510(k) Data Aggregation

    K Number
    K093233

    Validate with FDA (Live)

    Date Cleared
    2009-12-04

    (51 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K061101,K081928,K090073

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diagnostic Hybrids, Inc. device, D3 FastPoint L-DFA RSV/MPV Identification Kit is intended for the qualitative identification of respiratory syncytial virus and human metapneumovirus in nasal and nasopharyngeal swabs and aspirates/washes specimens from patients with signs and symptoms of respiratory infection by direct detection of immunofluorescence using monoclonal antibodies (MAbs).

    It is recommended that specimens found to be negative for respiratory syncytial virus after examination of the direct specimen result be confirmed by cell culture. Specimens found to be negative for human metapneumovirus after examination of the direct specimen results should be confirmed by an FDA-cleared human metapneumovirus molecular assay. Negative results do not preclude respiratory syncytial virus and human metapneumovirus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

    Device Description

    The D3 FastPoint L-DFA RSV/MPV Identification Kit uses a blend (called a "L-DFA Reagent'") of viral antigen-specific murine monoclonal antibodies that are directly labeled with either R-phycoerythin (PE) (respiratory syncytial virus) or fluorescein isothiocyanate (FITC) (human metapneumovirus) for the rapid identification of respiratory syncytial virus and human metapneumovirus in nasal and nasopharyngeal swabs and aspirates from patients with signs and symptoms of respiratory infection.

    The cells to be tested are derived from respiratory specimens from patients with signs and symptoms of respiratory infection. The cells are permeabilized and stained concurrently in a liquid suspension format with the L-DFA Reagent. After incubating at 35℃ to 37℃ for 5-minutes, the stained cell suspensions are rinsed with 1X PBS. The rinsed cells are pelleted by centrifugation and then re-suspended with the Resuspension Buffer and loaded onto a specimen slide well. The cells are examined using a fluorescence microscope. Cells infected with RSV will exhibit golden-yellow fluorescence due to the PE. Cells infected with hMPV will exhibit apple-green fluorescence due to the FITC. Non-infected cells will exhibit red fluorescence due to the Evans Blue counter-stain. Nuclei of intact cells will exhibit orange-red fluorescence due to the propidium iodide.

    AI/ML Overview

    Acceptance Criteria and Study for D3 FastPoint L-DFA RSV/MPV Identification Kit

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state numerical acceptance criteria for all performance metrics. However, for a 510(k) submission, implied acceptance is often "substantially equivalent" to predicate devices, and for clinical performance, high sensitivity and specificity are expected. The reproducibility study explicitly aims for 100% agreement.

    Note: The "acceptance criteria" are inferred based on standard expectations for diagnostic device performance and the detailed reporting of study outcomes, particularly the 100% agreement for reproducibility and the very high sensitivity/specificity for RSV. The lower sensitivity for hMPV in NP swab samples might be within acceptable limits given the context of medical device approval for challenging targets.

    Performance MetricAcceptance Criteria (Implied/Explicit)Reported Device Performance (D3 FastPoint L-DFA RSV/MPV Identification Kit)
    Reproducibility (Overall Agreement with Expected Result)100% (Implied by study design expecting full agreement)100% (280/280) across all sites and panel members
    Limit of Detection (LoD) - RSVThe lowest dilution at which at least 9/10 replicates are detected.100 infected cells/mL
    Limit of Detection (LoD) - hMPVThe lowest dilution at which at least 9/10 replicates are detected.100 infected cells/mL
    Analytical Reactivity (Inclusivity) - RSVDetection of various RSV strains at 10x LoD.All 3 tested RSV strains detected at 10x LoD.
    Analytical Reactivity (Inclusivity) - hMPVDetection of various hMPV strains at 10x LoD.All 4 tested hMPV strains detected at 10x LoD.
    Clinical Sensitivity (RSV - NP wash/aspirate)High sensitivity for diagnosis.98.6% (204/207) [95% CI: 95.8-99.7%]
    Clinical Specificity (RSV - NP wash/aspirate)High specificity for diagnosis.99.8% (462/463) [95% CI: 98.8-100%]
    Clinical Sensitivity (hMPV - NP wash/aspirate)High sensitivity for diagnosis.68.8% (55/80) [95% CI: 57.4-78.7%]
    Clinical Specificity (hMPV - NP wash/aspirate)High specificity for diagnosis.100.0% (614/614) [95% CI: 99.4-100%]
    Clinical Sensitivity (RSV - NP swab)High sensitivity for diagnosis.97.5% (39/40) [95% CI: 86.8-99.9%]
    Clinical Specificity (RSV - NP swab)High specificity for diagnosis.100.0% (647/647) [95% CI: 99.4-100%]
    Clinical Sensitivity (hMPV - NP swab)High sensitivity for diagnosis.54.5% (24/44) [95% CI: 38.8-69.9%]
    Clinical Specificity (hMPV - NP swab)High specificity for diagnosis.100.0% (632/632) [95% CI: 99.4-100%]

