510(k) Data Aggregation

    K Number
    K121040

    Validate with FDA (Live)

    Date Cleared
    2012-08-30

    (147 days)

    Product Code
    Regulation Number
    862.1600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K082462, K013850, K033063, K823480, K810615

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMARTLYTE is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium, chloride, calcium and lithium in serum, plasma, whole blood, pre-diluted urine samples. In addition, the analyzer can also measure sodium, chloride and calcium in dialysate samples.

    The SMARTLYTE Sodium Assay is intended to measure sodium in whole blood, serum, plasma, pre-diluted urine and dialysate on the SMARTLYTE Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

    The SMARTLYTE Potassium Assay is intended to measure potassium in whole blood, serum, plasma, urine and dialysate. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

    The SMARTLYTE Chloride Assay is intended to measure the level of chloride in whole blood, serum, plasma, urine and dialysate. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

    The SMARTLYTE Calcium Assay is intended to measure ionized calcium levels in whole blood, plasma, serum, and dialysate. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

    The SMARTLYTE Lithium Assay is intended to measure lithium (from the drug lithium carbonate) in whole blood, plasma, and serum. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).

    Device Description

    The SMARTLYTE Nat, K", CT, Ca*, Li Electrolyte Analyzer which can test serum, plasma, whole blood, pre-diluted urine samples, dialysate solutions and QC materials is identical to GEMLYTE Electrolyte Analyzer ( cleared under K082462) which tests serum, plasma, whole blood, pre-diluted urine samples, and QC materials. Both are microprocessor-controlled analyzers which utilize ion-selective electrodes for the measurement of sodium, potassium, chloride, calcium and lithium in serum, plasma, whole blood, pre-diluted urine samples, dialysate solutions and QC materials. The analyzer self-calibrates using Diamond Diagnostics Fluid Pack (510(k) 013850) every 4 hours through out the day or on request. Sodium, potassium, chloride and calcium are commonly measured for use in the diagnosis and management of patients with a broad range of renal, metabolic and cardiovascular disorders. Lithium is a drug used to treat mental illness. Mission controls (510 (k) 033063) are the recommended . quality control material to be used daily.

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and the study that proves the device meets them, based on your requested format:

    Acceptance Criteria and Device Performance Study for SMARTLYTE Electrolyte Analyzer

    The SMARTLYTE Electrolyte Analyzer is an automated, microprocessor-controlled device using ion-selective electrodes to measure sodium, potassium, chloride, calcium, and lithium in various sample types, including dialysate. The study aimed to demonstrate the substantial equivalence of the SMARTLYTE to its predicate device, the AVL 9180 (K961458), particularly for dialysate measurements, which represent a modification from previously cleared uses.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the predicate device's performance characteristics or defined in the "Reproducibility" section as maximum allowable CVs or SDs. The reported device performance is from the "Precision Dialysate" and "Dialysate Comparison" sections.

    Note on Acceptance Criteria: The document primarily compares the SMARTLYTE's performance to the predicate device's expected values and establishes its own internal precision targets. For the method comparison, "good correlation to predicate with correlation coefficients typically greater than 0.99 and slope values between 0.96 and 1.04" serves as a key acceptance criterion for the clinical claims (correlation with a predicate device). For precision, specific CV/SD targets are given.

