Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K221524

    Validate with FDA (Live)

    Date Cleared
    2022-06-24

    (29 days)

    Product Code
    Regulation Number
    882.4175
    Age Range
    0 - 100
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K960037

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sugita AVM Microclip Applier is exclusively designed to place Sugita AVM Microclips during surgical procedures.

    Device Description

    The Sugita AVM Microclip Applier has a spring handle design with a ratchet. The Sugita AVM Microclip Applier serves to open and close the Sugita AVM Microclips. Using the Sugita AVM Microclip Applier, the Sugita AVM Microclips are removed from their packaging and applied, repositioned or removed from the target tissue. The Sugita AVM Microclip Applier has no function of its own and may only be used in combination with the Sugita AVM Microclips. The Sugita AVM Microclip Applier has the same design and materials as the clip applier in the K960037. The body of the Sugita AVM Microclip Applier is manufactured from stainless steel and the jaws are manufactured from Elgiloy, a cobalt-chromium-molybdenum alloy.

    AI/ML Overview

    The provided document is a 510(k) summary for the Sugita AVM Microclip Applier (K221524). It details the device, its intended use, and its comparison to a predicate device. However, it does not describe an acceptance criteria table, a study proving device performance against acceptance criteria, or any of the detailed study information requested regarding sample sizes, expert ground truth, adjudication methods, MRMC studies, or standalone performance.

    The document focuses on demonstrating substantial equivalence to a predicate device (Aesculap Slim Clip Applier, K211572) based on similar indications for use, materials, design, and technological characteristics. The only "performance testing" mentioned is non-clinical sterilization and cleaning validation, which supports updated instructions for use, not a clinical performance study using patient data.

    Therefore, most of the requested information cannot be extracted from this document, as it pertains to a type of performance study that was not conducted or reported here.

    Here's what can be extracted and a clear indication of what is not present in the document based on your request:


    1. A table of acceptance criteria and the reported device performance

    Information Not Available: The document does not provide a table of acceptance criteria or reported device performance against such criteria. The submission focuses on demonstrating substantial equivalence through comparison of design, materials, and intended use, along with validating sterilization and cleaning parameters.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Information Not Available: No clinical test set or patient data was used for performance evaluation as described in this document. The "tests" performed were non-clinical validation of sterilization and cleaning parameters.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Information Not Available: No clinical test set requiring expert ground truth establishment was conducted or reported.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Information Not Available: No clinical test set requiring adjudication was conducted or reported.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Information Not Available: This device is a mechanical surgical applier, not an AI-powered diagnostic tool. Therefore, an MRMC study or AI assistance is not applicable and was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Information Not Available: This device is a mechanical surgical applier, not an algorithm. Therefore, standalone algorithm performance is not applicable and was not assessed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Information Not Available: No clinical ground truth was established as no clinical studies were performed for this 510(k) submission.

    8. The sample size for the training set

    Information Not Available: No training set was used as this is not an AI/machine learning device. The "training" for this submission would involve the development and validation of manufacturing processes, sterilization, and cleaning instructions.

    9. How the ground truth for the training set was established

    Information Not Available: No training set or associated ground truth, in the context of AI/machine learning, was established. The "ground truth" for the non-clinical validations would be successful completion of specified sterilization and cleaning cycles according to established scientific and regulatory standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211183

    Validate with FDA (Live)

    Date Cleared
    2022-03-02

    (316 days)

    Product Code
    Regulation Number
    882.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K960037

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sugita AVM Microclips are designed for occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures.

    Sugita AVM Microclips are not intended as a replacement for bipolar electrocoagulation of smaller vessels nor do they negate the need for immediate post-operative angiography following arteriovenous malformation surgery.