    2. Sample Size Used for the Test Set and Data Provenance

    • Reproducibility Test Set:
      • Sample Size: A reproducibility panel consisting of 5 members (low RSV, low hMPV, mixed RSV/hMPV, mixed hMPV/RSV, negative). Each panel member was tested daily in two separate runs for 5 days by 4 different laboratories, resulting in 40 total runs. This yielded 280 total tests (across all panel members and runs) with individual results reported for expected positive and negative wells.
      • Data Provenance: The study was conducted at four different laboratories. The document does not specify the country of origin but implies U.S. clinical laboratories (referencing "U.S. clinical laboratories" for clinical performance). It's a prospective study in the sense that the testing itself was performed to assess reproducibility.
    • Limit of Detection (LoD) Test Set:
      • Sample Size: Dilution series of infected model cells were used. For each virus (RSV and hMPV A1), 10 replicate microscope slides were prepared for each dilution level. The specific number of dilutions isn't explicitly stated as a single number but spanned from 1000 infected cells/mL down to 0.8 or 1.5 infected cells/mL, with 10 replicates for each dilution.
      • Data Provenance: Laboratory study, likely internal to the manufacturer or a contracted lab.
    • Analytical Reactivity (Inclusivity) Test Set:
      • Sample Size: 3 RSV virus strains and 4 hMPV virus strains were evaluated. For each strain, "low concentration infected cell suspensions (approximately 4% cells infected, 25 to 50 infected cells)" were prepared.
      • Data Provenance: Laboratory study.
    • Clinical Performance Test Set:
      • Sample Size: 1519 total respiratory specimens (nasal and nasopharyngeal swabs and aspirates/washes).
      • Data Provenance: Prospective studies at 4 geographically diverse U.S. clinical laboratories during the 2009 respiratory virus seasons (January 2009 - March 2009). The specimens were "excess, remnants of respiratory specimens that were prospectively collected from symptomatic individuals suspected of respiratory infection, and were submitted for routine care or analysis by each site, and that otherwise would have been discarded." Individual specimens were delinked from patient identifiers.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number or qualifications of experts for establishing ground truth as a separate role. Instead, the ground truth for clinical performance was established using a composite comparator method:

    • RSV: Direct Specimen Fluorescent Antibody (DSFA) test with an FDA-cleared predicate device, followed by viral culture confirmation of all negatives from the comparator DSFA test.
    • hMPV: DSFA with an FDA-cleared predicate device, followed by confirmation of all negative specimens from the comparator DSFA test using a validated hMPV real-time RT-PCR, which was then followed by bi-directional sequencing analysis.

    This implies that the "ground truth" was determined by the results of these established and confirmed laboratory methods, rather than by human expert consensus or adjudication of raw images/output from the test device solely.

    4. Adjudication Method for the Test Set

    For the clinical studies, an explicit "adjudication method" involving human experts reviewing conflicting results is not detailed. Instead, a composite comparator algorithm was used to define "true positive" and "true negative":

    • "True positive" for RSV was defined as any sample that tested positive by the comparator DSFA test or viral culture.
    • "True positive" for hMPV was defined as any sample that tested positive by the comparator DSFA test OR had bi-directional sequencing data meeting pre-defined quality acceptance criteria that matched hMPV sequences in GenBank.
    • "True negative" was defined as any sample that tested negative by both the comparator DSFA test and either viral culture (for RSV) or the hMPV real-time RT-PCR comparator assay (for hMPV).

    This approach essentially pre-defines how discordant results between screening and confirmatory tests contribute to the final ground truth, replacing a separate human adjudication step.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This study is an evaluation of an in-vitro diagnostic device (IVD), specifically a direct fluorescent antibody (DFA) test, which is read by trained laboratory personnel, but the study focuses on the device's performance against comparator methods, not on comparing reader performance with and without AI assistance (as it is not an AI device).

