Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K221900

    Validate with FDA (Live)

    Date Cleared
    2023-09-29

    (456 days)

    Regulation Number
    862.1120
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stat Profile Prime Plus Analyzer System is indicated for use by healthcare professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, Partial Pressure of Carbon Dioxide (pCO2), Partial Pressure of Oxygen (pO2), Hematocrit, Sodium, Chloride, Ionized Calcium, Ionized Magnesium, Gucose, and Lactate in heparinized capillary whole blood.

    Indication for Use: pH, pCO2, pO2 measurements are used in the diagnosis and treatment of life-threatening acid base disturbances.

    Hematocrit (Hct) measurements of the packed red blood cell volume are used to distinguish normal states, such as anemia and erythrocytosis.

    Glucose (Glu) measurement is used in the diagnosis and treatment of carbohydrate metabolism distuding diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

    Lactate (lactic acid) measurement is used to evaluate the acid-base status of patients suspected of having lactic acidosis.

    Sodium (Na) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance.

    Potassium (K) measurements are used in the diagnosis and treatment of disease conditions characterized by low or high potassium levels.

    Chloride (Cl) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

    Ionized Calcium (iCa) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

    Ionized Magnesium (iMg) measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium).

    Device Description

    The Stat Profile Prime Plus Analyzer System is an analyzer for use in hospital laboratory and point-of-care settings. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

    The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed.

    Primary Sensor Card Port:
    There are two options for the primary sensor card:

    • Primary Sensor Card 1 shall enable and report the following listed analytes: .
      • PO2, PCO2, pH, Hct, tHb, SO2, O2Hb, COHb, MetHb, HHb, Glu, Lactate, Sodium, o Potassium, Chloride, Calcium, Ionized Magnesium
    • Primary Sensor Card 2 shall enable and report the following listed analytes: .
      • PO2, PCO2, pH, Hct, tHb, SO2, Glu, Lactate, Sodium, Chloride, Calcium, Ionized o Magnesium

    Auxiliarv Sensor Card Port:
    The reporting of Creatinine and BUN parameters (or not reporting them) shall be determined by the selection of the Auxiliary Sensor Card

    • . Auxiliary Sensor Card 1 shall enable the Creatinine and BUN parameters
    • Auxiliary Sensor Card 2 shall be a "dummy" sensor card and will not report any parameters. .

    As with the predicate, the Stat Profile Prime Plus Analyzer is a blood gas, co-oximetry, electrolyte, chemistry, and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control is available, as well as an on-board Quality Management System (QMS), and an electronic monitoring approach that ensures the analyzer is working properly.

    The Stat Profile Prime Plus Analyzer accepts samples from syringes, open tubes, and capillary tubes. The sample size for analysis is 135 µL for the complete test panel or 90 µL for the capillary panel.

    Sample collection, preparation and application to the analyzer are the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.

    Stat Profile Prime Plus Analyzer System Components:
    The Stat Profile Prime Plus Analyzer System is comprised of the following components.

    • . Stat Profile Prime Plus Analyzer System
    • Primary Sensor Cartridge .
    • Auxiliary Sensor Cartridge .
    • Stat Profile Prime Plus Auto-Cartridge Quality Control Pack
    • Stat Profile Prime Plus Calibrator Cartridge
    • Stat Profile Prime Plus External Ampule Control
    • . IFU/Labeling

    Sample Types:
    The Stat Profile Prime Plus Analyzer System accepts lithium heparinized arterial, venous, and capillary whole blood.

