510(k) Data Aggregation

    K Number
    K060448

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-07-06

    (135 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lumenis One Family of Systems (and the delivery accessories that are used with them to deliver light and/or laser and/or RF energy) is indicated for use in surgical, aesthetic and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology as follows:

    Intense Pulsed Light (IPL) Wavelengths (515 - 1200 nm):
    The 515-1200 nm intense pulsed light wavelengths are indicated for:

    • The treatment of benign pigmented epidermal lesions, including dyschromia, hyperpigmentation, melasma, ephelides (freckles);
    • The treatment of tattoos;
    • The treatment of cutaneous lesions, including warts, scars and striae;
    • The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations;
    • The removal of unwanted hair from all skin types, and to effect stable long-term, or permanent, hair reduction in skin types I-V through selective targeting of melanin in hair follicles.

    Nd:YAG Laser Wavelength (1064 nm):
    The 1064 nm wavelength produced by the Nd: YAG laser is indicated for:

    • The coagulation and hemostasis of vascular lesions and soft tissue, including:
    • Treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1 - 4.0 mm diameter) of the leg:
    • The removal of unwanted hair from all skin types, and to effect stable long-term, or permanent, hair reduction in skin types I-V through selective targeting of melanin in hair follicies;
    • The non-ablative treatment of facial wrinkles.

    LightSheer Diode Laser Wavelength (800 mm):
    The 800 nm wavelength produced by the LightSheer diode laser is indicated for:

    • The removal of unwanted hair, and to effect stable long-term, or permanent', hair reduction through selective targeting of melani, in hair follicles;
    • The treatment of vascular lesions, including angiomas, hemangiomas telangiectasia and other benign vascular lesions;
    • The treatment of leg veins;
    • The treatment of benign pigmented lesion;
    • The treatment of pseudofolliculitis barbae;

    The LightSheer diode laser is intended for use on all skin types (Fitzpatrick skin types 1 - VI), including tanned skin.

    Aluma RF Energy (468 kHz):
    The Aluma 468 kHz energy is indicated for:

    • The Aluma Skin Renewal System is intended for use in Dermatologic and General Surgical procedures for the non-invasive treatment of wrinkles and thytids.
    Device Description

    Lumenis One systems are comprised of the following main components:

    • System console;
    • Control and Display;
    • One or more delivery handpieces, the Treatment Heads (up to three attached out of four handpieces available for each system);
    • Skin cooling technology integrated into the handpiece (as applicable);
    • Trigger button integrated into the handpiece; an additional safety button integrated into the laser handpieces;
    • Remote interlock connector (disables pulse emission when treatment room door is opened).
    AI/ML Overview

    This 510(k) summary explicitly states that no clinical data was provided for the Lumenis One Family of Systems. The device's substantial equivalence to predicate devices was based on shared indications for use and similar or identical technical specifications, not on new performance studies.

    Therefore, the following information cannot be found or accurately described from the provided text:

    • A table of acceptance criteria and reported device performance
    • Sample sizes used for test sets
    • Data provenance for test sets
    • Number of experts used to establish ground truth
    • Qualifications of experts
    • Adjudication methods
    • MRMC comparative effectiveness study results
    • Standalone performance results
    • Type of ground truth used
    • Sample size for the training set
    • How ground truth for the training set was established

    The document makes a direct statement that is crucial for understanding why these details are absent:

    "Clinical data therefore was not provided." (Section VII, page 2)

    The rationale for not providing clinical data is that:
    "The Lumenis One Family of Systems share the same indications for use and the same or similar technical specifications as the currently marketed predicate Lumenis Family of Intense Pulsed-Light (IPL) and IPL Nd: Y AG Laser Systems (K020839, K024093, K030342, K030527), the LightSheer Pulsed Diode Array Laser System (K973324, K974346, K982940, K001746, K003614), and the Aluma Skin Renewal System (K051214). The small differences that exist in certain technical specifications are not significant and they do not alter the safety or effectiveness of the Lumenis One Family of Systems." (Section VII, page 2)

    In summary, the device's acceptance criteria were implicitly met by demonstrating substantial equivalence to already cleared predicate devices, without the need for new clinical performance studies or establishing new acceptance criteria and demonstrating compliance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032846

    Validate with FDA (Live)

    Date Cleared
    2003-12-11

    (91 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spectra Hair Removal Laser is intended to effect temporary hair removal.

