The WallFlex Biliary Transhepatic Stent System is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms.
The WallFlex™ Biliary Transhepatic Stent System is comprised of two components: the implantable WallFlex Biliary Transhepatic stent and the delivery system.
The stent is offered uncovered or covered. The covered stents are offered as fully covered, or partially covered with a Permalume™ stent covering. The stent wires have a radiopaque core to improve radiopacity.
The stent is preloaded onto the delivery system, which has radiopaque marker bands used to aid in imaging during deployment of the stent. The delivery system accommodates a 0.035 in (0.89 mm) guidewire.
Here's an analysis of the provided text regarding the WallFlex™ Biliary Transhepatic Stent System, focusing on acceptance criteria and supporting studies:
Acceptance Criteria and Device Performance for WallFlex™ Biliary Transhepatic Stent System
This device appears to be a medical stent system, and as such, the "acceptance criteria" and "device performance" are primarily focused on the physical and biological characteristics of the device itself, rather than diagnostic accuracy or algorithmic performance usually seen in AI/software devices. The study conducted is a demonstration of substantial equivalence to existing predicate devices through a combination of biocompatibility and performance bench testing.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria/Tests | Reported Device Performance |
|---|---|---|
| Biocompatibility | Cytotoxicity | Passed (for delivery system), Stent identical to predicate (K061231, K081733, K083374, K083627) - no further testing. |
| Sensitization | Passed (for delivery system) | |
| Irritation | Passed (for delivery system) | |
| Acute Systemic Toxicity | Passed (for delivery system) | |
| USP Physicochemical Test | Passed (for delivery system) | |
| Material Mediated Rabbit Pyrogen Test | Passed (for delivery system) | |
| Performance Testing | Deployment | Successfully completed, achieved substantial equivalence. |
| Reconstrainment | Successfully completed, achieved substantial equivalence. | |
| Guidewire Passage | Successfully completed, achieved substantial equivalence. | |
| Trackability/Pushability | Successfully completed, achieved substantial equivalence. |
2. Sample Sizes Used for the Test Set and Data Provenance
Due to the nature of the device (a physical medical stent system) and the type of evaluation performed (bench testing for substantial equivalence), the concept of "test set sample size" and "data provenance" (country, retrospective/prospective) in the context of data-driven AI models does not directly apply here.
- The "test set" for biocompatibility comprised samples of the delivery system and relied on the stent component being identical to previously cleared devices. The exact number of samples for each biocompatibility test is not specified in the summary but would be standard for EN ISO 10993-1:2009.
- The "test set" for performance testing involved physical samples of the WallFlex™ Biliary Transhepatic Stent System. The exact number of units tested for each performance metric (deployment, reconstrainment, etc.) is not specified, but the summary states "comparative performance testing was successfully completed."
- There is no mention of country of origin for data or retrospective/prospective study design as this was a benchtop engineering and materials performance study.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable to this type of device submission. Ground truth, in the context of medical AI/software devices, refers to the definitive determination of a condition or diagnosis, often established by expert consensus or histological examination. For a physical stent system, the "truth" is whether the device meets its engineering and material specifications, which is determined by objective physical and chemical testing rather than expert interpretation of medical data.
4. Adjudication Method for the Test Set
Not applicable. The evaluation was based on objective physical and chemical testing, not subjective interpretation requiring adjudication of agreement between experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is not an AI/software diagnostic device, thus an MRMC study comparing human readers with and without AI assistance is irrelevant.
6. Standalone (Algorithm Only) Performance Study
Not applicable. This is a physical medical device, not an algorithm.
7. Type of Ground Truth Used
The "ground truth" for this device's evaluation was based on:
- Established Biocompatibility Standards: EN ISO 10993-1:2009 for the delivery system materials.
- Engineering and Design Specifications: For the performance characteristics (deployment, reconstrainment, guidewire passage, trackability/pushability). The success of these tests is determined by adherence to predefined, objective performance criteria.
- Identity to Predicate Device Materials: For the stent component's biocompatibility, the ground truth was "identity to the current WallFlex Biliary RX Stent."
