    Why did this record match?
    Reference Devices :

    K883780, K982902

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For Spinal Cord Stimulation: This system is indicated as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or limbs, including unilateral or bilateral pain. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.

    For Peripheral Nerve Stimulation: This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region. The trial devices are solely used for trial stimulation (no longer than 30 day) to determine efficacy before recommendation for a permanent (long term) device.

    Device Description

    The Nalu Neurostimulation System has been cleared by the FDA for spinal cord stimulation, and peripheral nerve stimulation (K201618), to provide therapeutic relief for chronic, intractable pain of the trunk and/or limbs including unilateral, bilateral nerve pain. The Nalu Neurostimulation therapy utilizes pulsed electrical current to create an energy field that acts on nerves in the spinal cord or peripheral nerve to inhibit the transmission of pain signals to the brain. The Nalu System is implanted only following a successful trial period using the Nalu Neurostimulation trial system. The Nalu Neurostimulation System consists of several components. The implantable pulse generator (IPG) provides electrical stimulation pulses that are transmitted through the leads, to the desired location, either on the spinal cord or peripheral nerve site. The leads are implantable and designed to deliver electrical pulses to the nerves via an array of four or eight cylindrical electrodes at the distal end. The Trial Therapy Disc or the Therapy Disc houses the battery and electronics for RF power and controls the IPG for therapy delivery via the remote programmer. Implantation of the Nalu IPG and lead components for spinal cord stimulation (SCS) or peripheral nerve stimulation (PNS) is performed via standard surgical tools and techniques, as described in (K201618).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria or performance metrics in the way one might expect for a diagnostic or AI-based device's clinical performance. Instead, the submission describes the Nalu Neurostimulation System and its substantial equivalence to predicate devices, focusing on technical specifications and safety rather than a measurable clinical outcome in a performance study.

    The closest to "acceptance criteria" are the parameters defining the device's operational range, and the "reported device performance" is that these parameters are within acceptable limits compared to predicate devices, demonstrating substantial equivalence.

    Acceptance Criteria (Implied)Reported Device Performance
    Pulse Width RangeSubject Device: 12 to 2000 µs (Predicate: 12 to 1000 µs; Reference: 50 to 1000 µs) - Analysis: Available programmable pulse width will be capped to maintain Maximum Phase Charge and Maximum Charge Density within limits of Medtronic Xtrel reference device limit. No impact to safety and effectiveness.
    Maximum Phase Charge (300 Ohms)Subject Device: 18.0 µC/pulse (Predicate: 10.2 µC/pulse; Reference: 18.0 µC/pulse) - Analysis: Same as Medtronic Xtrel reference device.
    Maximum Phase Charge (500 Ohms)Subject Device: 18.0 µC/pulse (Predicate: 10.2 µC/pulse; Reference: 14.2 µC/pulse) - Analysis: Same as Medtronic Xtrel reference device at 300 Ohms. Maximum Phase Charge constant in current controlled system and enforced below maximum reference device value (at 300 Ohms). No impact to safety and effectiveness.
    Maximum Phase Charge (800 Ohms)Subject Device: 18.0 µC/pulse (Predicate: 10.2 µC/pulse; Reference: 10.5 µC/pulse) - Analysis: Same as Medtronic Xtrel reference device at 300 Ohms. Maximum Phase Charge constant in current controlled system and enforced below maximum reference device value (at 300 Ohms). No impact to safety and effectiveness.
    Maximum Charge Density (300 Ohm)Subject Device: 146.94 µC/cm² (Predicate: 83.3 µC/cm²; Reference: 150.0 µC/cm²) - Analysis: Within maximum limit as set by Medtronic Xtrel reference device.
    Maximum Charge Density (500 Ohm)Subject Device: 146.94 µC/cm² (Predicate: 83.3 µC/cm²; Reference: 118.3 µC/cm²) - Analysis: Same as Medtronic Xtrel reference device at 300 Ohms. Maximum Phase Charge constant in current controlled system and enforced below maximum reference device value (at 300 Ohms). No impact to safety and effectiveness.
    Maximum Charge Density (800 Ohm)Subject Device: 146.94 µC/cm² (Predicate: 83.3 µC/cm²; Reference: 87.5 µC/cm²) - Analysis: Same as Medtronic Xtrel reference device at 300 Ohms. Maximum Phase Charge constant in current controlled system and enforced below maximum reference device value (at 300 Ohms). No impact to safety and effectiveness.
    Dosage Time RangeSubject Device: On spans 1 ms to 1000 ms, Off spans 1 ms to 2000 ms (Predicate: On spans 1 ms to 25 ms) - Analysis: This parameter has no impact on safety and effectiveness. Increasing the range allows the clinician to offer more flexibility to accommodate patient preferences.
    All other parameters (Frequency, Current/Voltage Regulated, Output Voltage/Current, Waveform, Pulse Shape, Maximum Current Density, Net Charge, Average Phase Power/Density, Pulse Delivery Mode, Current Path Options, Program Cycle, Pulse Pattern, Daily Therapy Time, Transmit Frequency)All "Same as predicate." or "Same as predicate/reference." indicating compliance with established safe and effective parameters.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a clinical performance study with a "test set" from real patient data in the context of an AI/diagnostic device. Instead, it refers to "testing" performed to support the safety and performance of the device. This testing appears to be primarily non-clinical (bench/laboratory) testing of electrical and software parameters.

