Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K993837

    Validate with FDA (Live)

    Date Cleared
    1999-12-16

    (34 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K944089, K961161, K974784, 984299

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVL OPTI Critical Care Analyzer is intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, glucose, urea (BUN), total hemoglobin content and oxygen saturation in samples of whole blood, serum, plasma or dialysate in either a traditional blood gas, clinical laboratory setting or point-of-care locations by personnel minimally qualified to perform and report these results.

    Device Description

    The AVL OPTI Critical Care Analyzer is a small [4.9 x 14.3 x 9.8 in, 10 lbs], instrument using optical fluorescence for the measurement of pH, PCO2, sodium, potassium, ionized calcium, choride. glucose and urea (BUN) of whole blood, plasma, serum or dialysate as appropriate. In addition, it uses scattering for the measurement of total hemoglobin and oxygen saturation. A disposable, single-use cassette containing six optical fluorescence sensors is packaged in a sealed foil pouch which bears a bar-coded label with calibration and identification information. The OPTI can perform up to 8 tests on a single sample, determined by the type of disposable cassette used.

    AI/ML Overview

    The provided text describes the AVL OPTI Critical Care Analyzer, focusing on its substantial equivalence to other marketed devices and its technological characteristics. However, it does not explicitly state specific acceptance criteria (e.g., performance thresholds like "accuracy must be within X%") in a table format, nor does it provide a detailed study that proves the device meets such criteria with quantitative results in the way you've requested.

    Instead, the document focuses on:

    • Demonstrating substantial equivalence: By comparing the AVL OPTI to existing devices and stating that there was "no significant difference in mean values (P<0.05) obtained on measurement by the AVL OPTI from those of the predicate devices."
    • Confirming compliance with standards: Mentioning EN 50081-1, FCC Class B, EN 50081-2, and IEC 1010-1.
    • General testing methods: Describing how precision, linearity, and interferences were evaluated.

    Given this, I cannot fully complete all sections of your request as the specific quantitative acceptance criteria and detailed study results (like sample sizes for test sets, experts, adjudication methods, effect sizes, standalone performance, and training set details) are not explicitly present in the provided text.

    However, I can extract the available information and structure it as per your request, indicating where information is missing.


    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter / MetricAcceptance Criteria (Explicitly Stated in Document)Reported Device Performance (Summary from Document)
    Comparative PerformanceNot explicitly stated as a quantitative threshold."no significant difference in mean values (P<0.05) obtained on measurement by the AVL OPTI from those of the predicate devices."
    PrecisionNot explicitly stated as a quantitative threshold."Typical Within-Run (Swr), Between-Day (Sdd) and Total (ST) precision were determined..." (Specific values for Swr, Sdd, ST are not provided in this summary)
    LinearityNot explicitly stated as a quantitative threshold."Linearity for the OPTI measurement has been established against reference materials or methods." (Specific range or R-squared values are not provided in this summary)
    InterferenceNot explicitly stated as a quantitative threshold."Representative samples taken the published guidelines for evaluation of interference substances and identified from literature were evaluated." (Specific findings are not provided)
    Regulatory ComplianceCompliance with relevant standards."The AVL OPTI Critical Care Analyzer has been tested and found to comply with EN 50081-1, FCC Class B, EN 50081-2 and IEC 1010-1."
    Safety and EffectivenessDemonstrated equivalence to predicate devices."safe, effective, and equivalent to those predicate devices to which it is compared."

    2. Sample size used for the test set and the data provenance

    • Sample Size (Test Set): Not explicitly stated. The document mentions "Specimens analyzed in these tests were remnant from patient specimens of both whole blood and plasma collected for routine analysis on existing instrumentation." It does not provide the number of patient specimens or samples.
    • Data Provenance: Retrospective, using "remnant from patient specimens... collected for routine analysis on existing instrumentation." Country of origin is not specified, but the submission is to the US FDA, implying clinical relevance to the US.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified. The document states "operated by personnel trained to perform and report these analyses," which refers to the clinical personnel using the device, not experts establishing ground truth for comparisons. Ground truth seems to be established by comparison to predicate devices, not by expert review of complex images or interpretations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not specified. The comparison relies on statistical analysis of mean values against predicate devices rather than human adjudication of agreement.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, this device is a standalone analyzer, not an AI-assisted diagnostic tool requiring human reader interpretation in the context of an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Yes, the device itself is a standalone analyzer. The performance described (precision, linearity, comparisons to predicate devices) represents the device's standalone analytical capabilities.

