Search Filters

Search Results

Found 19 results

510(k) Data Aggregation

    K Number
    K253023

    Validate with FDA (Live)

    Device Name
    BIOGRAPH One
    Date Cleared
    2026-01-15

    (118 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Magnetic Resonance Imaging (MRI) is a noninvasive technique used for diagnostic imaging. MRI with its soft tissue contrast capability enables the healthcare professional to differentiate between various soft tissues, for example, fat, water, and muscle, but can also visualize bone structures.

    Depending on the region of interest, contrast agents may be used.

    The MR system may also be used for imaging during interventional procedures and radiation therapy planning.

    The PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders, and cancer.

    The integrated system utilizes the MRI for radiation-free attenuation correction maps for PET studies. The integrated system provides inherent anatomical reference for the fused MR and PET images due to precisely aligned MR and PET image coordinate systems.

    Device Description

    BIOGRAPH One with software Syngo MR XB10 includes new and modified hardware and software compared to the predicate device, Biograph mMR with software syngo MR E11P-AP01. A high level summary of the new and modified hardware and software is provided below:

    Hardware

    New Hardware

    • Gantry offset phantom
    • SDB (Smart Distribution Box)

    New Coils

    • BM Contour XL Coil
    • BM Head/Neck Pro PET-MR Coil
    • BM Spine Pro PET-MR Coil
    • Transfer of up-to-date RF coils from the reference device MAGNETOM Vida.

    Modified Hardware

    • Main components such as:
      • Detector cassettes / DEA
      • Phantom holder
      • Gantry tube
      • Backplane
      • Magnet and cabling
      • Gradient coil
      • MaRS (measurement and reconstruction system)
      • MI MARS
      • PET electronics
      • RF transmitter TBX3 3T (TX Box 3)
    • Other components such as:
      • Cover
      • Filter plate
      • Patient table
      • RFCEL_TEMP

    Modified Coils

    • Body coil
    • Transfer of up-to-date RF coils from the reference device MAGNETOM Vida with some improvements.

    Software

    New Features and Applications

    • Fast Whole-Body workflows
    • Fast Head workflow
    • myExam PET-MR Assist
    • CS-Vibe
    • myExam Implant Suite
    • DANTE blood suppression
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS without diffusion function
    • BioMatrix Motion Sensor
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • ASNR recommended protocols for imaging of ARIA
    • Open Workflow
    • Ultra HD-PET
    • "MTC Mode"
    • OpenRecon 2.0
    • Deep Resolve Boost for TSE
    • GRE_PC
    • The following functions have been migrated for the subject device without modifications from MAGNETOM Skyra Fit and MAGNETOM Sola Fit:
      • 3D Whole Heart
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • Complex Averaging
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling
    • The following function has been migrated for the subject device without modifications from MAGNETOM Free.Max:
      • myExam Autopilot Spine
    • The following functions have been migrated for the subject device without modifications from MAGNETOM Sola:
      • myExam Autopilot Brain
      • myExam Autopilot Knee
    • Transfer of further up-to-date SW functions from the reference devices.

    New Software / Platform

    • PET-Compatible Coil Setup
    • Select&GO
    • PET-MR components communication

    Modified Features and Applications

    • HASTE_CT
    • FL3D_VIBE_AC
    • PET Reconstruction
    • Transfer of further up-to-date SW functions from the reference devices with some improvements.

    Modified Software / Platform

    • Several software functions have been improved. Which are:
      • PET Group
      • PET Viewing
      • PET RetroRecon
      • PET Status and Tune-up/QA

    Other Modifications and / or Minor Changes

    • Indications for use
    • Contraindications
    • SAR parameter
    • Off-Center Planning Support
    • Flip Angle Optimization (Lock TR and FA)
    • Inline Image Filter
    • Marketing bundle "myExam Companion"
    • ID Gain
    • Automatic System Shutdown (ASS) sensor (Smoke Detector)
    • Patient data display (PDD)
    AI/ML Overview

    The FDA 510(k) Clearance Letter for BIOGRAPH One refers to several AI/Deep Learning features. However, the provided document does not contain explicit acceptance criteria for these AI features in a table format, nor does it detail a comparative effectiveness study (MRMC study) for human readers. It primarily focuses on demonstrating non-inferiority to the predicate device through various non-clinical tests.

    Below is an attempt to extract and synthesize the information based on the provided text, while acknowledging gaps in the information regarding specific acceptance criteria metrics and clinical studies.

    Acceptance Criteria and Study Details for BIOGRAPH One AI Features

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria in a dedicated table format. Instead, it describes performance in terms of achieving "convergence of the training" and "improvements compared to conventional parallel imaging," or confirming "very similar metrics" to the predicate. The "acceptance criteria" are implied by these statements and the successful completion of the described tests.

    AI FeatureImplied Acceptance Criteria (Performance Goal)Reported Device Performance
    Deep Resolve Boost for FL3D_VIBE & SPACEConvergence of training and improvement compared to conventional parallel imaging for SSIM, PSNR, and MSE; no negative impact on image quality.Quantitative evaluations of SSIM, PSNR, and MSE metrics showed a convergence of the training and improvements compared to conventional parallel imaging. Inspection of test images did not reveal any negative impact to image quality. Function used for faster acquisition or improved image quality.
    Deep Resolve Sharp for FL3D_VIBE & SPACEImprovements across quality metrics (PSNR, SSIM, perceptual loss), increased edge sharpness, reduced Gibb's artifacts.Characterized by several quality metrics (PSNR, SSIM, perceptual loss). Tests show increased edge sharpness and reduced Gibb's artifacts.
    Deep Resolve Boost for TSE (First Mention)Very similar metrics (PSNR, SSIM, LPIPS) to predicate/modified network, outperforming conventional GRAPPA. No negative visual impact.Evaluation on test dataset confirmed very similar metrics (PSNR, SSIM, LPIPS) for the predicate and modified network, with both outperforming conventional GRAPPA. Visual evaluations confirmed no negative impact to image quality. Function used for faster acquisition or improved image quality.
    Deep Resolve Boost for TSE (Second Mention)Statistically significant reduction of banding artifacts, no significant changes in sharpness/detail, no difference in clinical suitability (radiologist evaluation).Statistically significant reduction of banding artifacts with no significant changes in sharpness and detail visibility. Radiologist evaluation revealed no difference in suitability for clinical diagnostics between updated and cleared predicate network.

    2. Sample Sizes Used for Test Set and Data Provenance

    The document primarily describes a validation dataset which serves as the "test set" for the AI models during development, and an additional "test dataset" for specific evaluations.

    • Deep Resolve Boost for FL3D_VIBE and SPACE:

      • Test Set Description: The "collaboration partners (testing)" data is mentioned as the source for testing, implying an external, independent test set. No specific number for this test set is provided beyond the 1265 measurements for training/validation.
      • Sample Size (Validation/Training): 27,679 3D patches from 1265 measurements.
      • Data Provenance: "in-house measurements (training and validation) and collaboration partners (testing)." The country of origin is not specified but is likely Germany (Siemens Healthineers AG) and/or China (Siemens Shenzhen Magnetic Resonance LTD.) where the manufacturing is listed.
      • Retrospective/Prospective: "Input data was retrospectively created from the ground truth by data manipulation and augmentation." This indicates retrospective data use.
    • Deep Resolve Sharp for FL3D_VIBE and SPACE:

      • Test Set Description: The document states, "The high-resolution datasets were split to 70% training and 30% validation datasets before training to ensure independence of them." This implies the 30% validation dataset is used as the test set.
      • Sample Size (Validation/Training): 27,679 3D patches from 1265 measurements (split into 70% training and 30% validation).
      • Data Provenance: "in-house measurements (training and validation) and collaboration partners (testing)."
      • Retrospective/Prospective: "Input data was retrospectively created from the ground truth by data manipulation." This indicates retrospective data use.
    • Deep Resolve Boost for TSE (First Mention - General Performance):

      • Test Set Description: The "evaluation on the test dataset" is mentioned. The validation set is 30% of the 500 measurements.
      • Sample Size (Validation/Training): Approximately 13,000 high resolution 3D patches from 500 measurements (split into 70% training and 30% validation).
      • Data Provenance: "in-house measurements."
      • Retrospective/Prospective: "Input data was retrospectively created from the ground truth by data manipulation." This indicates retrospective data use.
    • Deep Resolve Boost for TSE (Second Mention - Banding Artifacts):

      • Test Set Description: "Additional test dataset for banding artifact reduction: more than 2000 slices." This dataset was acquired after the release of the predicate network.
      • Sample Size: More than 2000 slices.
      • Data Provenance: "in-house measurements and collaboration partners."
      • Retrospective/Prospective: Not explicitly stated for this specific additional dataset, but the training/validation data for the predicate was retrospective.

    3. Number of Experts and Qualifications for Ground Truth

    • Radiologist Evaluation for Deep Resolve Boost for TSE (Second Mention): The document mentions "the radiologist evaluation revealed no difference in suitability for clinical diagnostics."

      • Number of Experts: Not specified (singular "radiologist" used, but typically multiple are implied for such evaluations).
      • Qualifications: "Radiologist." No specific years of experience or subspecialty are mentioned.
    • Other features: For Deep Resolve Boost/Sharp for FL3D_VIBE and SPACE, and Deep Resolve Boost for TSE (first mention), the ground truth is derived directly from acquired image data (see section 7). No independent human expert ground truth establishment for these.

    4. Adjudication Method (for Test Set)

    • Radiologist Evaluation for Deep Resolve Boost for TSE (Second Mention): The adjudication method is not specified in the document (e.g., 2+1, 3+1). It only states "the radiologist evaluation."

    • Other features: Adjudication methods are not applicable as human experts were not establishing ground truth for objective metrics.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, the document does not describe an MRMC comparative effectiveness study where human readers' performance with and without AI assistance is compared. The evaluation for Deep Resolve Boost for TSE mentions "radiologist evaluation" but not in a comparative MRMC study context.
    • Effect Size: Not applicable, as no MRMC study was performed.

    6. Standalone (Algorithm Only) Performance

    • Was standalone performance done? Yes, the performance testing for all Deep Resolve features (Boost and Sharp for FL3D_VIBE, SPACE, and TSE) was conducted "algorithm only" by evaluating metrics like PSNR, SSIM, MSE, and LPIPS, and then visual inspection/radiologist evaluation. These refer to the algorithm's direct output performance.

    7. Type of Ground Truth Used

    • Deep Resolve Boost for FL3D_VIBE and SPACE: "The acquired datasets (as described above) represent the ground truth for the training and validation."
    • Deep Resolve Sharp for FL3D_VIBE and SPACE: "The acquired datasets represent the ground truth for the training and validation." Input data was manipulated (cropped k-space) to create low-resolution input and high-resolution output/ground truth from the same dataset.
    • Deep Resolve Boost for TSE (First Mention): "The acquired datasets represent the ground truth for the training and validation." Input data was manipulated (cropped k-space) to create low-resolution input and high-resolution output/ground truth from the same dataset.
    • Deep Resolve Boost for TSE (Second Mention): "The acquired training/validation datasets... represent the ground truth for the training and validation." Input data was manipulated by undersampling k-space, adding noise, and mirroring k-space.
    • Summary: The ground truth for all AI features was derived from acquired, high-resolution original image data (retrospectively manipulated to simulate inputs). For Deep Resolve Boost for TSE (second mention), there was also an implicit "expert consensus" or "expert reading" component for the "radiologist evaluation" regarding clinical suitability.