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the studies presented are essentially standalone performance evaluations of the device. The D3 FastPoint L-DFA RSV/MPV Identification Kit is an immunofluorescent assay where a human technician observes fluorescent staining patterns under a microscope. However, the performance metrics (sensitivity, specificity) are for the device's ability to detect the viral antigens in specimens, without involving a study design where human readers using the device are compared to human readers using another method, or AI assistance. The results in the tables reflect the diagnostic performance of the kit itself when used according to its instructions.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for the clinical performance studies was based on a composite comparator method combining:

    • FDA-cleared predicate DFA devices
    • Viral culture (for RSV)
    • Validated real-time RT-PCR with bi-directional sequencing analysis (for hMPV)

    This is a form of reference standard composite, aiming for a highly accurate and confirmed diagnosis of the presence or absence of the target viruses.

    8. The Sample Size for the Training Set

    The document does not explicitly describe a separate "training set" or "validation set" in the context of machine learning. This is a traditional in vitro diagnostic device, not an AI/ML-based device.

    All the described studies (reproducibility, LoD, analytical reactivity, clinical performance) contribute to the overall evidence for the device. The 1519 clinical specimens (fresh nasal/nasopharyngeal wash/aspirate and swab specimens) can be considered the test set for evaluating clinical performance.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, there is no explicit "training set" mentioned in the context of machine learning. The studies assess the performance of the pre-developed D3 FastPoint L-DFA RSV/MPV Identification Kit using the specified ground truth methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091171

    Validate with FDA (Live)

    Date Cleared
    2009-09-11

    (142 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K061101, K081928, K090073

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diagnostic Hybrids, Inc. device, D3 FastPoint L-DFA Respiratory Virus Identification Kit is intended for the qualitative identification of influenza A virus, influenza B virus, respiratory syncytial virus, human metapneumovirus, adenovirus and to screen for the presence of parainfluenza virus types 1, 2, and 3 in nasal and nasopharyngeal swabs and aspirates/washes specimens from patients with signs and symptoms of respiratory infection by direct detection of immunofluorescence using monoclonal antibodies (MAbs).

    It is recommended that specimens found to be negative for influenza A virus, influenza B virus, respiratory syncytial virus, adenovirus or parainfluenza viruses after examination of the direct specimen result be confirmed by cell culture. Specimens found to be negative for human metapneumovirus after examination of the direct specimen results should be confirmed by an FDA cleared human metapneumovirus molecular assay. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

    Performance characteristics for influenza A virus detection and identification were established when influenza A (H3N2) and influenza A (H1N1) were the predominant influenza A strains circulating in the United States. Since influenza strains display antigenic drift and shift from year to year, performance characteristics may vary. If infection with a novel influenza A virus is suspected, based on clinical and epidemiological screening criteria communicated by public health authorities, collect specimens following appropriate infection control precautions and submit to state or local health departments, for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

    Device Description

    The D3 FastPoint L-DFA Respiratory Virus Identification Kit uses three blends (each called a "L-DFA Reagent") of viral antigen-specific murine monoclonal antibodies that are directly labeled with either R-PE (influenza A virus, respiratory syncytial virus, and parainfluenza virus) or fluorescein (influenza B virus, metapneumovirus, and adenovirus) for the rapid identification of respiratory viruses in nasal and nasopharyngeal swabs and aspirates from patients with signs and symptoms of respiratory infection.

    Kit Components:

    1. D3 FastPoint L-DFA Influenza A/Influenza B Reagent, 4.0-mL. One dropper bottle containing a mixture of PE-labeled murine monoclonal antibodies directed against influenza A virus antigens and FITC-labeled murine monoclonal antibodies directed against influenza B virus antigens. The buffered, stabilized, aqueous solution contains Evans Blue and propidium iodide as counter-stains and 0.1% sodium azide as preservative.
    2. D3 FastPoint L-DFA RSV/MPV Reagent, 4.0-mL. One dropper bottle containing a mixture of PE-labeled murine monoclonal antibodies directed against respiratory syncytial virus antigens and FITC-labeled murine monoclonal antibodies directed against metapneumovirus antigens. The buffered, stabilized, aqueous solution contains Evans Blue and propidium iodide as counter-stains and 0.1% sodium azide as preservative.
    3. D3 FastPoint L-DFA PIV/Adenovirus Reagent, 4.0-mL. One dropper bottle containing a mixture of PE-labeled murine monoclonal antibodies directed against parainfluenza virus types 1, 2, or 3 antigens and FITClabeled murine monoclonal antibodies directed against adenovirus antigens. The buffered, stabilized, aqueous solution contains Evans Blue and propidium iodide as counter-stains and 0.1% sodium azide as preservative.
    4. 40X PBS Concentrate, 25-mL. One bottle of 40X PBS concentrate containing 4% sodium azide (0.1% sodium azide after dilution to 1X using de-mineralized water).
    5. Re-suspension Buffer, 6.0-mL. One bottle of a buffered glycerol solution and 0.1% sodium azide.
    6. D3 FastPoint L-DFA Respiratory Virus Antigen Control Slides, 5-slides. Five individually packaged control slides containing 6 wells with cell culture-derived positive and negative control cells. Each positive well is identified as to the virus infected cells present, i.e., influenza A virus, influenza B virus, respiratory syncytial virus, metapneumovirus, parainfluenza virus, and adenovirus. The negative wells contain noninfected cells. Each slide is intended to be stained only one time.

    The cells to be tested are derived from respiratory specimens from patients with signs and symptoms of respiratory infection. The cells are permeabilized and stained concurrently in a liquid suspension format in 3 separate vials, each containing one of the 3 above reagents. After incubating at 35℃ to 37℃ for 5 minutes, the stained cell suspensions are rinsed with 1X PBS. The rinsed cells are pelleted by centrifugation and then re-suspended with the resuspension buffer and loaded onto a specimen slide well. The cells are examined using a fluorescence microscope. Cells infected with influenza A virus, respiratory syncytial virus, or parainfluenza virus types 1, 2 and 3 will exhibit goldenyellow fluorescence due to the PE. Cells infected with influenza B virus, metapnemovirus or adenovirus will exhibit apple-green fluorescence due to the FITC. Non-infected cells will exhibit red fluorescence due to the Evans Blue counter-stain. Nuclei of intact cells will exhibit orange-red fluorescence due to the propidium iodide.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the D3 FastPoint L-DFA Respiratory Virus Identification Kit, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria (e.g., "Sensitivity must be >X%"). Instead, it presents the results of reproducibility and clinical performance studies, implying that these results met the necessary standards for clearance. The "Reported Device Performance" column reflects the results from the clinical sensitivity and specificity studies presented in Tables 5.10-5.15 (for NP wash/aspirate) and 5.16-5.21 (for NP swab). For analytical performance, the reproducibility and LOD results are provided.

    Criterion TypeAcceptance Criteria (Implicit)Reported Device Performance (Summary)
    Analytical Performance
    ReproducibilityConsistent detection of viral antigens across different sites and runs.Influenza A/B Reagent: Total agreement 99.3% (278/280). RSV/hMPV Reagent: Total agreement 100% (280/280). HPIV/Adenovirus Reagent: Total agreement 100% (280/280).
    Limit of Detection (LoD)Detection of specific viral strains at low concentrations.Flu A: 50 infected cells/mL Flu B: 50 infected cells/mL RSV: 100 infected cells/mL hMPV A1: 100 infected cells/mL Adenovirus: 100 infected cells/mL HPIV-1: 100 infected cells/mL HPIV-2: 25 infected cells/mL HPIV-3: 50 infected cells/mL
    Analytical Reactivity (Inclusivity)Detection of various strains of targeted viruses.Detected all tested strains of Influenza A (13), Influenza B (7), RSV (3), hMPV (4), HPIV (3), and Adenovirus (10).
    Clinical Performance (NP Wash/Aspirate)Acceptable sensitivity and specificity compared to comparator methods.Flu A: Sensitivity 84.8%, Specificity 99.5% Flu B: Sensitivity 81.8%, Specificity 100.0% RSV: Sensitivity 98.6%, Specificity 99.8% Adenovirus: Sensitivity 92.3%, Specificity 100.0% HPIV: Sensitivity 92.0%, Specificity 99.3% hMPV: Sensitivity 68.8%, Specificity 100.0%
    Clinical Performance (NP Swab)Acceptable sensitivity and specificity compared to comparator methods.Flu A: Sensitivity 87.7%, Specificity 99.8% Flu B: Sensitivity 87.9%, Specificity 99.8% RSV: Sensitivity 97.5%, Specificity 100.0% Adenovirus: Sensitivity 100.0%, Specificity 100.0% (Note: Low prevalence, caution advised) HPIV: Sensitivity 92.9%, Specificity 100.0% hMPV: Sensitivity 54.5%, Specificity 100.0%