    Therefore, there is no mention of a sample size for a test set of patient data, nor data provenance (country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable and not provided. The study focuses on evaluating technical parameters against established medical device safety standards and comparison to predicate devices, not on a ground truth established by medical experts for diagnostic accuracy.

    4. Adjudication Method

    This information is not applicable and not provided, as the submission does not involve expert review or adjudication of clinical cases for diagnostic or performance accuracy.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This device is a neurostimulation system for pain relief, not a diagnostic imaging or AI-assisted diagnostic tool that would typically undergo MRMC studies to assess human reader improvement. The submission focuses on demonstrating substantial equivalence in technical characteristics and safety.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable. The "device" itself (Nalu Neurostimulation System) is a therapeutic electrical stimulator. The "software" changes described are modifications to its operational parameters. The performance testing referenced is likely to evaluate the hardware and software functionality as an integrated system, ensuring it operates within safe and effective limits. There isn't an "algorithm only" component in the sense of an AI interpreting inputs to provide diagnostic outputs.

    7. The type of ground truth used

    The "ground truth" for this submission appears to be regulatory and engineering standards, and the specifications and performance characteristics of legally marketed predicate and reference devices. The evaluation focuses on whether the subject device's technical specifications and functionality (particularly the updated software parameters) fall within the established safe and effective ranges of these comparable devices.

    8. The sample size for the training set

    This information is not applicable and not provided. The device is not an AI/ML model that would train on a dataset. The software changes are operational programming changes for the neurostimulator.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided, as there is no "training set" for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191435

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-09-06

    (99 days)

    Product Code
    Regulation Number
    882.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K152178, K171366, K934065, K883780

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region.

    The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.

    Device Description

    This submission will add 4 stimulation contact options to the predicate Nalu Neurostimulation System (also referred to as the "Nalu PNS System"). The Nalu PNS System is used for peripheral nerve stimulation to provide therapeutic relief for chronic, intractable pain of peripheral nerve origin. The Nalu PNS System incorporates a miniature implanted neurostimulator, powered by an externally worn Therapy Disc device. The Nalu Neurostimulation therapy utilizes pulsed electrical current to create an energy field that acts on the peripheral nerves to inhibit the transmission of pain signals to the brain. The Nalu PNS System may be implanted following a successful trial period using the Nalu PNS trial system.

    The leads that were cleared with the Nalu PNS System featured 8 stimulation contacts. This submission will provide an optional use of leads with 4 stimulation contacts.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Nalu Neurostimulation System for PNS" (specifically, the 4-contact version). The purpose of a 510(k) is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This document does NOT present a study that proves the device meets specific acceptance criteria through clinical performance data. Instead, it argues for substantial equivalence based on non-clinical performance testing because the changes are considered minor.

