Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy PULSE Thoracolumbar Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

    The DePuy PULSE Thoracolumbar Screw System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

    The ISOLA, VSP, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are pedicle screw systems intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

    The ISOLA, VSP, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

    The ISOLA, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are also a hook and sacral/iliac screw fixation system of the indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).

    The ISOLA, VSP, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems when used with pedicle screws are indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). Levels of fixation are for the thoracic, lumbar and sacral spine.

    The ISOLA and MOSS MIAMI Spine Systems when used as anterior thoracic/lumbar screw fixation systems, are indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).

    The MONARCH Spine System Dual Rod Connectors can be used to connect MONARCH Spine System Rods to rods of other DePuy Spine 4.75mm, 5.5mm, and 6.35mm diameter rod systems.

    The EXPEDIUM SFX Cross Connector System is designed to transversely connect two rods used in posterior spinal instrumentation constructs. The EXPEDIUM SFX Cross Connector System devices are intended for use with components of the commercially available EXPEDIUM, VIPER, VSP, ISOLA, MONARCH, MOSS MIAMI, and TiMX Spine Systems. Please refer to the labeling of the aforementioned spinal systems indications for use.

    The EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER 2 System, VIPER SAI, and VIPER PRIME is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic. Iumbar and sacral spine.

    The EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER 2 System, VIPER SAI, and VIPER PRIME is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

    When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER System, VIPER 2 System, VIPER PRIME is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER 2 System metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER System, VIPER 2 System, and VIPER PRIME are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The EXPEDIUM Growing System is indicated for patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The EXPEDIUM Growing System may be used with any cleared traditional 4.5 and 5.5 EXPEDIUM Spine Systems. The EXPEDIUM Growing Spine System is not intended to be used with 4.0mm diameter screws.

    When used in conjunction with CONFIDENCE High Viscosity Spinal Cement, the VIPER and EXPEDIUM Fenestrated Screw Systems are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The VIPER and EXPEDIUM Fenestrated Screw Systems augmented with the CONFIDENCE High Viscosity Spinal Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.

    Device Description

    The Cross over and E-Z Link Cross Connectors are designed to transversely connect two rods used in spinal instrumentation constructs. The connector minimizes torsional forces on the construct, thus reducing the micromotion and the probability of the construct shifting after placement. The CrossOver and E-Z Link Cross Connectors are indicated as part of the ISOLA Spinal System.

    The DePuy PULSE Thoracolumbar Screw System consists of a variety of rods, pedicle screws, connectors, setscrews and other connection components used to build a spinal construct. The DePuy PULSE Thoracolumbar Screw System is offered in titanium material in sizes ranging from 4.35mm to 12.0mm in shank diameter and 20mm to 130mm in length. Screw shanks with 7.0mm and smaller are assembled from the top of the outer head, while screws with 7.5mm and larger are assembled from the bottom using a flex ball. The implant components can be rigidly locked into a variety of configurations, with each construct being made for the individual case.

    The EXPEDIUM, VIPER and VIPER PRIME screws with fenestrations are designed with a cortical fix screw shank that is fully cannulated with lateral fenestrations at the distal end and may be used in conjunction with, or without the CONFIDENCE High Viscosity Spinal Cement. The cannulation and fenestrations allow for the injection of bone cement through the screws into the vertebral body in patients with severe osteoporosis or pathological fractures due to osteoporosis and tumor in the thoracic, lumbar and sacral spine.

    The EXPEDIUM SFX Cross Connector System is designed to transversely connect two rods used in constructs comprising spinal instrumentation components. It features an advanced top-loading design and unique snap-fit feature that simplify initial placement and tightening procedures which helps avoid impingement of surrounding anatomy. The EXPEDIUM SFX Cross Connector system incorporates a broad range of implant options in both and adjustable configurations for an optimal balance between dural clearance and implant profile.

    The EXPEDIUM VERSE Spine System is designed to provide intraoperative polyaxial to monoaxial conversion. It facilitates easier rod capture and provides a powerful and precise reduction mechanism. EXPEDIUM VERSE is a reduced profile thoracolumbar implant for use for with wide range of patient statures. EXPEDIUM VERSE is a selfcontained, efficient, and intuitive instrument system that is compatible with EXPEDIUM 5.5 rods, hooks and mono screws to enhance versatility.

    The ISOLA Spine System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The ISOLA Spine System consists of connectors, open and closed iliac screws, iliac bolts, and angled closed iliac screws. These components have been designed to allow for rigid fixation of the sacral and pelvic regions of the spine.