    8. Sample Size for the Training Set

    • Deep Resolve Boost for FL3D_VIBE and SPACE: 81% of 1265 measurements (for 27,679 3D patches).
    • Deep Resolve Sharp for FL3D_VIBE and SPACE: 70% of 1265 measurements (for 27,679 3D patches).
    • Deep Resolve Boost for TSE (First Mention): 70% of 500 measurements (for approx. 13,000 high resolution 3D patches).
    • Deep Resolve Boost for TSE (Second Mention): More than 23,250 slices (93% of the combined training/validation dataset from K213693).

    9. How the Ground Truth for the Training Set Was Established

    • Deep Resolve Boost for FL3D_VIBE and SPACE: The "acquired datasets" represent the ground truth. "Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further undersampling of the data by discarding k-space lines as well as creating sub-volumes of the acquired data."
    • Deep Resolve Sharp for FL3D_VIBE and SPACE: The "acquired datasets represent the ground truth." "Input data was retrospectively created from the ground truth by data manipulation. k-space data has been cropped such that only the center part of the data was used as input. With this method corresponding low-resolution data as input and high-resolution data as output / ground truth were created for training and validation."
    • Deep Resolve Boost for TSE (First Mention): Similar to Deep Resolve Sharp for FL3D_VIBE and SPACE: "The acquired datasets represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation. k-space data has been cropped such that only the center part of the data was used as input. With this method corresponding low-resolution data as input and high-resolution data as output / ground truth were created for training and validation."
    • Deep Resolve Boost for TSE (Second Mention): "The acquired training/validation datasets... represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further undersampling of the data by discarding k-space lines, lowering of the SNR level by addition of noise and mirroring of k-space data."

    In summary, for all AI features, the ground truth for training was established by using high-quality, originally acquired MRI data that was then retrospectively manipulated (e.g., undersampled, cropped, noise added) to create synthetic "lower quality" input data for the AI model to learn from, with the original high-quality data serving as the target output or ground truth.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K220939, K250436, K250443, K231617, K191040

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use / Indications for Use

    Indications for Use for MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit, MAGNETOM Viato.Mobile:

    The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

    The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

    Indications for Use for MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:

    The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

    The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

    Device Description

    The subject device, MAGNETOM Vida with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Vida with syngo MR XA60A (K231560).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • myExam 3D Camera
    • BM Contour XL Coil

    Modified Hardware:

    • RF Transmitter TBX3 3T (TX Box 3)
    • MaRS (Measurement and reconstruction system)

    Software

    New Features and Applications:

    • Brachytherapy Support for use with MR conditional applicators
    • CS Vibe
    • myExam Implant Suite
    • DANTE blood suppression
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • BioMatrix Motion Sensor
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • ASNR recommended protocols for imaging of ARIA
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • 3D Whole Heart
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • Complex Averaging
    • myExam Autopilot Spine
    • myExam Autopilot Brain and myExam Autopilot Knee
    • Open Workflow

    Modified features and applications:

    • GRE_PC
    • myExam RT Assist workflow improvements
    • Open Recon 2.0
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    The subject device, MAGNETOM Lumina with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Lumina with syngo MR XA60A (K231560). A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • myExam 3D Camera
    • BM Contour XL Coil

    Modified Hardware:

    • RF Transmitter TBX3 3T (TX Box 3)
    • MaRS (Measurement and reconstruction system)

    Software

    New Features and Applications:

    • CS Vibe
    • myExam Implant Suite
    • DANTE blood suppression
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • BioMatrix Motion Sensor
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • Complex Averaging
    • myExam Autopilot Spine
    • myExam Autopilot Brain and myExam Autopilot Knee
    • Compressed Sensing Cardiac Cine
    • Open Workflow

    Modified Features and Applications:

    • GRE_PC
    • Open Recon 2.0
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    The subject device, MAGNETOM Lumina with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Lumina with syngo MR XA60A (K231560). A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • myExam 3D Camera
    • BM Contour XL Coil

    Modified Hardware:

    • RF Transmitter TBX3 3T (TX Box 3)
    • MaRS (Measurement and reconstruction system)

    Software

    New Features and Applications:

    • CS Vibe
    • myExam Implant Suite
    • DANTE blood suppression
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • BioMatrix Motion Sensor
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • Complex Averaging
    • myExam Autopilot Spine
    • myExam Autopilot Brain and myExam Autopilot Knee
    • Compressed Sensing Cardiac Cine
    • Open Workflow

    Modified Features and Applications:

    • GRE_PC
    • Open Recon 2.0
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    The subject device, MAGNETOM Vida Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Vida with syngo MR XA60A (K231560).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • myExam 3D Camera
    • Beat Sensor
    • BM Contour XL Coil

    Modified Hardware:

    • RF Transmitter TBX3 3T (TX Box 3)
    • MaRS (Measurement and reconstruction system)
    • Host computers

    Software

    New Features and Applications:

    • Brachytherapy Support for use with MR conditional applicators
    • CS Vibe
    • myExam Implant Suite
    • DANTE blood suppression
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • BioMatrix Motion Sensor
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • ASNR recommended protocols for imaging of ARIA
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • GRE_PC
    • Open Recon 2.0
    • 3D Whole Heart
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • myExam Autopilot Spine
    • myExam Autopilot Brain and myExam Autopilot Knee
    • Deep Resolve for EPI
    • Deep Resolve for HASTE
    • Physiologging
    • Complex Averaging
    • Open Workflow

    Modified features and applications:

    • myExam RT Assist workflow improvements
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • myExam Angio Advanced Assist (Test Bolus)
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    The subject device, MAGNETOM Sola with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • BM Contour XL Coil

    Modified Hardware:

    • MaRS (Measurement and reconstruction system)

    Software

    New Features and Applications:

    • Brachytherapy Support for use with MR conditional applicators
    • CS Vibe
    • DANTE blood suppression
    • BioMatrix Motion Sensor
    • SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • ASNR recommended protocols for imaging of ARIA
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • 3D Whole Heart
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • Deep Resolve Swift Brain
    • myExam Autopilot Spine
    • Open Workflow
    • Complex Averaging
    • Open Workflow

    Modified features and applications:

    • myExam RT Assist workflow improvements
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • myExam Angio Advanced Assist (Test Bolus)
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    The subject device, MAGNETOM Sola with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • BM Contour XL Coil

    Modified Hardware:

    • MaRS (Measurement and reconstruction system)

    Software

    New Features and Applications:

    • Brachytherapy Support for use with MR conditional applicators
    • CS Vibe
    • DANTE blood suppression
    • BioMatrix Motion Sensor
    • SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • ASNR recommended protocols for imaging of ARIA
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • 3D Whole Heart
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • Deep Resolve Swift Brain
    • myExam Autopilot Spine
    • Open Workflow

    Modified features and applications:

    • myExam Implant Suite
    • GRE_PC
    • myExam RT Assist workflow improvements
    • Open Recon 2.0
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    The subject device, MAGNETOM Altea with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Altea with syngo MR XA61A (K232535).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • BM Contour XL Coil

    Modified Hardware:

    • MaRS (Measurement and reconstruction system)

    Software

    New Features and Applications:

    • CS Vibe
    • DANTE blood suppression
    • BioMatrix Motion Sensor
    • SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • Deep Resolve Swift Brain
    • myExam Autopilot Spine
    • Compressed Sensing Cardiac Cine
    • Open Workflow

    Modified features and applications:

    • myExam Implant Suite
    • GRE_PC
    • myExam RT Assist workflow improvements
    • Open Recon 2.0
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    The subject device, MAGNETOM Sola Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA70A (K250443).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • BM Contour XL Coil

    Modified Hardware:

    • MaRS (Measurement and reconstruction system)
    • Host computers

    Software

    New Features and Applications:

    • Brachytherapy Support for use with MR conditional applicators
    • CS Vibe
    • DANTE blood suppression
    • BioMatrix Motion Sensor
    • SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • ASNR recommended protocols for imaging of ARIA
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • myExam Implant Suite
    • GRE_PC
    • Open Recon 2.0
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • Deep Resolve Swift Brain
    • myExam Autopilot Spine
    • Open Workflow

    Modified features and applications:

    • myExam RT Assist workflow improvements
    • myExam Implant Suite
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    The subject device, MAGNETOM Sola Fit with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA70A (K250443).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • BM Contour XL Coil

    Modified Hardware:

    • MaRS (Measurement and reconstruction system)
    • Host computers

    Software

    New Features and Applications:

    • Brachytherapy Support for use with MR conditional applicators
    • CS Vibe
    • DANTE blood suppression
    • BioMatrix Motion Sensor
    • SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • ASNR recommended protocols for imaging of ARIA
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • myExam Implant Suite
    • GRE_PC
    • Open Recon 2.0
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • Deep Resolve Swift Brain
    • myExam Autopilot Spine
    • Open Workflow

    Modified features and applications:

    • myExam RT Assist workflow improvements
    • myExam Implant Suite
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE

    The subject device, MAGNETOM Viato.Mobile with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Viato.Mobile with syngo MR XA70A (K250443).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • BM Contour XL Coil

    Modified Hardware:

    • MaRS (Measurement and reconstruction system)
    • Host computers

    Software

    New Features and Applications:

    • CS Vibe
    • DANTE blood suppression
    • BioMatrix Motion Sensor
    • SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
    • RF pulse optimization with VERSE
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • ASNR recommended protocols for imaging of ARIA
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • myExam Implant Suite
    • GRE_PC
    • Open Recon 2.0
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • Deep Resolve Swift Brain
    • myExam Autopilot Spine
    • Open Workflow

    Modified features and applications:

    • myExam Implant Suite
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE

    With the subject software version, Syngo MR XB10, we are also introducing the following new 1.5T devices, which are part of our MAGNETOM Flow. Platform:

    MAGNETOM Flow.Elite
    MAGNETOM Flow.Neo
    MAGNETOM Flow.Rise

    The subject device, MAGNETOM Flow.Elite, MAGNETOM Flow.Neo and MAGNETOM Flow.Rise with software Syngo MR XB10, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA61A (K232535).

    A high-level summary of the new and modified hardware and software is provided below:

    New Hardware:

    • Magnet
    • MREF (Magnet Refrigerator)
    • Gradient system
    • Gradient Coil
    • RF System
    • System Cover
    • Patient Table
    • MaRS (Measurement and Reconstruction System)
    • Select&GO Display (TPAN_3G) and Control Panel (CPAN_2G)
    • Body Coil
    • Head/Neck Coil
    • BM Head/Neck Coil (with ComfortSound)
    • BM Contour S Coil
    • BM Contour M Coil
    • BM Contour L Coil
    • BM Contour XL Coil
    • Foot/Ankle Coil
    • BM Spine Coil
    • iTx Extremity 18 Flare
    • Multi-Index MR-RT Positioning (a part of "RT Pro Edition" marketing bundle) (not available for MAGNETOM Flow.Rise)

    Modified Hardware:

    • Gradient Power Amplifier (GPA)
    • SAR Monitoring
    • In-Vivo Shim

    Software

    New Features and Applications:

    • CS Vibe
    • BioMatrix Motion Sensor
    • SPAIR FatSat Improvements: SPAIR "Abdomen&Pelvis" mode and SPAIR Breast mode
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • Preview functionality for Deep Resolve Boost
    • EP2D_FID_PHS
    • EP_SEG_FID_PHS
    • AutoMate Cardiac (Cardiac AI Scan Companion)
    • DANTE blood suppression
    • SMS Averaging for TSE
    • SMS Averaging for TSE_DIXON
    • SMS for BLADE without diffusion function
    • Ghost reduction (Dual polarity Grappa (DPG))
    • Fleet Reference Scan
    • Deep Resolve Swift Brain
    • Quick Protocols
    • myExam Autopilot Spine
    • Open Workflow

    Modified features and applications:

    • myExam Implant Suite
    • GRE_PC
    • myExam RT Assist workflow improvements (not available for MAGNETOM Flow.Rise)
    • Open Recon 2.0
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE
    • SPACE Improvement: high bandwidth IR pulse
    • SPACE Improvement: increase gradient spoiling

    New (general) Software / Platform / Workflow:

    • Select&GO extension (coil-based Iso Centering, Patient Registration at the touch display, Start Scan at the touch display)
    • New Startup-Timer
    • myExam RT Assist (not available for MAGNETOM Flow.Rise)
    • myExam Brain RT-Autopilot (not available for MAGNETOM Flow.Rise)
    • Eco Power Mode Pro

    Modified (general) Software / Platform:

    • Improved Gradient ECO Mode Settings

    Furthermore, the following minor updates and changes were conducted for the subject devices MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit, MAGNETOM Sola, MAGNETOM Altea:

    • Off-Center Planning Support
    • Flip Angle Optimization (Lock TR and FA)
    • Inline Image Filter
    • Automatic System Shutdown (ASS) sensor (Smoke Detector)
    • ID Gain (re-naming)
    • Select&Go Display (Touch Display (TPAN))
    • Marketing bundle "myExam Companion"

    The following minor updates and changes were conducted for the subject devices MAGNETOM Sola Fit and MAGNETOM Viato.Mobile:

    • Off-Center Planning Support
    • Automatic System Shutdown (ASS) sensor (Smoke Detector)
    • ID Gain (re-naming)
    • Select&Go Display (Touch Display (TPAN))
    • Marketing bundle "myExam Companion"

    The following minor updates and changes were conducted for the subject devices MAGNETOM Flow.Elite, MAGNETOM Flow.Neo, MAGNETOM Flow.Rise:

    • Off-Center Planning Support
    • Flip Angle Optimization (Lock TR and FA)
    • Inline Image Filter
    • Automatic System Shutdown (ASS) sensor (Smoke Detector)
    • ID Gain (re-naming)
    • Marketing bundle "myExam Companion"
    • Marketing Bundle "RT Pro Edition"(not available for MAGNETOM Flow.Rise)
    AI/ML Overview

    This FDA 510(k) clearance letter pertains to several MAGNETOM MRI systems with software Syngo MR XB10. The document primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing of new and modified hardware and software features, particularly those involving Artificial Intelligence (AI) such as "Deep Resolve" functionalities.

    Here's an analysis of the acceptance criteria and the studies that prove the devices meet them, specifically for the AI features:

    1. Table of Acceptance Criteria and Reported Device Performance for AI Features

    The document does not explicitly state "acceptance criteria" for the AI features in a numerical format that would typically be seen for a device's performance metrics (e.g., minimum sensitivity, specificity). Instead, the acceptance criteria are implicitly defined by the evaluation methods and the "Test result summary" for each Deep Resolve feature, which aim to demonstrate equivalent or improved image quality compared to conventional methods.

    AI FeatureAcceptance Criteria (Implied)Reported Device PerformanceComments
    Deep Resolve Swift Brain- Quantitative quality metrics (PSNR, SSIM, NMSE) to demonstrate network impact.- Visual inspection to ensure no undetected artifacts.- Evaluation in clinical settings with collaboration partners.- "Impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM) and normalized mean squared error (NMSE)."- "Images were inspected visually to ensure that potential artefacts are detected that are not well captured by the metrics."- "Work-in-progress packages of the network were delivered and evaluated in clinical settings with collaboration partners."The results indicate successful performance in meeting these criteria, suggesting the AI feature performs as intended without negative impact on image quality and with acceptable quantitative metrics.
    Deep Resolve Boost for FL3D_VIBE and Deep Resolve Boost for SPACE- Quantitative evaluations (SSIM, PSNR, MSE) showing convergence of training and improvements over conventional parallel imaging.- Visual inspection to confirm no negative impact on image quality.- The function should allow for faster acquisition or improved image quality.- "Quantitative evaluations of structural similarity index (SSIM), peak signal-to-noise ratio (PSNR) and mean squared error (MSE) metrics showed a convergence of the training and improvements compared to conventional parallel imaging."- "An inspection of the test images did not reveal any negative impact to the image quality."- "The function has been used either to acquire images faster or to improve image quality."The results indicate successful performance, demonstrating quantitative improvements and confirming user benefit (faster acquisition or improved image quality) without negative visual impact.
    Deep Resolve Sharp for FL3D_VIBE and Deep Resolve Sharp for SPACE- Quantitative quality metrics (PSNR, SSIM, perceptual loss).- Rating and evaluation of image sharpness by intensity profile comparisons.- Demonstration of increased edge sharpness and reduced Gibb's artifacts.- "The impact of the Deep Resolve Sharp network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and perceptual loss."- "The tests include rating and an evaluation of image sharpness by intensity profile comparisons of reconstruction with and without Deep Resolve Sharp. Both tests show increased edge sharpness and reduced Gibb's artifacts."The results directly confirm improved image sharpness and reduced artifacts, meeting the implied performance criteria.
    Deep Resolve Boost for TSE- Similar metrics (PSNR, SSIM, LPIPS) to predicate (cleared) network, both outperforming conventional GRAPPA.- Statistically significant reduction of banding artifacts.- No significant changes in sharpness and detail visibility.- Radiologist evaluation confirming no difference in suitability for clinical diagnostics.- "The evaluation on the test dataset confirmed very similar metrics in terms of peak signal-to-noise ratio (PSNR), structural similarity index (SSIM) and learned perceptual image patch similarity metrics (LPIPS) for the predicate and the modified network with both outperforming conventional GRAPPA as the reference."- "Visual evaluations confirmed statistically significant reduction of banding artifacts with no significant changes in sharpness and detail visibility."- "In addition, the radiologist evaluation revealed no difference in suitability for clinical diagnostics between updated and cleared predicate network."This AI feature directly demonstrates equivalent or improved performance compared to the predicate, with specific mention of "radiologist evaluation" ensuring clinical suitability.

    2. Sample Size Used for the Test Set and Data Provenance

    Since the document distinguishes between training, validation, and testing datasets, the "test set" here refers to the data used for final evaluation of the AI model's performance.

    • Deep Resolve Swift Brain:

      • Test Set Sample Size: The document lists "Validation: 3,616 slices (1.5T validation); 6,048 slices (3T validation)" as part of the split. It also mentions "work-in-progress packages of the network were delivered and evaluated in clinical settings with collaboration partners," implying additional testing, but a specific numerical sample size for this external validation is not provided in detail. However, the initial splits serve as the primary "test set" for performance metrics mentioned.
      • Data Provenance: "in-house measurement," implying retrospective data collected at Siemens' facilities. The document notes that "attributes like gender, age and ethnicity are not relevant to the training data" due to network architecture, but no specific country of origin is stated beyond "in-house."
    • Deep Resolve Boost for FL3D_VIBE and Deep Resolve Boost for SPACE:

      • Test Set Sample Size: The document states 19% of 1265 measurements for validation. It also explicitly mentions "collaboration partners (testing)" indicating an external test set, but a specific numerical breakdown for this is not provided.
      • Data Provenance: "in-house measurements (training and validation) and collaboration partners (testing)." This suggests a mix of retrospective data potentially from various countries where Siemens has collaboration, though specific locations are not listed.
    • Deep Resolve Sharp for FL3D_VIBE and Deep Resolve Sharp for SPACE:

      • Test Set Sample Size: 30% of the 500 measurements are listed for validation, which serves as a test set. This equates to 150 measurements.
      • Data Provenance: "in-house measurements," implying retrospective data from Siemens' research facilities. Specific country not mentioned.
    • Deep Resolve Boost for TSE:

      • Test Set Sample Size: "Additional test dataset for banding artifact reduction: more than 2000 slices."
      • Data Provenance: "in-house measurements and collaboration partners" for training/validation. The "additional test dataset for banding artifact reduction" likely follows the same provenance. Retrospective data.

    3. Number of Experts Used and Qualifications for Ground Truth

    The document does not explicitly state the number of experts used to establish ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience") for any of the Deep Resolve features.

    However, for Deep Resolve Boost for TSE, it mentions:

    • "Visual evaluations confirmed statistically significant reduction of banding artifacts... "
    • "In addition, the radiologist evaluation revealed no difference in suitability for clinical diagnostics..."

    This indicates that radiologists were involved in the evaluation of the Deep Resolve Boost for TSE feature, presumably as experts to establish the clinical suitability. The exact number and their detailed qualifications are not provided. For other features, the ground truth is primarily based on the acquired raw data or manipulated versions of it, without explicit mention of expert review in the ground truth establishment process.


    4. Adjudication Method (for the test set)

    The document does not specify an adjudication method like "2+1" or "3+1" for establishing ground truth or evaluating the test set for any of the AI features. The ground truth for training and validation is derived from the "acquired datasets" which are considered the ground truth due to data manipulation and augmentation from these high-quality source images. For Deep Resolve Boost for TSE, a "radiologist evaluation" is mentioned, implying expert review without detailing a specific adjudication protocol.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to measure the improvement of human readers with AI assistance versus without AI assistance. The evaluations focus on the standalone performance of the AI algorithms in improving image quality metrics and, in one instance (Deep Resolve Boost for TSE, radiologist evaluation), the suitability for clinical diagnostics, rather than the impact on human reader performance.


    6. Standalone (Algorithm Only) Performance

    Yes, standalone (algorithm only) performance was done. The descriptions for each Deep Resolve feature focus entirely on the algorithm's performance in terms of quantitative image quality metrics (PSNR, SSIM, NMSE, MSE, LPIPS), visual inspection for artifacts, and improvements over conventional techniques. There is no mention of a "human-in-the-loop" component in the described performance evaluations for these AI features, except for the "radiologist evaluation" for Deep Resolve Boost for TSE which assessed clinical suitability of the output images, not reader performance with the AI.


    7. Type of Ground Truth Used

    • For Deep Resolve Swift Brain, Deep Resolve Boost for FL3D_VIBE & SPACE, and Deep Resolve Sharp for FL3D_VIBE & SPACE:

      • The ground truth used was the acquired datasets (raw MRI data). The input data for the AI models was then "retrospectively created from the ground truth by data manipulation and augmentation" (e.g., undersampling k-space, adding noise, cropping, creating sub-volumes, cropping k-space to simulate low-resolution input from high-resolution output). This means the AI models were trained to learn the mapping from manipulated (e.g., noisy, low-resolution, undersampled) inputs to the original, high-quality acquired image data.
    • For Deep Resolve Boost for TSE:

      • Similar to above, the "acquired training/validation datasets" were considered the ground truth. Input data was generated by "data manipulation and augmentation" (e.g., discarding k-space lines, lowering SNR, mirroring k-space data).

    In essence, the AI models are trained to restore or enhance images to resemble the high-quality, fully acquired MRI data that serves as the reference ground truth.


    8. Sample Size for the Training Set

    • Deep Resolve Swift Brain: 20,076 slices
    • Deep Resolve Boost for FL3D_VIBE and Deep Resolve Boost for SPACE: 81% of 1265 measurements. (This equates to approximately 1024 measurements).
    • Deep Resolve Sharp for FL3D_VIBE and Deep Resolve Sharp for SPACE: 70% of 500 measurements. (This equates to 350 measurements).
    • Deep Resolve Boost for TSE: More than 23,250 slices (93% of the total dataset).