    Therefore, many of the requested items related to clinical studies, ground truth establishment, expert adjudication, and MRMC studies are not applicable to this submission. The "acceptance criteria" here are demonstrating substantial equivalence through non-clinical testing.

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The core "acceptance criteria" for this 510(k) submission revolve around demonstrating substantial equivalence to a predicate device, primarily the Nalu Neurostimulation System (K183579). This is achieved by showing that the new device (4 Contact PNS System) has the same intended use, similar technological characteristics, and that any differences do not raise new questions of safety or effectiveness.

    The document demonstrates this through comparative tables and statements rather than a traditional pass/fail clinical study.

    Acceptance Criterion (Implicit for Substantial Equivalence)Reported Device Performance (4 Contact PNS System)
    Intended Use: Same as predicate.Same: Indicated for pain management in adults with severe intractable chronic pain of peripheral nerve origin. Not for craniofacial region. Trial devices for stimulation (max 30 days).
    Technological Characteristics: Similar to predicate, with differences not affecting safety/effectiveness.Differences: - Number of Electrodes: 4 (vs. 8 in primary predicate). - Lead Length: 25 cm, 40 cm (vs. 40 cm, 60 cm in primary predicate). - Electrode Array Length: 21 mm (vs. 52 mm in primary predicate). - Electrode Spacing: 3.0 mm (vs. 4.0 mm in primary predicate). - Cable Features: Coiled Wires (vs. Multilumen tube in primary predicate). - Lead Anchor: Integrated Lead Tines (vs. Separate molded silicone anchor with Ti locking mechanism in primary predicate). Performance: All differences are stated to "not affect safety and effectiveness of intended use" and are within parameters of other reference devices. Tested through non-clinical means.
    Therapeutic Attributes: Same as predicate.Same: All parameters like pulse frequency, pulse width, current/voltage regulation, output current/voltage, waveform, pulse shape, maximum phase charge/density, net charge, average phase power/density, pulse delivery mode, current path options, software level of concern, program cycle, pulse pattern, dosage time, transmit frequency are identical to the predicate (K183579).
    Biocompatibility: Meet standards.Met: Same biocompatibility reports as predicate supported this device. Materials and processes previously assessed.
    Sterilization: Meet standards.Met: Ethylene Oxide sterilization, same as predicate.
    Electrical & Mechanical Performance: Meet specifications, safe, and effective.Met: Verification testing included electrical and mechanical tests. Validation, performance, and usability testing demonstrated device meets user needs. Conforms to applicable standards (ISO 14708-1, ISO 14708-3, IEC 62366-1, EN ISO 14971, ISO 11607, ISO 11135-1, CISPR 11).
    Software Level of Concern: Moderate.Met: Same as predicate (moderate).
    Human Factors and Usability: Tested and met.Met: Testing performed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable. This submission relies on non-clinical performance testing (bench testing, verification, validation, biocompatibility) and a comparative analysis to a predicate device, as opposed to a clinical test set with a specific sample size of patients/cases.
    • Data Provenance: Not applicable. There is no clinical data from patients/cases mentioned. The testing is internal (Nalu Medical) and presumably took place in the U.S.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Not applicable. As there is no clinical test set requiring ground truth establishment by experts (e.g., for disease diagnosis or outcome evaluation), this information is not provided nor needed for this type of submission.

    4. Adjudication Method for the Test Set

    • Not applicable. No clinical test set or human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an implanted neurostimulator, not an AI-assisted diagnostic imaging device for human readers. No MRMC study was conducted or is relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device, not a standalone algorithm.

    7. The Type of Ground Truth Used

    • For Non-Clinical Testing: The "ground truth" implicitly used for non-clinical testing (electrical, mechanical, biocompatibility, sterilization) is based on established engineering principles, international standards (e.g., ISO, IEC, CISPR), and pre-defined specifications/tolerances for the device's performance. For biocompatibility, it's compliance with ISO 10993-1. For substantial equivalence, the "ground truth" is the established safety and effectiveness of the legally marketed predicate device.

