510(k) Data Aggregation

    K Number
    K060617

    Validate with FDA (Live)

    Date Cleared
    2006-04-05

    (28 days)

    Product Code
    Regulation Number
    862.1175
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K972669, K013173, K014099

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PTS PANELS HDL Cholesterol Test Strips are intended to measure high density lipoprotein cholesterol in whole blood. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

    Device Description

    PTS PANELS HDL Cholesterol Test Strips consist of chemical reagents in dry form. When whole blood is placed on the test strip, only the HDL cholesterol fraction passes through to the reaction layer. The HDL cholesterol reacts with a dried cholesterol reagent to produce color. The HDL concentration is proportional to the intensity of the color developed and is measured by a CardioChek brand analyzer.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria and a study demonstrating the device meets those criteria. It is a 510(k) summary for the PTS PANELS HDL Cholesterol Test Strips, focusing on its substantial equivalence to predicate devices and its intended use.

    However, I can extract information related to the device and its intended use based on the provided text.

    Here's an analysis of what is and is not available in the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Not Available: The document does not explicitly state acceptance criteria (e.g., specific accuracy, precision, or correlation thresholds) nor does it provide detailed performance data from a specific study against such criteria. The "Measuring Range" of 25-85 mg/dL HDL Cholesterol is mentioned, but this is a device specification, not an acceptance criterion related to a study outcome.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Available: The document does not describe any specific test set, its sample size, or its provenance (country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not Available: Since no specific clinical or performance study is detailed, there's no mention of experts establishing ground truth or their qualifications.

    4. Adjudication Method:

    • Not Available: No adjudication method is mentioned as there's no described study involving expert review.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Not Applicable/Not Available: This type of study is typically for image-based diagnostic aids involving human readers. The PTS PANELS HDL Cholesterol Test Strips are a dry chemistry test strip for quantitative measurement, not an imaging device requiring human interpretation in this context. Therefore, an MRMC study is not relevant, and no such study is described.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop):

    • Partially Addressed by Nature of Device: The device itself is a standalone diagnostic tool, as it directly measures HDL cholesterol. However, the document does not describe a study explicitly proving its standalone performance against a defined set of acceptance criteria. The entire 510(k) submission process inherently evaluates the standalone performance of the device against predicate devices, but specific details of such a study (sample size, methods, results) are not in this summary.

    7. Type of Ground Truth Used:

    • Implied (Not Explicitly Stated for a Study): For a device like this, the "ground truth" would typically be established by a reference laboratory method (e.g., a standardized enzymatic assay or a CDC-certified method) against which the device's measurements are compared. The document states the "Chemistry Method" is "Enzymatic (cholesterol esterase/oxidase) trinder method for cholesterol," which suggests a well-established biochemical principle. However, it does not detail how this ground truth was applied in a specific performance study for the modified device.

    8. Sample Size for the Training Set:

    • Not Available: The document does not describe any training set or its sample size. This device is a chemical assay, not an AI/ML algorithm that typically requires a training set in the conventional sense.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable/Not Available: As no training set is described, there's no information on how its ground truth was established.

    Summary of what is available from the document:

    • Device Name: PTS PANELS HDL Cholesterol Test Strips
    • Intended Use: To measure high density lipoprotein cholesterol in whole blood for the diagnosis and treatment of lipid disorders, atherosclerosis, and various liver and renal diseases.
    • Technology: Dry chemistry test strip for use with PTS reflectance photometer.
    • Measuring Range: 25-85 mg/dL HDL Cholesterol.
    • Chemistry Method: Enzymatic (cholesterol esterase/oxidase) trinder method for cholesterol.
    • Key Modification: Change in chemistry fractionation method from Phosphotungstic acid precipitation to Dextran sulfate inhibiting method, and removal of glass fiber from the blood separation membrane.
    • Predicate Devices: K990247 and K993377, the unmodified PTS PANELS HDL Cholesterol Test Strips from the same company.
    • Regulatory Class: Class I.

    To answer the prompt fully, more detailed study reports (e.g., analytical and clinical validation studies) would be required, which are typically found in the full 510(k) submission, not just the summary provided here. The provided text is a summary focusing on substantial equivalence to predicate devices, not a detailed performance study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013173

    Validate with FDA (Live)

    Date Cleared
    2001-10-22

    (28 days)

    Product Code
    Regulation Number
    862.1175
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K972669, K981493, K990247, K991894

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lipid Panel Test Strips are intended to measure cholesterol and triglycerides in whole blood on a BioScanner Plus analyzer. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism , or various endocrine disorders.

    The BioScanner Plus is intended to be used by healthcare professionals to measure blood analytes on a test strip.

    Device Description

    The BioScanner Plus is a reflectance photometer for in vitro diagnostic use with Polymer Technology System's (PTS) test strips to measure color developed in a chemical reaction that occurs when whole blood is placed on a test strip. The color is measured and converted into concentration for each analyte.

    The Lipid Panel Test Strips are dry phase test strips that are constructed from a plastic strip holder that holds chemically treated membranes. When whole blood is placed on the test strip, the membranes first separate and isolate the red blood cells, allowing the serum/plasma to flow to the reaction membranc and react to produce a color change. The Lipid Panel Test Strips are for in vitro diagnostic use with the BioScanner Plus reflectance photometer.

