptance Criteria (Implied / Stated)Device Performance (Reported)
    Internal Precision / Reproducibility (CV%)Less than 10%ALP: 1.7 - 4.3% ALT: 3.0 - 6.8% AST: 2.2 - 9.7% BUN: 3.0 - 3.8% CREA: 2.4 - 8.5% (All results within 10% CV)
    Matrix Comparison (Correlation Coefficient)High correlation (e.g., >0.99) with predicate/reference method for Serum vs. Plasma, Plasma vs. WB, WB vs. SerumALP: 0.9997 - 0.9998 ALT: 0.9997 - 0.9998 AST: 0.9987 - 0.9990 BUN: 0.9988 - 0.9990 CREA: 0.9964 - 0.9971 (All show strong correlation)
    Detection Limit (LoD/LoQ)LoD/LoQ should meet requirements for total error for each analyte.ALP: LoD/LoQ: 4.405 U/L ALT: LoD/LoQ: 5.228 U/L AST: LoD/LoQ: 11.690 (LoQ) / 12.089 (LoD) U/L BUN: LoD/LoQ: 1.526 mg/dL CREA: LoD/LoQ: 0.262 mg/dL (All LoD/LoQ met requirements for total error)
    Linearity (R2 )High linearity (e.g., >0.99) across the claimed measuring range.ALP: 0.9967 (range 30-2083 U/L vs. claimed 41-1500 U/L) ALT: 0.9999 (range 17-535 U/L vs. claimed 20-500 U/L) AST: 0.9984 (range 16-1021.3 U/L vs. claimed 20-1000 U/L) BUN: 0.9979 (range 1.8-126.9 mg/dL vs. claimed 2-120 mg/dL) CREA: 0.9993 (range 0.48-21.48 mg/dL vs. claimed 0.6-20 mg/dL) (All show strong linearity within and beyond claimed ranges)
    Endogenous InterferenceRecoveries between 90% and 110% of the neat value.No significant effect (>10% effect) from Hemolysis (60-66 mg/dL), Icterus (13-32 mg/dL), Lipemia (189-1032 mg/dL) for any tested analyte.
    Exogenous InterferenceNo significant interference (>10% shift)Most substances showed no significant interference. Ibuprofen showed 11.8% Inc. for CREA, Salicylic acid showed 16.3% Dec. for ALT, and Theophylline showed 14.4% Dec. for ALP.
    POC Method Comparison (Correlation Coefficient)High correlation (e.g., >0.99) with a comparative clinical analyzer.ALP: 0.9987 (WB), 0.9986 (Serum) ALT: 0.9980 (WB), 0.9983 (Serum) AST: 0.9988 (WB), 0.9986 (Serum) BUN: 0.9981 (WB), 0.9977 (Serum) CREA: 0.9975 (WB), 0.9975 (Serum) (All show strong correlation)
    POC Precision (CV%)Less than 10% between operators.All analytes across three concentration levels at three POC sites showed CVs less than 10%.
    Electrical Safety and EMCCompliance with IEC 60601-1-2.Passed tests according to IEC 60601-1-2.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Internal Precision/Reproducibility: 80 runs per level (quadruplicate testing a day for 20 working days) for each of the three serum levels for all 5 analytes. Data provenance is not specified (e.g., country of origin) but refers to "patient serum samples." Implied prospective collection for the study.
    • Matrix Comparison: 40 human samples. Data provenance not specified. Implied prospective collection for the study.
    • Detection Limit: LoB: 60 measurements of a near-zero sample over 10 days. LoD/LoQ: serum samples containing very low concentrations, tested in triplicate using 2 lots of reagent discs for 10 days. Data provenance not specified. Implied prospective collection.
    • Linearity: 9 intermediate dilutions created from high and low human serum pool samples, plus spiked samples. Data provenance not specified. Implied prospective collection.
    • Endogenous Interferences: Not explicitly stated, but implies multiple samples to test two different concentrations (normal and abnormal) of analytes against specified interference levels. Data provenance not specified. Implied prospective collection.
    • Exogenous Substances: Two concentrations (low and high level) of samples for each of the 10 potential interferents. Data provenance not specified. Implied prospective collection.
    • Point-of-Care (POC) Method Comparison: Over 120 heparinized venous whole blood and serum samples for each analyte. Data provenance not specified; likely collected from the three POC sites, implying prospective collection.
    • Point-of-Care (POC) Precision Studies: Three levels of human serum samples from POC sites, assayed in quadruplicate twice a day for 20 days. Data provenance not specified. Implied prospective collection.
    • Point-of-Care (POC) Whole Blood Precision: Not explicitly stated, but tabular data suggests multiple analyses (mean, SD, CV) for low, medium, and high samples across 3 POC sites and multiple operators (OP1, OP2, OP3).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This type of submission (510k for a clinical chemistry analyzer) does not typically involve human experts establishing a "ground truth" for the test set in the same way an image analysis or diagnostic AI device would. Instead, the ground truth is established by:

    • Reference Methods/Materials: Traceability to established reference methods (e.g., IFCC reference method for ALT/ALP/AST, CDC reference method for BUN, IDMS Reference Method for CREA) and reference materials (NIST SRM967).
    • Comparative Clinical Analyzers: For method comparison studies, the Beckman Coulter AU2700 clinical analyzer served as the comparative (reference) method.

    Therefore, the "experts" in this context are the established, validated, and traceable laboratory methods and instruments, rather than individual human practitioners.

    4. Adjudication Method for the Test Set

    Not applicable. Diagnostic test performance for clinical chemistry analyzers is typically evaluated by statistical comparison to a reference method or established clinical ranges, not by an adjudication process as seen in clinical trial settings for diagnostic imaging.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where multiple human readers interpret cases with and without AI assistance. This submission is for a clinical chemistry analyzer.