    8. The Sample Size for the Training Set

    • Not applicable. There is no "training set" as this is not a machine learning/AI device.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K152178, K934065, K883780, K000852

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region.

    The trial devices are solely used for trial stimulation (no longer thank 30 days) to determine efficacy before recommendation for a permanent (long term) device.

    Device Description

    The Nalu Neurostimulation System (also referred to as the "Nalu System") is used for peripheral nerve stimulation to provide therapeutic relief for chronic, intractable pain of peripheral nerve origin. The Nalu Neurostimulation system incorporates a miniature implanted neurostimulator, powered by an externally worn Therapy Disc device. Similar to the predicate StimQ, the Nalu Neurostimulation therapy utilizes pulsed electrical current to create an energy field that acts on the peripheral nerves to inhibit the transmission of pain signals to the brain. The Nalu System may be implanted following a successful trial period using the Nalu Neurostimulation trial system.

    The Nalu Neurostimulation System is comprised of 5 elements:

    1. Nalu Implantable Pulse Generator: The implantable pulse generator (IPG) provides electrical stimulation pulses that are transmitted through the leads to the desired peripheral nerve. The IPG is available in two different implant architectures: an “integrated" system with pre-attached leads and a "ported" system where leads may be attached, via connector ports. In addition, both of these versions are available in single or dual lead configurations. The hermetic IPG housing includes a ceramic enclosure and a feedthrough connected internally to a printed circuit board assembly. Wires leaving the IPG are encapsulated in polyurethane and a silicone over mold forms the final biocompatible surface of the IPG for direct patient tissue contact.
    2. Leads: Leads are implantable and are designed to deliver electrical pulses to the peripheral nerve via an array of eight cylindrical electrodes at the distal end. Leads may be integrated with or connected to the IPG. Both Trial and Permanent Implant leads are available for use. The leads use polyurethane insulation with Pt/Ir electrodes. The leads may be secured in place with the Nalu Lead Anchor.
    3. Surgical and Trial Tools: Implantation of the Nalu IPG and lead components for Peripheral Nerve Stimulation (PNS) is performed via standard PNS surgical techniques. The desired implant location is accessed via needle placement, followed by lead placement through an introducer. The leads are anchored and the IPG is placed in a subcutaneous pocket. Patient contacting materials include medical grade stainless steel, thermoplastic elastomers, ABS, silicone, and Urethane.
    4. Externally worn Therapy Discs: Two types of Therapy Disc are available. One is to be used during the trial phase (Trial Therapy Disc), and one is to be used after permanent IPG implantation (Therapy Disc). Both devices are worn by the patient using one of the Nalu-provided options The Therapy Discs house a rechargeable lithium ion battery, and electronics including a microcontroller running software for therapy control, patient interaction and communication with Nalu's Clinician Programmer and Remote Control devices. The Therapy Disc used to power and command the implant does so wirelessly using Radio Frequency (RF) and is held in place by an adhesive clip applied to the skin or a belt/cuff worn over clothing.
    5. Clinician Programmer and Remote Control: A Clinician Programmer Application is provided to configure the Trial Therapy Disc and Therapy Disc devices during surgery and programming. A Patient Remote Control Application is available to provide the patient with a convenient secondary option to control their system in addition to built-in controls on the Therapy Disc. The Clinician Programming Application runs on an Android tablet and communicates over a secure Bluetooth Low Energy link with the Trial Therapy Disc and Therapy Disc devices. The programmer is responsible for configuring the devices to deliver therapy according to clinician defined levels and patient preferences, and for managing patient and session records. The Patient Remote Control Application runs on iOS and Android platforms and offers basic control of the Trial Therapy Disc and Therapy Disc through a secure Bluetooth Low Energy link. The controls include selecting between clinician-defined therapy options (programs), turning stimulation on and off, and managing alerts.
    AI/ML Overview

    The provided text is a 510(k) premarket notification for the Nalu Neurostimulation System for Peripheral Nerve Stimulation. It describes the device, its intended use, and a comparison to predicate devices to demonstrate substantial equivalence. However, it explicitly states that no clinical performance data was deemed necessary for this submission.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from a clinical study, nor details about sample sizes, expert ground truth establishment, adjudication, MRMC studies, or standalone algorithm performance, as these were not part of the documented submission process for this device.

