510(k) Data Aggregation

    K Number
    K192517

    Validate with FDA (Live)

    Date Cleared
    2020-04-22

    (222 days)

    Product Code
    Regulation Number
    862.3220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K974234

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Psychemedics Microplate EIA For Cotinine in Hair is an in vitro diagnostic device for the qualitative detection of cotinine in human head and body hair as an aid in the detection of cotinine after use or exposure to tobacco products. The assay is intended for a single site and uses a cutoff calibrator of 200 pg cotinine/mg hair. This device is intended exclusively for Psychemedics use only and is not intended for sale to anyone.

    The Psychemedics Microplate EIA For Cotinine in Hair provides only a preliminary analytical test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Liquid Chromatography/Mass spectrometry/Mass spectrometry (LC/MS/MS) is the confirmatory method used by Psychemedics Corporation. The LC/ MS/MS analysis uses a cutoff, after extensive washing of 100 pg cotinine/mg hair with presence of hydroxycotinine at or above 10 pg/mg hair.

    Device Description

    The Psychemedics Microplate EIA For Cotinine in Hair consists of two parts; a pre-analytical hair treatment procedure (to convert the solid matrix of hair to a measurable liquid matrix) and the screening assay, the Psychemedics Microplate EIA for Cotinine. The screening portion of the test system consists of (1) microplate wells coated with cotinine conjugated to bovine serum albumin (BSA), polyclonal rabbit anti-cotinine, goat antirabbit secondary antibody conjugated to HRP (horseradish peroxidase), substrate [3, 3', 5, 5' tetramethylbenzidine (TMB)], HCl to acidify (and stop the reaction), and wash buffer for washing the plates. Absorbance in the wells is read with a microplate reader.

    Liquid Chromatography/Mass spectrometry/Mass spectrometry (LC/MS/MS) is the confirmatory method used by Psychemedics Corporation. The LC/MS/MS analysis uses a cutoff, after extensive washing of the hair, of 100 pg cotinine/mg hair with presence of hydroxycotinine at or above 10 pg/mg hair.

    AI/ML Overview

    Here's a structured summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance for Psychemedics Microplate EIA for Cotinine in Hair (K192517)

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device and provides performance data rather than explicitly stating pre-defined "acceptance criteria" as pass/fail thresholds for specific metrics. However, based on the performance testing results, we can infer the implied acceptance criteria.

    Performance MetricImplied Acceptance Criteria (Inferred)Reported Device Performance
    Sensitivity (EIA)No false negatives above the LC/MS/MS confirmation cutoff relative to the 200 pg/mg hair EIA cutoff.No samples that screened negative were false negatives (i.e., no negative EIA result confirmed above 200 pg cotinine/mg hair by LC/MS/MS).
    Specificity (EIA)Minimal cross-reactivity with related compounds; no interference from common substances.Specific cross-reactivity percentages documented (e.g., 50% for trans-3-hydroxycotinine, 0.4% for nicotine); 86 compounds showed no cross-reactivity/interference.
    Precision (EIA) - QualitativeAble to consistently classify samples at and around the cutoff (200 pg/mg hair) as negative or positive.Intra-Assay: 10/10 samples correctly classified for all levels (-100% to +100% of cutoff, except -25% level where 10/10 classified as negative).Inter-Assay: 50/50 samples correctly classified for all levels (-100% to +100% of cutoff, except -25% level where 50/50 classified as negative).
    Precision (LC/MS/MS) - Intra-Assay (around cutoff)%CV < 10% for concentrations around cutoff (50, 75, 100, 125, 150 pg/mg hair). Results averaged 98-100.1% of targets.%CV < 10% for all concentrations. Results averaged 98-100.1% of targets.
    Precision (LC/MS/MS) - Intra-Assay (over range)%CV < 10% for concentrations across assay range (10-1000 pg/mg hair). Results averaged 90.1-111.4% of targets.%CV < 10% for all concentrations. Results averaged 90.1-111.4% of targets.
    Precision (LC/MS/MS) - Inter-Assay%CV < 10% for concentrations for 5 days. Averages within 15% of targets.All sets of replicates had < 10% CV. Averages of all replicates were within 15% of target concentrations.
    Linearity (LC/MS/MS)Regression coefficient > 0.995; actual data points deviate < 10% from regression line.Regression coefficient > 0.995. Average value of 5 determinations within 15% of predicted Y value. Actual data points deviated < 10% from regression line, demonstrating linearity from 10 to 1000 pg/mg hair.
    Recovery (LC/MS/MS)Recovery > 90% from hair; Recovery > 65% for cotinine and > 55% for hydroxycotinine with SPE.Recovery from hair was > 91.6%; Recoveries of analytes and I.S. with SPE were > 65% for cotinine and > 55% for hydroxycotinine.
    Carryover (LC/MS/MS)No signal detected in negative sample following a high positive sample (up to 1000 pg/mg hair).No signal detected in negative sample after 1000 pg/mg sample.
    LLOQ/LOD Precision (LC/MS/MS)S:N ratio > 19:1 to 20:1. Results within 15% of target for LLOQ/LOD spiked samples.S:N ratios 161-302 for cotinine, 137-217 for hydroxycotinine (at 10 pg/mg hair). Results showed 109.5% and 110.9% of target, respectively.
    Cosmetic TreatmentsNegative samples remain negative; positive samples remain positive.All negative samples remained negative; all positive samples remained positive.
    Environmental ContaminationConfirmed positive samples (from contamination) should drop below LC/MS/MS cutoff after washing procedure.Heavily smoke-exposed hair, after extensive washing, resulted in no samples above the LC/MS/MS cutoff of 100 pg cotinine/mg hair.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Comparison Study (EIA vs. LC/MS/MS): A total of 82 samples were utilized.
      • Data Provenance: The document states "Samples were identified for inclusion in the study by screening with a screening assay for cotinine in serum (K974234)." It doesn't explicitly state the country of origin or if they were retrospectively or prospectively collected human hair samples, but implies they are human hair samples.
    • Precision Studies (EIA): Not explicitly stated as "test set," but 10 samples were used for intra-assay and 50 samples for inter-assay precision at various levels around the cutoff. These were "negative hair spiked with previously LC/MS/MS-validated calibrator and control spiking solutions."
    • Cosmetic Treatments: Sets of 5 cotinine-negative and 6 cotinine-positive samples (total of 11) were used.
    • Environmental Contamination: 7 hair samples were exposed to smoke for this specific test.
    • LC/MS/MS Recovery from Hair: 5 authentic samples.
    • LC/MS/MS Precision, Linearity, LLOQ/LOD: These studies involved spiked samples rather than a separate "test set" of authentic patient samples, with varying numbers of replicates (e.g., 5 analyses, 5 samples for 5 days).
    • Sample Stability during Shipping and Storage: 6 samples for shipping stability and another 6 samples for storage stability.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Experts:

