510(k) Data Aggregation

    K Number
    K241703

    Validate with FDA (Live)

    Date Cleared
    2025-02-24

    (256 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SD Implant System is indicated for use in partially or fully edentulous mandibular and maxillary areas, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. SD Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions. The multi-unit abutments are designed for use in multi-unit restorations. They should be used in cases where multiple implants are placed to support a prosthetic dental restoration not suitable for single-unit restorations. All digitally designed abutments for use with Customized Abutment are intended to be manufactured at an ARUM DENTISTRY validated milling center. Smaller diameter implant bodies (i.e., 3.3mm and 3.7mm) are indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions.

    Device Description

    SD Implant System consist of below:
    Fixture

    • SD Bone Level Fixture
      Abutment
    • Cover Screw
    • Healing Abutment
    • Scan Healing Abutment
    • Scan Healing Abutment Screw
    • Cemented Abutment
    • Angled Abutment
    • Master Fix
    • Master Fix Screw
    • Digital Abutment
    • Temporary Abutment
    • Multi Abutment
    • Multi Angled Abutment
    • Abutment Screw
    • Multi Scan Healing Cap
    • Multi Scan Healing Cap Screw
    • Multi Master Fix
    • Multi Master Fix Screw
    • Multi Ti Cylinder
    • Multi Digital Cylinder
    • Multi Temporary Cylinder

    SD Implant System is consisted with Fixtures (SD Bone Level Fixture) and Abutments (Cover Screw, Healing Abutment, Scan Healing Abutment, Cemented Abutment, Master Fix, Digital Abutment, Temporary Abutment, Multi Abutment, Multi Angled Abutment, Multi Scan Healing Cap, Multi Master Fix, Multi Ti Cylinder, Multi Digital Cylinder, Multi Temporary Cylinder)

    This product is a dental implant which is placed into alveolar bone to replace the function of missing teeth. To enhance the osseointegration with the alveolar bone, this titanium dental implant is treated with SLA (Sandblasted with Large-grit and Acid-etching). As a dental implant which is placed into alveolar bone to support dental prostheses such as artificial teeth used to rehabilitate a patient's masticatory function.

    An endosseous dental implant is a device made of a material Pure titanium (Conforming to ASTM F67) which will be placed in the alveclar bone to replace the function of missing tooth. The SD Implant System consists of dental implants, abutments for use in one or two-stage dental implant placement and restorations. The implant-abutment connection is tight and precise fitting with internal hex and morse taper bevel. Smaller diameter implant bodies (i.e., 3.3mm and 3.7mm) are indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions.

    The subject device abutments are not compatible with 3.8mm diameter implant bodies cleared in K213506 and K230725, nor with 3.2mm diameter implant bodies cleared in K240091.

    SD Implant System Abutment is compatible with the SD Implant Fixture and Some of the SD Implant System connections are compatible with NB 1 Compatible with SA Implant System, NB Implant System and NB Mini Implant System. (Cleared K213506, K230725, K240091)

    SD Implant System Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures, Separate sets of abutments are presented for compatibility with each Fixture. Dental Abutments are similar to other commercially available products based on the intended use, technology used, claims, material composition employed and performance characteristics.

    The Multi Abutments and Multi Angled Abutments intended to be placed onto osseointegrated dental implants to provide support for prosthetic superstructures at the gingival level. Multi Abutments are available as straight abutments, which have an integrated thread and can be placed directly onto the implant. Multi Angled Abutments are angled abutments, which can be attached to the implant with the corresponding abutment screw. The multi-unit abutments are intended for multi-unit restoration only.

    The Cover Screw, Healing Abutment, Scan Healing Abutment, Scan Healing Abutment Screw, Multi Scan Healing Cap and Multi Scan Healing Cap Screw are provided sterile by the manufacturer. All other abutments are provided non-sterile and should be sterilized by the end user before use. The multi-unit abutments included in this system are specifically intended to be used solely for multiunit restorations. The SD Master Fix, multi Master Fix, Digital abutments, and multi Digital cylinder are composed of two-piece abutment that is a titanium base at the bottom and zirconia superstructure (CAD/CAM patient specific superstructure) at the top. The zirconia superstructure is straight only and is not to be designed to provide an angle or divergence correction. All digitally designed custom abutments for use with SD Master Fix, multi Master Fix, Digital abutments, and multi Digital cylinder are to be sent to ARUM DENTISTRY validated milling center for manufacture. All superstructures are to be manufactured from zirconia (cleared K190112). Digitally designed CAD/CAM abutments must have a 0.5 mm minimum gingival height dimension. SD Master Fix, multi Master Fix, Digital abutments, and multi Digital cylinderare intended for single-unit or multi-unit restorations.

    For the Master Fix, Multi Master Fix, Digital Abutment, Multi Digital Cylinder the design parameters for the CAD/CAM zirconia superstructure are:

    Minimum wall thickness - 0.5 mm; Minimum post height for single-unit restorations (length above the abutment collar / gingival height) – 4.5 mm; Maximum gingival height - 5.0 mm; Minimum gingival height - 1.5 mm; Angulation - 0° and All zirconia superstructures are for straight abutments only.

    The Master Fix, Multi Master Fix, Digital Abutment, Multi Digital Cylinder are used as part of a twopiece abutment, where the base is premanufactured from titanium alloy (Ti-6AI-4V Eli) and the top half is a CAD/CAM zirconia superstructure, milled at a validated milling center. These pieces are cemented together to form the final abutment.

    The SD Implant System is intended for single use only. The Abutments are made of Ti-6AI-4V Eli (Conforming to ASTM F136) The SD Implant System connections compatible with NB 1 SA Implant System, NB Implant System and NB Mini Implant System

    AI/ML Overview

    The provided text is a 510(k) summary for a dental implant system (SD Implant System) and, as such, focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and performance data in the format typically used for performance studies of AI/ML-enabled devices.

    Therefore, I cannot extract the acceptance criteria and study details as requested for an AI/ML device.

    The document details the device's components, materials, dimensions, indications for use, and a comparison to predicate devices, along with non-clinical performance data like biocompatibility, sterilization validation, shelf-life, and mechanical testing (ISO 14801). It does not contain information about:

    • Acceptance Criteria for AI/ML performance: The document is for a physical dental implant system, not a software/AI device.
    • Reported device performance for AI/ML: There's no AI component described.
    • Sample size for test set and data provenance: Not applicable to this type of device submission.
    • Number of experts and qualifications for ground truth: Not applicable.
    • Adjudication method for ground truth: Not applicable.
    • MRMC comparative effectiveness study: Not applicable.
    • Standalone (algorithm only) performance: Not applicable.
    • Type of ground truth used (pathology, outcomes data, etc.): Not applicable in the context of AI. Ground truth here refers to engineering specifications met by physical testing.
    • Sample size for training set/How ground truth for training set was established: Not applicable, as there's no AI/ML training involved for a physical implant.

    The document primarily focuses on demonstrating that the physical dental implant system meets established performance standards and is comparable to previously cleared devices through non-clinical testing.