    The section "5.9. Clinical Performance Data" clearly states:
    "Nalu Medical determined that bench and non-clinical testing are sufficient to demonstrate that the Nalu Neurostimulation System is as safe and effective as the predicate device. Note that the predicate device did not need clinical evidence to obtain a determination of substantial equivalence."

    Based on the provided document, the device's acceptance was based on non-clinical performance (bench testing, animal studies, and compliance with standards).

    Here's a breakdown of what can be extracted from the document regarding the device's safety and effectiveness without clinical data:

    1. A table of acceptance criteria and the reported device performance:

    As no clinical study was performed for this submission, there are no "acceptance criteria" related to a clinical performance study with human subjects, nor "reported device performance" in terms of clinical outcomes. The device's performance was evaluated through non-clinical testing to demonstrate substantial equivalence to predicate devices.

    Table: Performance Evaluation based on Non-Clinical Testing and Substantial Equivalence

    Acceptance Criteria Category (based on Non-Clinical Testing)Reported Device Performance (Nalu Neurostimulation System)
    Functional and Electrical PerformanceVerified to meet target specifications over a range of operating and storage conditions through electrical testing.
    Mechanical PerformanceVerified to meet target specifications over a range of operating and storage conditions through mechanical testing.
    Software Verification & ValidationDemonstrated to perform as intended in the specified use conditions.
    BiocompatibilityDemonstrated to be biocompatible based on testing (cytotoxicity, sensitization, irritation/intracutaneous reactivity, systemic toxicity, implant studies, chemical characterization) according to ISO 10993-1:2009 and FDA guidance. This included testing for implant devices (IPG, Leads, Anchor, Extension: permanent contact), externally communicating devices (Needles, Sheaths, surgical tools: limited contact), and surface devices (Therapy Discs, adhesive clip: permanent intact skin contact).
    Sterilization ValidationValidated using Ethylene Oxide (ISO 11135-1:2014).
    Human Factors & UsabilityTesting performed on the device to ensure it met user needs as reflected in the functional specification, adhering to IEC 60601-1-6:2010+A1:2013 and IEC 62366-1:2015, and FDA guidance.
    Animal Study Findings (Pre-clinical Validation)Six (6) Nalu Neurostimulation IPGs and Lead systems implanted in a porcine model for 90 days. All devices performed as expected without incident. No device- or procedure-related complications or premature deaths. Data collected at 30, 60, and 90-day intervals. This study covered surgical usability, RF communication and stimulation, implanted device stability, and tissue response.
    Compliance with StandardsCompliance demonstrated with relevant standards (e.g., ISO 14708-1:2014, ISO 14708-3:2017, IEC 60601-1:2005:A2012, IEC 60601-1-11:2015, IEC 60601-1-2:2014, IEC 62304:2015, ISO 14971:2012/2007, ISO 11607-1/2:2006, CISPR 11, and FDA guidance on Cybersecurity).
    Substantial Equivalence to PredicateAll physical and therapeutic attributes are within or equivalent to the parameters of the StimQ Peripheral Nerve Stimulator (PNS) System (K171366) and other reference devices (Medtronic Mattrix, Medtronic Xtrel, ANS Renew). Differences in lead length, diameter, electrode array length, number of electrodes, electrode surface area, lead extension, anchor, configurations, software platforms, and transmit frequency were deemed not to affect safety and effectiveness of the intended use, based on engineering analysis and comparison to a range of previously cleared devices.