8. Sample Size for the Training Set
Not applicable. There is no training set in the context of an AI/machine learning model for this physical medical device. The "training" for such devices involves design, manufacturing, and R&D processes, not data-driven model training.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set as understood in AI/ML contexts.
Ask a specific question about this device
(94 days)
The WALLSTENT® Biliary Endoprosthesis is intended for use in the treatment of biliary strictures produced by malignant neoplasms.
The Wallstent® Biliary Endoprosthesis is indicated for the palliation of malignant neoplasms in the biliary tree.
The WALLSTENT® Biliary Endoprosthesis is a self-expanding prosthesis constructed of biomedical superalloy and an elastomeric polymer. Smaller diameter models may utilize a radiopaque core. The prosthesis is a braided wire structure that may be covered with an elastomeric polymer in selected models. The outward radial force along with the ends of the device serve to stabilize the prosthesis after implanted. The stent's purpose is to increase or maintain the inner lumen diameter of the biliary duct.
The stent is placed by means of a delivery system. The delivery system is a coaxial tubing assembly that constrains the prosthesis until it is released in a controlled manner. The release of the stent is accomplished by retracting the outer sheath. The prosthesis is packaged constrained on the delivery system ready for placement. The system is sterile and intended for single use only.
This document (K993232) is a 510(k) premarket notification for a modified medical device delivery system, not an AI/ML powered device. Therefore, the questions related to AI/ML specific aspects (like training/test sets, ground truth methodology, expert consensus, MRMC studies, standalone performance) are not applicable.
The submission focuses on demonstrating substantial equivalence of a new delivery system for an existing product (WALLSTENT® Biliary Endoprosthesis). The acceptance criteria and the study to prove it are based on in vitro testing comparing the new delivery system to the predicate device's delivery system.
Here's an interpretation of the relevant information provided:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria (Inferred from "substantially equivalent") | Reported Device Performance | Study Type |
|---|---|---|---|
| Deployment Forces | Comparable to current delivery systems | Demonstrated to be comparable | In vitro testing |
| Handling Characteristics | Comparable to current delivery systems | Demonstrated to be comparable | In vitro testing |
| Reduced Profile | Smaller French size than predicate device's delivery system | Achieved reduced profile | Design comparison |
Explanation:
The core acceptance criterion for this 510(k) is that the new Unistep™ Plus Delivery System with a reduced profile is "substantially equivalent" to the previously cleared Unistep™ Plus Delivery System (K964119). This substantial equivalence is specifically assessed in terms of deployment forces and handling characteristics. The reduced profile itself is a design change, and the 510(k) aims to show this change doesn't negatively impact performance.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated in the provided text. The document only mentions "results of in vitro testing."
- Data Provenance: In vitro testing. This means the testing was conducted in a laboratory environment, not on human or animal subjects. Details on the specific country of origin are not provided, but it would have been conducted by Boston Scientific Corporation.
- Retrospective or Prospective: Not applicable as it's in vitro testing.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
- This question is not applicable. The "ground truth" for in vitro testing of a physical device is typically based on engineering specifications, industry standards, and measurable physical properties, not expert human interpretation in the way AI/ML ground truth is established.
4. Adjudication Method for the Test Set
- This question is not applicable. As it's in vitro testing, there would be no subjective adjudication of results by human experts in the way AI/ML studies require. Test results would be objectively measured and compared against predefined engineering tolerances or equivalence criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- This question is not applicable. This is not an AI/ML device; it's a physical medical device delivery system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- This question is not applicable. This is not an AI/ML device.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
- This question is not applicable in the context of AI/ML ground truth. For this device, the "ground truth" for the in vitro testing would be the performance characteristics of the predicate delivery system (K964119) and established engineering/design specifications for safe and effective device function. The new delivery system's performance needed to fall within acceptable parameters defined by these benchmarks to be considered substantially equivalent.
8. The Sample Size for the Training Set
- This question is not applicable. This is not an AI/ML device, so there is no training set in the AI/ML sense.
9. How the Ground Truth for the Training Set Was Established
- This question is not applicable. As there is no AI/ML training set, there is no ground truth established for it.
