510(k) Data Aggregation

    K Number
    K112314

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-04-27

    (260 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OLPur H2H Hemodiafiltration (HDF) Module is indicated for use, with a UF controlled hemodialysis machine that provides ultrapure dialysate in accordance with current ANSI/AAMI/ISO standards, for treatment of patients with chronic renal failure as prescribed by a physician. The OLPur MD 220 Hemodiafilter is indicated for use for hemodiafiltration of patients with chronic renal failure. The OLPur MD 220 Hemodiafilter is only to be used with the OLpur H2H Hemodiafiltration Module. Selection of patients, as well as treatment operating parameters, are a medical decision and the responsibility of the prescribing physician.

    Device Description

    The H₂H™ Hemodiafiltration Module is a programmable electro-mechanical medical device designed to work in conjunction with and in close proximity to a hemodialysis machine during patient treatments. Upon installation, the H₂H Module is placed next to the hemodialysis machine (which can be on either side depending upon type and model of the hemodialysis machine) and is connected to the clinic's water supply, drain, and electricity.

    The H₂H Hemodiafiltration Module serves two basic functions. The Module's primary function is to generate on-line substitution fluid for hemodiafiltration. The H₂H Module does this by ultrafiltering dialysate, provided by the hemodialysis machine, through a reusable substitution filter. The secondary function of the H₂H Hemodiafiltration Module is to serve as a substitution filter reprocessing machine. These steps are performed when the H₂H Hemodiafiltration Module is not connected to the hemodialysis machine. More detailed descriptions of the device, its accessories and how the system functions can be found in Section 11 (Device Description) of this submission.

    AI/ML Overview

    The provided text is a 510(k) summary for the Nephros OLpūr® H₂H H₂H Hemodiafiltration (HDF) Module and OLpūr MD 220 Hemodiafilter. It focuses on device description, intended use, and technological characteristics compared to predicate devices for substantial equivalence.

    However, this document does not contain explicit acceptance criteria and a study proving the device meets those criteria in the typical format of a clinical trial report or performance study with detailed results against defined benchmarks. Instead, it refers to "Bench Analyses performed on the subject devices" in Section 18 and "clinical data supporting substantial equivalence" in Section 20 of the full application, which are not provided.

    Therefore, many of the requested details about acceptance criteria, study design, sample sizes, expert involvement, and ground truth cannot be extracted directly from the given 510(k) summary.

    Here's what can be gathered and what is missing based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated in terms of specific performance metrics with numerical targets within this summary. Substantial equivalence is the overall goal, meaning the device performs as safely and effectively as predicate devices.
    • Reported Device Performance: The summary states, "The proposed device has the same technological characteristics and is similar in design as compared to the predicate devices and already approved kidney dialysis therapy." This is a general statement of equivalency rather than specific performance numbers against acceptance criteria. The full submission's Section 18 and 20 would contain these details.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in this 510(k) summary.
    • Data Provenance: Not specified in this 510(k) summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not specified in this 510(k) summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not specified in this 510(k) summary.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This device is a medical module and hemodiafilter, not an AI-assisted diagnostic tool. Therefore, an MRMC study related to human readers improving with AI is irrelevant and not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This device is a physical medical device (hemodiafiltration module and filter), not a software algorithm. Therefore, "standalone algorithm" performance is not applicable in this context. Its performance would be evaluated during its operation in conjunction with a hemodialysis machine.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not specified in this 510(k) summary. Given the nature of the device, ground truth would likely refer to performance metrics like filtration efficiency, ultrafiltration control accuracy, biocompatibility, and microbiological purity, established through bench testing and potentially in vivo studies (though clinical data reference focuses on substantial equivalence).

    8. The sample size for the training set

    • Not applicable/specified. This is a physical medical device, not an AI/machine learning model that typically has a "training set."

    9. How the ground truth for the training set was established

    • Not applicable/specified for the same reasons as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K090885

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-06-29

    (90 days)

    Product Code
    Regulation Number
    876.5665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K983126

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DSU Dual Stage Ultrafilter is intended to be used to filter water or bicarbonate concentrate used in hemodialysis devices. The DSU Ultrafilter assists in providing hemodialysis quality water or bicarbonate concentrate. The device is not a complete water treatment system, but serves to remove biological contaminants. Therefore, it must be used in conjunction with other water treatment equipment (i.e., RO, DI, etc.).

    Device Description

    The DSU Dual Stage Ultrafilter is a hollow fiber filter that removes bacteria, viruses, endotoxin and particulate from water and bicarbonate concentrate used in hemodialysis.

    AI/ML Overview

    The provided text describes the DSU Dual Stage Ultrafilter, a device intended to filter water or bicarbonate concentrate used in hemodialysis. The information primarily focuses on its regulatory submission and equivalence to a predicate device. While it mentions various tests conducted, it does not provide detailed acceptance criteria with numerical thresholds or specific performance results in a comparative table format. Therefore, some information requested cannot be fully extracted from the provided text.

    Here's the information that can be extracted, and an indication of what is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Flow Rate versus Pressure DropNot specifiedTested
    Bicarbonate Composition EffectNot specifiedTested
    Pyrogen RemovalNot specifiedTested
    Virus Challenge TestNot specified (implied to meet requirements for substantial equivalence)Tested
    Bacteria Challenge TestNot specified (implied to meet requirements for substantial equivalence)Tested
    Chemical CompatibilityNot specifiedTested

    2. Sample Size for the Test Set and Data Provenance

    The document mentions that the DSU Dual Stage Ultrafilter "has been tested for performance." However, it does not specify the sample size used for the test set for any of the mentioned tests (Flow Rate, Bicarbonate Composition, Pyrogen Removal, Virus Challenge, Bacteria Challenge, Chemical Compatibility).

    The data provenance is also not explicitly stated. It's a regulatory submission to the FDA (USA), implying the testing was conducted by the manufacturer (Nephros Inc. in River Edge, NJ, USA). It does not indicate if the data is retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This device is a physical filter, not an AI/software device that relies on expert interpretation for ground truth. Therefore, this information is not applicable to this type of medical device submission. The performance is assessed through laboratory testing against physical and biological contaminants.

    4. Adjudication Method for the Test Set

    As this is a physical device with laboratory performance testing, adjudication methods typically used for human interpretation of medical images or data are not applicable. Test results would be objectively measured.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    This is not applicable as the device is a physical filter, not an AI-assisted diagnostic tool involving human readers.

    6. Standalone Performance Study

    Yes, the studies mentioned (Flow Rate, Bicarbonate Composition Effect, Pyrogen Removal, Virus Challenge Test, Bacteria Challenge Test, and Chemical Compatibility) represent standalone performance testing of the DSU Dual Stage Ultrafilter. The text states: "The DSU Dual Stage Ultra Filter has been tested for performance." This indicates that the algorithm (in this case, the device's physical mechanism) was evaluated directly without human intervention to complete its primary function.

    7. Type of Ground Truth Used

    The ground truth for the stated tests would be based on objective laboratory measurements and established scientific standards for water purification and filtration efficacy (e.g., bacterial count reduction, virus removal, pyrogen levels, flow rate consistency). For example, for the "Virus Challenge Test" and "Bacteria Challenge Test," the ground truth would be the known concentration of viruses or bacteria in the input water and the measured concentration in the output water after filtration, adhering to specific test protocols.

    8. Sample Size for the Training Set

    This information is not applicable as this is a physical medical device, not an AI/machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as this is a physical medical device, not an AI/machine learning algorithm.