    9. How the Ground Truth for the Training Set Was Established

    For all Deep Resolve features, the ground truth for the training set was established from acquired MRI datasets (either "in-house measurements" or from "collaboration partners"). These acquired datasets are implicitly considered the "true" or "high-quality" images. The AI models are designed to process inputs that mimic suboptimal acquisition conditions (e.g., undersampled k-space, lower SNR, lower resolution) and generate outputs that match these high-quality acquired images, which serve as the ground truth for learning. The process involved:

    • Retrospective creation: Input data was created retrospectively from the acquired ground truth data.
    • Data manipulation and augmentation: This involved techniques such as:
      • Discarding k-space lines (undersampling).
      • Lowering the SNR level by adding Gaussian noise to k-space data.
      • Uniformly-random cropping of training data.
      • Creating sub-volumes of acquired data.
      • Cropping k-space to generate low-resolution inputs corresponding to high-resolution ground truth.
      • Mirroring of k-space data.

    This approach demonstrates an unsupervised or self-supervised learning paradigm where the ground truth is derived directly from the complete and high-fidelity raw data, and the AI is trained to reconstruct or enhance images from degraded inputs to match this ideal ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251822

    Validate with FDA (Live)

    Date Cleared
    2025-11-20

    (160 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K232535, K191040

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MAGNETOM Free.Max:
    The MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross-sectional images that display, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body or extremities.

    Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images when interpreted by a trained physician or dentist trained in MRI yield information that may assist in diagnosis.

    The MAGNETOM MR system may also be used for imaging during interventional procedures when performed with MR-compatible devices such as MR Safe biopsy needles.

    When operated by dentists and dental assistants trained in MRI, the MAGNETOM MR system must only be used for scanning the dentomaxillofacial region.

    MAGNETOM Free.Star:
    The MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross-sectional images that display, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body or extremities.

    Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis.

    Device Description

    The subject devices MAGNETOM Free.Max and MAGNETOM Free.Star with software version syngo MR XA80A, consists of new and modified hardware and software features comparing to the predicate device MAGNETOM Free.Max and MAGNETOM Free.Star with software version syngo MR XA60A (K231617).

    New hardware features (Only for MAGNETOM Free.Max):

    • Dental coil
    • High-end host
    • syngo Workplace

    Modified hardware features:

    • MaRS
    • Select&GO Display (TPAN_3G)

    New Pulse Sequences/ Software Features / Applications:

    Only for MAGNETOM Free.Max:

    • EP_SEG_FID_PHS
    • EP2D_FID_PHS
    • EP_SEG_PHS
    • GRE_Proj
    • GRE_PHS
    • myExam Dental Assist
    • Select&GO Dental
    • Slice Overlapping

    For both MAGNETOM Free.Max and MAGNETOM Free.Star:

    • Eco Power Mode
    • Extended Gradient Eco Mode
    • System Startup Timer

    Modified Features and Applications:

    • myExam RT Assist (only for MAGNETOM Free.Max)
    • Deep Resolve for HASTE
    • Deep Resolve for EPI Diffusion
    • Select&GO for dental (only for MAGNETOM Free.Max)
    • Select&GO extension: Patient Registration and Start Scan
    • SPACE improvement: MTC prep module

    Other Modifications and Minor Changes:

    • MAGNETOM Free.Max Dental Edition marketing bundle (only for MAGNETOM Free.Max)
    • MAGNETOM Free.Max RT Pro Edition marketing bundle (only for MAGNETOM Free.Max)
    • Off-Center Planning Support
    • ID Gain
    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for MAGNETOM Free.Max and MAGNETOM Free.Star (K251822) offer high-level information regarding the devices and their comparison to predicate devices. However, it does not explicitly detail acceptance criteria (performance metrics with pass/fail thresholds) or a specific study proving the device meets those criteria for the overall device clearance.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices for general MR diagnostic imaging. The most detailed performance evaluation mentioned is for the AI feature "Deep Resolve Boost." Therefore, the response will focus on the information provided regarding Deep Resolve Boost, and address other points based on what is stated and what is not.


    Acceptance Criteria and Device Performance (Focusing on Deep Resolve Boost)

    Table 1. Deep Resolve Boost Performance Summary

    MetricAcceptance Criteria (Implicit from "significantly better")Reported Device Performance
    Structural Similarity Index (SSIM)Significantly better structural similarity with the gold standard than conventional reconstruction.Deep Resolve reconstruction has significantly better structural similarity with the gold standard than the conventional reconstruction.
    Peak Signal-to-Noise Ratio (PSNR) / Signal-to-Noise Ratio (SNR)Significantly better SNR than conventional reconstruction.Deep Resolve reconstruction has significantly better signal-to-noise ratio (SNR) than the conventional reconstruction, and visual evaluation confirmed higher SNR.
    Aliasing ArtifactsNot found to have caused artifacts.Deep Resolve reconstruction was not found to have caused artifacts.
    Image SharpnessSuperior sharpness compared to conventional reconstruction.Visual evaluation confirmed superior sharpness.
    Denoising LevelsImproved denoising levels.Visual evaluation confirmed improved denoising Levels (implicit in higher SNR and image quality).

    Note: The document does not provide numerical thresholds or specific statistical methods used to define "significantly better" for SSIM and PSNR. The acceptance criteria are implicitly derived from the reported positive performance relative to conventional reconstruction.

    Study Details for Deep Resolve Boost

    1. Sample Size used for the test set and the data provenance:

      • Test Data: A "set of test data" was used for quantitative metrics (SSIM, PSNR) and visual evaluation. This test data was a "retrospectively undersampled copy of the test data" which was also used for conventional reconstruction.
      • Provenance: "In-house measurements and collaboration partners."
      • Retrospective/Prospective: The process of creating the test data by manipulating (undersampling) retrospectively acquired data indicates a retrospective approach.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not specified. The document states, "Visual evaluation was performed by qualified readers."
      • Qualifications of Experts: "Qualified readers." No further specific qualifications (e.g., years of experience, specialty) are provided.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not specified. The document states, "Visual evaluation was performed by qualified readers." It does not mention whether multiple readers were used per case or how discrepancies were resolved.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study involving human readers with vs. without AI assistance was not explicitly described for the Deep Resolve Boost feature. The visual evaluation was focused on comparing images reconstructed with conventional methods versus Deep Resolve Boost, primarily to assess image quality attributes without explicit human performance metrics (e.g., diagnostic accuracy, reading time).
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance evaluation was done. The quantitative metrics (SSIM, PSNR) and the visual assessment of images reconstructed solely by the algorithm (Deep Resolve Boost) were performed to characterize the network's impact independently.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "acquired datasets represent the ground truth for the training and validation." Input data for testing was "retrospectively created from the ground truth by data manipulation and augmentation." This implies that the raw, fully sampled, and high-quality MRI acquisitions are considered the ground truth against which the reconstructed images (conventional and Deep Resolve Boost) are compared. This is a technical ground truth rather than a clinical ground truth like pathology.
    7. The sample size for the training set:

      • TSE: More than 25,000 slices.
      • HASTE: Pretrained on the TSE dataset and refined with more than 10,000 HASTE slices.
      • EPI Diffusion: More than 1,000,000 slices.
    8. How the ground truth for the training set was established:

      • "The acquired datasets represent the ground truth for the training and validation."
      • Input data for training was "retrospectively created from the ground truth by data manipulation and augmentation." This included "further under-sampling of the data by discarding k-space lines, lowering of the SNR level by addition of noise and mirroring of k-space data."
      • This indicates that the ground truth for training was derived from high-quality, fully sampled MRI acquisitions, which were then manipulated to simulate lower quality inputs for the AI to learn from.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250436

    Validate with FDA (Live)

    Date Cleared
    2025-06-16

    (122 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K231587, K231617, K223343, K191040

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross-sectional images, spectroscopic images and/or spectra, and that displays, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

    The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

    Device Description

    MAGNETOM Flow.Ace and MAGNETOM Flow.Plus are 60cm-bore MRI systems with quench pipe-free, sealed magnets utilizing DryCool technology. They are equipped with BioMatrix technology and run on Siemens' syngo MR XA70A software platform. The systems include Eco Power Mode for reduced energy and helium consumption. They have different gradient configurations suitable for all body regions, with stronger configurations supporting advanced cardiac imaging. Compared to the predicate device, new hardware includes a new magnet, gradient coil, RF system, local coils, patient tables, and computer systems. New software features include AutoMate Cardiac, Quick Protocols, BLADE with SMS acceleration for non-diffusion imaging, Deep Resolve Swift Brain, Fast GRE Reference Scan, Ghost reduction, Fleet Reference Scan, SMS Averaging, Select&GO extension, myExam Spine Autopilot, and New Startup-Timer. Modified features include improvements for Pulse Sequence Type SPACE, improved Gradient ECO Mode Settings, and Inline Image Filter switchable for users.

    AI/ML Overview

    The provided 510(k) clearance letter and summary describe the acceptance criteria and supporting studies for the MAGNETOM Flow.Ace and MAGNETOM Flow.Plus devices, particularly focusing on their AI features: Deep Resolve Boost, Deep Resolve Sharp, and Deep Resolve Swift Brain.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document uses quality metrics like PSNR, SSIM, and NMSE as indicators of performance and implicitly as acceptance criteria. Visual inspection and clinical evaluations are also mentioned.

    FeatureQuality Metrics (Acceptance Criteria)Reported Performance (Summary)
    Deep Resolve BoostPeak Signal-to-Noise Ratio (PSNR), Structural Similarity Index (SSIM)Most metrics passed.
    Deep Resolve SharpPSNR, SSIM, Perceptual Loss, Visual Rating, Image sharpness evaluation by intensity profile comparisonsVerified and validated by in-house tests, including visual rating and evaluation of image sharpness.
    Deep Resolve Swift BrainPSNR, SSIM, Normalized Mean Squared Error (NMSE), Visual InspectionAfter successful passing of quality metrics tests, work-in-progress packages were delivered and evaluated in clinical settings with collaboration partners. Potential artifacts not well-captured by metrics were detected via visual inspection.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document uses "Training and Validation data" and often refers to the datasets used for both. It is not explicitly stated what percentage or how many cases from these datasets were strictly reserved for a separate "test set" and what came from the "validation sets." However, given the separation in slice count, the "Validation" slices for Deep Resolve Swift Brain might be considered the test set.

    • Deep Resolve Boost:
      • TSE: >25,000 slices
      • HASTE: >10,000 HASTE slices (refined)
      • EPI Diffusion: >1,000,000 slices
      • Data Provenance: Retrospectively created from acquired datasets. Data covered a broad range of body parts, contrasts, fat suppression techniques, orientations, and field strength.
    • Deep Resolve Sharp:
      • Data: >10,000 high resolution 2D images
      • Data Provenance: Retrospectively created from acquired datasets. Data covered a broad range of body parts, contrasts, fat suppression techniques, orientations, and field strength.
    • Deep Resolve Swift Brain:
      • 1.5T Validation: 3,616 slices (This functions as a test set for 1.5T)
      • 3T Validation: 6,048 slices (This functions as a test set for 3T)
      • Data Provenance: Retrospectively created from acquired datasets.

    The document does not explicitly state the country of origin for the data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test sets. For Deep Resolve Swift Brain, it mentions "evaluated in clinical settings with collaboration partners," implying clinical experts were involved in the evaluation, but details are not provided. For Boost and Sharp, the "acquired datasets...represent the ground truth," suggesting the raw imaging data itself, rather than expert annotations on that data, served as ground truth.

    4. Adjudication Method for the Test Set

    The document does not describe a formal adjudication method (e.g., 2+1, 3+1). For Deep Resolve Swift Brain, it mentions "visually inspected" and "evaluated in clinical settings with collaboration partners," suggesting a qualitative assessment, but details on consensus or adjudication are missing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A formal MRMC comparative effectiveness study demonstrating human reader improvement with AI vs. without AI assistance is not described in the provided text. The studies focus on the AI's standalone performance in terms of image quality metrics and internal validation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Yes, standalone performance was done for the AI features. The "Test Statistics and Test Results Summary" for Deep Resolve Boost, Deep Resolve Sharp, and Deep Resolve Swift Brain describe the evaluation of the algorithm's output using quantitative metrics (PSNR, SSIM, NMSE) and visual inspection against reference standards, which is characteristic of standalone performance evaluation.