    Device Description

    The Spectra Hair Removal Laser is a semiconductor diode laser system that delivers infrared light at a wavelength of nominally 800 nm.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Spectra Hair Removal Laser, focusing on the acceptance criteria and study aspects you requested:

    Acceptance Criteria and Study Analysis for Spectra Hair Removal Laser (K032846)

    This 510(k) submission explicitly states that no performance data was required or submitted for the Spectra Hair Removal Laser. The basis for clearance was substantial equivalence to legally marketed predicate devices, and the indications for use were a subset of those claimed by the predicates. Therefore, the device did not undergo a specific study to prove it meets acceptance criteria, nor were explicit acceptance criteria defined within this document.

    Given this, many of the specific questions about the study design cannot be answered from the provided text.


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specifiedNo performance data provided

    Explanation: The 510(k) states: "None. The specifications and indications for use of the Spectra Hair Removal Laser are a subset of those claimed in one or more of the clearances for the above-listed predicate devices. Thus performance data were not required." This means the device's performance was considered acceptable based on its similarity to already cleared devices, not on specific, pre-defined quantitative criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size (Test Set): Not applicable, no performance study was conducted.
    • Data Provenance: Not applicable, no performance study was conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: Not applicable, no performance study was conducted.
    • Qualifications of Experts: Not applicable, no performance study was conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable, no performance study was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This device is a hair removal laser, not a diagnostic AI system with human-in-the-loop interaction for interpretation tasks.
    • Effect Size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Study: No. This device is a laser for hair removal, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: Not applicable, no performance study was conducted. The "ground truth" for this submission was the established safety and effectiveness of its predicate devices for temporary hair removal.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable, no AI/ML algorithm was involved, and no performance study was conducted for this specific device.

    9. How the ground truth for the training set was established

    • Ground Truth Establishment for Training Set: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K011747

    Validate with FDA (Live)

    Date Cleared
    2001-08-30

    (86 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K973324, K974346, K982940, K001746, K003614

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LC100 diode laser system is indicated for hair removal and permanent hair reduction. Permanent hair reduction is defined as a long-term, stable reduction in the number of re-growing hairs after a treatment regime.

    The LC100 is also indicated for photocoagulation of dermatological vascular lesions, including port wine stains, telangiectasia, angioma, hemangioma, spider nevi, and other benign vascular lesions; photothermolysis of blood vessels (treatment of facial and leg veins); removal of benign pigmented lesions, including lentigines, nevi, chloasma, and café-au-lait; incision/excision, ablation and coagulation (homeostasis) of soft tissue.

    The LC100 is indicated for use in patients with skin types I-VI, including tanned patients.

    Device Description

    The LC100 delivers infrared laser light with a wavelength of 810 nm, a selectable pulse duration of 50 - 1000 ms. The corresponding fluence is delivered through a 12x12 or 12x6 mm handpiece tip. The laser pulses are generated at a maximum pulse repetition frequency of 3.3 Hz.

    The complete system consists of a, chiller, a footswitch, and a handpiece containing diode bars connected to the power unit with an umbilical. In standard use, the handpiece is pressed against the patient's skin and a light pulse is delivered when the footswitch is depressed. The handpiece tip is water-Laser cooled to provide active skin cooling. parameters and other system features are controlled from the user interface panel on top of the laser unit, which provides an interface to the system computer.

    AI/ML Overview

    The provided document is a 510(k) summary for the Palomar LC100 Diode Array Laser System, which is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. This type of submission does not typically include the kind of detailed study data (like acceptance criteria tables, sample sizes for test/training sets, ground truth establishment, MRMC studies, or standalone performance) that would be present in a Clinical Study Report or a more extensive Premarket Approval (PMA) application.

    The document primarily focuses on establishing substantial equivalence by comparing the LC100 to predicate devices (Palomar SLP 1000 Diode Laser and LightSheer Diode Array Laser) through a description of its system and intended use. It explicitly states: "The differences in the specifications of the LC100 laser and the predicate device do not result in different performance or raise new questions of safety or efficacy. Based on the foregoing, the LC100 Diode Conclusion: Array Laser System is substantially equivalent to the legally-marketed claimed predicate device."

    Therefore, based solely on the provided text, I cannot provide the specific details requested in your prompt because they are not present in this 510(k) summary.

