Search Results
Found 19 results
510(k) Data Aggregation
(71 days)
Device Name:** SmartGuard Technology Predictive Low Glucose Technology
Regulation Number: 21 CFR 862.1356
Classification Name** | Interoperable Automated Glycemic Controller |
| Regulation Number | 21 CFR 862.1356
** | QJI | SAME |
| Regulation Name | Interoperable Automated Glycemic Controller (under 21 CFR 862.1356
| Regulation Name | Interoperable automated glycemic controller, insulin suspend (under 21 CFR 862.1356
Medical Devices" (September 2017) and the requirements defined by the iAGC special controls in 21 CFR 862.1356
SmartGuard Technology:
SmartGuard technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGM), alternate controller enabled (ACE) pumps, and digitally connected devices to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose (SG) values.
SmartGuard technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older, and Type 2 diabetes mellitus in persons 18 years of age and older requiring insulin.
SmartGuard technology is intended for single patient use and requires a prescription.
Predictive Low Glucose Technology:
Predictive Low Glucose technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGM), alternate controller enabled (ACE) pumps, and digitally connected devices to automatically suspend delivery of insulin when the sensor glucose (SG) value falls below or is predicted to fall below predefined threshold values.
Predictive Low Glucose technology suspends and resumes insulin delivery in Manual mode. Manual mode contains a bolus calculator that calculates an insulin dose based on user-entered data.
Predictive Low Glucose technology is intended for the management of type 1 diabetes mellitus in persons 7 years of age and older, and type 2 diabetes mellitus in persons 18 years of age and older requiring insulin.
Predictive Low Glucose technology is intended for single patient use and requires a prescription.
Device Description for SmartGuard Technology
SmartGuard Technology, also referred to as the Advanced Hybrid Closed Loop (AHCL) algorithm, is a software-only interoperable automated glycemic controller (iAGC) intended to adjust basal insulin delivery and perform automatic correction boluses based on user-entered inputs and sensor glucose (SG) values obtained from a compatible integrated continuous glucose monitor (iCGM) or an interoperable Medtronic continuous glucose monitor (CGM). SmartGuard Technology is embedded within the firmware of a compatible alternate controller enabled (ACE) pump, which functions as the host device.
The ACE pump receives SG data from compatible CGM/iCGM devices via Bluetooth Low Energy (BLE) and also collects user inputs through the ACE pump (for pumps that have a screen) or the MiniMed App (for screenless pumps). The AHCL algorithm does not directly communicate with CGMs/iCGMs; instead, the ACE pump transmits CGM/iCGM data and user-entered inputs to the AHCL algorithm and receives the algorithm's output commands.
In SmartGuard Mode, the AHCL algorithm uses adaptive control to adjust insulin delivery every five (5) minutes based on SG values. It delivers an auto basal insulin dose to maintain user-selectable targets of 100, 110, or 120 mg/dL, supports a temporary target of 150 mg/dL for up to 24 hours, and can automatically deliver correction boluses when appropriate.
Meal boluses are the responsibility of the user. The AHCL algorithm includes an integrated bolus calculator that determines user-initiated meal bolus doses based on available SG information, carbohydrate input, and other patient parameters.
The AHCL algorithm is a software-only device with no user interface. All therapy settings, user interactions, and AHCL-related alerts or alarms are managed through the compatible ACE pump or the MiniMed App. An optional App Manager device with the MiniMed App pre-installed is available as an alternative to a user's personal mobile device.
Device Description for Predictive Low Glucose Technology
Predictive Low Glucose Technology, also referred to as the Predictive Low Glucose Management (PLGM) algorithm, is a software-only interoperable automated glycemic controller (iAGC) that is intended to automatically suspend delivery of insulin when sensor glucose (SG) values fall below or are predicted to fall below predefined threshold values based on user-entered inputs. The PLGM algorithm is embedded within the firmware of a compatible alternate controller enabled (ACE) pump, which functions as the host device.
The ACE pump receives SG data from a compatible integrated continuous glucose monitor (iCGM) or interoperable Medtronic CGM via Bluetooth Low Energy (BLE) and collects user-entered inputs through the ACE pump (for pumps that have a screen) or the MiniMed App (for screenless pumps). The PLGM algorithm does not directly communicate with CGM/iCGM devices; instead, the ACE pump transmits CGM/iCGM data and user inputs to the algorithm and receives the resulting suspend and resume commands. PLGM functionality is available only in Manual mode. In this mode, users are responsible for all basal and bolus insulin delivery, including user-defined basal rate patterns and user-initiated boluses, with optional use of the bolus calculator and the max bolus limit feature. When enabled, the PLGM algorithm may suspend insulin delivery for a minimum of 30 minutes and up to 2 hours based on current or predicted SG values and will automatically resume delivery when criteria are met or when the maximum suspend duration is reached. Users may also manually resume delivery at any time.
The PLGM algorithm is a software-only device with no user interface. All therapy settings, user interactions, and PLGM-related alerts or alarms are managed through the compatible ACE pump or the MiniMed App. An optional App Manager device with the MiniMed App pre-installed is available as an alternative to a user's personal mobile device.
N/A
Ask a specific question about this device
(58 days)
Device Name:** SmartGuard technology; Predictive Low Glucose technology
Regulation Number: 21 CFR 862.1356
Class II | SAME |
| Regulation Name | Interoperable Automated Glycemic Controller (under 21 CFR 862.1356
Class II | SAME |
| Regulation Name | Interoperable automated glycemic controller (under 21 CFR 862.1356
Medical Devices (September 2017)" and the requirements defined by the iAGC special controls 21 CFR 862.1356
Furthermore, the subject devices meet all the iAGC Special Controls requirements defined in 21 CFR 862.1356
SmartGuard technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGMs), and alternate controller enabled (ACE) pumps to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose values.
SmartGuard technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older, and of Type 2 diabetes mellitus in persons 18 years of age and older requiring insulin.
SmartGuard technology is intended for single patient use and requires a prescription.
Predictive Low Glucose technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGMs), and alternate controller enabled (ACE) pumps to automatically suspend delivery of insulin when the sensor glucose value falls below or is predicted to fall below predefined threshold values.
Predictive Low Glucose technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older, and of Type 2 diabetes mellitus in persons 18 years of age and older requiring insulin.
Predictive Low Glucose technology is intended for single patient use and requires a prescription.
SmartGuard technology
SmartGuard technology, also referred to as Advanced Hybrid Closed Loop (AHCL) algorithm, is a software-only device intended for use by people with Type 1 diabetes, ages 7 years or older, and by people with Type 2 diabetes, ages 18 years or older. It is an interoperable automated glycemic controller (iAGC) that is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible interoperable Medtronic continuous glucose monitors (CGM) and compatible alternate controller enabled (ACE) pumps to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose (SG) values.
The AHCL algorithm resides on the compatible ACE pump, which serves as the host device. It is meant to be integrated in a compatible ACE pump and is an embedded part of the ACE pump firmware.
Inputs to the AHCL algorithm (e.g., SG values, user inputs) are received from the ACE pump (host device), and outputs from the AHCL algorithm (e.g., insulin delivery commands) are sent by the algorithm to the ACE pump. As an embedded part of the firmware, the AHCL algorithm does not connect to or receive data from compatible CGMs; instead, sensor glucose (SG) values or other inputs received by the ACE pump from compatible CGMs via Bluetooth Low Energy (BLE) technology are transmitted to the embedded AHCL algorithm.
The AHCL algorithm works in conjunction with the ACE pump and is responsible for controlling insulin delivery when the ACE pump is in Auto Mode. It includes adaptive control algorithms that autonomously and continually adapt to the ever-changing insulin requirements of each individual.
The AHCL algorithm requires specific therapy settings (target setpoint, insulin-to-carb ratios and active insulin time) that need to be established with the help of a health care provider (HCP) before activation. It also requires five (5) consecutive hours of insulin delivery history, a minimum of two (2) days of total daily dose (TDD) of insulin, a valid sensor glucose (SG) and blood glucose (BG) values to start automated insulin delivery.
When activated, the AHCL algorithm adjusts the insulin dose at five-minute intervals based on CGM data. A basal insulin dose (auto basal) is commanded by the AHCL algorithm to manage glucose levels to the user's target setpoint of 100 mg/dL, 110 mg/dL or 120 mg/dL. The user can also set a temporary target of 150 mg/dL for up to 24 hours. In addition, under certain conditions the algorithm can also automatically command correction boluses (auto correction bolus) without user input.
Meal boluses are the responsibility of the user. The AHCL algorithm includes an integrated bolus calculation feature for user-initiated boluses for meals. When the user inputs their carbohydrate intake, the AHCL algorithm automatically calculates a bolus amount based off available glucose information, entered carbohydrate amount and other patient parameters.
The AHCL algorithm contains several layers of "safeguards" (mitigations) to provide protection against over-delivery or under-delivery of insulin to reduce risk of hypoglycemia and hyperglycemia, respectively.
The AHCL algorithm is a software-only device and does not have a user interface (UI). The compatible ACE pump provides a UI to the user to configure the therapy settings and interact with the algorithm. The AHCL-related alerts/alarms are displayed and managed by the pump.
Predictive Low Glucose technology
Predictive Low Glucose technology, also referred to as the Predictive Low Glucose Management (PLGM) algorithm is a software-only device intended for use by people with Type 1 diabetes, ages 7 years or older, and by people with Type 2 diabetes, ages 18 years or older. It is an interoperable automated glycemic controller (iAGC) that is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible interoperable Medtronic continuous glucose monitors (CGM) and compatible alternate controller enabled (ACE) pumps to automatically suspend delivery of insulin when the sensor glucose value falls below or is predicted to fall below predefined threshold values.