    The MONARCH Spine System is both a rod-based and plate-based system designed to interface with various spinal anatomies. The plate-based system consists of pedicle screws, spine plates, transverse connectors. J-hooks, washers, nuts and set screws. The rod-based system consists of spinal rods, pedicle screws, set screws, various slotted connectors, and transverse connectors.

    The TiMX Low Back System is a construct that consists of pedicle and sacral screws, spine plates, nuts, washers, and transverse connectors. The TiMX Low Back System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of chronic instabilities and deformities of the thoracic, lumber, and sacral spine.

    The MOSS MIAMI Spine System is a closure mechanism which secures the rod to the screw in one single, simple step. The unique square threads balance the forces vertically, creating a secure assembly. It consists of anatomic hooks, monoaxial screws, polyaxial screws and a dual closure mechanism.

    The VIPER PRIME System is novel technique for percutaneous pedicle screw placement and posterior stabilization. This innovative technique eliminates the need for guidewires, Jamshidi needles and pedicle preparation instruments; utilizing a stylet that is fully controlled by the screwdriver. The VIPER PRIME System enables surgeons to target pedicles and insert the screw, without the need for instrument exchanges or reconfirmation of their trajectory.

    The VIPER Sacral-Alar-Iliac Screw is an implant designed for sacropelvic fixation. The VIPER SAI Screw is optimized for Sacral-Alar-Iliac placement vis-à-vis a favored angle polyaxial head, smooth shank and robust drive feature. The in-line nature of this anchor allows not only stabilization but also correction of pelvic sagittal or coronal deformities.

    The EXPEDIUM Spine System, VIPER System, and VIPER 2 System are metallic implants intended to provide immobilization and stabilization of spinal segments. They can be used for skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine; or for posterior non-cervical pedicle screw fixation in pediatric patients as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM and VIPER/VIPER2 systems are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The EXPEDIUM Spine System consist of longitudinal rods, monoaxial screws, polyaxial screws, uni-planar screws, reduction screws, cable/wire screws, bolts, slotted connectors, wires, hooks, reduction hooks, transverse connectors, SFX Cross Connector System, dual rod connectors, sacral extenders, lateral connectors, and washers. The VIPER and VIPER 2 Systems consist of cannulated polyaxial screws, monoaxial screws, uni-planar screws, reduction screws, and rods used in a percutaneous approach.

    The VSP Spine System is indicated for degenerative spondylolisthesis, in skeletally mature patients, with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion. Levels of fixation are for the thoracic, lumbar and sacral spine. The VSP Spine System is also indicated for pedicle screw fixation for severe spondylolisthesis (Grades 3 and 4) at L5-S1, in skeletally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion us attained. Levels of fixation are from L3-S1.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device, which seeks to demonstrate substantial equivalence to previously cleared devices. Therefore, the information provided focuses on comparative data rather than a standalone clinical study with acceptance criteria in the same manner as a de novo or PMA submission might require.

    However, based on the provided text, we can infer some "acceptance criteria" related to magnetic resonance compatibility and how the study "proves" the device meets them.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from ASTM Standards)Reported Device Performance
    Magnetically Induced Torque: Device should exhibit acceptable levels of torque in an MR environment (ASTM F2213)."Results demonstrated compatibility conditions of the subject devices in the MR environment." (Implies acceptable levels were met.)
    Magnetically Induced Displacement Force: Device should exhibit acceptable levels of displacement force in an MR environment (ASTM F2052)."Results demonstrated compatibility conditions of the subject devices in the MR environment." (Implies acceptable levels were met.)
    MR Image Artifacts: Device should produce acceptable levels of image artifacts (ASTM F2119)."Results demonstrated compatibility conditions of the subject devices in the MR environment." (Implies artifacts were within acceptable limits.)
    Radio Frequency Induced Heating: Device should not exhibit excessive heating in an MR environment (ASTM F2182)."Results demonstrated compatibility conditions of the subject devices in the MR environment." (Implies heating was within acceptable limits.)