    Measured Parameters:
    The Stat Profile Prime Plus Analyzer measures:

    • . pH
    • . Partial Pressure of Carbon Dioxide (pCO2)
    • Partial Pressure of Oxygen (pO2) ●
    • Hematocrit (Hct) ●
    • . Glucose (Glu)
    • . Lactate (Lac)
    • Sodium (Na) ●
    • Potassium (K)
    • Chloride (CI)
    • . Ionized Calcium (iCa)
    • . lonized Magnesium (iMg)
    AI/ML Overview

    The Nova Biomedical Stat Profile Prime Plus Analyzer System is undergoing a 510(k) premarket notification to expand its indications for use to include capillary whole blood specimen testing for pH, pCO2, pO2, Sodium (Na+), Potassium (K+), Chloride (Cl-), Ionized Calcium (Ca2+), Ionized Magnesium (Mg2+), Glucose, Lactate, and Hematocrit. The study described focuses on demonstrating the substantial equivalence of the Stat Profile Prime Plus Analyzer system to its predicate device, the Nova Biomedical Stat Profile pHOx Ultra Analyzer, specifically for capillary whole blood samples.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for substantial equivalence are primarily demonstrated through method comparison and precision studies. While explicit numerical acceptance criteria for each parameter (e.g., specific ranges for slope, intercept, r-value in method comparison, or max SD/CV% for precision) are not directly stated in the provided text as a standalone table, the conclusion sections for each study indicate that the device "met the clinical accuracy acceptance criteria" or "met the performance criteria for precision." The reported performance is shown in the tables below, which are the primary evidence for meeting the implicit acceptance criteria.

    Method Comparison (Clinical Accuracy - Comparison to Predicate Device)

    ParameterN (Combined)Altered Samples (Combined)Whole Blood Range (Combined)SlopeInterceptr
    pH249186.790-7.7290.98940.07360.9942
    pO2, (mmHg)251207.5-567.11.00060.83200.9976
    pCO2, (mmHg)245147.4-183.11.0075-0.59690.9968
    Hct, (%)2411018-550.99000.80110.9876
    Na, (mM)2431283.0-195.61.0129-2.22440.9885
    K, (mM)245141.34-18.530.99400.04160.9987
    Cl, (mM)2431264.5-191.60.99440.34940.9856
    Ca, (mM)247160.37-2.460.99000.01550.9932
    Mg, (mM)249180.13-1.220.96590.02140.9811
    Glu, (mg/dL)2451428-4520.99500.90410.9969
    Lac, (mM)243120.4-17.61.00010.01190.9989

    Precision (Laboratory and Point-of-Care Settings)

    The precision data is presented across multiple tables (Tables 4, 5, 6, 7, 8, 9, 10). Rather than reiterating all data here, the text explicitly states:

    • "The precision data for all samples in capillary mode met the within run and between analyzer imprecision specifications for the Prime Plus analyzers." (Summary of Capillary Mode Within Sample Precision)
    • "This study demonstrates the Stat Profile Prime Plus analyzer exhibits clinically acceptable imprecision specifications for pH, pCO2, pO2, sodium (Na+), chloride (C1-), potassium (K+), ionized calcium (Ca2+), ionized magnesium (Mg2+), glucose, lactate, and hematocrit measured by the Stat Profile Prime Plus Analyzer System in Capillary mode." (Conclusion of Within-Run Imprecision - Capillary Mode Fingerstick (External POC))
    • "The analyzer used for this evaluation met the performance criteria for within sample precision on capillary fingerstick specimens run by POC operators." (Conclusion of Within-Sample Imprecision - Capillary Mode Fingerstick (Internal POC))
    • "The Stat Profile Prime Plus analyzers provided consistently reliable performance throughout the evaluation study. The analyzers used for this evaluation met the acceptance criteria for precision." (Conclusion of Within-Run Imprecision - Capillary Mode)

    The acceptance criteria are therefore implicitly met by the reported r-values nearing 1.0 and slopes nearing 1.0 with intercepts near 0 for method comparison, and the CV% and SD values falling within acceptable limits (though the limits themselves are not numerically specified in the provided text).

