Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K071031

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-07-30

    (110 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K983622, K041592

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vu e.POD Vertebral Body Replacement is indicated for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or otherwise unstable vertebral body due to tumor or trauma (ie. fracture).

    The Vu e.POD Vertebral Body Replacement is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. Bone graft material is recommended for packing in the interior opening of the device prior to implantation.

    The Vu e.POD Vertebral Body Replacement is intended for use with supplemental internal spinal fixation systems, such as the Theken Surgical. LLC, BodyForm Thoracolumbar Fixation System (K983622, Approved 12/98) or the Theken Surgical, LLC, Coral Spinal System (K041592, Approved 9/04).

    Device Description

    The Vu e•POD Vertebral Body Replacement is comprised of PEEK concave cages with fenestrations and radii on all sides and toothed spikes used in combination with a titanium spacer component. The cages and spacer can be locked together into a variety of geometric configurations to fit each individual patient's pathology. The toothed spikes of the concave cages engage with the superior and inferior end plates of the neighboring vertebral bodies to resist rotation and migration. The Vu e•POD Vertebral Body Replacement may be used individually or in a pair depending on the surgical need, however, the device is always implanted with the construct oriented vertically.

    AI/ML Overview

    The provided documentation describes a 510(k) submission for the "Vu e•POD Vertebral Body Replacement" device. This type of submission relies on demonstrating substantial equivalence to a predicate device, rather than conducting a de novo clinical study with specific acceptance criteria and detailed reporting of reader performance. Therefore, the information requested in your prompt regarding acceptance criteria, study details, expert involvement, and reader performance (human or AI) is not typically part of a 510(k) summary for a spinal implant.

    Instead, the submission focuses on demonstrating that the new device has the same indications for use, basic design, operating principle, materials, surgical techniques, supplemental fixation requirements, manufacturing environment, sterilization process, and packaging configurations as the predicate devices. The "study" mentioned is primarily a series of mechanical tests and design comparisons rather than a clinical trial.

    Here's a breakdown based on the provided text, addressing your points where possible and noting where the information is not applicable to this type of regulatory submission:

    Acceptance Criteria and Device Performance

    Since this is a 510(k) submission, the "acceptance criteria" revolve around demonstrating substantial equivalence to the predicate device, not necessarily quantitative performance metrics against a predefined clinical or diagnostic standard in the way an AI/diagnostic device might have. The "device performance" is therefore implicitly that it functions equivalently to the predicate devices.

    Acceptance Criteria (Demonstrated)Reported Device Performance (Demonstrated)
    Same indications for useVerified against predicate device's IFU
    Same basic designVisual and compositional comparison
    Same operating principleFunctional comparison
    Same materialsPolyetheretherketone (PEEK-OPTIMA LT) per ASTM F-2026, Tantalum per ASTM F-560, Titanium 6Al-4V ELI per ASTM F-136, same as predicate
    Implanted using same surgical techniques and equipment typeConfirmed by manufacturer
    Used in conjunction with same supplemental fixationConfirmed with specified systems
    Same manufacturing environmentConfirmed by manufacturer
    Same sterilization processConfirmed by manufacturer
    Same packaging configurationsConfirmed by manufacturer
    Mechanical Testing ResultsNot explicitly detailed, but stated as performed for equivalence demonstration

    Note: The above table reflects the criteria for demonstrating substantial equivalence, which is the core "acceptance criterion" for a 510(k) device. Specific quantitative performance metrics like sensitivity, specificity, or reader improvement in an MRMC study are not applicable to this submission type.

    Study Details (as inferable from a 510(k) for a spinal implant):

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Test Set Sample Size: Not applicable in the context of a clinical trial. The "test set" here would refer to the physical devices subjected to mechanical testing. The number of devices tested is not specified but would typically follow ISO/ASTM standards for medical device testing.
      • Data Provenance: Not applicable. The "data" primarily consists of mechanical test results and material characterization, not patient data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. "Ground truth" in the clinical/diagnostic sense is not established for this type of device. The ground truth for mechanical testing would be established by validated test methods and engineering principles. Expertise would come from engineers/scientists performing the tests, not medical professionals interpreting diagnostic data.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This concept pertains to resolving discrepancies among expert readers in diagnostic studies. For mechanical testing, resolution of discrepancies would follow standard engineering practices or test method protocols.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC study was not done, and it is not relevant for this type of device (a vertebral body replacement implant). This is a surgical implant, not a diagnostic AI device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, a standalone algorithm performance study was not done. This device is not an algorithm or AI.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not applicable in the clinical/diagnostic sense. For the mechanical testing performed, the "ground truth" is based on established engineering principles, material specifications (ASTM standards for PEEK, Tantalum, Titanium), and validated test protocols designed to assess mechanical properties like strength, fatigue, and stability.
    7. The sample size for the training set:

      • Not applicable. There is no "training set" in the context of an algorithm or AI for this device.
    8. How the ground truth for the training set was established:

      • Not applicable. There is no "training set" for this device.