    However, the "POC Precision studies" did evaluate performance across multiple operators (9 operators) at three POC sites. This demonstrates inter-operator variability, which is a related concept to multi-reader studies in a laboratory context, but it does not measure an "improvement with AI vs. without AI assistance" since the device itself is the primary measurement tool, not an AI assistant to a human reader.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, the studies presented are generally "standalone" performance evaluations of the device (skyla/Minicare Clinical Chemistry Analyzer with Comprehensive Metabolic Panel) itself. The device automatically measures analytes and displays results; there isn't an "algorithm only" component separate from the integrated device performance. All performance data (precision, linearity, detection limits, interference) are solely based on the device's output.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance is established through:

    • Traceability to Reference Methods/Materials: For calibration and analytical accuracy (e.g., IFCC, CDC, IDMS reference methods, NIST reference materials).
    • Comparative Clinical Analyzers: The Beckman Coulter AU2700 served as the comparative method for method comparison studies, essentially acting as the "ground truth" or reference for evaluating the test device's performance on patient samples.
    • Known Concentrations: For studies like linearity, detection limits, and interference, samples with known or spiked concentrations are used.

    8. The Sample Size for the Training Set

    This document does not specify a separate "training set" in the context of machine learning or AI. This device is a traditional in-vitro diagnostic (IVD) clinical chemistry analyzer. Its development would involve internal validation and optimization processes by the manufacturer, which might loosely be considered "training," but it's not described as an AI model training set with a specific size or provenance.

    9. How the Ground Truth for the Training Set was Established

    As this is a traditional IVD device, the concept of a "training set" for an AI model's ground truth is not applicable in the way it would be for AI-powered diagnostic software. The "ground truth" for the development and internal validation of such a device is established through:

    • Chemical Principles and Reactions: The underlying scientific principles of colorimetry and specific reagent reactions form the fundamental 'ground truth' for measurement.
    • Calibration Standards: The device is calibrated using standards whose concentrations are traceable to recognized reference methods and materials, ensuring accurate quantitative measurements.
    • Quality Control Materials: Known quality control materials are used to ensure the device performs within expected ranges over time.

    These elements collectively serve as the basis for ensuring the device's accuracy and reliability during its design, development, and manufacturing phases.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972409

    Validate with FDA (Live)

    Date Cleared
    1997-08-28

    (63 days)

    Product Code
    Regulation Number
    862.2140
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | Common/Classification Name: Centrifugal Chemistry Analyzer for ClinicalUse21 CFR 862.2140

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AG" Chemistry Analyzer is indicated for in vitro diagnostic use for the quantitative determination of body fluid constituents using automated spectrophotometric methods.

    Device Description

    The AG" Automated Chemistry Analyzer is the latest in a series of automated blood chemistry analyzers for the clinical chemistry laboratory that have become more and more compact as the miniaturization of the technology of the various components has proceeded. The AG" weighs only 37% as much as the currently marketed device, the ASCA Automated Spin Chemistry Analyzer, and has a "footprint" only about one-third as large. The AG" is controlled with a personal computer with a good color monitor, floppy disk drive, hard drive, and printer. The cuvette rotor has a unique self-contained design. Forty cuvettes make up a single cuvette rotor. Each cuvette is made up of a sample shelf and reagent well. The design allows reactions to be completed in the rotor rather than having to be transferred upon mixing, by centrifugal force, to a separate cuvette or reading chamber. The pump uses a rack and pinion drive with a resolution of 3000 steps. The integrated pipettor/dilutor boasts an accuracy of & 1.0% at full stroke and a precision of & 0.05% CV within-run at full stroke. This provides excellent accuracy and precision over a wide range of volumes and speeds.

    AI/ML Overview

    This is an old 510(k) submission (1997) for a chemistry analyzer. The provided document is a 510(k) summary, which often lacks detailed study results and acceptance criteria as they would be found in the full submission.

    Based on the provided text, here's the information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria in a table format. It generally states that the results "supported the safety and effectiveness" and "demonstrated that it is substantially equivalent to the predicate device."

    Performance MetricAcceptance Criteria (Not explicitly stated, inferred from context)Reported Device Performance
    Electrical SafetyMust meet relevant safety standardsSupported safety
    EMCMust meet relevant compatibility standardsSupported safety
    PrecisionComparable to predicate device or within acceptable limitsSupported effectiveness
    Accuracy (Standard Samples)Comparable to predicate device or within acceptable limitsSupported effectiveness
    Clinical Sample PerformanceComparable to predicate deviceSupported effectiveness and substantial equivalence to predicate device

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify the sample size for any of the tests or the data provenance. It only mentions "standard samples" and "clinical samples."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable as the device is a chemistry analyzer performing quantitative determinations. Ground truth would likely be established by reference methods or comparison to the predicate device, not by expert interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable for a chemistry analyzer's performance testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a chemistry analyzer, not an AI-assisted diagnostic imaging device for human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a standalone chemistry analyzer, meaning its performance is its standalone performance without human input beyond sample loading and general operation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The phrase "ground truth" is not explicitly used, but for a chemistry analyzer, the accepted "truth" would come from:

    • Reference standards/calibrators: For accuracy testing with "standard samples."
    • Predicate device results: For comparison of "clinical samples," where the predicate device's results serve as the comparative truth.

    8. The sample size for the training set

    This device is not described as utilizing machine learning or AI in the contemporary sense, so there would be no explicit "training set" as understood in that context. Its development would involve engineering and calibration, not machine learning model training.

    9. How the ground truth for the training set was established

    Not applicable, as there's no mention of a training set for an AI model. For the development and calibration of the instrument, internal standards and reference materials would have been used.