    7. The Type of Ground Truth Used

    For Deep Resolve Boost, Deep Resolve Sharp, and Deep Resolve Swift Brain, the ground truth used was the acquired high-quality datasets themselves. The input data for training and validation was then retrospectively created from this ground truth by manipulating or augmenting it (e.g., undersampling k-space, adding noise, cropping, using only the center part of k-space). This means the original, higher-quality MR images or k-space data served as the reference for what the AI models should reconstruct or improve upon.

    8. The Sample Size for the Training Set

    • Deep Resolve Boost:
      • TSE: >25,000 slices
      • HASTE: pre-trained on the TSE dataset and refined with >10,000 HASTE slices
      • EPI Diffusion: >1,000,000 slices
    • Deep Resolve Sharp: >10,000 high resolution 2D images
    • Deep Resolve Swift Brain: 20,076 slices

    9. How the Ground Truth for the Training Set Was Established

    For Deep Resolve Boost, Deep Resolve Sharp, and Deep Resolve Swift Brain, the "acquired datasets (as described above) represent the ground truth for the training and validation." This implies that high-quality, fully acquired MRI data was considered the ground truth. The input data used during training (e.g., undersampled, noisy, or lower-resolution versions) was then derived or manipulated from this original ground truth. Essentially, the "ground truth" was the optimal, full-data acquisition before any degradation was simulated for the AI's input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233657

    Validate with FDA (Live)

    Device Name
    NAEOTOM Alpha
    Date Cleared
    2024-03-28

    (135 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K191040, K220450, K221727

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

    The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions.

    This CT system can be used for low dose lung cancer screening in high risk populations*.

    • As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
    Device Description

    The subject device NAEOTOM Alpha with software version SOMARIS/10 syngo CT VB10 is a Computed Tomography X-ray system which features two continuously rotating tube-detector systems, denominated as A- and B-systems respectively (dual source CT scanner system). The detectors' function is based on photon-counting technology. The NAEOTOM Alpha with SOMARIS/10 syngo CT VB10 produces CT images in DICOM format, which can be used by trained staff for postprocessing applications commercially distributed by Siemens and other vendors as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions. The computer system delivered with the CT scanner is able to run optional post-processing applications.

    The platform software for the NAEOTOM Alpha is syngo CT VB10 (SOMARIS/10 syngo CT VB10). It is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation.

    The software platform provides plugin software interfaces that allow for the use of specific commercially available post-processing software algorithms in an unmodified form from the cleared stand-alone post-processing version.

    AI/ML Overview

    The provided text describes the Siemens NAEOTOM Alpha CT Scanner System with software version SOMARIS/10 syngo CT VB10. While it details extensive non-clinical testing and verification/validation activities, it does not include acceptance criteria for specific AI/software performance metrics (e.g., sensitivity, specificity, accuracy) nor does it describe a comparative clinical study (MRMC or standalone AI performance) with human readers or clinical outcomes. The submission focuses on demonstrating substantial equivalence to predicate devices through engineering verification and validation of new and modified features, rather than a clinical performance study of an AI-powered diagnostic aid.

    Therefore, many of the requested details regarding AI performance acceptance criteria and a study proving device performance in a clinical AI context are not present in the provided document. The document primarily describes the general engineering and regulatory testing performed for a CT system and its software updates.

    However, I can extract information related to the non-clinical testing performed to support the modifications, as well as the types of ground truth used where applicable.

    Here's an attempt to answer the questions based only on the provided text, highlighting where information is absent:


    1. A table of acceptance criteria and the reported device performance

    The document defines "acceptance criteria" generally as part of the system validation tests (workflow and user manual tests, legal and regulatory tests) and system verification tests (system integration, functionality verification, image quality evaluation). For specific new/modified features, the acceptance criteria are generally qualitative or comparative relative to the predicate device, demonstrating comparability or improvement.

    Feature/Non-clinical Supportive TestingAcceptance Criteria (Implicit/Explicit)Reported Device Performance and Conclusion
    FAST 3D Camera / FAST Integrated WorkflowAccuracy of sub-features (FAST Isocentering, FAST Range, FAST Direction) should be comparable to the predicate device with syngo CT VA50."The FAST Isocentering accuracy of the subject device with syngo CT VB10 is comparable to the predicate device with syngo CT VA50, regardless of the camera mounting position." "For the FAST Range feature, the detection accuracy of all body region boundaries is comparable between the subject device with syngo CT VB10 and predicate device with syngo CT VA50." "The FAST Direction pose detection results are of comparable accuracy for subject and predicate device, regardless of the camera mounting position." "Overall, the SOMARIS/10 syngo CT VB10 delivers comparable accuracy to the SOMARIS/10 syngo CT VA50 predicate for the new FAST 3D Camera hardware."
    Multi-Purpose Table (Vitus)Provides sufficient freedom of movement for a mobile C-arm X-ray system to be used for clinical routine without any significant limitations for my needle Laser or 3D Camera."Based on the test results it can be concluded that a CT scanner, equipped with a Multi-Purpose (Vitus) Patient Table, which is installed with enhanced distance (674 mm) to the CT gantry and offers the iCT mode functionality, provides sufficient freedom of movement for a mobile C-arm X-ray system to be used for clinical routine without any significant limitations for my needle Laser or 3D Camera."
    ZeeFree (Cardiac Stack Artefact Correction)1. Reduction of stack misalignment artifacts (discontinuities in vessel structures, anatomical steps, doubling of anatomy). 2. No new artifacts introduced. 3. Equivalent image quality in quantitative standard physics phantom-based measurements (noise, homogeneity, high-contrast resolution, slice thickness, CNR) compared to non-corrected standard reconstruction. 4. Equivalent image quality in quantitative and qualitative phantom-based measurements with respect to metal objects. 5. Algorithm can be successfully applied to phantom data from motion phantom."If misalignment artefacts are identified in non-corrected standard ECG-gated reconstructed sequence or spiral images, the feature 'Cardiac Stack Artefact Correction' (SAC, marketing name: ZeeFree) enables optional stack artefact corrected images, which reduce the number of alignment artefacts." "The SAC reconstruction does not introduce new artefacts, which were previously not present in the non-corrected standard reconstruction." "The SAC reconstruction does realize equivalent image quality in quantitative standard physics phantom-based measurements (ACR, Gammex phantom) in terms of noise, homogeneity, high-contrast resolution, slice thickness and CNR compared to a non-corrected standard reconstruction." "The SAC reconstruction does realize equivalent image quality in quantitative and qualitative phantom-based measurements with respect to metal objects compared to a non-corrected standard reconstruction." "The SAC algorithm can be successfully applied to phantom data if derived from a suitable motion phantom demonstrating its correct technical function on the tested device."
    myNeedle Guide (with myNeedle Detection)1. Accuracy of automatic needle detection algorithm. 2. Reduction of necessary user interactions for navigating to a needle-oriented view."It has been shown that the algorithm was able to consistently detect needle-tips over a wide variety of scans in 90.76% of cases." "Further, the results of this bench test clearly shows that the auto needle detection functionality reduces the number of interactions steps needed to generate a needle-aligned view in the CT Intervention SW. Zero user interactions are required and a needle-aligned view is displayed right away after a new scan, if auto needle detection is switched on in the SW configuration."
    Quantum Spectral Imaging1. T3D reconstructions in Quantumpeak mode possible with sharpest available kernels. 2. Quantumpeak scan mode allows monoenergetic images from 40 to 190 keV. 3. Monoenergetic reconstructions free of artifacts. 4. Measured CT values precisely match reference values. 5. Accuracy of monoenergetic reconstructions in iodine and calcium inserts comparable or better than secondary predicate."The results showed that: with T3D reconstructions from Quantumpeak scan modes, high resolution images with sharp kernel up to Br98 are obtained. The resolution is comparable to other Highresolution scan modes of the NAEOTOM Alpha." "Monoenergetic reconstructions from Quantumpeak scan modes are free of artifacts. Measured CT values precisely match the reference values." "The accuracy of monoenergetic reconstructions in iodine and calcium inserts at the NAEOTOM Alpha is comparable or better than on the secondary predicate device SOMATOM Force."
    Quantum HD CardiacSubstantial equivalence in image quality (UHR and standard resolution spectral images) compared to single source spectral capable 120x0.2mm UHR scan mode."Based on the results it can be concluded that substantial equivalence in image quality is achieved by the images derived from the spectral capable cardiac acquisition mode 96x0.2mm for both, the high-resolution UHR and the standard resolution spectral image cases, compared to the single source spectral capable 120x0.2mm UHR scan mode."
    HD FoV (High Definition Field of View)HU accuracy in extended field of view region."In the phantom study, an HU value accuracy of about +/- 40 HU was achieved with skin-line accuracy of about +/- 3 mm." "HD FoV enables the reconstruction of images while significantly improving the visualization of anatomy in the regions outside the scan field of view of 50 cm."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified in terms of patient counts for clinical validation, as this document focuses on non-clinical (phantom and bench) testing. For the "myNeedle Guide" feature, it states "a wide variety of scans" were used, and the success rate was "90.76% of cases" but doesn't quantify the number of cases. Phantom studies are mentioned for other features without specific numbers of scans/readings.
    • Data Provenance: The testing described is non-clinical (phantom, bench, and system integration/verification testing). There is no mention of human patient data or its country of origin. The manufacturing site is Siemens Healthcare GmbH in Forchheim, Germany, implying the testing likely occurred there.
    • Retrospective or Prospective: Not applicable as it's non-clinical testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.
    • Role of Experts: For the non-clinical tests, "ground truth" generally refers to the known physical properties of the phantoms or the expected performance based on engineering specifications. Human experts are mentioned as trained staff who would use the device, but not as part of a formal ground truth establishment process for the performance studies presented.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable, as no multi-reader human-based test set or clinical study is described. The assessment of performance is based on measurements from phantoms and internal engineering verification.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No MRMC comparative effectiveness study is mentioned. The document focuses on demonstrating that new/modified features of the CT system are comparable to or improve upon the predicate device through non-clinical testing. The device is a CT system, not an AI-powered diagnostic assist that would typically be evaluated with MRMC. The "myNeedle Detection" feature is a software algorithm within the CT system to aid in procedures, but its evaluation is described as a bench test of its detection accuracy and reduction of user interaction steps, not an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: The performance data provided is for the CT system and its integrated software features, including algorithms like ZeeFree and myNeedle Detection. The "algorithm only" performance is implicitly covered in the bench testing of these features (e.g., myNeedle Detection achieving 90.76% detection accuracy). However, this is not presented as a "standalone AI" product in the sense of a distinct AI diagnostic algorithm being submitted for clearance. It's an integrated feature of the CT system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: For the non-clinical tests described, the ground truth is primarily based on:
      • Known physical properties of phantoms: Used for evaluating image quality metrics (noise, homogeneity, resolution, CNR, HU accuracy, etc.) and quantitative measurements.
      • Engineering specifications and expected functional behavior: For features like FAST 3D Camera accuracy and Multi-Purpose Table movement.
      • Reference values: For monoenergetic reconstructions in Quantum Spectral Imaging.
      • Manual verification/observation: For testing user interaction steps in myNeedle Guide.

    8. The sample size for the training set

    • Training Set Sample Size: Not specified. The document does not describe the development or training of AI models. It focuses on the verification and validation of specific software features within the CT system.

