510(k) Data Aggregation

    K Number
    K983509

    Validate with FDA (Live)

    Date Cleared
    1999-02-03

    (119 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K955033

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DURASUL acetabular insert is a snap-in acetabular shell liner that is manufactured from ultra-high molecular weight polyethylene (UHMWPE). The insert is available in three styles and is intended to be used in conjunction with an Inter-Op acetabular shell to replace the acetabulum during total hip arthroplasty. Inter-Op acetabular components (shell/insert combination) may be implanted with or without bone cement.

    Diagnostic indications for use of this device include:

    • patient conditions of noninflammatory degenerative joint disease (NIDJD); e.g., avascular . necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis.
    • those patients with failed previous surgery where pain, deformity, or dysfunction persists. .
    • revision of previously failed hip arthroplasty. ●
    Device Description

    Sulzer Orthopedics currently markets a polyethylene shell liner (insert) in conjunction with its Inter-Op Acetabular System. The UHMWPE insert was originally cleared via 510(k) K955033 and is available in three styles, all of which mate with various shell designs in the system.

    The purpose of this submission is to describe a different UHMWPE treatment process that was developed and patented by Massachusetts General Hospital (MGH) and Massachusetts Institute of Technology (MIT). This treatment produces a highly cross-linked polyethylene and reduces free radicals that can lead to oxidative degradation. This in turn provides a highly wear resistant surface which has been demonstrated through hip simulation.

    Sulzer Orthopedics has licensed this process for use in the manufacture of Inter-Op acetabular inserts. Sulzer Orthopedics does not intend to change the intended use of this device, nor do we intend to alter the basic geometrical design of the three insert styles. Moreover, while there is a change in process to enhance the wear properties of UHMWPE, there is no change to the raw material (GUR 1150/1050 conforming to ASTM F 648).

    AI/ML Overview

    The provided text describes a medical device, the Sulzer Orthopedics Inter-Op™ DURASUL™ Acetabular Insert, and its clearance process. However, it does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of diagnostic performance (e.g., accuracy, sensitivity, specificity).

    The submission is for a 510(k) premarket notification for a Class II medical device, which is a hip joint prosthetic component. The "study" mentioned is an in vitro wear test, not a diagnostic study with human subjects, and the acceptance criteria are related to mechanical performance rather than diagnostic accuracy.

    Therefore, many of the requested fields cannot be filled as they pertain to diagnostic device studies.

    Here's a breakdown of the available information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Criteria (Implicit)Reported Device Performance (DURASUL™)Comparator (Control) Performance
    Wear Properties (Hip Simulation)Reduced wear compared to existing insertsUndetectable wear over 5 million cyclesSteady average wear rate of approximately 12 mg/million cycles
    Mechanical PropertiesNot compromising fatigue and creep behaviorFatigue and creep behavior essentially same as standard materialFatigue and creep behavior of standard material
    Locking StrengthNot compromising shell/liner interface locking strengthLocking strength same as currently available deviceLocking strength of currently available device

    2. Sample size used for the test set and the data provenance:

    • Sample Size (Test Set): Not applicable for a diagnostic performance study. The wear test refers to "DURASUL inserts" and "CONTROL inserts," implying multiple samples were tested, but specific numbers are not provided.
    • Data Provenance: In vitro hip simulator test. Countries of origin for data or retrospective/prospective nature are not relevant here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is an in vitro mechanical performance study, not a diagnostic study requiring expert ground truth.

    4. Adjudication method for the test set:

    • Not applicable as this is an in vitro mechanical performance study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI algorithm. The performance evaluation was focused on the material properties of the insert itself in a simulated environment.

    7. The type of ground truth used:

    • Mechanical/Physical Measurements: The "ground truth" for the wear study was the direct measurement of material loss (wear) over a specified number of cycles in a hip simulator. For fatigue, creep, and locking strength, it was based on standard mechanical testing protocols and measurements.