    7. The type of ground truth used

    • Ground Truth Type: Comparison to predicate devices (other commercially available and legally marketed 'Combi Analyzers' and 'Point of Care' analyzers) served as the primary "ground truth" for demonstrating substantial equivalence. Additionally, "reference materials or methods" were used for linearity, and "N.I.S.T. traceable pH buffers" for pH standardization.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable/not specified. The document describes a "new device" that is an "improved design" of an existing AVL OPTI 1 pH/Blood Gas Analyzer. The testing described focuses on verification and validation of the new device's performance, not the training of a machine learning model.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable/not specified, as this is not an AI/ML device that requires a training set in the conventional sense. The device's operation is based on established fluorescence optode technology and optical scattering, calibrated against known standards and verified through traditional analytic methods.
    Ask a Question

    Ask a specific question about this device

    K Number
    K984299

    Validate with FDA (Live)

    Date Cleared
    1998-12-11

    (10 days)

    Product Code
    Regulation Number
    862.1145
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K944089, K961161, K974784

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVL OPTI Critical Care Analyzer is intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium and chloride, total hemoglobin content and oxygen saturation in samples of whole blood, serum, plasma or dialysate in either a traditional blood gas, clinical laboratory setting or point-of-care locations by personnel minimally qualified to perform and report these results.

    The AVL OPTI Critical Care Analyzer intended to be used for the measurement of pH, PCO2, PO2, ctHb, SO2, Na', K', Cl' and Ca* in whole blood, serum, plasma and aqueous dialysate solutions as appropriate by minimally trained personnel qualified to perform and to report these values in either a traditional blood gas, clinical laboratory setting or point-of-care locations by personnel minimally qualified to perform and report these results.

    For Professional Use Only
    For In Vitro Diagnostic Use

    Device Description

    The AVL OPTI Critical Care Analyzer is a small [4.9 x 14.3 x 9.8 in. 10 lbs], instrument using optical fluorescence for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium and choride of whole blood, plasma, serum or dialysate as appropriate. In addition, it uses optical reflectance for the measurement of total hemoglobin and oxygen saturation. A disposable, single-use cassette containing six optical fluorescence sensors is packaged in a sealed foil pouch which bears a bar-coded label with calibration and identification information. The OPTI can perform up to 8 tests on a single sample, determined by the type of disposable cassette used.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the AVL OPTI Critical Care Analyzer, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for precision, linearity, or interference. Instead, it describes compliance through comparative studies and general statements of no significant difference. Here's a summary of the reported performance:

    ParameterAcceptance Criteria (explicitly stated)Reported Device Performance
    PrecisionNot explicitly stated"Typical Within-Run (Swr), Between-Day (Sdd) and Total (ST) precision were determined from two runs per day with 2 replicates per run for 20 days on two AVL OPTI instruments using samples: serum, reduced bovine hemoglobin solution and three levels of aqueous quality control solution." (No specific numerical results provided)
    LinearityNot explicitly stated"Wherever possible, linearity for the OPTI measurement has been established against reference materials or methods. Linearity for pH of whole blood is established by measurement of blood specimens which were tonometered to various CO2 values, and measured on an AVL 995 pH/Blood Gas Analyzer standardized to N.I.S.T. traceable pH buffers, and on OPTI Critical Care Analyzers." (No specific numerical results provided)
    InterferencesNot explicitly stated (implied compliance with guidelines)"Representative samples taken the published guidelines for evaluation of interference substances and identified from literature were evaluated." (No specific numerical results or identified interferences provided)
    Clinical CorrelationNo significant difference (P<0.05) to predicate devices"In all evaluations, there was no significant difference in mean values (P<0.05) obtained on measurement by the AVL OPTI from those of the predicate devices."
    Safety and EffectivenessEquivalent to predicate devices"Analysis of the comparative measurement presented in the 510(k) for this device, together with the linearity and precision data collected during these clinical and non-clinical trials demonstrates that the AVL OPTI Critical Care Analyzer with the additional analytes: chloride and ionized calcium, is safe, effective, and equivalent to those predicate devices to which it is compared."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes used for the precision, linearity, interference, or clinical correlation tests. It mentions "samples: serum, reduced bovine hemoglobin solution and three levels of aqueous quality control solution" for precision and "remnant from patient specimens of both whole blood collected for routine analysis on existing instrumentation" for clinical tests.
    • Data Provenance: The data appears to be prospective for the clinical tests, as it uses "remnant from patient specimens... collected for routine analysis." The country of origin is not explicitly stated, but given the submitter's address (Roswell, GA) and the FDA submission, it implicitly refers to studies conducted in the United States or under US regulatory guidelines.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish ground truth. For the clinical correlation study, the comparison was made against "predicate devices" operated by "personnel trained to perform and report these analyses." The AVL 995 pH/Blood Gas Analyzer standardized to N.I.S.T. traceable pH buffers was used for linearity of pH as a reference method.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none). The clinical evaluation involved direct comparison to predicate devices, and the statistical analysis focused on the "mean values (P<0.05)."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. The device is an automated critical care analyzer, not an AI-assisted diagnostic imaging tool that would typically involve human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are essentially standalone performance evaluations of the AVL OPTI Critical Care Analyzer. It is an automated instrument that directly measures analytes, and the studies assess its precision, linearity, interference, and agreement with predicate devices without human interpretation as part of its core function.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance evaluations primarily involved:

    • Reference Materials/Methods: For linearity, this included "reference materials or methods" and specifically an "AVL 995 pH/Blood Gas Analyzer standardized to N.I.S.T. traceable pH buffers."
    • Predicate Devices: For clinical correlation, the ground truth was established by the measurements from "predicate devices" (e.g., NOVA Biomedical STAT 5, Chiron 865, AVL 995, etc.).
    • Quality Control Solutions: For precision, "aqueous quality control solution" was used.

    8. The Sample Size for the Training Set

    The document does not mention any "training set" or "training data" in the context of machine learning or AI. This device is an automated analyzer based on optical fluorescence and reflectance, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a "training set" or a machine learning component for this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974784

    Validate with FDA (Live)

    Date Cleared
    1998-02-24

    (64 days)

    Product Code
    Regulation Number
    862.1600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K944089, K961161

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVL OPTI Critical Care Analyzer is intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, total hemoglobin content and oxygen saturation in samples of whole blood, serum, and plasma in either a traditional blood gas, clinical laboratory setting or point-of-care locations by personnel minimally qualified to perform and report these results.

    Device Description

    The AVL OPTI Critical Care Analyzer is a small [4.9 x 14.3 x 9.8 in, 10 lbs], instrument using optical fluorescence for the measurement of pH, PCO2, PO2, sodium and potassium of whole blood, plasma or serum as appropriate. In addition, it uses optical reflectance for the measurement of total hemoglobin and oxygen saturation. A disposable, single-use cassette containing six optical fluorescence sensors is packaged in a sealed foil pouch which bears a bar-coded label with calibration and identification information. The OPTI can perform up to 7 tests on a single sample, determined by the type of disposable cassette used.

    AI/ML Overview

    The AVL OPTI Critical Care Analyzer is a medical device intended for measuring pH, PCO2, PO2, sodium, potassium, total hemoglobin content, and oxygen saturation in blood, serum, and plasma samples. The device uses optical fluorescence for some measurements and optical reflectance for others.

    Here's a breakdown of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document explicitly states: "In all evaluations, there was no significant difference in mean values ( P<0.05) obtained on measurement by the AVL OPTI." This serves as the primary acceptance criterion and the reported performance.