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Test Set:

      • Total Specimens Evaluated: 1519
      • Provenance: Prospectively collected excess remnants of respiratory specimens from symptomatic individuals suspected of respiratory infection.
      • Country of Origin: 4 geographically diverse U.S. clinical laboratories.
      • Retrospective or Prospective: Prospective studies (January 2009 - March 2009).
      • Specific Breakdown for Clinical Performance Tables:
        • NP Wash/Aspirate (Sites 1, 2, and 3 combined): Number of specimens varies per virus (e.g., 637 for Flu A, 694 for hMPV).
        • NP Swab (Sites 3 and 4 combined): Number of specimens varies per virus (e.g., 689 for Flu A, 675 for hMPV).
    • Analytical Test Set (Reproducibility):

      • Sample Size: 5 randomized panel members for each of the 3 panels (Influenza A/B, RSV/hMPV, HPIV/Adenovirus). Each panel was tested daily in two separate runs for 5 days by four different laboratories (40 total runs per virus group). This means 40 replicates for each panel member for a given virus.
      • Provenance: Proficiency-level antigen control slides with infected cells.
    • Analytical Test Set (Limit of Detection):

      • Sample Size: 10 replicate microscope slides for each dilution level of 8 characterized respiratory virus isolates.
      • Provenance: Dilution series of infected model cells.
    • Analytical Test Set (Analytical Reactivity/Inclusivity):

      • Sample Size: Low concentration infected cell suspensions (approximately 4% cells infected, 25-50 infected cells) for various viral strains for each of the 3 reagents.
      • Provenance: Culture isolates of various influenza A (13 strains), influenza B (7 strains), RSV (3 strains), hMPV (4 strains), HPIV (3 strains), and Adenovirus (10 strains).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish the ground truth for the clinical test set. Instead, it defines "True" positive and "True" negative based on a composite comparator method:

    • For Influenza A, Influenza B, RSV, Parainfluenza, and Adenovirus: Direct Specimen Fluorescent Antibody (DSFA) test with an FDA cleared device, and viral culture confirmation of all negatives (as determined by the comparator DSFA test).
    • For Human Metapneumovirus (hMPV): DSFA with an FDA cleared device, and confirmation of all negative specimens (as determined by the comparator DSFA test) using a validated hMPV real-time RT-PCR followed by bi-directional sequencing analysis.

    4. Adjudication Method for the Test Set

    The adjudication method used for establishing the ground truth for the clinical test set was a composite comparator method. This means results from multiple established methods (DSFA, viral culture, and for hMPV, RT-PCR with sequencing) were combined to determine the "true" status of a specimen. There is no mention of a specific expert panel adjudication method like "2+1" or "3+1" for interpreting these comparator results in case of discrepancies; rather, the definition of "true" positive/negative indicates the hierarchy of methods (e.g., positive by DSFA or viral culture is "true" positive).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is a diagnostic kit read by a human using a fluorescence microscope, but the study focuses on the kit's performance against comparator methods, not on human reader improvement with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the D3 FastPoint L-DFA Respiratory Virus Identification Kit is a diagnostic kit that relies on a human reading the results under a fluorescence microscope. It is not an AI algorithm. The performance presented is of the kit as used by a human.

    7. The Type of Ground Truth Used

    The ground truth used for the clinical test set was a composite comparator method, which included:

    • FDA cleared Direct Specimen Fluorescent Antibody (DSFA) tests
    • Viral Culture
    • Validated hMPV real-time RT-PCR followed by bi-directional sequencing analysis (for hMPV only)
    • NCBI GenBank database matching with acceptable E-values for bi-directional sequencing data.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI development. This device is a diagnostic reagent kit, not an AI algorithm. The studies conducted are for analytical and clinical validation of the kit itself.

    9. How the Ground Truth for the Training Set was Established

    As there is no "training set" in the context of this diagnostic kit, this question is not applicable. The ground truth for the comparator methods was established using established laboratory techniques and FDA-cleared devices, as described in point 7.