    8. The sample size for the training set:

    • Not applicable. This is not a machine learning model.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not a machine learning model.

    Summary of Device and Study Focus:

    The provided document describes the modification of an existing acetabular insert (a component of a total hip replacement) to improve its wear characteristics. The "study" mentioned is an in vitro hip simulation wear test comparing the new DURASUL™ insert to the previous version (CONTROL). The success criteria revolve around the mechanical performance and material properties of the UHMWPE, specifically its wear resistance, fatigue, creep behavior, and locking strength. There is no diagnostic component to this device or its evaluation as described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970300

    Validate with FDA (Live)

    Date Cleared
    1997-07-03

    (160 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Patient conditions of non-inflammatory degenerative joint disease (NIDJD), e.g., 1. avascular necrosis, osteoarthritis and inflammatory joint disease (IJD), e.g., rheumatoid arthritis.
    • Those patients with failed previous surgery where pain, deformity, or dysfunction 2. persists.
    • Revision of previously failed arthroplasty. 3.

    The Natural-Hip™ System Porous Stem with HA/CSTi™ is intended only for use without bone cement.

    Device Description

    The Natural-Hip System Porous Stem with HA/CSTi is a straight stem manufactured from Ti-6Al-4V alloy conforming to ASTM F-136. The stem is available in both a collared and collarless design. The proximal anterior surface of the stem employs anterior angulation or build-up to match the anatomic angulation of a natural femur. The superior portion of the stem has a built-in anteversion of nine degrees to match the anatomic anteversion of the femur. Circumferential hydroxylapatite coated Cancellous Structured Titanium™ (CSTi™) porous coating provides biological fixation in a cementless application. The distal portion of the Natural-Hip System Porous Stem with HA/CSTi employs ribs and flutes on the anterior and posterior sides to enhance cortical contact, thereby providing rotational stability to the hip stem. In addition, the distal portion of the stem has a flared coronal slot which reduces stiffness of the stem.

    AI/ML Overview

    This 510(k) submission describes a hip implant device, not an AI/ML-enabled medical device. Therefore, the information typically requested for AI/ML device acceptance criteria and studies (such as sample sizes for test and training sets, expert qualifications, ground truth methods, MRMC studies, and standalone performance) is not applicable in this context.

    The document focuses on demonstrating substantial equivalence to predicate devices based on product description, diagnostic indications, and material composition. The FDA's letter (K970300) confirms this substantial equivalence but places limitations on labeling and promotional claims.

    Here's an attempt to interpret the request within the context of a traditional medical device submission, focusing on "acceptance criteria" as related to the device's intended use and substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance

    For traditional medical devices like this hip implant, "acceptance criteria" are typically met by demonstrating that the device is substantially equivalent to legally marketed predicate devices in terms of intended use, technological characteristics, and safety and effectiveness. The "reported device performance" is essentially the determination of substantial equivalence by the FDA.

    CriterionReported Device Performance (FDA's Determination)
    Intended Use Equivalence: The device is intended for the same uses as legally marketed predicate devices.Met: The Natural-Hip™ System Porous Stem with HA/CSTi™ is intended "to replace the anatomy of the femur in cases of total hip or hemi-hip replacement," and for "cementless (press-fit) application" for conditions like non-inflammatory degenerative joint disease (NIDJD), inflammatory joint disease (IJD), failed previous surgery, and revision. These indications are consistent with the predicate devices listed. (Refer to pages 1 and 5).
    Technological Characteristics Equivalence: The device has similar technological features (e.g., design, materials, manufacturing).Met: The device is described as "a straight stem manufactured from Ti-6Al-4V alloy conforming to ASTM F-136." It features "hydroxylapatite coated Cancellous Structured Titanium™ (CSTi™) porous coating" for biological fixation. The submission explicitly states the device is substantially equivalent to other Intermedics Orthopedics, Inc. products (Natural-Hip System Porous Stem, APR Universal Hip System) and competitive systems (Howmedica Precision Osteolock, Osteonics Omnifit with HA). The FDA's decision is "based on consideration of the specific design of stem and coating composition detailed in this application." (Refer to pages 1 and 2).
    Safety and Effectiveness Equivalence: The device is as safe and effective as legally marketed predicate devices.Met (with limitations): The FDA determined the device is substantially equivalent, allowing marketing. However, a significant limitation was placed: "You may not label or in any way promote these devices for enhanced clinical or radiographic performance, enhanced biological fixation and/or long-term stable fixation. The data presented support equivalence with no additional claims over a conventional porous-coated uncemented hip prosthesis (i.e., biological fixation, only)." This indicates the device meets the baseline safety and effectiveness of its predicates for biological fixation but cannot claim superiority or augmented performance. (Refer to page 2).