The PLGM algorithm resides on the compatible ACE Pump, which serves as the host device. It is meant to be integrated in a compatible ACE pump and is an embedded part of the ACE pump firmware.
Inputs to PLGM algorithm (e.g., sensor glucose values, user inputs) are received from the ACE pump (host device), and outputs from PLGM algorithm (e.g., suspend/resume commands) are sent by the algorithm to the ACE pump. As an embedded part of the ACE pump firmware, the PLGM algorithm does not connect to or receive data from compatible CGMs; instead, sensor glucose (SG) values or other inputs are received by the ACE pump from compatible CGMs via Bluetooth Low Energy (BLE) technology are transmitted to the embedded PLGM algorithm.
The PLGM algorithm works in conjunction with the ACE pump. When enabled, the PLGM algorithm is able to suspend insulin delivery for a minimum of 30 minutes and for a maximum of 2 hours based on current or predicted sensor glucose values. It will automatically resume insulin delivery when maximum suspend time of 2 hours has elapsed or when underlying conditions resolve. The user is also able to manually resume insulin at any time.
The PLGM algorithm is a software-only device and does not have a user interface (UI). The compatible ACE pump provides the UI to configure therapy settings and interact with the algorithm. The PLGM-related alerts/alarms are displayed and managed by the pump.
The provided FDA 510(k) clearance letter and summary concern the Medtronic SmartGuard Technology (Advanced Hybrid Closed Loop algorithm, AHCL) and Predictive Low Glucose Technology (Predictive Low Glucose Management algorithm, PLGM). The document details the devices' descriptions, indications for use, comparison to predicate devices, and summaries of non-clinical and clinical performance data.
However, it's important to note that this document primarily focuses on establishing substantial equivalence to previously cleared predicate devices and does not explicitly state specific acceptance criteria (performance targets) for clinical efficacy metrics (e.g., specific HbA1c reduction percentages or time in range targets needed for clearance) and does not present the study results in a direct "acceptance criteria vs. reported performance" table format for those specific targets. Instead, it highlights that the clinical data "confirmed the safety and effectiveness" and "demonstrated improved glycemic outcomes" or "non-inferiority," and that "the results also confirm that use...was associated with improved glucose control."
The document also provides details about the clinical studies without explicitly labelling them as "the study that proves the device meets the acceptance criteria" in the way a clinical trial protocol would specify primary and secondary endpoints and their statistical targets. Instead, it justifies substantial equivalence through the provided study data.
Therefore, the response below will extract the most relevant information based on your request, presenting the outcomes demonstrated by the studies as "reported device performance" where specific metrics are given, and noting the absence of explicit, pre-defined acceptance criteria targets in the clearance letter itself.
Acceptance Criteria and Study to Prove Device Meets Criteria
The FDA 510(k) summary for Medtronic's SmartGuard Technology and Predictive Low Glucose Technology establishes substantial equivalence to predicate devices. While the document asserts the safety and effectiveness, it does not explicitly define quantitative "acceptance criteria" for specific performance metrics in a pass/fail sense within this summary. Instead, it relies on demonstrating improved or non-inferior clinical outcomes compared to baseline or predicate performance. The clinical studies described confirm the safety and effectiveness and demonstrate associations with improved glucose control, which implicitly means the performance was deemed acceptable by the FDA for substantial equivalence.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
As explicit quantitative acceptance criteria (e.g., "HbA1c must reduce by X%") are not stated in this 510(k) summary, the table below presents the demonstrated clinical outcomes as "Reported Device Performance" highlighting the positive findings that supported clearance.
| Acceptance Criteria (Implied / Not Explicitly Stated) | Reported Device Performance (AHCL Algorithm) |
|---|---|
| Safety and Effectiveness | T1D (AHCL with Simplera Sync CGM - K251217): Confirmed safety and effectiveness. Demonstrated improved glycemic outcomes (reduction in HbA1c) compared to baseline, superiority for time in range, and non-inferiority for reduction in HbA1c. T2D (AHCL with Simplera Sync CGM & Guardian 4 CGM - P160017/S124): Confirmed safety and effectiveness. Use of AHCL SmartGuard was associated with improved glucose control. No device-related serious adverse events reported. - Phase 1 (Guardian 4 CGM): Significant Time-in-Range (TIR) of 80.9% (70-180 mg/dL). - Phase 2 (Simplera Sync sensor): Significant TIR of 85.4% (70-180 mg/dL). T1D (AHCL with Guardian 4 CGM & Lyumjev/Fiasp Insulins - P160017/S125): Confirmed safety and effectiveness. Use of Lyumjev and Fiasp with the MiniMed 780G Auto Correction feature was associated with improved glucose control in all age groups. No device-related serious adverse events reported for Fiasp. Lyumjev study reported one non-device related serious adverse event during screening, but none during run-in or study period for device use. |
| Glycemic Control Improvement | Demonstrated improved glycemic outcomes (HbA1c reduction, increased Time in Range). Specific TIR percentages for T2D were 80.9% (Phase 1) and 85.4% (Phase 2), significantly exceeding ADA recommendations (implicitly the target). In silico simulations for T2D showed statistical significance above ADA recommended TIR targets. |
| Hypoglycemia Reduction/Management | PLGM Algorithm: In silico studies for PLGM showed "percentage time in hypoglycemia <70 mg/dL fell within the margin" for the adult age group, indicating equivalency in time spent below 70 mg/dL with PLGM use. The clinical study for PLGM in K251217 (MiniMed 640G System) evaluated safety. |
| No Device-Related Serious Adverse Events (SAEs) | Generally reported "no device-related serious adverse events" across clinical trials for both AHCL and PLGM technologies when used in conjunction with the specified insulins and CGMs. |
2. Sample Sizes Used for the Test Set and Data Provenance
The primary "test sets" for clinical effectiveness and safety were the patient cohorts in the described clinical studies.
- AHCL with Simplera Sync CGM (Type 1 Diabetes): The sample size details for this study (originally in K251217) are not explicitly provided in this 510(k) summary. However, it confirmed safety and effectiveness in Type 1 Diabetes patients.
- AHCL with Simplera Sync CGM & Guardian 4 CGM (Type 2 Diabetes) (P160017/S124):
- Phase 1 (Guardian 4 CGM): N = 95 subjects.
- Phase 2 (Simplera Sync sensor): N = 302 subjects (66 "transition", 236 "naive").
- Data Provenance: Multi-center, single-arm study. The document does not specify countries but implies clinical trial settings. Given the general nature of FDA submissions, it would typically involve US-based and possibly international sites. Clinical data is generally considered prospective for such trials.
- AHCL with Guardian 4 CGM and Lyumjev Insulin (Type 1 Diabetes) (P160017/S125):
- ITT Population: N = 101 (Age 7-17 Years), N = 110 (Age 18-80 Years).
- Data Provenance: Single-arm, multi-center, home clinical investigation. The document does not specify countries but implies clinical trial settings. Prospective.
- AHCL with Guardian 4 CGM and Fiasp Insulin (Type 1 Diabetes) (P160017/S125):
- ITT Population: N = 107 (Age 7-17 Years), N = 116 (Age 18-80 Years).
- Data Provenance: Global multi-center, single-arm study. Countries mentioned: Australia, Canada, United States. Prospective.
- PLGM Algorithm: Evaluated for safety in a multi-center, single-arm, in-clinic study of the MiniMed 640G System. The sample size for this study (originally in K251217) is not explicitly provided in this 510(k) summary.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document refers to "clinical data" and "multi-center, single-arm studies" involving patients. For automated glycemic control devices, the "ground truth" for glucose values is typically established by laboratory reference methods (e.g., YSI glucose analyzer) performed by trained clinical staff as part of the study protocol, not by experts determining a ground truth in the interpretative sense. The efficacy endpoints (HbA1c, Time in Range, hypoglycemia) are derived from objective measurements, not subjective expert assessment of an image or signal.
There is no mention of "experts" in the context of establishing ground truth for glucose values or clinical outcomes. Clinical trials are monitored by clinical investigators (physicians, endocrinologists) and their teams, who ensure data integrity and protocol adherence, but they are not establishing a "ground truth" in the way radiologists might for AI image analysis.
4. Adjudication Method for the Test Set
Adjudication methods (e.g., 2+1, 3+1) are typically used in studies where subjective interpretation is involved, such as in reading medical images. For automated glycemic control devices, clinical outcomes are based on objective measurements (e.g., sensor glucose, lab-measured HbA1c). Therefore, the concept of an adjudication method as described does not directly apply to the clinical performance data presented here. Device-related adverse events would be reported and reviewed by the study investigators and likely an independent Data Safety Monitoring Board (DSMB), but this isn't an "adjudication method" for the primary clinical endpoints. The document does not explicitly describe an adjudication method for the objective clinical endpoints.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. MRMC studies are typically used to evaluate the performance of diagnostic devices or AI algorithms that assist human readers (e.g., radiologists interpreting images). The SmartGuard and PLGM technologies are automated insulin delivery algorithms, not diagnostic tools that human readers interpret. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" does not apply here. The studies evaluate the algorithm's direct impact on glycemic control in patients.
6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance
Yes, the core of the evaluation involves the "standalone" or "algorithm-only" performance in controlling glucose levels. The AHCL algorithm automatically adjusts basal insulin delivery and delivers correction boluses. The PLGM algorithm automatically suspends insulin delivery. While these systems require user input for meal boluses and setup, and interaction with alerts/alarms, the "SmartGuard technology" and "Predictive Low Glucose technology" themselves are software algorithms that function autonomously based on sensor glucose values to manage insulin delivery. The clinical studies assess the efficacy and safety of these algorithms in action within the device system (pump + CGM).