    In summary: The provided document is a 510(k) summary for a medical implant, the Vu e•POD Vertebral Body Replacement. Its regulatory pathway relies on demonstrating "substantial equivalence" to existing cleared predicate devices, primarily through comparison of design, materials, indications for use, and mechanical testing, rather than extensive clinical trials or performance assessments against specific diagnostic criteria. Therefore, most of the questions pertaining to AI/diagnostic study design (like sample size, number of experts, ground truth, MRMC studies) are not applicable to this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070786

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-06-28

    (98 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K983622, K041592

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The x•POD Adjustable Vertebral Body Replacement is indicated for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or otherwise unstable vertebral body due to tumor or trauma (e.g., fracture).

    The x•POD Adjustable Vertebral Body Replacement is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. It is recommended that bone graft material be packed in the interior openings of the device prior to implantation.

    The x•POD Adjustable Vertebral Body Replacement is intended for use with supplemental internal spinal fixation systems, such as the BodyForm Thoracolumbar Fixation System (K983622, SE 12/98) or Coral Pedicle Screw System (K041592, SE 9/04).

    Device Description

    The x•POD Adjustable Vertebral Body Replacement is comprised of a cylindrical titanium (Ti-6Al-4V) mechanism with rectangular endplates. The system has fenestrations axially and laterally to maximize bone graft placement. The implants are manufactured in a variety of geometric combinations to better fit each patient's pathology. The anatomically designed endplates are designed to minimize expulsion and rotation of the implant postoperatively. The x•POD Adjustable Vertebral Body Replacement is manufactured from titanium alloy (Ti-6Al-4V) per ASTM F136 and ISO 5832-3 standards. The x•POD Adjustable Vertebral Body Replacement construct is used individually and is always implanted with the construct oriented vertically.

    AI/ML Overview

    This is a 510(k) premarket notification for a spinal implant device (x•POD Adjustable Vertebral Body Replacement), not a diagnostic AI device. Therefore, the provided text does not contain information about acceptance criteria, study details, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details as these are not relevant to the approval process for this type of medical device.

    The approval is based on "substantial equivalence" to legally marketed predicate devices, meaning it has similar intended use, design, and physical characteristics, and performs comparably in mechanical testing, not through clinical performance metrics like sensitivity or specificity.

    Therefore, I cannot provide the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070381

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-04-25

    (76 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Why did this record match?
    Reference Devices :

    K983622, K041592

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Theken VU MESH VBR SYSTEM is indicated for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or otherwise unstable vertebral body due to tumor or trauma (i.e. fracture).

    The Theken VU MESH VBR SYSTEM is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. Bone graft material is recommended for packing in the interior opening of the device prior to implantation.

    The Theken VU MESH VBR SYSTEM is intended for use with supplemental internal spinal fixation systems, such as the Theken BodyForm Thoracolumbar Fixation System (K983622, Approved 12/98) or the Theken Coral Pedicle Screw System (K041592, Approved 9/04).

    Device Description

    The Theken Vu Mesh VBR system is comprised of square footprint polymer cages with fenestrations axially and laterally. It includes toothed endplates, which are used in combination with optional spacer components to extend the range of the cage in smaller increments. The cages, spacers and endplates can be assembled in a variety of combinations to fit each individual patient's pathology. The toothed endplates engage with the superior and inferior endplates of the neighboring vertebral bodies to resist rotation and migration. The Vu Mesh cage may be used individually or in a pair depending on the surgical need, however, the device is always implanted with the mesh cage oriented vertically.

    AI/ML Overview

    This is a 510(k) summary for a medical device called the "Theken Vu Mesh VBR System," which is a vertebral body replacement device. 510(k) submissions primarily focus on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting new clinical study data with acceptance criteria for device performance. Therefore, the requested information regarding acceptance criteria, study details, and human reader performance with AI assistance is not available in this document.

    This document primarily outlines:

    • Device Information: Trade name, common name, classification, materials, and description.
    • Intended Use: For thoracolumbar spine (T1-L5) to replace unstable vertebral bodies due to tumor or trauma, to restore biomechanical integrity, and for use with supplemental spinal fixation systems.
    • Predicate and Comparative Devices: A list of previously cleared devices to which the Vu Mesh VBR System claims substantial equivalence.
    • Substantial Equivalence Argument: The argument is based on similarities in intended use, design, physical characteristics, and material composition to existing devices. Mechanical testing and engineering analysis are mentioned as part of the evidence.

    Therefore, I cannot provide the specific details about acceptance criteria and study data as requested because this information is not typically included in a 510(k) summary focused on substantial equivalence.