    The only modification that was made to the BioScanner to produce the BioScanner Plus was: A fourth optical read window was added to the three optical read windows in the BioScanner.

    The only modification that was made to the Cholesterol, HDL Cholesterol and Triglycerides test strips to produce the Lipid Panel Test Strips is that the reaction membranes from the three individual test strips were assembled together on one test strip holder that holds all three at once.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BioScanner Plus Lipid Panel Test Strips, based on the provided 510(k) summary:

    This document is a 510(k) summary for a medical device modification, not a standalone study report. As such, it primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing detailed acceptance criteria or a comprehensive study analysis of a novel device. Therefore, some information requested, particularly regarding detailed acceptance criteria, specific performance metrics, sample sizes for test sets, expert-derived ground truth, and MRMC studies, is not explicitly present in this type of submission. The focus is on comparing the modified device's performance to the predicate device's established performance.

    1. Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly state numerical acceptance criteria (e.g., specific accuracy, precision, or correlation thresholds). Instead, the criterion for acceptance is "substantial equivalence" to the predicate device. This implies that the modified device (BioScanner Plus and Lipid Panel Test Strips) must perform comparably to the predicate device (BioScanner and its individual Cholesterol, HDL Cholesterol, and Triglycerides Test Strips).

    The "performance" described is the functionality and analytical method, with the implicit understanding that the analytical performance (accuracy, precision) would be equivalent to the predicate. The changes made were structural (four optical windows instead of three) and organizational (combining three test strips into one Lipid Panel Test Strip), not fundamental changes to the chemical reaction or measurement principle.

    MetricAcceptance Criteria (Implicit, based on substantial equivalence)Reported Device Performance
    Determination of Total Cholesterol, HDL Cholesterol, and TriglyceridesEquivalent to predicate deviceDetermines Total Cholesterol, HDL Cholesterol, and Triglycerides concentrations in whole blood
    Method of Red Blood Cell SeparationEquivalent to predicate deviceUses the same method of red blood cell separation
    Reaction MembranesEquivalent to predicate deviceUses the same reaction membranes for blood separation and color development
    Measurement PrincipleEquivalent to predicate deviceUses a reflectance photometer to convert the intensity of color produced in a colorimetric chemical reaction into results
    EEPROM Chip FunctionalityEquivalent to predicate deviceIncorporates a lot-specific EEPROM chip for curve information and a failsafe feature

    2. Sample Size for the Test Set and Data Provenance

    This information is not provided in the 510(k) summary. 510(k) summaries for device modifications often do not include the detailed sample sizes of validation studies if the modifications are considered minor and don't significantly alter the fundamental performance. The focus is on demonstrating that the changes do not negatively impact performance compared to the predicate, often through bridging studies rather than full-scale clinical trials.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the 510(k) summary. Given that this is an in vitro diagnostic device measuring standard analytes, ground truth would typically be established using validated laboratory reference methods, not necessarily by human expert consensus as might be the case for image-based diagnostic aids.

    4. Adjudication Method for the Test Set

    This information is not provided in the 510(k) summary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study was not done. This type of study is typically performed for diagnostic aids where human interpretation is involved (e.g., radiology AI). The BioScanner Plus is an automated in vitro diagnostic device for quantitative measurements; direct human reader involvement in the measurement process for improvement with AI assistance is not applicable here.

    6. Standalone Performance Study (Algorithm Only)

    The 510(k) summary describes a standalone device (BioScanner Plus with Lipid Panel Test Strips). However, it does not detail a specific "algorithm only" study in the sense of comparing an AI algorithm's performance in isolation. The device itself is the "algorithm only" in the context of its chemical and optical measurement system. The performance is assessed by comparing its output (analyte concentrations) to accepted reference methods or the predicate device, not by comparing an AI algorithm's output to human expert readings.

    7. Type of Ground Truth Used for the Test Set

    The type of ground truth used would almost certainly be reference laboratory methods (e.g., enzymatic assays, gas chromatography-mass spectrometry, or other established analytical methods) for Cholesterol, HDL Cholesterol, and Triglycerides. This is standard practice for validating in vitro diagnostic devices. However, the specific methods are not detailed in this summary.

    8. Sample Size for the Training Set

    This information is not provided in the 510(k) summary. The BioScanner Plus is a reflectance photometer, which uses a chemical reaction and optical detection, not an AI or machine learning algorithm that typically requires a distinct "training set" in the common sense. The "training" in this context refers to the calibration curves established for each lot of test strips.

    9. How the Ground Truth for the Training Set Was Established

    The "ground truth" for the "training set" (calibration) in this device is established via a multipoint curve that is set for each lot of test strips. This involves measuring known concentrations of the analytes (standards) to generate a calibration curve. This curve information is then programmed into a lot-specific electronically erasable programmable read-only memory (EEPROM) chip. The exact methods for determining the "known concentrations" (standards) are not detailed in this summary but would involve highly accurate reference methods.