    Here's what I can extract or infer based on the nature of a 510(k) submission:

    • Acceptance Criteria and Reported Device Performance: Not explicitly stated in a quantitative table. The "acceptance criterion" for a 510(k) is typically substantial equivalence to a predicate device, meaning it performs as safely and effectively as the predicate without raising new questions of safety or efficacy. The "reported device performance" is essentially that it is considered substantially equivalent to the predicate devices for its stated indications for use.
    • Study Proving Acceptance Criteria: The "study" is the submission itself, which argues for substantial equivalence by comparing the LC100's specifications and intended use to those of the predicate devices. This is generally a comparative analysis rather than a traditional clinical study with detailed performance metrics.
    • Sample Size for Test Set and Data Provenance: Not applicable/not provided. This is not a study that involved a test set in the sense of a clinical trial.
    • Number of Experts and Qualifications: Not applicable/not provided. Ground truth establishment with experts is not detailed for this type of submission.
    • Adjudication Method: Not applicable/not provided.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable/not provided. These are complex studies not typically required or included in a 510(k) for a laser system.
    • Standalone Performance: Not applicable/not provided as a distinct study. The comparison is against predicate devices.
    • Type of Ground Truth Used: Not applicable/not provided as this is not a diagnostic device requiring external validation of its output against a 'ground truth.'
    • Sample Size for Training Set: Not applicable/not provided. This device is not an AI/ML algorithm that would have a training set.
    • How Ground Truth for Training Set was Established: Not applicable/not provided.

    In summary, the provided document is a regulatory submission for substantial equivalence, not a detailed clinical performance study. The information requested regarding explicit acceptance criteria, detailed study design, sample sizes, and ground truth establishment is not typically found in this type of 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991935

    Validate with FDA (Live)

    Date Cleared
    2000-01-27

    (232 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K963249, K974536, 980517, 982980, K982940

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EpiLight® and PhotoDerm® HR are used for the removal of unwanted hair. EpiLight® and PhotoDerm® HR are also intended to effect stable long-term, or permanent hair reduction in skin types I-V through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as a long-term stable reduction in the number of hairs regrowing after a treatment regime.

    Device Description

    EpiLight® and PhotoDerm® HR are electro-optical medical devices designed for effective photothermal treatment of unwanted hair and its removal.

    AI/ML Overview

    The provided text is a 510(k) summary for the EpiLight® and PhotoDerm® HR devices, which are electro-optical medical devices for hair removal. The document focuses on establishing substantial equivalence to predicate devices rather than providing detailed acceptance criteria and study results for a new AI-powered device. Therefore, much of the requested information regarding AI device performance, sample sizes, ground truth establishment, and MRMC studies is not available in the provided text.

    Based on the information given, here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly define acceptance criteria in terms of specific quantitative metrics (e.g., sensitivity, specificity, accuracy) for a new device's performance. Instead, it relies on demonstrating "substantial equivalence" to predicate devices through technical comparisons and clinical trials.

    The primary performance claim for the EpiLight® and PhotoDerm® HR is:

    • Permanent hair reduction: Defined as a long-term stable reduction in the number of hairs regrowing after a treatment regime.

    The document states that a technical comparison and clinical trials were performed to establish substantial equivalence. However, no specific performance metrics resulting from these studies are provided in this summary.

    Acceptance Criteria (Implicit from Intended Use)Reported Device Performance
    Removal of unwanted hairAchieved (by substantial equivalence to predicates)
    Stable long-term, or permanent hair reduction in skin types I-V through selective targeting of melanin in hair folliclesAchieved (by substantial equivalence to predicates)
    Permanent hair reduction: a long-term stable reduction in the number of hairs regrowing after a treatment regime.Achieved (by substantial equivalence to predicates)

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "clinical trials were performed by ESC Medical Systems," but it does not provide any details about:

    • The sample size of the test set (number of participants or observations).
    • The data provenance (e.g., country of origin, retrospective or prospective nature of the data).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide any information regarding experts used to establish ground truth.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The devices described are medical devices for hair removal, not AI-assisted diagnostic tools requiring reader interpretation. Therefore, there is no discussion of human reader improvement with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The devices are physical medical devices for hair removal, not AI algorithms.

    7. The Type of Ground Truth Used

    Given the nature of the device (hair removal), the "ground truth" for "permanent hair reduction" would most likely be based on clinical observation and measurement of hair regrowth over time following a treatment regime. However, the document does not explicitly state the method for establishing this ground truth (e.g., photographic assessment, hair count, patient self-reporting, etc.).

    8. The Sample Size for the Training Set

    This question is not applicable as the document describes a physical medical device, not an AI algorithm that undergoes training.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable.

    Ask a Question

    Ask a specific question about this device