Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243505

    Validate with FDA (Live)

    Date Cleared
    2025-05-30

    (199 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K162716, K121599, K231125

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 5008X Hemodialysis System is intended for use in acute and chronic hemodialysis therapy. Therapy options include hemodialysis (HD), hemodiafiltration (HDF), hemofiltration (HF), and isolated ultrafiltration (ISO).

    The 5008X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration.

    The 5008X™ Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine. The bibag system is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. The bibag system is intended to be used as one component in the preparation of dialysate according to a physician's prescription in a 3-stream proportioning hemodialysis machine equipped with the bibag module.

    The 5008X™ Machine is available equipped with a Crit-Line Clip (CLiC) Monitor System.

    The optional CLiC system is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

    The DIASAFE®plusUS is intended for the preparation of ultrapure dialysate and sterile, nonpyrogenic substitution fluid from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system).

    The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.

    The DIASAFE®plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE®plusUS.

    The blood tubing sets are intended to be used only with the Fresenius Medical Care 5008X Hemodialysis Device, which is indicated for intermittent hemodialysis treatment for patients with acute kidney injury or chronic kidney disease in a healthcare facility. Therapy options include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.

    The HDF blood tubing sets are indicated for use with a prescribed dialyzer. For use with a compatible dialyzer, as per the labeling.

    The Fresenius Twister Access Flow Reversing Connector (AFRC) is indicated for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices in order to obtain an access flow measurement.

    The blood tubing set is intended to be used with the Crit-Line Clip Monitor System (CLiC) on the 5008X Hemodialysis Device.

    The CLiC is used to non-invasively measure hematocrit, oxygen saturation, and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

    CitraSure is intended for thermal disinfection of the Fresenius Medical Care 5008X Hemodialysis System when delivered and diluted per the 5008X heat disinfection program.

    Device Description

    The 5008X Hemodialysis System consists of the following:

    • The 5008X Hemodialysis Machine
    • Two (2) DIASAFE®plusUS Filters (hereinafter referred to as the "Diasafe filter(s)")
    • The 5008X Bloodlines are available in five (5) configurations:
      • 5008X HD/HDF – Standard Blood Tubing Set (hereinafter referred to as the "HD/HDF Standard Bloodline")
      • 5008X HD/HDF – With Twister Blood Tubing Set (hereinafter referred to as the "HD/HDF Twister Bloodline")
      • 5008X HD/HDF – With CLiC Blood Tubing Set (hereinafter referred to as the "HD/HDF CLiC Bloodline")
      • 5008X HD Pre-Flush – Standard Blood Tubing Set (hereinafter referred to as the "HD Standard Bloodline")
      • 5008X HD Pre-Flush – With Twister Blood Tubing Set (hereinafter HD Twister Bloodline")
    • CitraSure™ Disinfectant

    The 5008X Hemodialysis Machine is an electromechanical device. Software controls the machine during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.

    The Diasafe filter is a non-sterile dialysis fluid filter that produces ultrapure dialysate and sterile, non-pyrogenic substitution fluid as defined in ISO 23500-5. The filter reduces microbial contaminants including endotoxins in the dialysate during hemodialysis treatment.

    The 5008X Bloodlines are single-use, ethylene oxide (EO) sterilized bloodlines.

    CitraSure™ disinfectant is a liquid chemical disinfectant for chemical heat disinfection of the 5008X Hemodialysis Machine.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the 5008X Hemodialysis System pertains to a medical device for renal treatment. The document does not describe an AI/ML device, a comparative effectiveness study (MRMC), or a standalone AI algorithm performance study. Therefore, sections of your request related to those topics (e.g., sample size for test/training sets, data provenance, number of experts, adjudication methods, ground truth types, human reader improvement with AI) are not applicable and cannot be extracted from this document.

    However, I can provide the acceptance criteria and a summary of the performance testing conducted to prove the device meets these criteria based on the provided text for the various components of the 5008X Hemodialysis System.

    Acceptance Criteria and Device Performance Summary for 5008X Hemodialysis System

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document lists "Key Performance Specifications/Characteristics" which serve as the acceptance criteria for each component. The "Reported Device Performance" column summarizes the claims made in the document regarding the device meeting these specifications through various tests.