7. Type of Ground Truth Used
The ground truth for evaluating device performance in these studies is based on objective clinical measurements and patient outcomes. Specifically:
- Continuous Glucose Monitoring (CGM) sensor values: These are the primary input to the algorithm. While not explicitly stated as the ground truth for the algorithm's inputs, the accuracy of these values is critical and would have been established independently for the compatible CGMs.
- Laboratory-measured HbA1c: A standard clinical biomarker for average blood glucose over 2-3 months.
- Time-in-Range (TIR): Percentage of time spent with sensor glucose values within a target range (e.g., 70-180 mg/dL), derived from CGM data.
- Time-below-range (TBR): Percentage of time spent with sensor glucose values below a predefined threshold (e.g., <70 mg/dL), derived from CGM data.
- Adverse Events (AEs) and Serious Adverse Events (SAEs): Collected during clinical trials to assess safety.
These are physiological and clinical outcome data, not expert consensus or pathology reports in the typical sense.
8. The Sample Size for the Training Set
The document does not provide details on the sample size used for training the algorithm. This 510(k) summary focuses on the clinical data for validation of the finalized algorithms. The training set would be data used during the development phase of the algorithms, which is typically proprietary and not disclosed in 510(k) summaries. It would likely involve a large dataset of glucose profiles and insulin delivery patterns.
9. How the Ground Truth for the Training Set Was Established
Similarly, the document does not describe how the ground truth for the training set was established. For algorithms predicting glucose or controlling insulin, development and training would typically rely on:
- Retrospective or prospective real-world glucose and insulin data: Collected from individuals with diabetes using CGMs and insulin pumps, potentially under controlled conditions or in free-living settings.
- Validated glucose measurements: Such as YSI or other reference laboratory methods for blood glucose, and accurate CGM data.
- Clinical expert knowledge: Incorporating understanding of diabetes physiology, insulin pharmacokinetics/pharmacodynamics, and desired glycemic targets.
- Mathematical models of glucose metabolism: To simulate physiological responses and generate synthetic data for training.
The ground truth would be the actual glucose values and the physiological responses to insulin delivery, enabling the algorithm to learn patterns and predict future glucose trends or optimal insulin dosing.
Ask a specific question about this device
(249 days)
Grenoble, 38000
France
Re: K251152
Trade/Device Name: DBLG2
Regulation Number: 21 CFR 862.1356
** Interoperable Automated Glycemic Controller (iAGC)
Device Classification: Class II, 21 CFR 862.1356
** Interoperable Automated Glycemic Controller (iAGC)
| Device Classification: Class II, 21 CFR 862.1356 |
|---|
| Regulatory Information |
| Product Code |
| Regulation Name |
| , Interoperable Automated Glycemic Controller |
DBLG2, a mobile application with algorithm technology, is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) insulin infusion pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when glucose values are predicted to exceed a predefined threshold. To do this, the DBLG2 software takes into account the patient's profile, glycemia (current and predicted), announced meals and physical activities.
DBLG2 is intended for the management of type 1 diabetes mellitus in persons 12 years of age and greater.
DBLG2 is intended for single patient use.
DBLG2 is Rx - For Prescription Use Only.
Diabeloop DBLG2 is an Android application installed on patient personal mobile phone, intended for managing glucose levels in people with type 1 diabetes, using a hybrid closed loop approach (automated insulin delivery). It is designed to be connected with a compatible Automated Controller Enabled (ACE) insulin pump and integrated Continuous Glucose Monitors (iCGM).
DBLG2 has a regulation algorithm to automatically manage the patient's blood glucose level. It takes as input glycemia value received from the CGM, personal patient medical parameters and patient input related to meals and physical activities, and it calculates every 5 minutes the amount of insulin to deliver in order to keep the patient in the normoglycemia bounds. It sends this information to the pump that automatically delivers this quantity of insulin.
The software can ask the pump to deliver:
- A meal bolus
- A correction bolus (small amount of insulin)
- A basal rate over a given period of time.
The software can also ask the patient to take a calculated amount of carbohydrates if the system determines that the patient would go into hypoglycemia even if the insulin basal rate is brought down to zero.
DBLG2 acts mostly by modulating the basal rate of insulin delivery, but in some cases can deliver, automatically, correction boluses. It includes a patient-confirmed meal bolus calculator that simplifies meal dosing by allowing the patient to enter their meal carbohydrate amount while the system retrieves the patient's personalized insulin dosing parameters from their profile. The system calculates and displays a recommended meal bolus dose, which the patient must review and confirm before delivery is initiated.
In addition, DBLG2 has a self-learning module that applies improvements to the patient's algorithm parameters, based on estimated glycemia history and insulin delivery quantities, from the patient's history.
DBLG2 is designed to be secure, with by-design and structural security mechanisms that prevent from both hypoglycemia and hyperglycemia:
-
Hypoglycemia: by detecting an existing or upcoming hypoglycemia:
- The algorithm cuts down insulin delivery if a risk of hypoglycemia exists within the next fifteen minutes, the system will ask the user to take carbohydrates (by an alert).
- If the patient is in hypoglycemia below 55mg/dL (= 3.1 mmol/L), an alarm is triggered.
-
Hyperglycemia: the algorithm orders the delivery of insulin correction bolus to reduce the glycemia.
The software also guarantees patient's safety by returning to the pre-programmed basal pattern if the automatic regulation cannot be done for any reason (including loss of communication with the CGM or pump) or if the patient wishes to return to manual control by stopping the automatic regulation.
The algorithm includes appropriate alerts/alarms in case of any malfunction of one of the components.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) clearance letter for Diabeloop DBLG2:
1. Table of Acceptance Criteria and Reported Device Performance:
The document explicitly states that the DBLG2 software's clinical performance has been demonstrated through "consistent results in 6 prospective clinical trials." While specific numerical acceptance criteria (e.g., minimum Time in Range percentage, maximum hypoglycemia duration) are not clearly detailed in the provided text as a formal table of acceptance criteria, the document implies that the device is deemed safe and effective based on improvements in several glycemic outcomes compared to current treatment or a predicate device.
| Acceptance Criteria (Implied) | Reported Device Performance (DBLG2) |
|---|---|
| Improved Glycemic Outcomes | Demonstrated "various improvements in glycemic outcomes" including: |
| - Time in Range (TIR) 70-180 mg/dL | Improved |
| - Time in Hypoglycemia | Improved |
| - Time in Hyperglycemia | Improved |
| - Mean Continuous Glucose Monitoring (CGM) value | Improved |
| - HbA1c | Improved |
| - Occurrence of severe metabolic episodes | Improved |
| User Satisfaction | High satisfaction reported in various studies |
| Safety and Effectiveness for Intended Users/Uses | Human factors and clinical validation demonstrated that Diabeloop software performed as designed and intended for the intended users, uses, and use environments. |
| Substantial Equivalence to Predicate | "DBLG2 software is substantially equivalent to the predicate Tidepool Loop cleared in K203689. The differences... do not raise different questions of safety or effectiveness." |
2. Sample Size Used for the Test Set and Data Provenance:
- Test Set Description: The clinical performance is supported by "6 prospective clinical trials."
- Sample Size:
- Adults: Total of 15,325 patient-weeks of data collected.
- Adolescents (with T1D): Total of 1,594 patient-weeks of data collected.
- Data Provenance: The trials were "prospective clinical trials." The provided text doesn't explicitly state the country of origin for the data collection, but the submitter is based in Grenoble, France, which suggests these studies could have been conducted in Europe or internationally.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
The document does not provide details on the number of experts used or their specific qualifications (e.g., "radiologist with 10 years of experience") for establishing ground truth within the clinical trials. The phrase "ground truth" generally refers to definitive, objectively measured outcomes. For a glycemic control system, the primary ground truth would be the continuous glucose monitoring (CGM) data itself, along with lab-measured HbA1c, and recorded hypoglycemic/hyperglycemic events, which are direct physiological measurements rather than expert interpretations.
4. Adjudication Method for the Test Set:
The document does not specify any adjudication method (like 2+1, 3+1, none) for the test set. Given the nature of objective physiological measurements (CGM, HbA1c), manual adjudication of "ground truth" by multiple experts is less common than in, for example, image interpretation studies. The "ground truth" is largely the direct physiological data collected.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
No, an MRMC comparative effectiveness study was not done. MRMC studies are typically relevant for diagnostic aids where human readers interpret medical images or data, and the AI's assistance is being evaluated regarding its impact on their diagnostic accuracy. This device is an automated glycemic controller, not a diagnostic interpretation tool, so an MRMC study is not applicable. Therefore, there is no mention of an effect size for human reader improvement with AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:
Yes, the studies evaluate the performance of the DBLG2 software, which is an "algorithm technology" designed to "automatically increase, decrease, and suspend delivery of basal insulin" and "deliver correction boluses." While it integrates with user input (meal announcements, physical activities) and requires patient confirmation for meal boluses, the core "automatic regulation" of insulin delivery based on CGM readings and predictions represents its standalone algorithmic performance within the hybrid closed-loop system. The clinical trials assess the overall system performance, which inherently includes the algorithm's performance without constant human override of every automated decision.
7. The Type of Ground Truth Used:
The ground truth used is primarily physiological outcomes data, specifically:
- Continuous Glucose Monitoring (CGM) values, from which Time in Range (TIR), time in hypoglycemia, time in hyperglycemia, and mean CGM values are derived.
- HbA1c measurements.
- Documented occurrence of severe metabolic episodes.
- Patient feedback on satisfaction.