    Device Description

    The Sugita AVM Microclips were cleared by the FDA through 510(k) K960037. The Sugita AVM Microclips are manufactured from Elgiloy, a cobalt-chromiummolybdenum alloy. They are designed for the occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures. The Sugita AVM Microclips are available in the following sizes: 2mm, 3mm, 4mm and 5mm. Other than the modifications identified above, there have been no other changes to the Sugita AVM Microclips.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Subject Device K211183)Reported Device Performance (Sugita AVM Microclips)
    MRI Safety - Magnetic Field Interaction (Force)Force no greater than gravitational field. (Tested according to ASTM F2052-15 - translational attraction determined to be 14°)
    MRI Safety - Magnetic Field Interaction (Torque)Magnetically induced torque is not substantial and requires no further evaluation. (Tested according to ASTM F2213-17 – torque showed no movement or alignment)
    MRI Safety - MRI Related HeatMaximum Whole Body Averaged SAR 2-W/Kg for 60 min. continuous RF exposure. (Tested according to ASTM F2182-19 – maximum temperature rise of 1.5°C after 15 min. of continuous scan)
    MRI Safety - Imaging ArtifactThe presence of this implant produces an imaging artifact. Therefore, should carefully select pulse sequence parameters if the implant is located in the area of interest. (Tested according to ASTM F2119-2013 – image artifact extends approx. 10mm from implant with 3-Tesla gradient echo pulse sequence)
    MRI Safety - Spatial Gradient Magnetic FieldTesting supports using a value of 2,000 gauss/cm (20-T/m) for the MRI related labeling of the Sugita AVM Microclips. (5mm Sugita AVM Microclips were attached to a porcine blood vessel in order to conduct a digital force gauge-based Pull-Test.)
    Steam Sterilization ValidationValidated steam sterilization parameters are provided in the product labeling. (Tested according to ISO 17665-1 to validate steam sterilization parameters for the Sugita AVM Microclips using an FDA cleared nonwoven wrap to a sterility assurance level of (SAL) of 10-6.)
    MR Conditional ConditionsMR Conditional for Use in Both 1.5T and 3T MR Systems. (Demonstrated through the above MRI safety testing, which included tests for 1.5T and 3T MR systems where applicable.) The predicate device was only cleared for 1.5T MR Systems, making this an improvement and a new acceptance criterion for the modified device.
    Maintenance of Device Performance (Shelf-life)Performance data are not needed to establish maintenance of device performance over the shelf life of the Sugita AVM Microclips because the materials of construction of these products are highly stable metal alloys with a long history of use in surgical implants. (The device is provided non-sterile and constructed of inert materials, implying no degradation over time for its core functionality before use, as the device itself doesn't possess a shelf-life.)

    2. Sample Size for the Test Set and Data Provenance

    The provided text does not specify exact sample sizes for each MRI compatibility test beyond implying "Sugita AVM Microclips" were used for testing. For the Spatial Gradient Magnetic Field testing, it mentions "5mm Sugita AVM Microclips" were attached to "a porcine blood vessel." The provenance of this data is not explicitly stated as retrospective or prospective, nor is the country of origin. It can be inferred that these are prospective in-vitro tests conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. The tests performed are engineering-based performance tests (MRI compatibility, sterilization validation) and do not involve human expert interpretation or ground truth establishment in the traditional clinical sense. These tests rely on established scientific and engineering standards (e.g., ASTM, ISO).

    4. Adjudication Method for the Test Set

    Not applicable. As the tests are objective engineering measurements against predefined standards, there is no need for an adjudication method by human experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical medical implant (microclips), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study is irrelevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests relies on established engineering and scientific standards and measurements as defined by:

    • ASTM F2503-13 (Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment)
    • Guidance for Industry and Food and Drug Administration Staff (Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, document issued on May 20, 2021)
    • ASTM F2052-15 (Standard Test Method for Measurement of Magnetically Induced Translational Force on Medical Devices in the Magnetic Resonance Environment)
    • ASTM F2213-17 (Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment)
    • ASTM F2182-19 (Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging)
    • ASTM F2119-2013 (Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants)
    • ISO 17665-1 (Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices)

    8. The Sample Size for the Training Set

    Not applicable. This device undergoes physical performance testing, not machine learning model training.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1