    The primary "ground truth" for the comparison study and for confirming the EIA results is the Liquid Chromatography/Mass Spectrometry/Mass Spectrometry (LC/MS/MS) method. This is a laboratory-based analytical method, not directly dependent on expert human interpretation of images or clinical signs. The document does not mention human experts establishing ground truth for the test set.

    4. Adjudication Method for the Test Set:

    Not applicable, as the ground truth is established by the LC/MS/MS analytical method, which is a quantitative measurement, not subject to human adjudication in the traditional sense of multiple readers interpreting a case. The LC/MS/MS confirms positive cotinine levels above a certain cutoff (100 pg cotinine/mg hair with presence of hydroxycotinine at or above 10 pg/mg hair).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. The device is an in vitro diagnostic assay, not an imaging device or AI algorithm requiring human interpretation. The comparison is between the new EIA screening method and the confirmatory LC/MS/MS method.

    6. If a Standalone Study (algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone study of the device (Psychemedics Microplate EIA for Cotinine in Hair) was performed in comparison to the LC/MS/MS confirmatory method, which serves as the "gold standard" ground truth. The EIA itself operates independently without direct human intervention in the interpretation of its results (beyond reading the microplate reader output). The final confirmed analytical result is obtained by LC/MS/MS.

    7. The Type of Ground Truth Used:

    The ground truth used is an alternative chemical method, specifically Liquid Chromatography/Mass Spectrometry/Mass Spectrometry (LC/MS/MS), which is considered a highly specific and sensitive analytical technique for cotinine detection. This is a form of laboratory/analytical ground truth. The LC/MS/MS analysis uses a cutoff of 100 pg cotinine/mg hair (after extensive washing) with the presence of hydroxycotinine at or above 10 pg/mg hair.

    8. The Sample Size for the Training Set:

    The document does not explicitly mention a "training set" in the context of machine learning. The device is a traditional enzyme immunoassay (EIA). The performance studies involve validation of the assay's characteristics (precision, linearity, cross-reactivity, etc.) using spiked samples, negative controls, and a limited number of authentic samples for comparison studies. There isn't a "training set" in the sense of data used to train an AI model.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no specific "training set" for an AI model. The calibrators and control materials for the EIA and LC/MS/MS are prepared using drug stocks from commercial vendors with certificates of analysis, and their concentrations are confirmed by LC/MS/MS.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033601

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-02-10

    (88 days)

    Product Code
    Regulation Number
    862.3200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K974234

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LabOne Micro-Plate Cotinine EIA is a competitive micro-plate immunoassay for the qualitative and semi-quantitative determination of cotinine in oral fluid collected with the OraSure® Oral Fluid Specimen Collection Device. LabOne Micro-Plate Cotinine EIA is used as an aid in the detection of cotinine after use of tobacco products or other products containing nicotine. The test is for in Vitro use only.

    The LabOne Microplate Cotinine Calibrators are a device intended for medical purposes for use with the LabOne Microplate Cotinine assay to establish points of reference that are used in determination of values in the measurement of cotinine in saliva.

    The LabOne Microplate Cotinine Controls are intended for use as an assayed quality control matrix to monitor the precision and accuracy of the laboratory testing procedures for cotinine.

    Device Description

    LabOne Micro-Plate Cotinine EIA is a solid phase competitive enzyme immunoassay for the qualitative and semi-quantitative analysis of cotinine in oral fluid specimens collected with the OraSure® Oral Specimen Collection Device.

