Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K014195

    Validate with FDA (Live)

    Date Cleared
    2002-03-19

    (88 days)

    Product Code
    Regulation Number
    862.3550
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K973645

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Boston Scientific Corporation (BSC) Kayak™ Hydrophilic Guide Wires are intended to facilitate the placement of a catheter during diagnostic or interventional procedures. The wire can be torqued to facilitate navigation through tortuous arteries and/or avoid unwanted side branches. This device is not intended for use in the coronary arteries or the neurovasculature.

    Device Description

    The Kayak™ Hydrophilic Guide Wires are steerable guide wires available in a nominal diameter of 0.035 and 0.038 inches and nominal lengths of 80, 150 and 180 centimeters. The available tip flexibilities will be • Standard, • Stiff • Long Taper. The distal three centimeters of all models are available in straight and pre-formed angled tip design. The Kayak™ Hydrophilic Guide Wire consists of the following three components: • nickel titanium core • polyurethane jacket • hydrophilic coating. The full length of the guide wire has a polyurethane jacket containing tungsten encasing the core wire. The full length of the guide wire has a hydrophilic coating for lubricity.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and study information for the Kayak™ Hydrophilic Guide Wires:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Tensile StrengthNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
    Torque StrengthNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
    Cantilever StiffnessNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
    Tip ProlapseNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
    Peel TestNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
    Lubricity and DurabilityNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
    BiocompatibilityNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
    Product Shelf LifeNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"

    Note: The document states that the "Test results demonstrate that the Kayak™ Hydrophilic Guide Wires met all evaluation acceptance criteria and are adequate for their intended use." However, the specific quantitative or qualitative acceptance criteria for each test are not detailed in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for each non-clinical test. It only lists the types of tests performed. The data provenance is described as "in vitro testing and evaluation," meaning the tests were conducted in a laboratory setting, not on human subjects. There is no information regarding the country of origin of the data provided beyond Boston Scientific being located in Maple Grove, Minnesota, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable as the described study is a non-clinical, in vitro performance test of a medical device (guide wire). There is no "ground truth" established by human experts in the context of diagnostic performance or clinical outcomes. The "ground truth" here is the physical and mechanical properties of the device, measured against predefined engineering specifications and standards.

    4. Adjudication Method for the Test Set

    This information is not applicable as the study is a non-clinical, in vitro performance test. Adjudication methods are typically relevant in clinical studies or diagnostic evaluations where expert consensus is needed to resolve discrepancies in interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This document describes non-clinical, in vitro testing of a medical device's physical and functional properties, not a study evaluating human reader performance with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone algorithm-only performance study was not conducted. This document describes the performance of a physical medical device (guide wire), not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests described is based on engineering specifications, material science properties, and performance standards for guide wires. For example, tensile strength would be measured and compared against a specified minimum force the wire must withstand before breaking. Lubricity would be measured against a standard for friction.

    8. The Sample Size for the Training Set

    This information is not applicable as the study is a non-clinical performance test of a physical device. There is no "training set" in the context of machine learning or AI algorithm development.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1