    9. How the ground truth for the training set was established

    • Training Set Ground Truth Establishment: Not specified, as no training set or AI model development is described in this document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232322

    Validate with FDA (Live)

    Date Cleared
    2024-03-22

    (232 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K213693, K202014, K191040

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGNETOM system is indicated for use as a magnetic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, and that displays the internal structure and/or function of the head or extremities. Other physical parameters derived from the images may also be produced. Additionally, the MAGNETOM system is intended to produce Sodium images for the head and Phosphorus spectroscopic images and/or spectra for whole body, excluding the head. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

    Device Description

    MAGNETOM Terra and MAGNETOM Terra.X with software syngo MR XA60A include new and modified hardware and software compared to the predicate device, MAGNETOM Terra with software syngo MR E12U. A high level summary of the new and modified hardware and software is provided below: Hardware: New Hardware (Combiner (pTx to sTx), MC-PALI, GSSU control unit, 8Tx32Rx Head coil), Modified Hardware (Main components such as: Upgrade of GPA, New Host computer hardware, New MaRS computer hardware, Upgrade the SEP, The new shim cabinet ASC5 replaces two ACS4 shim cabinets; Other components such as: RFPA, Use of a common MR component which provides basic functionality that is required for all MAGNETOM system types, The multi-nuclear (MNO) option has been modified, OPS module, Cover with UI update on PDD). Software: New Features and Applications (Static B1 shimming, TrueForm (1ch compatibility mode), Deep Resolve Boost, Deep Resolve Gain, Deep Resolve Sharp, Bias field correction (marketing name: Deep RxE), The new BEAT pulse sequence type, BLADE diffusion, The PETRA pulse sequence type, TSE DIXON, The Compressed Sensing (CS) functionality is now available for the SPACE pulse sequence type, The Compressed Sensing (CS) functionality is now available for the TFL pulse sequence type, IDEA, The Scientific Suite), Modified Features and Applications (EP2D DIFF and TSE with SliceAdjust, The Turbo Flash (TFL)), Modified Software / Platform (Stimulation monitoring, "dynamic research labeling"), Other Modifications and / or Minor Changes (Intended use, SAR Calculation and Weight limit reduction for 31P/1H TxRx Flex Loop Coil, X-upgrade for MAGNETOM Terra to MAGNETOM Terra.X, Provide secure MR scanner setup for DoD (Department of Defense) -Information Assurance compliance).

    AI/ML Overview

    The provided text describes the acceptance criteria and supporting study for the AI features (Deep Resolve Boost, Deep Resolve Sharp, and Deep RxE) within the MAGNETOM Terra and MAGNETOM Terra.X devices.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    AI FeatureAcceptance CriteriaReported Device Performance
    Deep Resolve BoostCharacterization by several quality metrics such as peak signal-to-noise ratio (PSNR) and structural similarity index (SSIM). Visual inspection to ensure potential artifacts are detected. Successful passing of quality metrics tests. Work-in-progress packages delivered and evaluated in clinical settings. (Implicit: No misinterpretation, alteration, suppression, or introduction of anatomical information, and potential for faster image acquisition and significant time savings).The impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR) and structural similarity index (SSIM). Additionally, images were inspected visually to ensure that potential artifacts are detected that are not well captured by the metrics listed above. After successful passing of the quality metrics tests, work-in-progress packages of the network were delivered and evaluated in clinical settings with cooperation partners. In a total of seven peer-reviewed publications, the investigations covered various body regions (prostate, abdomen, liver, knee, hip, ankle, shoulder, hand, and lumbar spine) on 1.5T and 3T systems. All publications concluded that the work-in-progress package and the reconstruction algorithm can be beneficially used for clinical routine imaging. No cases have been reported where the network led to a misinterpretation of the images or where anatomical information has been altered, suppressed, or introduced. Significant time savings are reported.
    Deep Resolve SharpCharacterization by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and perceptual loss. Verification and validation by in-house tests including visual rating and evaluation of image sharpness by intensity profile comparisons. (Implicit: Increased edge sharpness).The impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and perceptual loss. In addition, the feature has been verified and validated by in-house tests. These tests include visual rating and an evaluation of image sharpness by intensity profile comparisons of reconstruction with and without Deep Resolve Sharp. Both tests show increased edge sharpness.
    Deep RxE1. During training, the loss (difference to ground truth) is monitored, and the training step with the lowest test loss is taken as the final trained network. 2. Automated unit-tests are set up to test the consistency of the generated output to a previously defined reference output. 3. During verification, the performance of the network is tested on a phantom against the ground truth with a maximal allowed NRMSE of 11% (for 2D network) and 8.7% (for 3D network). 4. The trained final network was used in the clinical study. (Implicit: Increases image homogeneity in a reproducible way on the receive profile, and images acquired with Deep RxE are rated better for image quality in the clinical study).1. During training, the loss, as the difference to a ground truth, is monitored and the training step with the lowest test loss is taken as the final trained network. 2. Automated unit-tests are set up to test the consistency of the generated output to a previously defined reference output. 3. During verification, the performance of the network is tested on a phantom against the ground truth with a maximal allowed NRMSE of 11% (11% for the 2D network and 8.7% for the 3D network were achieved). 4. The trained final network was used in the clinical study. The tests show that Deep RxE increases image homogeneity in a reproducible way on the receive profile. Images acquired with Deep RxE (DL bias field correction) are rated better for image quality than the ones acquired without it in the clinical study that was conducted.

    Note on Acceptance Criteria: The document directly states acceptance criteria for Deep RxE (e.g., NRMSE < 11%). For Deep Resolve Boost and Deep Resolve Sharp, the "acceptance criteria" are more implicitly derived from the described validation and evaluation metrics and outcomes (e.g., "successful passing of quality metrics tests," "increased edge sharpness," "no misinterpretation").

    2. Sample Size Used for the Test Set and Data Provenance

    AI FeatureTest Set Sample SizeData Provenance
    Deep Resolve Boost1,874 2D slices (from validation set)In-house measurements and collaboration partners. (Retrospective, as input data was retrospectively created from ground truth by data manipulation and augmentation).
    Deep Resolve Sharp2,057 2D slices (from validation set)In-house measurements. (Retrospective, as input data was retrospectively created from ground truth by data manipulation).
    Deep RxE23,992 2D slices / 404 3D volumes (validation and test set)All data from two 7T MR systems (MAGNETOM Terra and MAGNETOM Terra.X). (Implied retrospective, as data was separated into independent sets).

    Patient characteristics (Gender/Age) were recorded for Deep RxE: female: 56%, male: 41%, phantom: 3%. Age range 20-80 years. Not recorded for other features.
    Ethnicity was not recorded for any feature. The document states that due to network architecture, attributes like gender, age, and ethnicity are not relevant to training data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Deep Resolve Boost & Deep Resolve Sharp: The document does not mention the use of experts for ground truth establishment for the test set regarding these features. Images were visually inspected and quality metrics were used.
    • Deep RxE: The document mentions that images acquired with Deep RxE were "rated better for image quality than the ones acquired without it in the clinical study that was conducted." This implies expert evaluation, but the number of experts or their qualifications for the test set ground truth for Deep RxE is not explicitly stated.
      • Separately, for the overall device clearance, "radiologist's evaluation reports from two U.S. board-certified radiologists have been provided" for software modifications and new hardware. This is a general statement for the device and not specifically linked to the ground truth of the AI features' test sets.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (like 2+1, 3+1) for establishing ground truth for the test sets of these AI features. For Deep Resolve Boost and Sharp, the ground truth was derived from the acquired datasets themselves, which were then manipulated to create input data. For Deep RxE, the ground truth for phantom testing was a "previously defined reference output" or the acquired data, and for the clinical study, images were "rated better" but the adjudication process for this rating isn't detailed.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is directly mentioned specifically for the AI features (Deep Resolve Boost, Deep Resolve Sharp, or Deep RxE) that compares human readers with vs. without AI assistance. The document alludes to radiologists evaluating images with new software features or comparing images from subject/predicate devices, and for Deep Resolve Boost, it mentions clinical settings and "seven peer-reviewed publications" concluding beneficial use for clinical routine, with reports of "significant time savings." For Deep RxE, images were "rated better for image quality," which implies a reader study, but no details on methodology, number of readers, or specific effect size are provided to quantify human reader improvement with AI assistance.

    6. Standalone (Algorithm Only) Performance

    Yes, standalone performance was conducted for all three AI features:

    • Deep Resolve Boost: Characterized by PSNR and SSIM, and visual inspection.
    • Deep Resolve Sharp: Characterized by PSNR, SSIM, perceptual loss, visual rating, and intensity profile comparisons.
    • Deep RxE: Performance of the network tested on a phantom against ground truth (maximal allowed NRMSE of 11% for 2D, 8.7% for 3D achieved). Unit-tests and a two-step test procedure involving validation on unseen data and RMS error calculation against ground truth.

    These indicate evaluation of the algorithm's performance without a human in the loop, beyond initial visual inspections by evaluators.

    7. Type of Ground Truth Used

    • Deep Resolve Boost: The acquired datasets themselves, retrospectively manipulated through data manipulation and augmentation (under-sampling, lowering SNR, mirroring k-space data) to create input data.
    • Deep Resolve Sharp: The acquired datasets themselves, retrospectively manipulated through data manipulation (cropping k-space data) to create corresponding low-resolution input and high-resolution output/ground truth.
    • Deep RxE: For phantom testing, a "previously defined reference output" was used, and for other evaluations, the acquired datasets themselves were used for comparison against bias field correction methods (homodyne filtering, N4, UNICORN).

    8. Sample Size for the Training Set

    AI FeatureTraining Set Sample Size
    Deep Resolve Boost24,599 2D slices
    Deep Resolve Sharp11,920 2D slices
    Deep RxE119,955 2D slices / 2007 3D volumes

    9. How the Ground Truth for the Training Set Was Established

    • Deep Resolve Boost: The "acquired datasets represent the ground truth for the training and validation." Input data was "retrospectively created from the ground truth by data manipulation and augmentation."
    • Deep Resolve Sharp: The "acquired datasets represent the ground truth for the training and validation." Input data was "retrospectively created from the ground truth by data manipulation."
    • Deep RxE: The document states that "During training the loss, as the difference to a ground truth, is monitored." The method of establishing this initial ground truth is implicitly the raw acquired data from the 7T MRI scanners, as the network aims to correct for B1 inhomogeneities. It also states "All data from the two MR systems were separated into independent training, validation and test datasets," implying the raw or processed raw data served as reference for training.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232856

    Validate with FDA (Live)

    Date Cleared
    2023-12-01

    (77 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K220783, K191040, K231157

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Syngo Carbon Clinicals is intended to provide advanced visualization tools to prepare and process the medical image for evaluation, manipulation and communication of clinical data that was acquired by the medical imaging modalities (for example, CT, MR, etc.)

    The software package is designed to support technicians and physicians in qualitative and quantitative measurements and in the analysis of clinical data that was acquired by medical imaging modalities.

    An interface shall enable the connection between the Syngo Carbon Clinicals software package and the interconnected software solution for viewing, manipulation, communication, and storage of medical images.

    Device Description

    Syngo Carbon Clinicals is a software only Medical Device, which provides dedicated advanced imaging tools for diagnostic reading. These tools can be called up using standard interfaces any native/syngo based viewing applications (hosting applications) that is part of the SYNGO medical device portfolio. These tools help prepare and process the medical image for evaluation, manipulation and communication of clinical data that was acquired by medical imaging modalities (e.g., MR, CT etc.)