    Feature/Test ObjectiveAcceptance Criteria (Specification/Characteristic)Reported Device Performance (Summary of Test Results)
    5008X Machine
    Maximum Blood Flow Rate600 mL/minSystem Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Maximum Dialysate Flow Rate1000 mL/minSystem Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Maximum Substitution Flow RateAutoSub plus: 400 mL/min; Manual substitution: 600 mL/minSystem Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Net Fluid Removal0–4000 mL/hr; Accuracy: ± (1%UF + 0.15% of balanced fluid volume)System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Dialysis TimeIntermittent, typical time 4 hoursSystem Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Dialysis Fluid CompositionVolumetric, selectable: Acid 125–151 mEq/L Na+; Bicarbonate 25–40 mEq/L; Monitoring conductivity average accuracy: ± 1.5%System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Dialysis Fluid TemperatureRange 34°C–39°C. Fixed alarm window 33°C (or 32°C during BTM recirculation measurement and preparation) or above 40°C.System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Heparin Administration Rate0.1 to 9.9 mL/hr; Accuracy: ± 0.5mL or ± 6%System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy.
    Electrical Safety and EMCIEC 60601-1-2 Edition 4.1 2020-09 complianceEMC testing conducted in accordance with IEC 60601-1-2.
    Software Verification & ValidationFunctional and Performance Verification, Regression Testing, Code ReviewsUnit, integration, and system level software verification testing performed to demonstrate efficacy and confirm operation.
    Human Factors ValidationSafe and effective use (per FDA guidance)Human Factors Validation Testing conducted.
    Diasafe Filter
    Bacterial and Endotoxin FiltrationProduces sterile, non-pyrogenic substitution fluid from dialysis fluid with max incoming water quality < 200 CFU/mL (bacteria) and < 2 EU/mL (endotoxin)Performance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Maximum Number Disinfection Cycles100 CitraSure cycles; 13 Pure Bright Bleach cyclesPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Use Life90 days maximum or if max disinfection cycles reachedPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    5008X Bloodlines
    Maximum Blood Flow Rate600 mL/minBlood Pump Endurance test performed. Results support substantial equivalence, safety, and efficacy.
    Minimum Labeled Arterial Pressure−300 mmHgStructural Integrity and Positive/Negative Leak Test performed. Results show the bloodlines can withstand 1.5X labeled max positive/negative pressures and pressures from faulty conditions.
    Maximum Labeled Venous Pressure+500 mmHgStructural Integrity and Positive/Negative Leak Test performed. Results show the bloodlines can withstand 1.5X labeled max positive/negative pressures and pressures from faulty conditions.
    Arterial Pump Segment8.0 mm/12.0 mm (ID/OD)Pump Segment Performance test evaluated characteristics over range of inlet pressures and flow rates.
    Twister ComponentFacilitates measurement of access flow by reversing blood flow while maintaining closed circuitTwister Rotation Test (180° ± 1°) and Twister Torque Test (≤ 7 inch-pound) performed.
    CLiC Chamber (Optional) – Hematocrit (HCT)Standard deviation X2 ≤ 3%, average bias ≤ 1% (of average HCT) on control blood chambers; measurement range 10–60 HCTFunctional CLiC Chamber Test (Hematocrit) performed.
    CLiC Chamber (Optional) – Oxygen Saturation (O2 Sat)Standard deviation X2 ≤ 3%, average bias ≤ 2% (of average O2 Sat) on control blood chambers; measurement range 30%–100% (when HCT ≥ 10)Functional CLiC Chamber Test (O2 Sat) performed.
    Structural Integrity (Bloodlines & Accessories)Withstand 1.5X max positive/negative pressures; Withstand pressures from faulty machine conditionsTesting performed.
    Pump Segment PerformanceEvaluate characteristics over range of inlet pressures (0 to -250 mmHg), flow rates, and treatment timeTesting performed.
    Blood Pump EnduranceNo break or detachment at max blood flow rates & pressures for 12 hoursTesting performed.
    Simulated UseNo tubing failures (kinking, collapsing, disconnection) under simulated use for ≥ 4 hours; components removable without external leaksTesting performed.
    Needle Access PortWithstand 1.5X recommended max positive/negative pressures after multiple accessesTesting performed.
    Luer Connectors (Performance)Meet ISO 80369-7 2021, Section 6Testing performed.
    Luer Connectors (Dimensions)Meet ISO 80369-7: 2021, Section 5Testing performed.
    Luer Connectors (Raw Material)Nominal modulus of elasticity > 700 MPaTesting performed.
    Transparency of Transducer ProtectorClear to allow visual inspection of blood contaminationTesting performed.
    Transducer Protector Leak TestMaintain secure and leak-free connection when subjected to 2X max labeled pressureTesting performed.
    Arterial Pressure Dome Leak TestMaintain secure and leak-free connection when subjected to 2X max labeled pressureTesting performed.
    CLiC Blood Chamber Lens Gap0.078 ± 0.005 in.Testing performed to verify optical property.
    Tubing Compliance TestCapable of being occlusively clamped by venous line clamp of 5008X machineTesting performed.
    Clamp Occlusion TestAir and liquid tightness occlusion at desired pressure and timeTesting performed.
    Check Valve Cracking PressureLower than 400 mbarTesting performed.
    Check Valve Pressure DropSpecific range when substitution fluid pumped at certain flow rateTesting performed.
    Tube TransparencyObserve interface of air and liquid during passage of air bubblesTesting performed.
    Tube Resistance to KinkingNo kinking at worst-case bend radius when packaged (evaluate % flow reduction)Testing performed.
    Tube ClampingResist clamp occlusions by subjecting to desired pressure and cyclesTesting performed.
    Tube PatencyMaintain patency after clamping and reopening (measure flow reduction)Testing performed.
    Air Bubble Trapping (Venous Chamber)Able to trap incoming air bubbles larger than 20 µLTesting performed.
    Recirculation Connector PackagingBurst strength, peel strength, dye penetration, microbial resistanceTesting performed.
    Viral Retentiveness (Transducer Protector)Prevent passage of bacteriophage (PhiX174) from patient to machine side up to 600 mmHg for 1 hrTesting performed.
    Sterility (Bloodlines)Sterility assurance level (SAL) 10-6Sterilized by 100% ethylene oxide (EO).
    EO Residuals (Bloodlines)< 4.6 mg/device for EO and EChTested per AAMI/ANSI/ISO 10993-7:2008/(R)2012.
    Bacterial Endotoxin (Pyrogenicity) (Bloodlines)Non-pyrogenic (< 20 EU/device)Tested per ANSI/AAMI/ST72:2019.
    Sterile Barrier Testing (Bloodlines)Maintain sterility of fluid pathStructural integrity test adapted from ISO 8637-2:2018 performed on samples after aging and distribution.
    CitraSure Disinfectant
    Minimum Disinfection Temperature70ºCPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Minimum Contact Time10 minutesPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Use Life90 DaysPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.
    Shelf Life2 yearsPerformance testing conducted. Results support substantial equivalence, safety, and efficacy.