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison Test Set (Capillary Mode):

      • For each measured parameter, the sample size (N) ranged from 118 to 123 at the ER site and 123 to 128 at the Hemodialysis site. The combined sample size (N) for each parameter ranged from 241 to 251.
      • Provenance: This was a prospective clinical study conducted at two external Point-of-Care (POC) sites within the United States (an Emergency Room and a Hemodialysis Unit). Some samples (less than 10%, indicating "Altered Samples" ranging from 5 to 10 for each site) were altered to cover the full dynamic range. These were "de-identified and discarded arterial blood specimens" for the external precision study (implicitly reflecting human samples, though the exact origin beyond "external POC site" is not specified beyond being collected from patients).
    • Precision Test Set (Capillary Mode):

      • Within Run Precision (Internal Lab): 20 replicates for each parameter, tested on two Prime Plus analyzers from venous blood transferred to capillary tubes. This appears to be lab-based, controlled samples.
      • Within Sample Precision (Internal Lab): 2 replicates from 30 different donors (Total N=60 for each analyte) of capillary whole blood. This implies human subjects.
      • Within-Run Imprecision (External POC): Sample analysis involved transferring discarded arterial blood specimens from a lithium heparin syringe to three balanced heparin capillary tubes. The number of unique discarded specimens is not explicitly stated but "each whole blood specimen" suggests multiple, distinct specimens were used.
      • Within-Sample Imprecision (Internal POC - Fingerstick): Capillary whole blood was collected via fingerstick puncture from individuals, with 2 replicates for each. N=60 for all sample pairs. This explicitly involves human subjects/donors.
      • Within-Run Imprecision (Internal Study - Lab): 5 different concentrations of deidentified venous whole blood specimens per analyte. Each concentration was run on 3 Prime Plus analyzers, 5 days, 1 run/day, 8 replicates/run/level. This totals 120 (5 concentrations * 3 analyzers * 5 days * 8 replicates) data points per analyte for the "N" value in Table 10. These are likely controlled lab samples simulating human blood.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The provided text does not explicitly state the number of experts used or their specific qualifications for establishing ground truth.

    • For the method comparison study, the predicate device (Nova Stat Profile pHOx Ultra Analyzer) serves as the "ground truth" or reference method for comparison. The performance of this predicate device itself is assumed to be established and accepted.
    • For the precision studies, the intrinsic analytical performance of the device is assessed, rather than against a human expert's interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable as the device measures objective chemical and physical parameters rather than interpreting images or clinical signs that would require human adjudication. The "ground truth" is the measurement from the predicate device or the inherent value in the sample for precision studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, Effect Size

    This information is not applicable as the device is an in-vitro diagnostic (IVD) analyzer for quantitative measurements, not an AI imaging or diagnostic algorithm requiring human reader performance studies. The study focuses on instrument performance and equivalence rather than human reader improvement with AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the Loop Performance) Was Done

    Yes, the studies conducted (method comparison and precision) are standalone performance evaluations of the device's accuracy and precision in measuring the analytes. There is no "human-in-the-loop" aspect to the analytical performance being evaluated; the device provides direct quantitative measurements.

    7. The Type of Ground Truth Used

    • Method Comparison: The "ground truth" or reference standard for comparison was the predicate device, the Nova Stat Profile pHOx Ultra Analyzer. This is a comparative method where the new device's performance is assessed against an already legally marketed and accepted device.
    • Precision Studies: The "ground truth" for precision is the measured value itself and its statistical variation across multiple runs or samples. It's an assessment of the device's inherent reproducibility and repeatability, not against an external truth source like pathology or outcomes data. Human samples (venous and capillary whole blood) were used to test performance under realistic conditions.