    Deployment Scenario: Syngo Carbon Clinicals is a plug-in that can be added to any SYNGO based hosting applications (for example: Syngo Carbon Space, syngo.via etc...). The hosting application (native/syngo Platform-based software) is not described within this 510k submission. The hosting device decides which tools are used from Syngo Carbon Clinicals. The hosting device does not need to host all tools from the Syngo Carbon Clinicals, a desired subset of the provided tools can be used. The same can be enabled or disabled thru licenses.

    AI/ML Overview

    The provided text is a 510(k) summary for Syngo Carbon Clinicals (K232856). It focuses on demonstrating substantial equivalence to a predicate device through comparison of technological characteristics and non-clinical performance testing. The document does not describe acceptance criteria for specific device performance metrics or a study that definitively proves the device meets those criteria through clinical trials or quantitative bench testing with specific reported performance values.

    Instead, it relies heavily on evaluating the fit-for-use of algorithms (Lesion Quantification and Organ Segmentation) that were previously studied and cleared as part of predicate or reference devices, and ensuring their integration into the new device without modification to the core algorithms. The non-clinical performance testing for Syngo Carbon Clinicals focuses on verification and validation of changes/integrations, and conformance to relevant standards.

    Therefore, many of the requested details about acceptance criteria and reported device performance cannot be extracted directly from this document. However, I can provide information based on what is available.


    Acceptance Criteria and Study for Syngo Carbon Clinicals

    Based on the provided 510(k) summary, formal acceptance criteria with specific reported performance metrics for the Syngo Carbon Clinicals device itself are not explicitly detailed in a comparative table against a clinical study's results. The submission primarily relies on the equivalency of its components to previously cleared devices and non-clinical verification and validation.

    The "study" proving the device meets acceptance criteria is fundamentally a non-clinical performance testing, verification, and validation process, along with an evaluation of previously cleared algorithms from predicate/reference devices for "fit for use" in the subject device.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned, a direct table of specific numerical acceptance criteria and a corresponding reported device performance from a clinical study is not present. The document describes acceptance in terms of:

    Feature/AlgorithmAcceptance Criteria (Implicit)Reported Device Performance
    Lesion Quantification Algorithm"Fit for use" in the subject device, with design mitigations for drawbacks/limitations identified in previous studies of the predicate device."The results of phantom and reader studies conducted on the Lesion Quantification Algorithm, in the predicate device, were evaluated for fit for use in the subject device and it was concluded that the Algorithm can be integrated in the subject device with few design mitigations to overcome the drawbacks/limitations specified in these studies. These design mitigations were validated by non-Clinical performance testing and were found acceptable." (No new specific performance metrics are reported for Syngo Carbon Clinicals, but rather acceptance of the mitigations).
    Organ Segmentation Algorithm"Fit for use" in the subject device without any modifications, based on previous studies of the reference device."The results of phantom and reader studies conducted on the Organ Segmentation Algorithm, in the reference device, were evaluated for fit for use in the subject device. And it was concluded that the Algorithm can be integrated in the subject device without any modifications." (No new specific performance metrics are reported for Syngo Carbon Clinicals).
    Overall Device FunctionalityConformance to specifications, safety, and effectiveness comparably to the predicate device."Results of all conducted testing were found acceptable in supporting the claim of substantial equivalence." (General statement, no specific performance metrics). Consistent with "Moderate Level of Concern" software.
    Software Verification & ValidationAll software specifications met the acceptance criteria."The testing results support that all the software specifications have met the acceptance criteria." (General statement).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated for specific test sets in this document for Syngo Carbon Clinicals. The evaluations of the Lesion Quantification and Organ Segmentation algorithms refer to "phantom and reader studies" from their respective predicate/reference devices, but details on the sample sizes of those original studies are not provided here.
    • Data Provenance: Not specified. The original "phantom and reader studies" for the algorithms were likely internal to the manufacturers or collaborators, but this document does not detail their origin (e.g., country, specific institutions). The text indicates these were retrospective studies (referring to prior evaluations).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Number of Experts: Not specified. The document mentions "reader studies" were conducted for the predicate/reference devices' algorithms, implying involvement of human readers/experts, but the number is not stated.
    • Qualifications of Experts: Not specified. It can be inferred that these would be "trained medical professionals" as per the intended user for the device, but specific qualifications (e.g., radiologist with X years of experience) are not provided.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not specified for the historical "reader studies" referenced. This document does not detail the methodology for establishing ground truth or resolving discrepancies among readers if multiple readers were involved.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Comparative Effectiveness Study: The document itself states, "No clinical studies were carried out for the product, all performance testing was conducted in a non-clinical fashion as part of verification and validation activities of the medical device." Therefore, no MRMC comparative effectiveness study for human readers with and without AI assistance for Syngo Carbon Clinicals was performed or reported in this submission. The device is a set of advanced visualization tools, not an AI-assisted diagnostic aid that directly impacts reader performance in a comparative study mentioned here.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Standalone Performance: The core algorithms (Lesion Quantification and Organ Segmentation) were evaluated in "phantom and reader studies" as part of their previous clearances (predicate/reference devices). While specific standalone numerical performance metrics for these algorithms (e.g., sensitivity, specificity, accuracy) are not reported in this document, the mention of "phantom" studies suggests a standalone evaluation component. The current submission, however, evaluates these previously cleared algorithms for "fit for use" within the new Syngo Carbon Clinicals device, implying their standalone performance was considered acceptable from their original clearances.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: Not explicitly detailed. The referenced "phantom and reader studies" imply that for phantoms, the ground truth would be known (e.g., physical measurements), and for reader studies, it would likely involve expert consensus or established clinical benchmarks. However, the exact method for establishing ground truth in those original studies is not provided here.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not specified in this 510(k) summary. The document mentions that the deep learning algorithm for organ segmentation was "cleared as part of the reference device syngo.via RT Image suite (K220783)." This implies that any training data for this algorithm would have been part of the K220783 submission, not detailed here for Syngo Carbon Clinicals.

    9. How the Ground Truth for the Training Set was Established

    • Ground Truth for Training Set: Not specified in this 510(k) summary. As with the training set size, this information would have been part of the original K220783 submission for the organ segmentation algorithm and is not detailed in this document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231157

    Validate with FDA (Live)

    Date Cleared
    2023-07-19

    (86 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K211459, K191040

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    syngo.CT Lung CAD device is a computer-aided detection (CAD) tool designed to assist radiologists in the detection of solid and subsolid pulmonary nodules during review of multi-detector computed tomography (MDCT) from multivendor examinations of the chest. The software is an adjunctive tool to alert the radiologist to regions of interest (ROI) that may be otherwise overlooked.

    The syngo.CT Lung CAD device may be used as a concurrent first reader followed by a full review of the case by the radiologist or as second reader after the radiologist has completed his/her initial read.

    The syngo.CT Lung CAD device may also be used in "solid-only" mode, where potential (or suspected) sub-solid and/or fully calcified CAD findings are filtered out.

    The software device is an algorithm which does not have its own user interface component for displaying of CAD marks. The Hosting Application incorporating syngo. CT Lung CAD is responsible for implementing a user interface.

    Device Description

    Siemens Healthcare GmbH intends to market the syngo.CT Lung CAD which is a medical device that is designed to perform CAD processing in thoracic CT examinations for the detection of solid pulmonary nodules (between 3.0 mm and 30.0mm) and subsolid nodules (between 5.0 mm and 30.0mm) in average diameter. The device processes images acquired with multi-detector CT scanners with 16 or more detector rows recommended.

    The syngo.CT Lung CAD device supports the full range of nodule locations (central, peripheral) and contours (round, irregular).

    The syngo.CT Lung CAD sends a list of nodule candidate locations to a visualization application, such as syngo MM Oncology, or a visualization rendering component, which generates output images series with the CAD marks superimposed on the input thoracic CT images to enable the radiologist's review. syngo MM Oncology (FDA clearanceK211459 and subsequent versions ) is deployed on the syngo.via platform (FDA clearance K191040 and subsequent versions), which provides a common framework for various other applications implementing specific clinical workflows (but are not part of this clearance) to display the CAD marks. The syngo.CT Lung CAD device may be used either as a concurrent first reader, followed by a review of the case, or as a second reader only after the initial read is completed

    AI/ML Overview

    The provided text describes the Siemens syngo.CT Lung CAD (Version VD30) and its substantial equivalence to its predicate device (syngo.CT Lung CAD Version VD20). The primary change in VD30 is the introduction of a "solid-only" mode. The acceptance criteria and study details are primarily focused on demonstrating that the VD30 in "solid-only" mode is not inferior to VD20 in standard mode, and that VD30 in standard mode is not inferior to VD20 in standard mode. Since the document primarily focuses on demonstrating non-inferiority to a predicate device, explicit acceptance criteria values (e.g., minimum sensitivity thresholds) are not explicitly stated as numerical targets. Instead, the acceptance is based on statistical non-inferiority.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied for Non-inferiority)Reported Device Performance (Summary)
    For VD30 (solid-only mode) vs. VD20 (standard mode):
    - Sensitivity of VD30 in solid-only mode is not inferior to VD20 in standard mode.The standalone accuracy has shown that the sensitivity of VD30 in solid-only mode is not inferior to VD20 in standard mode.
    - Mean number of false positives (FPs) per subject is significantly lower with VD30 in solid-only mode.The mean number of false positives per subject is significantly lower with VD30 in solid-only mode.
    - The 2 CAD systems overlap in True Positives (TPs) and FPs.(Implied as part of showing non-inferiority and lower FPs).
    For VD30 (standard mode) vs. VD20 (standard mode):
    - Sensitivity of VD30 in standard mode is not inferior to VD20 in standard mode.The sensitivity of VD30 in standard mode is not inferior to VD20 in standard mode.
    - Mean number of FPs per subject of VD30 in standard mode is not inferior to VD20 in standard mode.The mean number of FPs per subject of VD30 in standard mode is not inferior to VD20 in standard mode.

    2. Sample size used for the test set and the data provenance

    • Sample Size: 712 CT thoracic cases.
    • Data Provenance: Retrospectively collected data from 3 sources:
      • The UCLA study (232 cases)
      • The original PMA study (145 cases)
      • Additional cases (335 cases)

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document differentiates ground truth establishment based on the data source:

    • UCLA data: Reference standard (ground truth) was determined as part of the reader study for the predicate device (K203258). The number and qualifications of experts are not explicitly stated for this subset in the provided text for VD30, but it refers to the predicate clearance.
    • PMA study cases: 18 readers were used. Qualifications are not explicitly stated, but 9 of the 18 readers were needed for declaring a true nodule.
    • Additional cases: 7 readers were used. Qualifications are not explicitly stated, but 4 of the 7 readers were needed for declaring a true nodule.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    The adjudication method varied based on the data source:

    • PMA study cases: 9 out of 18 readers were needed for declaring a true nodule. This suggests a majority consensus from a large panel.
    • Additional cases: 4 out of 7 readers were needed for declaring a true nodule. This also suggests a majority consensus.
    • UCLA data: "Reference standard for the UCLA data was determined as part of the reader study (K203258)." Specific adjudication details for this subset are not provided in this document but are referenced to the predicate device's clearance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A MRMC comparative effectiveness study involving human readers with and without AI assistance is not explicitly described in this document. The statistical analysis performed was a standalone performance analysis to demonstrate substantial equivalence between two CAD versions (VD30 vs VD20), focusing on the algorithm's performance metrics (sensitivity, FPs).

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance analysis was done. The document states: "The standalone performance analysis was designed to demonstrate the substantial equivalence between syngo.CT Lung CAD VD30A (VD30) and the predicate device syngo.CT Lung CAD VD20."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth was established through expert consensus/reader review.

    • For PMA cases: 9 out of 18 readers' consensus.
    • For additional cases: 4 out of 7 readers' consensus.
    • For UCLA data: Reference standard from a reader study.