8. The Sample Size for the Training Set:
The document does not explicitly state the sample size used for the training set for the DBLG2 algorithm. The provided information focuses on the data used for clinical validation/testing. The DBLG2 system also includes a "self-learning module that applies improvements to the patient's algorithm parameters, based on estimated glycemia history and insulin delivery quantities, from the patient's history," which implies ongoing, personalized "training" within each patient's use. However, the initial development and training data for the core algorithm are not detailed here.
9. How the Ground Truth for the Training Set was Established:
The document does not detail how the ground truth for the training set was established. Given the nature of the device, it is highly likely that for the initial algorithm development and training, retrospective or simulated physiological data, potentially from a variety of sources (e.g., historical patient data, simulated diabetes models), would have been used. The "self-learning module" uses the patient's own glycemia history and insulin delivery quantities to refine parameters, establishing its internal "ground truth" from these real-time individual patient data points.
Ask a specific question about this device
(176 days)
01720
Re: K251779
Trade/Device Name: Omnipod 5 algorithm
Regulation Number: 21 CFR 862.1356
Panel (s):** | Clinical Chemistry |
| Product Code(s): | QJI |
| Regulation Numbers: | 21 CFR 862.1356
• Special Controls: evaluation of the Special Controls for this device (regulation 21 CFR 862.1356
The Omnipod 5 algorithm is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and pause delivery of insulin based on current and predicted glucose values. The Omnipod 5 algorithm is intended for the management of type 1 diabetes mellitus in persons 2 years of age and older and type 2 diabetes mellitus in persons 18 years of age and older. The Omnipod 5 algorithm is intended for single patient use and requires a prescription.
The Omnipod 5 algorithm is a software-only device that enables automated insulin delivery by taking in glucose inputs from a connected iCGM and calculating insulin micro-bolus outputs for delivery by a connected ACE Pump.
The Omnipod 5 algorithm is part of the Omnipod 5 Automated Insulin Delivery System, which also includes the Omnipod 5 ACE Pump and the SmartBolus Calculator regulated devices. The Omnipod 5 ACE Pump and the SmartBolus Calculator are functionally independent from the Omnipod 5 algorithm. The Omnipod 5 algorithm is intended to be digitally connected to a third party iCGM, the Omnipod 5 ACE Pump, and the SmartBolus Calculator.
The Omnipod 5 algorithm software is installed on both the Omnipod 5 Pod and Omnipod 5 App (either installed on a compatible smartphone or on the Insulet provided Controller) to comprise the Omnipod 5 System. As all regulated devices are required for the Omnipod 5 System, a prescription for the Omnipod 5 System and Omnipod 5 Pods includes:
• Omnipod Pods with Omnipod 5 algorithm software and
• Omnipod 5 Application with Omnipod 5 algorithm and SmartBolus Calculator software provided on Insulet provided Controller and when installed on the user's personal mobile device.
The Omnipod 5 System is a hybrid closed loop system and therefore can operate in either open loop (Manual Mode; Omnipod 5 algorithm disabled) or closed loop (Automated Mode; Omnipod
5 algorithm enabled). When Automated Mode is turned on, the Omnipod 5 algorithm (installed on the Pod) controls insulin delivery based on recent CGM values.
N/A
Ask a specific question about this device
(130 days)
/Device Name: SmartGuard technology
Predictive Low Glucose technology
Regulation Number: 21 CFR 862.1356
Class II | SAME |
| Regulation Name | Interoperable Automated Glycemic Controller (under 21 CFR 862.1356
Class II | SAME |
| Regulation Name | Interoperable automated glycemic controller (under 21 CFR 862.1356
Medical Devices (September 2017)" and the requirements defined by the iAGC special controls 21 CFR 862.1356
Furthermore, the subject devices meets all the iAGC Special Controls requirements defined in 21 CFR 862.1356
SmartGuard technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGMs), and alternate controller enabled (ACE) pumps to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose values.
SmartGuard technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older requiring insulin.
SmartGuard technology is intended for single patient use and requires a prescription.
Predictive Low Glucose technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGMs), and alternate controller enabled (ACE) pumps to automatically suspend delivery of insulin when the sensor glucose value falls below or is predicted to fall below predefined threshold values.
Predictive Low Glucose technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older requiring insulin.
Predictive Low Glucose technology is intended for single patient use and requires a prescription.
SmartGuard Technology, also referred to as Advanced Hybrid Closed Loop (AHCL) algorithm, is a software-only device intended for use by people with Type 1 diabetes for ages 7 years or older. It is an interoperable automated glycemic controller (iAGC) that is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible interoperable Medtronic continuous glucose monitors (CGM) and compatible alternate controller enabled (ACE) pumps to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose (SG) values.
The AHCL algorithm resides on the compatible ACE pump, which serves as the host device. It is meant to be integrated in a compatible ACE pump and is an embedded part of the ACE pump firmware.
Inputs to the AHCL algorithm (e.g., SG values, user inputs) are received from the ACE pump (host device), and outputs from the AHCL algorithm (e.g., insulin delivery commands) are sent by the algorithm to the ACE pump. As an embedded part of the firmware, the AHCL algorithm does not connect to or receive data from compatible CGMs; instead, sensor glucose (SG) values or other inputs received by the ACE pump from compatible CGMs via Bluetooth Low Energy (BLE) technology are transmitted to the embedded AHCL algorithm.
The AHCL algorithm works in conjunction with the ACE pump and is responsible for controlling insulin delivery when the ACE pump is in Auto Mode. It includes adaptive control algorithms that autonomously and continually adapt to the ever-changing insulin requirements of each individual.
The AHCL algorithm requires specific therapy settings (target setpoint, insulin-to-carb ratios and active insulin time) that need to be established with the help of a health care provider (HCP) before activation. It also requires five (5) consecutive hours of insulin delivery history, a minimum of two (2) days of total daily dose (TDD) of insulin, a valid sensor glucose (SG) and blood glucose (BG) values to start automated insulin delivery.
When activated, the AHCL algorithm adjusts the insulin dose at five-minute intervals based on CGM data. A basal insulin dose (auto basal) is commanded by the AHCL algorithm to manage glucose levels to the user's target setpoint of 100 mg/dL, 110 mg/dL or 120 mg/dL. The user can also set a temporary target of 150 mg/dL for up to 24 hours. In addition, under certain conditions the algorithm can also automatically command correction boluses (auto correction bolus) without user input.
Meal boluses are the responsibility of the user. The AHCL algorithm includes an integrated bolus calculation feature for user-initiated boluses for meals. When the user inputs their carbohydrate intake, the AHCL algorithm automatically calculates a bolus amount based off available glucose information, entered carbohydrate amount and other patient parameters.
The AHCL algorithm contains several layers of "safeguards" (mitigations) to provide protection against over-delivery or under-delivery of insulin to reduce risk of hypoglycemia and hyperglycemia, respectively.
The AHCL algorithm is a software-only device and does not have a user interface (UI). The compatible ACE pump provides a UI to the user to configure the therapy settings and interact with the algorithm. The AHCL-related alerts/alarms are displayed and managed by the pump.
Predictive Low Glucose Technology, also referred to as the Predictive Low Glucose Management (PLGM) algorithm is a software-only device intended for use by people with Type 1 diabetes for ages 7 years or older. It is an interoperable automated glycemic controller (iAGC) that is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible interoperable Medtronic continuous glucose monitors (CGM) and compatible alternate controller enabled (ACE) pumps to automatically suspend delivery of insulin when the sensor glucose value falls below or is predicted to fall below predefined threshold values.
The PLGM algorithm resides on the compatible ACE Pump, which serves as the host device. It is meant to be integrated in a compatible ACE pump and is an embedded part of the ACE pump firmware.
Inputs to PLGM algorithm (e.g., sensor glucose values, user inputs) are received from the ACE pump (host device), and outputs from PLGM algorithm (e.g., suspend/resume commands) are sent by the algorithm to the ACE pump. As an embedded part of the ACE pump firmware, the PLGM algorithm does not connect to or receive data from compatible CGMs; instead, sensor glucose (SG) values or other inputs are received by the ACE pump from compatible CGMs via Bluetooth Low Energy (BLE) technology are transmitted to the embedded PLGM algorithm.
The PLGM algorithm works in conjunction with the ACE pump. When enabled, the PLGM algorithm is able to suspend insulin delivery for a minimum of 30 minutes and for a maximum of 2 hours based on current or predicted sensor glucose values. It will automatically resume insulin delivery when maximum suspend time of 2 hours has elapsed or when underlying conditions resolve. The user is also able to manually resume insulin at any time.
The PLGM algorithm is a software-only device and does not have a user interface (UI). The compatible ACE pump provides the UI to configure therapy settings and interact with the algorithm. The PLGM-related alerts/alarms are displayed and managed by the pump.
The provided FDA 510(k) clearance letter and supporting documentation detail the acceptance criteria and the studies conducted to prove that Medtronic's SmartGuard Technology and Predictive Low Glucose Technology meet these criteria.
It's important to note that the provided text focuses on demonstrating substantial equivalence to a predicate device, as is typical for 510(k) submissions, rather than establishing de novo acceptance criteria for an entirely novel device. The "acceptance criteria" here refer to the performance benchmarks that demonstrate safety and effectiveness comparable to the predicate and compliance with regulatory special controls.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
The acceptance criteria are generally implied by the comparative data presented against the predicate device (Control-IQ Technology) and the compliance with "iAGC Special Controls requirements defined in 21 CFR 862.1356." The clinical study primarily aims to demonstrate non-inferiority or beneficial outcomes in key glycemic metrics compared to baseline (run-in period).
Table of Acceptance Criteria and Reported Device Performance
Given that this is a 510(k) submission showing substantial equivalence, the "acceptance criteria" are largely derived from the performance of the predicate device and clinical guidelines (e.g., ADA guidelines for Time Below Range). While specific numerical thresholds for acceptance are not explicitly listed as "acceptance criteria" in a table format within the provided text, the results presented serve as the evidence that these implicit criteria are met.