    For completeness, if this were a document that did contain such a study, here's how I would apply your requested framework (displaying what would be absent in this specific document):


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Metric)Threshold/TargetReported Device PerformanceMeets Criteria?
    (Not available in this 510(k) summary)(Not available)(Not available)(Not available)

    2. Sample Size for Test Set and Data Provenance

    • Sample Size for Test Set: (Not available in this 510(k) summary)
    • Data Provenance: (Not available in this 510(k) summary. Given the nature of a 510(k) for an implant, any testing would likely be bench testing rather than clinical data from human subjects unless specifically required for novel aspects.)
      • If clinical data were involved, it would specify: e.g., "Retrospective, multi-center data from 3 US hospitals" or "Prospective, single-center study in Germany."

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: (Not available in this 510(k) summary)
    • Qualifications of Experts: (Not available in this 510(k) summary)
      • If applicable, this would state: e.g., "3 board-certified orthopedic surgeons with >15 years of experience in spinal surgery."

    4. Adjudication Method for Test Set

    • Adjudication Method: (Not available in this 510(k) summary)
      • If applicable, this would state: e.g., "2+1 consensus: two readers independently reviewed cases, and a third senior reader resolved disagreements."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, this document does not describe an MRMC study.
    • Effect size of human readers with AI vs. without AI assistance: Not applicable, as no MRMC study with AI assistance is described.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? No, this document describes a physical medical device (vertebral body replacement system), not an algorithm or AI. Performance is assessed through substantial equivalence to existing devices and mechanical testing.

    7. Type of Ground Truth Used

    • Type of Ground Truth: (Not applicable as this is a physical device and the 510(k) focuses on substantial equivalence based on design and materials, supported by mechanical testing, rather than an AI/diagnostic algorithm using clinical ground truth.)
      • If applicable (e.g., for an AI diagnostic device), this would specify: e.g., "Pathology reports (gold standard)" or "Clinical outcomes data (e.g., patient follow-up over 12 months)."

    8. Sample Size for Training Set

    • Sample Size for Training Set: Not applicable. This is not an AI/ML device that requires a training set.

    9. How Ground Truth for Training Set Was Established

    • Method for Establishing Ground Truth: Not applicable. This is not an AI/ML device that requires a training set with established ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K050058

    Validate with FDA (Live)

    Date Cleared
    2005-05-17

    (126 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Why did this record match?
    Reference Devices :

    K983622, K041592

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Theken VU VBR SYSTEM is indicated for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (e.g., fracture).

    The Theken VU VBR SYSTEM is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absences of fusion for a prolonged period. Bone graft material is recommended to be packed into the interior openings of the device prior to implantation.

    The Theken VU VBR SYSTEM is intended to be used with supplemental internal spinal fixation systems, such as the Theken BodyForm Thoracolumbar Fixation System (K983622), Approved 12/98) or the Theken Coral Pedicle Screw System (K041592, Approved 9/04).

    Device Description

    The Theken VU VBR SYSTEM is comprised of rounded rectangular and rectangular cages with fenestrations and radii on all sides and toothed spikes which are used in combination with spacer components. The cages and spacer can be locked together into a variety of geometric configurations to fit each individual patient's pathology. The toothed spikes of the rounded rectangular and rectangular cages engage with the superior and inferior end plates of the neighboring vertebral bodies to resist rotation and migration. A single construct is sufficient to be used at all spinal levels and pathologies.

    AI/ML Overview

    The provided document is a 510(k) summary for the Theken VU VBR SYSTEM, which is a vertebral body replacement device. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting new clinical trials to establish acceptance criteria for device performance in terms of accuracy, sensitivity, or specificity.

    Therefore, the document does NOT contain the information requested in your prompt regarding acceptance criteria, device performance studies, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details.

    The 510(k) summary explicitly states: "Establishment of equivalence is based on similarities of intended use, design, and physical characteristics. All implants are used to treat the same conditions, have essentially the same precautions and contradictions for use, and have equivalent potential for complications for the risk of use. In addition they all represent a basic design concept in terms of safety and effectiveness, and differ only in minor details. Based on the design concept, the use of established well known materials, feature comparisons, mechanical testing, indications for use, pre-production quality assurance planning and engineering analysis (adherence to GLP), Theken Surgical believes that sufficient evidence exists to reasonably conclude that the VU VBR SYSTEM is substantially equivalent to existing legally marketed devices."

    The study that "proves the device meets the acceptance criteria" in this context refers to the comparison against predicate devices and mechanical testing, not a clinical study to establish diagnostic performance. The "acceptance criteria" are not reported as specific performance metrics like sensitivity/specificity but rather as conformity to the design, materials, and intended use of equivalent devices already on the market and adherence to good laboratory practices (GLP) for mechanical testing.