    8. The Sample Size for the Training Set

    The provided text does not mention a training set as this is not a machine learning or AI-driven device in the sense of requiring an explicit training phase with labeled data in the way an imaging algorithm would. This is an analytical instrument based on established sensor technology and algorithms. Therefore, discussions of training sets and their sample sizes are typically not relevant for this type of device submission. The device uses "the same sensor technology, measurement algorithms, formulations of the internal and external controls, and calibrator cartridge" as its predicate, implying a well-established design.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is discussed or implied to be applicable for this type of analytical device in the provided context, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193246

    Validate with FDA (Live)

    Date Cleared
    2020-01-24

    (60 days)

    Product Code
    Regulation Number
    862.1120
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K173797

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stat Profile Prime Plus Analyzer System is intended for use by healthcare professionals in clinical laboratory settings and for point-of-care usage for quantitative determination of pH, Partial Pressure of Carbon Dioxide (pCO2) and Partial Pressure of Oxygen (pO2) in heparinized arterial and venous whole blood.

    pH, pCO2 and pO2 Measurements are used in the diagnosis and treatment of life-threatening acid base disturbances.

    Device Description

    The Stat Profile Prime Plus Analyzer System is a low cost, low maintenance analyzer for hospital laboratory and point-of-care settings. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

    The Stat Profile Prime Plus Analyzer has slots to accommodate two sensor cartridges (Primary and Auxiliary). The analyzer will determine the configuration of the system by detecting which sensor cards are installed.

    As with the predicate, the Stat Profile Prime Plus Analyzer is a blood gas, co-oximetry, electrolyte, chemistry, and hematology analyzer with an enhanced test menu and multiple quality control options. Both traditional internal and external quality control is available, as well as an on-board Quality Management System (QMS), and an electronic monitoring approach that insures the analyzer is working properly at all times.

    The Stat Profile Prime Plus Analyzer accepts samples from syringes and open tubes. The minimum sample size for analysis is 135 µL.

    Sample collection, preparation and application to the same as for the previously cleared predicate. The end user can select which analytes are to be tested in the panel.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the demonstration of "substantial equivalence" to the predicate device (Nova Stat Profile pHOx Ultra Analyzer System and K173797 - Stat Profile Prime Plus Analyzer System). The performance of the proposed device is compared to the predicate device and established CLSI guidelines for method comparison and imprecision.

    Criterion TypeAcceptance Criteria (Implicitly from CLSI/Substantial Equivalence)Reported Device Performance (Stat Profile Prime Plus Analyzer)
    Method Comparison (POC vs Lab)Agreement between POC measurements and laboratory measurements for pH, pCO2, and pO2, demonstrating substantial equivalence to the predicate and clinical acceptability.pH: N=432, Range 6.832 - 7.931, Slope 0.9930, Intercept 0.0500, r 0.9976. 95% CI of Bias: MDL (7.1-7.6) for Lower Limit (7.099, 7.595) and Upper Limit (7.102, 7.601) pO2: N=432, Range 11.5 - 555.2, Slope 1.0109, Intercept -1.5391, r 0.9989. 95% CI of Bias: MDL (40-160) for Lower Limit (38.4, 159.7) and Upper Limit (40.6, 160.7) pCO2: N=428, Range 14.0 - 199.2, Slope 0.9848, Intercept 0.9958, r 0.9963. 95% CI of Bias: MDL (20-75) for Lower Limit (19.6, 74.5) and Upper Limit (20.9, 75.2)
    Total ImprecisionAcceptable Total SD and %CV for pH, pCO2, and pO2 across multiple levels of quality control materials and different POC personnel, according to CLSI EP5-A2T.Level 1 (pH 7.223, pCO2 60.0 mmHg, pO2 76.5 mmHg): pH Total SD 0.008; pCO2 Total SD 3.6, %CV 5.9; pO2 Total SD 2.4, %CV 3.1 Level 2 (pH 7.428, pCO2 37.2 mmHg, pO2 112.2 mmHg): pH Total SD 0.006; pCO2 Total SD 2.3, %CV 6.1; pO2 Total SD 3.0, %CV 2.6 Level 3 (pH 7.627, pCO2 20.1 mmHg, pO2 148.9 mmHg): pH Total SD 0.008; pCO2 Total SD 1.0, %CV 4.8; pO2 Total SD 3.4, %CV 2.3
    Within-Run Whole Blood PrecisionAcceptable Within-Run SD and %CV for pH, pCO2, and pO2 using fresh, native whole blood samples by POC operators, demonstrating consistent performance in replicate measurements.pH: SD values ranging from 0.003 to 0.011 across 5 samples.pCO2 (mmHg): %CV values ranging from 0.70% to 1.62% across 5 samples.pO2 (mmHg): %CV values ranging from 0.80% to 4.73% across 5 samples. (Specific mean, SD, %CV provided for each of the 5 samples in Table 3)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Method Comparison:
      • N = 432 for pH and pO2 (venous & arterial whole blood specimens combined)
      • N = 428 for pCO2 (venous & arterial whole blood specimens combined)
    • Sample Size for Total Imprecision: 40 runs (duplicate measurements each day for 20 days) for each of 3 levels of quality control materials.
    • Sample Size for Within-Run Whole Blood Precision: 10 replicate measurements for each of 5 different native whole blood samples at each site.
    • Data Provenance: The studies were conducted at 3 Point-of-Care (POC) sites including a Cardiothoracic Intensive Care Unit (CTICU), an Emergency Department (ED), and a Respiratory Therapy Lab (RT). The data provenance refers to real-world samples tested by clinical staff in typical POC settings. The report does not specify the country of origin, but given it's an FDA submission, it's typically understood to be within the U.S.
    • Nature of Data: The method comparison used "discarded blood gas specimens" and "quality control materials" in addition to "fresh, native whole blood samples" for within-run precision. This indicates retrospective usage of discarded patient samples and prospective collection for precision studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The ground truth for the method comparison study was established by "laboratory measurements" (presumably by a predicate or established laboratory analyzer). The document does not specify the number of experts or their qualifications who performed these laboratory measurements. The term "POC personnel" refers to the users of the device under test, not the ground truth establishment.