For the purpose of this response, I will synthesize the implied performance targets from the "Pivotal Study Observed Results" and "Supplemental Clinical Data" sections and present the device's reported performance against them.
Table 1: Implied Acceptance Criteria (via Predicate Performance/Clinical Guidelines) and Reported Device Performance for SmartGuard Technology (AHCL Algorithm)
| Performance Metric (Implied Acceptance Criteria) | Device Performance (SmartGuard Technology) - Reported | Comparison and Interpretation |
|---|---|---|
| HbA1c Reduction | Adults (18-80 yrs): Baseline: 7.4% ± 0.9. End of Study: 6.7% ± 0.5. | Shows a mean reduction of 0.7%, indicating improved glycemic control comparable to or better than predicate expectations. |
| Pediatrics (7-17 yrs): Baseline: 7.7% ± 1.0. End of Study: 7.3% ± 0.8. | Shows a mean reduction of 0.4%, indicating improved glycemic control. | |
| Percentage of subjects with HbA1c < 7% | Adults (18-80 yrs): From 30.9% (baseline) to 68.9% (end of study). | Significant increase in subjects achieving ADA target, demonstrating effectiveness. |
| Pediatrics (7-17 yrs): From 19.6% (baseline) to 36.9% (end of study). | Improvement in target achievement. | |
| Time in Range (TIR) 70-180 mg/dL | Adults (18-80 yrs): Increase from 66.5% ± 12.6 (run-in) to 80.2% ± 8.1 (Stage 3). | Substantial improvement (13.7 percentage points), demonstrating effective glucose management. Exceeds typical goals for AID systems. |
| Pediatrics (7-17 yrs): Increase from 54.4% ± 15.7 (run-in) to 71.4% ± 9.9 (Stage 3). | Significant improvement (17 percentage points), demonstrating effective glucose management. | |
| Time Below Range (TBR) < 70 mg/dL | Adults (18-80 yrs): Decrease from 1.7% ± 1.9 (run-in) to 1.5% ± 1.4 (Stage 3). | Maintained low rates of hypoglycemia, indicating safety. Well within ADA guideline of <4%. |
| Pediatrics (7-17 yrs): Increase from 1.6% ± 1.7 (run-in) to 1.9% ± 1.4 (Stage 3). | No significant increase, maintained low rates of hypoglycemia, indicating safety. Well within ADA guideline of <4%. | |
| Time Below Range (TBR) < 54 mg/dL | Adults (18-80 yrs): Decrease from 0.3% ± 0.5 (run-in) to 0.2% ± 0.4 (Stage 3). | Maintained very low rates of severe hypoglycemia. Well within ADA guideline of <1%. |
| Pediatrics (7-17 yrs): Increase from 0.3% ± 0.6 (run-in) to 0.4% ± 0.3 (Stage 3). | Maintained very low rates of severe hypoglycemia. Well within ADA guideline of <1%. | |
| Time Above Range (TAR) > 180 mg/dL | Adults (18-80 yrs): Decrease from 31.8% ± 13.1 (run-in) to 18.2% ± 8.4 (Stage 3). | Significant reduction, indicating improved hyperglycemia management. |
| Pediatrics (7-17 yrs): Decrease from 44.0% ± 16.1 (run-in) to 26.7% ± 10.1 (Stage 3). | Significant reduction, indicating improved hyperglycemia management. | |
| Severe Adverse Events (SAEs) related to device | Adults (18-80 yrs): 3 SAEs reported, but not specified if device-related. The "Pivotal Safety Results" section for ages 18-80 states "three serious adverse events were reported...". The "Clinical Testing for Predictive Low Glucose Technology" states that for PLGM, "there were no device related serious adverse events." Given this context, it's highly probable the SmartGuard SAEs were not device-related and the submission emphasizes no device-related SAEs across both technologies' studies. | Absence of device-related SAEs is a critical safety criterion. |
| Pediatrics (7-17 yrs): 0 SAEs (stated implicitly: "There were 0 serious adverse events..."). | Absence of device-related SAEs is a critical safety criterion. | |
| Diabetic Ketoacidosis (DKA) Events | Reported as 0 for SmartGuard Technology. | Absence of DKA events is a critical safety criterion. |
| Unanticipated Adverse Device Effects (UADEs) | Reported as 0 for SmartGuard Technology. | Absence of UADEs is a critical safety criterion. |
Table 2: Implied Acceptance Criteria and Reported Device Performance for Predictive Low Glucose Technology (PLGM Algorithm)
| Performance Metric (Implied Acceptance Criteria) | Device Performance (PLGM Algorithm) - Reported | Comparison and Interpretation |
|---|---|---|
| Avoidance of Threshold (≤ 65 mg/dL) after PLGM activation | 79.7% of cases (pediatric study). | Demonstrates effectiveness in preventing severe hypoglycemia. |
| Mean Reference Glucose Value 120 min post-suspension | 102 ± 34.6 mg/dL (adult study). | Indicates effective recovery from suspension without significant persistent hypoglycemia. |
| Device-related Serious Adverse Events | 0 reported. | Critical safety criterion. |
| Diabetic Ketoacidosis (DKA) Events related to PLGM | 0 reported. | Critical safety criterion. |
| Unanticipated Adverse Device Effects (UADEs) | 0 reported. | Critical safety criterion. |
Study Details
1. Sample Sizes and Data Provenance
Test Set (Clinical Studies):
-
SmartGuard Technology (AHCL Algorithm) - Pivotal Study:
- Adults (18-80 years): 110 subjects enrolled (105 completed).
- Pediatrics (7-17 years): 112 subjects enrolled (107 completed).
- Total: 222 subjects enrolled (212 completed).
- Provenance: Multi-center, single-arm study conducted across 25 sites in the U.S. This implies prospective data collection, specifically designed for this regulatory submission. Home-setting study.
-
Predictive Low Glucose Technology (PLGM Algorithm) - Clinical Testing:
- Adults (14-75 years): 71 subjects. In-clinic study.
- Pediatrics (7-13 years): 105 subjects. In-clinic evaluation.
- Total: 176 subjects.
- Provenance: Multi-center, single-arm, in-clinic studies. Location not explicitly stated but part of a US FDA submission, implying US or international sites adhering to FDA standards. Prospective data.
Training Set:
- SmartGuard Technology & Predictive Low Glucose Technology (Virtual Patient Model):
- Sample Size: Not explicitly stated as a number of "patients" but referred to as "extensive validation of the simulation environment" and "virtual patient (VP) model."
- Data Provenance: In-silico simulation studies using Medtronic Diabetes' simulation environment. This is synthetic data generated by computational models, validated against real patient data.
2. Number of Experts and Qualifications for Ground Truth (Test Set)
The clinical studies for both SmartGuard and PLGM technologies involved direct measurement of glucose values via CGM and blood samples (YSI for PLGM study, and HbA1c for SmartGuard study). These are objective physiological measures, not subjective interpretations requiring external expert consensus for "ground truth."
- For SmartGuard Technology: Glucose values were measured by the Simplera Sync CGM and HbA1c by laboratory tests. These are considered objective measures of glycemic control.
- For Predictive Low Glucose Technology: Hypoglycemia induction was monitored with frequent sample testing (FST) and frequent blood sampling for glucose measurements (likely laboratory-grade methods like YSI [Yellow Springs Instrument]).
- Expert involvement: While healthcare professionals (investigators, study coordinators, endocrinologists, nurses) were undoubtedly involved in conducting the clinical studies, managing patient care, and interpreting results, their role was not to establish "ground truth" through consensus or adjudication in the sense of image review. The ground truth was physiological measurements.
3. Adjudication Method for the Test Set
Not applicable in the typical sense of subjective clinical assessments (e.g., radiology image interpretation). Ground truth was established by direct physiological measurements (CGM data, HbA1c, YSI/FST blood glucose). The clinical studies were single-arm studies where subject outcomes were measured, not comparative assessments where multiple readers adjudicate on decisions.
4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not described. The clinical studies were single-arm (evaluating the device's performance in a standalone setting against a baseline or predefined safety/efficacy metrics) and did not involve human readers interpreting data from the device to make clinical decisions and then comparing human performance with and without AI assistance. Instead, the AI (algorithm) directly controlled insulin delivery and was evaluated based on patient physiological outcomes.
5. Standalone (Algorithm Only) Performance
Yes, the core of the evaluation is the standalone performance of the algorithms (SmartGuard AHCL and PLGM) in managing glucose, as they are "software-only devices" that reside on the ACE pump firmware. The clinical studies directly measured the physiological impact of the algorithm's actions on glucose levels, HbA1c, and safety parameters in a human population.
6. Type of Ground Truth Used
-
Clinical Studies (SmartGuard & PLGM): Objective Physiological Measurements
- Sensor Glucose (SG) values: From compatible CGMs (Simplera Sync CGM, Guardian 4 CGM)
- HbA1c: Laboratory measurements of glycosylated hemoglobin.
- Frequent Sample Testing (FST) / Blood Glucose (BG) values: Clinical laboratory measurements (e.g., YSI) to confirm hypoglycemia during PLGM induction.
- Adverse Events (AEs): Clinically reported and documented events.
- These are considered the definitive "ground truth" for evaluating glycemic control and safety.
-
In-Silico Simulation Studies: Virtual Patient Model Outputs
- The "ground truth" for these simulations is the metabolic response of the validated virtual patient models. This computational modeling is used to extend the clinical evidence to various parameter settings and demonstrate equivalence to real-world scenarios.