    ParametersMatrixAcceptance Criteria (Predicate/Target)Reported Device Performance (SMARTLYTE)
    Precision
    Na+DialysateWithin Run C.V. ≤ 1% Between Run C.V. ≤ 2%Within Run C.V.: 0.32-0.47% Between Run C.V.: 0.32-0.54%
    K+DialysateWithin Run C.V. ≤ 2% Between Run C.V. ≤ 3%Within Run C.V.: 0.28-0.87% Between Run C.V.: 0.88-1.27%
    Cl-DialysateWithin Run C.V. ≤ 2% Between Run C.V. ≤ 3%Within Run C.V.: 0.24-0.89% Between Run C.V.: 0.53-1.44%
    Ca++DialysateWithin Run SD ≤ 0.02 Between Run SD ≤ 0.06Within Run SD: 0.0087-0.0155 (CV not calculated in source) Between Run SD: 0.0076-0.0226 (CV not calculated in source)
    Na+UrineWithin Run C.V. ≤ 5% Between Run C.V. ≤ 5% (Predicate: ≤ 7%)(Not explicitly reported in this document for Urine; previously cleared K082462)
    K+UrineWithin Run C.V. ≤ 5% Between Run C.V. ≤ 5%(Not explicitly reported in this document for Urine; previously cleared K082462)
    Cl-UrineWithin Run C.V. ≤ 5% Between Run C.V. ≤ 5%(Not explicitly reported in this document for Urine; previously cleared K082462)
    LinearityDialysateR-squared ≈ 0.99 (implied by "good correlation")Sodium: 0.9995 Potassium: 0.9976 Chloride: 0.9998 Calcium: 0.9972
    Slope ≈ 1 (implied)Sodium: 0.9806 Potassium: 1.0031 Chloride: 0.9996 Calcium: 0.9857
    Intercept ≈ 0 (implied)Sodium: 2.53 Potassium: -0.37 Chloride: 1.60 Calcium: 0.06
    Method Comparison (to predicate AVL 9180)DialysateR > 0.99, Slope 0.96-1.04 (Stated by manufacturer)Sodium: R=0.9989, Slope=1.0183 Potassium: R=0.9996, Slope=0.9882 Chloride: R=0.9966, Slope=0.9825 Calcium: R=0.9956, Slope=1.0021
    Measuring RangeDialysateNa: 40-205 mEq/L K: 1.7-15 mEq/L Cl: 50-200 mEq/L Ca: 0.3-5.5 mmol/L (Predicate range)Na: 40-200 mEq/L (Linearity confirmed: 24-205); Method comparison supported: 40-205 mEq/L K: 1.7-15 mEq/L (Linearity confirmed: 1.4-15.1); Method comparison supported: 1.7-11 mEq/L Cl: 50-200 mEq/L (Linearity confirmed: 40-202); Method comparison supported: 50-205 mEq/L Ca: 0.3-5.5 mmol/L (Linearity confirmed: 0.4-5.0); Method comparison supported: 0.3-5.5 mmol/L

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision (Dialysate):
      • Within-run: 30 replicates for each of three dialysate samples (low, mid, high concentrations) per measurand.
      • Between-run (Total precision): 40 replicates (4 groups of 10 replicates over 10 days) for each of three dialysate samples per measurand.
    • Linearity (Dialysate):
      • Sodium: 34 samples
      • Potassium: 36 samples
      • Chloride: 34 samples
      • Calcium: 48 samples
    • Method Comparison (Dialysate):
      • Sodium: 43 samples
      • Potassium: 56 samples
      • Chloride: 51 samples
      • Calcium: 43 samples

    Data Provenance: The document implies these were prospective studies conducted to support the 510(k) submission for the specific device modifications. No explicit country of origin is stated for data collection, but given it's a US FDA submission, it's typically assumed to be North American or adhering to international standards accepted by the FDA. The samples were "dialysates collected" and some were "spiked or diluted" to span the measurement ranges, suggesting a mix of clinical and artificially prepared samples. Prior data for whole blood, serum, plasma, and urine were "previously cleared (K082462)".

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is an in-vitro diagnostic device measuring analytes. The "ground truth" for the test set is established by the reference methods or the predicate device that the SMARTLYTE is compared against, not by human expert assessment of images or clinical cases. The reference for comparison would be the AVL 9180 predicate device.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (like 2+1, 3+1 for consensus among experts) are used in studies involving human interpretation or subjective assessments. This study evaluates the analytical performance of an IVD device against a predicate or defined analytical targets.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study involves multiple human readers evaluating cases with and without AI assistance to measure a change in human performance (e.g., diagnostic accuracy, reading time). The SMARTLYTE is an automated IVD device; it does not involve human readers interpreting output in the same way an imaging AI would.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described (precision, linearity, method comparison) represent standalone performance of the SMARTLYTE device. The device operates automatically to measure analyte concentrations. The "human-in-the-loop" aspect is limited to operating the device and collecting samples, not interpreting an AI's output to make a diagnosis.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance studies can be defined as:

    • Reference Intervals/Expected Values: For linearity studies, expected values based on known dilutions of stock solutions serve as the ground truth.
    • Predicate Device Measurements: For the method comparison study, the measurements from the predicate device (AVL 9180) on the same dialysate samples served as the comparative ground truth.
    • Internal Precision Targets: For the precision studies, the acceptance criteria (max CV/SD) are internal performance targets, and the reported values are measured against these.

    8. The Sample Size for the Training Set

    Not applicable. The SMARTLYTE Electrolyte Analyzer is an IVD device based on ion-selective electrode technology, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. Its performance is based on its hardware, chemistry, and pre-programmed algorithms, which are calibrated and verified through analytical studies.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device's technology. The device uses established physical and chemical principles (ion-selective electrodes). Calibrations are performed using certified calibration fluids.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121140

    Validate with FDA (Live)

    Date Cleared
    2012-08-22

    (128 days)

    Product Code
    Regulation Number
    862.1665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K031159, K033063

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PROLYTE Electrolyte Analyzer Na*/K*/Cli/Li is designed for clinical laboratory use by laboratory professionals to assess the levels of Sodium, Potassium, Chloride and Lithium found in whole blood, serum, plasma, and urine of patients. The analysis is performed in-vitro, and neither the analyzer nor any of its components come in contact with the patient.

    This analyzer is used by trained laboratory technicians in clinical laboratories to aid in the diagnosis and treatment of paiients with electrolyte imbalance as well as monitoring the lithium levels for those patients taking lithium. These locations routinely conform to CLA regulations, and conduct daily quality control programs.

    The PROLYTE is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium, chloride and lithium in whole blood, plasma, serum, and urine samples.

    The PROLYTE Sodium Assay is intended to measure sodium in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer Na 7K 7CI/Li. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

    The PROLYTE Potassium Assay is intended to measure potassium in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer Na */K */C//Li. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

    The PROLYTE Chloride Assay is intended to measure the level of chloride in whole blood, plasma, serum, and urine on the PROLYTE Electrolyte Analyzer Na*/K*/Cli/Li. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

    The PROLYTE Lithium test system is intended to measure lithium (from the drug lithium carbonate) in whole blood, plasma, or serum on the PROLYTE Electrolyte Analyzer Na*/K*/Cl'/Li. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).

    It is For In Vitro Diagnostic Use.

    Device Description

    The PROLYTE Electrolyte Analyzer Na*/K'/Cl/Li* is intended to be a direct replacement for the PROLYTE Electrolyte Analyzer Na K * (K102959). The PROLYTE Na * , K , CI / Li * Electrolyte Analyzer has all the Elookofyo Khayzerman Proyte Electrolyte Analyzer Na/K'/Cl (K102959), with the added feature that the CI ISE sensor may be replaced by a Li* ISE sensor by the end-user.

    The PROLYTE Electrolyte Analyzer Na*/K*/Cli/Lif is designed for clinical laboratory professionals to assess the levels of Sodium, Potassium, Chloride and Lithium found in whole blood, plasma, and serum of patients. Sodium, Potassium, and Chloride are also assessed in urine of patients. The analysis and scrum of pairons. Ocularly be analyzer nor any of its components come in contact with the patient.

    This bench top analyzer is used by trained laboratory technicians in clinical laboratories to aid in the ring bonen top and freatment of patients with electrolyte imbalance as well for the monitoring drug levels in those patents taking Lithium. These locations routinely conform to CLIA regulations, and conduct daily control programs.

    It uses the Diamond Diagnostics Fluid Pack (K031159) which contains in a sealed package the two calibrants required for calibration along with a flush solution and a waste container. The analyzer can be programmed to self-calibrate at set intervals or on request. The analyzer establishes a slope for the electrode by means of the two calibrants. A value for a sample is determined by direct comparison to the calibrants. Mission Controls (K033063) are the recommended quality control material to be used daily.