    While specific percentage or numerical thresholds for accuracy or precision are not provided as acceptance criteria in this summary, the following non-clinical tests were conducted to establish performance:

    Acceptance Criteria CategoryReported Device Performance
    Accuracy (Clinical Correlation)No significant difference in mean values (P<0.05) compared to predicate devices in a clinical setting.
    PrecisionTypical Within-Run (Swr), Between-Day (Sdd), and Total (ST) precision were determined from two runs per day with 2 replicates per run for 20 days. Specific numerical values for Swr, Sdd, and ST are not provided in this summary.
    LinearityEstablished against reference materials or methods. For pH, linearity was established by measuring tonometered blood specimens against an AVL 995 pH/Blood Gas Analyzer standardized to NIST-traceable pH buffers, and on three OPTI Critical Care Analyzers. Specific ranges or R-squared values for linearity are not provided.
    InterferencesEvaluated against representative samples identified from literature and published guidelines (NCCLS Document EP7-P). No specific findings regarding the impact of interferences on performance are detailed in this summary, beyond stating they were evaluated.
    ComplianceComplies with EN 50081-1, FCC Class B, EN 50081-2, and IEC 1010-1.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not explicitly stated with a specific number. The text mentions "Specimens analyzed in these tests were remnant from patient specimens of both whole blood and serum collected for routing analysis on existing instrumentation." This implies a varied and likely significant number of patient samples.
    • Data Provenance: The data is described as "remnant from patient specimens... collected for routing analysis on existing instrumentation." This indicates the data is retrospective and derived from actual patient samples. The country of origin of the data is not explicitly stated, but the submitter's address is Roswell, GA, USA, suggesting the studies were likely conducted in the US.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Number of Experts: Not explicitly stated.
    • Qualifications of Experts: Not explicitly stated. The ground truth was established by "predicate devices" and "reference materials or methods" (for linearity). The clinical testing was performed by "personnel trained to perform and report these analyses," implying qualified laboratory or healthcare professionals.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not explicitly mentioned. The comparison was made against predicate devices, with the conclusion being "no significant difference in mean values (P<0.05)." This suggests a statistical comparison rather than an expert adjudication process in the traditional sense of resolving discrepancies between human readers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Was it done?: No, an MRMC comparative effectiveness study, in the typical sense of evaluating human readers' improvement with or without AI assistance, was not done. This device is an automated analyzer, not an AI-assisted diagnostic tool for human interpretation of images or other complex data. The comparison was between the device and predicate devices.

    • Effect size of human readers' improvement with AI vs without AI assistance: Not applicable, as this was not an AI-assisted human reading study.

    6. Standalone (Algorithm Only) Performance Study:

    • Was it done?: Yes, the performance described (precision, linearity, and clinical correlation) is inherently a standalone performance of the AVL OPTI Critical Care Analyzer. The device is an automated system, and its measurements were compared directly against established methods and other devices, not as a component assisting a human in real-time.

    7. Type of Ground Truth Used:

    • Ground Truth Type:
      • For clinical correlation: Measurements from "predicate devices" (NOVA Biomedical STAT 5, Chiron 865, AVL 995, SenDx 100, Diametrics IRMA, I-STAT 200, AVL 9181, IL 943 Flame Photometer). These predicate devices represent established and accepted methods for measuring the analytes.
      • For linearity: "Reference materials or methods" and an "AVL 995 pH/Blood Gas Analyzer standardized to N.I.S.T. traceable pH buffers."
      • For precision: Evaluation against "samples plasma, reduced bovine hemoglobin solution and three levels of aqueous quality control solution."

    8. Sample Size for the Training Set:

    • Sample Size for Training Set: The document does not describe a "training set" in the context of a machine learning or AI model, as this device predates widespread AI development for medical devices. The device's calibration refers to internal calibration mechanisms within the disposable cassette, not a data-driven training process in the modern AI sense. "Training" for this device would refer to its engineering and development processes rather than a data subset for model training.

    9. How the Ground Truth for the Training Set Was Established:

    • How Ground Truth for Training Set Was Established: Not applicable as there is no mention of a "training set" in the AI/machine learning context. The calibration of the device is done via the disposable cassette, which "contains all the elements needed for calibration," suggesting pre-programmed or established calibration parameters.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1