Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K042769

    Validate with FDA (Live)

    Date Cleared
    2004-11-24

    (50 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K992727, K960548

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adseal is a biocompatible root canal sealer for permanent sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points). Adseal is intended for use by qualified healthcare personnel trained in its use.

    Device Description

    Adseal root canal sealer is a two component paste:paste device based upon epoxy-amine resin chemistry. This sealer is easy to mix and adapts closely to the walls of the prepared root canal and provides outstanding long-term dimensional stability with minimal shrinkage upon setting. The device consists of tow components, the epoxy resin paste (Paste A) and the aminecontaining paste (Paste B); portions of which are mixed prior to insertion into the root canal. This two component system reacts via an epoxide-amine chemical reaction to cause setting. It may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points). Paste A and Paste B are contained, separately, with the chambers of a two component plastic syringe, packaged with a disposable applicator.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the Adseal Root Canal Sealer, focusing on acceptance criteria and supporting studies:

    This 510(k) summary is for a traditional medical device (a root canal sealer), and as such, the nature of its "acceptance criteria" and "study" is different from what one might expect for an AI/ML powered device. For such devices, "acceptance criteria" are typically based on established performance standards for the material, and the "study" is a series of laboratory tests demonstrating compliance with those standards. There is no AI component involved here, so questions related to AI performance, ground truth, expert adjudication, or MRMC studies are not applicable.

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are derived from the international standard ISO 6876, "Dental root canal sealing materials." The 510(k) implicitly indicates that Adseal meets these criteria by stating "Similar Physical Properties" and then listing those properties under the Adseal column in the substantial equivalence comparison table.

    Acceptance Criteria (from ISO 6876)Reported Device Performance (Adseal)
    FluidityMeets ISO 6876 requirements
    Working TimeMeets ISO 6876 requirements
    Film ThicknessMeets ISO 6876 requirements
    RadiopacityMeets ISO 6876 requirements
    Solubility & DisintegrationMeets ISO 6876 requirements
    BiocompatibilityFree from toxicity per ISO/TR 7405 Agar diffusion test

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in a 510(k) summary for a material-based device. The "test set" would refer to the samples of the material tested in the laboratory. The document doesn't specify the number of samples or batches tested. The provenance is likely Korea, as the manufacturer is Meta Dental Co. 82-06 Grand Avenue Elmhurst, NY 11373 (US presence but likely manufacturing in Korea, common for medical device companies) and the ultimate letter is addressed to "Meta Biomed Company Limited." The testing would have been prospective to demonstrate compliance with the standard for the specific formulation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. For a root canal sealer, "ground truth" refers to the objective physical and chemical properties of the material, as measured and evaluated according to the specified international standard (ISO 6876). This does not involve expert readers or diagnostic accuracy assessments.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no adjudication method in the context of laboratory testing for material properties. The tests are designed to produce objective, measurable results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a material, not an AI-powered diagnostic tool. There are no "human readers" in this context, nor an AI component.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance claims of this device is the objective measurements and observations performed in a laboratory setting according to the methodologies prescribed by ISO 6876 Dental root canal sealing materials. For biocompatibility, it's defined by the outcome of an "Agar diffusion test" per ISO/TR 7405.

    8. The sample size for the training set

    Not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. There is no "training set" or associated ground truth establishment for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1