7. Sample Size for the Training Set
The document does not explicitly state a numerical "sample size" for a distinct "training set" of patients in the traditional machine learning sense for the algorithms themselves. The algorithms are likely developed and refined using a combination of:
- Physiological modeling: Based on established understanding of glucose-insulin dynamics.
- Historical clinical data: From previous similar devices or general diabetes patient populations (though not specified in this document for algorithm training).
- Clinical expertise: Incorporated into the algorithm design.
- The "Virtual Patient Model" itself is a form of simulated data that aids in development and testing. The validation of this model is mentioned as "extensive validation" and establishment of "credibility," implying a robust dataset used to verify its accuracy against real patient responses.
It's typical for complex control algorithms like these to be developed iteratively with physiological models and potentially large historical datasets, but a specific "training set" size for a machine learning model isn't detailed.
8. How the Ground Truth for the Training Set was Established
As noted above, a distinct "training set" with ground truth in the conventional sense of labeling data for a machine learning model isn't described. The development of control algorithms often involves:
- Physiological Simulation: The ground truth for this is the accurate metabolic response as modeled mathematically.
- Clinical Expertise & Design Principles: The ground truth is embedded in the scientific and medical principles guiding the algorithm's control logic.
- Validation of Virtual Patient Model: The "equivalency was demonstrated between Real Patients (RPs) and Virtual Patients (VPs) in terms of predetermined characteristics and clinical outcomes." This suggests that real patient data was used to validate and establish the "ground truth" for the virtual patient model itself, ensuring it accurately mirrors human physiology. This validated virtual patient model then serves as a crucial tool for in-silico testing.
Ask a specific question about this device
(68 days)
92130
Re: K250798
Trade/Device Name: Control-IQ+ technology
Regulation Number: 21 CFR 862.1356
Classification Name: Interoperable Automated Glycemic Controller
Regulation Number: 21 CFR 862.1356
Pump Type** | Alternate controller enabled insulin pump | Identical |
| Classification | 21 CFR 862.1356
Control-IQ+ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold.
Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater.
Control-IQ+ technology is intended for single patient use and requires a prescription.
The Subject Device, Control-IQ+ technology ("Control-IQ+") is a software-only device intended for the management of type 1 and type 2 diabetes mellitus. The device controls insulin delivery from a compatible alternate controller enabled insulin pump (ACE pump) based on inputs provided by a compatible integrated continuous glucose monitor (iCGM) and inputs provided by the user (e.g., carbohydrate intake, exercise, and sleep schedule). Control-IQ+ technology is meant to be installed on a compatible ACE pump.
Control-IQ+ technology has three different modes: Normal, Sleep, and Exercise. The glucose targets are not individually customizable in these modes but can change based on the mode selected. During Normal mode, Control-IQ+ technology aims to control glucose within a target range of 112.5 – 160 mg/dL, during Sleep mode the target range is 112.5 – 120 mg/dL, and during Exercise mode the target range is 140 – 160 mg/dL.
Control-IQ+ technology includes an integrated feature whereby iCGM values are automatically populated into the glucose field of the integrated bolus calculator when Control-IQ+ technology is active (i.e., the device is operating in closed-loop mode). This feature is disabled when Control-IQ is turned off.
Control-IQ+ technology requires users to input their weight and their total daily insulin requirement, which should be established with the help of a health care provider before using the device.
This FDA 510(k) clearance letter describes the acceptance criteria and study for the Control-IQ+ technology, an interoperable automated glycemic controller.
It's important to note that this device is a software-only device (Control-IQ+ technology), and the primary change described in this 510(k) is the addition of a new compatible insulin (Lyumjev U-100 Insulin) for use with the existing Control-IQ technology. The clearance relies heavily on the substantial equivalence to a predicate device (K243823, Control-IQ+ technology) and a clinical study demonstrating the safety and effectiveness of the new compatible insulin with the existing Control-IQ system.
Here's the breakdown of the acceptance criteria and study proving device performance:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria" for the Control-IQ+ technology's performance in terms of glucose control metrics (e.g., time in range, hypoglycemia rates) for this specific 510(k) submission. Instead, the "acceptance" for this submission appears to be based on demonstrating non-inferiority or better safety when using the new insulin (Lyumjev) with the already cleared Control-IQ system, compared to established benchmarks.
The primary "performance" mentioned is related to safety, specifically the rates of severe hypoglycemia and DKA.
| Acceptance Criteria (Implied) | Reported Device Performance (with Lyumjev) |
|---|---|
| Frequency of severe hypoglycemia comparable to or lower than T1D Exchange clinic registry data | Rates of severe hypoglycemia were lower than in the T1D Exchange clinic registry data. |
| Frequency of DKA events comparable to or lower than T1D Exchange clinic registry data | Rates of DKA were lower than in the T1D Exchange clinic registry data. |
| Well-tolerated with few adverse effects | The use of Lyumjev with t:slim X2 insulin pump with Control-IQ technology was well tolerated with few adverse effects. |
| No increase in hypoglycemia | No increase in hypoglycemia observed. |
| Statistical comparison meeting prespecified success criteria | The statistical comparison met the prespecified success criteria. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: 179 participants with type 1 diabetes.
- 70 adults (18-75 years old)
- 109 pediatric participants (6-17 years old)
- Data Provenance: The study was a "single-arm prospective safety trial." While the exact country of origin isn't specified, FDA clearances typically involve studies conducted in the US or under protocols recognized by the FDA. The T1D Exchange clinic registry is a US-based registry, suggesting a likely US context for the study.
- Retrospective/Prospective: Prospective. The study involved an initial (~16-day) Humalog Lead-in Period and a subsequent (13-week) Lyumjev Treatment Period.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This type of device (automated glycemic controller) does not typically involve expert review for "ground truth" in the same way an imaging or diagnostic AI might. The "ground truth" for glucose control is direct physiological measurements (iCGM readings) and clinical outcomes (hypoglycemia, DKA). Therefore, specific numbers or qualifications of experts for establishing ground truth are not applicable in this context. The study design itself serves to establish the performance and safety against clinical outcomes.
4. Adjudication Method for the Test Set
Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving subjective expert interpretation of data (e.g., radiologists reviewing images). For a system controlling insulin delivery based on CGM data, adjudication of such a type is not applicable. Clinical events (severe hypoglycemia, DKA) are typically adjudicated by an independent clinical endpoint committee or medical monitors based on predefined criteria, but the document does not specify this level of detail.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done.
MRMC studies are relevant for diagnostic aids where human readers interpret data (e.g., images) with and without AI assistance to measure improvement in reader performance. Control-IQ+ technology directly controls insulin delivery; it is not an assistive diagnostic tool for human interpretation.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)
The device is an "Interoperable Automated Glycemic Controller," meaning it operates to "automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values." It also delivers "correction boluses." This indicates that the core function is standalone (algorithm-only) in its closed-loop operation. While users can input data (carbohydrate intake, exercise, sleep schedule) and the device is intended for "single patient use" with a "prescription," the control logic itself functions automatically without continuous human intervention in real-time decision-making for insulin delivery adjustment. The study evaluates the system performance which includes this automated functionality.
7. The Type of Ground Truth Used
The ground truth used for this study was primarily:
- Physiological data: iCGM readings for glucose values.
- Clinical Outcomes Data: Rates of severe hypoglycemia and DKA events. These were compared against "reported frequencies from the T1D Exchange clinic registry" as a benchmark for safety.
8. The Sample Size for the Training Set
The document does not provide information regarding the sample size used for the training set of the Control-IQ+ technology. This 510(k) is for a modification (new compatible insulin) to an already cleared device, implying the core algorithm was trained and validated previously.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set of the original Control-IQ algorithm was established. Given the nature of an automated glycemic controller, it would typically involve extensive simulations, in-silico testing, and potentially prior clinical trials where continuous glucose monitoring (CGM) data, insulin delivery data, and corresponding blood glucose measurements were collected and used to train and validate the control algorithms. However, this specific 510(k) submission focuses on the safety and effectiveness of a new component (Lyumjev insulin) with the existing system rather than the initial foundational algorithm development.
Ask a specific question about this device
(74 days)
California 92130
Re: K243823
Trade/Device Name: Control-IO+ technology Regulation Number: 21 CFR 862.1356
Automated Glycemic Controller |
| Regulation Number | 21 CFR 862.1356
Control-IQ+ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold.
Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater.
Control-IQ+ technology is intended for single patient use and requires a prescription.
Control-IQ+ technology (Control-IQ+, the device) is a software-only device intended for the management of type 1 and type 2 diabetes mellitus. The device controls insulin delivery from a compatible alternate controller enabled insulin pump (ACE pump) based on inputs provided by a compatible integrated continuous glucose monitor (iCGM) and inputs provided by the user (e.g., carbohydrate intake, exercise, and sleep schedule). Control-IQ+ technology is meant to be installed on a compatible ACE pump.
Control-IQ+ technology has three different modes: Normal, Sleep, and Exercise. The glucose targets are not individually customizable in these modes but can change based on the mode selected. During normal mode, Control-IQ+ technology aims to control glucose within a target range of 112.5 - 160 mg/dL. During sleep mode, this range is changed to 112.5-120 mg/dL. and it is changed to 140-160 mg/dL during exercise mode.
Control-IQ+ technology includes an integrated feature whereby iCGM values are automatically populated into the glucose field of the integrated bolus calculator when Control-IQ + technology is active (i.e., the device is operating in closed-loop mode). This feature is disabled when Control-IQ is turned off.
Control-IQ+ technology requires users to input their weight and their total daily insulin requirement, which should be established with the help of a health care provider before using the device.
Here's an analysis of the provided text to extract information about acceptance criteria and the study proving the device meets them, presented in the requested format.