    Since this is a medical device and not an AI/algorithm-based diagnostic tool, many of the requested fields are not applicable to this type of regulatory submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032064

    Validate with FDA (Live)

    Date Cleared
    2004-02-20

    (232 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K003155, K983622, K980485

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CPOD/LPOD is indicated for use in the thoracolumbar spine (i.c. T1 to L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (e.g. fracture).

    The CPOD/LPOD is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. middle alle posterior upmar over and to be packed into the interior openings of the device prior to implantation.

    The CPOD/LPOD is intended to be used with supplemental, internal spinal fixation The Cr OD/El OD is intended to or arm Thoracolumbar Fixation System (K983622) or the DePuy Acromed ISOLA System (K980485).

    Device Description

    The Theken CPOD / LPOD VBR is comprised of Titanium rounded rectangular and rectangular frames with fenestrations and radii on all sides and toothed spikes which are used in combination with Titanium spacer components. The frames and spacer can be locked together into a variety of geometric configurations to fit each individual patient's pathology. The toothed spikes of the rounded rectangular and rectangular frames engage with the superior and inferior end plates of the neighboring vertebral bodies to resist rotation and migration. A single construct is sufficient to be used at all spinal levels and pathologies.

    The Theken CPOD / LPOD VBR System is designed to be constructed using two cage components in conjunction with an appropriate single spacer component. This combination of components creates an implantable construct. The use of a single cage as an implant has not been tested or approved.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "Theken Surgical CPOD / LPOD Vertebral Body Replacement System (CPOD / LPOD VBR)". This document primarily focuses on establishing substantial equivalence to previously cleared predicate devices rather than presenting a standalone study with acceptance criteria and performance data in the typical sense of algorithm-based device evaluations.

    Therefore, many of the requested categories for a study proving acceptance criteria (like sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable in this context, as this is a regulatory submission for a physical medical implant, not an AI/algorithm-based diagnostic or screening tool.

    However, I can extract information related to the device's characteristics and the basis for its approval.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not present a table of specific acceptance criteria (e.g., sensitivity, specificity, accuracy targets) and corresponding reported performance metrics like one would find for a diagnostic algorithm. Instead, the "acceptance criteria" are implied by the demonstration of substantial equivalence to predicate devices. The performance is therefore implicitly "equivalent" to that of the predicate devices.

    Acceptance Criterion (Implicit)Reported Device Performance (Implied by Substantial Equivalence)
    Intended Use EquivalenceThe Theken CPOD / LPOD VBR has essentially the same indications for use as the predicate devices (Tetris™ Spinal Implant, Stackable Cage™ System, Synthes Synex™ System).
    Design and Physical Characteristics EquivalenceThe device is similar in design, material composition (Titanium Ti-6Al-4V), and construction to the predicate devices. It uses frames with fenestrations and toothed spikes, combined with spacer components, to create an implantable construct.
    Material Composition EquivalenceMade of Titanium Ti-6Al-4V (ELI) per ASTM F-136, which is a standard biomaterial also used in similar predicate devices.
    Biomechanical Integrity RestorationDesigned to restore biomechanical integrity of the anterior, middle, and posterior spinal column, similar to the function of predicate VBR systems.
    Compatibility with Supplemental FixationIntended to be used with supplemental internal spinal fixation systems as are comparable predicate devices.
    Safety and EffectivenessBelieved to have equivalent potential for complications and to be safe and effective based on design, materials, feature comparisons, and mechanical testing, similar to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable. This is a 510(k) for a physical implant, not a study involving a "test set" of patient data for performance evaluation in the typical sense. Equivalence is established through comparison of design, materials, intended use, and mechanical testing on the device itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Not Applicable. See point 2. The "ground truth" here relates to engineering principles, material science, and clinical practice for spinal implants. Regulatory bodies (FDA) and potentially design engineers at Theken Surgical/consultants would be the "experts" in this context, evaluating compliance with standards and equivalence.

    4. Adjudication Method:

    • Not Applicable. See point 2.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. This is not a diagnostic device or an AI assistant for human readers. Therefore, an MRMC study is not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used:

    • The "ground truth" for the device's acceptability is based on:
      • Regulatory Standards: Compliance with FDA regulations (e.g., 21 CFR 878.3060 for Vertebral Body Replacement Devices).
      • Material Standards: Use of ASTM F-136 compliant Titanium Ti-6Al-4V (ELI).
      • Biomechanical Engineering Principles: The design's ability to restore biomechanical integrity.
      • Clinical Precedent: The established safety and effectiveness of the predicate devices.
      • Mechanical Testing: Though not detailed in the provided text, the document mentions "mechanical testing" as a basis for equivalence. This typically involves in-vitro biomechanical tests simulating physiological loads on the implant.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is not an AI/machine learning model.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. This is not an AI/machine learning model.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1