    Study that Proves the Device Meets the Acceptance Criteria:

    The "study" in this context is the 510(k) Premarket Notification itself, which presents information to the FDA to demonstrate substantial equivalence to predicate devices. This typically involves:

    • Comparison to Predicate Devices: A detailed comparison of the new device's intended use, technological characteristics (materials, design, manufacturing process), and performance data (if applicable, e.g., mechanical testing for implants) to one or more legally marketed predicate devices.
    • Performance Data (Non-Clinical): For an implant, this generally involves engineering tests, material characterization, and biocompatibility assessments to ensure the device performs as intended and is safe. The document implies such data were submitted, as the FDA based its decision on "the specific design of stem and coating composition." However, specific test results are not detailed in this summary.

    Regarding your specific numbered points (which are more applicable to AI/ML devices):

    1. Sample size used for the test set and the data provenance: Not applicable in the AI/ML sense. Mechanical testing or biocompatibility studies would use specific samples of the device and materials, but this isn't "test set" data for an algorithm.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an implant isn't established by experts in this manner. Its performance is evaluated through engineering standards and clinical observation (post-market, often).
    3. Adjudication method: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable. This is not an imaging or diagnostic device that would involve human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical implant.
    6. The type of ground truth used: For a physical implant, "ground truth" relates to its conformity to engineering specifications, material properties, biocompatibility, and mechanical performance under simulated physiological conditions. Ultimately, long-term clinical outcomes (post-market surveillance) provide the most significant "ground truth" regarding its effectiveness in patients.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    In summary, for a traditional medical device like a hip implant, the "acceptance criteria" are fundamentally about demonstrating substantial equivalence to pre-existing, legally marketed devices. The "study" is the 510(k) submission itself, supported by non-clinical performance data (e.g., material properties, mechanical testing) rather than clinical studies or AI algorithm performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970166

    Validate with FDA (Live)

    Date Cleared
    1997-04-08

    (82 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Natural-Hip System CoCr Revision Stem is intended to replace the anatomy of the femur in cases of total hip or hemi-hip replacement. The Natural-Hip System CoCr Revision Stem is intended for cemented application only and is intended for single use only. The Natural-Hip System CoCr Revision Stem is intended for use in treatment of the following: Patient conditions of non-inflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis and inflammatory joint disease (IJD), e.g., rheumatoid arthritis. Those patients with failed previous surgery where pain, deformity or dysfunction persists. Revision of previously failed arthroplasty.