It's important to note that the provided text is an FDA 510(k) clearance letter and a 510(k) summary. These documents primarily focus on demonstrating substantial equivalence to a predicate device, rather than providing a detailed technical breakdown of novel acceptance criteria for algorithm performance in the same way one might expect for a new AI/ML device where performance metrics (e.g., sensitivity, specificity, accuracy) are the primary basis of clearance.
For Control-IQ+ technology, the key change is the expansion of the intended use to include Type 2 diabetes patients, supported by a pivotal clinical study. The algorithm itself is stated to be "unchanged," implying that its core performance characteristics (e.g., glucose control targets, basal rate adjustments) are those already established for Type 1 diabetes with the predicate device. Therefore, the "acceptance criteria" here largely pertain to the clinical safety and effectiveness in the new population, rather than the internal algorithmic performance metrics.
Acceptance Criteria and Device Performance Study for Control-IQ+ Technology
Given that the device is a refined version of an already cleared product ("Control-IQ technology" K232382) and the core algorithm is "unchanged," the acceptance criteria for this 510(k) submission (K243823) are primarily focused on demonstrating the safety and effectiveness of expanding its indications to include Type 2 diabetes patients. The primary evidence for this is a clinical study.
1. Table of Acceptance Criteria and Reported Device Performance
As this is a 510(k) focusing on an expanded user population for an existing algorithm rather than a de novo submission for a novel AI algorithm with specific performance cutoffs, the acceptance criteria are framed in terms of clinical outcomes rather than typical AI performance metrics like sensitivity/specificity.
| Acceptance Criterion (Implicitly Derived from Study Goal) | Reported Device Performance (Result of Pivotal Clinical Study) |
|---|---|
| Primary Outcome: Safety and Effectiveness in Type 2 Diabetes Adults, measured by change in HbA1C. The goal is to show a significant improvement in HbA1C for the Control-IQ+ arm compared to the control arm. | Significant improvement in change in HbA1C (specific magnitude not detailed in this document, but implied to be sufficient for a "safe and effective" claim). Control-IQ+ technology was shown to be "safe and effective in patients with Type 2 diabetes." |
| Usability/Human Factors for the new Type 2 diabetes population. | The Summative Human Factors validation "ensured individuals can safely and effectively perform critical tasks associated with the use of Control-IQ+ technology." |
| Continued adherence to Special Controls of the Predicate Device. | "Evaluation and adherence to the Special Controls of the Predicate Device (K232382) demonstrates continued assurance of the safety and effectiveness of the Subject Device." |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set (Clinical Study):
- Sample Size: 319 subjects
- Data Provenance: Not explicitly stated (e.g., country of origin), but implies a multi-site prospective clinical trial. The study was a "randomized controlled trial."
- Retrospective/Prospective: Prospective
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. For a clinical trial of an automated glycemic controller, the "ground truth" for evaluating effectiveness is typically objective clinical measurements (e.g., HbA1C tests, CGM data, adverse events) rather than expert consensus on diagnostic images. Clinical outcomes are inherently the "ground truth." Expert physicians would design the trial, manage patients, and interpret results, but they aren't establishing a "ground truth" per se in the way radiologists might do for disease detection.
4. Adjudication Method for the Test Set
This information is not provided. For a randomized controlled trial measuring objective metrics like HbA1C, formal adjudication in the sense of a diagnostic agreement (e.g., 2+1 radiology consensus) is generally not applicable, as the outcomes are quantitative and measured directly.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- MRMC Study: No, an MRMC study was not done. This type of study is relevant for diagnostic imaging AI, where human readers (e.g., radiologists) interpret images with or without AI assistance.
- Effect Size of Human Improvement with AI: Not applicable, as this device is an automated glycemic controller, not an AI assisting human interpretation of medical images. The clinical study compares the device's performance directly against a control group (CGM with continued insulin injections), rather than evaluating human reader performance with and without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
- The "Clinical Testing" section describes a pivotal clinical study where Control-IQ+ technology's performance (an automated insulin delivery system) was evaluated in patients. While it's implied that the system operates autonomously based on CGM inputs and user settings, the study measures the system's performance in a real-world setting with human users interacting with the device, not a purely standalone algorithmic evaluation in a simulated environment. The study is effectively evaluating the device's performance as intended for clinical use.
7. The Type of Ground Truth Used
- The primary ground truth for the clinical study was outcomes data, specifically the change in HbA1C (Glycated Hemoglobin), which is a key clinical endpoint for diabetes management. Other clinical measures from the iCGM (e.g., glucose values, time in range) and safety outcomes would also serve as ground truth for evaluating the device's performance.
8. The Sample Size for the Training Set
- This information is not provided in the document. The document states that "The design of the Control-IQ+ algorithm is unchanged" from its predicate. This implies that the training of the core algorithm was done prior to the predicate device's clearance and is not being re-evaluated for this submission. The 510(k) is primarily about demonstrating safety and effectiveness in the expanded user population, not re-training or re-validating the core algorithm's initial development.
9. How the Ground Truth for the Training Set Was Established
- This information is not provided as the algorithm itself is stated to be unchanged from the predicate device. Details about the original algorithm's development, including its training data and how its ground truth was established, would have been part of the predicate device's (K232382) submission.
Ask a specific question about this device
(67 days)
Massachusetts 01720
Re: K241777
Trade/Device Name: SmartAdjust™ Technology Regulation Number: 21 CFR 862.1356
|
| Regulation Numbers: | 21 CFR 862.1356
Special Controls: Evaluation of the Special Controls for this device (regulation 21 CFR 862.1356) assures
SmartAdjust™ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and pause delivery of insulin based on current and predicted glucose values. SmartAdjust™ technology is intended for the management of type 1 diabetes mellitus in persons 2 years of age and older and type 2 diabetes mellitus in persons 18 years of age and older. SmartAdjust™ technology is intended for single patient use and requires a prescription.
SmartAdjust™ technology is a software-only device that enables automated insulin delivery by taking in glucose inputs from a connected iCGM and calculating insulin micro-bolus outputs for delivery by a connected ACE Pump.
SmartAdjust™ technology is part of the Omnipod 5 Automated Insulin Delivery System, which also includes the Omnipod 5 ACE Pump and the SmartBolus Calculator regulated devices. The Omnipod 5 ACE Pump and the SmartBolus Calculator are functionally independent from SmartAdjust™ technology. SmartAdjust™ technology is intended to be digitally connected to a third party iCGM, the Omnipod 5 ACE Pump, and the SmartBolus Calculator.
The SmartAdjust™ technology software is installed on both the Omnipod 5 Pod and Omnipod 5 Controller (which contains the Omnipod 5 App), the 2 physical components that make up the Omnipod 5 System.
The Omnipod 5 System is a hybrid closed loop system and therefore can operate in either open loop (Manual Mode; SmartAdjust™ technology disabled) or closed loop (Automated Mode; SmartAdjust™ technology enabled). When Automated Mode is turned on, the SmartAdjust™ algorithm (installed on the Pod) controls insulin delivery based on recent CGM values.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
This document describes the 510(k) premarket notification for the SmartAdjust™ Technology, which is an interoperable automated glycemic controller. The submission aims to expand the indications for use to include individuals with Type 2 Diabetes Mellitus aged 18 years and older, in addition to the existing indication for Type 1 Diabetes Mellitus.
1. A table of acceptance criteria and the reported device performance
The provided text doesn't explicitly list a table of "acceptance criteria" with numerical targets in a typical performance study format. However, it does report on a primary safety endpoint from a clinical study which effectively serves as a key performance metric for the expanded indication.
Here's a table based on the information provided, focusing on the clinical study's primary safety endpoint:
| Acceptance Criteria (Implied from Clinical Study Focus) | Reported Device Performance (Clinical Study Results) |
|---|---|
| Non-inferiority in change in HbA1c (margin 0.3%) | Mean change in HbA1c was -0.8% (P<0.001 for non-inferiority) |
| No safety concerns associated with device use | No safety concerns identified during the study |
| No unexpected or serious adverse device effects | No unexpected or serious adverse device effects reported |
| No DKA events | No DKA events reported |
| No hospitalizations or emergency visits for severe hypoglycemic events | No hospitalizations or emergency visits for severe hypoglycemic events reported |
| No instances of hyperosmolar hyperglycemic syndrome | No instances of hyperosmolar hyperglycemic syndrome reported |
2. Sample sized used for the test set and the data provenance
- Sample Size for Test Set (Clinical Study): 343 participants
- Data Provenance: The document doesn't explicitly state the country of origin. It indicates it was a "clinical study" performed by Insulet, implying it was prospective within the context of validating the device for T2DM.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The clinical study evaluated patient outcomes (HbA1c, adverse events) rather than relying on expert consensus for ground truth on specific assessments.
4. Adjudication method for the test set
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically relevant for interpretative devices where human experts read cases and the AI provides assistance. The SmartAdjust™ technology is an automated glycemic controller that directly delivers insulin based on algorithms, not an interpretive aid for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the clinical study essentially assessed the standalone performance of the SmartAdjust™ algorithm (within the Omnipod 5 System) in a human-in-the-loop context. While users interact with the system, the key performance metrics (HbA1c, safety events) are a direct consequence of the algorithm's automated insulin delivery decisions. The document states: "When Automated Mode is turned on, the SmartAdjust™ algorithm (installed on the Pod) controls insulin delivery based on recent CGM values." The clinical study then validated this control for T2DM.
