510(k) Data Aggregation

    K Number
    K042089

    Validate with FDA (Live)

    Date Cleared
    2004-11-24

    (113 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K030059, K022747, K994292, K002045

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage Water Purification System is a stand-alone water treatment system designed to pretreat and purify water for use in hemodialysis applications. The purified water can be used in preparation of dialysate, equipment rinsing and disinfection, and dialyzer reprocessing. It is intended to be used in a variety of environments, including acute or chronic care facility or home.

    Device Description

    The NxStage Water Purification System is a stand-alone water System designed to pre-treat and purify water for use with hemodialysis applications. The purified water can be used in preparation of dialysate, equipment rinsing and disinfection, and dialyzer reprocessing. The System is intended to be used in a variety of environments, including an acute or chronic care facility or the home. There are no known contraindications with the use of the NxStage Water Purification System.

    The System is single patient use and is comprised of three major components: 1) a pretreatment system which consists of a check valve, sediment filter and air vent to remove trapped air, sediment, and any impurities, 2) a Pump Controller unit which contains the control and safety software that controls the flow rate, monitors pressure, water quality and system operation and 3) a replaceable resin system that removes organic and inorganic chemical contaminants as well as microbiological contaminants such as bacteria and endotoxins. The System flow rate is specified by the end user.

    With proper use, the System will produce water that meets or exceeds the requirements of the standard issued by the American National Standard Institute and the Association for the Advancement of Medical Instrumentation: ANSI/AAMI RD62:2001, Water Treatment Equipment for Hemodialysis Applications.

    AI/ML Overview

    Here's an analysis of the provided text regarding the NxStage Water Purification System, focusing on acceptance criteria and study details.

    Important Note: The provided document is a 510(k) Summary of Safety & Effectiveness and an FDA clearance letter. These documents typically summarize testing and do not contain detailed study protocols, raw data, or comprehensive statistical analyses. Therefore, some information, especially regarding specific sample sizes for training sets, adjudication, or detailed effect sizes for MRMC studies, is not present in this type of document and cannot be extracted. The focus of this document is on demonstrating substantial equivalence to predicate devices and meeting established standards.


    Acceptance Criteria and Device Performance

    The core acceptance criteria for the NxStage Water Purification System are defined by its ability to meet or exceed the requirements of the American National Standard Institute and the Association for the Advancement of Medical Instrumentation (ANSI/AAMI) RD62:2001, Water Treatment Equipment for Hemodialysis Applications.

    Table 1: Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard)Reported Device Performance
    ANSI/AAMI RD62:2001, Water Treatment Equipment for Hemodialysis Applications: Specifies permissible levels of various impurities (e.g., chemical contaminants, microbiological contaminants like bacteria and endotoxins) in water used for hemodialysis applications."With proper use, the System will produce water that meets or exceeds the requirements of the standard issued by the American National Standard Institute and the Association for the Advancement of Medical Instrumentation: ANSI/AAMI RD62:2001, Water Treatment Equipment for Hemodialysis Applications."
    System Functionality Indicators:
    Removal of trapped air, sediment, and impurities.Accomplished by a pretreatment system (check valve, sediment filter, air vent).
    Control of flow rate.Managed by a Pump Controller unit.
    Monitoring of pressure, water quality, and system operation.Managed by a Pump Controller unit with control and safety software.
    Removal of organic and inorganic chemical contaminants.Achieved by a replaceable resin system.
    Removal of microbiological contaminants (bacteria and endotoxins).Achieved by a replaceable resin system.

    Study Information

    Based on the provided text, the study conducted was a non-clinical performance testing to demonstrate that the NxStage Water Purification System is substantially equivalent to predicate devices and meets the specified ANSI/AAMI standard.

    1. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    * Test Set Sample Size: The document does not specify a numerical sample size for the test set (e.g., number of water samples tested, number of systems tested). It broadly refers to "Performance testing." For a water purification system, this would typically involve testing multiple units over a period of time with various water sources.
    * Data Provenance: The document does not specify the country of origin of the data. It is internal testing conducted by "NxStage Medical." The study is described as "performance testing," which implies it was prospective testing of the device's functionality.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    * N/A. This device is a water purification system, not an imaging or diagnostic device that requires expert interpretation for establishing ground truth. Ground truth for water purity is established through chemical and microbiological laboratory analysis processes against industry standards.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    * N/A. Adjudication methods like 2+1 or 3+1 are relevant for subjective interpretations, often in medical imaging. For a water purification system, the performance is assessed against objective, measurable chemical and microbiological standards, which do not typically involve human adjudication in this manner.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    * No. This is not applicable to a water purification system. An MRMC study is relevant for diagnostic devices involving human interpretation.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    * Yes, in essence. The performance testing demonstrated the device's ability to purify water to specified standards "with proper use" of the system itself. While human interaction is required to operate and maintain the system, the purification process and its effectiveness are tested in a standalone manner (device's output measured against standards), independent of human "interpretation" of the water quality. The "Pump Controller unit which contains the control and safety software" functions as the "algorithm only" component controlling the purification process.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    * Objective Laboratory Analysis against Industry Standards. The ground truth for water purity is established by quantitative chemical and microbiological analysis of the treated water samples, compared directly to the specified maximum contaminant levels outlined in the ANSI/AAMI RD62:2001 standard.

    7. The sample size for the training set:
    * N/A. The concept of a "training set" is typically associated with machine learning or AI models. This device is a physical water purification system with a controller and safety software. While the software itself would have undergone development and testing, there's no mention or implication of machine learning that would necessitate a "training set" in the conventional sense.

    8. How the ground truth for the training set was established:
    * N/A. As mentioned above, the concept of a training set and its associated ground truth derivation (like for AI development) does not apply to this type of device based on the provided information.