    Explanation: The "acceptance criteria" here are indirectly established by the selection of specific ASTM standards for magnetic resonance compatibility. The study "proves" the device meets these criteria by performing tests according to these standards and reporting that the "results demonstrated compatibility conditions." While the exact quantitative thresholds are not provided in this summary, the FDA's clearance implies that the performance met the requirements set by these recognized standards for MR safety.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document states that "non-clinical testing was conducted in alignment with the following standards." This usually means a set of representative devices or components are tested, but the specific number is not provided in this summary.
    • Data Provenance: The nature of the tests (magnetic resonance compatibility) indicates laboratory testing of the devices themselves, rather than human subject data. Therefore, "country of origin of the data" and "retrospective or prospective" do not directly apply in the traditional sense of clinical studies. The tests were performed to demonstrate that the device materials and design are safe within an MR environment.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This submission is for device clearance based on non-clinical (laboratory) testing for MR compatibility. Therefore, there were no experts used to establish ground truth for a test set in the clinical sense. The "ground truth" is governed by the objective measurements defined by the ASTM standards for MR compatibility.


    4. Adjudication Method for the Test Set

    Not applicable. As this is non-clinical laboratory testing, there is no adjudication method involving human interpretation or consensus. The data points are quantitative measurements against predefined ASTM standard limits.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. An MRMC comparative effectiveness study was not done. This type of study is relevant for assessing the impact of AI on diagnostic performance with human readers, which is not the purpose of this submission. This submission is for the clearance of physical medical devices (spinal systems) and their magnetic resonance compatibility labeling.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. A standalone algorithm-only performance study was not done. This submission concerns physical medical devices and their MR compatibility, not an algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests was defined by the specifications and limits within the cited ASTM standards (ASTM F2213, ASTM F2052, ASTM F2119, ASTM F2182) for magnetic resonance compatibility. These standards outline methodologies and acceptable thresholds for torque, displacement force, image artifacts, and heating.


    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" in the context of this 510(k) submission, as it relates to physical device testing for MR compatibility, not machine learning or AI algorithm development.


    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set, there is no ground truth to establish for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182478

    Validate with FDA (Live)

    Date Cleared
    2019-01-15

    (127 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K141253, K031585, K973836, K962628

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Whistler Modular Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The Whistler Modular Pedicle Screw System is a multiple component, posterior spinal fixation system which consists of pedicle screws, rods, and locking cap set screws. All of the components are available in a variety of sizes to match more closely to the patient's anatomy. All implant components are made from titanium alloy (ASTM F136). The Whistler Modular Pedicle Screw System is provided in both non-sterile and sterile forms. All implants are intended for single use only.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for the Whistler Modular Pedicle Screw System. It describes the device, its indications for use, and a comparison to predicate devices, but it does not contain information regarding an AI/ML-driven device or studies proving its performance against acceptance criteria in the context of diagnostic accuracy, image analysis, or similar AI-related applications.

    Instead, this document focuses on a physical medical device (pedicle screw system) and its mechanical performance testing to demonstrate substantial equivalence to existing devices. The performance data listed (Static Axial Compression Bending, Static Torsion, Dynamic Axial Compression Bending, Static A-P load) are standard mechanical tests for spinal implants, not AI performance metrics.

    Therefore, I cannot fulfill the request to describe acceptance criteria and a study that proves a device meets acceptance criteria in an AI/ML context using the provided text. The document is entirely about a hardware implant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112684

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-01

    (229 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K962628, K973836, K052123, K993067

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mont Blanc Pedicle Screw System is intended for noncervical pedicle fixation from the T1 to S1 vertebrae in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion.

    Device Description

    The Mont Blanc Pedicle Screw System is comprised of a variety of monoaxial and polyaxial pedicle screws and rods. The rods are provided straight and intended to interface with the screws, which are traditional saddle design. This device is intended to be used with bone graft to provide immobilization and stabilization of a spinal segment as an adjunct to fusion. The Mont Blanc Pedicle Screw System is fabricated from wrought Ti-6Al-4V (ISO 5832-3).

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the "Mont Blanc Pedicle Screw System." It describes the device, its intended use, and the basis for its substantial equivalence to previously cleared devices.

    However, the questions you've asked are about "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of device performance, sample sizes for test and training sets, expert review, and ground truth establishment.

    Based on the provided document, the device described is a pedicle screw system, which is an orthopedic implant. For this type of device, "performance testing" typically refers to mechanical testing to ensure the device's structural integrity and durability, rather than a clinical study evaluating diagnostic accuracy or algorithmic performance as would be the case for an AI/ML device.

    Therefore, the information you are requesting about

    1. Acceptance criteria and reported device performance (in terms of sensitivity, specificity, etc.)
    2. Sample sizes for test set and data provenance
    3. Number of experts and qualifications for ground truth
    4. Adjudication method
    5. MRMC comparative effectiveness study
    6. Standalone algorithm performance
    7. Type of ground truth (pathology, outcomes data, etc.)
    8. Sample size for training set
    9. How training set ground truth was established

    is not present in the provided 510(k) summary.