    Device Description

    The Natural-Hip System CoCr Revision Stem is fabricated from either forged or wrought cobalt chromium alloy which conforms to the American Standards for Testing and Materials (ASTM) standards F-799 or F-1537, respectively. The Natural-Hip System CoCr Revision Stem features a threaded Sulzer 12/14 configured neck trunnion for attachment to IOI's currently marketed femoral heads, including Biolox and Zirconia heads, employing a Sulzer 12/14 configured bore. The Natural-Hip System CoCr Revision Stem is available in both a collared and collarless design. The proximal one-third of the stem's surface employs normalization steps and is roughened via the process of grit blasting. The normalization steps and grit blasted surface enhance cement compression and bonding for optimal fixation of the hip stem in the femoral canal. The proximal surface of the Natural-Hip System CoCr Revision Stem also employs polymethylmethacrylate (PMMA) cement centralizers to provide even cement mantle when the stem is implanted in the femur. The distal portion of the Natural-Hip System CoCr Revision Stem is bowed to mimic the anatomy of the femur. The Natural-Hip System CoCr Revision Stem employs a hole in the distal portion of the stem to allow the use of a PMMA centralizer for correct distal alignment.

    AI/ML Overview

    This is a 510(k) summary for a medical device, specifically a hip joint replacement component. It does not describe any AI/ML device or software, therefore, it does not contain information about acceptance criteria or studies related to AI/ML device performance.

    The document discusses the physical characteristics, intended use, and predicate devices for a "Natural-Hip™ System CoCr Revision Stem." It is purely about the hardware of a hip implant.

    Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, or ground truth establishment because this document does not pertain to an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K964357

    Validate with FDA (Live)

    Date Cleared
    1997-01-27

    (87 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Natural-Hip System CoCr Stem is intended to replace the anatomy of the femur in cases of total hip or hemi-hip replacement. The Natural-Hip System is intended for cemented application only. The general indications associated with the use of Natural-Hip System CoCr Stem in total hip arthroplasty include: Patient conditions of non-inflammatory degenerative joint disease (NIDJD), e.g., 1. avascular necrosis, ostooarthritis and inflammatory joint disease (JJD), e.g., rheumatoid arthritis. 2. Those patients with failed previous surgery where pain, deformity or dysfunction persiss. Revision of previously failed arthroplasty. 3.

    Device Description

    The Natural-Hip System CoCr Stem is a straight stem employing a proximal wedge shaped design. The stem is available in both a collarless design. The proximal one-third of the stem's surface is grit blasted. Normalization steps are also located on the proximal portion of the stem. The grit blasted surface and normalization steps enhance cement compression and bonding for optimal fixation of the hip stern in the femoral canal. The Natural-Hip System CoCr Stem employs a Sulzer 12/14 configured neck trumion for attachment to Intermedics Orthopedics, Inc.'s femoral heads, including Biolox and Zirconia ceramic heads, featuring a Sulzer 12/14 configured bore. The Natural-Hip System CoCr Stem has a hole in the distal portion of the stem to allow the use of a distal centralizer for correct distal alignment. The stern also employs proximal PMMA centralizers which, along with the distal centralizer, provide for an even cement mantle. This device is intended for single use only.

    AI/ML Overview

    This document is a 510(k) summary for a medical device (Natural-Hip™ System CoCr Stem), not a study report or clinical trial outcome document. Therefore, it does not contain the information required to answer your questions about acceptance criteria and study details.

    The provided text focuses on:

    • Device Description: What the device is, its materials, design features, and intended use.
    • Intended Use/Indications: The medical conditions it's designed to address and how it's used (e.g., cemented application).
    • Substantial Equivalence: Listing similar devices already on the market to establish that this new device is just as safe and effective.

    There is no information in this document about:

    1. Acceptance criteria and reported device performance: This document does not describe specific performance metrics or thresholds for success.
    2. Sample size and data provenance: No study data is presented.
    3. Number and qualifications of experts for ground truth: Not applicable as no ground truth establishment is mentioned.
    4. Adjudication method: Not applicable.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable.
    6. Standalone performance study: Not applicable.
    7. Type of ground truth used: Not applicable.
    8. Sample size for the training set: Not applicable (this device is a physical implant, not an AI algorithm).
    9. How ground truth for the training set was established: Not applicable.

    In essence, this document is a regulatory submission for a physical orthopedic implant seeking clearance based on its similarity to existing devices, not a report detailing the results of a diagnostic performance study.