    2. Sample size used for the test set and the data provenance:

    • Test Set (Non-clinical):
      • Animal Study: 6 Nalu Neurostimulation IPGs and Lead systems were implanted. Data provenance is a pre-clinical study conducted by Nalu Medical (presumably in the USA, as it's an FDA submission for a US company). It's a prospective study.
      • Bench Testing: Quantities of devices or components used for electrical, mechanical, biocompatibility, sterilization, and human factors testing are not specified in numerical terms within this document. The provenance is internal testing by Nalu Medical.
    • Training Set (Not applicable for this submission method): No training set data is referenced, as this is a 510(k) based on substantial equivalence to predicates, not a de novo clearance requiring clinical studies or algorithmic training data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for clinical ground truth. For the animal study, the "ground truth" was the observed performance of the device in a living model and tissue response, assessed by the study investigators. The qualifications of these experts are not detailed in the document.
    • For engineering and performance testing, the ground truth is established by the design specifications, validated test methods, and industry standards, implemented and evaluated by qualified engineers and testers.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable for clinical studies. For the non-clinical and animal studies, methods like peer review of study protocols, data analysis, and report generation would be in place, but a "2+1" or "3+1" adjudication method typically refers to radiologist consensus in image-based AI studies, which is not relevant here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study would be relevant for AI-powered diagnostic devices, which is not the case for this neurostimulation system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a neurostimulation system, not an AI algorithm for diagnosis or interpretation. Its software controls the device's function, and its performance is evaluated in the context of device operation, not as a standalone diagnostic tool. The software was subject to verification and validation tests as per IEC 62304.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the animal study: The ground truth was based on direct observation of device performance, surgical usability, stability, tissue response, and lack of complications in the porcine model.
    • For bench testing: Ground truth was based on established engineering specifications, validated test methods, and compliance with national and international standards.

    8. The sample size for the training set:

    • Not applicable. This submission is based on substantial equivalence and non-clinical data, not on an algorithm trained with a specific dataset.

    9. How the ground truth for the training set was established:

    • Not applicable. (See point 8).
    Ask a Question

    Ask a specific question about this device

    K Number
    K183047

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-22

    (140 days)

    Product Code
    Regulation Number
    882.5880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K141399, K934065, K883780, K000852

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nalu Neurostimulation System is indicated as the sole mitigating agent or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or limbs, including unilateral or bilateral pain.

    The trial devices are solely used for trial stimulation (no longer thank 30 days) to determine efficacy before recommendation for a permanent (long term) device.

    Device Description

    The Nalu Neurostimulation system (also referred to as the "Nalu System") is used for spinal cord stimulation to provide therapeutic relief for chronic, intractable pain of the trunk and/or limbs including unilateral or bilateral pain. The Nalu Neurostimulation system incorporates a miniature implanted neurostimulator, powered by an externally worn device. Similar to the predicate Stimwave system, the Nalu Neurostimulation therapy utilizes pulsed electrical current to create an energy field that acts on nerves in the spinal cord to inhibit the transmission of pain signals to the brain. The Nalu System is implanted only following a successful trial period using the Nalu Neurostimulation trial system.

    The Nalu Neurostimulation System is comprised of 5 elements: Nalu Implantable Pulse Generator, Leads, Surgical and Trial Tools, Externally worn Therapy Discs, and Programmer, Remote.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Nalu Neurostimulation System, indicating that clinical performance data was not required for substantial equivalence. Therefore, there is no study described in this document that proves the device meets specific acceptance criteria based on human clinical data.

    The acceptance criteria described in the document relate to non-clinical performance testing (bench testing, biocompatibility, and animal testing) to demonstrate safety and effectiveness relative to a predicate device.

    Here's an breakdown of the information requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Since no clinical acceptance criteria or performance metrics were provided in the document due to the nature of a 510(k) without clinical studies, I will summarize the general approach to substantial equivalence and the scope of non-clinical testing.

