Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171844

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-07-18

    (28 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K844067

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Inspire ICE, a ceramic orthodontic bracket, is intended for use during orthodontic treatment of malocclusions. A malocclusion is imperfect positioning of the teeth when the jaws are closed.

    Device Description

    Inspire ICE, is a single-use orthodontic bracket made from monocrystalline alumina. The brackets are intended for use during orthodontic treatment of malocclusions. A malocclusion is imperfect positioning of the teeth when the jaws are closed. After following standard office protocols for tooth preparation, the brackets are bonded to the teeth with an orthodontic adhesive or cement. An archwire is then threaded through the bracket's archwire slot and is held in place with a ligature tie. These ligatures are tightened, or ligated, around the 'wings' of the bracket and over the archwire. To aid in optimal teeth movement, the individual brackets can also be ligated to other brackets or other orthodontic devices. After the teeth have moved into their proper positions, the orthodontic / dental professional will remove the brackets with the Inspire ICE Debonding Pliers. This step is also known as debonding.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Inspire ICE™ aka Inspire ICE™ Clear Brackets, an orthodontic plastic bracket. The document details the device's characteristics and its substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the non-clinical performance data and the comparison to predicate devices, demonstrating the device performs "as safe, as effective, and performs as well as the predicate devices". Specific quantitative acceptance criteria or pass/fail thresholds for each test are not explicitly stated in this summary. The table below summarizes the reported non-clinical performance data categories.

    Test ElementAcceptance Criteria (Implied)Reported Device Performance (Inspire ICE)
    Bond StrengthDemonstrates comparable bond strength to predicate devices and meets relevant standards.Testing results performed. Implied to be acceptable based on the conclusion of substantial equivalence.
    Tie Wing StrengthDemonstrates comparable tie wing strength to predicate devices and meets relevant standards.Testing results performed. Implied to be acceptable based on the conclusion of substantial equivalence.
    Torque StrengthDemonstrates comparable torque strength to predicate devices and meets relevant standards.Testing results performed. Implied to be acceptable based on the conclusion of substantial equivalence.
    BiocompatibilityMeets the requirements of ISO 10993 series for biological evaluation of medical devices.Testing results performed, utilizing relevant ISO 10993 standards. The finished device is considered to have met biological evaluation and risk assessment requirements.
    Stability TestingDemonstrates stability over time under specified conditions.Testing results performed. Implied to be acceptable based on the conclusion of substantial equivalence.
    Material (Alumina)Material composition similar to predicate, ensuring equivalent performance and safety.Made from monocrystalline alumina, identical to predicate and predicate reference materials.
    Physical DimensionsConformity to established ranges/values for orthodontic brackets (In/Out, Torque, Angulation, Rotation).In/Out: .035" - .050" Torque: +17° thru -22° Angulation: Up to +13° Rotation: Up to 4° These are within or comparable to the ranges of predicate devices.
    Manufacturing Method (Grinding)Manufacturing method consistent with predicate, ensuring similar structural integrity.Grinding, identical to predicate and predicate reference manufacturing methods.
    Ligation (Non Self-Ligating)Ligation mechanism consistent with predicate, ensuring proper orthodontic function.Non Self-Ligating, identical to predicate and predicate reference ligation methods.
    Debonding PerformanceCompatible with a specific debonding instrument, ensuring safe removal.Debonding Pliers (REF: 866-4020) are used, similar to the predicate reference and comparable/compatible with predicate device debonding.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the non-clinical performance testing (Bond Strength, Tie Wing Strength, Torque Strength, Biocompatibility, Stability testing). It mentions that "internal test methods" were utilized alongside applicable ISO standards.

    The data provenance is from non-clinical performance data (laboratory testing) conducted by the manufacturer, Ormco Corporation, or accredited research service companies. The country of origin of the data is not explicitly stated, but the submitter (Ormco Corporation) is based in Orange, California, USA, suggesting the testing was likely conducted in the USA or by labs adhering to international standards (ISO). The testing is prospective for the Inspire ICE device, comparing its performance to already marketed predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document describes non-clinical performance testing and a comparison to predicate devices for substantial equivalence. It does not involve a "test set" in the context of human interpretation or diagnostic accuracy where expert ground truth would be established by panels of human experts. Instead, the "ground truth" for material properties and mechanical performance would be derived from:

    • Established international standards (e.g., ISO 27020, ISO 10993 series).
    • Validated laboratory testing methodologies.

    Therefore, the concept of "number of experts" or their qualifications for establishing ground truth as typically understood in studies involving diagnostic accuracy is not directly applicable here. The experts involved would be engineers, material scientists, and toxicologists familiar with the respective ISO standards and testing procedures.

    4. Adjudication Method for the Test Set

    As this study focuses on non-clinical performance data and comparison to predicate devices, there is no "adjudication method" in the context of resolving discrepancies in expert interpretations (e.g., 2+1, 3+1). The performance is assessed against established technical specifications and international standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is an orthodontic bracket, a physical medical device, not a software algorithm that assists human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical orthodontic bracket, not an algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for this submission is based on:

    • Established performance specifications and standards: Primarily ISO 27020 for orthodontics brackets and tubes, and the ISO 10993 series for biological evaluation of medical devices.
    • Validated laboratory testing methodologies: Used to measure Bond Strength, Tie Wing Strength, Torque Strength, Biocompatibility, and Stability.
    • Comparison to legally marketed predicate devices: The "ground truth" for demonstrating safety and effectiveness relies on showing that the new device performs equivalently to previously cleared devices which have an established history of safe and effective use.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical orthodontic bracket and does not involve AI or machine learning models requiring a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set is involved.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1