510(k) Data Aggregation

    K Number
    K090180

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-06-17

    (145 days)

    Product Code
    Regulation Number
    870.2360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K911529

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tender-Trode® Prewired electrode is a disposable, noninvasive electrode that is intended to be used for resting ECG monitoring on neonatal and pediatric patients in a hospital setting. These electrodes can be used for short-term or long-term applications, up to 24 hours. The Tender-Trode of prewired electrodes are single-use, non-sterile, disposable and are to be used on intact (uninjured) skin.

    Device Description

    A conductive solid gel electrode consisting of a conductive adhesive gel, a silver/silver chloride plated ABS eyelet, an adhesive cloth substrate, vinyl label, 24" wire w/.060 dia. female socket and polystyrene release liner.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Tender-Trode® Prewired ECG Electrodes, addressing your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Standard)Reported Device Performance
    PerformanceEC12:2000 for Disposable ECG ElectrodesMeets or exceeds the performance specifications established in EC12:2000. Test results indicate conformance.
    LabelingEC12:2000 AAMI standard for labelingMeets EC12:2000 AAMI standard for labeling.
    Shelf LifeEC12:2000 AAMI standard for shelf lifeData obtained in accelerated shelf life studies substantiates claimed shelf life.
    PackagingEC12:2000 AAMI standard for packagingMeets EC12:2000 AAMI standard for packaging.
    SafetyEC12:2000 AAMI standard for safetyMeets EC12:2000 AAMI standard for safety.
    BiocompatibilityTripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact (Cytotoxicity, Sensitization, Primary Skin Irritation)Materials were found to be non-irritating, non-cytotoxic, and non-sensitizing.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify the exact sample size used for the test set (i.e., the number of electrodes tested for performance, shelf life, or biocompatibility). It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective) for these tests, beyond indicating that accelerated shelf life studies were performed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable and not provided in the document. The "ground truth" concept, especially involving expert consensus, is typically relevant for interpretative devices like those using AI in medical imaging. For a physical device like an ECG electrode, ground truth is established through adherence to engineering standards and validated laboratory testing, not expert interpretation of results.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided in the document. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies where multiple human readers or algorithms interpret data to reach a consensus, often in diagnostic imaging. For a physical device demonstrating adherence to standards, the "adjudication" is inherent in the standardized testing methodologies themselves and the objective measurements obtained.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done and is not described in this document. This type of study is specifically designed to compare the diagnostic performance of human readers, often with and without AI assistance, across multiple cases. The Tender-Trode® is a physical medical device (an ECG electrode), not a diagnostic AI system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm only) performance study was not done. This concept applies to AI algorithms. The device being reviewed is a physical electrode. Its performance is assessed through its physical and electrical characteristics as determined by standardized tests, not by an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the Tender-Trode® Prewired ECG Electrodes is established by adherence to internationally recognized consensus standards, specifically EC12:2000 for Disposable ECG Electrodes and ISO 10993-1 for Biocompatibility. For biocompatibility, this ground truth is further supported by specific biological tests (Cytotoxicity, Sensitization, and Primary Skin Irritation) with objective endpoints.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. The concept of a "training set" refers to data used to train machine learning models. The Tender-Trode® is a physical medical device, not an AI algorithm, and therefore does not have a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and not provided. As stated above, there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033010

    Validate with FDA (Live)

    Date Cleared
    2003-10-10

    (14 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K030489, K960732, K911529, K000206

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brainz Instruments Ltd BRM2 Brain Monitor is an Electroencephalograph as per 21 CFR §882.1400 (a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head).

    The BRM2 Brain Monitor is intended to monitor the state of the brain by acquisition of electroencephalogram (EEG) signals, in the intensive care unit, operating room, and for clinical research.

    Device Description

    The BRM2 Brain Monitor is a two-channel EEG device, consisting of skin electrodes, Sensor Lead, Acquisition Unit, Isolation Unit, Serial and Power Cables, Monitor, and Roll-Pole. All components other than the skin electrodes and Sensor Lead are the same as described in 510(k) K030489 for the BRM2 Brain Monitor.