Ask a specific question about this device
(118 days)
The Cordis PERFLEX Stainless Steel Stent and Delivery System is intended for use in the palliation of malignant neoplasms in the biliary tree.
The Cordis PERFLEX Stainless Steel Stent and Delivery System consists of the following components:
- The balloon-expandable PERFLEX Stainless Steel Stent; .
- A delivery balloon catheter (either the Cordis PowerFlex or OPTA 5 Balloon . Catheter, depending on the stent size); and,
- . An insertion tool.
The Cordis PERFLEX Stainless Steel Stent and Delivery System features a balloonexpandable, stainless steel, welded wire stent that is delivered by a balloon catheter. The stent is provided premounted upon its associated balloon catheter. The Cordis PowerFlex and OPTA 5 Balloon Catheters are used for the delivery of the PERFLEX Stainless Steel Stents. The Cordis PERFLEX Stainless Steel Stent and Delivery System is advanced over a guidewire through a sheath lumen to an obstruction site in the biliary tree where the balloon is inflated to expand the stent. The balloon is then deflated and removed.
Also included with the Cordis PERFLEX Stainless Steel Stent and Delivery System is a stainless steel insertion tool that can be placed within the hemostatic valve of a commercially available introducer sheath to protect the stent / balloon catheter assembly during its insertion. Once the stent/balloon assembly passes through the valve, the insertion tool is removed.
The Cordis PERFLEX Stainless Steel Stent and Delivery System is provided sterile and is intended for single use only.
This document, a 510(k) summary for the Cordis PERFLEX Stainless Steel Stent and Delivery System, focuses on demonstrating substantial equivalence to previously cleared devices rather than establishing novel safety and effectiveness criteria through a new clinical study with acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, specific study design, and performance metrics is not present in this submission.
Here's a breakdown of what can be extracted and what is not applicable:
1. A table of acceptance criteria and the reported device performance
This information is not provided in the document. The 510(k) summary for the Cordis PERFLEX Stainless Steel Stent and Delivery System states: "The safety and effectiveness of the Cordis PERFLEX Stainless Steel Stent and Delivery System have been demonstrated via data collected from nonclinical tests and analyses." It does not provide specific acceptance criteria or reported device performance metrics from a clinical study. The submission focuses on demonstrating substantial equivalence to predicate devices, implying that the established safety and effectiveness of those predicates are sufficient.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The submission references "nonclinical tests and analyses," which are typically bench tests or animal studies, not human clinical trials. Thus, there is no "test set" in the context of human data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. As there is no human "test set" or clinical study with ground truth established by experts, this is not applicable.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the document. Not applicable as there is no human test set requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not provided in the document. MRMC studies are typically for AI/diagnostic imaging devices. This device is a medical implant (stent), and its approval process does not involve MRMC studies or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not provided in the document. This is not an AI algorithm but a physical medical device; therefore, "standalone algorithm performance" is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This information is not provided in the document beyond the mention of "nonclinical tests and analyses." For a physical device like a stent, ground truth in nonclinical tests would likely involve engineering specifications, materials testing results, and potentially animal model outcomes demonstrating mechanical integrity, patency, and biocompatibility, as compared to predicate devices.
8. The sample size for the training set
This information is not provided in the document. There is no mention of a "training set" in the context of machine learning or AI. For nonclinical tests, the "sample size" would refer to the number of devices tested, but this detail is not given.
9. How the ground truth for the training set was established
This information is not provided in the document. Not applicable, as there is no training set in the context of AI. For nonclinical tests, the "ground truth" would be established by validated measurement techniques and adherence to engineering standards.
Summary of the Document's Approach to Demonstrating Safety and Effectiveness:
The document leverages the concept of substantial equivalence to predicate devices. This means that instead of conducting a new clinical study with specific acceptance criteria, and reporting performance against those criteria, the manufacturer is arguing that their new device is fundamentally similar in design, materials, components, method of delivery, and intended use to devices already cleared by the FDA. The "performance data" mentioned refers to nonclinical tests that confirm the new device's characteristics align with those of the predicate devices and meet established engineering and biocompatibility standards. The FDA's clearance (K980653) confirms that they found the device to be substantially equivalent.