    4. Adjudication Method for the Test Set

    The document does not describe any specific "adjudication method" in the sense of multiple experts reviewing results and reaching a consensus. For method comparison studies of this type (quantitative measurements), the reference method (laboratory measurement) typically serves as the "ground truth" against which the new device's performance is compared, using statistical methods (slope, intercept, correlation, bias). No expert adjudication is typically involved beyond ensuring the reference method itself is properly calibrated and operated.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly done as described in the context of human readers improving with AI assistance. This device is an analyzer, not an AI-assisted diagnostic tool that interprets images or signals requiring human interpretation. The study involved different POC personnel operating the device, which is a form of multi-user testing to assess "imprecision performance" in a real-world setting, but it's not an MRMC study comparing human performance with and without AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the device operates in a standalone manner to measure pH, pCO2, and pO2. The "Point-of-Care (POC) study" assessed the device's performance when operated by healthcare professionals in clinical settings, but the device itself generates the quantitative measurements automatically. The "bench testing" mentioned, performed prior to the POC study, also represents standalone performance. The values in the tables are direct measurements from the device, not interpretations aided by humans.

    7. The Type of Ground Truth Used

    • For Method Comparison Studies: The ground truth was established by "laboratory measurements" using presumably a reference laboratory analyzer or the predicate device.
    • For Imprecision Performance: The ground truth was based on the expected values of the "quality control materials" used.
    • For Within-Run Whole Blood Precision: The study evaluated the device's consistency when repeatedly measuring "fresh, native whole blood samples", where the ground truth is simply the intrinsic value of the sample, and the measurement aims to show reproducibility.

    8. The Sample Size for the Training Set

    This document only describes performance testing for regulatory submission (510(k)). It does not mention or provide details about a "training set" as would be relevant for machine learning algorithms. This device is a diagnostic instrument based on established electrochemical principles, not an AI/ML device that requires a training set in the conventional sense.

    9. How the Ground Truth for the Training Set was Established

    As no training set is described for this type of device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1