    AI/ML Overview

    The provided text describes a 510(k) summary for the LabOne Micro-Plate Cotinine EIA, which is a device for detecting cotinine in oral fluid. However, the document does NOT contain detailed information about specific acceptance criteria and the comprehensive study results to "prove" the device meets them in the way described by the requested table and points.

    Specifically, the document states:
    "Performance characteristic studies on precision, analytical sensitivity, interference and antibody cross-reactivity showed that the LabOne Micro-Plate Cotinine EIA is substantially equivalent to the OraSure Micro-Plate Cotinine EIA."
    And:
    "Results tested from self-claimed patient specimens and diluted samples with both the LabOne Micro-Plate Cotinine EIA and the OraSure Micro-Plate Cotinine EIA also showed that the sample results from this two test systems are substantially equivalent when using self-claimed specimen results and urine results as references."

    This indicates that studies were performed, but the actual acceptance criteria (e.g., minimum sensitivity, specificity, or precision values) and the specific reported device performance values for these criteria are not provided in the summary. The core of the submission is to demonstrate "substantial equivalence" to a predicate device (OraSure Micro-Plate Cotinine EIA, K974234), rather than proving it meets standalone, predefined performance metrics with detailed study data.

    Therefore, many of your requested points cannot be directly extracted from the provided text.

    Here's an attempt to answer what can be inferred from the document, along with explanations for what cannot:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred/Generic)Reported Device Performance (Inferred from "Substantially Equivalent")
    Precision requirementsDeemed "substantially equivalent" to OraSure Micro-Plate Cotinine EIA
    Analytical Sensitivity requirementsDeemed "substantially equivalent" to OraSure Micro-Plate Cotinine EIA
    Interference limitsDeemed "substantially equivalent" to OraSure Micro-Plate Cotinine EIA
    Antibody Cross-reactivity limitsDeemed "substantially equivalent" to OraSure Micro-Plate Cotinine EIA
    Agreement with predicate device (OraSure Micro-Plate Cotinine EIA) for self-claimed patient specimens and diluted samples"Substantially equivalent" when using self-claimed specimen results and urine results as references.

    Explanation: The document focuses on declaring substantial equivalence to the predicate device, K974234. It states that "performance characteristic studies on precision, analytical sensitivity, interference and antibody cross-reactivity showed that the LabOne Micro-Plate Cotinine EIA is substantially equivalent" and that "sample results from this two test systems are substantially equivalent when using self-claimed specimen results and urine results as references." However, the specific quantitative acceptance criteria (e.g., "precision CV < X%") and the actual performance metrics achieved (e.g., "precision CV was Y%") are not reported in this summary.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated. The document mentions "self-claimed patient specimens and diluted samples" were tested, but the number of these samples is not provided.
    • Data Provenance: Not explicitly stated. However, given the context of a US FDA submission, it's highly likely the data would be derived from studies relevant to the US regulatory environment. The terms "self-claimed patient specimens" suggest human biological samples, but their origin (country, etc.) is not specified.
    • Retrospective or Prospective: Not explicitly stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not mentioned.
    • Qualifications of Experts: Not mentioned.
    • Explanation: For an in vitro diagnostic (IVD) device like this, ground truth would typically be established through a reference method or clinical status, rather than expert interpretation of images or other subjective data. So, the concept of "experts" in the context of radiologists, for example, is not applicable here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable/not mentioned. This type of adjudication is typically for subjective assessments (e.g., reading medical images). For an IVD like this, ground truth is usually objectively determined via a reference standard or clinical outcome.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This is an in vitro diagnostic device, not an imaging device requiring human reader interpretation in the same way. The device's performance is compared to a predicate device and reference methods, not human readers.
    • Effect size of human readers improvement: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Yes, in essence. The device itself (LabOne Micro-Plate Cotinine EIA) is a diagnostic test kit that measures cotinine. Its "performance" refers to how accurately it detects or quantifies cotinine, irrespective of human interpretation in the loop in the way AI applications are often discussed. The comparison data described ("Performance characteristic studies on precision, analytical sensitivity, interference and antibody cross-reactivity") are effectively standalone performance evaluations against established analytical standards and the predicate device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: The document mentions "using self-claimed specimen results and urine results as references." This implies that the presence or absence of cotinine, or its concentration, was established using:
      • Self-claimed patient information: Patients' own statements about tobacco use.
      • Urine results: Testing of urine samples (likely using a validated reference method for cotinine) as an independent corroborator.
      • Additionally, the "performance characteristic studies" would rely on controlled samples with known cotinine concentrations or known interfering substances, which serve as a form of ground truth for analytical validity.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable/Not mentioned. For a traditional immunoassay kit like this, there isn't a "training set" in the sense of machine learning algorithms. The device's components (reagents, antibodies, etc.) are developed and optimized, but not "trained" on data in the same way.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable, as there isn't a "training set" in this context. The development of such a kit would involve significant R&D to select and optimize reagents that provide the desired biochemical specificity and sensitivity. The "ground truth" during this development would be the known concentration of cotinine (or other analytes) in characterization samples.