7. The type of ground truth used
The ground truth for the clinical study was based on clinical outcomes data, specifically:
- HbA1c measurements (a standard biochemical marker for long-term glucose control)
- Observation and reporting of adverse events (DKA, severe hypoglycemia, hyperosmolar hyperglycemic syndrome)
8. The sample size for the training set
The document does not provide information on the sample size used for the training set of the SmartAdjust™ algorithm. This information is typically not included in a 510(k) summary for a device like an automated insulin delivery system, as the algorithm's development and validation might involve proprietary datasets and methods outside the scope of the premarket notification for an expanded indication.
9. How the ground truth for the training set was established
The document does not provide information on how the ground truth for the training set (if any specific to machine learning) was established. Given the nature of a glycemic control algorithm, training might involve simulated data, retrospective patient data with known glucose values and insulin needs, or data from prior clinical trials. However, the document does not specify this. The clinical study described in the summary is for validation of the algorithm's performance with the expanded indication, not for its initial training.
Ask a specific question about this device
(265 days)
Massachusetts 02144
Re: K232741
Trade/Device Name: SmartAdjust™ technology Regulation Number: 21 CFR 862.1356
QJI |
| Regulation Numbers: | 21 CFR 862.1356
to the specifications for glucose sensor performance for connected iCGMs in accordance with 21 CFR 862.1356
|
| iCGMPerformanceSpecificationsper 21 CFR862.1356
SmartAdjust™ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and pause delivery of insulin based on current and predicted glucose values. SmartAdjust™ technology is intended for the management of type 1 diabetes mellitus in persons 2 years of age and older. SmartAdjust™ technology is intended for single patient use and requires a prescription.
SmartAdjust™ technology is a software-only device that enables automated insulin delivery by taking in glucose inputs from a connected iCGM and calculating insulin micro-bolus outputs for delivery by a connected ACE Pump.
SmartAdjust™ technology is part of the Omnipod 5 Automated Insulin Delivery System, which also includes the Omnipod 5 ACE Pump and the SmartBolus Calculator regulated devices. The Omnipod 5 ACE Pump and the SmartBolus Calculator functions are functionally independent from SmartAdjust™ technology. SmartAdjust™ technology is intended to be digitally connected to a third party iCGM, the Omnipod 5 ACE Pump, and the SmartBolus Calculator.
The SmartAdjust™ technology software is installed on both the OP5 Pod and OP5 Controller (which contains the OP5 App), the 2 physical components that make up the Omnipod 5 System.
The Omnipod 5 System is a hybrid closed loop system and therefore can operate in either open loop (Manual Mode; SmartAdjust™ technology disabled) or closed loop (Automated Mode; SmartAdjust™ technology enabled). When Automated Mode is turned on, the SmartAdjust™ alqorithm (installed on the Pod) controls insulin delivery based on recent CGM values.
This document is an FDA 510(k) clearance letter and summary for the Insulet SmartAdjust™ technology. It states that the device is substantially equivalent to a previously cleared predicate device (K220394), with the only change being an update to the specifications for glucose sensor performance for compatible integrated continuous glucose monitors (iCGMs). This means the core algorithm and its performance were already established with the original clearance.
Therefore, the document does not contain the detailed study information typically found in a new device submission or a clinical trial report that directly proves the device meets specific acceptance criteria in a new clinical study. Instead, it relies on the substantial equivalence argument, implying that the previous studies for the predicate device, or studies demonstrating the updated iCGM performance, are sufficient.
Based on the provided text, here's what can be extracted and what cannot:
Information Available:
- Device Name: SmartAdjust™ technology
- Predicate Device: SmartAdjust™ technology (K220394)
- Change: Updated iCGM Performance Specifications per 21 CFR 862.1356(1)(iv).
- Conclusion: The subject device is substantially equivalent to its predicate. The differences do not raise new questions of safety and effectiveness.
Information NOT Available (because this is a substantial equivalence submission for a minor update, not a new clinical trial submission document):
- A table of acceptance criteria and the reported device performance: This document does not detail specific performance metrics or acceptance criteria for a new study, as it's an update to an already cleared device. The performance requirements were presumably met by the predicate device and the updated iCGM.
- Sample sizes used for the test set and the data provenance: Not provided.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided.
- Adjudication method: Not applicable/provided for this type of submission.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable/provided.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The device is an algorithm/software, but the clinical performance data related to its efficacy (how well it manages glucose) would typically be from human-with-device studies from the original predicate submission. No new standalone study details are here.
- The type of ground truth used: Not specified, but for a glycemic controller, this would typically involve actual blood glucose measurements.
- The sample size for the training set: Not provided.
- How the ground truth for the training set was established: Not provided.
Reconstruction of "Acceptance Criteria" based on the document's type:
Since this is a 510(k) for an update, the "acceptance criteria" revolve around demonstrating that the change (updated iCGM specifications) does not negatively impact safety or effectiveness, and that the device still meets the requirements for its classification.
Table of "Acceptance Criteria" and "Reported Device Performance" as implied by a 510(k) for an update:
| Acceptance Criteria (Implied for this 510(k) update) | Reported Device Performance (as stated in the 510(k) summary) |
|---|---|
| 1. Equivalence in Intended Use and Indications for Use The updated device must have the | The subject device has the same intended use and indications for use as the predicate device. |
| 2. Equivalence in Technological Characteristics No new questions of safety/effectiveness raised by changes. | There are no changes to the design or technology of SmartAdjust™ technology itself, other than the updated iCGM performance specifications. The differences do not raise any different questions about safety and effectiveness. |
| 3. Compliance with Special Controls (21 CFR 862.1356) The device must continue to meet the specific requirements for an Interoperable Automated Glycemic Controller. | The subject device has been shown to meet the special controls for an Interoperable Automated Glycemic Controller. |
| 4. Performance with updated iCGM Specifications The device's performance (safety and effectiveness in automating insulin delivery) must remain acceptable when integrated with iCGMs meeting the updated performance specifications. | Implied by the statement that the "differences in the performance specifications for compatible iCGMs do not raise different questions of safety and effectiveness." The previous clinical data for the original predicate device (K220394) would have supported the device's efficacy with compatible iCGMs. This update likely references test data from the iCGMs themselves or in-silico/bench testing to confirm continued compatibility. |
| 5. Substantial Equivalence to Predicate The overall assessment must confirm substantial equivalence. | SmartAdjust™ technology is substantially equivalent to its predicate. |
Study Information (based on the context of a 510(k) for an update):
- Sample size used for the test set and data provenance: No specific sample size for new clinical testing is mentioned. The approval hinges on the existing data for the predicate device and the updated iCGM meeting its own specifications. The data provenance for the original predicate would have included clinical trial data (likely prospective).
- Number of experts and qualifications, Adjudication method, MRMC study, Standalone performance: These details are not relevant to this specific 510(k) summary, as it's establishing equivalence based on a minor technical update, not presenting new clinical efficacy data for the core algorithm from scratch. Such studies would have been part of the original K220394 submission.
- Type of ground truth used: For the underlying technology (which is unchanged here), the ground truth for an automated glycemic controller would be actual blood glucose measurements, measured by a reference method (e.g., lab venous plasma glucose, or accurate point-of-care devices).
- Training Set Sample Size and Ground Truth Establishment: Not mentioned, as this is an established, already-trained algorithm.
In summary, this document is a regulatory approval for a minor update to an already-cleared medical device software, rather than a detailed report of a new clinical study. The "proof" lies in the argument of substantial equivalence to the predicate device, with the specific change (iCGM specifications) not introducing new safety or effectiveness concerns.
Ask a specific question about this device
(269 days)
Washington, District of Columbia 20001
Re: K232603
Trade/Device Name: CamAPS FX Regulation Number: 21 CFR 862.1356
CamAPS FX is a mobile app intended for managing glucose levels in people with type 1 diabetes, aged 2 years and over, using a hybrid closed loop approach (automated basal insulin delivery with manual bolusing for meals). Additional age and other restrictions may apply depending on the chosen continuous glucose monitor and insulin pump.
CamAPS FX requires an insulin pump and a continuous glucose monitor (CGM) to fulfil its intended use. The list of supported insulin pumps and CGMs is provided in the User Manual.
CamAPS FX is indicated for use in pregnancy complicated by type 1 diabetes provided that the linked continuous glucose monitoring system is suitable for use in pregnancy.
CamAPS FX is for prescription use only.
CamAPS FX is a mobile app.
This FDA 510(k) clearance letter (K232603) for the CamAPS FX device does not contain the detailed acceptance criteria and study information typically found in a clinical study report or a more comprehensive summary of safety and effectiveness.
The document primarily focuses on the regulatory clearance of the device, stating that it is substantially equivalent to legally marketed predicate devices and outlining the indications for use. It details regulatory compliance aspects such as general controls, special controls, the Quality System regulation, and reporting requirements.
Therefore, I cannot extract the requested information from the provided text. To answer your questions, one would need access to the full 510(k) submission, including the performance data and clinical study summaries.
Based on the provided text, the answer to all your questions about acceptance criteria and the study that proves the device meets them would be: "Information not available in the provided document."
The document mentions:
- Device Name: CamAPS FX
- Regulation Number/Name: 21 CFR 862.1356, Interoperable automated glycemic controller
- Regulatory Class: Class II
- Indications for Use: Managing glucose levels in people with type 1 diabetes, aged 2 years and over, using a hybrid closed loop approach (automated basal insulin delivery with manual bolusing for meals). Also indicated for use in pregnancy complicated by type 1 diabetes, provided the linked CGM is suitable. Requires an insulin pump and CGM. Prescription use only.
However, it does not provide any details regarding:
- A table of acceptance criteria and reported device performance.
- Sample sizes or data provenance for test sets.
- Number and qualifications of experts for ground truth.
- Adjudication method.
- Multi-reader multi-case comparative effectiveness study results or effect sizes.
- Standalone algorithm performance.
- Type of ground truth used.
- Training set sample size.
- How training set ground truth was established.