    General Acceptance Criteria (Implied for 510(k) clearance): The Nalu Neurostimulation System is substantially equivalent to the predicate device (Stimwave Freedom SCS System K170141) in terms of:

    • Intended Use
    • Technological Characteristics
    • Principles of Operation

    Reported Device Performance (Non-Clinical):

    Acceptance Criteria CategoryReported Device PerformanceJustification/Outcome
    Substantial Equivalence (Overall)Nalu Neurostimulation System vs. Stimwave Freedom SCS System (K170141)No significant differences in physical and therapeutic attributes that would raise new questions of safety or effectiveness. Differences exist but do not alter intended therapeutic use or affect safety/effectiveness.
    BiocompatibilityCompliance with ISO 10993-1:2009Components categorized by contact type (Implant, Externally Communicating, Surface) and duration, and subjected to testing including cytotoxicity, sensitization, irritation/intracutaneous reactivity, systematic toxicity, implant studies, and chemical characterization. Biocompatibility was demonstrated.
    Animal StudyEvaluation in porcine model over 90 daysSix Nalu Neurostimulation IPGs and Lead systems implanted. Evaluated surgical usability, RF communication/stimulation, implanted device stability, and tissue response. All devices performed as expected with no device- or procedure-related complications or premature deaths. Preliminary validation of safety and use.
    Electrical, Mechanical, Software PerformanceVerification TestingDevice met target specifications over a range of operating and storage conditions.
    Usability/Human FactorsValidation and Usability TestingDevice demonstrated to meet user needs as reflected in the functional specification.
    SterilizationEthylene Oxide sterilization validationPerformed to ISO 11135-1:2014 standards.
    Software Life Cycle ProcessesCompliance with IEC 62304:2015Processes followed for medical device software. (Software Level of Concern: Moderate)
    Risk ManagementCompliance with EN ISO 14971:2012, ISO 14971:2007Application of risk management to medical devices.
    PackagingCompliance with ISO 11607-1:2006/Amd 1:2014 and -2:2006/Amd 1:2014Packaging for terminally sterilized medical devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set (Non-Clinical):
      • Animal Study: Six Nalu Neurostimulation IPGs and Lead systems were used in a porcine model. This was a prospective animal study. The specific country of origin for the animal study is not mentioned.
      • Bench Testing: The document does not specify exact sample sizes for each bench test (electrical, mechanical, software), but it implies sufficient testing was conducted to meet relevant international standards and guidance documents. This is typically prospective testing conducted in a laboratory setting.
      • Biocompatibility Testing: The number of samples for biocompatibility testing is not specified but would follow established protocols for ISO 10993. This is typically prospective testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable for clinical studies. For the non-clinical tests described:
      • Animal Study: The study was conducted by researchers/veterinary staff, but the "ground truth" was established by direct observation and measurement of device performance, tissue response, and lack of complications in the porcine model. No external experts for ground truth establishment were specifically mentioned in this context.
      • Bench/Biocompatibility/Software Testing: Ground truth is established by engineering specifications, validated test methods, and compliance with recognized standards. Expertise would come from internal design and quality engineers, and potentially third-party testing laboratories, but not in the "expert medical consensus" sense.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable for clinical studies. For the non-clinical tests described, adjudication methods like 2+1 or 3+1 are not used. Results are based on objective measurements and compliance with predefined specifications and standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. The document explicitly states: "Nalu Medical determined that bench and non-clinical testing are sufficient to demonstrate that the Nalu Neurostimulation system is as safe and effective as the predicate device. Note that the predicate device did not need clinical evidence to obtain a determination of substantial equivalence." This is a premarket notification (510(k)) that established substantial equivalence based on non-clinical data, not clinical performance or AI assistance for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. The Nalu Neurostimulation System is an implantable medical device for pain relief, not an AI or imaging diagnostic tool. Therefore, standalone algorithm performance is not relevant to this submission.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Non-Clinical Ground Truth:
      • Bench Testing: Engineering specifications, validated test methods, and compliance with national and international standards (e.g., ISO 14708, IEC 60601 series, ISO 11135, CISPR 11).
      • Biocompatibility Testing: Standards-based evaluation (ISO 10993-1) of material interactions with biological systems.
      • Animal Study: Direct observation of surgical usability, RF communication, device stability, and histopathological tissue response in the porcine model, as interpreted by veterinary and scientific experts.

    8. The sample size for the training set

    • Not applicable. This device is hardware for neurostimulation, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. As stated above, this device does not use an AI/ML algorithm with a training set.