    The document states:

    • Performance Testing: "Testing performed indicates the Mont Blanc Pedicle Screw System is substantially equivalent to predicate devices. Testing included mechanical testing per ASTM F1717, including static and dynamic compression bending and static torsion." This refers to laboratory-based mechanical tests, not clinical studies.
    • Acceptance Criteria: These would be defined by the ASTM F1717 standard for pedicle screw systems (e.g., minimum loads for static and dynamic failure, or deformation limits) but are not explicitly detailed in the summary.
    • Device Performance: The documented performance is that it met the requirements of ASTM F1717, showing it is "substantially equivalent" to predicate devices. Specific numerical results from these tests (e.g., actual breaking strength values) are not provided in this summary.

    In summary, this document does not contain the type of information you are asking for, which is typically found in submissions for AI/ML-driven diagnostic devices or devices that rely on complex data analysis and expert interpretation. It describes a traditional, passive mechanical implant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090230

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-03-06

    (35 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

    The EXPEDIUM™ Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

    Device Description

    The subject EXPEDIUM Spine System components consist of 4.5mm and 6.35mm rods and are available in various geometries and sizes.

    AI/ML Overview

    The provided text describes the EXPEDIUM Spine System, a medical device intended for spinal immobilization and stabilization. However, it does not contain specific acceptance criteria, detailed study designs, or performance metrics in a way that would allow for a complete answer to your request regarding acceptance criteria and a study proving the device meets them.

    The document is a 510(k) summary and FDA clearance letter, which primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed performance study against pre-defined acceptance criteria.

    Here's a breakdown of what can be gleaned and what is missing, structured as requested:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not Explicitly Stated"Performance data per ASTM F 1717 were submitted to characterize the subject EXPEDIUM™ Spine System components addressed in this notification."

    Explanation: The document states that "Performance data per ASTM F 1717 were submitted." ASTM F 1717 is a standard specification for spinal implant constructs in vitro, likely involving mechanical testing (e.g., fatigue, static strength). However, the specific acceptance criteria (e.g., "must withstand X cycles at Y load," or "static strength must exceed Z N-m") and the device's measured performance against those criteria are not provided in this summary. The summary confirms that data was submitted to characterize the device, implying that a study was conducted, but it does not present the results or the pass/fail thresholds.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for any test set. The ASTM F 1717 standard would typically define the number of constructs to be tested, but this detail is not present in the summary.
    • Data Provenance: Not specified. It's an in-vitro mechanical test, so "country of origin of data" is less relevant than for clinical data, but the location of the testing facility is not provided. Retrospective or prospective classification is not applicable to an in-vitro mechanical test.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This question is not applicable to the device and performance data described. The device is a spinal implant system, and the performance data cited (ASTM F 1717) refers to mechanical properties, not diagnostic performance or clinical outcomes requiring expert interpretation to establish ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. As the performance data pertains to mechanical testing (ASTM F 1717), there is no need for expert adjudication. Mechanical tests follow standardized protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a spinal implant system, not a diagnostic imaging or AI-assisted device. Therefore, MRMC studies involving human readers and AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This device is a physical spinal implant system, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable in the typical sense of ground truth for diagnostic devices. For mechanical testing (ASTM F 1717), the "ground truth" is established by the physical properties and behavior of the device components under controlled test conditions, measured against the specifications of the standard. There is no expert consensus, pathology, or outcomes data used to establish "ground truth" for this type of performance assessment.

    8. The sample size for the training set

    • Not applicable. There is no "training set" for a physical medical device undergoing mechanical performance testing. This concept applies to machine learning algorithms.

    9. How the ground truth for the training set was established

    • Not applicable. (See point 8).

    Summary of available information:

    The 510(k) summary indicates that the EXPEDIUM Spine System is a Class III medical device intended for spinal immobilization and stabilization. It states that "Performance data per ASTM F 1717 were submitted to characterize the subject EXPEDIUM™ Spine System components." This implies that a study was conducted to evaluate the mechanical properties of the device according to the ASTM F 1717 standard. However, the specific details of this study, including the acceptance criteria, the actual performance results, sample sizes, or any form of "ground truth" that would be applicable to diagnostic or AI devices, are not present in the provided document. The document primarily serves to demonstrate substantial equivalence to previously cleared predicate devices, allowing it to proceed to market without a full PMA.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1