    Why did this record match?
    Reference Devices :

    K931149/A, K933886, K942096/S1, K942097/S1

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sysmex® CA-1000 & CA-5000 are intended for use as automated blood plasma coagulation analyzers.

    Device Description

    Automated blood plasma coagulation analyzers. The systems were described in details in premarket notifications, document control numbers: K931149/A, K933886, K942096/S1 and K942097/S1. The devices belong to the same family of instruments and they are equivalent in their technological features and performance.

    AI/ML Overview

    The acceptance criteria and device performance information for the Sysmex® Automated Coagulation Analyzer CA-1000 and CA-5000 are detailed below, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes a comparison between new software versions (24 for CA-1000, 13 for CA-5000) and previous software versions or predicate devices. The acceptance criteria are implicitly good correlation and agreement between the new and predicate devices/software, indicated by high correlation coefficients and regression equations close to Y=X.

    TestAcceptance Criteria (implicit)Reported Device Performance (r)Regression Equation
    Prothrombin Time (PT), secondsHigh correlation (e.g., r > 0.95)0.999Y = 0.97X + 0.1
    Activated Partial Thromboplastin Time (APTT), secondsHigh correlation (e.g., r > 0.95)0.978Y = 0.99X + 0.7
    Fibrinogen (Clauss), mg/dLHigh correlation (e.g., r > 0.95)0.995Y = 0.96X + 10.4
    Derived FibrinogenHigh correlation (e.g., r > 0.95)0.951Y = 1.02X - 9.8
    Factor VII AssayHigh correlation (e.g., r > 0.95)0.996Y = 0.95X + 1.0
    Factor VIII AssayHigh correlation (e.g., r > 0.95)0.995Y = 0.92X + 0.6
    Thrombin TimeHigh correlation (e.g., r > 0.95)0.951Y = 1.00X - 0.3

    The acceptance criteria are inferred based on the strong correlations and regression equations presented, indicating that the new software versions perform comparably to the established predicate devices/software.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The sample sizes for each test ranged from 19 to 49 samples, as reported in the "Sample Number (n)" column of the table. Specifically:
      • PT: 39 samples
      • APTT: 37 samples
      • Fibrinogen (Clauss): 41 samples
      • Derived Fibrinogen: 19 samples
      • Factor VII Assay: 49 samples
      • Factor VIII Assay: 38 samples
      • Thrombin Time: 42 samples
    • Data Provenance: The samples consisted of "plasma samples with each representative analyte of the core coagulation assays." This group "represented approximately even numbers of males and females, consisted of approximately 40 samples." The study evaluated specimens from "apparently healthy individuals and from patients with different pathological conditions." The document does not specify the country of origin or whether the data was retrospective or prospective, only that it was a "clinical study."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This type of device is an automated coagulation analyzer. Ground truth is established by comparing the results of the new device/software against the results of a predicate device or established laboratory methods. Therefore, the concept of "experts" establishing ground truth in the traditional sense (e.g., radiologists interpreting images) is not directly applicable. The "ground truth" here is the measurement obtained by the predicate device or a recognized standard method.

    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are typically used for subjective assessments where multiple human interpreters might disagree. For an automated coagulation analyzer, the "ground truth" (or reference standard) is typically derived from laboratory measurements using an established method or a predicate device. Therefore, no multi-expert adjudication method is relevant or described for this type of test.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This study focuses on the performance comparison between the proposed device/software and a predicate device/software, not on the improvement of human readers with AI assistance. The device is an automated analyzer, not an AI-assisted diagnostic tool for human interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, this study represents a standalone (algorithm only) performance evaluation. The coagulation analyzer is an automated system, and the study compares its measurements generated by different software versions to each other, without human interpretation being part of the primary measurement process.

    7. The Type of Ground Truth Used

    The ground truth for this comparison was the measurements obtained from the predicate device or the current software versions of the Sysmex® Automated Coagulation Analyzers. The studies were "method comparison evaluations" comparing "Software Version 21 versus Software Version 24" for the CA-1000 and CA-5000 systems. This indicates that the results of the established, previously cleared software/device served as the reference for evaluating the new software/device.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" for the software. This is common for traditional laboratory instruments where the software's algorithms are based on established biochemical principles and calibrated during development rather than "trained" on a large dataset in the machine learning sense. The clinical study described evaluates the performance of the already developed software.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, a "training set" in the context of machine learning is not applicable here. The algorithms within the coagulation analyzer software are developed based on known coagulation assay principles and calibrated using standard reference materials and methods, not by establishing ground truth from a large, labeled training dataset in the same way an AI model would be.