    8. The sample size for the training set

    The document does not explicitly state the sample size for the training set. It mentions that the algorithm is based on Convolutional Networks (CNN) and that the lung segmentation algorithm for VD30 in particular is "trained on lung CT data including comorbidities for robustness," but the specific number of cases for this training set is not provided.

    9. How the ground truth for the training set was established

    The document does not explicitly describe how the ground truth for the training set was established. It only states that the lung segmentation algorithm was "trained on lung CT data" and that the overall algorithm uses CNNs, implying supervised learning, which would require ground truth annotations. However, the method of obtaining these annotations is not detailed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230561

    Validate with FDA (Live)

    Date Cleared
    2023-03-21

    (21 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K191040, K180563

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Syngo Carbon Space is a software intended to display medical data and to support the review and analysis of medical images by trained medical professionals.

    Syngo Carbon Space "Diagnostic Workspace" is indicated for display, rendering, post-processing of medical data (mostly medical images) within healthcare institutions, for example, in the field of Radiology, Nuclear Medicine and Cardiology.

    Syngo Carbon Space "Physician Access" is indicated for display and rendering of medical data within healthcare institutions.

    Device Description

    Syngo Carbon Space is a software only medical device which is intended to be installed on recommended common IT Hardware. The hardware is not seen as part of the medical device. Syngo Carbon Space is intended to support reviews and analysis of medical images by trained medical practitioners. The software is used in Radiology for reading images and throughout the healthcare institutions for image & result distribution.

    Syngo Carbon Space is a medical device, provided in two variants/options.

    • Diagnostic Workspace (Fat/Thick Client) -
    • -Physician Access (Thin/Web Client)

    In any scenario, both the options can be installed/run on the same machine and be used simultaneously.

    Syngo Carbon Space Diagnostic Workspace provides a reading workspace for Radiology that supports display of medical image data & documents and connects intelligent work tools (diagnostic and non-diagnostic software elements) to enable easy access to the data needed, easy access to external tools and creation of actionable results.

    Syngo Carbon Space Physician Access provides a zero-footprint web application for enterprise-wide viewing of DICOM, non-DICOM, multimedia data and clinical documents to facilitate image and result distribution in the healthcare institution.

    AI/ML Overview

    The provided text is a 510(k) summary for the Syngo Carbon Space (VA30A) device, a medical image management and processing system. The core of this submission is to demonstrate substantial equivalence to a predicate device (Syngo Carbon Space VA20A).

    However, the provided text does not contain the detailed clinical or non-clinical performance test results with acceptance criteria and reported performance values that would typically be presented in a table. It states that "No clinical studies were carried out for the product, all performance testing was conducted in a non-clinical fashion as part of verification and validation activities of the medical device." and "There are no changes to the algorithm and its performance that requires a new bench testing for the subject device. The results/summary from the predicate device is still applicable for the subject device."

    Therefore, I cannot populate the table of acceptance criteria and reported device performance from the provided text directly. The document focuses on demonstrating that the new version of the software (VA30A) maintains the same safety and effectiveness as the previous version (VA20A) by highlighting identical intended use, indications for use, contraindications, and core functionalities. It also details minor enhancements to existing features (like measurement tools and patient jacket functionality) and updates to supported operating systems and browsers, which were validated through non-clinical performance testing.

    Here's a breakdown of the requested information based on the provided text, with explicit notes on what is NOT available:


    1. A table of acceptance criteria and the reported device performance

    This information is NOT explicitly detailed in the provided document. The document states that "all performance testing was conducted in a non-clinical fashion as part of verification and validation activities" and "The testing results support that all the software specifications have met the acceptance criteria." However, it does not enumerate specific acceptance criteria or the quantitative results of these non-clinical tests for the VA30A or its predicate beyond a high-level statement of conformance.

    The document emphasizes that there are "no changes to the algorithm and its performance that requires a new bench testing for the subject device. The results/summary from the predicate device is still applicable for the subject device." This implies that the performance demonstrated by the predicate device (VA20A) is considered valid for the subject device (VA30A), but the specific performance metrics and their acceptance criteria are not provided in this summary.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: NOT specified. The document mentions "non-clinical performance testing" and "verification and validation activities" but does not provide details on the sample size of data or tests used for these validations.
    • Data Provenance: NOT specified. Given that no clinical studies were performed, the data would have been synthetic or from internal testing environments. The origin (country) or nature (retrospective/prospective) of any test data is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • NOT applicable/specified. Since no clinical studies were performed and the testing was non-clinical and focused on software verification and validation, there is no mention of experts establishing ground truth for a clinical test set. The validation would have been against pre-defined software requirements or simulated scenarios.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • NOT applicable/specified. As no clinical studies or reader studies are reported, there's no mention of an adjudication method for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was NOT done. The document explicitly states: "No clinical studies were carried out for the product". The device is a "Medical Image Management and Processing System" and doesn't appear to include AI-assisted diagnostic capabilities (the text mentions "No automated diagnostic interpretation capabilities like CAD are included. All image data are to be interpreted by trained personnel.").

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, in spirit, standalone non-clinical testing was performed for software verification. While not described as an "algorithm only" performance study in the context of diagnostic accuracy, the entire submission is based on "non-clinical performance testing" and "verification and validation activities" to ensure the software functions as intended and meets specifications. The document states: "Performance tests were conducted to test the functionality of the device Syngo Carbon Space. These tests have been performed to assess the functionality of the subject device."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • NOT applicable/specified in the context of clinical ground truth. For non-clinical software testing, ground truth would be established by specified software requirements, functional correctness, performance benchmarks (e.g., speed, data integrity, display accuracy against known inputs), and adherence to standards (DICOM, HL7). The document doesn't detail how specific "ground truths" were established for these technical tests.

    8. The sample size for the training set

    • NOT applicable/specified. As this is not a submission for an AI/ML algorithm that requires a training set for model development, there is no mention of a training set size. The device is a software system for image management and processing.

    9. How the ground truth for the training set was established

    • NOT applicable/specified. (See point 8)
    Ask a Question

    Ask a specific question about this device

    K Number
    K223363

    Validate with FDA (Live)

    Date Cleared
    2023-01-12

    (70 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K191040

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MAMMOVISTA B.smart is a dedicated softcopy review environment for both screening and diagnostic mammography as well as digital breast tomosynthesis. Its user interface and workflow have been optimized to support experienced mammography and tomosynthesis reviewers in both screening and diagnostic reading. Efficiency and reading quality are supported by various specialized features. MAMMOVISTA B.smart provides visualization and image enhancement tools to aid a qualified radiologist in the review of digital mammography and digital breast tomosynthesis datasets, as well as other modalities of breast images.

    Device Description

    MAMMOVISTA B.smart is an optional software application for the Siemens Healthineers syngo.via platform (K191040). MAMMOVISTA B.smart is an image viewing and processing software environment dedicated to breast image display. It is designed to provide the performance required for the high data volume of digital tomosynthesis and the display of multi-modality breast images, such as those from MRI and ultrasound. Individual workflows can be adapted to either screening or diaqnostic purposes.

    MAMMOVISTA B.smart runs on a PC and can be used for Mammography image review together with monitors cleared for Mammography diagnostics. The software solution provides for the display of DICOM compatible information, such as breast density and CAD (Computer Aided Diagnostics) markers.

    AI/ML Overview

    The provided text describes MAMMOVISTA B.smart (VB70), a software device for mammography image review. However, it does not explicitly state acceptance criteria or a dedicated study proving performance against such criteria. The submission is a 510(k) premarket notification for substantial equivalence, comparing the new VB70 version to a predicate device, MAMMOVISTA B.smart VB60 (K212621).

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of quantitative acceptance criteria or device performance metrics for the VB70 version beyond a feature-by-feature comparison to the predicate device. The performance is described in terms of functional equivalence and safety.

    Feature/CriterionAcceptance Criteria (Implied)Reported Device Performance (VB70)
    Functional EquivalenceFunctions identically to predicate device (VB60)."MAMMOVISTA B.smart VB70 has the same indications for use as the predicate device. ...The new software design was completed in accordance with Quality Management System Design Controls comparable to the processes available for the predicate device. The scope of internationally recognized standards compliance is the same as it was for the predicate device." "Verification and validation testing demonstrate that the MAMMOVISTA B.smart performs as intended."
    SafetyNo new safety risks compared to predicate device."It is Siemens' opinion that the MAMMOVISTA B.smart does not introduce any new potential safety risks and is substantially equivalent to the MAMMOVISTA B.smart VB60." Risk analysis completed and controls implemented.
    Compliance with StandardsConforms to relevant software and medical device standards.Complies with IEC 62366-1 2015 Ed 1.0, IEC 62304 2015, Ed.1.1, and NEMA PS 3.1 - 3.20 2016.
    DICOM CompatibilityCompatible with DICOM 3.0 and various modalities.Same as predicate, supports MG, MG Tomo, MR, CR, CT, DR, NM, US, SC, PET.
    Display of CAD MarkersAbility to display third-party CAD markers.Yes, same as predicate.
    Display/Processing of DBTAbility to display and process Digital Breast Tomosynthesis images.Yes, same as predicate.
    Display of Breast DensityAbility to display breast density values.Yes, same as predicate.
    Configuration/SettingsWorkflow, layout, image viewing, and tool settings function as intended. Minimal impact on safety/effectiveness for new settings.Includes automatic study grouping, diagnostic display responsibility, client compatibility check, image rendering performance, layout settings, ReportFlow settings, custom image text settings, image navigation settings, image viewing preferences, image tool settings, workflow settings, screening case detection, double blind reading. "The new settings do not impact safety and effectiveness."
    MR SupportMR Layouts and functionality (e.g., color overlay, time curve analyzer) function as intended. Minimal impact on safety/effectiveness for new MR features.New MR layouts (MR.Kaiser, MR.MPR, MR.DWI, MR.FollowUp), color overlay, time curve analyzer. "The new MR features do not impact safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states: "Non-clinical tests (integration and functional) were conducted on the MAMMOVISTA B.smart during product development." It further notes, "Siemens did not conduct any clinical tests for the subject device." Therefore, the "test set" in this context refers to software testing and verification/validation, not a clinical data set for performance evaluation in a medical context.

    • Sample Size for Test Set: Not specified, as it refers to internal software testing, not a clinical population.
    • Data Provenance: Not applicable for a software-only 510(k) submission based on substantial equivalence and non-clinical testing. No patient data or clinical images are mentioned as being part of a "test set" for performance evaluation in the clinical sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. This was a non-clinical software verification and validation study, not a clinical performance study requiring expert ground truth for medical diagnoses.

    4. Adjudication Method for the Test Set:

    Not applicable. This was a non-clinical software verification and validation study, not a clinical performance study involving human adjudication of medical findings.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. The document explicitly states: "Siemens did not conduct any clinical tests for the subject device." The device is a "softcopy review environment" that provides "visualization and image enhancement tools to aid a qualified radiologist," meaning it's a viewing workstation, not an AI or CAD device that provides interpretations or assists directly with diagnostic accuracy in a quantifiable way like an AI algorithm.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No. The device is a viewing and processing software. It is not an algorithm designed to provide standalone diagnostic interpretations. Its purpose is to "aid a qualified radiologist in the review" of images.

    7. The Type of Ground Truth Used:

    Not applicable. As a software viewing platform, the concept of "ground truth" (pathology, expert consensus, outcomes data) for its own performance is not directly relevant. Its performance is related to its ability to display images correctly, adhere to DICOM standards, and provide tools as specified, which are verified through non-clinical software testing.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a software viewing platform, not an AI/Machine Learning algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set mentioned or implied for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2