    2. Sample Size for the Test Set and the Data Provenance

    As this is not an AI/ML device, the concept of "test set" in the context of diagnostic AI algorithms (e.g., patient images or clinical data) is not directly applicable. The performance data presented refers to engineering and quality assurance tests on the physical devices/systems.

    • Sample Size for Physical Device Testing: The document does not specify exact sample sizes for each individual engineering performance test (e.g., how many bloodlines were tested for structural integrity, how many machines for fluid flow accuracy). It generally states "Testing performed" or "Testing conducted."
    • Data Provenance: Not applicable in the context of patient data provenance (e.g., country of origin, retrospective/prospective). These are laboratory and simulated use tests on hardware components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of those Experts

    Not applicable. The "ground truth" for the performance of a hemodialysis system is based on direct physical measurements against defined engineering specifications and regulatory standards (e.g., flow rates, pressure tolerances, material properties, sterility levels), not expert interpretation of complex data like medical images.

    4. Adjudication Method for the Test Set

    Not applicable for a medical device's engineering performance testing. Adjudication methods like 2+1 or 3+1 typically refer to resolving discrepancies in expert interpretations (e.g., radiologists reviewing medical images for ground truth labeling in AI studies).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. The document explicitly states: "No clinical studies were performed for the 5008X Hemodialysis System." MRMC studies inherently involve human readers (clinicians) assessing cases, often with or without AI assistance, to measure diagnostic performance and comparative effectiveness. This type of study was not conducted as this is not an AI-assisted diagnostic device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. The device described is a physical hemodialysis system, not a standalone AI algorithm. While it contains software, the performance evaluation is for the integrated system, and there is no mention of a standalone AI algorithm being assessed.

    7. The Type of Ground Truth Used

    For the various performance tests, the "ground truth" is defined by:

    • Engineering Specifications and Standards: Numeric values for flow rates, pressures, temperatures, accuracies, dimensions, and material properties derived from established engineering principles and relevant international standards (e.g., ISO 8637-2:2018 for bloodlines, ISO 23500-5 for dialysate quality, IEC 60601-1-2 for EMC).
    • Regulatory Requirements: Compliance with FDA guidance documents and regulations (e.g., for biocompatibility, sterility, software verification and validation, human factors).
    • Absence of Failure: Demonstrating that components do not break, kink, leak, or fail under specified conditions (e.g., blood pump endurance, simulated use).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set. The software for the 5008X Hemodialysis Machine would have undergone development, verification, and validation, but this is distinct from AI model training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI training set, there is no ground truth established in this context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1