    Changes to the BRM2 Brain Monitor consist of using the Neonatal Sensor Set for the EEG skin electrodes, which has two signal electrode pairs and a reference electrode all pre-wired to a single connector. The Sensor Lead has been modified to add a compatible connector for the Neonatal Sensor Set. Products used for skin preparation at the EEG electrode application sites have been made available in a convenience kit.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that indicates the device meets them, based on the provided text:

    Important Note: The provided document is a 510(k) summary for a medical device (BRM2 Brain Monitor) seeking substantial equivalence. As such, the level of detail regarding acceptance criteria, study methodologies, and exact performance metrics is often less comprehensive than what would be found in a full scientific publication or clinical trial report. The information is focused on demonstrating equivalence to predicate devices, rather than establishing de novo performance.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria / Standards MetReported Device Performance/Conclusion
    Electrical SafetyIEC 60601-1 (general safety international standard)Device (BRM2 Brain Monitor and Neonatal Sensor Set) meets the requirements.
    UL 2601-1 (relevant USA deviations for general safety)Device meets the requirements.
    IEC 60601-2-26 (particular requirements for electroencephalographs)Device meets the requirements.
    Biocompatibility(Not explicitly named, but implied for device components)Non-clinical testing to cover biocompatibility was carried out. (Conclusion: meets design and performance requirements)
    Product Shelf Life(Not explicitly named)Non-clinical testing to cover product shelf life was carried out. (Conclusion: meets design and performance requirements)
    Electrode PerformanceAAMI EC12 (standard for disposable ECG electrodes)Neonatal Sensor Set meets the requirements.
    FDA / IEC requirements for safety of electrode lead wiresNeonatal Sensor Set meets the requirements.
    Clinical EffectivenessAcceptable performance of electrodes in general use and over the 24-hour maximum recommended usage period for the Neonatal Sensor Set."Clinical testing was carried out to demonstrate the effective duration of use for the Neonatal Sensor Set. This established acceptable performance in general use of the electrodes and in particular over the 24-hour maximum recommended usage period."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not explicitly state the sample size used for the clinical testing of the Neonatal Sensor Set.
    • Data Provenance: The document does not specify the country of origin for the clinical testing data. It mentions "Clinical testing was carried out," implying it was a prospective study, but doesn't explicitly state "retrospective" or "prospective."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This information is not provided in the document. The clinical test was to demonstrate the effective duration of use for the Neonatal Sensor Set, implying observation of electrode function and patency rather than a diagnostic accuracy study requiring expert adjudication of ground truth for EEG interpretation.

    4. Adjudication Method for the Test Set

    • This information is not provided as the nature of the clinical test described (electrode duration of use) does not typically involve an adjudication method in the way a diagnostic accuracy study would.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. The document describes non-clinical and clinical testing focused on the device's technical performance and electrode efficacy, not a comparison of human reader performance with and without AI assistance. The device itself is an electroencephalograph, a signal acquisition device, not an AI-powered diagnostic interpretation tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    • Not applicable / No. This device is an electroencephalograph, which is a hardware device for measuring and recording brain activity. It does not contain an AI algorithm for standalone performance evaluation that would be distinct from human interpretation. The "performance" being evaluated relates to the quality of the signal acquisition and the durability of the electrodes.

    7. The Type of Ground Truth Used

    • For the electrical safety, biocompatibility, and shelf-life testing, the "ground truth" is adherence to established international and national standards (e.g., IEC 60601-1, UL 2601-1, AAMI EC12).
    • For the clinical testing of the Neonatal Sensor Set's effective duration of use, the "ground truth" would likely be direct observation of electrode integrity, signal quality over time, and possibly absence of adverse skin reactions (though not explicitly detailed beyond "acceptable performance"). It relies on objective measurements and clinical assessment, not expert consensus on diagnostic interpretation.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a hardware system for signal acquisition. It does not use a "training set" in the context of machine learning or AI algorithms. The design and validation are based on engineering principles and regulatory standards.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable for the reasons stated in point 8.