    AI/ML Overview

    The provided text describes the PROLYTE Electrolyte Analyzer Na+/K+/Cl-/Li+ and its performance based on non-clinical and clinical testing, primarily for the Lithium (Li+) assay as other analytes were previously cleared.

    Here's an analysis of the acceptance criteria and study details based on the provided input:


    Acceptance Criteria and Device Performance for PROLYTE Electrolyte Analyzer (K121140)

    1. Table of Acceptance Criteria and Reported Device Performance (Lithium only)

    The acceptance criteria are implied by the "Spec, BLD" (Specification for Blood) rows in the precision tables and the comparison against predicate device performance.

    Precision (for Lithium: Blood, Plasma, Serum)

    ParameterAcceptance Criteria (Predicate / Specification)Reported Device Performance (Worst Case Recorded)Pass/Fail
    Within Run Precision (S.D.)
    Whole Blood (0.3 – 2.0 mEq/L reference range)≤ 0.03 mEq/L (Specific BLD)0.010 mEq/L (Mid-range)Pass
    Plasma (0.3 – 2.0 mEq/L reference range)≤ 0.03 mEq/L (Specific BLD)0.01336 mEq/L (High range)Pass
    Serum (0.3 – 2.0 mEq/L reference range)≤ 0.03 mEq/L (Specific BLD)0.01166 mEq/L (High range)Pass
    Run to Run Precision (S.D.)
    Whole Blood (0.3 – 2.0 mEq/L reference range)≤ 0.09 mEq/L (Specific BLD)0.014 mEq/L (V Low range)Pass
    Plasma (0.3 – 2.0 mEq/L reference range)≤ 0.09 mEq/L (Specific BLD)0.02444 mEq/L (High range)Pass
    Serum (0.3 – 2.0 mEq/L reference range)≤ 0.09 mEq/L (Specific BLD)0.0425 mEq/L (High range)Pass

    Linearity (Lithium)

    ParameterAcceptance Criteria (Implied by Predicate)Reported SlopeInterceptPass/Fail
    Whole BloodHigh R² (near 1), Slope near 10.9854-0.02850.9979Pass
    PlasmaHigh R² (near 1), Slope near 10.9985-0.03150.9997Pass
    SerumHigh R² (near 1), Slope near 11.00680.0190.9998Pass

    Method Comparison (Lithium vs. Predicate GEMLYTE K082462)

    ParameterAcceptance Criteria (Implied correlation)Reported SlopeInterceptPass/Fail
    Whole BloodHigh R² (near 1), Slope near 11.00960.07190.9947Pass
    PlasmaHigh R² (near 1), Slope near 10.98060.12920.9971Pass
    SerumHigh R² (near 1), Slope near 10.99770.01330.9881Pass

    Note on Precision "F" (Fail) for Plasma and Serum within-run precision at V High:
    The tables show "F" for Plasma and Serum at the "V High" concentration for within-run precision (Spec, BLD = 0.03). However, the narrative states, "Lithium demonstrated precision within the limits defined below in the reference range, 0.3 to 2.0 mEq/L." The "V High" concentrations (4.41513 and 4.30820 mEq/L) are outside this stated reference range. The "Spec, BLD" of ≤ 0.03 appears to be the acceptance criterion for the reference range (0.3 - 2.0 mEq/L). The reported standard deviations for these V High values, while exceeding 0.03, may be acceptable for concentrations outside the strict clinical reference range or may indicate a different specification applied to the extreme high end which is not explicitly stated. Given the overall conclusion of safety and effectiveness, it's implied these were deemed acceptable.


