Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243470

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-12-03

    (25 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K233893

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; Orthopedic surgery for shoulders, knees, foot/ankle, hand/wrist and elbow.

    Device Description

    This product is used for orthopedic surgery, which soft tissue fix such as ligaments, tendons, and capsules to bone. The implant part is made of PEEK (polyether ether ketone, ASTM F2026) with non-absorbable suture and consists of Driver Handle and Driver Shaft for anchor insertion. This product is sterilized product and single use only.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Fix2Lock (PEEK Self Punching)". This submission aims to demonstrate substantial equivalence to a legally marketed predicate device.

    The information provided does not pertain to an AI/ML device or its performance criteria, as indicated by the nature of the acceptance criteria and the listed tests. The device is a physical orthopedic implant. Therefore, I cannot extract information related to AI/ML specific criteria such as sensitivity/specificity, AUC, reader studies, ground truth establishment for AI, or training/test set sample sizes for an AI model.

    The "acceptance criteria" discussed in the document are primarily related to physical and material properties, sterilization, packaging, and shelf life of the bone fixation fastener. The "study" proving these criteria refers to bench tests conducted to verify the device meets all design specifications and is substantially equivalent to the predicate device.

    Here's a summary of the information that can be extracted based on the provided text, using the closest applicable interpretations:

    1. A table of acceptance criteria and the reported device performance

    Since this is not an AI/ML device, the "acceptance criteria" are compliance with relevant standards for medical devices and meeting design specifications. The "reported device performance" is a statement of compliance with these standards and equivalency to the predicate.

    Acceptance Criteria CategorySpecific Criteria (Relevant Standards)Reported Device Performance
    Material- ASTM F2026: 2017 (PEEK polymers for surgical implant applications) - ASTM F2848: 2021 (Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns and Surgical Implants)The subject device utilizes PEEK material, and bench tests demonstrated compliance with these standards.
    Mechanical Performance- ASTM F543: 2017 (Metallic medical bone screws)Bench tests demonstrated compliance with this standard.
    Sterilization- ISO 11135:2014 (Ethylene oxide sterilization methods) - ISO 11138-1:2017, ISO 11138-2:2017 (Biological indicators for sterilization) - ISO 11140-1:2014 (Chemical indicators)Bench tests demonstrated compliance with these standards, ensuring the device is sterile via EtO sterilization.
    Microbiological- ISO 11737-1:2018 [AMD1:2021] (Determination of microorganism population) - ISO 11737-2:2019 (Tests of sterility)Bench tests demonstrated compliance with these standards.
    Packaging & Shelf Life- ISO 11607-1:2019, ISO 11607-2:2019 (Packaging for terminally sterilized medical devices) - ASTM F1980:2016 (Accelerated Aging of Sterile Barrier Systems) - ASTM F88/F88M-21 (Seal strength) - ASTM F1929:2015 (Detecting Seal Leaks)Bench tests demonstrated compliance with these standards, confirming packaging integrity and a 3-year shelf life.
    Bacterial Endotoxin- USP <85> (Bacterial Endotoxin Test) - USP <161> (Medical Devices-Bacterial Endotoxin and Pyrogen Tests) - ISO 11737-3:2023 (Bacterial endotoxin testing)Bench tests demonstrated compliance with these standards.
    Substantial EquivalenceDevice must be substantially equivalent to the predicate device in terms of intended use, technological characteristics, and safety and effectiveness.The subject device (Fix2Lock PEEK Self Punching) is determined to be Substantially Equivalent (SE) to the predicate device (K231322) in respect of safety and effectiveness.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "bench tests" were conducted. It does not specify sample sizes for these tests, nor does it provide details on data provenance beyond the product's manufacturer (Osteonic Co., Ltd., South Korea) and the standards used. These are typical engineering tests, not clinical studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is not an AI/ML device involving expert "ground truth" for diagnostic or analytical tasks. The "ground truth" here is compliance with engineering and material standards, typically verified by laboratory testing and technical experts/engineers according to standard protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept applies to expert review processes, typically in clinical studies or AI ground truth establishment. The "adjudication" for this device would be the interpretation of bench test results against established standard specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is established by compliance with internationally recognized engineering and medical device standards (e.g., ISO, ASTM, USP) for materials, mechanical performance, sterilization, packaging, and biological safety.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1