    2. Sample Size Used for the Test Set and Data Provenance

    • Precision Test Set (Lithium):

      • Within Run: 30 replicates for each of 5 concentration levels (V Low, Low, Mid, High, V High) for whole blood, plasma, and serum matrices. (Total of 30 samples/level * 5 levels = 150 measurements per matrix).
      • Run to Run (Total Precision):
        • Plasma/Serum: 40 replicates (measured twice each morning and twice each afternoon for 10 days) for most concentration levels. Some levels had 41 or 44 replicates.
        • Whole Blood: 4 groups of 10 replicates, separated by calibration (total of 40 measurements).
      • Data Provenance: Not explicitly stated, but the context of an FDA 510(k) submission for IVD devices typically implies data collected in a controlled laboratory setting, often in the US, following GLP/GCP guidelines for clinical studies. It is prospective data collection specifically for these tests.
    • Linearity Test Set (Lithium):

      • Whole Blood: 100 samples
      • Plasma: 38 samples
      • Serum: 44 samples
      • Data Provenance: Not explicitly stated, but derived from serially diluted stock solutions. Prospective data.
    • Method Comparison Test Set (Lithium):

      • Whole Blood: 94 patient samples
      • Plasma: 100 patient samples
      • Serum: 91 patient samples
      • Data Provenance: "patient samples" implies retrospective or prospective collection from clinical pathology/laboratory settings. The country of origin is not specified, but typically for US regulatory submissions, the data is often generated in-country or by international labs adhering to recognized standards. It's not explicitly stated if it's retrospective or prospective collection, but method comparison studies often use prospectively collected "real-world" patient samples or archived samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This is a medical device for measuring analytes (Sodium, Potassium, Chloride, Lithium) in patient samples. The "ground truth" (or reference method) for such devices is typically established through:

    • Reference laboratory methods: For quantitative measurements, this refers to established, highly accurate, and precise analytical methods.
    • Comparison to a legally marketed predicate device: The method comparison study directly uses the predicate device's measurements as the reference for comparison, implying the predicate itself is the "ground truth" for regulatory purposes in this context.

    No 'experts' in the sense of clinicians or radiologists are typically used to establish ground truth for this type of quantitative IVD assay. The accuracy is determined by analytical methods and comparison to a proven device, not expert consensus on qualitative interpretation.


    4. Adjudication Method for the Test Set

    Not applicable. Adjudication typically applies to qualitative or semi-quantitative assessments where multiple human interpreters might disagree (e.g., image interpretation). For a quantitative analyzer measuring discrete chemical levels, the output is a numerical value, and the ground truth relies on established analytical methods or comparison to a predicate device, not human subjective assessment needing adjudication.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is an In Vitro Diagnostic (IVD) device for quantitative chemical analysis, not an imaging device or AI-assisted diagnostic tool that involves human readers interpreting cases. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study or analysis of human reader improvement with AI assistance is not relevant to this device type.


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the studies presented (precision, linearity, method comparison) are inherently standalone performance studies for the device (the "algorithm" in a broader sense of the instrument's measurement principle and processing) without direct human intervention in the measurement process itself once initiated. The device provides a direct numerical output.


    7. Type of Ground Truth Used

    • Precision: Internal consistency of the device, measured against statistical definitions (standard deviation, coefficient of variation). The "truth" is the mean of multiple measurements.
    • Linearity: Expected values based on serial dilution of known stock solutions. The "truth" is the theoretical concentration.
    • Method Comparison: Measurements from a legally marketed predicate device (GEMLYTE K082462). This is the primary "ground truth" for demonstrating substantial equivalence. For the analytes (Na, K, Cl, Li) this is a direct quantitative comparison.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI development. This device uses ion-selective electrode technology, which is a well-established analytical principle based on electrochemical measurements, not a machine learning model that requires a discrete training phase. The "training" of the device is analogous to calibration, which uses known calibrant solutions.


    9. How the Ground Truth for the Training Set Was Established

    As noted above, a "training set" for an AI model is not applicable here. However, for the operational aspect of the device (calibration), the "ground truth" is established by:

    • The use of "Fluid Pack (K031159) which contains in a sealed package the two calibrants required for calibration." These calibrants would have precisely known concentrations of the target analytes.
    • The analyzer "establishes a slope for the electrode by means of the two calibrants." This process essentially "trains" the electrodes to accurately translate electrochemical signals into analyte concentrations.